Mechanical Behavior of a Spring

advertisement
Mechanical Behavior of a Spring
R. Hooke, DePotentia Restitutiva (1678)
We have measured the strength “k” of a mechanical spring using both static and
dynamic methods. In the static method, we explored Hooke’s Law, F(x) = -kx, by
recording stretch length for various masses hung on a spring and obtained ks = 1.45 ±
0.05 N/m. In the dynamic method, we measured oscillation period for the same set-up
and obtained kd =1.51 ± 0.03 N/m. These two methods agree within their uncertainty, but
the dynamic method is more precise. The static method revealed a built-in tension, such
that the “rest length” L0 ~ 10cm was less than the physical length Lm=12cm for no mass.
The dynamic method revealed that the mass of the spring itself contributes an effective
mass of m0= 110 ± 20 g.
Introduction
In this lab the mechanical behavior of a spring is investigated. Springs are characterized
by a linear restoring force and exhibit simple harmonic motion. This behavior occurs in many
other physical systems, such as 3D solids, waves, electrical circuits, etc. The concept is central to
many subjects, and is discussed at length in physics, electrical engineering and mechanical
engineering textbooks. Here we will measure the force constant using both static and dynamic
methods and compare with Hooke’s law.
Theory
An ideal spring follows a linear force-distance relation and is massless. The static
behavior then is described by Hooke’s Law:
F(X) = -ks (X-X0),
Eq. 1
where F is the force exerted by the spring (opposing the external stretch force), (X-X0) is
the stretch distance, measured from the resting position X0 where the force is zero, and ks
is the static spring constant. In this lab, the force is generated by calibrated weights hung
on a vertically suspended spring. Any system described by Hooke’s law will exhibit
“simple harmonic motion" if it is displaced from X0 and released. The system will
oscillate at a frequency given by
2 = kd/m,
Eq. 2
where is the angular frequency and kd is a dynamic spring constant. For perfect
systems, we expect kd = ks. Real springs have a non-zero mass, but still exhibit a single
oscillation frequency, given by





2 = kd/(m+m0)
where "m" is the applied mass and m0 is an effective mass. It can be shown by
Lagrangian method, that m0 is 1/3 of the physical spring mass (Analytic Mechanics,
Fowles and Cassiday).
Sample Lab: Springs
p1/6
9/20/2010
Eq. 3
Procedure
For the static measurements, calibrated weights were hung from the spring and the
overall length between the centers of the two end-hooks was measured using a meter stick, by
sighting carefully to minimize parallax. Data are taken first by adding and then by removing
weights to establish uncertainties.
For the dynamic measurements, the elapsed time for 3 complete cycles of oscillation was
recorded for each mass. Both the start and stop times were marked at the top of the oscillation,
after “synchronizing” to the pace of the motion. The amplitude was kept below 3 cm, to avoid
non-linearity. A small decay of amplitude after 3 beats was noticed, but is not expected to affect
the average period.
Results
Fig. 1 Spring length vs applied mass. The circles are for adding
weights, the squares for subtracting weights.
Figure 1 shows the data for the static measurement, which are tabulated in the appendix
as Table A1. The “adding” and “subtracting” data agree within ± 0.5 cm, which
establishes the random uncertainties. The data show a linear trend, with no systematic
deviation, except for applied mass below m = 0.4kg where the length is constant because the
coils are touching. This region is excluded from the analysis (see below). The remaining data are
fit to equation 1, using plot variables L = (x-x0) and F = mg. Thus, the fit line is L (m) = -g/k *
Sample Lab: Springs
p2/6
9/20/2010
m, with slope S = 6.75 ± 0.24 cm/kg, and intercept I = 9.8 ± 0.24 cm. This yields a static spring
constant ks = 1.45 ± 0.05 N/m. Calculations are shown in the appendix.
Figure 2. Linearized plot of oscillation period vs mass.
Figure 2 shows the dynamic behavior, which is tabulated in the appendix, Table A2.
These data are fit with a linearized plot to equation 3, using plot variable T2, where T2= (2pi/)2
= (2pi)2/k * (m+m0). The fit region is limited to m > 0.4kg, as for the static data. The fit line
yields slope S = 26.1 ± 0.5 sec2/kg and intercept I = 2.84 ± 0.5 sec2, corresponding to a dynamic
spring constant kd = 1.51 ± 0.03 N/m and effective mass m0 = 110 ± 20 g. Calculations are
shown in the appendix.
Discussion
For the static measurement, the fit to Equation 1 is good. That is, there is no
systematic deviation from the trend. All physical systems that follow this behavior shall
henceforth be said to follow Hooke’s Law. We note that, for applied mass less than 0.4kg,
the spring has a fixed length of 12 cm. This can be understood as a pre-loaded tension,
corresponding to a “rest length” of L0 = 10.2 ± 0.14 cm which is shorter than the physical
length of Lm = 12 cm. The scatter in the data (random uncertainty) is about ± 0.5cm. This
is attributed to parallax errors in reading the length. This could be improved using a
travelling microscope, or more simply with a mirror assembly next to the ruler to allow
auto-collimation of the sighting. It was expected that the stretch length might drop below
the trendline for large applied force, due to material limitations (real springs get stronger),
but this was not observed. This behavior might become visible with reduced scatter.
For the dynamic measurement, the fit to Equation 3 is good, meaning there is no
systematic deviation from the trend, within the scatter of approximately 0.7 sec.
Concerning this scatter: Some uncertainty is expected due to errors of start and stop times,
Sample Lab: Springs
p3/6
9/20/2010
but these are minimized by “synching” to the motion before timing, and are estimated to
be ~ 0.3 sec. This could be improved, in any case, using a photo-gate. We suspect that the
scatter mainly arises from variation in the amplitude of the motion, either due to variable
starting displacement or due to a variable waiting time before the start of timing. This
could be improved by more diligent procedure.
The results for the static vs dynamic method are shown in Table 1. We find that
these two methods agree within their uncertainties. The dynamic method is more precise.
Method
Static
Dynamic
Value for k (N/m)
1.45 ± 0.05
1.51 ± 0.03
Table 1. Comparison of static vs dynamic results for spring constant.
Sample Lab: Springs
p4/6
9/20/2010
Appendix
Mass
(kg)
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
1.10
1.20
1.30
1.40
1.50
Length (cm)
increasing
12.0
12.0
12.0
12.8
13.3
13.9
14.8
15.1
16.1
16.4
17.4
17.9
18.7
19.0
19.6
Length (cm)
decreasing
12.0
12.0
12.0
12.5
13.3
14.0
14.4
15.2
15.9
16.4
17.1
18.0
18.5
19.1
19.9
Table A1. Length of spring for various applied masses.
Mass (kg)
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
1.10
1.20
1.30
1.40
1.50
3*T (sec)
0.0
0.0
0.0
11.2
12.2
12.7
13.6
14.6
15.3
16.0
16.8
17.7
18.1
18.9
19.5
T2 (sec2)
0.0
0.0
0.0
14.0
16.5
18.1
20.4
23.8
25.9
28.3
31.4
34.7
36.3
39.7
42.5
Table A2. Oscillation period vs applied mass.
Sample Lab: Springs
p5/6
9/20/2010
Sample Lab: Springs
p6/6
9/20/2010
Download