Scleral Lens Case Studies: Keratoconus

advertisement

contamac.com

Scleral Lens Case Studies: Keratoconus

Case Study A: Patient: LN Female Age:32

History: This 32 year old Caucasian female was diagnosed in 2010 with keratoconus in both eyes. The left eye shows the keratoconus to be more advanced. She has tried numerous RGP lenses over the past few years with little success. She wore soft lenses in 2010 but her vision was poor.

Topography

Plan

Right Eye: SSF 13

RADIUS DIAMETER

8.23

8.43

8.70

12.80

14.40

SAG

Rx: -4.75/-1.25 X 180°

8.60

12.00

13.80

14.80

15.60

4.1339mm

C.T.: 0.40

Left Eye: SSF 7

RADIUS DIAMETER

7.18

7.38

8.70

12.80

14.40

SAG

Rx: -12.50/-1.50 X 180°

8.60

12.00

13.80

14.80

15.60

4.7173mm

C.T.: 0.40

Fitting regular 9.8 diameter hard lenses seemed to be very unstable especially on the LEFT eye so the decision was made to go the scleral lens route.

The above lenses were fitted.

Result

Good clearance but with some bearing on the limbus (o.u). This bearing also had an effect on corneal function as the patient presented limbal hyperaemia after 2 hours of wear.

The sagittal depths of the above lenses showed good corneal clearance but showed inadequate limbal clearance under both lenses due to the landing zone coming in too flat over the limbus.

The Zeiss Visante OCT showed good clearance but the bearing of the lens on the limbal zone is clearly visible.

Making the 13.8 & 14.8mm diameter curves steeper and blending the 2 zones over this area would create less touch.

Another option is to increase the diameter of these two zones in order to move the clearance over the limbus.

In the two OCT images above, we can see the sharp transition area between the 13.8 & 14.8 diameter curves. This could also influence a good fit and therefore the introduction of an aspheric landing zone was needed. We did the above adjustments to the lenses manually but still the hyperaemia persisted.

Larger lenses were then fitted to both eyes with the following parameters:

Right Eye: SWH 11

RADIUS DIAMETER

8.40

7.6556

8.45

13.25 x 14.25

14.55

Rx: -4.25

8.90

10.80

13.80

16.50

17.50

C.T.: 0.38

Left Eye: SWH 7 (mod)

RADIUS DIAMETER

7.30

7.212

8.40

13.25 x 14.25

14.55

Rx: -11.50/-1.50 X 168°

8.90

11.10

13.80

16.50

17.50

C.T.: 0.38

The larger 17.5mm diameter lenses are an improvement to the previous lenses and sit well over the limbus with no evidence of hyperaemia.

As seen in the 16.5 diameter zone we decided on a toric curve for better fit on the sclera. “Observation of most scleral patients so far show more blanching in the horizontal area over the sclera than in the vertical and a toric curve makes much more sense to insure a good aligned fit with the sclera in 360 ĚŠ. Studies have shown that the sclera is flatter nasally and the temporal sclera steeper than the inferior sclera.” – Van der Worp – 2009

The above OCT images show a much more acceptable landing zone and good clearance. The larger overall diameter lenses are not giving any hyperaemia in the limbal area even after a few hours of wear.

Slit Lamp Observations

FIGURE1: Shows an impression ring after lens removal. It is of no consequence as there is no scleral blanching when the lens is in situ. Limbal staining is slightly visible horizontally. As explained before, to alleviate staining, increase limbal clearance through steepening the landing zone in the 16.5 diameter and blending the transition between the 13.8 and 16.5 diameter zone.

Limbal pressure can also be alleviated by increasing the optic zone slightly.

FIGURE 2: Right scleral lens showing good clearance over the centre of the cornea.

FIGURE 3: Left Scleral lens showing good clearance over the centre of the left cornea with a thinning in the clearance visible at the position of the cone.

FIGURE 4: Right scleral lens still showing a small amount of clearance over the limbus. We need to aim for clearance of about 100 microns over the limbus.

Observing the scleral lens in different gaze positions

FIGURE 5: Left scleral lens showing clearance over the limbal area.

FIGURE 6: Right scleral lens showing no blanching in down gaze.

FIGURE 7 : Left scleral lens showing no blanching on down gaze.

FIGURE 8: Right scleral lens showing a small amount of blanching on the landing zone. If the bigger veins are observed, there is no blanching.

This part of the lens needs to be made steeper.

FIGURE 9: Left scleral lens showing no blanching on left gaze.

The patient wears the lenses for up to 12 hours per day very comfortably.

Our patient is very happy with the comfort and presents with a very satisfactory V.A of the following:

Right: 6/7.5 +2

Left: 6/7.5 -2

Case Study B: Patient: KB Female Age:21

History . This patient is a black female of 21 years old. She presented with keratoconus at the age of 10 with hydrops, leading to scaring on the left cornea. There was only light perception in her left eye and a corneal transplant was done in 2004.The best vision with a spectacle correction is:

Right: 6/12

Left: 6/9

Topography

OD: OS:

Examination

Conventional hard lenses did not stay on the left eye and were forced out on the blink. The corneal graft on the left eye also posed a challenge to obtain a good RGP fit. A regular post-surgery RGP was an option for the left eye but since patient comfort was of high priority, a scleral lens was fitted.

Plan

The first lenses fitted had the following parameters:

Right Eye: SSF 11

RADIUS DIAMETER

7.84

8.04

8.70

12.85

14.50

SAG

Rx: -0.25

8.60

12.00

13.80

14.80

15.60

4.3125mm

C.T.: 0.40

Left Eye: SSF 6.5

RADIUS DIAMETER

7.10

7.30

8.70

12.85

14.50

SAG

Rx: -5.00

8.60

12.00

13.80

14.80

15.60

4.7173mm

C.T.: 0.40

Result

Figures 1 and 2 (following page) both show good clearance. 200-300 microns is adequate clearance even though the

Visante OCT shows a bit less than 200 microns.

Vision was very good with these lenses: RIGHT: 6/7.5+1

FIGURE1 FIGURE 2

The patient complained of discomfort in the left eye after wearing the lenses for a while. On examination I found that the inferior edge of the lens was pressing the bottom eyelid inward causing the discomfort. This was not apparent in the right eye.

A decision was made to fit a larger lens in the left eye to overcome the discomfort.

A SWH 17.5mm diameter lens was fitted on the left eye with the following parameters:

Left Eye: SWH 4

RADIUS DIAMETER

7.30

6.6586

8.45

13.25

14.50

SAG @ 13.80

Rx: -4.75

8.60

11.10

13.80

16.50

17.50

4.300mm

C.T.: 0.38

FIGURE 3

The Visante OCT shows better clearance over the central cornea in Figure 3 and the patient finds more comfort with the bigger lens diameter. There is no longer bottom lid irritation as there is now less lid influence on the larger lens.

Slit Lamp Observations

Right Eye

Image 1: Showing good clearance over the cornea but with some touch on the limbus superior nasally.

Image 2: Central corneal clearance can be seen using the thin slit

(parallel-piped) on your slit-lamp without a blue light. The clearance is about 200microns in the centre over the cornea and still sufficient for optimal corneal health. It is easy to see how the clearance increases towards the limbus.

Image 3: Now showing good clearance of the limbus.

Left Eye

Image 4: Still showing a decrease in clearance superior nasally.

Image 5: The borders of the corneal graft are clearly visible without the lens.

Image 6: Good clearance over the centre of the cornea.

Image 7: Showing good clearance over most of the cornea except for the superior limbus. This touch over the limbus could be due to upper lid pressure on the lens and is only visible in the straight-ahead gaze.

Image 8 : On downward gaze the pressure disappears and therefore it is so important to examine the clearance in all directions of gaze.

Image 9: Even in the upward gaze there is sufficient clearance over the limbus.

Image 10: The decrease in clearance over the superior limbus in straight-ahead gaze is once again visible using the parallel-piped setting.

The patient presented very good vision and she is very happy with the comfort:

R: 6/6

L: 6/6-2

As seen from the Visante OCT scans, the sag can be increased more to insure even more clearance over the cornea.

An extra pair of lenses were made for the patient where the sagittal depth was increased by 91.2 microns OD &

100microns OS. This was to insure that the lenses maintain a clearance over the cornea and the graft of the left eye even after a few hours of wear.

Final Conclusion

Scleral lenses can be a very successful option for a variety of patients. These lenses could help post-corneal graft patients, keratoconics, irregular corneas, post-traumatic corneas and even children with very high refractive errors. Comfort is really one of the most spectacular benefits of scleral lenses and many keratoconus sufferers would prefer these lenses if they had a choice after trying normal RGPs versus scleral lenses.

A study has also found that 69 per cent of patients who were referred for keratoplasty could be successfully fitted with contact lenses without surgery (Smiddy et al 1988) and if your patient could choose which lens to do this with, I can guarantee that the scleral lenses, manufactured by the Contact Lens Laboratory of South Africa, would be the patient’s as well as the optometrist’s lens of choice.

OPTOMETRIST:

Lizelle Dorfling-Smith

B.Optom (RAU)

CAS (Boston) Advanced Contact Lenses

Email: lizelledsmith@gmail.com

Download