pH / ION METER Application note: A04-001A A Comparison of Ion-Selective Electrode Analysis Methods Introduction The measurement of ion concentrations in solution is a requirement in many industries. The use of an ionselective electrode (ISE) is the technique of choice for many users as it is portable, relatively inexpensive and can be performed by operators with only a minimal amount of training. A number of analysis methods have been developed to quantify the analyte concentration of a solution using an ISE. This application note will investigate three of these techniques – direct potentiometry, standard addition and sample addition. Direct potentiometry is the preferred analysis method for many ISE users. The technique does not require quantitative measurements or complex mathematical calculations to be performed by the operator. It requires the user to construct a calibration curve of electrode response versus analyte concentration. The curve is used to quantify the analyte concentration of an unknown sample, directly from the electrode response of the sample. The standard addition technique calculates the analyte concentration in an unknown sample by measuring the change in an electrode’s response when a small volume of solution, with known concentration, is added to a large volume of sample. The sample addition method is used in exactly the same way as standard addition, except that a small volume of sample is added to a large volume of standard. This application note will compare the results obtained with these three analysis methods, when determining the concentration of a prepared fluoride solution. A discussion of the advantages and disadvantages of each technique will also be made using observations from this and other published studies. Methods The Measurement of Analyte Concentration by Direct Potentiometry Two standard solutions of 100ppm and 10ppm were prepared from a 1000ppm fluoride standard (part code 0250087) by diluting 25.0ml and 2.5ml respectively to 250.0ml with deionised water. 25.0ml aliquots of the 1000ppm, 100ppm and 10ppm standards were taken and 5.0ml of ionic strength adjustment buffer (ISAB - part code 0250107) were added to each to ensure that the ionic strength of each sample was consistent . A fluoride combination ISE (part code 924 305) was used to measure the response of each solution at 20˚C and a plot of electrode response vs. Log10 concentration was plotted. A 200ppm working sample solution was prepared by diluting 50.0ml of the 1000ppm standard to 250.0ml with deionised water. 5.0ml of ISAB were added to five separate 25.0ml aliquots of the working sample solution and the electrode response was recorded for each sample solution. The calibration graph was used to determine the fluoride concentration of the sample. The Measurement of Analyte Concentration by Standard Addition A 100ppm standard solution was prepared from a 1000ppm fluoride standard (part code 0250087) by diluting 25.0ml to 250.0ml with deionised water. A 25.0ml aliquot was taken from each of the 1000ppm and 100ppm standards and 5.0ml of ISAB (part code 0250107) were added. The electrode response of each solution was recorded with a fluoride combination ISE (part code 924 305) and the slope of the electrode’s response was calculated. Three 25.0ml aliquots of the 200ppm working sample, prepared previously, had 5.0ml of ISAB added and the electrode response recorded. A 3.0ml aliquot of the 1000ppm fluoride standard was added to each sample and the electrode response was recorded for a second time. The Measurement of Analyte Concentration by Sample Addition Two standard solutions of 100ppm and 10ppm were prepared from a 1000ppm fluoride standard (part code 0250087) by diluting 25.0ml and 2.5ml respectively to 250.0ml with deionised water. A 25.0ml aliquot was taken from the 100ppm and 10ppm standards. Each aliquot had 5.0ml of ISAB (part code 0250107) added. The electrode response of each solution was recorded with a fluoride combination ISE (part code 924 305) and the slope of the electrode response was calculated. Three 25.0ml aliquots of the 10ppm standard had 5.0ml of ISAB added and the electrode response of each solution was recorded. A 3.0ml aliquot of the 200ppm working sample solution, prepared previously, was added to each solution and the electrode response was recorded for a second time. jenwayhelp@bibby-scientific.com www.jenway.com Tel: 01785 810433 Results The sample concentration values in Table 2 were calculated using the following equation: Direct Potentiometry The electrode response values that were recorded for each of the standard calibration solutions used in the direct potentiometry technique are shown in Table.1. Standard Concentration (ppm) -300 -350 Standard Addition To obtain the highest level of accuracy, the standard addition technique requires the electrode response values of two standard solutions to be measured. If an estimate of the unknown sample’s concentration can be made, the concentrations of the two standard solutions should bracket this estimated value. -400 Standard Concentration (ppm) -450 y = -56.75x - 328.27 2 R =1 -550 0 0.5 1 1.5 2 2.5 LOG(10) Concentration + 328 .27 The electrode response values of the selected standard solutions and the slope of a line plot of electrode response vs. Log10 concentration are shown in Table 3. Fluoride ISE Calibration Graph -500 (mV) Stable electrode response values were obtained in less than 60 seconds for all sample and standard solutions. Electrode Response (mV) 10 -385.2 100 -441.4 1000 -498.7 Slope -56.75 2 R 1.00 Table 1: The electrode response values of the fluoride standard solutions. Electrode Response (mV) Electrode Response Conc. (ppm) = 10^ − 56 .75 3 3.5 Figure 1: A graph of the standard electrode response of the fluoride combination ISE. The slope, intercept and correlation coefficient were calculated by performing a linear regression analysis of the data in Excel. The results are displayed in Table 1 and Figure 1. The values obtained indicate that the electrode was functioning correctly. Five sample solutions had their electrode response values measured. The recorded values, mean value and standard deviation are displayed in Table 2. Sample Electrode Response (mV) Sample Conc. (ppm) 1 -460.2 211.2 2 -459.9 208.7 3 -460.1 210.4 4 -459.8 207.8 5 -459.7 207.0 Electrode Response (mV) 100 -441.4 1000 -498.7 Slope -57.3 Table 3: Electrode response and calculated slope values used in the standard addition procedure. The slope value is used in the equation to calculate an unknown sample’s analyte content. The electrode response values of the three sample solutions were measured before and after the addition of a known volume of 1000ppm standard solution. The recorded values are displayed in Table 4. Sample Electrode Response E1 (mV) Electrode Response E2 (mV) Sample Conc. (ppm) 1 -460.2 -467.7 205.4 2 -460.3 -467.7 207.9 3 -460.3 -467.8 205.4 Mean 206.2 Std.Dev. 1.4434 Table 4: Electrode response values before and after standard addition, with the calculated concentration values. 208.5 Mean 1.2677 Std.Dev. Table 2: Electrode response values of the sample solutions with the calculated concentration values. jenwayhelp@bibby-scientific.com www.jenway.com Tel: 01785 810433 The sample concentration values in Table 4 were calculated using the following equation (1): Vs Cs × ( Vu + Vs ) Cu = ^ (E -E ) m Vu 10 - ( Vu + Vs ) 2 Sample Electrode Response E1 (mV) Electrode Response E2 (mV) Sample Conc. (ppm) 1 -385.3 -410.3 207.6 2 -385.1 -410.1 206.4 3 -385.4 -410.3 205.1 1 Where: Cu = concentration of the unknown sample (ppm) Cs = concentration of the standard (1000ppm). Vs = volume of standard (3.0ml). Vu = volume of sample + ISAB (30.0ml). E1 = electrode potential before addition (mV). E2 = electrode potential after the addition (mV). m = the electrode slope (-57.3mV). Stable electrode response values were obtained in less than 60 seconds for all solutions. 206.4 Std.Dev. 1.2503 Table 6: Electrode response values before and after sample addition, with the calculated concentration values. The sample concentration values in Table 6 were calculated using the following equation (1): Vs + Vu ^ (E -E ) m Vs Cu = Cs × - × 10 Vu Vs 2 1 Where: Sample Addition To obtain the highest level of accuracy, the sample addition technique requires the electrode response values of two standard solutions to be measured. If an estimate of the unknown sample’s concentration can be made, the concentrations of the two standard solutions should bracket the range of concentrations that will be measured during the analysis. The electrode response values of the selected standard solutions and the slope of a line plot of electrode response versus Log10 concentration are shown in Table 5. Standard Concentration (ppm) Mean Electrode Response (mV) 10 -385.2 100 -441.4 Slope -56.2 Table 5: Electrode response and calculated slope values used in the sample addition procedure. The electrode response values of three 10ppm standard solutions were measured before and after the addition of a known volume of the 200ppm working sample solution. The recorded values are displayed in Table 6. Cu = concentration of the unknown sample (ppm) Cs = concentration of the standard (10ppm). Vs = volume of standard + ISAB (30.0ml). Vu = volume of sample (3.0ml). E1 = electrode potential before addition (mV). E2 = electrode potential after the addition (mV). m = the electrode slope (-56.2mV). Stable electrode response values were obtained in less than 60 seconds for all solutions. Conclusions The results from this investigation are summarised in Table 7. Technique Results (ppm) Mean (ppm) Std. Dev. Direct Potentiometry 211.2, 208.7, 210.4, 207.8, 207.0 208.5 1.2677 Standard Addition 205.4, 207.9, 205.4 206.2 1.4434 Sample Addition 207.6, 206.4, 205.1 206.4 1.2503 ANOVA (α α=0.05) F(calc) 4.1188 F(critical) 4.4590 Table 7: Results and statistics obtained from the three analysis techniques under test. jenwayhelp@bibby-scientific.com www.jenway.com Tel: 01785 810433 The data collected in this study shows that the techniques of direct potentiometry, standard addition and sample addition do not give significantly different results, when compared at the 95% confidence interval using two-way ANOVA. By demonstrating that these techniques produce statistically equivalent results, when testing a sample with a simple sample matrix, the decision on which analytical technique to use for a particular assay can usually be based on the practical advantages and disadvantages of each technique. Calibration curves can also be used repeatedly when analysts are confident that the probe in use is performing consistently, allowing preparation times to be reduced further. References (1) Rundle C. A Beginners Guide to Ion-Selective Electrode Measurements. 2000, chapter 10b. (2) Vesely J, Weiss D, Stulik K. Analysis with Ionselective Electrodes. Ellis Horwood 1978, pg 101 –103. The techniques of standard and sample addition benefit from having the electrode immersed in solution throughout the procedure. This virtually eliminates the error that is caused by a change in the reference electrode’s liquid junction potential when an electrode is moved between different samples. Eliminating this error can significantly improve the precision of an assay, particularly in samples with complex or dirty matrices (2). The standard and sample addition techniques also allow the user to determine the slope of the electrode at a point that is very close to the sample’s concentration. A standard with a concentration that lies within the range of the samples under test is prepared and tested using the same test procedure. The electrode slope is re-calculated and the new slope value is used to determine the unknown sample’s concentration. The main advantage of this fine tuning is that it allows these techniques to give reliable results in the non-linear range of the electrode. This may be a significant advantage when an analyst only has access to old or worn electrodes (1). An additional advantage of the sample addition technique is that it is ideal for applications where only small amounts of sample are available or where the sample is dirty or viscous in nature. The practical advantages of being able to perform measurements when only a small amount of sample is available are clear, as these applications would otherwise require expensive micro-electrodes (2), whilst the ability to minimise the matrix effects of dirty samples by dissoluting a small volume of sample in a large volume of standard, ensures that any interferences in these types of sample are minimised. The techniques of standard and sample addition offer the analyst a number of practical advantages when compared to direct potentiometry and while the addition techniques require accurate volumetric measurements and more complicated mathematical calculations, the practical advantages should be thoroughly considered when deciding on which method to use in new ISE applications. Direct potentiometry is still however the preferred method of analysis in many applications as it allows a wide range of sample concentrations to be analysed without the planning and preparation steps required in the standard and sample addition techniques. jenwayhelp@bibby-scientific.com www.jenway.com Tel: 01785 810433