APPLICATION FOR LOW VOLTAGE DIRECTIVE On Behalf of Shenzhen New Huayi Instrument Co., Ltd Digital Clamp Meter Model No.: MS2008A, MS2008B, MS2108, MS2108A Prepared for : Shenzhen New Huayi Instrument Co., Ltd Address: West Wing, 9/F., Block 3, Saige Science and Technology Industrial Garden, Hua Qiang Bei Rd. Shenzhen Prepared By : Shenzhen Certification Technology Service Co., Ltd. Address: 2F, Building B, East Area of Nanchang Second Industrial Zone, Gushu 2nd Road, Bao'an District, Shenzhen 518126, P.R. China Date of Test: May 23-27, 2013 Date of Report: May 28-29, 2013 Report Number: STS130523255 Version Number: REV0 TEST REPORT IEC 61010-1 / EN 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements Report Reference No..................... : STS130523255 Tested by (name + signature) ..........: Jonson Cai ...................................................... Supervised by (name + signature) ...: Noman Lee ...................................................... Approved by (name + signature) ......: Jack Xu ...................................................... Date of issue.....................................: May 29, 2013 Testing Laboratory.................... : Shenzhen Certification Technology Service Co., Ltd. Address ............................................: 2F, Building B, East Area of Nanchang Second Industrial Zone, Gushu 2nd Road, Bao'an District, Shenzhen 518126, P.R. China Testing location/procedure ...............: TL Address ............................................: Same as above Applicant’s name ...................... : Shenzhen New Huayi Instrument Co., Ltd Address ............................................: West Wing, 9/F., Block 3, Saige Science and Technology Industrial Garden, Hua Qiang Bei Rd. Shenzhen RMT SMT WMT TMP Test specification: Standard ...........................................: IEC 61010-1:2010 EN 61010-1: 2010 Test procedure .................................: CE-LVD Non-standard test method................: N.A. Test Report Form No......................: IEC/EN 61010_1F TRF Originator..................................: VDE Testing and Certification Institute Master TRF.......................................: 2011-03 Copyright © 2011 Worldwide System for Conformity Testing and Certification of Electrotechnical Equipment and Components (IECEE), Geneva, Switzerland. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context. Test item description .....................: Digital Clamp Meter Trademark ........................................: HYelec Model/Type reference ........................: MS2008A, MS2008B, MS2108, MS2108A Model difference .................................: See below general information Manufacturer.......................................: Shenzhen New Huayi Instrument Co., Ltd Address...............................................: West Wing, 9/F., Block 3, Saige Science and Technology Industrial Garden, Hua Qiang Bei Rd. Shenzhen Rating(s) ...........................................: Power supply 1.5Vdc X 3 AAA battery CAT III 600 V, max. 600A TRF No. IEC61010_1F TRF originator: VDE Page 2 of 77 Report No. STS130523255 Test item particulars ..................................................: Type of item tested .................................................... : Measurement equipment Description of equipment function .............................: The clamp meter is a digital clamp AC current meter integrated with voltage measurement, resistance measurement, diode test and audible continuity. DC current measurement, inrush current , capacitance and temperature measurement are specific functions for certain model. Measurement (installation) category..........................: CAT III 600V Pollution degree............................................................: Pollution degree 2 Protection class..........................................................: Class II Environmental rating..................................................: Temperature: 0°C - 40°C Equipment mobility ....................................................: Portable or hand-held Connection to mains supply ......................................: No (battery powered) Operating conditions..................................................: Overall size of the equipment (W x D x H)................: Mass of the equipment (kg)......................................: Marked degree of protection to IEC 60529 ...............: Continuous 250mm×99mm×43mm(mian part) 0.340 N/A Accessories and detachable parts included in the N/A evaluation ..................................................................: Options ......................................................................: No Test case verdicts: Test case does not apply to the test object ................. : N(/A) Test object does meet the requirement ....................... : P(Pass) Test object does not meet the requirement ................. : F(Fail) Testing ......................................................................... : Date of receipt of test item ........................................... : May 23, 2013 Date (s) of performance of tests................................... : May 23-05, 2013 General remarks: This report shall not be reproduced, except in full, without the written approval of the issuing testing laboratory. The test results presented in this report relate only to the item(s) tested. "(see remark #)" refers to a remark appended to the report. "(see Annex #)" refers to an annex appended to the report. "(see Form A.#)" refers to a table appended to the report. Throughout this report a comma (point) is used as the decimal separator. General product information: Series models MS2008A, MS2008B, MS2108 and MS2108A are identical in most internal configuration, the model difference is electrical measurement function and range as bellow: MS2008A Measure function: DC/AC voltage, AC current, resistance, continuity, diode. MS2008B Measure function: DC/AC voltage, AC current, resistance, continuity, diode, temperature. MS2108 Measure function: DC/AC voltage, DC/AC current, resistance, capacitance, diode, continuity MS2108 Measure function: DC/AC voltage, DC/AC current, resistance, capacitance, diode, continuity, inrush current TRF No. IEC61010_1F TRF originator: VDE Page 3 of 77 Report No. STS130523255 Copy of marking plate: 1) Marking label for model MS2008A Digital Clamp Meter Model No.: MS2008A Power supply: 1.5Vdc X 3 AAA battery CAT III 600 V, max. 600A Shenzhen New Huayi Instrument Co., Ltd Made In China 2) Below warning statement and symbol were molded on appliance TRF No. IEC61010_1F TRF originator: VDE Page 4 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict P TABLE : 1 - Documents attached to this report Document No. Document description Page Numbers Appendix 2 Photo documentation 68-77 P TABLE: 2 - Test equipment list Item Manufacturer Equipment - - Model No. Calibration date 1 Last Comments Due *Note: Appendix 1 (Page 66-67) 1) or interval between calibrations. P TABLE: 3 – List of components and circuits relied on for safety Unique component reference or location Manufacturer Part number (NOTE 1) RATING (NOTE 2) (including drawing reference if required) Evidence of acceptance (NOTE 3) Enclosure, clamp enclosure, battery cover, terminal holder, rotatable probe socket, function switch knob Chi Mei Corporation PA-765A(+) V-0, 85 ℃, ABS, CTI > 400 V UL E56070 and tested with appliance PCB MEIZHOU KEJIE INTEGRATED CIRCUIT CO LTD Bayer Thai Co. Ltd. In Taiwan KJ-2 94V-0, 130 ℃ PC-110+ V-0, thickness min 1.5 CTI II Toshiba Silicone Co. Ltd. In Japan TSE221-4U HB UL ZPMV2,E255 694 UL ZPMV2 and Tested in appliance Tested in appliance Shenzhen Weilin Electronics Co. Ltd. Symbio Inc. WMZ12A 0.250A, 1.5W -- 130 o C, single layer dielectric: 3320 Vrms / 60 s Lens (PC) Function Selecting Rubber Keypad (silicone rubber) PTC Insulating tape On winding surface of jaw NOTE 1 - List all manufacturers concerned. NOTE 2 - Electrical, mechanical, flammability, etc. NOTE 3 - Licence number, file number or other documentary evidence of acceptance TRF No. IEC61010_1F Tested in appliance UL and Tested in appliance Page 5 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test 4.4 Testing in SINGLE FAULT CONDITIONS 4.4.1 Fault tests 4.4.2 Application of SINGLE FAULT CONDITIONS 4.4.2.1 SINGLE FAULT CONDITIONS not Result - Remark P (see Form A.1 and A.2) covered by 4.4.2.1 to Verdict P P (see Form A.1 and A.2) — 4.4.2.14 4.4.2.2 PROTECTIVE IMPEDANCE N 4.4.2.3 PROTECTIVE CONDUCTOR N 4.4.2.4 Equipment or parts for short-term or intermittent operation N 4.4.2.5 Motors N 4.4.2.6 Capacitors N 4.4.2.7 MAINS transformers N 4.4.2.7.2 Short circuit N 4.4.2.7.3 Overload N 4.4.2.8 Outputs N 4.4.2.9 Equipment for more than one supply N 4.4.2.10 Cooling N 4.4.2.11 Heating devices N 4.4.2.12 Insulation between circuits and parts N 4.4.2.13 Interlocks N 4.4.2.14 Voltage selectors N 4.4.3 Duration of tests (see Form A.1 and A.2) P 4.4.4 Conformity after application of fault conditions (see Form A.1; A.2; A.8, A.14) P 5 MARKING AND DOCUMENTATION P 5.1.1 General P Required equipment markings are: — visible: P From the exterior; or P After removing a cover; or No such parts used N Opening a door Ditto N After removal from a rack or panel Ditto N Not put on parts which can be removed by an operator Letter symbols (IEC 60027) used P See the copy of marking plate Graphic symbols (IEC 61010-1: Table 1) used 5.1.2 Identification P P See the copy of marking plate — Equipment is identified by: P a) Manufacturer’s or supplier’s name or trademark HYelec P TRF No. IEC61010_1F Page 6 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 5.1.3 Requirement + Test Result - Remark b) Model number, name or other means MS2008A, MS2008B, MS2108, MS2108A P Manufacturing location identified Single manufacturing location, see user’s manual. N MAINS supply Not operated by AC mains N Equipment is marked as follows: Ditto N a) Nature of supply: Battery operated N b) Verdict 1) a.c. RATED MAINS frequency or range of frequencies.................................................. : N 2) d.c. with symbol 1 N RATED supply voltage(s) or range..................... : Battery operated N c) Max. RATED power (W or VA) or input current ... : N The marked value not less than 90 % of the maximum value N If more than one voltage range: N Separate values marked; or N Values differ by less than 20 % d) OPERATOR-set (see Form A.3) N N for different RATED supply voltages: Indicates the equipment set voltage N Portable equipment indication is visible from the exterior N Changing the setting changes the indication N e) Accessory MAINS socket-outlets accepting standard MAINS plugs are marked: 5.1.4 5.1.5 No socket outlet N With the voltage if it is different from the MAINS supply voltage ................................................... : N For use only with specific equipment N If not marked for specific equipment it is marked with: N The maximum rated current or power; or N Symbol 14 with full details in the documentation N Fuses No fuse used N Operator replaceable fuse marking (see also 5.4.5) ........................................................ : N TERMINALS, connections and operating devices P Where necessary for safety, indication of purpose of TERMINALS, connectors, controls and indicators marked P If insufficient space, symbol 14 used P Push-buttons and actuators of emergency stop devices and indicators: N used only to indicate a warning of danger or TRF No. IEC61010_1F N Page 7 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark the need for urgent action N coloured red N coded as specified in IEC 60073 N Supplementary means of coding provided, if meaning of colour relates (see IEC 60073): 5.1.5.1 N to safety of persons; or N safety of the environment N TERMINALS N MAINS supply TERMINAL identified N Other TERMINAL marking: N N (symbol 5 used) a) FUNCTIONAL EARTH TERMINALS b) PROTECTIVE CONDUCTOR TERMINALS: N Symbol 6 is placed close to or on the TERMINAL; or N Part of appliance inlet N of control circuits (symbol 7 used) N d) HAZARDOUS LIVE TERMINALS supplied from the interior N c) 5.1.5.2 Verdict TERMINALS Standard MAINS socket outlet; or N RATINGS marked; or N Symbol 14 used N Measuring circuit TERMINALS P Unless clear indication that below the limits of 50 V a.c. or 120 V d.c. to earth: Required markings are adjacent to TERMINALS; OR If insufficient space: P -- On the RATING plate or scale plate; or N TERMINAL is marked with symbol 14 P a) For CAT I measurement circuits: -- RATED voltage ....................................................: N Current marked if applicable ..............................: N Symbol 14 marked N b) For CAT II, CAT III or CAT IV measurement circuits: RATED voltage ....................................................: -CAT III 600 V P Current marked if applicable ..............................: 600A P Appropriate measurement category marked (CAT II, CAT III or CAT IV); or ...........................: No marking required for: CAT III P TERMINALS other than those permanently connected and not ACCESSIBLE with appropriate information in installation manual (see 5.4.3) TRF No. IEC61010_1F N N Page 8 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test 5.1.6 For specific connection to other equipment TERMINALS only, and means for identifying provided Switches and circuit breakers 5.1.7 Result - Remark Verdict N No switches and circuit breakers used N If disconnecting device, off position clearly marked N If push-button used as power supply switch: N Symbol 9 and 15 used for on-position N Symbol 10 and 16 used for off-position N Pair of symbols 9, 15 and 10, 16 close together N Equipment protected by DOUBLE INSULATION or P REINFORCED INSULATION Protected throughout (symbol 11 used) See the copy of marking plate Only partially protected (symbol 11 not used) 5.1.8 5.2 5.3 Field-wiring TERMINAL boxes P N No terminal boxes used N If TERMINAL or ENCLOSURE exceeds 60 °C: N Cable temperature RATING marked ....................... : N Marking visible before and during connection or beside TERMINAL N Warning markings P Visible when ready for NORMAL USE P Are near or on applicable parts P Symbols and text correct dimensions and colour: — a) symbols min 2,75 mm and text 1,5 mm high and contrasting in colour with background P b) symbols and text moulded, stamped or engraved in material min. 2,0 mm high and N 0.5 mm depth or raised if not contrasting in colour N If necessary marked with symbol 14 N Statement to isolate or disconnect if access by using a tool to HAZARDOUS LIVE parts is permitted N Durability of markings P The required markings remain clear and legible in (see Form A.4) P NORMAL USE 5.4 Documentation 5.4.1 General P Equipment is accompanied by documentation for Provided in user’s manual. safety purposes for OPERATOR or RESPONSIBLE BODY P Safety documentation for service personnel authorized by the manufacturer N Documentation necessary for safe operation is provided in printed media or P in electronic media if available at any time N TRF No. IEC61010_1F Page 9 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test — a) intended use P b) technical specification P c) name and address of manufacturer or supplier P See 5.4.2 to 5.4.5 P e) information to mitigate residual RISK (see also subclause 17) P f) accessories for safe operation of the equipment specified N g) guidance provided to check correct function of the equipment, if incorrect reading may cause a HAZARD from harmful or corrosive substances of HAZARDOUS live parts P h) instructions for lifting and carrying N Warning statements and a clear explanation of warning symbols: — Provided in the documentation; or P Information is marked on the equipment N Equipment ratings P Documentation includes: P a) Supply voltage or voltage range .......................: Powered by: 3 pieces of 1.5 V AAA battery. Frequency or frequency range....................... : DC P N Power or current rating................................... : N b) Description of all input and output connections in accordance to 6.6.1 a) N RATING of insulation of external circuits in accordance to 6.6.1 b) P d) Statement of the range of environmental conditions (see 1.4) P e) Degree of protection (IEC 60529) N f) if impact rating less than 5 J: N IK code in accordance to IEC 62262 marked or N symbol 14 of table 1 marked, with N c) RATED 5.4.3 Verdict Documentation includes: d) Information specified in 5.4.2 to 5.4.6 5.4.2 Result - Remark energy level and test method stated Equipment installation N Hand-held equipment N Documentation includes instructions for: N a) assembly, location and mounting requirements N b) protective earthing N c) connections to supply N d) permanently connected equipment: N 1) Supply wiring requirements TRF No. IEC61010_1F N Page 10 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark 2) If external switch or circuit-breaker, requirements and location recommendation N e) ventilation requirements N f) N special services (e. g. air, cooling liquid) N g) Instructions relating to sound level 5.4.4 Equipment operation See below. P Instructions for use include: P a) identification and description of operating controls P b) positioning for disconnection N c) instructions for interconnection N d) specification of intermittent operation limits N e) explanation of symbols used P f) P replacement of consumable materials g) cleaning and decontamination Use soft dry cloth without any solvents or water. P N h) Listing of any poisonous or injurious gases and quantities 5.4.5 Verdict reduction procedures relating to flammable liquids (see 9.5) N reduction procedures relating burn from surfaces permitted to exceed limits of 10.1 N i) RISK j) RISK Additional precautions for IEC 60950 conforming equipment in regard to moistures and liquids N A statement about protection impairment if used in a manner not specified by the manufacturer N Equipment maintenance P Instructions for RESPONSIBLE BODY include: — Instructions sufficient in detail permitting safe maintenance and inspection and continued safety: P Instruction against the use of detachable MAINS supply cord with inadequate rating N Specific battery type of user replaceable batteries Battery type 6F22 mentioned in the user manual. P Any manufacturer specified parts Test leads P Rating and characteristics of fuses N Instructions include following subjects permitting safe servicing and continued safety: N a) product specific RISKS may affect service personnel N b) protective measures for these RISKS N c) verification of the safe state after repair N TRF No. IEC61010_1F Page 11 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark 5.4.6 Integration into systems or effects resulting from special conditions N Aspects described in documentation N (see Form A.5) Verdict 6 PROTECTION AGAINST ELECTRIC SHOCK P 6.1 General P 6.1.1 Requirements — Protection against electric shock maintained in and SINGLE FAULT CONDITION P NORMAL CONDITION ACCESSIBLE parts not HAZARDOUS LIVE All accessible parts are not hazards live Voltage, current, charge or energy below the limits in and in SINGLE FAULT CONDITION between: P — NORMAL CONDITION ACCESSIBLE 6.1.2 parts and earth P two ACCESSIBLE parts on same piece of the equipment within a distance of 1,8 m N Conformity is checked by the determination of 6.2 and 6.3 followed by the tests of 6.4 to 6.11 N Exceptions N Following HAZARDOUS LIVE parts may be accessible to an OPERATOR: N a) parts of lamps and lamp sockets after lamp removal N b) parts to be replaced by operator only by the use of tool and warning marking N Those parts not HAZARDOUS LIVE 10 s after interruption of supply (see Forms A.6 ) N Capacitance test if charge is received from internal capacitor (see Forms A.6 and A.7) N 6.2 Determination of accessible parts (see Form A.6) P 6.2.1 General P Unless obviously determination of accessible parts as specified in 6.2.2 to 6.2.4 P Examination P - with jointed test finger (as specified B.2) N - with rigid test finger (as specified B.1) and a force of 10 N P Openings above parts that are HAZARDOUS LIVE N - test pin with length of 100 mm and 4 mm in diameter applied N Openings for pre-set controls N - test pin with length of 100 mm and 4 mm in diameter applied N 6.2.2 6.2.3 6.2.4 TRF No. IEC61010_1F Page 12 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test 6.3 Limit values for ACCESSIBLE parts 6.3.1 Levels in NORMAL CONDITION 6.3.2 Result - Remark Verdict P (see Form A.7) P a) Voltage limits less than 33 V r.m.s. and 46,7 V peak or 70 V d.c. P for wet locations voltage limits less than 16 V r.m.s. and 22,6 V peak or 35 V d.c. N Voltages are not HAZARDOUS LIVE the levels of: — b) Current less than 0,5 mA r.m.s. for sinusoidal, 0,7 mA peak non sinusoidal or mixed frequencies or 2 mA d.c. when measured with measuring circuit A.1 or A.2 if less than 100 Hz N for wet locations measuring circuit A.4 used N or N c) Levels of capacitive charge or energy less: N 1) 45 µC for voltages up to 15 kV peak or d.c. or line A of Figure 3 N 2) 350 mJ stored energy for voltages above 15 kV peak or d.c. N Levels in SINGLE FAULT CONDITION (see Form A.7) P a) Voltage limits less than 33 V r.m.s. and 46,7 V peak or 70 V d.c. P for wet locations voltage limits less than 16 V r.m.s. and 22,6 V peak or 35 V d.c. N Voltages are not HAZARDOUS LIVE the levels of: — b) Current less than 0,5 mA r.m.s. for sinusoidal, 0,7 mA peak non sinusoidal or mixed frequencies or 2 mA d.c. when measured with measuring circuit A.1 or A.2 if less than 100 Hz N for wet locations measuring circuit A.4 used N or N c) Levels of capacitive charge or energy less: N 1) 45 µC for voltages up to 15 kV peak or d.c. or line A of Figure 3 N 2) 350 mJ stored energy for voltages above 15 kV peak or d.c. N 6.4 Primary means of protection P 6.4.1 ACCESSIBLE parts prevented from being HAZARDOUS LIVE by one or more of following means: P a) ENCLOSURES or PROTECTIVE BARRIERS b) BASIC INSULATION P (see 6.4.2) N (see 6.4.3) N c) Impedance (see 6.4.4) 6.4.2 ENCLOSURES or PROTECTIVE BARRIERS - meet rigidity requirements of 8.1 TRF No. IEC61010_1F (see Form A.13) P P Page 13 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict - meet requirements for BASIC INSULATION, if protection is provided by insulation N - meet requirements of 6.7 for CREEPAGE and CLEARANCES between ACCESSIBLE parts and HAZARDOUS live parts, if protection is provided N by limited access 6.4.3 BASIC INSULATION (see Form A.13) - meet CLEARANCE, CREEPAGE DISTANCE and solid insulation requirements of 6.7 6.4.4 Impedance N (see Form A.12) Impedance used as primary means of protection meets all of following requirements: a) limits current or voltage to level of 6.3.2 b) RATED for maximum WORKING VOLTAGE and the amount of power it will dissipate c) CLEARANCE, CREEPAGE DISTANCE between terminations of the impedance meet requirements of BASIC INSULATION of 6.7 N N — (see Form A.7) N N (see Form A.13) N 6.5 Additional means of protection in case of SINGLE FAULT CONDITION P 6.5.1 ACCESSIBLE parts are prevented from becoming HAZARDOUS live by the primary means of protection P and supplemented by one of: N (see 6.5.2) a) PROTECTIVE BONDING b) SUPPLEMENTARY INSULATION P (see 6.5.3) c) automatic disconnection of the supply (see 6.5.5) N d) current- or voltage-limiting device (see 6.5.6) N Alternatively one of the single means of protection is used: N e) REINFORCED INSULATION (see 6.5.3) P f) PROTECTIVE IMPEDANCE (see 6.5.4) N 6.5.2 PROTECTIVE BONDING (see Form A.9, A.10 and A.11) N 6.5.2.1 ACCESSIBLE conductive parts, may become HARZARDOUS LIVE in SINGLE FAULT CONDITION: N Bonded to the PROTECTIVE CONDUCTOR TERMINAL; or N Separated by conductive screen or barrier bonded to N PROTECTIVE CONDUCTOR TERMINAL 6.5.2.2 Integrity of PROTECTIVE BONDING a) PROTECTIVE BONDING consists of directly connected structural parts or discrete conductors or both; and withstands thermal and dynamic stresses N N b) Soldered connections: N Independently secured against loosening N Not used for other purposes N TRF No. IEC61010_1F Page 14 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark N c) Screw connections are secured d) PROTECTIVE BONDING not Verdict N interrupted; or N exempted as removable part carries MAINS SUPPLY INPUT connection e) Any moveable PROTECTIVE BONDING connection specifically designed, and meets 6.5.2.4 N f) N No external metal braid of cables used (not regarded as PROTECTIVE BONDING) g) IF MAINS SUPPLY PASSES THROUGH: N Means provided for passing protective conductor; N Impedance meets 6.5.2.4 N h) Protective conductors bare or insulated, if insulated, green/yellow N Exceptions: N 1) earthing braids; N 2) internal protective conductors etc.; N Green/yellow not used for other purposes N TERMINAL suitable for connection CONDUCTOR, and meets 6.5.2.3 6.5.2.3 N of a PROTECTIVE PROTECTIVE CONDUCTOR TERMINAL N a) Contact surfaces are metal N b) Appliance inlet used N c) For rewirable cords and PERMANENTLY CONNECTED EQUIPMENT, PROTECTIVE CONDUCTOR TERMINAL is close to MAINS supply TERMINALS N d) If no MAINS supply is required, any PROTECTIVE CONDUCTOR TERMINAL: N Is near terminals of circuit for which protective earthing is necessary N External if other terminals external N e) Equivalent current-carrying capacity to MAINS supply TERMINALS (see Form A.9) N If plug-in, makes first and breaks last N g) If also used for other bonding purposes, protective conductor: N f) Applied first; N Secured independently; N Unlikely to be removed by servicing N h) PROTECTIVE CONDUCTOR of measuring circuit: 1) Current RATING equivalent to measuring circuit TERMINAL; 2) PROTECTIVE BONDING: TRF No. IEC61010_1F N N N Page 15 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Not interrupted; or i) FUNCTIONAL EARTH TERMINALS allow Verdict N N independent connection j) N If a binding screw used for PROTECTIVE CONDUCTOR TERMINAL: Suitable size for bond wire N Not smaller than M 4 (No. 6) N At least 3 turns of screw engaged N Passes tightening torque test (see Form A.9) N k) Contact pressure not capable being reduced by deformation of materials 6.5.2.4 Impedance of PROTECTIVE BONDING of plugconnected equipment (see Form A.10) Impedance between PROTECTIVE CONDUCTOR TERMINAL and each ACCESSIBLE part where PROTECTIVE BONDING is specified, is: 6.5.2.5 N N — less than 0,1 Ohm; or N less than 0,2 Ohm if equipment is provided with non detachable cord N Bonding impedance of PERMANENTLY CONNECTED (see Form A.10) N (see Form A.11) N EQUIPMENT 6.5.2.6 Transformer PROTECTIVE BONDING screen Transformer provided with screen for protective bonding: 6.5.3 6.5.4 N screen bonding consists of directly connected structural parts or discrete conductors or both; and withstands thermal and dynamic stresses (see 6.5.2.2 a ) N screen bonding with soldered connection (see 6.5.2.2 b ) is: N - Independently secured against loosening N - Not used for other purposes N SUPPLEMENTARY and REINFORCED INSULATION N - meet CLEARANCE, CREEPAGE DISTANCE and solid insulation requirements of 6.7 P PROTECTIVE IMPEDANCE (see Form A.12) N Limits current or voltage to level of 6.3.1 in NORMAL and to level of 6.3.2 in SINGLE FAULT CONDITION CLEARANCE, CREEPAGE DISTANCE between terminations of the impedance meet requirements of DOUBLE or REINFORCED INSULATION of 6.7 N (see Form A.13) N The protective impedance consists of one or more of (see Table 3 and Form A.12) the following: — a) appropriate single component suitable for safety and reliability for protection, it is: N TRF No. IEC61010_1F Page 16 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict 1) RATED twice the maximum WORKING VOLTAGE N 2) resistor RATED for twice the power dissipation for maximum WORKING VOLTAGE N b) combination of components N Single electronic device not used as PROTECTIVE N IMPEDANCE 6.5.5 6.5.6 Automatic disconnection of the supply N a) RATED to disconnect the load within time specified in Figure 2 N b) RATED for the maximum load conditions of the equipment N Current- or voltage limiting devices (see Form A.12) Device complies with all of: a) RATED to limit the current or voltage to the level of 6.3.2 b) RATED N (see Form A.8) CLEARANCE, CREEPAGE DISTANCE between terminations of the impedance meet requirements of SUPPLEMENTARY INSULATION of 6.7 N N for the maximum working voltage; and RATED for the maximum operational current if applicable c) N N (see Form A.13) N 6.6 Connections to external circuits P 6.6.1 Connections do not cause ACCESSIBLE parts of the following to become HAZARDOUS LIVE in NORMAL CONDITION or SINGLE FAULT CONDITION: P P - the equipment P Protection achieved by separation of circuits; or N short circuit of separation does not cause a HAZARD P Instructions or markings for each terminal include: N a) 6.6.2 - the external circuits RATED conditions b) Required RATING of external circuit INSULATION N TERMINALS for external circuits P TERMINALS which receive a charge from an internal capacitor are not HAZARDOUS LIVE after 10 s of interrupting supply connection 6.6.3 P for TERMINAL (see Form A.7) N Circuits with terminals which are HAZARDOUS LIVE N These circuits are: N Not connected to ACCESSIBLE conductive parts; or N Connected to ACCESSIBLE conductive parts, but are not MAINS circuits and have one TERMINAL contact at earth potential N TRF No. IEC61010_1F Page 17 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 6.6.4 Requirement + Test Result - Remark Verdict No ACCESSIBLE conductive parts are HAZARDOUS LIVE N ACCESSIBLE terminals for stranded conductors N No RISK of accidental contact because: N Located or shielded N Self-evident or marked whether or not connected to ACCESSIBLE conductive parts N ACCESSIBLE TERMINALS will not work loose N 6.7 Insulation requirements 6.7.1 The nature of insulation P 6.7.1.1 Insulation between ACCESSIBLE parts or between separate circuits consist of CLEARANCES, CREEPAGE DISTANCES and solid insulation if provided as protection against a HAZARD P 6.7.1.2 CLEARANCES P Required CLEARANCES reflecting factors of 6.7.1.1 6.7.1.3 6.7.1.5 (see Form A.5) P P Equipment rated for operating altitude greater than 2000 m correction factor of Table 3 of 61010-1 applied P CREEPAGE DISTANCES P Required CLEARANCES reflecting factors of 6.7.1.1 6.7.1.4 see Form A.5 (see Form A.5) P CTI material group reflected by requirements P CTI test performed P Solid insulation N Required CLEARANCES reflecting factors of 6.7.1.1 (see Form A.5) N Requirements for insulation according to type of circuit (see Form A.5) N a) 6.7.2 MAINS circuits of OVERVOLTAGE CATEGORY II up to nominal supply voltage of 300 V N b) 6.7.3 Secondary circuits separated from circuits defined in a) by transformer N c) K.1 MAINS circuits of OVERVOLTAGE CATEGORY III and IV or OVERVOLTAGE CATEGORY II over 300 V N d) K.2 Secondary circuits separated from circuits defined in a) by transformer N e) K.3 Circuits having one or more of: N 1) maximum TRANSIENT OVERVOLTAGE is limited to known level below the level of MAINS CIRCUIT N 2) maximum TRANSIENT OVERVOLTAGE above the level of MAINS CIRCUIT N 3) WORKING VOLTAGE is the sum of more than one circuit or a mixed voltage N 4) WORKING VOLTAGE includes recurring peak voltage, may include non-sinusoidal or nonperiodic waveform N TRF No. IEC61010_1F Page 18 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark 5) WORKING VOLTAGE with a frequency above 30 kHz 6.7.2 Verdict N Insulation for MAINS CIRCUITS of OVERVOLTAGE II with a nominal supply voltage up to 300 V N CATEGORY 6.7.2.1 CLEARANCES and CREEPAGE DISTANCES (see Form A.13) N Values for MAINS CIRCUITS of table 4 are met N Coatings to achieve reduction to POLLUTION DEGREE I comply with requirements of Annex H N 6.7.2.2 Solid insulation N 6.7.2.2.1 Withstands electrical and mechanical stresses in normal use and all RATED environmental conditions of 1.4 N Equipment passed voltage tests of 6.8.3 with values of Table 5 Complies as applicable: a) 6.7.2.2.2 6.7.2.2.3 ENCLOSURE or PROTECTIVE BARRIER Clause 8 (see Form A.14) N N N b) moulded and potted parts requirements of 6.7.2.2.2 N c) inner layers of printed wiring boards requirements of 6.7.2.2.3 N d) thin-film insulation requirements of 6.7.2.2.4 N Moulded and potted parts N Conductors between same two layers are separated by at least 0,4 mm after moulding is completed N Inner insulation layers of printed wiring boards N Separated by at least 0,4 mm between same two layers N REINFORCED INSULATION have adequate electric strength; one of following methods used: N a) thickness at least 0,4 mm N b) insulation is assembled of minimum two separate layers, each RATED for test voltage of Table 5 for BASIC INSULATION N c) insulation is assembled of minimum two separate layers, where the combination is rated for test voltage of Table 5 for REINFORCED N INSULATION 6.7.2.2.4 Thin-film insulation N Conductors between same two layers are separated by applicable CLEARANCES and CREEPAGE DISTANCES N REINFORCED INSULATION have adequate electric strength; one of following methods used: N a) thickness at least 0,4 mm N TRF No. IEC61010_1F Page 19 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict N b) insulation is assembled of min two separate layers, each RATED for test voltage of Table 5 for BASIC INSULATION (see Form A.14) c) insulation is assembled of min three separate layers, where the combination of two layers passed voltage tests of 6.8.3 with values of Table 5 for REINFORCED INSULATION N 6.7.3 Insulation for secondary circuits derived from MAINS of OVERVOLTAGE CATEGORY II up to 300 V N 6.7.3.1 Secondary circuits where separation from MAINS CIRCUITS is achieved by a transformer providing: — - REINFORCED INSULATION N - DOUBLE INSULATION N - screen connected to the PROTECTIVE CONDUCTOR N TERMINAL 6.7.3.2 CLEARANCES N a) meet the values of Table 6 for BASIC INSULATION and SUPPLEMENTARY INSULATION; or N N twice the values of Table 6 for REINFORCED INSULATION or — b) pass the voltage tests of 6.8 with values of Table 6; with following adjustments: (see Form A.14) N 1) values for REINFORCED INSULATION are 1,6 times the values for BASIC INSULATION N 2) if operating altitude is greater than 2000 m values of CLEARANCES multiplied with factor of Table 3 N 3) minimum CLEARANCE is 0,2 mm for POLLUTION 2 and 0,8 mm for POLLUTION DEGREE 3 N DEGREE 6.7.3.3 CREEPAGE DISTANCES N Based on WORKING VOLTAGE meets the values of Table 7 for BASIC and SUPPLEMENTARY INSULATION N Values for REINFORCED INSULATION are twice the values of BASIC INSULATION N Coatings to achieve reduction to POLLUTION DEGREE I comply with requirements of Annex H N 6.7.3.4 Solid insulation N 6.7.3.4.1 Withstands electrical and mechanical stresses in normal use and all RATED environmental conditions of 1.4 N a) Equipment passed voltage test of 6.8.3.1 for 5 s with VALUES of Table 6 for BASIC and (see Form A.14) N SUPPLEMENTARY INSULATION values for REINFORCED INSULATION are 1,6 times the values of BASIC INSULATION TRF No. IEC61010_1F N Page 20 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict b) if WORKING VOLTAGE exceeds 300 V, equipment passed voltage test of 6.8.3.1 for 1 min with a test voltage of 1,5 times working voltage for BASIC or SUPPLEMENTARY INSULATION (see Form A.14) N value for REINFORCED INSULATION WORKING VOLTAGE 6.7.3.4.2 6.7.3.4.3 6.7.3.4.4 N are twice the Complies as applicable: N 1) ENCLOSURE or protective barrier Clause 8 N 2) moulded and potted parts requirements of 6.7.3.4.2 N 3) inner layers of printed wiring boards requirements of 6.7.3.4.3 N 4) thin-film insulation requirements of 6.7.3.4.4 N Moulded and potted parts N Conductors between same two layers are separated by applicable distances of Table 8 N Inner insulation layers of printed wiring boards N Separated by at least by applicable distances of Table 8 between same two layers N REINFORCED INSULATION have adequate electric strength; one of following methods used: N a) thickness at least applicable distance of Table 8 N b) insulation is assembled of minimum two separate layers, each RATED for test voltage of Table 6 for BASIC INSULATION N c) insulation is assembled of min two separate layers, where the combination is rated for 1,6 times the test voltage of Table 6 N Thin-film insulation N Conductors between same two layers are separated by applicable CLEARANCES and CREEPAGE DISTANCES N REINFORCED INSULATION have adequate electric strength; one of following methods used: N a) thickness at least applicable distance of Table 8 N b) insulation is assembled of min two separate layers, each RATED for test voltage of Table 6 for N BASIC INSULATION c) insulation is assembled of min three separate layers, where the combination of two layers passed voltage tests with 1,6 time values of Table 6: 6.8 (see Form A.14) N a.c. test of 6.8.3.1; or N d.c. test of 6.8.3.2 for circuits stressed only by d.c. voltages N Procedure for dielectric strength tests TRF No. IEC61010_1F (see Form A.5 and A.14) P Page 21 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test 6.9 Constructional requirements for protection against electric shock P 6.9.1 If a failure could cause a HAZARD: N a) Security of wiring connections N b) Screws securing removable covers N c) Accidental loosening P d) CREEPAGE and CLEARANCES not reduced below the values of basic insulation by loosening N Material not to be used for safety relevant insulation: P Easily damaged materials not used P Non-impregnated hydroscopic materials not used P Colour coding N Green-and-yellow insulation shall not be used except: N a) protective earth conductors; N b) protective bonding conductors; N c) potential equilization conductors; N d) functional earth conductors N 6.9.2 6.9.3 Result - Remark 6.10 Connection to MAINS supply source and connections between parts of equipment 6.10.1 MAINS supply cords N RATED for N maximum equipment current (see 5.1.3c) Battery operated, not connect to the mains Verdict N Cable complies with IEC 60227 or IEC 60245 N Heat-resistant if likely to contact hot parts N Temperature RATING (cord and inlet) ...................... : N Green/yellow used only for connection to PROTECTIVE N CONDUCTOR TERMINALS Detachable cords with IEC 60320 MAINS connectors: — Conform to IEC 60799; or N Have the current RATING of the MAINS connector N 6.10.2 Fitting of non-detachable MAINS supply cords N 6.10.2.1 Cord entry N Inlet or bushing smoothly rounded; or N Insulated cord guard protruding >5D N Cord anchorage N Protective earth conductor is the last to take the strain N a) Cord is not clamped by direct pressure from a screw N b) Knots are not used N 6.10.2.2 TRF No. IEC61010_1F Page 22 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test N d) No failure of cord insulation in anchorage with metal parts N e) Not to be loosened without a tool N N Cord replacement does not cause a HAZARD and method of strain relief is clear Push-pull and or torque test (see Form A.15) N Plugs and connectors N MAINS supply plugs, connectors etc., conform with relevant specifications N If equipment supplied at voltages below 6.3.2.a) or from a sole source: — Plugs of supply cords do not fit MAINS sockets above rated SUPPLY voltage N MAINS type plugs used only for connection to MAINS supply N Plug pins which receive a charge from an internal capacitor (see Form A.7) N Accessory MAINS socket outlets: — a) Marking if accepts a standard MAINS plug (see 5.1.3e) N b) Input has a protective earth conductor if outlet has EARTH TERMINAL CONTACT N 6.11 Disconnection from supply source 6.11.1 Disconnects all current carrying conductors 6.11.2 Exceptions 6.11.3 Requirements according to type of equipment 6.11.3.1 PERMANENTLY CONNECTED EQUIPMENT and multiphase equipment: 6.11.3.2 Verdict c) Cannot push the cord into the equipment to cause a HAZARD f) 6.10.3 Result - Remark Battery operated, not connect to the mains N N Supplied by low energy source (battery). P N Hand-held equipment N Employs switch or circuit-breaker N If switch or circuit-breaker is not part of the equipment, documentation requires: — a) Switch or circuit-breaker to be included in building installation N b) Suitable location easily reached N c) Marking as disconnecting for the equipment N Single-phase cord-connected equipment N Equipment is provided with one of the following: N a) Switch or circuit-breaker N b) Appliance coupler (disconnectable without tool) N TRF No. IEC61010_1F Page 23 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict c) Separable plug (without locking device) N Disconnecting devices N Electrically close to the SUPPLY N Switches and circuit-breakers N When used as disconnection device: — Meets IEC 60947-1 and IEC 60947-3 N Marked to indicate function ..................................... : N Not incorporated in MAINS cord N Does not interrupt PROTECTIVE EARTH CONDUCTOR N Appliance couplers and plugs N Where an appliance coupler or separable plug is used as the disconnecting device (see 6.11.3.2): N Readily identifiable and easily reached by the operator N Single-phase portable equipment cord length not more than 3 m N PROTECTIVE EARTH CONDUCTOR connected first and disconnected last N 7 PROTECTION AGAINST MECHANICAL HAZARDS P 7.1 Equipment does not cause a mechanical HAZARD in NORMAL nor in SINGLE FAULT CONDITION P Conformity is checked by 7.2 to 7.7 P 6.11.4 6.11.4.1 6.11.4.2 7.2 Sharp edges Smooth and rounded, no hazards P Easily touched parts are smooth and rounded P Do not cause injury during NORMAL USE and P Do not cause injury during SINGLE FAULT CONDITION P 7.3 Moving parts 7.3.1 HAZARDS from moving parts limited to a tolerable level with the conditions specified in 7.3.2 and 7.3.5 N RISK assessment in accordance with 7.3.3 carried out N Exceptions N Access to HAZARDOUS moving parts permitted under following circumstances: N a) obviously intended to operate on parts or materials outside of the equipment N 7.3.2 inadvertent touching of moving parts minimized by equipment design (e .g. guards or handles) b) If operator access is unavoidable outside normal use following precautions have been taken: TRF No. IEC61010_1F No hazardous moving part P N N Page 24 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 7.3.3 7.3.4 Requirement + Test Result - Remark Verdict 1) Access requires TOOL N 2 ) Statement about training in the instructions N 3 ) Warning markings on covers prohibiting access by untrained operators N or symbol 14 with full details in documentation N RISK assessment for mechanical HAZARDS to body parts N RISK is reduced to a tolerable level by protective measures as specified in Table 12 N Minimum protective measures: N A. Low level measures N B. Moderate measures N C. Stringent measures N Limitation of force and pressure (see Form A.16) N Following levels are met in normal and single fault condition: N Continuous contact pressure below 50 N / cm² with force below 150 N N Temporary force below 250 N for an area at least of 3 cm² for a maximum duration of 0,75 s N 7.3.5 Gap limitations between moving parts 7.3.5.1 Access normally allowed N If levels of 7.3.4 exceeded and body part may be inserted minimum gap as specified in Table 13 assured in NORMAL and in SINGLE FAULT CONDITION N Access normally prevented N Maximum gap as specified in Table 14 assured in and in SINGLE FAULT CONDITION N 7.3.5.2 (see Form A.16) N NORMAL 7.4 Stability Hand-held appliance N Equipment not secured to building structure is physical stable N Stability maintained after opening of drawers etc. by automatic means, or N warning marking requires the application of means N Compliance checked by following tests as applicable: — a) 10° tilt test for other than handheld equipment N b) multi-directional force test for equipment exceeds height of 1 m and mass of 25 kg N c) downward force test for floor-standing equipment N d) overload test with 4 times maximum load for castor or support that supports greatest load N e) castor or support that supports greatest load removed from equipment N TRF No. IEC61010_1F Page 25 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test 7.5 Provisions for lifting and carrying N 7.5.1 Equipment more than 18 kg : — Has means for lifting or carrying; or N Directions in documentation N Handles or grips N Handles or grips withstand four times weight N Lifting devices and supporting parts N Rated for maximum load; or N tested with four times maximum static load N Wall mounting N Mounting brackets withstand four times weight N Expelled parts N Equipment contains or limits the energy N Protection not removable without the aid of a tool N 8 RESISTANCE TO MECHANICAL STRESSES P 8.1 Equipment does not cause a HAZARD when subjected to mechanical stresses in NORMAL USE P Normal protection level is 5 J P Levels below 5 J but not less than 1 J are acceptable if all of following criteria are met: N a) lower level justified by RISK assessment of manufacturer N b) equipment installed in its intended application is not easily touched N c) only occasional access during NORMAL USE N d) IK code in accordance to IEC 62262 marked or symbol 14 used with full information in the documentation N For non-metallic ENCLOSURES rated below 2 °C ambient temperature value chosen for minimum rated temperature N Impact energies between IK values, the IK code marked for nearest lower value N Conformity is checked by performing following tests: — 1) static test of 8.2.1 P 2) impact test of 8.2.2 with 5 J except for HAND- N 7.5.2 7.5.3 7.6 7.7 Result - Remark Verdict HELD EQUIPMENT if impact energy not selected to 5 J alternate method of IEC 62262 used N 3) drop test of 8.3.1 or 8.3.2 except for FIXED and EQUIPMENT with mass over 100 kg P TRF No. IEC61010_1F Page 26 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict Equipment rated with an impact rating of IK 08 that obviously meets the criteria N After the tests inspection with following results: — - HAZARDOUS LIVE ACCESSIBLE P parts above the limits of 6.3.2 not - insulation pass the voltage tests of 6.8 (see Form A.24) P i) no leaks of corrosive and harmful substances P ii) ENCLOSURE shows HAZARD no cracks resulting in a P iii) CLEARANCES not less than their permitted values P iv) insulation of internal wiring remains undamaged P not damaged or loosened N v) PROTECTIVE BARRIERS vi) No moving parts exposed, except permitted by 7.3 N vii) no damage which could cause spread of fire P 8.2 ENCLOSURE rigidity test P 8.2.1 Static test P - 30 N with 12 mm rod to each part of ENCLOSURE P - in case of doubt test conducted at maximum RATED ambient temperature N Impact test P Impact applied to any part of ENCLOSURE causing a if damaged P Impact energy level and corresponding IK code .... : P Non-metallic ENCLOSURES cooled to minimum RATED ambient temperature if below 2 °C N 8.3 Drop test P 8.3.1 Other than HAND-HELD and DIRECT-PLUG-IN EQUIPMENT N Tests conducted with a drop height or angle of...... : N HAND-HELD and DIRECT-PLUG-IN EQUIPMENT P Non-metallic ENCLOSURES cooled to minimum RATED ambient temperature if below 2 °C P Drop test conducted with an height of 1 m P PROTECTION AGAINST THE SPREAD OF FIRE P No spread of fire in NORMAL and SINGLE FAULT P 8.2.2 HAZARD 8.3.2 9 9.1 CONDITION MAINS supplied equipment meets requirements of 9.6 additionally P Conformity is checked by minimum one or a combination of the following (see Figure 11): (see Form A.17) P a) Fault test of 4.4; or (see Form A.1 and Form A.2) P TRF No. IEC61010_1F Page 27 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 9.2 Requirement + Test Result - Remark b) Application of 9.2 (eliminating or reducing the sources of ignition); or P c) Application of 9.2 (containment of fire within the equipment) P Eliminating or reducing the sources of ignition within the equipment P a) 1) Limited-energy circuit (see 9.4); or 2) BASIC INSULATION provided for parts of different potential; or Bridging the insulation does not cause ignition Battery is considered limitedenergy circuit P (see Form A.5 and A.14) N (see Form A.2) N N b) Surface temperature of liquids and parts (see 9.5) c) No ignition in circuits designed to produce heat 9.3 9.3.1 (see Form A.2) N Containment of the fire within the equipment, should it occur N a) Energizing of the equipment is controlled by an operator held switch N b) ENCLOSURE is conform with constructional requirements of 9.3.1; and N Requirements of 9.5 are met N Constructional requirements N a) Connectors and insulating material have flammability classification V-2 or better (see Table: 3 or Form A.18) N b) Insulated wires and cables are flame retardant (VW-1 or equivalent) (see Table: 3 or Form A.18) N c) ENCLOSURE meets following requirements: (see Form A.17) N 1) Bottom and sides in arc of 5 ° (see Figure 13) to non-limited circuits (9.4) meets: N i) no openings; or N ii) perforated as specified in Table 16; or N iii) metal screen with a mesh; or N N iv) baffles as specified in Figure 12 2) Material of ENCLOSURE and any baffle or flame barrier is made of: N Metal (except magnesium); or Non-metallic materials have flammability classification V-1 or better N (see Table: 3 or Form A.18) 3) ENCLOSURE and any baffle or flame barrier have adequate rigidity 9.4 Verdict Limited-energy circuit N N (see Form A.19) P a) Potential not more than 30 r.m.s. and 42.4 V peak, or 60 V dc P b) Current limited by one of following means: P TRF No. IEC61010_1F Page 28 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test 1) Inherently or by impedance (see Table 17); or 9.5 Result - Remark the output of battery is conside red inherently limited-energy circuit Verdict P 2) Over current protective device (see Table 18); or N 3) A regulating network limits also in SINGLE FAULT CONDITION (see Table 17) N c) Is separated by at least BASIC INSULATION N Fuse or a nonadjustable electromechanical device is used N Requirements for equipment containing or using flammable liquids No flammable liquid inside. N Flammable liquids contained in or specified for use with equipment do not cause spread of fire (see Form A.20) N RISK is reduced to a tolerable level : N a) The temperature of surface or parts in contact with flammable liquids is 25 °C below fire point N b) The quantity of liquid is limited N c) Flames are contained within the equipment N N Detailed instructions for RISK-reduction provided 9.6 Overcurrent protection 9.6.1 MAINS supplied equipment protected BASIC INSULATION between MAINS parts of opposite polarity provided Apparatus not energized by mains. N N (see Form A.14) N Devices not in the protective conductor N Fuses or single-pole circuit-breakers not fitted in neutral (multi-phase) N PERMANENTLY CONNECTED EQUIPMENT N Overcurrent device: N Fitted within the equipment; or N Specified in manufacturer's instructions N Other equipment N Protection within the equipment N 10 EQUIPMENT TEMPERATURE LIMITS AND RESISTANCE TO HEAT P 10.1 Surface temperature limits for protection against burns P 9.6.2 9.6.3 Easily touched surfaces within the limits in NORMAL and in SINGLE FAULT CONDITION: - at an specified ambient temperature of 40 °C TRF No. IEC61010_1F (see Form A.21A) P P Page 29 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 10.2 Requirement + Test Verdict - for equipment rated above 40 °C ambient temperature limits not exceeded raised by the difference to 40 °C N Heated surfaces necessary for functional reasons exceeding specified values: N Are recognizable as such by appearance or function; or N Are marked with symbol 13 N Guards are not removable without tool N Temperatures of windings Limits not exceeded in: 10.3 Result - Remark N (see Form A.21B) N NORMAL CONDITION N SINGLE FAULT CONDITION N Other temperature measurements P Following measurements conducted if applicable: (see Form A.21A) P a) Value of 60 °C of field-wiring terminal box not exceeded N b) Surface of flammable liquids and parts in contact with this liquids N c) Surface of non-metallic ENCLOSURES P d) Parts made of insulating material supporting parts connected to MAINS supply N e) Terminals carrying a current more than 0,5 A N 10.4 Conduct of temperature test P 10.4.1 Tests conducted under reference test conditions and (see Form A.21A) manufacturer’s instructions P 10.4.2 Temperature measurement of heating equipment P Tests conducted in test corner 10.4.3 (see Form A.21A) Equipment intended for installation in a cabinet or wall Equipment built in as specified in installation instructions P P (see Form A.21A) P 10.5 Resistance to heat 10.5.1 Integrity of CLEARANCE and CREEPAGE DISTANCES (see Form A.13) P 10.5.2 Non-metallic ENCLOSURES (see Form A.22) P 10.5.3 P Within 10 min after treatment: — Equipment subjected to suitable stresses of 8.2 and 8.3 complying with criteria of 8.1 P Insulating material Plastic for enclosure. P a) Parts supporting parts connected to MAINS supply Battery operated, not connect to the mains N b) TERMINALS carrying a current more than 0.5 A TRF No. IEC61010_1F N Page 30 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict Examination of material data; or N in case of doubt: N 1) Ball pressure test; or N 2) Vicat softening test of ISO 306 N 11 PROTECTION AGAINST HAZARDS FROM FLUIDS P 11.1 Protection to OPERATORS and surrounding area provided by EQUIPMENT N All fluids specified by manufacturer considered N 11.2 Cleaning (see Form A.24) N 11.3 Spillage (see Form A.24) N 11.4 Overflow (see Form A.24) N 11.5 Battery electrolyte Alkaline or carbon-zinc battery used, considered as self-safe. P Battery electrolyte leakage presents no HAZARD Ditto P 11.6 Specially protected equipment (see Form A.24) N 11.7 Fluid pressure and leakage No fluids inside the appliance N 11.7.1 Maximum pressure ............................................... : (see Form A.25) N N Maximum pressure of any part does not exceed PRATED 11.7.2 Leakage and rupture at high pressure No fluids inside the appliance. N Fluid containing parts subjected to hydraulic test if: (see Form A.25) N a) product of pressure and volume > 200 kPal; and N b) pressure > 50 kPa N Parts of refrigerating systems meets pressure-related requirements of IEC 60335-24 or IEC 60335-24 N 11.7.3 Leakage from low-pressure parts 11.7.4 Overpressure safety device N Does not operate in NORMAL USE N a) Connected as close as possible to parts intended to be protected N b) Easy access for inspection, maintenance and repair N c) Adjustment only with TOOL N d) No discharge towards person N e) No HAZARD from deposit of discharged material N f) N Adequate discharge capacity No shut-off valve between overpressure safety device and protected parts TRF No. IEC61010_1F (see Form A.25) N N Page 31 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark 12 PROTECTION AGAINST RADIATION, INCLUDING No any radiation will emit from LASER SOURCES, AND AGAINST SONIC AND appliance. ULTRASONIC PRESSURE N 12.1 Equipment provides protection N 12.2 Equipment producing ionizing radiation N 12.2.1 Ionizing radiation 12.2.1.1 Equipment meets the following requirements: N a) if intended to emit radiation meets requirements of 12.2.1.2; or N (see Form A.26) 12.2.1.3 12.2.2 12.3 12.4 N N tested, classified and marked in accordance to IEC 60405 12.2.1.2 Verdict b) if only emits stray radiation meets requirements of 12.2.1.3 N Equipment intended to emit radiation N Effective dose rate of radiation measured .............. : N If dose rate exceeds 5 µSv/h marked with the following: N a) Symbol 17 (ISO 361) N b) Abbreviations of the radionuclides .................. : N c) With maximum dose at 1 m; or ....................... : N with dose rate value between 1 µSv/h and 5 µSv/h in m .......................................................... : N Equipment not intended to emit radiation N Limit for unintended stray radiation of 1 µSv/h at any easily reached point kept ........................................ : N Accelerated electrons N Compartments opened only by the use of a TOOL N Ultraviolet (UV) radiation N No unintentional HAZARDOUS escape of UV radiation: — - checked by inspection; and N - evaluation of RISK assessment documentation N Micro-wave radiation N 2 Power density does not exceed 10 W/m ............. : N 12.5 Sonic and ultrasonic pressure N 12.5.1 Sound level 12.5.2 (see Form A.27) N No HAZARDOUS sound emission N Maximum sound pressure level measured and calculated for maximum sound power level as specified in ISO 3746 or ISO 9614-1 N Instruction describes measures for protection N Ultrasonic pressure TRF No. IEC61010_1F (see Form A.27) N Page 32 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement + Test Result - Remark Verdict Equipment not intended to emit ultrasound does not exceed limit of 110 dB between 20 kHz and 100 kHz N Equipment intended to emit ultrasound: N Outside useful beam does not exceed limit of 110 dB between 20 kHz and 100 kHz N If inside useful beam above values exceeded: N Marked with Symbol 14 of Table 1 N and following information in the documentation: N a) dimensions of useful beam N b) area where ultrasonic pressure exceed 110 dB N c) maximum sound pressure inside beam area N Laser sources N Equipment meets requirements of IEC 60825-1 N 13 PROTECTION AGAINST LIBERATED GASES, EXPLOSION AND IMPLOSION P 13.1 Poisonous and injurious gases N No poisonous or injurious gases or substances liberated in NORMAL CONDITION N Attached data/test reports demonstrate conformity N 13.2 Explosion and implosion N 13.2.1 Components N Components liable to explode: — Pressure release device provided; or N Apparatus incorporates operator protection (see also 7.7) N Pressure release device: — Discharge without danger N Cannot be obstructed N 12.6 13.2.2 Batteries and battery charging Used non-rechargeable battery.(see Form A.28) N If explosion or fire HAZARD could occur: No such hazard. — Protection incorporated in the equipment; or P Instructions specify batteries with built-in protection N In case of wrong type of battery used: P No HAZARD; or P Warning by marking and within instructions N Equipment with means to charge rechargeable batteries: — Warning against the charging of non-rechargeable batteries; and N TRF No. IEC61010_1F Page 33 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 13.2.3 Requirement + Test Result - Remark Verdict Type of rechargeable battery indicated; or N Symbol 14 used N Battery compartment design N Single component failure N Polarity reversal test N Implosion of cathode ray tubes No such item in appliance N If maximum face dimensions > 160 mm................. : — Intrinsically protected and correctly mounted; or N ENCLOSURE provides protection: N If non-intrinsically protected: — Screen not removable without TOOL N If glass screen, not in contact with surface of tube N 14 COMPONENTS AND SUBASSEMBLIES P 14.1 Where safety is involved, components and subassemblies meet relevant requirements 14.2 Motors N 14.2.1 Motor temperatures N Does not present a HAZARD when stopped or prevented form starting; or 14.2.2 14.3 14.4 (see Table 3) (see Form A.21) N Protected by over-temperature or thermal protection device conform with 14.3 N Series excitation motors N Connected direct to device, if over-speeding causes a HAZARD N Overtemperature protection devices No such device in appliance N Devices operating in a SINGLE FAULT CONDITION (see Form A.29) N a) Reliable function is ensured N b) RATED to interrupt maximum current and voltage N c) Does not operate in NORMAL USE N If self-resetting device used to prevent a HAZARD, protected part requires intervention before restarting N Fuse holders No such parts used No access to HAZARDOUS LIVE parts 14.5 N MAINS voltage selecting devices N N No such parts used Accidental change not possible N N 14.6 MAINS transformers tested outside equipment (see Forms A.30 and A.31) N 14.7 Printed circuit boards Approved P TRF No. IEC61010_1F Page 34 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 14.8 Requirement + Test Result - Remark Verdict Data shows conformity with V-1 of IEC 60695-11-10 or better; or P Test shows conformity with V-1 of IEC 60695-11-10 (see Form A.18) or better N Not applicable for printed wiring boards with limitedenergy circuits (9.4) N Circuits or components used as transient overvoltage limiting devices N Test conducted between each pair of MAINS SUPPLY (see Form A.32) N TERMINALS No HAZARD resulting from rupture or overheating of the component: N - no bridging of safety relevant insulation N - no heat to other parts above the self-ignition points N 15 PROTECTION BY INTERLOCKS 15.1 Interlocks are designed to remove a HAZARD before OPERATOR exposed N 15.2 Prevention of reactivation N 15.3 Reliability N Single fault unlikely to occur; or N Cannot cause a HAZARD N 16 HAZARDS RESULTING FROM APPLICATION N 16.1 Reasonably foreseeable misuse N No hazards arising from settings not intended and not described in the instructions N Other cases of reasonably foreseeable misuse addressed by risk assessment N Ergonomic aspects N Factors giving rise to a hazard the risk assessment is reflecting those aspects: N a) limitation of body dimensions N b) displays and indicators N c) accessibility and conventions of controls N d) arrangement of terminals N Risk assessment N Risk assessment conducted, if hazard might arise and not covered by Clauses 6 to 16 N Tolerable risk achieved by iterative documented process covering the following: N 16.2 17 TRF No. IEC61010_1F No interlock N Page 35 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause ANNEX F Requirement + Test Result - Remark Verdict a) Risk analysis N Identifies hazards and estimates risk N b) Risk evaluation N Plan to judge acceptability of resulting risk level based on the estimated severity and likelihood of a risk N c) Risk reduction N Initial risk reduced by counter measures; N Repeated risk evaluation without new risks introduced N Risks remaining after risk assessment addressed in instructions to responsible body: N Information contained how to mitigate these risks N Following principles in methods of risk reduction applied by manufacturer in given order: N 1) Risks eliminated or reduced as far as possible N 2) Protective measures taken for risks that cannot be eliminated N 3) User information about residual risk due to any defect of the protective measures N Indication of particular training is required N Specification of the need for personal protective equipment N Conformity checked by evaluation of the risk assessment documentation N ROUTINE TESTS Manufacturer ´s declaration TRF No. IEC61010_1F Carried out by manufacturer. N N Page 36 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 4.4.2 TABLE: Summary of SINGLE FAULT CONDITIONS Subclause Title Result — Remark Does not apply 4.4.2.1 SINGLE FAULT CONDITIONS not covered by 4.4.2.2 to 4.4.2.14 4.4.2.2 PROTECTIVE IMPEDANCE X 4.4.2.3 PROTECTIVE CONDUCTOR X 4.4.2.4 Equipment or parts for short-term or intermittent operation 4.4.2.5 Motors X – stopped while fully energized X – prevented from starting X – one phase interrupted (multi-phase) X 4.4.2.6 Capacitors X 4.4.2.7 MAINS transformers Attach drawing of MAINS transformers showing all protective devices (see Forms A.30 and A.31) X 4.4.2.8 Outputs X 4.4.2.9 Equipment for more than one supply X 4.4.2.10 Cooling X – air holes closed X – fans stopped X – coolant stopped X – loss of cooling liquid X Heating devices X – timer overridden X – temperature controller overridden X 4.4.2.11 Form A.1 Carried out X X 4.4.2.12 Insulation between circuits and parts 4.4.2.13 Interlocks X 4.4.2.14 Voltage selectors X X List below all SINGLE FAULT CONDITIONS not covered by 4.4.2.2 to 4.4.2.14: 13.2.2 Battery short X 13.2.2 Battery reverse X Supplementary information: X means YES (see Form A.2 for details of tests) TRF No. IEC61010_1F Verdict Comments P Page 37 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 4.4 TABLE: Testing in SINGLE FAULT CONDITION – Results Test subclause 13.2.2 Fault No. Result — Remark Fault description Verdict Form A.2 Td 4.4.3 (NOTE) P 1 Battery reverse and short 5minutes How was test terminated Comments The apparatus has no display and can't work. No hazard 16.2 2 Measuring Function select capacitance, then Measure 600 V (30kVA) 2minutes Display 00, No hazard. Yes 16.2 3 Measuring Function select Ohm, then Measure 600 V (30kVA) 2minutes Display XXXX, No hazard. Yes 16.2 4 Measuring Function select DC current, then Measure 600 V (30kVA) 2minutes Display number 796-798, No hazard. Yes 16.2 5 Measuring Function select AC current, then Measure 600 V (30kVA) 2minutes Measuring Function select AC current, then Measure 600 V (30kVA) Yes NOTE Td = Test duration in hh:mm:ss Record dielectric strength test on Form A.14 and temperature tests on Form A.21. Record in the comments column for each test whether carried out during or after SINGLE FAULT CONDITION. Supplementary information: TRF No. IEC61010_1F Meets 4.4.4 Yes Page 38 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark 5.1.3c) TABLE: MAINS supply Verdict Form A.3 V N Marked rating.........................: -- Phase ....................................: -- Frequency .............................: -- Hz — Current ..................................: -- A — Power ...................................: -- W — Power ...................................: -- VA — — Test Voltage Frequency Current Power in Power in No. V Hz A W VA 1 2 3 Note – Measurements are only required for marked ratings. Supplementary information: TRF No. IEC61010_1F — Comments Page 39 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 5.3 TABLE: Durability of markings Result — Remark Verdict Form A.4 Marking method (see NOTE) Agent 1) Adhesive label A Water 2) Ink printed B Isopropyl alcohol 70% 3) Laser marked C (specify agent) 4) Filmcoated (plastic foil control panel) D (specify agent) 5) Imprinted on plastic (moulded in) E (specify agent) NOTE – Where applicable include print method, label material, ink or paint type, fixing method, adhesive and surface to which marking is fixed. Marking location Marking method (see above) Identification (5.1.2) 1 MAINS supply (5.1.3) -- Fuses (5.1.4) -- terminals and operating devices (5.1.5.2) -- Switches and circuit breakers (5.1.6) -- Double/reinforced equipment (5.1.7) -- Field wiring Terminal boxes (5.1.8) -- Warning marking (5.2) -- Battery charging (13.2.2) -- Method 1 Test agent Remains legible Label loose Curled edges Verdict Verdict Verdict Yes / No Yes / No Yes / No A/B Supplementary information: TRF No. IEC61010_1F Comments P / F / NA P Page 40 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark 6 TABLE: Protection against electric shock - Block diagram of system Form A.5 Pollution degree ....... : 2 Verdict Overvoltage category ...................: II Location or Insulation type Maximum working description (NOTE 1) voltage (NOTE 2) CREEPAGE Distance (NOTE 3) PWB mm CTI Other mm CTI CLEARANCE (NOTE 3) Test voltage mm (NOTE 2) V LCD circuit RI (through lens) to accessible part 600V AC/DC 10.9 400 ≥ 10.9 CTI < 600 5312Vrms AC Internal circuit to knob RI 600V AC/DC 11.0 400 ≥ 11.0 CTI < 600 5312Vrms AC Jaw to accessible part RI 600V AC/DC 16.0 400 ≥ 12.5 CTI < 600 5312Vrms AC NOTE 1 – Type of insulation: BI = BASIC INSULATION DI = DOUBLE INSULATION PI = PROTECTIVE IMPEDANCE RI = Reinforced INSULATION SI = Supplementary INSULATION Supplementary Information: TRF No. IEC61010_1F P NOTE 2 - Types of voltage Peak impulse test voltage (pulse) r.m.s. d.c. peak Comments NOTE 3 - INSTALLATION CATEGORIES (OVERVOLTAGE CATEGORIES) or POLLUTION DEGREES which differ from these should be shown under "Comments". Page 41 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark Verdict 6 TABLE: Protection against electric shock - Block diagram of system Form A Battery + polarity to accessible part (through enclosure gap) RI NOTE 1 – Type of insulation: BI = BASIC INSULATION DI = DOUBLE INSULATION PI = PROTECTIVE IMPEDANCE RI = Reinforced INSULATIONSI = Supplementary INSULATION 600V AC/DC NOTE 2 - Types of voltage Peak impulse test voltage (pulse) r.m.s. d.c. peak 11.35 400 ≥ CTI < 600 11.00 P 5312Vrms AC NOTE 3 - INSTALLATION CATEGORIES (OVERVOLTAGE CATEGORIES) or POLLUTION DEGREES which differ from these should be shown under "Comments". Supplementary Information: Limit: CAT III 600V material group II Cr (RI) 10.5 mm CL(RI) 10.5 mm 6.2 TABLE: List of ACCESSIBLE parts Form A.6 P 6.1.2 Exceptions See clause 6.1.2 — 6.2 Determination of accessible parts See below — Item Description Determination method (NOTE 5) 1 Exterior enclosure Test finger, test pin Exception under 6.1.2 (NOTE 4) --- NOTE 1 – Test fingers and pins are to be applied without force unless a force is specified (see 6.2.1) NOTE 2 – Special consideration should be given to inadequate insulation and high voltage parts (see 6.2) NOTE 3 – Parts are considered to be ACCESSIBLE if they could be touched in the absence of any covering which is not considered toprovide suitable insulation (see note to paragraph 1 of 6.4). NOTE 4 – Capacitor test may be required (see Form A.7). NOTE 5 – The determination methods are: visual; rigid test finger; jointed test finger; pin 3 mm diameter; pin 4 mm diameter. Supplementary information TRF No. IEC61010_1F Page 42 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 6 TABLE: Values in NORMAL CONDITION 6.1.2 Exceptions 11.2 Cleaning and decontamination — 6.3.1 Values in NORMAL CONDITION (see NOTE 1) 11.3 Spillage — 6.6.2 Terminals for external circuit 11.4 Overflow — 6.10.3 Plugs and connections Item Result — Remark Internal 67 circuit to accessible part V peak 86 Form A.7 Current V d.c. -- Test mA circuit r.m.s. A1/A2/A3 A1 0.015 (7.6 /500) Capacitance mA peak 0.124 (67 /500) 10 s / 5 s test (NOTE) mA d.c. μC mJ V μC mJ -- -- -- -- -- -- Comments While measuring AC 600Vrms NOTE – A 10 s test is specified in 6.1.2 a) b). A 5 s test is specified in 6.10.3. The capacitance level versus voltage below the limits given from figure 3 of IEC 61010-1. Supplementary information: battery supplied and is considered limited –energy circuit TRF No. IEC61010_1F P — Voltage V r.m.s. Verdict Page 43 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 6.3.2 TABLE: Values in SINGLE FAULT CONDITION Item Subclause and (see Form A.6) fault No. (see Form A.2) Internal circuit to accessible part -- Result — Remark Form A.8 Voltage V r.m.s. 67 Verdict Transient (see NOTE) Current V peak V d.c. V s Test circuit A1/A2/A3 86 -- -- -- A1 mA r.m.s. mA peak 0.158 0.134 (7.9/5 (66.9/ 00) 500) Capacitance mA d.c. μF (see NOTE) -- -- NOTE – Transient voltages must be below the limits given from Figure 2 and the capacitance below the limits from figure 3 of IEC 61010-1. Supplementary information: battery supplied and is considered limited –energy circuit TRF No. IEC61010_1F P Comments While measuring AC 600Vrms Page 44 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 6.7 TABLE: CLEARANCES and CREEPAGE DISTANCES 6.9.101 Insulation for JAWS and JAW OPENING 8 Location (see Form A.5) Mechanical resistance to shock and impact Measured (initial – 6.7) Verdict LCD circuit (through lens) to accessible part Form A5 Internal circuit to knob Jaw to accessible part Battery + polarity to accessible part (through enclosure gap) Result — Remark CREEPAGE DISTANCE (mm) Clearance (mm) 10.9 10.9 P 11.0 11.0 16.0 11.35 Verdict P P Test at max. Mechanical tests (note) RATED Measured after test (if required) Drop (8.2) Normal Handheld/ Plug-in 30 P N/A N/A P 40℃ 10.9 10.9 P P 30 P N/A N/A P 40℃ 11.0 11.0 P 12.5 P 30 P N/A N/A P 40℃ 16.0 12.5 P 11.0 P 30 P N/A N/A P 40℃ 11.35 11.0 (6.7N) ambient (10.5.1) CAT III 600V material group IIIa Cr (RI) 12 mm CL(RI) 10.5 mm CAT III 600V material group II Cr (RI) 10.5 mm CL(RI) 10.5 mm CREEPAGE DISTANCE (mm) CLEARANCE (mm) Rigidity (8.1) Static Dynamic Applied force NOTE – Refer to Form A.14 for dielectric strength tests following the above tests. RI means reinforce insulation. TRF No. IEC61010_1F P Form A.13 Verdict Comment s Page 45 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark Form A.14 Verdict 6.8 TABLE: Dielectric strength tests P 4.4.4.1 b) Conformity after application of fault conditions1 P 6.4 Protection in NORMAL CONDITION P 6.5.2 DOUBLE INSULATION and REINFORCED INSULATION P 6.6.1 Connections to external circuits P 6.7.3.1 c) CLEARANCE values – General: reduced CLEARANCES for homogeneous construction N/A 6.10.2.5 Fitting of non-detachable MAINS SUPPLY cords1 N/A 8 Mechanical resistance to shock and impact 9.1 a) 2) Eliminating or reducing the sources of ignition within the equipment N/A 9.3 c) Limited-energy circuit N/A 11.2 Cleaning1 N/A 11.3 Spillage1 N/A 11.4 Overflow1 N/A P 11.6 Specially protected equipment1 1 Record the fault, test or treatment applied before the dielectric strength test N/A Test site altitude ................................................: 13 m — Test voltage correction factor (see Table 10)....: NO — Location or references from Forms A.2 and A.5 Probe to enclosure Clause or sub-clause 6.4 6.52 4.4.4.1b) 8 Supplementary information: TRF No. IEC61010_1F Humidity Yes/No YES Working voltage V Test voltage r.m.s./peak/d.c V MAX 600Vrms AC/DC 5312Vrms Comments Verdict RI P Page 46 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 9 Item TABLE: Protection against the spread of fire Source of HAZARD or area of the equipment considered (circuit, component, liquid etc.) 1 Plastic enclosure 2 PCB Supplementary information: TRF No. IEC61010_1F Result — Remark Verdict Form A.16 Protection Method (9a, 9b or 9c) Protection details P Verdict Flammability of V-0 P Flammability of V-0 P Page 47 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark 10. TABLE : Temperature Measurements 10.1 Surface temperature limits - NORMAL CONDITION and 10.2 Temperature of windings- NORMAL CONDITION and 10.3 Other temperature measurements Verdict Form A.21A Operating conditions: Measure AC 600V Continuously. Frequency ............... : DC Hz Test room ambient temperature (ta) .................................: Voltage.................... : 4.5 V Part / Location Test duration ...........: 25 °C 1 h tc °C 41 tmax °C 80 Verdict Enclosure(front) tm °C 26 Enclosure(rear) 26 41 80 P LCD Display 25 40 80 P 45 min Comments P NOTE 1 - tm = measured temperature tc = tm corrected (tm–ta+ 40 °C or max. RATED ambient) tmax = maximum permitted temperature NOTE 2 - see also 14.1 with reference to component operating conditions NOTE 3 - Record values for NORMAL CONDITION and / or SINGLE FAULT CONDITION in this Form use additional form if necessary NOTE 4 - see Form A.21B for details of winding temperature measurements Supplementary information: TRF No. IEC61010_1F P Page 48 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark 10.2 TABLE: Temperature of windings Resistance method Temperature Measurements 4.4.2.7 MAINS transformers 14.2.1 Motor temperatures Verdict Form A.21B N Operating conditions .... : Frequency ................... : Hz Test room ambient temperature (ta1/ta2) ................................ : Voltage........................ : V Part / Designation Rcold Ω / °C (initial / final) Test duration ........... : Rwarm Current A Ω tr K h tc °C tmax °C Verdict min Comments NOTE 1- Rcold = initial resistance Rwarm = final resistance tr = temperature rise tc = tr corrected (tc= tr - { ta2 - ta1} + [40 °C or max RATED ambient]) tmax = maximum permitted temperature NOTE 2 - Indicate insulation class (IEC 60085) under comments (optional) NOTE 3 - Record values for NORMAL CONDITION and / or SINGLE FAULT CONDITION in this Form use additional form if necessary Supplementary information: 10.5.2 TABLE: Resistance to heat of non-metallic ENCLOSURES Form A.22 Test method used....................................................: N — Non operative treatment..........................................: [ ] Empty ENCLOSURE ...................................................: [ ] Operative treatment.................................................: [ ] Temperature during tests ........................................: — ENCLOSURE samples tested were............................: — Description Material Dielectric strength test (6.8)…………………………: Comments Verdict V r.m.s./peak/d.c. NOTE – Within 10 minutes of the end of treatment suitable tests in acc. to 8.2 and 8.3 must be conducted and pass criteria of 8.1. Supplementary information: TRF No. IEC61010_1F Page 49 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause 10.5.2 Requirement — Test Result — Remark TABLE: Resistance to heat of non-metallic enclosures Test method used: Non operative treatment.........................................: Empty ENCLOSURE...................................................: Operative treatment................................................: Temperature during tests .......................................: ENCLOSURE samples tested were ...........................: Verdict Form A.21 P A — [O] P [ ] P [ ] N/A — 70°C — Description Material Comments Verdict Enclosure PA765A+ V-0 No damage P Dielectric strength test (6.8)....................................: Supplementary information: TRF No. IEC61010_1F 5312 AC V r.m.s. Page 50 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 13.2.2 TABLE: Batteries Result — Remark Verdict Form A.28 P Battery load and charging circuit diagram: Battery type ............................................................: 1.5 V AAA battery — Battery manufacturer/model/catalogue No.............: No specified — Battery ratings ........................................................: 1.5 V — Reverse polarity instalment test No hazard P Single component failures Component Battery Supplementary information: TRF No. IEC61010_1F Verdict Open circuit Short circuit P P Page 51 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark 14.3 TABLE: Overtemperature protection devices Verdict Form A.29 Reliability test Component NOTE: NSR = non-self-resetting (10 times) NR = non-resetting (1 time) SR = self-resetting (200 times) Supplementary information: TRF No. IEC61010_1F Type (NOTE) Verdict Comments N Page 52 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark Verdict 4.4.2.7 TABLE: MAINS transformer 4.4.2.7.2 Short circuit N 14.6 MAINS transformers tested outside equipment N Form A.30 N Type ..............................: — Manufacturer .................: — Test in equipment Test on bench Test repeated inside equipment (see 14.6) Optional – Insulation class (IEC 60085) of the lowest rated winding : Winding identification Type of Protector for winding (Note 1) Elapsed time Current, A primary secondary Winding temperature, °C primary (see Note 2) secondary Tissue paper / cheesecloth OK ? (Pass / Fail) Voltage tests (see Note 3) Primary to secondary ______ V ____ Primary to core ______ V ____ Secondary to secondary ______ V ____ Secondary to core ______ V ____ Verdict Note 1: Note 2: Note 3: Primary fuse Secondary fuse Overtemperature protection Impedance protection Indicate method of measurement - PF / ( ) A - SF / ( ) A - OP / ( ) °C -Z TC = with thermocouple R = resistance method If resistance method is used, record resistance in cold and warm condition in FormA.20B! Record the voltage applied and the type of voltage (r.m.s. / d.c. / peak) and for results use NB = no breakdown or B = breakdown Supplementary information: TRF No. IEC61010_1F — Page 53 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test Result — Remark Verdict 4.4.2.7 TABLE: MAINS transformer 4.4.2.7.3 Overload tests (for MAINS transformers) N 14.6 MAINS transformers tested outside equipment N Form A.31 N Type ................................ : — Manufacturer................... : — Test in equipment Test on bench Test repeated inside equipment (see 14.6) Optional – Insulation class (IEC 60085) of the lowest rated winding : Winding identification Type of Protector for winding (Note 1) Elapsed time Current, A primary secondary Winding temperature, °C primary (see Note 2) secondary Tissue paper / cheesecloth OK ? (Pass / Fail) Voltage tests (see Note 3) Primary to secondary ______ V ____ Primary to core ______ V ____ Secondary to secondary ______ V ____ Secondary to core ______ V ____ Verdict Note 1: Note 2: Note 3: Primary fuse Secondary fuse Overtemperature protection Impedance protection Indicate method of measurement - PF / ( ) A - SF / ( ) A - OP / ( ) °C -Z TC = with thermocouple R = resistance method If resistance method is used, record resistance in cold and warm condition in FormA.20B! Record the voltage applied and the type of voltage (r.m.s. / d.c. / peak) and for results use NB = no breakdown or B = breakdown Supplementary information: TRF No. IEC61010_1F — Page 54 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 14.8 TABLE: Transient overvoltage limiting devices Component / Designation Result — Remark Overvoltage MAINS voltage Test voltage Category V rms V Test room ambient temperature ….: Verdict Form A.32 tm °C tc °C tmax °C Rupture Circuit breaker Verdict Yes / No tripped Comments °C NOTE - tm = measured temperature tc = tm corrected (tm–ta+ 40 °C or max. RATED ambient) tmax = maximum permitted temperature Conformity is checked by applying 5 positive and 5 negative impulses with the applicable impulse withstand voltage, spaced up to 1 min apart, from a hybrid impulse generator (see IEC 61180-1). Supplementary information: TRF No. IEC61010_1E N Page 55 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement – Test Result — Remark Annex H TABLE: Qualification of conformal coating for Verdict Addition to Form A.xx N N protection against pollution Technical properties Manufacturer ................................................. : — Type............................................................... : — Meet requirements of ANSI / UL 746E .......... : [yes / no] Manufacturer declaration of coating material : [yes / no] Operating temperature of coating.................. : [ ] °C Comparative tracking index (CTI).................. : [ ] Insulation resistance...................................... : [ ] Ω Dielectric strength.......................................... : [ ] V UV resistance (if required)............................. : [yes / no] Flammability rating ........................................ : Preparation of the test specimens Item Test conditioning Parameter [yes / no] Td h 1 Scratch resistance Visual inspection 2 Cold 24 3 Dry heat 48 4 Rapid temp. change 5 Damp heat 6 Adhesion of coading 24 5N Visual inspection 7 Humidity 8 Insulation resistance Visual inspection NOTE Td = Test duration time Supplementary information: TRF No. IEC61010_1E 48 >= 100 Ω Samples 1 2 3 4 Verdic t 5 6 Comments Page 56 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement – Test Result — Remark 6.7.2.2.2 TABLE: Reliability of potted components Verdict Addition to Form A.14 N Temperature Cycling Test Manufacturer................................................ : Type ............................................................. : Construction................................................. : Potting compound........................................ : CREEPAGE distances measured ................... : CLEARANCES measured................................ : Thickness through insulation ....................... : Adhesive test Pass/Fail ............................... : Test temperature T °C ................................. : Cycles at U= AC 500 V Leakage current (500 V) mA Number of cycles Date 1. Cycle from to 2. Cycle from to 3. Cycle from to 4. Cycle from to 5. Cycle from to 6. Cycle from to 7. Cycle from to 8. Cycle from to 9. Cycle from to 10. Cycle from to 68 h / 125 °C 1h/ 25 °C 2h/ 0 °C 1h/ 25 °C After Cycling Test : Humidity conditioning 48 h Requirements for dielectric strength (s. insulation diagram) Basic insulation ______ V r.m.s. Additional insulation Reinforced insulation ______ V r.m.s. ______ V r.m.s. Supplementary information: TRF No. IEC61010_1E Test voltage V r.m.s Verdict Page 57 of 77 Report No. STS130523255 IEC/EN 61010-1 Clause Requirement — Test 6. TABLE: Working voltage of Switch Mode Power Supply Location / Measuring track Input supply voltage…….: Supplementary information: TRF No. IEC61010_1F Result — Remark Insulation (Form A.5) V RMS voltage V Hz Peak voltage V Verdict Addition to Form A.5 Required Measured Required Measured Verdict cl cl cp cp mm mm mm mm Comments N Page 58 of 77 Report No. STS130523255 TEST REPORT IEC 61010-2-032 / EN 61010-2-032 Safety requirements for electrical equipment for measurement, control, and laboratory use Particular requirements for hand-held and hand-manipulated current sensors for electrical test and measurement Testing Laboratory .............. : Shenzhen Certification Technology Service Co., Ltd. Address ............................................: 2F, Building B, East Area of Nanchang Second Industrial Zone, Gushu 2nd Road, Bao'an District, Shenzhen 518126, P.R. China Testing location/procedure ...............: TL Address ............................................: Same as above RMT SMT WMT TMP Applicant’s name ................. : Shenzhen New Huayi Instrument Co., Ltd Address ............................................: West Wing, 9/F., Block 3, Saige Science and Technology Industrial Garden, Hua Qiang Bei Rd. Shenzhen Test specification: Standard ...........................................: IEC 61010-2-032: 2012 EN 61010-2-032: 2012 Test procedure .................................: CE-LVD Non-standard test method................: N.A. Test Report Form No......................: IEC/EN 61010_2_032B TRF Originator..................................: Intertek Semko AB Master TRF.......................................: 2012-12 Copyright © 2011 Worldwide System for Conformity Testing and Certification of Electrotechnical Equipment and Components (IECEE), Geneva, Switzerland. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context. Test item description .....................: Digital Clamp Meter Trademark ........................................: HYelec Model/Type reference ........................: MS2008A, MS2008B, MS2108, MS2108A Model difference .................................: See below general information Manufacturer.......................................: Shenzhen New Huayi Instrument Co., Ltd Address...............................................: West Wing, 9/F., Block 3, Saige Science and Technology Industrial Garden, Hua Qiang Bei Rd. Shenzhen Rating(s) ...........................................: Power supply 1.5Vdc X 3 AAA battery CAT III 600 V Page 59 of 77 Report No. STS130523255 IEC/EN 61010-2-032 Clause Requirement + Test 5 Marking and documentation P 5.1.2 Identification P 5.1.2 aa) 1) If designed for specific equipment it is clearly indicated or If information only in documentation, marked with symbol 14 For general use N Noted as above N 2) Type A CURRENT SENSOR marked with symbol 102 3) Type B and C CURRENT SENSORS marked with symbol 101 The marking above is adjacent to any CAT marking Type A P 5.1.5.101 Result - Remark N On the unit Voltage and current RATINGS of jaws RATED circuit to earth voltage for uninsulated conductor Verdict P P 600V CAT III Nature is marked unless valid for both AC and DC P P Marking of CAT adjacent voltage 600V CAT III P Value and nature of maximum rated current Max 600 A ~ P 5.4.4 Equipment operation P 5.4.4 a) Identification of operating controls and their use P 5.4.4 b) P 5.4.4 c) Instructions for connecting to accessories and other equipment indication of suitable accessories and detachable parts Limits of intermittent operation N 5.4.4 d) Explanation of safety related symbols used P 5.4.4 e) Instructions for replacement of replaceable parts P 5.4.4 f) Instructions for cleaning and decontamination P Instructions for the application and removal of the P 5.4.4 g) 5.4.4 h) 5.4.4 i) 5.4.4 j) 5.4.4 k) 5.4.4 l) 6 P CURRENT SENSOR Instructions to de-energize or adopt safe procedures when working on hazardous live installations with type B and C CURRENT SENSORS Warning to use individual protective equipment if working in installations with ACCESSIBLE HAZARDOUS LIVE parts Instructions about the tactile indication A warning not to use FLEXIBLE CURRENT SENSOR if the contrasting inner colour is visible A warning not to use the sensor if the wear indicator in the jaw is visible A statement that if used in a manner not specified by the manufacturer, safety may be impaired Protection against electric shock N P P NOT FLEXIBLE CURRENT SENSOR N N P P Page 60 of 77 Report No. STS130523255 IEC/EN 61010-2-032 Clause Requirement + Test 6.1.2 Exceptions 6.1.2 aa) Conductive parts within a jaw opening, provided that they meet 6.9.101 6.7 CLEARANCES and CREEPAGE DISTANCES P 6.7.3 Circuits other than MAINS CIRCUITS N 6.7.3.1 General N 6.7.3.1 c) Reduction of CLEARANCES by homogenous construction not permitted N 6.9 P 6.9.101 Constructional requirements for protection against electric shock and prevention of short-circuits Insulation requirements for jaws and jaw openings 6.9.101.1 General P Tests performed after pre-treatment P Pre-treatment of the JAW OPENING P Procedure used P Protection against touching the HAZARDOUS LIVE conductor P 6.9.101.2 6.9.101.3 Type A CURRENT SENSORS have barrier or tactile indicator Cover at least 50% of the perimeter Result - Remark P See form A5 6.9.101.4 6.9.101.5 P P Have barrier P Full perimeter covered P Extend along two opposite sides CLEARANCE and CREEPAGE meet the requirements for DOUBLE or REINFORCED INSULATION Verdict P See Form A.5 and A.13 P HAND-HELD or hand-manipulated parts P Separated by DOUBLE or REINFORCED INSULATION from: - ACCESSIBLE magnetic circuit P N - HAZARDOUS LIVE conductor P - output and input circuit and their leads N Insulation of a flexible CURRENT SENSOR N Provided with wear indicator Not a flexible current sensor N DOUBLE or REINFORCED INSULATION when new N At least BASIC INSULATION when reached the wear indicator If not provided with wear indicator, DOUBLE or REINFORCED INSULATION provided after typical lifetime wear N N Page 61 of 77 Report No. STS130523255 IEC/EN 61010-2-032 Clause 6.9.101.6 6.9.101.7 Requirement + Test Result - Remark Verdict Treatment used....................................................: N Pull test for endcaps of flexible CURRENT SENSORS N Pull test for endcaps of flexible CURRENT SENSORS N Displacement measured......................................: N Repeated test, if applicable .................................: N No damage N CleARANCE and CREEPAGE not have been reduced below the limits of 6.7.4 Dielectric strength test acc. to 6.8 N Protection against short-circuits caused by JAWS and JAW OPENINGS Enclosure provides at least BASIC INSULATION in closed position Dielectric strength test acc. to 6.8 P Type A and B CURRENT SENSORS have additional protection providing BASIC INSULATION during insertion and removal Dielectric strength test acc. to 6.8 N s. Form A.5 and Form A. 13 P s. Form A.14 P s. Form A.13 P s. Form A.14 P 8 Mechanical resistance to shock and impact P 8.1.2 Dynamic test P Test on 3 samples P If RATED below 2°C, cooled before test P Dielectric strength test acc. to 6.8 without humidity preconditioning s. Form A.14 P 14 Components 14.101 Signal and measuring leads meet 61010-031 16 Test and measurement equipment P 16.101 CURRENT SENSORS with internal current transformers P Protection against hazard caused by high voltage generated during interruption N The test probes were not evaluated by Intertek No interruption -- N Page 62 of 77 Report No. STS130523255 IEC/EN 61010-2-032 Clause 6 Requirement + Test Result - Remark TABLE: Protection against electric shock - Block diagram of system Form A.5 Location or Insulation type Maximum working CREEPAGE DISTANCE (NOTE 3) description (NOTE 1) voltage (NOTE 2) PWB mm Internal metal parts of the JAWS in closed position and the outer surfaces. P Measurement category (overvoltage category) .. : CAT III 600 V Pollution degree ...... : 2 Live parts in clamp to accessible surface Verdict CTI Other mm CTI CLEARANCE (NOTE 3) mm RI 600 Vac -- -- 16.0 >400 12.5 BI 600 Vac -- -- 6.0 >400 6.0 NOTE 1 – Type of insulation: BI = BASIC INSULATION DI = DOUBLE INSULATION PI = PROTECTIVE IMPEDANCE RI = Reinforced INSULATIONSI = Supplementary INSULATION NOTE 2 - Types of voltage Peak impulse test voltage (pulse) r.m.s. d.c. peak Supplementary Information: CAT III 600V material group II Cr (RI) 10.5 mm, Cl (RI) 10.5 mm CAT III 600V material group II Cr (BI) 5.5 mm, Cl (BI) 5.5 mm Test voltage Comments (NOTE 2) V 5312 Vr.m.s 3320 Vr.m.s Clause: 6.9.101.3 Clause 6.9.101.7 NOTE 3 - INSTALLATION CATEGORIES (OVERVOLTAGE CATEGORIES) or POLLUTION DEGREES which differ from these should be shown under "Comments". Page 63 of 77 Report No. STS130523255 IEC/EN 61010-2-032 Clause Requirement + Test Result - Remark Verdict Form A.13 P 6.7 TABLE: CLEARANCES and CREEPAGE DISTANCES 6.9.101 Insulation for JAWS and JAW OPENING P 8 Mechanical resistance to shock and impact P Location (see Form A.5) Measured (initial – 6.7) Verdict Live part inside jaw and jaw surface Measured after test (if required) RATED CREEPAGE DISTANCE (mm) Clearance (mm) Applied force (6.7N) Current jaw of hazardous live conductor and hand held part Test at max. Mechanical tests (note) Rigidity Drop (8.1) Static (8.2) Dynamic Normal ambient (10.5.1) CREEPAGE CLEARANCE DISTANCE (mm) Verdict Comments (mm) Handheld/ Plug-in 16.0 12.5 10 √ N/A √ √ 16.0 12.5 P P 6.0 6.0 10 √ N/A √ √ 6.0 6.0 P P NOTE – Refer to Form A.14 for dielectric strength tests following the above tests. Page 64 of 77 Report No. STS130523255 IEC/EN 61010-2-032 Clause Requirement + Test Result - Remark Verdict 6.8 TABLE: Dielectric strength tests 4.4.4.1 b) Conformity after application of fault conditions1 P 6.9.101.3 Protection against touching the HAZARDOUS LIVE conductor P 6.9.101.4 HAND-HELD or hand-manipulated parts P 6.9.101.5 Insulation of a flexible CURRENT SENSOR N/A 6.9.101.6 Pull test for endcaps of flexible CURRENT SENSORS N/A 6.9.101.7 Protection against short-circuits caused by the JAWS and JAW OPENING P 8.1.2 Dynamic test P Form A.14 P 1 Record the fault, test or treatment applied before the dielectric strength test Test site altitude.................................................: 13 m Test voltage correction factor (see Table 10) ....: None Location or references from Form A.5 Current jaw of hazardous live conductor and hand held part Current jaw of hazardous live conductor and hand held part Current jaw of hazardous live conductor and hand held part Live part in jaw and jaw surface Clause or sub-clause Humidity Yes/No 4.4.4.1 b), 8.1.2 No Working voltage V Test voltage r.m.s./peak/d.c V Comments — — Verdict P 600VAC/DC 5312V r.m.s 6.9.101.3, 6.9.101.4 6.9.101.7 P RI 600VAC/DC Yes P 6.9.101.7 Supplementary information: 3320 Vr.m.s BI P Page 65 of 77 Report No. STS130523255 IEC/EN 61010-2-032 Clause Requirement + Test 16.1 TABLE: Current measuring circuits 16.101 Current sensors with internal current transformers Result - Remark Verdict Form A.31 N N These tests are performed with all types and models of current transformers without internal protection, and which are specified by the manufacturer for use with the equipment a) Current transformers Type/Model RATED current A Supplementary information: Test current A Interrupt Yes / No Verdict Comments No Equipment Manufacturer Page 66 of 77 Appendix 1 Equipment List Model No. Report No. STS130523255 Serial No. Calibration date Calibration due date 34970A MY44012054 2012.10.25 2013.10.24 34970A MY44011615 2012.10.25 2013.10.24 TT-K-30-1000 / 2012.10.25 2013.10.24 GDS-250 080943 2012.10.25 2013.10.24 101A-3 3711A 3711A 3711A 3711A TDS3012B 8716C 8716C YD2511 115C 45 45 45 YD2654B 7611 31446 A06BI03017 A06BI02095 A06BI03015 A06BH02122 YT204842 870806042 870806037 11-2250 96721596 7662018 8095018 6792039 54B-053 1330848 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 YD9820A 20A-1750 2012.10.25 2013.10.24 YD2650A BCSS-F-6 NK-300 YT211 088 081050 67420 840156 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2013.10.24 2013.10.24 2013.10.24 2013.10.24 CTH-608 20953 2012.10.25 2013.10.24 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 52 Data Acquisition / Agilent Switch Unit Data Acquisition / Agilent Switch Unit Thermocouple wire OMEGA Temp. & Humid. Gongwen Chamber Oven Chamber Rongfeng DC Electronic Load ARRAY DC Electronic Load ARRAY DC Electronic Load ARRAY DC Electronic Load ARRAY Oscilloscope Tektronix Digital Power Meter Qingzhi Digital Power Meter Qingzhi Ohm Meter Yang Zi Multi Meter Fluke Desktop Multi Meter Fluke Desktop Multi Meter Fluke Desktop Multi Meter Fluke Grounding Bond Meter Yang Zhi Leakage Current Meter EXTECH Insulation Resistance Yang Zhi Tester Hi-Pot Tester Yang Zhi Electronic Scale Balance Push-Pull Scale Japan Instrumentation Digital Caliper YATO Electronic ThermoShanghai Qixiang Hygrometer Goniometer Wenzhou Tumbling Barrel Zhilitong Audio Generator LWDQGS Noise Generator Ningbo Zhongce Plug Torque Tester Zhilitong Test Probe 13 Zhilitong Test Probe 41 Zhilitong Finger Nail Probe Zhilitong Test Finger Probe B Zhilitong Rigid Finger Probe Zhilitong Test Probe Zhilitong Test Probe Zhilitong Test Probe D Zhilitong Test Probe Zhilitong Test hook Zhilitong Accessibility Probe Zhilitong UL Finger Probe Zhilitong Steel Ball Zhilitong Ball Pressure Tester SINNA Ball Pressure Tester SINNA Hammer SINNA Torque Driver Kanon Timer San He Glow Wire Test Set SINNA Needle Flame Test Set SINNA DC Power Supply Manson JZC-B2 GT-1 TAG-101 DF1681 LJ-1 TP13 TP41 FN-1 TF-B RFP D4L100 TP-C TP-D FG2C TH-1 ZA-1 ULP-01 GQ-1 SN3407 SN3407 SN3406 12LTDK SH396 ZRS-2 ZY-2 SIM-9106 15032 G010308 308909 071001107 LJ010908 D3L15 D30L80 D12N30 D12J3 D12N50 60065-913 60065-915 60065-914 D12L80 W8L180T1 A1310 D5L97 G121008 08051808 08082302 08083102 08G338 T024 08091118 08091125 G360800228 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 2013.10.24 53 hardened steel pin Zhilitong SC30 R25N30 2012.10.25 2013.10.24 54 platform scale Shanghai TGT-100 526 2012.10.25 2013.10.24 58 Tape line Jiabibei JB-5R1022 262629 2012.10.25 2013.10.24 59 Test rod Zhilitong TZ-14 D40N5 2012.10.25 2013.10.24 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 Page 67 of 77 Report No. STS130523255 Equipment List No Equipment Manufacturer Model No. Serial No. Calibration date Calibration due date 60 61 62 63 Vibration tester Surge tester Digital Power Meter dust chamber Draught-proof enclosure Hammer Hammer Hammer Data Acquisition / Switch Unit Leakage Current Tester Temp. & Humid. Chamber shengshiwei Ceprei Qingzhi Gongwen SW-TF 1065A 8713B1 SC-500 20100228 0503Y01 870909080 100311 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2013.10.24 2013.10.24 2013.10.24 2013.10.24 Tengbo TB180 Q100226 2012.10.25 2013.10.24 Zhilitong Zhilitong Zhilitong CJ-3 CJ-3 CJ-3 C031026 C031027 C031028 2012.10.25 2012.10.25 2012.10.25 2013.10.24 2013.10.24 2013.10.24 Agilent 34970A US37013205 2012.10.25 2013.10.24 Simpson 228 WHTH-100040-880 7173286 2012.10.25 2013.10.24 100631 2012.10.25 2013.10.24 72 Oscillating tube damsion DMS-E01 2012.10.25 2013.10.24 73 74 spray nozzle immersion tester Lihui kunshang LH56 IPX7-1 2012.10.25 2012.10.25 2013.10.24 2013.10.24 75 Test Probe 18 Aodesaichuang AUTO-18 2012.10.25 2013.10.24 76 Test Probe 19 Aodesaichuang AUTO-19 2012.10.25 2013.10.24 Agilent 34970A MY44052414 2012.10.25 2013.10.24 64 65 66 67 68 69 70 Weihuang 2011DNSE010401 63125 SK2018M5 auto110721-1801 auto110721-1902 Agilent 34970A MY44052411 2012.10.25 2013.10.24 79 80 81 82 Data Acquisition / Switch Unit Data Acquisition / Switch Unit Digital Power Meter Desktop Multi Meter Desktop Multi Meter Oscillograph Probe Yokogawa Agilent Agilent Agilent WT210 34401A 34401A P4100 2012.10.25 2012.10.25 2012.10.25 2012.10.25 2013.10.24 2013.10.24 2013.10.24 2013.10.24 83 Hi-Pot Tester ME I RUIKE RK2672D 2012.10.25 2013.10.24 Qingzhi 8713B1 91K223105 MY44008459 MY44008472 2012010701 RK72D111130010 871109016 2012.10.25 2013.10.24 ZHAOXIN KXN-6030D KXN.PS.JPS 2012.10.25 2013.10.24 Aigu German 10DPSK 10x 356019 12234 2012.10.25 2012.10.25 2013.10.24 2013.10.24 APC AF-11010G F310120052 2012.10.25 2013.10.24 Tianya N509 / 2012.10.25 2013.10.24 Long An LA-ORT28 LA-201206001 2012.10.25 2013.10.24 KINGPO 7006-27B-1 / 2012.10.25 2013.10.24 KINGPO 7006-28A-1 / 2012.10.25 2013.10.24 KINGPO 7006-27C-1 / 2012.10.25 2013.10.24 KINGPO 7006-50-1 / 2012.10.25 2013.10.24 KINGPO 7006-51A-2 / 2012.10.25 2013.10.24 Tektronix FEILING TDS3012B FL8618 C010353 / 2012.10.25 2012.10.25 2013.10.24 2013.10.24 Agilent 34970A MY44006829 2012.10.25 2013.10.24 OMEGA Ceprei Dongguan lixiong Inventfine Instrument Co., Ltd. TT-J-30-1000 410B LX-1211 / 1207AG10 / 2012.10.25 2013.02.28 2013.02.28 2013.10.24 2014.02.27 2014.02.27 CH338 1301004 2013.02.28 2014.02.27 77 78 84 Digital Power Meter Switching Mode Power 85 Supply 86 Torque Driver 87 Magnifying glass Regulated Power 88 Supply 89 Air Pressure Gauge Step Temperature 90 Room "GO“ Gauge for E27 91 Caps "NOT GO“ Gauge for 92 E27 Caps "GO“ Gauge for 93 dimension "S1" of E27 Caps Gauge for E27 Caps for 94 testing contact making Gauge for E27 Caps for testing protection 95 against accidental contact during insertion 96 Oscilloscope 97 Single wing drop tester Data Acquisition / 98 Switch Unit 99 Thermocouple wire 100 Touch current tester 101 Cord oscillating tester Lampholder digital 102 torsion meter Page 68 of 77 Appendix 2 Photo documentation Photo 1 View: overview Model: MS2008A [ ] front [ ] rear [ ] right side [√ ] left side [ ] top [ ] bottom [ ] internal Photo 2 View: Model: MS2008A [ ] front [ ] rear [√ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Report No. STS130523255 Page 69 of 77 Photo documentation Photo 3 View: Model: MS2008A [√ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Photo 4 View: Model: MS2008A [ ] front [√ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Report No. STS130523255 Page 70 of 77 Photo documentation Photo 5 View: Model: MS2008A [ ] front [√ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Photo 6 View: Model: MS2008A [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Report No. STS130523255 Page 71 of 77 Photo documentation Photo 7 View: Model: MS2008A [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Photo 8 View: Model: MS2008A [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Report No. STS130523255 Page 72 of 77 Photo documentation Photo 9 View: Model: MS2008A [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Photo 10 View: Model: MS2008A [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Report No. STS130523255 Page 73 of 77 Photo documentation Photo 11 View: overview Model: MS2008B [ ] front [ ] rear [ ] right side [√ ] left side [ ] top [ ] bottom [ ] internal Photo 12 View: Model: MS2008B [ ] front [ ] rear [√ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Report No. STS130523255 Page 74 of 77 Photo documentation Photo 13 View: Model: MS2008B [√ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Photo 14 View: Model: MS2008B [ ] front [√ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Report No. STS130523255 Page 75 of 77 Photo documentation Photo 15 View: Model: MS2008B [ ] front [√ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ ] internal Photo 16 View: Model: MS2008B [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Report No. STS130523255 Page 76 of 77 Photo documentation Photo 17 View: Model: MS2008B [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Photo 18 View: Model: MS2008B [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Report No. STS130523255 Page 77 of 77 Photo documentation Photo 19 View: Model: MS2008B [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Photo 20 View: Model: MS2008B [ ] front [ ] rear [ ] right side [ ] left side [ ] top [ ] bottom [ √ ] internal Report No. STS130523255