I Direct Determination of Aerobic P-ower James A. Davis, PhD California State University at Long Beach M easurement of aer?bic power or maximal oxygen uptake (V02max) is one of the oldest and most common measurements in exercise physiology. This chapter outlines the historical development of the equipment used to make the measurement, the exercise test modes and protocols, and the criteria used to determine achievement of V02max. The last section of the chapter deals with reference values for V02max. The parameters of aerobic function are aerobic power or maximal oxygen uptake (V02max), work efficiency, time constant for V02 kinetics, and the lactate threshold (Whipp, Davis, Torres, & Wasserman, 1981). Maximal V02 is an important parameter because it represents the upper limit of aerobic exercise tolerance. Endurance activities are performed at some fraction of V02max. If V02max is low, then the level of endurance performance is necessarily constrained. Whole-body oxygen uptake can be determined from cardiovascular measurements or from respiratory measurements, using equations based on the Fick principle. From cardiovascular measurements, V02 = e.o. (Ca02 - CV02), where CO. is cardiac output, Ca02 is content of O2 in arterial blood, and CV02 is content of O2 in mixed venous blood. From respiratory measurements, V02 = VA(FP2 - FP2) where VAis alveolar ventilation, F,02 is the fraction of O2 in inspired gas, and Fti02 is the fraction of O2 in the mean alveolar gas; this equation is not rigorous because it does not account for the normally small differences between inspired and expired alveolar volumes. Thus, the determinants ofV02 are the heart (CO.), the mechanical properties of the lungs and the chest wall (VA)'diffusion of O2from the alveolus into the pulmonary capillary blood (FX02 and Ca02), and the extraction of O2 from the capillary blood by the muscle cell (CV02)' A person with a high V02max necessarily has good function in each of these determinants. Conversely, a sedentary person has relatively poor function for each determinant, which results in a low V02max. If a person has pathology associated with one of these determinants (e.g., coronary artery disease that would limit the cardiac output during exercise), then the V02max will be very low. Hence, one of the important reasons for measuring V02 during graded exercise testing (eXT) is to establish whether the V02max is normal. Two issues are unresolved regarding the measurement of V02max. The first is the criteria used to establish whether V02max has been achieved during eXT. The most widely accepted criterion is that the V02 plateaus during the later stages of the eXT as the work rate continues to increase. Many subjects, however, clearly reach their limit of tolerance during eXT without demonstrating a plateau in V02. The second unresolved issue is the recommendation that if a subject does not demonstrate a plateau in V02, then the term should be V02peak, not V02max. A later section of this chapter more fully discusses both of these unresolved issues. . 9 ~~av~isL- _ Equations First, it is necessary to present the measurements that need to be made. The basic equation is as follows: (1) where VI is inspired ventilation, VE is expired ventilation, FP2 is the fraction of O2 in the inspired gas, and FfP2 is the fraction of O2 in the mixed expired gas. However, it is generally assumed that inspired nitrogen (VlIN 2) equals expired nitrogen (VEF~2)' Thus, Vf~2= VE'eN2 . = V[ iN2 VFEFN [ where FiN 2 = 1- or (F E °+ 2 F E CO 2 ) (2) 2 . Replacement of V[ with VE F-N E F[N2 2 ( ) in equation 1 yields (3) (4) Under normoxic (room air) conditions, dry FP2 equals 0.2093 and dry FIN2 equals 0.7904. The ratio of dry inspired O2 to dry inspired N2 is 0.265. Hence, equation (4) can be simplified to V02 = V E [FeN2 (0.265) - FE02] (5) Therefore, the quantities to measure for determination of V02 at the mouth are VE, FE02, and FEC02• • By convention, V02 is expressed under standard temperature and pressure dry (STPD) gas conditions. Because VEis typically measured under atmospheric temperature and pressure saturated (ATPS) gas conditions, it must be converted to STPD gas conditions. The STPD correction factor can be calculated using the following equation: STPD correction factor = 273 oK 273°K+TA )x (PB -PH2o) 60 7 (6) where K is degrees Kelvin, TAis the gas temperature at the time of measurement, PB is the barometric pressure, and PH20 is the water vapor pressure of saturated gas at TA.Values of PH20 at various values ofT, can be found in many exercise physiology textbooks. .•....., ..- .•.•...... 0 , ( , :-. '.1i1U·~i!.~~'J~':~~?"'::': r ... Hlstorlcel Development of V02 Measurement One of the first open-circuit equipment configurations used to measure V02 during exercise was the Tissot spirometer-volumetric gas analyzer system. The Tissot spirometer measured expired ventilation. An aliquot sample of the mixed expired gas was drawn from the spirometer and analyzed by a volumetric gas analyzer (e.g., Haldane or Scholander) for fractional concentrations of O2 and CO2, A slight variation of the Tissot spirometervolumetric gas analyzer system was the Douglas bag-volumetric gas analyzer system. In later years, lightweight meteorological balloons replaced the heavy, bulky Douglas bags and a gas meter replaced the Tissot spirometer for measurement of gas volume. Over time, electronic gas analyzers replaced the volumetric gas analyzers. This system is called the meteorological balloon-electronic gas analyzer system.', The next development in equipment configuration was the semiautomated system (Wilmore &. Costill, 1974). In this system, the subject's expired air first went into a mixing chamber and then into .agas meter. A pump pulled approximately 500 ml of gas per minute from the mixing chamber and sent it to a 2 L latex bag; a second pump pulled the gas from the bag and sent it to the electronic analyzers for determination ofFE02 and FEC02• The system used three bags. While the gas analyzers were sampling the contents of one bag, a second bag was being filled with gas from the mixing chamber and a third bag was being evacuated. The bags were part of a spinner device that was manually rotated 120 each minute so that the evacuated bag would move to a new position and be filled with gas from the mixing chamber. The other bags would likewise move to new positions. A problem with this mixing-chamber system is that the measurement of ventilation and the mixed expired gas fractions occur at different times. Ventilation is measured without any delay, but measurement of the mixed expired gases is delayed. This delay has two components. One is the transport delay between the subject's mouth and the analyzers. The other delay is due to the response time of the analyzers. Unless the mixed expired gas fractions are not changing, the calculated V02 will be in error for this system. The magnitude of the error depends on the degree of temporal misalignment between the measurements of ventilation and the mixed expired gas fractions. Fortunately, at heavy exercise, the ventilation is usually large enough to cause the temporal misalignment to be quite small, resulting in little error in the calculated V02max value. 0 J ~. 1 --.-.-.---'----.-----.--- -. ~-'>":' - Bowing the semiautomated system was the mated system developed by Beckman Instrunts (Wilmore, Davis, ~ Norton, 1976) called 'metabolic measuring <tart(MMC). This system -sured ventilation with a turbine volume transer (Davis & Larnarra. 1984), had a mixing ber, and measured the mixed expired gas .ons with O2 and CO2 electronic gas analyzers. e MMC did not have the latex bag arrangement f the semiautomated system. The gas drawn from e mixing chamber was sent directly to the O2 and 2gas analyzers. A small computer was an attrace feature of the MMC. The computersampled the alog signals from the turbine volume transducer d the electronic gas analyzers over a duration of ically 1 min. The computer would then calculate e standard variables of interest (e.g., V02, VC02, d R, the respiratory exchange ratio) and output em via a printer a few seconds after the end of the mpling interval. The MMC had many advantages over its predecesrs. It eliminated hand calculation of the variables f interest and manual turning of the spinner device er each sampling interval, and it provided nearly . nline analysis of V02 and VC02. A limitation of the MMC was that it made no attempt to align ventilation and the mixed expired gas fractions temporally. However, the second generation of the MMG called the MMC Horizon, was programmed to time align these signals so that the accuracy of the computed V02 and VC02 would be independent of the ventilation magnitude (Jones, 1984). The equipment configurations just described collected or sampled many breaths over some time interval, typically 1 min. The calculated V02 represented the average V02 during the sampling interval. These types of equipment configurations are sufficient for the measurement of V02max. But other parameters of aerobic function, most notably the time constant for V02 kinetics, require a greater density of data. Hence, an equipment configuration was designed to provide online, breath-by-breath measurement of ventilation and the gas concentrations (Beaver, Wasserman, & Whipp, 1973). Now, more than 12 companies manufacture over 20 automated systems. In a comprehensive review of automated systems, Macfarlane (2001) provides a table show-. ing the details of these systems. Of the 22 listed as laboratory-based systems, 12 are breath-by-breath systems, 7 are mixing-chamber systems, and 3 can provide either breath-by-breath or mixing-chamber measurements. The manufacturers of current automated systems use a variety of technologies to measure V£,FE02, and __ ... • _, .<I ~----.D.ir.ect.D.eterminatLan-QLAeIDb.k:..P..oYie.r~ .••••• ~i...:=- . ._~~. FEC62:~Measurement of VE is made with a turbine device, a pneumotachometer, Pitot tubes, or a hot wire sensor. Technically, these four devices measure flow. Volume is found by integrating the area under the flow curve. A few manufacturers use a mass spectrometer to measure both F-02 and F-C0 2, but E E most of them use separate analyzers to measure the fractional concentrations of each gas. Most manufacturers use infrared radiation absorption technology for the FeC02 measurement and a polargraphic electrode, a paramagnetic analyzer, or a fuel cell for the Fe02 measurement. Several issues are specific to automated systems. One is the temporal alignment of the ventilation and gas concentration measurements. If misalignment is present, especially in breath-by-breath systems, large errors (as much as 30% at high breathing frequencies) can occur in the calculation of V02 (Proctor & Beck, 1996). Some of the sensors for the VE measurement and the Fe02 and FEC02 measurements generate nonlinear outputs given linear inputs. Automated systems typically solve these nonlinear hardware problems with software corrections. Knowing the temperature and humidity of gas collected in a meteorological balloon is easy, but it is not so easy to know this information during each minute of a GXT for the flow and gas concentration sensors of an automated system. Hughson, Northey, Xing, Dietrich, and Cochrane (1991) have shown that errors in the gas temperature measurement, or the assumed value of that measurement, can result in errors in the computed V02, for example, a 3.6% error in V02 for a 2 C error in the temperature measurement or the assumed value of that measurement. A number of independent studies have been published dealing with the validity and reliability of automated systems. A summary of these studies presented by Macfarlane (2001) provides evidence that several of these systems are both valid and reliable. Finally, the cost of a typical automated system-$20,000 to $25,000-is a significant issue. Given the complexities of these systems, some investigators purchase maintenance contracts with their systems, which further increase the cost. Of the equipment configurations described, only the automated systems and the meteorological balloon-electronic gas analyzer system are used routinely today. Fay, Londeree, LaFontaine, and Volek (1989) used the latter system to measure V02max in female distance runners. Expired gas was directed through 3.175 em ID smooth rubber tubing, corrugated plastic tubing, and a three-way valve into meteorological balloons. The gas was collected continuously in 30 s intervals after the subject indicated 0 .lL.....f2oyj5.-- _ tures of any system are desirable. First, the system should not have any leaks. Second, the resistance to inspit'ation-6r'~explration caused by the system - -, should be less than 5 cm H20 pressure at any ventilation. Third, the device used to measure gas volume should provide measurements to within 3% of the true value. Fourth, the device or devices used to measure fractional gas concentrations should be able to measure both O2 and CO2 to within 0.0003 of their true values. that she could only run 90 s longer. An aliquot sample was' taken from each meteorological balloon and measured-for FE02 with a paramagnetic gas analyzer and F CO2 with an infrared gas analyzer. Reference gases for calibration of the electronic gas analyzers were verified with the Scholander apparatus. Expired gas volumes were measured with a gas meter. Oxygen uptake was calculated using equation (5), presented earlier in this chapter. McArdle, Karch, and Pechar (1973) analyzed the test-retest reliability data for the V02max measurement by the meteorological balloon-electronic gas analyzer system. They performed duplicate cycle ergometer GXTson 15 college-age males. The mean ±SDV02max values for the first and second tests were 4.157 ± 0.445 and 4.146 ± 0.480 L· min-I, respectively. The standard error of estimate was low (0.094 L· min-I), and the correlation coefficient was high (.959). Hence, the meteorological balloon-electronic gas analyzer system can provide excellent test-retest reliability. Whether one uses a meteorological balloonelectronic gas analyzer system or an automated system to measure V02max, several general fea- Exercise Test Modes The most popular modes for graded exercise testing are the cycle ergometer and the motor-driven treadmill. Both modes can be controlled manually or by computer. For cycle ergometers, the braking mechanism is done either by friction or electronically. The electronic cycle ergo meters are more expensive and more difficult to calibrate. Maximal V02 values in typical people are approximately 10% higher when measured during treadmill running compared with cycle ergometry (see figure 2.1) (Davis & Kasch, 1975). 5~--------------------------------------~ 4 ~ 'c 'E dx(1J E 3 0 C\I 0 .> .E 2 '0 (1J Q) f!:: y=x O~-----''------.-------.-------r------~ 4 5 o 2 3 Cycle ergometer V02 max (L • rnirr") Figure 2.1 Maximal V02 measured during treadmill running compared with cycle ergometry. Data from Davis and Kasch. 1975. I I I II I I ! I 'I t I i I 1 _____.__. . . .. . . D.iLecLD.eierruio.atioD-ULAe.mbkk>wer 13 Exercise Test Protocols The most widely used exercise testing protocols are continuous and graded. The study of Buchfuhrer et al. (1983) is worthy ofreview before specific GXTprotocols are discussed. Buchfuhrer et al. examined the possibility that the V02max measurement was protocoldependent. They found that "fast" protocols (i.e., those with large work rate increments per minute) and" slow" protocols (i.e., those with small work rate increments per minute) caused underestimations of the true V02max value, which was found using "intermediate speed" protocols. Buchfuhrer et al. (1983) suggested that the fast protocols caused subjects to terminate the GXTearly because they had insufficient muscle strength to accommodate the large work rate increases during the final stages of the test. Two reasons probably explain why Buchfuhrer et al. (1983) found low V02max values for the slow protocols. First, these protocols, which lasted an average of 18 min for cycle ergometry and 26 min for treadmill exercise, would likely result in a significant increase in core temperature. This increase, in turn, would result in a redistribution of the cardiac output so that less blood (and, therefore, less O2) would be going to the exercising musculature and more blood would be going to the cutaneous circulation in an effort to dissipate heat. Lessblood flow (and therefore less O2 delivery) to the working muscles at maximal work rates would explain the lower V02max values found for the slow protocols. A second plausible explanation for this finding is subject motivation. The slow protocols are particularly exhausting, requiring high motivation on the part of the subject to deal with the high levels of lactate and heat associated with heavy, prolonged exercise. Buchfuhrer et al. (1983) found the highest V02max values with GXTprotocols that lasted 8 to 12 min. Using the results of Buchfuhrer et al. (1983), GXT protocols should be designed in such a way to cause the test to end somewhere between 8 and 12 min. An example will demonstrate how the increment size can be found. Consider a subject with a predictedV02maxof3,OOO ml min' for a cycle ergorneterGXT.According to Wasserman and Whipp (1975), the relationship between V02 in ml mirr ' and work rate (WR) in watts (W) for cycle ergometry is given by the following linear regression equation: V02 = 10 ml-mirr' . WR + 500 ml-mirr" (7) Solving for the work rate that would result in a predicted V02 of3,000 ml- mirr ' yields 250 W. The predicted increment size per minute that would produce a test duration of 8 min is 31 W (250 Wj8 min). For a test duration of 12 min, the predicted increment size per minute would be 21 W (250 Wj12 min). A prudent choice of the increment size per minute would be 25 W, which would be predicted to produce a test duration of approximately 10 min. The most widely used cycle ergometer GXT protocols have a warm-up period of approximately 4 min. The work rate during the warm-up period is typically unloaded cycling or a light work rate such as 15 W. Immediately after 4 min of warm-up, the work rate is incremented by x W each minute until the subject reaches his or her limit of tolerance; x is the increment size that is predicted to produce a test duration somewhere between 8 and 12 min from the time when the work rate increments begin. These cycle ergometer protocols can be used to test the entire spectrum of subjects, from elite athletes to patients with cardiopulmonary disease. Only the increment size needs to be adjusted. Regarding the pedal frequency during cycle ergometry GXT, Hermansen and Saltin (1969) found that 60 rpm gave higher V02max values than did 50, 70, or 80 rpm. Hence, the pedal frequency for cycle ergometry GXT should be 60 rpm. The Balke test (Balke & Ware, 1959) is a widely used treadmill GXTprotocol. It is basically a treadmill walking test; the speed is 3.3 mph (5.3 kph) until a grade of 25% is reached. Thereafter, the grade is constant and the speed is increased 0.2 mph (0.32 kph) each minute. The grade is 0% for the first minute. The grade is raised to 2% at the end of the first minute and increased 1% per minute thereafter until it reaches 25%. For low-fit subjects, this protocol elicits valid V02max values. For fitter subjects, however, the test duration is very long. Also, these subjects are required to walk at grades above 20% during the last few minutes of the test and, according to McArdle et al. (1973), complain of severe local discomfort in the lower back and calf muscles, which may limit their ability to achieve maximal work rates. McArdle et al. (1973) compared V02max values measured using the Balke treadmill test to those measured using a continuous running treadmill test in reasonably fit male subjects. They found that the Balke test yielded V02max values that were approximately 4% lower than those obtained on the continuous running test. The Balke test could be modified to have a faster walking speed, which would likely result in increased V02max values for fitter subjects. Maksud and Coutts (1971) provide a typical running treadmill protocol. Ii begins with the subject running at 6 mph (9.7 kph) on the level (0% grade) ~It:t~i\he 1-2.3 continuous test and 67.3 min for the discontinuous test. Both the subjects and the investigators preferred the continuous test to the discontinuous test because of the long test duration of the discontinuous test. Over the years, cardiologists have designed a number of clinical treadmill protocols to test patients with heart disease. Two of these protocols are those of Bruce (Bruce, Kusumi, &. Hosmer, 1973) and Ellestad (1986)-see figures 2.2 and 2.3, respectively. The Bruce protocol is the most widely used clinical treadmill protocol. for 2 min. Thereafter, the grade is increased by 2.5% each 2 min; the speed is constant throughout the test. Each of the protocols just described is continuous. Some GXT protocols are discontinuous. An example is the treadmill test of Mitchell, Sproule, and Chapman (1958). The test begins with the subject walking at 3 mph (4.8 kph) for 10 min at 10% grade. The subject then rests for 10 min. Next, the subject runs for 2.5 min at 6 mph (9.7 kph) up a 2.5% grade. After another 10 min rest period, the subject runs for 2.5 min at 6 mph (9.7 kph) up a 5.0% grade. This procedure (rest followed by a 2.5% grade increase) continues until the subject reaches his or her limit of tolerance. The V02max obtained using a continuous protocol is the same as that obtained using a discontinuous protocol. McArdle et al. (1973) compared the continuous Maksud and Coutts treadmill protocol with the discontinuous Mitchell et al. (1973) treadmill protocol in 15 college-aged males. The mean V02max values for the continuous and discontinuous protocols were similar, 4.109 and 4.145 Lvmirr ', respectively. What differed markedly was the test duration: Criteria . for Achievement of_V.Q_2-ma~ .__.. The most widely accepted criterion for the achievement of V02max during GXT is a plateau in V02 as the work rate continues to increase. Typically, however, less than 50% of subjects tested demonstrate a plateau. Cumming and Borysyk (1972) administered GXT to 65 men aged 40 to 65 years. Only 43% of 24 22 1 20 I 18 ~ ~ CD "0 5.5 14.2 14 I 12 cO •.... (J 6.0 15.0 16 10 I 3.4 mph 2.5 1.7 8 6 4 2 a Figure 2.2 Adapted, a 3 6 9 12 Time (min) 15 18 21 Bruce treadmill protocol for patients with heart disease. by permission, from the American Williams, and Wilkins), 20. College of Sports Medicine, _ 1986, ACSM's guidelines for exercise testing and prescription, 3"' ed. (Lippincott, . ----------------------- 15 6.0 II 7.0 8.0 --~ ~ Q) I "0 10 ~ 1.7 mph 3_0 4.0 5.0 CD 5 O+-------r----+----+-----~r_--_r----r_--_r~ o 3 5 7 10 14 16 12 Time (min) Figure 2.3 Ellestad treadmill protocol for patients with heart disease. Adapted, by permission, from the American College of Spons Medicine, 1986, ACSM's guidelines for exercise testing and prescription, Williams, and Wilkins), 20. 3'd ed. (Lippincott, these men met the plateau requirement. Freedson et al. (1986) found that less than 40% of 301 adults undergoing GXT demonstrated a plateau. Cumming and Friesen (1967); Cunningham, Van Waterschoot, Paterson, Lefcoe, and Sangal (1977); and Astrand (1952) found that less than 50% of young boys who underwent GXT demonstrated a plateau. Indeed, Noakes (1988) has pointed out that the original investigators who developed the plateau criterion failed to find a true plateau in V02 as the work rate continued to increase. Three other criteria often used to defend the achievement of V02max are (a) blood lactate concentration in the first 5 min of recovery >8 mrnol/L, (b) respiratory exchange ratio at test termination >1.00, and (c) heart rate at test termination >90% of age-predicted maximum (220 - age). The third criterion is the least rigorous because of the wellknown large variation in maximal heart rate at any given age. It has been suggested that if a subject fails to demonstrate a plateau in V02 as the work rate continues t.o increase, he or she then reached a V02peak, not a teau criterion for achievement of V02max (Noakes, 1988), however, many investigators accept that the subject has achieved V02max if he or she meets the plateau criterion or two of the three secondary criteria. For patients, especially those with heart disease, a special term has been developed, namely, symptomlimited V02max_ This quantity is simply the highest V02 measured before the patient had to stop exercising because of symptoms like severe angina. Thoden (1991) has developed a novel solution to identifying the "true" V02maxvalue_ He uses a protocol with athletes that has two phases (see figure 2.4) preceded by a 5 min warm-up. The progressive phase is a GXTthat is initiated at about 30% of the athlete's predicted V02max value and progresses at about 10 to 15% of that value for each work rate increment. The duration of each work rate increment is 2 min. The typical duration of the progressive phase is 8 to 12 min. Following the progressive phase, the athlete recovers at low exercise intensity until his or her heart rate returns to about 100 beats -mirr ', which typically takes 5 to 15 min. At this point, the verification phase V02ffiax. Given the concerns.raised regarding the pla- begins. This phase is a constant-load (square-wave) .:~';~':.. __".. 1~a'lis, _ -------------- 7,.':- testat a work rate one increment higher than that achieved during the progression phase. The typical duration is 3 to 5;min. An increase of <2% compared with that obtained in the progressive phase indicates that VOzmax has been achieved. If the highest V02 during the verification phase is ~2% above that found during the progressive phase, then the verification test is repeated at the next work rate prescribed by the protocol. This procedure continues until the increase is <2%. The "true" VOzmax is taken as the highest VOz found in either the progressive phase or the verification phase. The V02 response to the progressive phase will be linear up to the maximum after a short delay at the start of the phase. For the squarewave forcing function of the verification phase, the V02 response will be exponential up to the maximum. Thoden (1991) reports excellent test-retest reliability for this protocol. For 15 subjects with VOzmax ranging from 46 to 76 ml- min-I. kg', no significant difference occurred from test to retest. The correlation coefficient was .95 between the two tests. Many factors are known to influence VOzmax. Bed rest causes it to go down, whereas endurance exercise training increases it (Saltin et al., 1968). When V02max is expressed in the units of L· min-I, large people (those with increased height or weight) typically have higher values than small people do. Women typically have lower V02max values than men do. Old adults generally have lower VOzmax values than young adults do. These factors (i.e., extent of physical activity in leisure time, height, weight, gender, and age) can be used to predict V02max with some degree of confidence. Jones, Makrides, Hitchcock Chypchar, and McCartney (1985) performed cycle ergometer GXTson 50 male and 50 female subjects of various fitness levels who ranged in age from 15 to 71 years. From the data collected, they developed the following multiple linear regression equation: 2100 H~OO- 1500- ,...-- 'c 'E • E - - 1200 OJ ==Q) ~ .::£ 0 S 900 - 600 - ~ - 300 o o I I I I I I I I I T 2 4 6 8 10 12 14 16 18 20 22 Time (min) Figure 2.4 An example of a cycleergometer protocol used by Thoden (1991) to determine "true" V0 max in athletes. The protocol is preceded by a 5 min warm-up (not shown). The protocol has three parts, namely, a progressive phase, ~ recoveryperiod, and the verification phase, The protocol shown in this figure is specificto athletes with predicted V02max valuesbetween 2 and 4 L·min-I. 2 - .....-.---.-.--------.- ..-- __ . . V02max (L'mirr ') = 0.025 (Ht) - 0.023 (Age)0.542 (Gender) + 0.019 (Wt) + 0.15 (Lei) - (8) 2.32 Lrnirr ' where Ht is standing height in em, Age is in years, and Wt is body weight in kg. For male subjects, the gender code is O. For female subjects, the gender code is l. Lei is leisure time spent per week in physical activity. The four grades of leisure activity are as follows: grade 1 for <1 h per wk, grade 2 for 1 to 3 h per wk, grade 3 for 3 to 6 h per wk, and grade 4 for >6 h per wk. An example illustrates how this equation can be used to predict V02max. Assume that a 44-year-old male subject is 183 cm tall, that he weighs 70 kg, and that he jogs 2.5 h per week (hence, his leisure activity grade is 2). Plugging these numbers into the above equation yields a predicted V02max of 2.87 L· mirr '. The multiple correlation coefficient for this equation is .892 and the standard error of estimate is 0.415 L· min'. This latter statistic can be used to compute the upper and lower 95% confidence limits of the predicted value. In the above example, the predicted V02max value of 2.87 L· mirr ' would have lower and upper 95% confidence limits of2.05 and 3.69 L· mirr ', respectively. Recently, Davis, Storer, Caiazzo, and Pham (2002) developed multiple linear regression equations that allow computation of the lower reference limit for V02max. The unique feature of these equations is that they were developed using sedentary people. Hence, if the measured V02max value of a subject falls below the lower reference limit, it cannot be argued that the low measured value was due to sedentary living. For each gender, three equations with different predictor variables were developed. Shown below are the prediction equations for men including the squared multiple correlation coefficient (R2),the standard error of estimate (SEE), and the one-sided 95% confidence interval (95% CI) for each equation. The units for the predictor variables of age, height (Ht), mass, and fat-free mass (FFM) are years, em, kg, and kg, respectively. VOimax (Lrnin") = - 0.0282 (Age) + 0.0205 (Ht) + 0.3200 L'rnin"; R2= .580; SEE = 0.398 Lrnirr"; . and 95% CI = 0.660 Lrnin' (9) V02max (Lrnin+) = - 0.0296 (Age) + 0.0167 (Mass) + 2.6538 Lrnirr '; R2= .640; SEE = 0.369 L'min'"; and 95% CI = 0.612 L'min-' (10) ..J2ir.e.cLD.e.tenninatiQ[LQtAerQblc ...P..Q.w.eL-J.I V02max (L'rnirr") = - 0.0262 (Age) + 0.0266 (FFM) + 2.1154 L'mirr"; R2= .657; SEE = 0.359 Lrnin": and 95% CI = 0.595 Lrnirr" (11) The corresponding prediction equations for women are shown below. V02max (Lrnirr") = - 0.0171 (Age) + 0.0160 (Ht) - 0.2740 Lrnirr"; R2= .500; SEE = 0.286 L'rnirr"; and 95% CI = 0.474 Lrnin? (12) V02max (Lrnirr") = - 0.0199 (Age) + 0.0135 (Mass) + 1.6267 L'mirr"; R2=.557; SEE = 0.269 L'rnirr"; 95% CI = 0.446 L'rnirr" (13) V02max (L'rnirr") = - 0.0183 (Age) + 0.0230 (FFM) + 1.3405 L'mirr"; R2 = .554; SEE = 0.270 Lmirr ': 95% CI = 0.448 Lrnirr" (14) These equations were developed on a relatively large sample (115 men and 115women) that ranged in age from 20 to 70 years. The sample 'included about 23 subjects in each age decade. The GXT. mode was the cycle ergometer. Thus, the predicted V02max values are specific to cycleergometer GXT. The equations can be used to predict V02max for treadmill GXT simply by multiplying the predicted value by 1.10 to raise that value by 10%. An example will illustrate how the lower reference limit can be calculated. The predicted V02max for a 40-year-old male who is 175 cm tall is 2.780 L· rnirr ' using the equation for men with the predictor variables of age and height. The lower 95% confidence limit (lower reference limit) is calculated as the predicted V02max minus the one-sided 95% confidence interval (0.660 L· rnirr ']. Thus, the lower reference limit for this subject is 2.120 L· mirr-' (2.780 - 0.660 L· mirr"}. £ummary If the only aerobic functionparameter of interest is V02max, then the meteorological balloon-electronic gas analyzer system is the simplest an'd, most inexpensive of the systems currently used to make the measurement. But if the user wants to measure all the