Lucy® 3D QA Phantom Datasheet pdf Format (1

advertisement
Phantoms
•
END-TO-END
STEREOTACTIC QA
Perform image transfer
QA, dosimetry and
machine QA within the
exact coordinate system
of commercially available
stereotactic head frames
and frameless mask systems
•PHYSICS TOOLS AT YOUR FINGERTIPS
The Lucy 3D QA Phantom minimizes clinically relevant
errors by quantifying system variances in the imaging
chain to within 0.1mm. Increase your confidence in your
system to correctly image known volumes, distances and
geometries. Confirm that CT/MRI/Angioscans and image
fusion are accurately recreated within the TPS, assuring
precise target dose delivery
•SRS PATIENT DOSIMETRY QA
Obtain absolute, relative and point dose dosimetry QA
measurements at isocenter and at exact positions off
Elekta Fraxion™
Stereotactic Frame Interface
for Fraxion systems
isocenter, using your current ion chambers, film, diodes,
gels, MOSFETs and TLDs
•RADIATION ALIGNMENT MACHINE QA
Assess optical and geometric isocenter, laser alignment,
BrainLAB
Stereotactic Frame Interface
and non-invasive Mask Systems
indexed table positioning alignment and positioning
coordinates for a comprehensive evaluation of geometric
accuracy. Precisely determine CBCT and MV/kV
isocenter alignment.
•LOCKS INTO SRS FRAMES
Leksell Gamma Knife®
Stereotactic Frame Interface
Quickly and precisely position the Lucy 3D QA Phantom
within the manufacturer’s exact coordinate system
using the Stereotactic Frame Interface, available for
all commercial SRS systems. Works for invasive and
non-invasive systems.
Varian / Radionics™ / CRW/BRW
Stereotactic Frame Interface
Phantoms
Features
• Make QA setup easy by locking the Lucy 3D QA Phantom
to your exact SRS coordinate system
• Effectively evaluates six degrees of freedom positioning
systems in 3D space with convenient image markers
• Commission your system quickly by using the Lucy 3D QA Phantom to confirm specification upon delivery of equipment
• Evaluate manufacturer’s automatic patient positioning
systems with film, ion chambers, diodes, gels, MOSFETs
and TLDs
• Use as a QA tool as described in the performance validation
references 1,2,3,4,5 and in AAPM Task Group reports
including TG-142
• Check cone beam CT performance in 3D space
• Ideally suited for QC and dosimetry measurements with
framed and frameless SRS systems
• Evaluate CT/MRI/Angiofusion at 20 points in 3D space
Configurations
Lucy 3D QA Phantom
Lucy MRI Imaging QA Accessory Package
• Spherical Lucite Phantom
• MRI Marker Cylinders, set of four
• Precision Leveling and Rotational Alignment Base
• MRI Volume Insert with three known geometries
Stereotactic Frame Interfaces
• MRI Signal Generator
• Elekta Fraxion
Radiation Alignment Pointers
• BrainLAB
• Includes 5 mm tungsten sphere tip and 3 mm aluminum
sphere tip
• Leksell Gamma Knife
• Varian and Radionics CRW/BRW frame
• Tomotherapy Hi-Art System
• Precision Leveling and Rotational Alignment Base for
frameless systems and extracranial work
• Automate Winston-Lutz test with the Lucy 3D QA Phantom
and PIPSpro Software using a filmless procedure
Accessories
• IGRT kV and X-Ray Angiography Marker Cylinders, set of four
Lucy Dosimetry QA Accessory Package
• Dosimetry Film Cassette for fourteen 2.5 x 2.5 inch films
• Dosimetry Insert for small volume ion chamber of your choice
• Other accessories are available for many ion chambers, TLD dosimetry, diode detectors, MOSFET detectors and
MRI compatible bases for phantom positioning
• Dosimetry Film Cassette for three 3 x 3 inch films
• Target/Treatment Verification Film Cassette
Dosimetry Instruments
Lucy CT Imaging QA Accessory Package
• MAX 4000 Electrometer
• CT Marker Cylinders, set of four
• Exradin A16 MicroChamber
• CT Electron Density Insert
• Exradin A14SL MicroChamber
• CT Volume Insert with three known geometries
• CT Grid Insert for spatial distortion
Leksell Gamma Knife® is a registered trademark of Elekta Instruments, Inc. CyberKnife® is a registered trademark of Accuray Incorporated.
TomoTherapy Hi-Art System® is a trademark of TomoTherapy, Inc. Specifications subject to change without notice.
performance validation
1. R. Ramasesham, M. Heydarian, “Comprehensive quality assurance for stereotactic radiosurgery treatments”, Phys. Med. Biol. 48(2003) N119-N205.
2. R. Ramani, et.al., “A QA phantom for dynamic stereotactic radiosurgery: Quantitative measurements”, Med. Phys. 22(8), August 1995.
3. R. Ramani, et.al., “The use of Radiochromic film in treatment verification of dynamic stereotactic radiosurgery”, Med. Phys. 21(3), 1994.
4. AAPM Radiation Therapy Committee TG53: “Quality assurance for clinical radiotherapy treatment planning”, Med. Phys. 25(10), October 1998.
5. AAPM Radiation Therapy Committee TG66: “Quality assurance for computed-tomography simulators and the computed-tomography-simulation
process”, Med. Phys. 30(10), October 2003
3120 Deming Way Middleton WI 53562-1461 USA
800-261-4446 . ph 608-831-0025 . fax 608-831-2202
www.standardimaging.com
© 2010 Standard Imaging, Inc. 1284-21, 9/10