GE Healthcare Life Sciences Fraction collector F9-C and F9-R Operating instructions Original instructions Table of Contents Table of Contents 1 2 3 4 5 6 Introduction .......................................................................................................... 5 1.1 1.2 1.3 About this manual ................................................................................................................................ Important user information ............................................................................................................. Regulatory information ...................................................................................................................... 6 7 9 Safety instructions ............................................................................................... 12 2.1 2.2 2.3 2.4 Safety precautions ............................................................................................................................... Labels ......................................................................................................................................................... Emergency procedures ...................................................................................................................... Recycling information ......................................................................................................................... 13 20 23 25 Fraction collector overview ............................................................................... 26 3.1 Fraction collector F9-C ...................................................................................................................... 3.1.1 Function .................................................................................................................................................. 3.1.2 Fraction collector F9-C illustrations ............................................................................................ 3.1.3 Cassettes, Cassette tray and racks ............................................................................................ 3.2 Fraction collector F9-R ...................................................................................................................... 3.2.1 Function .................................................................................................................................................. 3.2.2 Fraction collector F9-R illustrations ............................................................................................ 3.2.3 Tubes ....................................................................................................................................................... 27 28 29 33 37 38 39 42 Installation ............................................................................................................ 44 4.1 4.1.1 4.1.2 4.1.3 4.2 4.2.1 4.2.2 4.3 4.4 4.5 Room and site requirements ........................................................................................................... Delivery and storage ......................................................................................................................... Placement and space requirements .......................................................................................... Site requirements ............................................................................................................................... Connect to ÄKTA pure ......................................................................................................................... Power and communication ........................................................................................................... Connect tubing .................................................................................................................................... Prepare waste tubing for Fraction collector F9-C ................................................................. Set the delay volume .......................................................................................................................... Performance test .................................................................................................................................. 45 46 47 50 51 52 58 60 62 64 Operation .............................................................................................................. 65 5.1 5.2 Fraction collector F9-C ....................................................................................................................... Fraction collector F9-R ....................................................................................................................... 66 73 Maintenance ......................................................................................................... 80 6.1 6.2 81 83 Fraction collector F9-C ....................................................................................................................... Fraction collector F9-R ....................................................................................................................... Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Table of Contents 7 Reference information ........................................................................................ 84 A Declaration of Hazardous Substances (DoHS) ................................................ 85 Index ....................................................................................................................... 88 4 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 1 Introduction 1 Introduction About this chapter This chapter contains important user information, description of safety notices, regulatory information and a general description of the intended use of Fraction collector F9-C and Fraction collector F9-R. In this chapter This chapter contains the following sections: Section See page 1.1 About this manual 6 1.2 Important user information 7 1.3 Regulatory information 9 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 1 Introduction 1.1 About this manual 1.1 About this manual Purpose of this document The Operating Instructions provide you with the instructions needed to install, operate and maintain Fraction collector F9-C and Fraction collector F9-R in a safe way. Typographical conventions Software items are identified in the text by bold italic text. A colon separates menu levels, thus File:Open refers to the Open command in the File menu. Hardware items are identified in the text by bold text (e.g., Power switch). 6 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 1 Introduction 1.2 Important user information 1.2 Important user information Read this before operating Fraction collector F9-C and Fraction collector F9-R All users must read the entire Operating Instructions before installing, operating or maintaining Fraction collector F9-C and Fraction collector F9-R. Always keep the Operating Instructions at hand when operating Fraction collector F9-C and Fraction collector F9-R. Do not operate Fraction collector F9-C and Fraction collector F9-R in any other way than described in the user documentation. If you do, you may be exposed to hazards that can lead to personal injury and you may cause damage to the equipment. Intended use Fraction collector F9-C and Fraction collector F9-R are automated fraction collectors intended for the collection of fractions from ÄKTA™ pure purification runs. ÄKTA pure and included modules are intended for research use only, and shall not be used in clinical procedures, or for diagnostic purposes. Prerequisites In order to operate Fraction collector F9-C and Fraction collector F9-R safely and according to the intended purpose the following prerequisites must be met: • You should understand the concepts of liquid chromatography. • You should be familiar with the ÄKTA pure instrument and system and have read the ÄKTA pure Operating Instructions. • You must read and understand the Safety chapter of these Operating Instructions. • The system must be installed according to the instructions in Chapter 4 Installation, on page 44. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 7 1 Introduction 1.2 Important user information Safety notices This user documentation contains WARNINGS, CAUTIONS and NOTICES concerning the safe use of the product. See definitions below. Warnings WARNING WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury. It is important not to proceed until all stated conditions are met and clearly understood. Cautions CAUTION CAUTION indicates a hazardous situation which, if not avoided, could result in minor or moderate injury. It is important not to proceed until all stated conditions are met and clearly understood. Notices NOTICE NOTICE indicates instructions that must be followed to avoid damage to the product or other equipment. Notes and tips 8 Note: A note is used to indicate information that is important for trouble-free and optimal use of the product. Tip: A tip contains useful information that can improve or optimize your procedures. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 1 Introduction 1.3 Regulatory information 1.3 Regulatory information Introduction This section lists the directives and standards that are fulfilled by Fraction collector F9-C and Fraction collector F9-R. Manufacturing information The table below summarizes the required manufacturing information. For further information, see the EC Declaration of Conformity document. Requirement Content Name and address of manufacturer GE Healthcare Bio-Sciences AB, Björkgatan 30, SE 751 84 Uppsala, Sweden Place and date of declaration See EC Declaration of Conformity Identity of person authorized to sign DoC See EC Declaration of Conformity CE Conformity This product complies with the European directives listed in the table, by fulfilling the corresponding harmonized standards. The Declaration of Conformity is available on request. Directive Title 2006/42/EC Machinery Directive (MD) 2004/108/EC Electromagnetic Compatibility (EMC) Directive Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 9 1 Introduction 1.3 Regulatory information CE Marking The CE marking and the corresponding Declaration of conformity is valid for the fraction collector when it is: • used as a stand-alone unit, or • connected to other CE marked instruments, or • connected to other products recommended or described in the user documentation, and • used in the same state as it was delivered from GE Healthcare, except for alterations described in the user documentation. International standards This product fulfills the requirements of the following standards: Standard Description Notes EN ISO 12100 Safety of machinery. General principles for design. Risk assessment and risk reduction. EN ISO standard is harmonized with EU directive 2006/42/EC EN 61010-1, IEC 610101, UL 61010-1, CAN/CSA C22.2 No. 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use. No notes EN 61326-1, IEC 613261 Electrical equipment for measurement, control and laboratory use EMC requirements EN standard is harmonized with EU directive 2004/108/EC (Emission according to CISPR 11, Group 1, class A) FCC 47 CFR Part 15 10 Federal Communications Commission (FCC) title 47 of the Code of Federal Regulations (CFR), Part 15, Radio Frequency Devices Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 1 Introduction 1.3 Regulatory information FCC statement Fraction collector F9-C and Fraction collector F9-R comply with FCC 47 CFR Part 15b. Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Regulatory compliance of connected equipment Any equipment connected to Fraction collector F9-C and Fraction collector F9-R should meet the safety requirements of EN 61010-1/IEC 61010-1, or other relevant harmonized standards. Within EU, connected equipment must be CE marked. Environmental Conformity Directive Title 2011/65/EU Restriction of Hazardous Substances (RoHS) Directive 2002/96/EC Waste Electrical and Electronic Equipment (WEEE) Directive Regulation (EC) No 1907/2006 Registration, Evaluation, Authorization and restriction of CHemicals (REACH) ACPEIP Administration on the Control of Pollution Caused by Electronic Information Products, China Restriction of Hazardous Substances (RoHS) Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 11 2 Safety instructions 2 Safety instructions About this chapter This chapter describes safety precautions and emergency shutdown procedures for Fraction collector F9-C and Fraction collector F9-R. The labels on the fraction collectors and information regarding recycling are also described. In this chapter This chapter contains the following sections: Section 12 See page 2.1 Safety precautions 13 2.2 Labels 20 2.3 Emergency procedures 23 2.4 Recycling information 25 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 2 Safety instructions 2.1 Safety precautions 2.1 Safety precautions About this section The safety precautions in this section are grouped in the following categories: • General precautions, on page 13 • Flammable liquids, on page 14 • Personal protection, on page 15 • Installing and moving the fraction collector, on page 16 • Operation, on page 16 • Maintenance, on page 17 Unless stated otherwise, the precautions apply to both Fraction collector F9-C and Fraction collector F9-R. General precautions WARNING Do not operate Fraction collector F9-C and Fraction collector F9-R in any other way than described in the Operating Instructions. WARNING Operation and user maintenance of Fraction collector F9-C and Fraction collector F9-R should be performed by properly trained personnel only. WARNING Do not use any accessories not supplied or recommended by GE Healthcare. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 13 2 Safety instructions 2.1 Safety precautions WARNING Do not use Fraction collector F9-C and Fraction collector F9-R if it is not working properly, nor if it has suffered any damage, for example: • damage to the power cord or its plug • damage caused by dropping the equipment • damage caused by splashing liquid onto it Flammable liquids CAUTION Explosion hazard during fractionation of flammable liquids. Do not use a covered fraction collector (Fraction collector F9-C) to fractionate flammable liquids. When using an open fraction collector (Fraction collector F9-R), do not place the fraction collector in a closed compartment and make sure that the room ventilation meets the local requirements. Fraction collector F9-C CAUTION Fire Hazard. Do not fractionate flammable liquids using Fraction collector F9-C. When running RPC methods, or other procedures using organic solvents, collect fractions through the outlet valve or Fraction collector F9-R. CAUTION Explosion hazard if flammable liquid leaks during cleaning of the flow path. When cleaning the flow path of Fraction collector F9-C with a flammable liquid like ethanol, carefully inspect the flow path, including the waste tubing, to make sure there will be no leakage. 14 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 2 Safety instructions 2.1 Safety precautions Fraction collector F9-R CAUTION Fractionation with flammable liquids using Fraction collector F9-R. Before you start the run, make sure that the tubes in the fraction collector are of correct size according to fraction size settings. CAUTION Moving tubes with flammable liquids. Move tubes filled with flammable liquids carefully, to prevent spillages that may come in contact with an ignition source. Personal protection CAUTION Always use appropriate personal protective equipment during operation and maintenance of Fraction collector F9-C and Fraction collector F9-R. CAUTION Hazardous substances. When using hazardous chemical and biological agents, take all suitable protective measures, such as wearing protective glasses and gloves resistant to the substances used. Follow local and/or national regulations for safe operation, maintenance and decommissioning of the equipment. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 15 2 Safety instructions 2.1 Safety precautions Installing and moving the fraction collector WARNING UniNet cable. Only use UniNet cables delivered or approved by GE Healthcare. CAUTION Heavy object. Use proper lifting equipment, or use two or more persons when moving Fraction collector F9-C. All lifting and moving must be performed in accordance with local regulations. NOTICE Vents on Fraction collector F9-C. To ensure adequate ventilation, keep papers and other objects away from the vents of the fraction collector. Operation CAUTION Risk of breaking test vials. Do not use excessive force to press vials with incorrect dimensions into the fraction collector. Glass vials may break and cause injuries. CAUTION Avoid spillage and overflow. Make sure that the equipment is prepared according to the settings in the method to be run. 16 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 2 Safety instructions 2.1 Safety precautions Fraction collector F9-C WARNING Moving parts. Do not open the door of Fraction collector F9-C while the fraction collector is active. If you need to access the fraction collector, press Pause on the chromatography instrument, and make sure that no parts are moving before opening the door. CAUTION Maximum weight on the top of Fraction collector F9-C. Do not place more than 30 kg on top of Fraction collector F9-C. Fraction collector F9-R CAUTION Replacement of the fraction collector bowl in Fraction collector F9-R. Do not replace the bowl during a run. CAUTION Replacing tubes. Do not replace the tubes in the fraction collector during a run. Maintenance CAUTION Electrical shock hazard. All repairs should be done by service personnel authorized by GE Healthcare. Do not open any covers or replace parts unless specifically stated in the user documentation. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 17 2 Safety instructions 2.1 Safety precautions CAUTION Disconnect power. Always switch off power to Fraction collector F9-C and Fraction collector F9-R, before cleaning any of its components, unless stated otherwise in the user documentation. This is done by switching off the chromatographic instrument that provides Fraction collector F9-C and Fraction collector F9-R with power. CAUTION Hazardous chemicals and biological agents. Before maintenance, service and decommissioning, wash Fraction collector F9-C and Fraction collector F9-R with a neutral solution to make sure that any hazardous solvents and biological agents have been flushed out from the fraction collector. CAUTION Always use appropriate personal protective equipment when decommissioning the equipment. CAUTION Cleaning Fraction collector F9-C and Fraction collector F9-R before decommissioning. • Wipe Fraction collector F9-C and Fraction collector F9-R with a damp tissue using a cleaning agent so that no hazardous solvents or biological agents remain on the surface. Do not forget to clean the interior of Fraction collector F9-C. • Perform a system CIP using a neutral solution. Make sure that any hazardous solvents or biological agents are flushed out from the fraction collector. NOTICE Cleaning. Keep Fraction collector F9-C and Fraction collector F9R dry and clean. Wipe regularly with a soft damp tissue and, if necessary, a mild cleaning agent. Let Fraction collector F9-C and Fraction collector F9-R dry completely before use. 18 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 2 Safety instructions 2.1 Safety precautions CAUTION Explosion hazard if flammable liquid leaks during cleaning of the flow path. When cleaning the flow path of Fraction collector F9-C with a flammable liquid like ethanol, carefully inspect the flow path, including the waste tubing, to make sure there will be no leakage. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 19 2 Safety instructions 2.2 Labels 2.2 Labels About this section This section describes the safety labels that are attached to Fraction collector F9-C and Fraction collector F9-R. Labels on Fraction collector F9-C The label below is located both on the back of the fraction collector and on the top wall of its interior. Fraction collector F9-C safety label It is not allowed to fractionate flammable liquids using Fraction collector F9-C due to the risk of build-up of flammable and explosive gases. When running RPC methods, or other procedures using organic solvents, collect fractions through the Outlet valve or with Fraction collector F9-R. The safety label below is located on the inside of the door of Fraction collector F9-C. 3^=>CUaPRcX^]PcT U[P\\PQ[T[X`dXSb 20 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 2 Safety instructions 2.2 Labels Labels on Fraction collector F9-R The label is located on the back of the fraction collector. Label information The label contains information on the serial number, regulatory compliance symbols, recycling symbols, and specifications. The label information is explained in the table below. Label text Meaning Code no Fraction collector code number Serial no Fraction collector serial number Mfg Year Manufacturing year Voltage Supply voltage Max Power Max power consumption Protection class Protection class, ingress protection according to IEC 60529 Warning! Read the Operating Instruction before using the module. Do not open any covers or replace parts unless specifically stated in the Operating Instructions. The equipment complies with the applicable requirements for Australia and New Zealand Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 21 2 Safety instructions 2.2 Labels Label text Meaning The fraction collector complies with applicable European directives This symbol indicates that electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of equipment. This symbol indicates that the product might contain hazardous materials in excess of the limits established by the Chinese standard SJ/T11363-2006 Requirements for Concentration Limits for Certain Hazardous Substances in Electronic Information Products. 22 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 2 Safety instructions 2.3 Emergency procedures 2.3 Emergency procedures About this section The ÄKTA instrument supplies the Fraction collector F9-C and Fraction collector F9-R with power. This section describes how to perform an emergency shutdown of the Fraction collector F9-C and Fraction collector F9-R by shutting down the ÄKTA instrument. This section also describes the results in the event of power failure or network interruption. Emergency shutdown In an emergency situation, stop the run by either pausing the run or switching off the instrument as described below: If you want to... then... pause the run • press the Pause button on the Instrument control panel: Note: The buttons on the Instrument control panel may be locked. This is an option available in the System settings. or • click the Pause icon in UNICORN™: Result: All pumps in the instrument are stopped. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 23 2 Safety instructions 2.3 Emergency procedures If you want to... then... switch off the instrument • press the Power switch to the 0 position, or • disconnect the power cord from the wall socket. Result: The run is interrupted immediately. Note: The sample and data may be lost as a result of switching off the power. Power failure In case of a power failure or if communication is lost, the run is interrupted immediately and all moving parts in the fraction collector are stopped. 24 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 2 Safety instructions 2.4 Recycling information 2.4 Recycling information About this section This section describes the procedures for disposal and recycling of Fraction collector F9-C and Fraction collector F9-R. Decommissioning and disposal of the equipment When taking Fraction collector F9-C and Fraction collector F9-R out of service: • The equipment must be decontaminated. • The components must be separated and recycled according to national and local environmental regulations CAUTION Always use appropriate personal protective equipment when decommissioning the equipment. Disposal of electrical components Waste comprising electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of equipment. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 25 3 Fraction collector overview 3 Fraction collector overview About this chapter This chapter gives overviews of Fraction collector F9-C and Fraction collector F9-R. In this chapter This chapter contains the following sections: Section 26 See page 3.1 Fraction collector F9-C 27 3.2 Fraction collector F9-R 37 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1 Fraction collector F9-C About this section This section shows an overview of Fraction collector F9-C. Technical details are found in the ÄKTA pure System handbook. In this section This section contains the following subsections: Section See page 3.1.1 Function 28 3.1.2 Fraction collector F9-C illustrations 29 3.1.3 Cassettes, Cassette tray and racks 33 - Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 27 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.1 Function 3.1.1 Function The fraction collector collects fractions from ÄKTA pure purification runs. The fraction collector can be used for: • Fixed volume fractionation • Peak fractionation • Combined fixed volume fractionation and peak fractionation Fraction collector F9-C has the following functions for reducing sample spill during fractionation: 28 • Drop Sync • Accumulator Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.2 Fraction collector F9-C illustrations 3.1.2 Fraction collector F9-C illustrations Introduction This section provides illustrations of Fraction collector F9-C. The main features and components are indicated. Front view The illustration below shows the main parts of the exterior of Fraction collector F9-C. 1 2 3 4 Part Description 1 Fractionation indicator 5 6 Symbol indicating that fractionation is ongoing. Do not open the door while the indicator is lit. 2 Door 3 Window Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 29 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.2 Fraction collector F9-C illustrations Part Description 4 Door handle 5 Tubing connector for outlet valve tubing 6 Vents Rear view The illustration below shows the rear view of Fraction collector F9-C. 1 1 3 30 2 Part Description 1 Vents 2 UniNet-9 D-type connector (for communication and power supply) 3 Waste tube Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.2 Fraction collector F9-C illustrations Interior The illustration below shows the main parts of the inside of Fraction collector F9-C. 3 4 5 6 7 8 9 2 10 11 1 12 14 13 Part Description 1 Fractionation arm guide rail 2 Fractionation arm main rail 3 Lamp 4 Tubing guide 5 Tubing connection 6 Dispenser head 7 Fractionation arm 8 Tubing guide 9 Height exclusion bar 10 Waste funnel 11 Waste tube 12 Tray catch Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 31 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.2 Fraction collector F9-C illustrations Part Description 13 Waste groove, in case of overflow 14 Tray guides Dispenser head The illustration below shows the Dispenser head of Fraction collector F9-C. 1 2 3 4 6 32 5 Part Description 1 Dispenser head 2 Nozzle 3 Dispenser head cover 4 Accumulator (back part of Dispenser head) 5 Drop sync sensor 6 Type code reader Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.3 Cassettes, Cassette tray and racks 3.1.3 Cassettes, Cassette tray and racks Introduction Fractions can be collected in deep well plates and in tubes of different sizes. A number of cassettes, trays and racks for different tubes and deep well plates are available. The cassettes are placed on a tray with six cassette positions. The cassette type codes, tray type code and rack type codes are scanned by the type code reader to determine the type of cassettes, trays and racks used. For more information on requirements that need to be fulfilled by the tubes and deep well plates to be used in Fraction collector F9-C, see the ÄKTA pure System handbook . Available cassettes, trays and racks The following Cassettes and racks are available: • Cassette 3 ml tubes (for 40 tubes) • Cassette 8 ml tubes (for 24 tubes) • Cassette 15 ml tubes (for 15 tubes) • Cassette 50 ml tubes (for 6 tubes) • Cassette for deep well plate (24, 48, 96 wells) • Cassette tray (for six cassettes) • Rack for 50 ml tubes (for 55 tubes) • Rack for 250 ml bottles (for 18 bottles) For information on dimension requirements for tubes and deep well plates to be used in the fraction collector, see Fraction collector tubes, on page 35 and Deep well plates, on page 35 respectively. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 33 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.3 Cassettes, Cassette tray and racks Illustration of Cassettes on the Cassette tray The illustration below shows Cassettes placed on the Cassette tray. Part Description 1 Cassette tray 2 Cassette position number 3 Cassette 4 Cassette type code 5 Tubes or deep well plates placed in a Cassette Illustrations of Fraction collector F9-C tray and racks The illustrations below show the Cassette tray, the Rack for 50 ml tubes and the Rack for 250 ml bottles. The fronts of the tray and the racks are marked with the GE-logotype. In the Cassette tray, the cassette positions are marked 1 to 6. 34 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.3 Cassettes, Cassette tray and racks Cassette tray Rack for 50 ml tubes Rack for 250 ml bottles Note: The tray and racks are inserted into the fraction collector with the GE-logotype facing outwards. Note: Do not use the Cassette tray when a rack for tubes or bottles is placed in the fraction collector. Fraction collector tubes The tubes used in Fraction collector F9-C must fulfill the requirements listed in the table below. Examples of manufacturers are also listed in the table. Tube size [ml] Diameter [mm] Height [mm] Min. Max. Min. Max. Max. volume [ml] Examples of manufacturers 3 10.5 11.5 50 56 3 NUNC 8 12 13.3 96 102 8 BD Biosciences, VWR 15 16 17 114 120 15 BD Biosciences 50 28 30 110 116 50 BD Biosciences Deep well plates Requirements The deep well plates used in Fraction collector F9-C must fulfill the requirements listed in the table below. Property Specification No. of wells 24, 48, or 96 Shape of wells Square, not cylindrical Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 35 3 Fraction collector overview 3.1 Fraction collector F9-C 3.1.3 Cassettes, Cassette tray and racks Property Specification Well volume 10, 5, or 2 ml Approved deep well plates The plates listed in the table below are tested and approved by GE Healthcare to be used with Fraction collector F9-C. Plate type Manufacturer Part no. 96 deep well plate Eppendorf™ 951033405/ 0030 501.306 BD Biosciences 353966 Greiner Bio-One 780270 Porvair Sciences 219009 Seahorse Bioscience™ S30009 Whatman™ 7701-5200 Seahorse Bioscience S30004 Whatman 7701-5500 Seahorse Bioscience S30024 Whatman 7701-5102 48 deep well plate 24 deep well plate 36 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.2 Fraction collector F9-R 3.2 Fraction collector F9-R About this section This section shows an overview of Fraction collector F9-R. Technical details are found in the ÄKTA pure System handbook. In this section This section contains the following subsections: Section See page 3.2.1 Function 38 3.2.2 Fraction collector F9-R illustrations 39 3.2.3 Tubes 42 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 37 3 Fraction collector overview 3.2 Fraction collector F9-R 3.2.1 Function 3.2.1 Function The fraction collector collects fractions from ÄKTA pure purification runs. The fraction collector can be used for: • Fixed volume fractionation • Peak fractionation • Combined fixed volume fractionation and peak fractionation Fraction collector F9-R has the following function for reducing sample spill during fractionation: • 38 Drop Sync Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.2 Fraction collector F9-R 3.2.2 Fraction collector F9-R illustrations 3.2.2 Fraction collector F9-R illustrations Introduction This subsection provides illustrations of Fraction collector F9-R. The main features and components are indicated. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 39 3 Fraction collector overview 3.2 Fraction collector F9-R 3.2.2 Fraction collector F9-R illustrations Front view The illustration below shows the main parts of the Fraction collector. 3 4 2 5 6 1 7 8 40 Part Function 1 Lock knob 2 Stationary part of delivery arm 3 Delivery arm 4 Tubing connector 5 Tube sensor 6 Collection tubes 7 Tube rack 8 Base unit Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.2 Fraction collector F9-R 3.2.2 Fraction collector F9-R illustrations Connector panel The illustration below shows the main parts of the connector panel on the fraction collector. 1 22 Part Function 1 Node ID switch 2 UniNet-9 F-type connector (for communication and power supply) Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 41 3 Fraction collector overview 3.2 Fraction collector F9-R 3.2.3 Tubes 3.2.3 Tubes Introduction For Fraction collector F9-R the fractions are collected in tubes of different sizes. For more information on requirements that need to be fulfilled by the tubes to be used in Fraction collector F9-R, see the ÄKTA pure System handbook. Available tubes Tubes with the following diameter can be used with Fraction collector F9-R: • 12 mm • 18 mm • 30 mm The tubes can have a tube length between 50-180 mm. 42 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 3 Fraction collector overview 3.2 Fraction collector F9-R 3.2.3 Tubes Illustration of the Fraction collector F9-R Tube rack Each tube rack is made up of a combination of a Bowl, Tube support, Tube guide and Tube holder. For more information on the assembly of the tube rack, see Assemble the Tube rack, on page 74. For information on which Tube rack to use, see the ÄKTA pure System handbook. Part Function 1 Single cutout 2 L-shaped cutout 3 Bowl 4 Tube support 5 Tube guide 6 Tube holder Note: Note that the tube guide has both single and L-shaped cutouts, while the tube holder only has single cutouts. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 43 4 Installation 4 Installation About this chapter This chapter provides required information to enable users and service personnel to install, move and transport Fraction collector F9-C and Fraction collector F9-R. In this chapter Section 44 See page 4.1 Room and site requirements 45 4.2 Connect to ÄKTA pure 51 4.3 Prepare waste tubing for Fraction collector F9-C 60 4.4 Set the delay volume 62 4.5 Performance test 64 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.1 Room and site requirements 4.1 Room and site requirements About this section This section describes the site planning and the preparations necessary for the installation of Fraction collector F9-C and Fraction collector F9-R. The purpose is to provide planners and technical staff with the data needed to prepare the laboratory for the installation. In this section Section See page 4.1.1 Delivery and storage 46 4.1.2 Placement and space requirements 47 4.1.3 Site requirements 50 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 45 4 Installation 4.1 Room and site requirements 4.1.1 Delivery and storage 4.1.1 Delivery and storage About this section This section describes the requirements for receiving the delivery box and storing Fraction collector F9-C and Fraction collector F9-R before installation. When you receive the delivery • Record on the receiving documents if there is any apparent damage on the delivery box. Inform your GE Healthcare representative of such damage. • Move the delivery box to a protected location indoors. Storage requirements The delivery box should be stored in a protected place indoors. The following storage requirements must be fulfilled for the unopened box: 46 Parameter Allowed range Ambient temperature, storage -25°C to +60°C Relative humidity up to 90% atmospheric humidity at 40°C for 48 hrs Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.1 Room and site requirements 4.1.2 Placement and space requirements 4.1.2 Placement and space requirements About this section This section describes different options of where to place Fraction collector F9-C and Fraction collector F9-R and the space required. Location Place the fraction collector on a clean, flat and stable area that is able to support the weight of the fraction collector. The following locations are recommended for the Fraction collector F9-C and the Fraction collector F9-R: • on the bench to the left of the ÄKTA pure instrument • on a shelf below the bench Note: Longer tubing length increases back pressure and band broadening in the chromatographic process. Place the fraction collector so that the total tubing length is minimized. Space requirements The illustration below shows the space recommended for the system including Fraction collector F9-C. 63 cm 60 cm 39 cm 54 cm Min 10 cm Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 80 cm 47 4 Installation 4.1 Room and site requirements 4.1.2 Placement and space requirements The illustration below shows the space recommended for the system including Fraction collector F9-R. 63 cm 60 cm 32 cm Note: 54 cm Min 10 cm 80 cm Fraction collector F9-R requires 10 cm free space above the fractionation arm for handling of tubes. Fraction collector weight 48 Item Weight Fraction collector F9-C 21 kg Fraction collector F9-R 5 kg Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.1 Room and site requirements 4.1.2 Placement and space requirements Fraction collector dimensions The outer dimensions of Fraction collector F9-C are shown in the illustration below. 320 mm 585 mm 390 mm The outer dimensions of Fraction collector F9-R are shown in the illustration below. 250 mm 400 mm 320 mm Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 49 4 Installation 4.1 Room and site requirements 4.1.3 Site requirements 4.1.3 Site requirements About this section This section describes the site requirements for installation of Fraction collector F9-C and Fraction collector F9-R. Operating conditions The following requirements must be fulfilled: • The instrument is intended for indoor use only. • The room must have exhaust ventilation. • The instrument should not be exposed to direct sunlight. • Dust in the atmosphere should be kept to a minimum. Allowed operating conditions are specified in the table below. Parameter Allowed range Ambient temperature, operating 4°C to 35°C Ambient temperature, storage -25°C to +60°C Relative humidity, operating 20% to 95%, non condensing Altitude Maximum 2000 m Pollution degree 2 Note: 50 Do not use buffers with a freezing point close to or below the temperature in the room or in the cold room cabinet. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.2 Connect to ÄKTA pure 4.2 Connect to ÄKTA pure About this section This section contains information on how to set up tubing and a connection for power and communication between Fraction collector F9-C or Fraction collector F9-R and the ÄKTA pure instrument. It is possible to install up to two fraction collectors in the following combinations when using ÄKTA pure: • one Fraction collector F9-C and one Fraction collector F9-R, or • two Fraction collector F9-R. If two fraction collectors are to be used, the following applies to the second one: • it must be a Fraction collector F9-R, • it has to be configured as Fraction collector F9-R, 2nd, • the node ID has to be changed, as described in Check/change Node ID, on page 56. In this section Section See page 4.2.1 Power and communication 52 4.2.2 Connect tubing 58 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 51 4 Installation 4.2 Connect to ÄKTA pure 4.2.1 Power and communication 4.2.1 Power and communication Introduction This subsection describes how to set up the power and communication connection between Fraction collector F9-C or Fraction collector F9-R and the ÄKTA pure instrument. Connect fraction collector The following cables are used to connect the fraction collectors to the ÄKTA pure instrument. Use the specified cable and follow the instructions below. Fraction collector Cable Fraction collector F9-C UniNet-9 cable, D-type Fraction collector F9-R UniNet-9 cable, F-type WARNING UniNet cable. Only use UniNet cables delivered or approved by GE Healthcare. 52 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.2 Connect to ÄKTA pure 4.2.1 Power and communication Step Action 1 Switch off the power to the ÄKTA pure instrument. 2 Remove the jumper from the UniNet-9 port to be used on the rear of the ÄKTA pure instrument (port 7 or 8 for Fraction collector F9-C, port 1 to 6 for Fraction collector F9-R). Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 53 4 Installation 4.2 Connect to ÄKTA pure 4.2.1 Power and communication Step Action 3 Connect the UniNet-9 cable between the UniNet-9 ports on the back of the fraction collector and on the back of the ÄKTA pure instrument. NOTICE Do not use the Test port on the ÄKTA pure instrument. 2 Fraction collector F9-C 4 Fraction collector F9-R Make sure that all unused UniNet-9 ports on the ÄKTA pure instrument are plugged with jumpers. Software configuration When the fraction collector has been installed, the System properties for the system has to be updated in UNICORN. The system will restart automatically when the configuration has been changed and the system can be reconnected. From hereon, UNICORN refers to UNICORN 6.3 or other compatible versions of the software. The examples given in these Operating Instructions refer to UNICORN 6.3. Follow the instructions below to update the system in UNICORN. 54 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.2 Connect to ÄKTA pure 4.2.1 Power and communication Step Action 1 • In the Administration module, choose Tools:System Properties or click the System Properties icon to open the dialog. Result: The System Properties dialog is displayed. • Select a system in the System Properties dialog. Note: Only active systems can be edited. • Click the Edit button. Result: The Edit dialog is displayed. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 55 4 Installation 4.2 Connect to ÄKTA pure 4.2.1 Power and communication Step Action 2 In the Edit dialog: • Select Fraction collectors from the Component types list. Result: All available fraction collectors are shown in the Component selection list. Note: Instrument modules are referred to as Components in UNICORN. Note: An outlet valve has to be selected in order to select a fraction collector. • Select the Fraction collector check box. • Choose what kind of fraction collector to add from the drop down list box. • Note: Fraction collector 2 is only available in the Component selection list if Fraction collector is already selected. 3 Click the OK button to apply the changes. Check/change Node ID The Node ID for Fraction collector F9-C cannot be changed. The Node ID for Fraction collector F9-R needs to be changed if • 56 the fraction collector will be used as secondary fraction collector, fraction collector, 2nd. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.2 Connect to ÄKTA pure 4.2.1 Power and communication • the fraction collector is to be used as primary fraction collector after it has been used as secondary. The Node ID for Fraction collector F9-R is set by positioning the arrow of the rotating switch at the back of the Fraction collector. Use a screwdriver to set the arrow of the switch to the desired number. Fraction collector Order Node ID Fraction collector F9-C 1st 0 Fraction collector F9-R 1st 0 2nd 1 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 57 4 Installation 4.2 Connect to ÄKTA pure 4.2.2 Connect tubing 4.2.2 Connect tubing Introduction This subsection contains information on how to connect tubing from ÄKTA pure to Fraction collector F9-C and Fraction collector F9-R. Fraction collector F9-C Fraction collector F9-C is delivered with all internal tubing in place. The tubing between the fraction collector and ÄKTA pure need to be installed. Follow the instructions in the table below to connect the tubing from the ÄKTA pure instrument to the fraction collector. 58 Step Action 1 Connect the tubing Frac to the Frac port on the outlet valve on the ÄKTA pure instrument. 2 Connect the other end of the Frac tubing to the inlet port on the fraction collector. 3 Adjust the delay volume setting in UNICORN, see Section 4.4 Set the delay volume, on page 62 for more details. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.2 Connect to ÄKTA pure 4.2.2 Connect tubing Fraction collector F9-R Follow the instructions in the table below to connect the tubing from the fraction collector to the ÄKTA pure instrument. Step Action 1 Lift out the Tubing holder (4) from the Delivery arm (1). 3 5 6 4 2 Loosen the nut of the Tubing holder. Do not remove the Tubing holder nut (5) from the Tubing holder. 3 Insert the tubing through the Tubing holder. 4 Place the Tubing holder with the tubing over the Tube adjustment cavity (2) of the Delivery arm. Push the tubing down against the bottom of the Tube adjustment cavity, and then fingertighten the Tubing holder nut. This ensures the correct length of the exposed tubing end (6). 5 Re-install the tubing holder in the Delivery arm. 6 For Fraction collector F9-R: Connect the tubing from the fraction collector to the port Frac on the outlet valve. For Fraction collector F9-R, 2nd: Connect the tubing from the second fraction collector to: 7 • port Out 10 on Outlet valve V9-O • port Out 1 on Outlet valve V9-Os Adjust the delay volume setting in UNICORN to the volume of the tubing, see Section 4.4 Set the delay volume, on page 62 for more details. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 59 4 Installation 4.3 Prepare waste tubing for Fraction collector F9-C 4.3 Prepare waste tubing for Fraction collector F9-C The waste outlet is located on the rear of the fraction collector. See instructions below for how to prepare the waste tubing. Instructions Follow the instructions below to prepare the waste tubing. Step Action 1 Insert the waste tubing in a vessel. NOTICE The maximum level of the vessel for the waste tubing must be lower than the fraction collector. 60 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.3 Prepare waste tubing for Fraction collector F9-C Step Action 2 Cut the waste tubing to appropriate length. It is important that the tubing is not bent and will not be submerged in liquid during the run. Note: If the tubing is too short, replace it with new tubing. Do not lengthen the tubing as this might cause obstruction of the tubing. 3 Securely fasten the tubing to the waste vessel. CAUTION Make sure that the waste vessel will hold all the produced volume of the run. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 61 4 Installation 4.4 Set the delay volume 4.4 Set the delay volume Introduction The delay volume settings are used to make sure that the fractions collected during fractionation, using the outlet valve or the fraction collector, correspond to the fractions indicated in the chromatogram. Set the delay volume Follow the instructions below to set the delay volume between the UV monitor and the outlet valve and between the UV monitor and the fraction collector. Refer to the ÄKTA pure System handbook for more information about delay volumes. Step 1 Action Select System:Settings in the System Control module. Result: The System Settings dialog opens. 2 62 • Select Tubing and Delay Volumes and select Delay volume: Monitor to outlet valve. • Type in the volume in the Volume field and click OK. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 4 Installation 4.4 Set the delay volume Step Action 3 In the System Settings dialog: • Select Delay volume: Monitor to outlet valve or Monitor to frac according to the configuration you are using. • Type in the volume in the Volume field and click OK. Note: The system will use the delay volume appropriate to the configuration used and ignore other settings (e.g. the value for Monitor to outlet valve will be ignored if you are using a fraction collector). It is however recommended to set all delay volumes so that the volumes remain correct if you change fractionation method. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 63 4 Installation 4.5 Performance test 4.5 Performance test Before taking the Fraction collector F9-C and Fraction collector F9-R instrument into use, run a performance test to check the function of the equipment. See the ÄKTA pure System handbook for further instructions. 64 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5 Operation About this chapter This chapter describes how to prepare and assemble the fraction collector before a run. In this chapter Section See page 5.1 Fraction collector F9-C 66 5.2 Fraction collector F9-R 73 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 65 5 Operation 5.1 Fraction collector F9-C 5.1 Fraction collector F9-C About this section This section describes how to prepare and assemble Fraction collector F9-C before a run. The fraction collector is connected to ÄKTA pure and controlled by UNICORN. Control of the fraction collector can be achieved automatically in a method run, or manually. CAUTION Fire Hazard. Do not fractionate flammable liquids using Fraction collector F9-C. When running RPC methods, or other procedures using organic solvents, collect fractions through the outlet valve or Fraction collector F9-R. Note: The tray and racks can tilt slightly when not fully inserted into the fraction collector and may harm the fractionation arm. The tilt is due to the height difference between the door and the floor of the fraction collector. The tendency to tilt is affected by the placement and weight of the cassettes, tubes or bottles Prepare the fraction collector Before starting to prepare Fraction collector F9-C, check the fractionation settings in the method to be run. Perform the steps described below according to the settings in the method. • Insert the Cassette tray or a rack for tubes or bottles. • Change the System Settings in UNICORN to set the fractionation mode and other settings for fraction collection. How to insert a tray or a rack is shown below. For information on how to change the System Settings before a run, see the ÄKTA pure System handbook. 66 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5.1 Fraction collector F9-C Prepare and insert the Cassette tray Follow the instructions below to add cassettes to the Cassette tray and insert the tray into the fraction collector. Step Action 1 If you are to use cassettes with the QuickRelease function, open the cassettes. For more information on the QuickRelease function see the ÄKTA pure System handbook. 2 Place the tubes and deep well plates in the cassettes. Make sure that the deep well plates are rotated so that the well marked A1 is positioned above the A1 marking on the cassette. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 67 5 Operation 5.1 Fraction collector F9-C 68 Step Action 3 Close the cassettes that have the QuickRelease function. 4 Place the cassettes on the Cassette tray. Make sure that the cassette type code (see illustration below) faces the front of the tray marked with the GE logo. 5 Open the door of the fraction collector using the handle. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5.1 Fraction collector F9-C Step Action 6 Insert the tray into the fraction collector: • make sure that the front of the tray (marked with the GE logo) faces outwards • position the tray and slide it into the fractionation collector until it reaches the end • make sure that the Tray catch snaps into closed position, as shown below. Note: The tray can tilt slightly when not fully inserted into the fraction collector and may harm the fractionation arm. The tilt is due to the height difference between the door and the floor of the fraction collector. The tendency to tilt is affected by the placement and weight of the cassettes. Note: A height exclusion bar ensures that the tubes and deep well plates are correctly positioned and cannot damage the dispenser head. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 69 5 Operation 5.1 Fraction collector F9-C Step Action 7 Close the door. Make sure that it closes properly. Result: After the door has been closed, the fractionation arm scans the cassette type code of each cassette to identify the cassette types. If deep well plates are used, the instrument also identifies the types of deep well plates. Note: If the rack is inserted with the front of the rack facing the wrong way it will not be possible to close the door. Prepare and insert a rack for tubes or bottles Follow the instructions below to insert a rack for tubes or bottles into the fraction collector. 70 Step Action 1 Place the tubes or bottles in the rack. 2 Open the door of the fraction collector using the handle. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5.1 Fraction collector F9-C Step Action 3 Insert the rack into the fraction collector: • make sure that the front of the rack (marked with the GE logo) faces outwards • position the rack and slide it into the fractionation collector until it reaches the end • make sure that the Tray catch snaps into closed position, as shown below. Note: The rack can tilt slightly when not fully inserted into the fraction collector and may harm the fractionation arm. The tilt is due to the height difference between the door and the floor of the fraction collector. The tendency to tilt is affected by the placement and weight of the tubes or bottles. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 71 5 Operation 5.1 Fraction collector F9-C Step Action Note: A height exclusion bar ensures that the tubes and deep well plates are correctly positioned and cannot damage the Dispenser head. Note: Do not use the Cassette tray when the Rack for 50 ml tubes or the Rack for 250 ml bottles is placed in the fraction collector. 4 Close the door. Make sure that it closes properly. Note: If the rack is inserted with the front of the rack facing the wrong way it will not be possible to close the door. 72 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5.2 Fraction collector F9-R 5.2 Fraction collector F9-R About this section This section describes how to prepare and assemble Fraction collector F9-R before a run. The Fraction collector F9-R is connected to ÄKTA pure and controlled by UNICORN. Control of the fraction collector can be achieved automatically in a method run, or manually. Prepare the fraction collector Before starting to prepare Fraction collector F9-R, check the fractionation settings in the method to be run. Perform the steps described below according to the settings in the method. • Assemble the tube rack • Insert collection tubes • Adjust the Delivery arm • Change the System Settings in UNICORN to set the fractionation mode and other settings for fraction collection. It is shown how to assemble the tube rack, insert collection tubes and adjust the delivery arm below. For more information on how to change the System Settings before a run, see the ÄKTA pure System handbook. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 73 5 Operation 5.2 Fraction collector F9-R Assemble the Tube rack Follow the instructions below to assemble the Tube rack. See the ÄKTA pure System handbook for information about which Tube rack parts to use. Step Action 1 Insert the Tube support (4), if required, into the bowl (3). The circular marks on the Tube support should face down. Note: When assembling a Tube rack, Single cutouts (1) and L-shaped cutouts (2), are used for various inserts depending on the length of the collection tubes. See ÄKTA pure System handbook for detailed information. 2 74 Insert the Tube guide (5) with the tube position numbers upwards. The Tube guide should rest about 1 cm above the Tube support. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5.2 Fraction collector F9-R Step Action 3 Insert the Tube holder (6) with the tube position numbers upwards: • Check that tube position 1 (7) is directly above tube position 1 (7) of the Tube guide. 7 • Push the flexible bowl out at each rib and snap the Tube holder under the top lip of the rib (8). 8 Note: Do not force the tube holder into place as this may damage the lip. • 4 Check that the surface of the Tube holder is level. Gently move the Delivery arm (10) out to the outer stop. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 75 5 Operation 5.2 Fraction collector F9-R Step Action 5 Place the Tube rack (9) over the Central spindle (11) and pull the spring loaded Drive sleeve (12) out so the Tube rack comes to rest. 9 10 12 11 Insert collection tubes Insert sufficient collection tubes in to the Tube rack, starting at position 1, pushing each one down as far as they will go. All the tubes must be of the same length and diameter and there should be no empty positions in the sequence. 76 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5.2 Fraction collector F9-R Adjust the Delivery arm Follow the instructions in the table below to adjust the Delivery arm. Step Action 1 Gently lift and then lower the Delivery arm (1), and allow it to move in so the Tube sensor (2) touches the collection tubes of the outer track. 1 2 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 77 5 Operation 5.2 Fraction collector F9-R Step Action 2 Adjust the height of the Delivery arm as follows: • Loosen the lock knob (3) 3 • Adjust the height so that the horizontal mark (5) on the Tube sensor (6) is at the same level as the top of the flat collection tubes and approximately 2 mm over the top of the flanged collection tubes (4). 4 5 6 • 78 Lock the Delivery arm at this height with the lock knob. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 5 Operation 5.2 Fraction collector F9-R Step Action 3 Check that the Tube sensor (9) is in the correct position for the used tubes (10). The eluent tubing (7) should be above the center of the collection tube. Use the Sensor control (8) to position the Tube holder. 7 8 9 10 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 79 6 Maintenance 6 Maintenance About this chapter This chapter describes maintenance and provides instructions for how to clean Fraction collector F9-C and Fraction collector F9-R. For further information, see the ÄKTA pure System handbook. CAUTION Explosion hazard if flammable liquid leaks during cleaning of the flow path. When cleaning the flow path of Fraction collector F9-C with a flammable liquid like ethanol, carefully inspect the flow path, including the waste tubing, to make sure there will be no leakage. In this chapter Section 80 See page 6.1 Fraction collector F9-C 81 6.2 Fraction collector F9-R 83 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 6 Maintenance 6.1 Fraction collector F9-C 6.1 Fraction collector F9-C Maintenance interval Clean the fraction collector when required, for example if liquid has been spilled in the fraction collector chamber. The internal tubing of the fraction collector may need to be replaced for maintenance or for process purposes. Information on how and when to replace the internal tubing can be found in the ÄKTA pure System handbook. Required material The following material is required: • Wash bottle • Water or 20% ethanol • Cloth Instruction Follow the instruction below to clean the interior of the Fraction collector. The locations of the components of the fraction collector are shown in Section 3.1.2 Fraction collector F9-C illustrations, on page 29. Step Action 1 In System Control, select Manual:Execute Manual Instructions:Fraction collection:Frac cleaning position . Click Execute. Result: The Dispenser head moves to cleaning position, and the Instrument display states System pause. 2 Open the Frac door and remove the rack 3 Wash the Cassette tray or rack and the Cassettes (if applicable), with water and a mild cleaning agent. 4 Lift off the Waste groove and wash it with water and a mild cleaning agent. Refit the Waste groove. 5 Wipe off the interior of the fraction collector using a damp cloth. Wipe off stains using a mild cleaning agent or 20% ethanol. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 81 6 Maintenance 6.1 Fraction collector F9-C Step Action 6 Wipe off the Dispenser head and its diode windows (the Drop sync sensor and the Type code reader) using a wash bottle with water or 20% ethanol and a cloth. 7 Let the fraction collector dry completely before starting a run. 8 Close the door of the fraction collector. Result: Automatic scanning is performed. 9 In the System Control module, press the End icon in the toolbar. Result: The Dispenser head moves to home position. 82 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 6 Maintenance 6.2 Fraction collector F9-R 6.2 Fraction collector F9-R Maintenance interval Clean the Fraction collector when required, for example in case of liquid spill. Required material The following material are required: • Water or 20% ethanol • Cloth Instruction Follow the instructions below to clean the instrument externally. Step Action 1 Check that no run is in progress. 2 Switch off the instrument. 3 Wipe the surface with a damp cloth. Wipe off stains using a mild cleaning agent or 20% ethanol. Wipe off any excess. 4 Let the Fraction collector F9-R dry completely before restart. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 83 7 Reference information 7 Reference information See the ÄKTA pure Product information and the ÄKTA pure System handbook for information on the following: • Technical specifications • Chemical resistance For information on operating conditions, see Section 4.1.3 Site requirements, on page 50. Environmental ranges Parameter Data Storage and transport temperature range -25°C to +60°C Chemical environment See the ÄKTA pure System handbook. Equipment noise level Equipment Acoustic noise level Fraction collector F9-C < 60 dB A Fraction collector F9-R < 60 dB A Heat output Equipment Heat output Fraction collector F9-C Typically 6 W Maximum 18 W Fraction collector F9-R 84 2W Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA A Declaration of Hazardous Substances (DoHS) Appendix A Declaration of Hazardous Substances (DoHS) Introduction The following product pollution control information is provided according to SJ/T113642006 Marking for Control of Pollution caused by Electronic Information Products. 㫈㗽SJ/T11364-2006ᇵ䟄≰㋾ℶ❐㻰㩢㘶Ⓟ㪖幕尐㻑ᇶ䔈㙟∪Ⱁₚ㦘␂㻰㩢 㘶Ⓟ 㡈槱䤓≰㋾ Symbols used in pollution control label 䟄≰㋾ℶ❐㻰㩢㘶Ⓟ㪖㉦広㢝 Label Meaning This symbol indicates the product contains hazardous materials in excess of the limits established by the Chinese standard SJ/T11363-2006 Requirements for Concentration Limits for Certain Hazardous Substances in Electronic Information Products. The number in the symbol is the Environment-friendly Use Period (EFUP), which indicates the period during which the toxic or hazardous substances or elements contained in electronic information products will not leak or mutate under normal operating conditions so that the use of such electronic information products will not result in any severe environmental pollution, any bodily injury or damage to any assets. The unit of the period is “Year”. In order to maintain the declared EFUP, the product shall be operated normally according to the instructions and environmental conditions as defined in the product manual, and periodic maintenance schedules specified in Product Maintenance Procedures shall be followed strictly. Consumables or certain parts may have their own label with an EFUP value less than the product. Periodic replacement of those consumables or parts to maintain the declared EFUP shall be done in accordance with the Product Maintenance Procedures. This product must not be disposed of as unsorted municipal waste, and must be collected separately and handled properly after decommissioning. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 85 A Declaration of Hazardous Substances (DoHS) Label Meaning 年㪖㉦嫷㢝㦻ℶ❐⚺㦘怔扖SJ/T11363-2006ᇵ䟄≰㋾ℶ❐₼㦘㹡 㦘⹂䓸德䤓棟 摞尐㻑ᇶ₼棟摞䤓㦘㹡㦘⹂䓸德ᇭ㪖㉦₼䤓㟿ⷦ㦻 ℶ❐䤓䘾≬∎䞷㦮᧨嫷㢝㦻 ℶ❐⦷㷲デ∎䞷䤓㧰ↅₚ᧨㦘㹡㦘⹂䓸 德ₜ↩♠䞮⮥㽓㒥䴐♧᧨䞷㓆∎䞷㦻ℶ❐ ₜ↩⺈䘾⬒抯㒟₴摜㻰㩢 㒥⺈␅ⅉ愺ᇬ徱ℶ抯㒟₴摜㗮⹂䤓㦮棟ᇭ◤⇜ᇭ ≬幐㓏䟂㢝䤓䘾≬∎䞷㦮棟᧨ㄣ㖘ℶ❐㓚␛₼㓏屓⸩䤓䘾⬒㧰ↅ ✛㡈㽤扪嫛㷲 デ∎䞷᧨ㄅ₴㫋挄⸗ℶ❐冃≽㓚␛₼屓⸩䤓㦮冃≽✛ ≬␊尐㻑ᇭ ℶ❐₼䤓䀗劦ↅ✛㩟K榅捷ↅ♾厌㦘␅◤䕻䤓䘾≬∎䞷㦮棟㪖㉦᧨ ㄅ₣␅䘾≬∎ 䞷㦮棟㦘♾厌㹣㠃₹ℶ❐㦻愺䤓䘾≬∎䞷㦮棟䩼ᇭㄣ Ⓙ㦮㖘ℶ❐冃≽䲚ㄞ㦃㗱挲 K䀗劦ↅ✛榅捷ↅ᧨ⅴ≬幐㓏䟂㢝䤓㠃 ₹ℶ❐䤓䘾≬∎䞷㦮棟ᇭ 㦻ℶ❐⦷∎䞷⺎✌兢㧮㢅ₜ♾⇫㣽抩䞮㿊⨒⧍⮓䚕᧨ㄣ嬺◤䕻㟅 楕ⰴ⠓⮓䚕 List of hazardous substances and their concentrations ℶ❐₼㦘㹡㦘⹂䓸德㒥⏒侯䤓⚜䱿♙⚺摞 Indication for each major part if substance exceeds limit Value Meaning O Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in SJ/T11363-2006. 嫷䯉年㦘㹡㦘⹂䓸德⦷年捷ↅ㓏㦘⧖德㧟㠨₼䤓⚺摞⧖⦷SJ/T113632006 㪖屓⸩䤓棟摞尐 㻑ⅴₚ X Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement in SJ/T11363-2006. • Data listed in the table represents best information available at the time of publication 嫷䯉年㦘㹡㦘⹂䓸德咂⺠⦷年捷ↅ䤓㩟⧖德㧟㠨₼䤓⚺摞怔⒉ SJ/T11363-2006 㪖屓⸩䤓 棟摞尐㻑 • 86 㷳嫷㓏⒦㟿㗽♠を㢅㓏厌噆㈦䤓㦏∂≰㋾ Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA A Declaration of Hazardous Substances (DoHS) List of hazardous substances Component name 捷ↅ⚜䱿 1 Hazardous substance 㦘㹡㦘⹂䓸德㒥⏒侯 Pb Hg Cd Cr6+ PBB PBDE 杔 㻭 柘 ⏼ↆ杻 ⮩䅃勣啾 ⮩䅃ℛ啾搩 29-0277-43 1 0 0 0 0 0 0 29-0113-621 0 0 0 0 0 0 The product has not been tested as per the Chinese standard SJ/T11363-2006 Requirements for Concentration Limits for Certain Hazardous Substances in Electronic Information Products. Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 87 Index Index A ÄKTA pure Cassettes and racks, 33 preparation, 66, 73 C Labels, 20 Fraction collector F9-C instrument label, 21 M Cassettes, 33 CE Conformity, 9 Marking, 10 Manufacturing information, 9 N Notes and tips, 8 D Delay volume set, 62 Dimensions instrument, 49 E Emergency procedures emergency shutdown, 23 F Fractionation assemble tube rack, 43 Fraction collector F9-C Cassettes and racks, 33 cleaning, 81 connect tubing, 58 requirements on deep well plates, 35 requirements on tubes, 35 Tubes, 42 Fraction collector F9-R cleaning, 83 connect tubing, 59 Delivery arm, 77 Insert collection tubes, 76 Tube rack, 74 I Installation performance test, 64 Instrument overview Fraction collector F9-R, 40 88 L O Operating conditions, 50 P Prepare the system ÄKTA pure, 66, 73 waste tubing, 60 Prerequisites, 7 Purpose of this document, 6 R Racks, 33 Recycling disposal, 25 Recycling procedures, 25 Reference information recycling, 25 Regulatory information international standards, 10 regulatory compliance of connected equipment, 11 S Safety notices, 8 Safety precautions, 13 emergency procedures, 23 labels, 20 Storage, 46 T Tube rack assembly, 43 Typographical conventions, 6 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA Index U W User information, important, 7 Waste tubing prepare, 60 Weight instrument, 48 Fraction collector F9-C and Fraction collector F9-R Operating Instructions 29-0256-06 AA 89 For local office contact information, visit www.gelifesciences.com/contact GE, imagination at work and GE monogram are trademarks of General Electric Company. GE Healthcare Bio-Sciences AB Björkgatan 30 751 84 Uppsala Sweden Seahorse Bioscience is a trademark of Seahorse Bioscience Corporation. www.gelifesciences.com/AKTA ÄKTA, UNICORN and Whatman are trademarks of GE Healthcare companies. Eppendorf is a trademark of Eppendorf AG. © 2012 General Electric Company – All rights reserved. First published Nov. 2012 All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information. UNICORN: Any use of this software is subject to GE Healthcare Standard Software End-User License Agreement for Life Sciences Software Products. A copy of this Standard Software End-User License Agreement is available on request. GE Healthcare Europe GmbH Munzinger Strasse 5, D-79111 Freiburg, Germany GE Healthcare UK Limited Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK GE Healthcare Bio-Sciences Corp. 800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327, USA GE Healthcare Japan Corporation Sanken Bldg. 3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073, Japan imagination at work 29-0256-06 AA 11/2012