Steri-Cult R Automated CO2 Incubator

Steri-Cult® R Automated CO2 Incubator
Features and Technical Specifications
Specifications
Automation
Plate Storage. . .
Capacity . . . . . .
Access Time . . .
Robot Interface .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
7 x 28 position hotels
196 shallow-well microplates
20 seconds maximum
RS-232, V24 9600 baud
Temperature
Control . . . . . . . . . . . . . . . . . ±0.1C @ 37C (98.6F)
Range . . . . . . . . . . . . . . . . . . 10C above ambient to
50C (122F)
Sensor. . . . . . . . . . . . . . . . . . Thermistor
Controller . . . . . . . . . . . . . . . Microprocessor
Setpoint . . . . . . . . . . . . . . . . Digital
Display . . . . . . . . . . . . . . . . . Digital LED
Readability and Setability . . . . 0.1C
Uniformity . . . . . . . . . . . . . . . ±0.5C
Temperature Safety
Sensor. . . . . . . . . . . . . . . . . . Thermistor
Controller . . . . . . . . . . . . . . . Microprocessor
Setability . . . . . . . . . . . . . . . . 0.1C
CO2
Control . . . . . . . . . . . . . .
Range . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . .
Inlet Pressure . . . . . . . . .
Filter . . . . . . . . . . . . . . . .
Sensor. . . . . . . . . . . . . . .
Controller . . . . . . . . . . . .
Display . . . . . . . . . . . . . .
Readability and Setability .
Tracking Alarm. . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
±0.1% @ 5.0%
0-20%
Automatic
15 PSIG (1.0 bar)
0.2 micron, disposable
Dual-beam IR
Microprocessor
Digital LED
0.1%
User programmable high/low
Controlled RH
RH . . . . . . . . . . . . . . .
Control . . . . . . . . . . . .
Sensor. . . . . . . . . . . . .
Controller . . . . . . . . . .
Readability & Setability
Humidity Reservoir . . .
Tracking Alarm. . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Ambient to 95% RH
±2.0%
Capacitive
Microprocessor
1%
1.0 gal. (3.8 liters)
User programmable high/low
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Unit Heat Load
3387 . . . . . . . . . . . . . . . . . . . 595 BTUH (175 Watts)
3388 . . . . . . . . . . . . . . . . . . . 647 BTUH (190 Watts)
Dimensions
Exterior . . . . . . . . . . . . . . . . . 35.0"W x 39.4"H x 31.7"F-B
(88.9cm x 100.1cm x 80.5cm)
Interior . . . . . . . . . . . . . . . . . 20.8"W x 32.8"H x 24.6"F-B
(52.8cm x 83.3cm x 62.5cm)
Transfer Height . . . . . . . . . . . 8.1" (20.6cm)
Construction
Interior Volume . . . . . . . . . . . 9.7 cu. ft. (274.7 liters)
Interior . . . . . . . . . . . . . . . . . Type 304 mirror finish,
stainless steel
Exterior . . . . . . . . . . . . . . . . . 18 gauge, cold-rolled steel,
powder coated
Inner Door . . . . . . . . . . . . . . . 1/4" (0.6cm) fully tempered
safety glass with cam action
latch
Outer Door Gasket . . . . . . . . Four-sided, molded, magnetic
vinyl
Inner Door Gasket . . . . . . . . . Feather, silicone
People of Science Working
with People of Science
A Leader in Controlled Environment
Equipment
One Convenient Source, Total Solutions
Whether you are storing, preparing, culturing, or analyzing
samples, Thermo has the research tools you depend on to
succeed in the lab. We have been providing high quality,
reliable controlled environment laboratory products for
a wide variety of clinical, educational, biotechnological,
pharmaceutical, and industrial applications for more
than 50 years.
We offer the following:
Thermo’s expertise extends beyond controlled
environment equipment. In fact, you may be surprised
by the breadth of products and services we offer.
Our life and laboratory sciences and services, optical
technologies, measurement and control, turnkey
operations, and even capital equipment financing
specialists are eager to share their knowledge and work
as your partner...for life.
The benefit? We’re people of science working with people
of science. And you will quickly discover the value of one
convenient source providing total solutions tailored to your
specific needs.
• Forma Cell Culture Incubators
• Forma In Vitro Fertilization (IVF) Incubators and
CryoMed® Controlled Rate Freezers
Electrical
3387 . . . . . . . . . . . . . . . . . . . 115V, 50/60 Hz, 5.2 FLA
(Operating range 90-125V)
3388 . . . . . . . . . . . . . . . . . . . 230V, 50/60 Hz, 3.6 FLA
(Operating range 180-250V)
Power Switch . . . . . . . . . . . . 2 Pole
Plug. . . . . . . . . . . . . . . . . . . . NEMA 5-15P (3387),
CEE 7/7 (3388)
Convenience Receptacle . . . . 115V standard/220V export
(75 Watts maximum)
Alarm Contacts . . . . . . . . . . . Deviation of temperature,
CO2, RH, and power failure.
Accessed using RJ11 jack on
rear of unit.
• Forma Environmental Chambers
Estimated Weight
Net . . . . . . . . . . . . . . . . . . . . 480 lbs. (217.7 kg)
Shipping (Motor) . . . . . . . . . . 550 lbs. (249.5 kg)
• Forma Lab Refrigerators
Controlled Environment
• Forma Ultra-Low Temperature Freezers
Equipment Solutions
• Forma Laminar Airflow Equipment
• Forma Biological Safety Cabinets
for the Lab
• Forma Cryopreservation Equipment
• Forma Blood Bank Equipment
• Forma Circulating Baths
• Forma Monitor/Alarm System
• Forma Orbital Shakers
• Forma Programmable Centrifuges
• Forma Lab Washers/Dryers
• Forma Lab Freezers
• Forma Chromatography Refrigerators
Fittings
Drain Port . . . . . . . . . . . . . . . 3/8" barbed fitting with shutoff
Access Port. . . . . . . . . . . . . . 1.4" (3.6cm) with removable
silicone plug
CO2 Inlet . . . . . . . . . . . . . . . . 1/4" hose (barbed)
These specifications are for informational purposes only and are subject to change without notice. Thermo makes no warranties, expressed or implied,
in this product summary.
5
TRANSFER STATION
DIMENSIONS
Measurements displayed are in inches.
• Custom-Designed Equipment
Steri-Cult® R Automated CO2 Incubator
• Laboratory Services
About Thermo
Thermo provides scientific instruments, equipment,
software, and solutions that help laboratory and industrial
customers advance scientific knowledge, enable drug
discovery, improve manufacturing processes, and protect
people and the environment.
With annual revenue exceeding $2 billion, Thermo is a
global leader employing 11,000 people in 30 countries
worldwide. The company is based in Waltham,
Massachusetts.
© 2003 Thermo Electron Corporation. All rights reserved. Thermo Electron Corporation, question everything, and Analyze. Detect. Measure. Control are trademarks of Thermo Electron Corporation. Steri-Cult, Forma,
HEPA2, and CryoMed are trademarks or registered trademarks of Thermo Forma. CRS CataLyst-5, CRS CataLyst Express, CRS F3, CRS Dimension4, and CRS POLARA are trademarks of Thermo CRS.
Rev. 0, 07/03 5M
Controlled Environment Equipment
P.O. Box 649 • Marietta, OH 45750
740-373-4763 • Fax: 740-373-6770
Toll Free USA and Canada 866-984-3766
www.thermo.com/forma
Analyze • Detect • Measure • Control™
Airborne and waterborne contamination can affect
product yields and reliability. Some cell-based assays
can take several weeks to culture. You can’t afford
to lose time and productivity if they are destroyed
by contaminants. Steri-Cult R is the only automated
incubator on the market with airborne and waterborne
contamination control.
compatible, automated storage and plate handling system and our well known Steri-Cult CO2
Incubator—the name that brought leading edge technology, such as infrared CO2 sensors,
microprocessor control, directed airflow, and HEPA filtration, to cell culture incubators. Steri-Cult has
been the leading name in high volume, contamination-resistant tissue culture for more than 15 years.
This is a key advantage in assay development, where contamination can bring the whole process to a
The Steri-Cult R’s design integrates three methods
of contamination control: an external humidification
system, HEPA filtration, and careful chamber design.
Contamination-Resistant External Humidification System
stop, costing you time and money.
By uniting our cell culture incubator engineering and manufacturing expertise, highly regarded
laboratory automation systems experience, and the talent of a leading automation designer, Thermo
Electron Corporation developed a specialized drug discovery product that reflects years of solid
research and rigorous testing. The Steri-Cult R automated system is ideal for your demanding
ADME/Tox assays, such as Caco2, as well as other cell-based assays and general automation-
The humidification system is located outside the
incubator chamber, reducing the risk of waterborne
contamination that can cause product loss, incubator
downtime, and wasted time and money.
There’s no humidity pan to check or handle, no relying
on a pan’s mechanical switch to warn you that the
water is low. And no need to disassemble/reassemble
automation components (with the hope that
everything will realign properly) to clean an interior
source of humidification.
Best in Class: Fastest Speed and High Capacity
To protect your
cultures, our
unique HEPA
Filter Airflow
System, located
inside the
chamber,
employs
laminar airflow
techniques and handles primary sources of
contamination (biological contaminants and airborne
particulates) found in most laboratory settings.
Steri-Cult R
provides storage
for standard
shallow-well
microplates or
deepwell
microplates.
The 196 microplate
capacity makes
it a cost-effective
system for your
high throughput
needs by
incubating
more plates,
speeding up the process, and lowering the price per
plate processed.
• Gentle, directed airflow ensures continuous
cleaning during incubator operation to keep
airborne particulates from contaminating and
ruining cultures, costing time and money.
• Microbiological filters protect the gas inlet and
sample port.
Three transfer station configurations allow you to
tailor the equipment to your specific needs: standard
stationary (fixed), optional swapping (swaps positions
of two plates), and optional plate reorientation
(rotates plate from landscape to portrait).
• Entire chamber volume of air is filtered every 60
seconds to prevent contamination. And Class 100
air quality is achieved within five minutes of door
closing. (Testing protocol available.)
Straightforward Integration: Low Risk, High Value
We removed the barriers. It’s easy for you to take
advantage of the Steri-Cult R’s faster access time and
high capacity.
100000
Market-Leading Contamination Control to protect your cellbased assays and cultures, reducing the risk of a costly, time consuming
break in the development process
Best in Class Speed and Capacity for fastest processing and
low cost per plate processed
An Integrator’s Dream due to careful incubator design and
2
adherence to industry hardware and software standards
IDEAL CULTURING AND
HIGH THROUGHPUT
Steri-Cult has long been known for
optimum temperature uniformity and
stability. It is a favorite in biopharm
research labs and high volume clinical
and research facilities. Now we’ve
combined that trusted incubator with a
high capacity, automated, handling
carousel and plate mover that offer
superior speed; industry-standard
software; and a convenient plate
presentation location.
• At-a-glance water level monitoring eliminates the
need to open the incubator’s front doors, reducing
the risk of contamination. Blue backlighting flashes
when water is low, a subtle reminder to check the
humidity supply.
• Easy access
to water bottle
and snap-open
humidification
chamber
ensures ease
of maintenance
and less
disturbance to
your assays.
10000
Air Classification
Particles >= 0.5 micron
compatible incubation applications. Take full advantage of all that the new Steri-Cult R offers.
• Built at our ISO 9001-2000 registered facility, the
standardized incubator is CE Marked and CSA Certified
to meet worldwide requirements, attesting to the
incubator's electrical and mechanical safety and
compliance with international requirements related to
emission of, and susceptibility to, RF energy and
electrostatic discharge.
1000
100
10
1
0
5
10
15
20
25
Particulate count reduction in a Steri-Cult R
CO2 Incubator with Class 100 HEPA filtration
• Optional HEPA2™ VOC filter removes volatile organic
chemicals (vapors) as well as airborne contaminants
and particulates for added product protection.
Easy-to-Clean Chamber Design
• Humidification is 100% HEPA filtered before
entering the incubator chamber to minimize the
risk of contamination.
• Chamber’s mirror finish, stainless steel interior
with 100% coved corners is easy to clean.
• HEPA filter is located inside the incubator chamber
for optimum filtering, easy access, and simple
replacement without the use of tools.
• User-friendly alphanumeric command set is available
so integrators can program plate storage and retrieval
when the Steri-Cult R is used with a variety of
automation systems.
• Transfer station options ensure greater integration
flexibility.
Accessories
Stock No.
303000
3
4
– contribute to the Steri-Cult R’s full compatibility
with Thermo’s CRS CataLyst-5™, CRS CataLyst
Express™, CRS F3™, and CRS Dimension4™
robots and easy integration into CRS POLARA™
laboratory systems
Description
Microplate hotel, 28 places up to
2.3cm each
Deepwell microplate hotel, 12 places up
to 5.0cm each
Stationary (fixed) transfer station
Plate reorientation transfer station
Swapping transfer station
HEPA filter replacement kit,* HEPA and
in-line filters
HEPA2 VOC filter replacement kit,* HEPA
VOC filter and in-line filters
Built-in CO2 gas guard, monitors
supply and switches to another cylinder,
factory installed
303001
303002
303003
303004
1900160
1900161
1900153
*Customer installed filters are easily replaced without tools.
HEPA filter is rated a minimum 99.97% efficient at 0.3 microns.
90.0
• We made systems compatibility straightforward by
implementing industry standards for the software interface;
plate presentation hardware; and RS-232, V24 9600 baud
robot control interface.
Time (minutes) after Door Closing
• Flush-mount access gate and gate cover
• Filter built in
to water fill
bottle’s lid allows clean air to enter the bottle
while keeping out contaminants.
– permit easy access to the plate (portrait orientation)
and ensure that nothing hinders external robot
contact with the transfer station
88.8
RH
The new Steri-Cult R Automated CO2 Incubator is a winning combination of a sophisticated, robot-
Continuous Air Filtering to Achieve Class 100 Air Quality
Temperature
Continuous Contamination Control: Protecting
Your Valuable Assays
37.0
Temp
36.8
RH
1.0 1.2
Time (minutes)
Small size of the automatic access gate and short 10 second
cycle minimize disruption of the chamber environment.
5.0
4.8
CO2
Steri-Cult® R Automated CO2 Incubator
Features and Technical Specifications
A Winning Combination for Your Drug Discovery Needs
1.0 1.2
Time (minutes)
Advanced dual-beam IR control system ensures
highly stable CO2, even during gate openings.
Airborne and waterborne contamination can affect
product yields and reliability. Some cell-based assays
can take several weeks to culture. You can’t afford
to lose time and productivity if they are destroyed
by contaminants. Steri-Cult R is the only automated
incubator on the market with airborne and waterborne
contamination control.
compatible, automated storage and plate handling system and our well known Steri-Cult CO2
Incubator—the name that brought leading edge technology, such as infrared CO2 sensors,
microprocessor control, directed airflow, and HEPA filtration, to cell culture incubators. Steri-Cult has
been the leading name in high volume, contamination-resistant tissue culture for more than 15 years.
This is a key advantage in assay development, where contamination can bring the whole process to a
The Steri-Cult R’s design integrates three methods
of contamination control: an external humidification
system, HEPA filtration, and careful chamber design.
Contamination-Resistant External Humidification System
stop, costing you time and money.
By uniting our cell culture incubator engineering and manufacturing expertise, highly regarded
laboratory automation systems experience, and the talent of a leading automation designer, Thermo
Electron Corporation developed a specialized drug discovery product that reflects years of solid
research and rigorous testing. The Steri-Cult R automated system is ideal for your demanding
ADME/Tox assays, such as Caco2, as well as other cell-based assays and general automation-
The humidification system is located outside the
incubator chamber, reducing the risk of waterborne
contamination that can cause product loss, incubator
downtime, and wasted time and money.
There’s no humidity pan to check or handle, no relying
on a pan’s mechanical switch to warn you that the
water is low. And no need to disassemble/reassemble
automation components (with the hope that
everything will realign properly) to clean an interior
source of humidification.
Best in Class: Fastest Speed and High Capacity
To protect your
cultures, our
unique HEPA
Filter Airflow
System, located
inside the
chamber,
employs
laminar airflow
techniques and handles primary sources of
contamination (biological contaminants and airborne
particulates) found in most laboratory settings.
Steri-Cult R
provides storage
for standard
shallow-well
microplates or
deepwell
microplates.
The 196 microplate
capacity makes
it a cost-effective
system for your
high throughput
needs by
incubating
more plates,
speeding up the process, and lowering the price per
plate processed.
• Gentle, directed airflow ensures continuous
cleaning during incubator operation to keep
airborne particulates from contaminating and
ruining cultures, costing time and money.
• Microbiological filters protect the gas inlet and
sample port.
Three transfer station configurations allow you to
tailor the equipment to your specific needs: standard
stationary (fixed), optional swapping (swaps positions
of two plates), and optional plate reorientation
(rotates plate from landscape to portrait).
• Entire chamber volume of air is filtered every 60
seconds to prevent contamination. And Class 100
air quality is achieved within five minutes of door
closing. (Testing protocol available.)
Straightforward Integration: Low Risk, High Value
We removed the barriers. It’s easy for you to take
advantage of the Steri-Cult R’s faster access time and
high capacity.
100000
Market-Leading Contamination Control to protect your cellbased assays and cultures, reducing the risk of a costly, time consuming
break in the development process
Best in Class Speed and Capacity for fastest processing and
low cost per plate processed
An Integrator’s Dream due to careful incubator design and
2
adherence to industry hardware and software standards
IDEAL CULTURING AND
HIGH THROUGHPUT
Steri-Cult has long been known for
optimum temperature uniformity and
stability. It is a favorite in biopharm
research labs and high volume clinical
and research facilities. Now we’ve
combined that trusted incubator with a
high capacity, automated, handling
carousel and plate mover that offer
superior speed; industry-standard
software; and a convenient plate
presentation location.
• At-a-glance water level monitoring eliminates the
need to open the incubator’s front doors, reducing
the risk of contamination. Blue backlighting flashes
when water is low, a subtle reminder to check the
humidity supply.
• Easy access
to water bottle
and snap-open
humidification
chamber
ensures ease
of maintenance
and less
disturbance to
your assays.
10000
Air Classification
Particles >= 0.5 micron
compatible incubation applications. Take full advantage of all that the new Steri-Cult R offers.
• Built at our ISO 9001-2000 registered facility, the
standardized incubator is CE Marked and CSA Certified
to meet worldwide requirements, attesting to the
incubator's electrical and mechanical safety and
compliance with international requirements related to
emission of, and susceptibility to, RF energy and
electrostatic discharge.
1000
100
10
1
0
5
10
15
20
25
Particulate count reduction in a Steri-Cult R
CO2 Incubator with Class 100 HEPA filtration
• Optional HEPA2™ VOC filter removes volatile organic
chemicals (vapors) as well as airborne contaminants
and particulates for added product protection.
Easy-to-Clean Chamber Design
• Humidification is 100% HEPA filtered before
entering the incubator chamber to minimize the
risk of contamination.
• Chamber’s mirror finish, stainless steel interior
with 100% coved corners is easy to clean.
• HEPA filter is located inside the incubator chamber
for optimum filtering, easy access, and simple
replacement without the use of tools.
• User-friendly alphanumeric command set is available
so integrators can program plate storage and retrieval
when the Steri-Cult R is used with a variety of
automation systems.
• Transfer station options ensure greater integration
flexibility.
Accessories
Stock No.
303000
3
4
– contribute to the Steri-Cult R’s full compatibility
with Thermo’s CRS CataLyst-5™, CRS CataLyst
Express™, CRS F3™, and CRS Dimension4™
robots and easy integration into CRS POLARA™
laboratory systems
Description
Microplate hotel, 28 places up to
2.3cm each
Deepwell microplate hotel, 12 places up
to 5.0cm each
Stationary (fixed) transfer station
Plate reorientation transfer station
Swapping transfer station
HEPA filter replacement kit,* HEPA and
in-line filters
HEPA2 VOC filter replacement kit,* HEPA
VOC filter and in-line filters
Built-in CO2 gas guard, monitors
supply and switches to another cylinder,
factory installed
303001
303002
303003
303004
1900160
1900161
1900153
*Customer installed filters are easily replaced without tools.
HEPA filter is rated a minimum 99.97% efficient at 0.3 microns.
90.0
• We made systems compatibility straightforward by
implementing industry standards for the software interface;
plate presentation hardware; and RS-232, V24 9600 baud
robot control interface.
Time (minutes) after Door Closing
• Flush-mount access gate and gate cover
• Filter built in
to water fill
bottle’s lid allows clean air to enter the bottle
while keeping out contaminants.
– permit easy access to the plate (portrait orientation)
and ensure that nothing hinders external robot
contact with the transfer station
88.8
RH
The new Steri-Cult R Automated CO2 Incubator is a winning combination of a sophisticated, robot-
Continuous Air Filtering to Achieve Class 100 Air Quality
Temperature
Continuous Contamination Control: Protecting
Your Valuable Assays
37.0
Temp
36.8
RH
1.0 1.2
Time (minutes)
Small size of the automatic access gate and short 10 second
cycle minimize disruption of the chamber environment.
5.0
4.8
CO2
Steri-Cult® R Automated CO2 Incubator
Features and Technical Specifications
A Winning Combination for Your Drug Discovery Needs
1.0 1.2
Time (minutes)
Advanced dual-beam IR control system ensures
highly stable CO2, even during gate openings.
Airborne and waterborne contamination can affect
product yields and reliability. Some cell-based assays
can take several weeks to culture. You can’t afford
to lose time and productivity if they are destroyed
by contaminants. Steri-Cult R is the only automated
incubator on the market with airborne and waterborne
contamination control.
compatible, automated storage and plate handling system and our well known Steri-Cult CO2
Incubator—the name that brought leading edge technology, such as infrared CO2 sensors,
microprocessor control, directed airflow, and HEPA filtration, to cell culture incubators. Steri-Cult has
been the leading name in high volume, contamination-resistant tissue culture for more than 15 years.
This is a key advantage in assay development, where contamination can bring the whole process to a
The Steri-Cult R’s design integrates three methods
of contamination control: an external humidification
system, HEPA filtration, and careful chamber design.
Contamination-Resistant External Humidification System
stop, costing you time and money.
By uniting our cell culture incubator engineering and manufacturing expertise, highly regarded
laboratory automation systems experience, and the talent of a leading automation designer, Thermo
Electron Corporation developed a specialized drug discovery product that reflects years of solid
research and rigorous testing. The Steri-Cult R automated system is ideal for your demanding
ADME/Tox assays, such as Caco2, as well as other cell-based assays and general automation-
The humidification system is located outside the
incubator chamber, reducing the risk of waterborne
contamination that can cause product loss, incubator
downtime, and wasted time and money.
There’s no humidity pan to check or handle, no relying
on a pan’s mechanical switch to warn you that the
water is low. And no need to disassemble/reassemble
automation components (with the hope that
everything will realign properly) to clean an interior
source of humidification.
Best in Class: Fastest Speed and High Capacity
To protect your
cultures, our
unique HEPA
Filter Airflow
System, located
inside the
chamber,
employs
laminar airflow
techniques and handles primary sources of
contamination (biological contaminants and airborne
particulates) found in most laboratory settings.
Steri-Cult R
provides storage
for standard
shallow-well
microplates or
deepwell
microplates.
The 196 microplate
capacity makes
it a cost-effective
system for your
high throughput
needs by
incubating
more plates,
speeding up the process, and lowering the price per
plate processed.
• Gentle, directed airflow ensures continuous
cleaning during incubator operation to keep
airborne particulates from contaminating and
ruining cultures, costing time and money.
• Microbiological filters protect the gas inlet and
sample port.
Three transfer station configurations allow you to
tailor the equipment to your specific needs: standard
stationary (fixed), optional swapping (swaps positions
of two plates), and optional plate reorientation
(rotates plate from landscape to portrait).
• Entire chamber volume of air is filtered every 60
seconds to prevent contamination. And Class 100
air quality is achieved within five minutes of door
closing. (Testing protocol available.)
Straightforward Integration: Low Risk, High Value
We removed the barriers. It’s easy for you to take
advantage of the Steri-Cult R’s faster access time and
high capacity.
100000
Market-Leading Contamination Control to protect your cellbased assays and cultures, reducing the risk of a costly, time consuming
break in the development process
Best in Class Speed and Capacity for fastest processing and
low cost per plate processed
An Integrator’s Dream due to careful incubator design and
2
adherence to industry hardware and software standards
IDEAL CULTURING AND
HIGH THROUGHPUT
Steri-Cult has long been known for
optimum temperature uniformity and
stability. It is a favorite in biopharm
research labs and high volume clinical
and research facilities. Now we’ve
combined that trusted incubator with a
high capacity, automated, handling
carousel and plate mover that offer
superior speed; industry-standard
software; and a convenient plate
presentation location.
• At-a-glance water level monitoring eliminates the
need to open the incubator’s front doors, reducing
the risk of contamination. Blue backlighting flashes
when water is low, a subtle reminder to check the
humidity supply.
• Easy access
to water bottle
and snap-open
humidification
chamber
ensures ease
of maintenance
and less
disturbance to
your assays.
10000
Air Classification
Particles >= 0.5 micron
compatible incubation applications. Take full advantage of all that the new Steri-Cult R offers.
• Built at our ISO 9001-2000 registered facility, the
standardized incubator is CE Marked and CSA Certified
to meet worldwide requirements, attesting to the
incubator's electrical and mechanical safety and
compliance with international requirements related to
emission of, and susceptibility to, RF energy and
electrostatic discharge.
1000
100
10
1
0
5
10
15
20
25
Particulate count reduction in a Steri-Cult R
CO2 Incubator with Class 100 HEPA filtration
• Optional HEPA2™ VOC filter removes volatile organic
chemicals (vapors) as well as airborne contaminants
and particulates for added product protection.
Easy-to-Clean Chamber Design
• Humidification is 100% HEPA filtered before
entering the incubator chamber to minimize the
risk of contamination.
• Chamber’s mirror finish, stainless steel interior
with 100% coved corners is easy to clean.
• HEPA filter is located inside the incubator chamber
for optimum filtering, easy access, and simple
replacement without the use of tools.
• User-friendly alphanumeric command set is available
so integrators can program plate storage and retrieval
when the Steri-Cult R is used with a variety of
automation systems.
• Transfer station options ensure greater integration
flexibility.
Accessories
Stock No.
303000
3
4
– contribute to the Steri-Cult R’s full compatibility
with Thermo’s CRS CataLyst-5™, CRS CataLyst
Express™, CRS F3™, and CRS Dimension4™
robots and easy integration into CRS POLARA™
laboratory systems
Description
Microplate hotel, 28 places up to
2.3cm each
Deepwell microplate hotel, 12 places up
to 5.0cm each
Stationary (fixed) transfer station
Plate reorientation transfer station
Swapping transfer station
HEPA filter replacement kit,* HEPA and
in-line filters
HEPA2 VOC filter replacement kit,* HEPA
VOC filter and in-line filters
Built-in CO2 gas guard, monitors
supply and switches to another cylinder,
factory installed
303001
303002
303003
303004
1900160
1900161
1900153
*Customer installed filters are easily replaced without tools.
HEPA filter is rated a minimum 99.97% efficient at 0.3 microns.
90.0
• We made systems compatibility straightforward by
implementing industry standards for the software interface;
plate presentation hardware; and RS-232, V24 9600 baud
robot control interface.
Time (minutes) after Door Closing
• Flush-mount access gate and gate cover
• Filter built in
to water fill
bottle’s lid allows clean air to enter the bottle
while keeping out contaminants.
– permit easy access to the plate (portrait orientation)
and ensure that nothing hinders external robot
contact with the transfer station
88.8
RH
The new Steri-Cult R Automated CO2 Incubator is a winning combination of a sophisticated, robot-
Continuous Air Filtering to Achieve Class 100 Air Quality
Temperature
Continuous Contamination Control: Protecting
Your Valuable Assays
37.0
Temp
36.8
RH
1.0 1.2
Time (minutes)
Small size of the automatic access gate and short 10 second
cycle minimize disruption of the chamber environment.
5.0
4.8
CO2
Steri-Cult® R Automated CO2 Incubator
Features and Technical Specifications
A Winning Combination for Your Drug Discovery Needs
1.0 1.2
Time (minutes)
Advanced dual-beam IR control system ensures
highly stable CO2, even during gate openings.
Steri-Cult® R Automated CO2 Incubator
Features and Technical Specifications
Specifications
Automation
Plate Storage. . .
Capacity . . . . . .
Access Time . . .
Robot Interface .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
7 x 28 position hotels
196 shallow-well microplates
20 seconds maximum
RS-232, V24 9600 baud
Temperature
Control . . . . . . . . . . . . . . . . . ±0.1C @ 37C (98.6F)
Range . . . . . . . . . . . . . . . . . . 10C above ambient to
50C (122F)
Sensor. . . . . . . . . . . . . . . . . . Thermistor
Controller . . . . . . . . . . . . . . . Microprocessor
Setpoint . . . . . . . . . . . . . . . . Digital
Display . . . . . . . . . . . . . . . . . Digital LED
Readability and Setability . . . . 0.1C
Uniformity . . . . . . . . . . . . . . . ±0.5C
Temperature Safety
Sensor. . . . . . . . . . . . . . . . . . Thermistor
Controller . . . . . . . . . . . . . . . Microprocessor
Setability . . . . . . . . . . . . . . . . 0.1C
CO2
Control . . . . . . . . . . . . . .
Range . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . .
Inlet Pressure . . . . . . . . .
Filter . . . . . . . . . . . . . . . .
Sensor. . . . . . . . . . . . . . .
Controller . . . . . . . . . . . .
Display . . . . . . . . . . . . . .
Readability and Setability .
Tracking Alarm. . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
±0.1% @ 5.0%
0-20%
Automatic
15 PSIG (1.0 bar)
0.2 micron, disposable
Dual-beam IR
Microprocessor
Digital LED
0.1%
User programmable high/low
Controlled RH
RH . . . . . . . . . . . . . . .
Control . . . . . . . . . . . .
Sensor. . . . . . . . . . . . .
Controller . . . . . . . . . .
Readability & Setability
Humidity Reservoir . . .
Tracking Alarm. . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Ambient to 95% RH
±2.0%
Capacitive
Microprocessor
1%
1.0 gal. (3.8 liters)
User programmable high/low
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Unit Heat Load
3387 . . . . . . . . . . . . . . . . . . . 595 BTUH (175 Watts)
3388 . . . . . . . . . . . . . . . . . . . 647 BTUH (190 Watts)
Dimensions
Exterior . . . . . . . . . . . . . . . . . 35.0"W x 39.4"H x 31.7"F-B
(88.9cm x 100.1cm x 80.5cm)
Interior . . . . . . . . . . . . . . . . . 20.8"W x 32.8"H x 24.6"F-B
(52.8cm x 83.3cm x 62.5cm)
Transfer Height . . . . . . . . . . . 8.1" (20.6cm)
Construction
Interior Volume . . . . . . . . . . . 9.7 cu. ft. (274.7 liters)
Interior . . . . . . . . . . . . . . . . . Type 304 mirror finish,
stainless steel
Exterior . . . . . . . . . . . . . . . . . 18 gauge, cold-rolled steel,
powder coated
Inner Door . . . . . . . . . . . . . . . 1/4" (0.6cm) fully tempered
safety glass with cam action
latch
Outer Door Gasket . . . . . . . . Four-sided, molded, magnetic
vinyl
Inner Door Gasket . . . . . . . . . Feather, silicone
People of Science Working
with People of Science
A Leader in Controlled Environment
Equipment
One Convenient Source, Total Solutions
Whether you are storing, preparing, culturing, or analyzing
samples, Thermo has the research tools you depend on to
succeed in the lab. We have been providing high quality,
reliable controlled environment laboratory products for
a wide variety of clinical, educational, biotechnological,
pharmaceutical, and industrial applications for more
than 50 years.
We offer the following:
Thermo’s expertise extends beyond controlled
environment equipment. In fact, you may be surprised
by the breadth of products and services we offer.
Our life and laboratory sciences and services, optical
technologies, measurement and control, turnkey
operations, and even capital equipment financing
specialists are eager to share their knowledge and work
as your partner...for life.
The benefit? We’re people of science working with people
of science. And you will quickly discover the value of one
convenient source providing total solutions tailored to your
specific needs.
• Forma Cell Culture Incubators
• Forma In Vitro Fertilization (IVF) Incubators and
CryoMed® Controlled Rate Freezers
Electrical
3387 . . . . . . . . . . . . . . . . . . . 115V, 50/60 Hz, 5.2 FLA
(Operating range 90-125V)
3388 . . . . . . . . . . . . . . . . . . . 230V, 50/60 Hz, 3.6 FLA
(Operating range 180-250V)
Power Switch . . . . . . . . . . . . 2 Pole
Plug. . . . . . . . . . . . . . . . . . . . NEMA 5-15P (3387),
CEE 7/7 (3388)
Convenience Receptacle . . . . 115V standard/220V export
(75 Watts maximum)
Alarm Contacts . . . . . . . . . . . Deviation of temperature,
CO2, RH, and power failure.
Accessed using RJ11 jack on
rear of unit.
• Forma Environmental Chambers
Estimated Weight
Net . . . . . . . . . . . . . . . . . . . . 480 lbs. (217.7 kg)
Shipping (Motor) . . . . . . . . . . 550 lbs. (249.5 kg)
• Forma Lab Refrigerators
Controlled Environment
• Forma Ultra-Low Temperature Freezers
Equipment Solutions
• Forma Laminar Airflow Equipment
• Forma Biological Safety Cabinets
for the Lab
• Forma Cryopreservation Equipment
• Forma Blood Bank Equipment
• Forma Circulating Baths
• Forma Monitor/Alarm System
• Forma Orbital Shakers
• Forma Programmable Centrifuges
• Forma Lab Washers/Dryers
• Forma Lab Freezers
• Forma Chromatography Refrigerators
Fittings
Drain Port . . . . . . . . . . . . . . . 3/8" barbed fitting with shutoff
Access Port. . . . . . . . . . . . . . 1.4" (3.6cm) with removable
silicone plug
CO2 Inlet . . . . . . . . . . . . . . . . 1/4" hose (barbed)
These specifications are for informational purposes only and are subject to change without notice. Thermo makes no warranties, expressed or implied,
in this product summary.
5
TRANSFER STATION
DIMENSIONS
Measurements displayed are in inches.
• Custom-Designed Equipment
Steri-Cult® R Automated CO2 Incubator
• Laboratory Services
About Thermo
Thermo provides scientific instruments, equipment,
software, and solutions that help laboratory and industrial
customers advance scientific knowledge, enable drug
discovery, improve manufacturing processes, and protect
people and the environment.
With annual revenue exceeding $2 billion, Thermo is a
global leader employing 11,000 people in 30 countries
worldwide. The company is based in Waltham,
Massachusetts.
© 2003 Thermo Electron Corporation. All rights reserved. Thermo Electron Corporation, question everything, and Analyze. Detect. Measure. Control are trademarks of Thermo Electron Corporation. Steri-Cult, Forma,
HEPA2, and CryoMed are trademarks or registered trademarks of Thermo Forma. CRS CataLyst-5, CRS CataLyst Express, CRS F3, CRS Dimension4, and CRS POLARA are trademarks of Thermo CRS.
Rev. 0, 07/03 5M
Controlled Environment Equipment
P.O. Box 649 • Marietta, OH 45750
740-373-4763 • Fax: 740-373-6770
Toll Free USA and Canada 866-984-3766
www.thermo.com/forma
Analyze • Detect • Measure • Control™
Steri-Cult® R Automated CO2 Incubator
Features and Technical Specifications
Specifications
Automation
Plate Storage. . .
Capacity . . . . . .
Access Time . . .
Robot Interface .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
7 x 28 position hotels
196 shallow-well microplates
20 seconds maximum
RS-232, V24 9600 baud
Temperature
Control . . . . . . . . . . . . . . . . . ±0.1C @ 37C (98.6F)
Range . . . . . . . . . . . . . . . . . . 10C above ambient to
50C (122F)
Sensor. . . . . . . . . . . . . . . . . . Thermistor
Controller . . . . . . . . . . . . . . . Microprocessor
Setpoint . . . . . . . . . . . . . . . . Digital
Display . . . . . . . . . . . . . . . . . Digital LED
Readability and Setability . . . . 0.1C
Uniformity . . . . . . . . . . . . . . . ±0.5C
Temperature Safety
Sensor. . . . . . . . . . . . . . . . . . Thermistor
Controller . . . . . . . . . . . . . . . Microprocessor
Setability . . . . . . . . . . . . . . . . 0.1C
CO2
Control . . . . . . . . . . . . . .
Range . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . .
Inlet Pressure . . . . . . . . .
Filter . . . . . . . . . . . . . . . .
Sensor. . . . . . . . . . . . . . .
Controller . . . . . . . . . . . .
Display . . . . . . . . . . . . . .
Readability and Setability .
Tracking Alarm. . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
±0.1% @ 5.0%
0-20%
Automatic
15 PSIG (1.0 bar)
0.2 micron, disposable
Dual-beam IR
Microprocessor
Digital LED
0.1%
User programmable high/low
Controlled RH
RH . . . . . . . . . . . . . . .
Control . . . . . . . . . . . .
Sensor. . . . . . . . . . . . .
Controller . . . . . . . . . .
Readability & Setability
Humidity Reservoir . . .
Tracking Alarm. . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Ambient to 95% RH
±2.0%
Capacitive
Microprocessor
1%
1.0 gal. (3.8 liters)
User programmable high/low
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Unit Heat Load
3387 . . . . . . . . . . . . . . . . . . . 595 BTUH (175 Watts)
3388 . . . . . . . . . . . . . . . . . . . 647 BTUH (190 Watts)
Dimensions
Exterior . . . . . . . . . . . . . . . . . 35.0"W x 39.4"H x 31.7"F-B
(88.9cm x 100.1cm x 80.5cm)
Interior . . . . . . . . . . . . . . . . . 20.8"W x 32.8"H x 24.6"F-B
(52.8cm x 83.3cm x 62.5cm)
Transfer Height . . . . . . . . . . . 8.1" (20.6cm)
Construction
Interior Volume . . . . . . . . . . . 9.7 cu. ft. (274.7 liters)
Interior . . . . . . . . . . . . . . . . . Type 304 mirror finish,
stainless steel
Exterior . . . . . . . . . . . . . . . . . 18 gauge, cold-rolled steel,
powder coated
Inner Door . . . . . . . . . . . . . . . 1/4" (0.6cm) fully tempered
safety glass with cam action
latch
Outer Door Gasket . . . . . . . . Four-sided, molded, magnetic
vinyl
Inner Door Gasket . . . . . . . . . Feather, silicone
People of Science Working
with People of Science
A Leader in Controlled Environment
Equipment
One Convenient Source, Total Solutions
Whether you are storing, preparing, culturing, or analyzing
samples, Thermo has the research tools you depend on to
succeed in the lab. We have been providing high quality,
reliable controlled environment laboratory products for
a wide variety of clinical, educational, biotechnological,
pharmaceutical, and industrial applications for more
than 50 years.
We offer the following:
Thermo’s expertise extends beyond controlled
environment equipment. In fact, you may be surprised
by the breadth of products and services we offer.
Our life and laboratory sciences and services, optical
technologies, measurement and control, turnkey
operations, and even capital equipment financing
specialists are eager to share their knowledge and work
as your partner...for life.
The benefit? We’re people of science working with people
of science. And you will quickly discover the value of one
convenient source providing total solutions tailored to your
specific needs.
• Forma Cell Culture Incubators
• Forma In Vitro Fertilization (IVF) Incubators and
CryoMed® Controlled Rate Freezers
Electrical
3387 . . . . . . . . . . . . . . . . . . . 115V, 50/60 Hz, 5.2 FLA
(Operating range 90-125V)
3388 . . . . . . . . . . . . . . . . . . . 230V, 50/60 Hz, 3.6 FLA
(Operating range 180-250V)
Power Switch . . . . . . . . . . . . 2 Pole
Plug. . . . . . . . . . . . . . . . . . . . NEMA 5-15P (3387),
CEE 7/7 (3388)
Convenience Receptacle . . . . 115V standard/220V export
(75 Watts maximum)
Alarm Contacts . . . . . . . . . . . Deviation of temperature,
CO2, RH, and power failure.
Accessed using RJ11 jack on
rear of unit.
• Forma Environmental Chambers
Estimated Weight
Net . . . . . . . . . . . . . . . . . . . . 480 lbs. (217.7 kg)
Shipping (Motor) . . . . . . . . . . 550 lbs. (249.5 kg)
• Forma Lab Refrigerators
Controlled Environment
• Forma Ultra-Low Temperature Freezers
Equipment Solutions
• Forma Laminar Airflow Equipment
• Forma Biological Safety Cabinets
for the Lab
• Forma Cryopreservation Equipment
• Forma Blood Bank Equipment
• Forma Circulating Baths
• Forma Monitor/Alarm System
• Forma Orbital Shakers
• Forma Programmable Centrifuges
• Forma Lab Washers/Dryers
• Forma Lab Freezers
• Forma Chromatography Refrigerators
Fittings
Drain Port . . . . . . . . . . . . . . . 3/8" barbed fitting with shutoff
Access Port. . . . . . . . . . . . . . 1.4" (3.6cm) with removable
silicone plug
CO2 Inlet . . . . . . . . . . . . . . . . 1/4" hose (barbed)
These specifications are for informational purposes only and are subject to change without notice. Thermo makes no warranties, expressed or implied,
in this product summary.
5
TRANSFER STATION
DIMENSIONS
Measurements displayed are in inches.
• Custom-Designed Equipment
Steri-Cult® R Automated CO2 Incubator
• Laboratory Services
About Thermo
Thermo provides scientific instruments, equipment,
software, and solutions that help laboratory and industrial
customers advance scientific knowledge, enable drug
discovery, improve manufacturing processes, and protect
people and the environment.
With annual revenue exceeding $2 billion, Thermo is a
global leader employing 11,000 people in 30 countries
worldwide. The company is based in Waltham,
Massachusetts.
© 2003 Thermo Electron Corporation. All rights reserved. Thermo Electron Corporation, question everything, and Analyze. Detect. Measure. Control are trademarks of Thermo Electron Corporation. Steri-Cult, Forma,
HEPA2, and CryoMed are trademarks or registered trademarks of Thermo Forma. CRS CataLyst-5, CRS CataLyst Express, CRS F3, CRS Dimension4, and CRS POLARA are trademarks of Thermo CRS.
Rev. 0, 07/03 5M
Controlled Environment Equipment
P.O. Box 649 • Marietta, OH 45750
740-373-4763 • Fax: 740-373-6770
Toll Free USA and Canada 866-984-3766
www.thermo.com/forma
Analyze • Detect • Measure • Control™