Boston University/ Boston Medical Center Analytical X-Ray Safety Manual [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 BOSTON UNIVERSITY ANALYTICAL X-RAY EQUIPMENT SAFETY PROGRAM Page Table of Contents 1.0 Purpose 5 2.0 Scope and Applications 5 3.0 Organization of X-ray Safety Program 3.1 Permitting and Registration 3.2 Training and Instruction 3.3 Enforcement Actions 3.4 Deficiency Correction Analytical X-ray Equipment: 4.1 X-ray Permit Holder’s Responsibilities 4.2 X-ray Supervisor’s Responsibilities 4.3 Operators' Responsibilities 4.4 Minors General Equipment and Safety Requirements 5.1 Safety Devices 5.2 Warning Devices 5.3 Ports 5.4 Labeling 5.5 Shutters 5.6 Warning Lights 5.7 Radiation Source Housing 5.8 Generator Cabinet 5.9 Separate Room 5.10 Permanent Shields 5.11 Set Up Procedures 5.12 Test Safety Devices 5.13 Viewing Devices and Alignment 5.14 X-ray Use Log Radiation Survey 5 7 7 8 8 9 9 10 11 11 11 11 11 12 12 12 12 13 13 13 13 13 13 14 14 14 Personal Monitoring and Control of X-ray Exposure 7.1 Proper Use of Dosimetry Devices 15 15 4.0 5.0 6.0 7.0 Page | 2 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 Table of Contents Page Appendix A: Definitions 17 Appendix B: Application for a Permit to Possess and Use Equipment that Emits Ionizing Radiation (XR-1) Appendix C: X-ray User Certification Form (XR-2) 21 Appendix D: Template SOP (XR-3) 28 Appendix E: X-ray Laboratory Inspection Checklist (XR-4) 38 Appendix F: X-ray Equipment Inventory (XR-5) 41 26 Page | 3 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 Page | 4 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 1.0 Purpose Analytical X-ray equipment is used for X-ray diffraction analysis, fluorescence analysis, or direct X-ray transmission analysis of materials. These analytical X-ray systems are comprised of components that utilize X-rays to determine elemental composition, or to examine the microstructure of materials. These analytical X-ray devices are used for nonmedical purposes. The primary objective of this program is to keep occupational exposures to radiation As Low As Reasonably Achievable (ALARA) while utilizing these types of equipment. This safety program establishes Boston University and Boston Medical Center procedures for the safe operation of analytical X-ray equipment and the associated potential radiation hazards. 2.0 Scope and Application Analytical X-ray equipment has become a major tool in research and quality control programs. Despite the advances in operating techniques and equipment design, the most common hazards are due to operators' errors and equipment malfunctions. The potential exposures to the primary beam are of a major concern when evaluating potential radiation exposures. Exposures to the primary beam in a typical analytical X-ray unit may be as great as 100,000 R/min. This program applies to all departments, supervisors, employees, students, visiting scientists, and any personnel from other organizations who work with or near analytical X-ray equipment at Boston University. Operational procedures will be established for each X-ray unit to maintain radiation exposures As Low As Reasonably Achievable (ALARA) with due considerations as to the feasibility and nature of the research being conducted. 3.0 Organization of X-ray Safety Program Radiation Safety Committee: The Boston University (BU) and Boston Medical Center (BMC) Radiation Safety Committee is responsible for the establishment and continuing review of the Radiation Safety Program at Boston University and Boston Medical Center (BU/BMC). The Committee is also responsible for the BU and BMC compliance with the radiation safety regulations promulgated by the State, Federal, and local agencies for ionizing radiationproducing equipment. Such equipment must be registered with the BU/BMC Division of Medical Physics and Radiation Safety (DMPRS). Boston University has a separate Laser Safety Committee and Program which deals with the use of laser equipment. Chief Medical Physicist: Page | 5 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 The Chief Medical Physicist oversees the implementation of the analytical X-ray equipment program including surveys, procedures and incident review. Radiation Safety Officer: The Radiation Safety Officer oversees the radiological safety of the analytical X-ray equipment. Boston University Division of Medical Physics and Radiation Safety The Boston University Division of Medical Physics and Radiation Safety (DMPRS) provides services to assist departments, supervisors, students and technicians in maintaining a comprehensive analytical X-ray equipment safety program including: 1. Training and instruction in the safety procedures and practices required for all persons who work with or near analytical X-ray equipment. 2. Managing the RIMS website that allows X-ray Permit Holders to review information specific to their labs such as Authorized User status, training status, and lab profile information at http://www.bu.edu/rims Page | 6 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 3.1 3. Maintaining a current listing of analytical X-ray equipment and its authorized users. 4. Evaluating each installation as to the control of radiation exposures including recommendations for placement of radiation warning signs and/or warning devices. 5. Performing routine annual radiation safety inspections of analytical X-ray equipment. 6. Review and approval of modifications to X-ray apparatus that affects radiation protection including X-ray tube housing, cameras, shielding, and safety interlocks. 7. Providing personal monitoring badges and area monitors. 8. Investigating any unusual radiation exposures to personnel and taking remedial action, if necessary. 9. Assisting in achieving compliance with all applicable federal, state, and local rules and regulations. 10. Registering X-ray units with the Massachusetts Department of Public Health as per section 105 CMR 120.000. Permitting and Registration of users of Analytical X-ray equipment with the Division of Medical Physics and Radiation Safety All analytical X-ray equipment must be registered with DMPRS using X-ray Equipment Inventory Form XR-5. The supervisor of the analytical X-ray equipment must apply for a Permit to Possess and Use Equipment that Emits Ionizing Radiation using Form XR-1. Permits are issued by Radiation Safety Committee upon recommendation by Radiation Generating Devices Subcommittee. Permits are revived annually by DMPRS. No person shall begin using analytical X-ray equipment without first completing the required DMPRS training and SOP specific training which is to be documented on the X-ray User Certification Form, XR-2. 3.2 Training and Instruction Page | 7 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 Individuals requiring X-ray Safety training may register for training on-line via RIMS at http://www.bu.edu/rims/ a. Permit Holders, Lab Staff, and Students All permit holders, laboratory staff, and students working with or frequenting a lab using analytical X-ray equipment must complete the Basic Analytical X-ray Training on RIMS. The permit holder must provide to staff and students any additional training specific to the instrument and its standard operating procedures. b. Auxiliary Workers All custodians, trades and non-technical workers who will work in or frequent an X-ray restricted area must attend the Radiation Safety Awareness program presented by DMPRS. c. Visitors A visitor is defined as someone who enters the lab on an infrequent basis solely for the purposes of observation. A visitor may not operate the Analytical X-ray Equipment d. Exceptions The Radiation Safety Officer (RSO) or Chief Medical Physicist may exercise discretion in modifying the training requirements for each training category as appropriate. 3.3 Enforcement Actions The Radiation Safety Committee can suspend or revoke an Analytical X-ray Permit which is not in compliance with BU/BMC policies and regulations. When an Analytical X-ray Permit is suspended or revoked, authorized users of that equipment may not utilize the unit and the equipment may be locked out at the discretion of the RSO. A suspended authorization will not be reinstated until the Committee is assured that reasonable measures have been instituted to prevent recurrence of the deficiencies. A revoked permit will require new permit application 3.4 Deficiency Correction Laboratories using analytical X-ray equipment are inspected by DMPRS annually. Additional inspections may be conducted at the discretion of the Radiation Safety Officer or Chief Medical Physicist, for example, as a component of an X-ray incident Page | 8 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 investigation. The results of the inspection will be documented with copies sent to the Permit Holder. Inspection findings will be reported to the Radiation Generating Device subcommittee for further recommendation 4.0 Analytical X-ray Equipment The inventory of the analytical X-ray equipment is available in DMPRS 4.1 Permit Holder's Responsibilities Each analytical X-ray equipment Permit Holder is responsible for: 1. Providing the Boston University DMPRS with a complete XR-1 form (Application for Permit to Possess and Use Equipment that Emits Ionizing Radiation at BU). See Appendix B. DMPRS will review each proposed installation and corresponding operating procedures. A permit must be obtained from the Radiation Safety Committee before the X-ray equipment is operated. 2. All operations carried out with the equipment. 3. Ensuring that all the personnel under their supervision are registered with and receive initial training from the DMPRS in analytical X-ray equipment safety. 4. Compliance with the specific recommendations made by the DMPRS, and also the general equipment and safety requirements listed in Section 7.0. 5. Ensuring that only authorized users will enter the areas that are restricted due to the use of the analytical X-ray units. 6. Providing specific hands-on training to the authorized user(s) for each analytical X-ray unit, in accordance with the approved Standard Operating Procedures (SOP). A template SOP may be found in Appendix D, Form XR-3. Completed SOPs for each piece of X-ray equipment will be maintained in the X-ray Manual and updated annually. An up-to-date list of Authorized Users shall be posted on the lab’s permit. 7. Ensuring that the project has a properly operating survey instrument that has been calibrated within the last twelve months. 8. Informing DMPRS of acquisitions, transfers, servicing or decommissioning of Xray equipment. The X-ray inventory shall be amended as needed to reflect such changes. (Appendix F, Form XR-5). Page | 9 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 9. 10. Notifying the Radiation Safety Officer (RSO) and Chief Medical Physicist of the Boston University Division of Medical Physics and Radiation Safety when: a) it is necessary to alter safety devices, such as bypassing interlocks. The exception would be generic bypassing for test purposes that has been authorized by DMPRS. b) it is known or suspected that a radiation exposure of personnel may have occurred. c) an existing unit is moved or beam path is altered. d) there are changes in operating parameters such as kV and mA beyond that which were approved by DMPRS. e) there are changes in the approved shielding. f) there is any major service performed on the X-ray unit by lab personnel. g) an outside service technician is contracted. Such technician must be registered with the MDPH and approved by the RSO or Chief Medical Physicist. h) planning to procure new analytical X-ray equipment or to dispose, sell or gift unwanted equipment. Procurement or removal of such equipment requires the advance authorization of the BU RSO. Emergency Procedures Written emergency procedures shall be established for each X-ray analytical unit by the Permit Holder and shall be posted in a conspicuous location near each Xray unit. These should include the telephone number(s) of the Permit Holder, the BU Division of Medical Physics and Radiation Safety, the BU Campus Police Department, and instructions to follow in the event of a known or suspected accident involving radiation exposure 4.2 X-ray Supervisor’s Responsibility 1. Assuming the responsibilities of the X-ray Permit Holder in the absence of that individual. Page | 10 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 4.3 Operators’ Responsibility Each authorized user of the analytical X-ray equipment is responsible for: 4.4 1. Completing the analytical X-ray safety training provided by the BU DMPRS prior to operating the analytical X-ray equipment. 2. Receiving specific hands-on training with each analytical X-ray instrument provided by the X-ray supervisor in accordance with the written SOP. 3. Wearing the assigned personal monitoring dosimeter (s) (see Section 7.0) 4. Documentation of items 1, 2 and 3 on the X-ray User Authorization Form, XR-2 (see Appendix C) that is maintained in the X-ray Safety Manual in the laboratory. Minors Minors are not allowed to operate analytical X-ray equipment. The Permit Holder shall be responsible for ensuring that the presence of any minor in the laboratory is in full compliance with the Boston University Personnel Policy Manual (see http://www.bu.edu/hr/files/documents/employee-handbook.pdf, page 8). 5.0 General Equipment and Safety Requirements As defined in 105 CMR 120.603. 5.1 SAFETY DEVICE A device which prevents the entry of any portion of an individual's body into the primary X-ray beam path or which causes the beam to be shut off upon entry into its path shall be provided on all open beam configurations. Exception to this requirement may be granted through the Department of Public Health with the prior approval of the BU RSO and Chief Medical Physicist. 5.2 WARNING DEVICES Open beam configurations shall be provided with a readily discernible indication of: a.) X-ray tube "on-off" status located near the radiation source housing if the Page | 11 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 primary beam is controlled in this manner; and/or b.) Shutter "open-closed" status located near each port on the radiation source housing if the primary beam is controlled in this manner. Warning devices shall be labeled so that their purpose is easily identifiable. On equipment installed after February 9, 1996, warning devices shall have fail-safe characteristics. 5.3 PORTS Unused ports on radiation source housings shall be secured in the closed position in a manner that will prevent casual opening. 5.4 LABELING All analytical X-ray equipment shall be labeled with readily discernible sign(s) bearing the radiation symbol and the words: a.) "CAUTION-HIGH INTENSITY X-RAY BEAM" on X-ray source housing; and b.) CAUTION RADIATION-THIS EQUIPMENT PRODUCES RADIATION WHEN ENERGIZED" near any switch that energizes an X-ray tube if the radiation source is an X-ray tube; or c.) CAUTION-RADIOACTIVE MATERIAL" on the source housing if the radiation source is a radionuclide. 5.5 SHUTTERS On open beam configurations installed after February 9, 1996, each port on the radiation source housing shall be equipped with a shutter that cannot be opened unless a collimator or a coupling has been connected to the port. 5.6 WARNING LIGHTS An easily visible warning light labeled with the words "X-RAY ON" shall be located: a.) near any switch that energizes an X-ray tube and shall be illuminated only when the tube is energized; or b.) In the case of a radioactive source, near any switch that opens a housing shutter and shall be illuminated only when the shutter is open. On equipment installed after February 9, 1996, warning lights shall have fail-safe Page | 12 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 characteristics. 5.7 RADIATION SOURCE HOUSING Each radiation source housing shall be subject to the following requirements: a.) Each X-ray tube housing shall be equipped with an interlock that shuts off the tube if it is removed from the radiation source housing or if the housing is disassembled. b.) Each radioactive source housing or port cover or each X-ray tube housing shall be so constructed that, with all shutters closed, the radiation measured at a distance of 5 centimeters from its surface is not capable of producing a dose in excess of 2.5 millirems in one hour. For systems utilizing X-ray tubes, this limit shall be met at any specified tube rating. 5.8 GENERATOR CABINET Each X-ray generator shall be supplied with a protective cabinet, which limits leakage radiation measured at a distance of 5 centimeters from its surface such that it is not capable of producing a dose in excess of 0.25 millirem in one hour. 5.9 SEPARATE ROOM The analytical X-ray equipment must be placed in a separate room from other work areas whenever practical. 5.10 PERMANENT SHIELDS Properly installed permanent shields should be used in preference to temporary shielding. When temporary shielding is necessary, it must be securely fastened. 5.11 SET-UP PROCEDURES Set-up procedures will be carried out with the X-ray beam off or with shutters closed as much as possible. If the latter, a survey shall be performed before the starting set-up. 5.12 TEST SAFETY DEVICES All safety devices such as interlocks, shutters, warning lights, etc. will be tested Page | 13 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 upon each use to ensure proper operation. 5.13 VIEWING DEVICES AND ALIGNMENT Particular attention should be given to viewing devices to ensure that lenses and other transparent components attenuate the radiation beam to minimal levels when alignment involves working near the primary beam. The beam current should be reduced when a fluorescent alignment tool is used; dimming the room light will permit a significant reduction in beam current. The fluorescent alignment tool should be long enough to permit the analytical X-ray equipment user's hand to be kept a safe distance from the beam. The Division of Medical Physics and Radiation Safety can be consulted for more detailed advice. 5.14 X-RAY USE LOG A use log for each X-ray unit is required for all open beam analytical X-ray units. It is recommended but not required for other analytical x-ray units. The use log should document the authorized user, date, and duration of each use of the equipment. Completed logs should be available for review by DMPRS. 6.0 Radiation Surveys Analytical X-ray equipment USERS are required to make and keep records of the following surveys with a GM survey meter: 1. Upon equipment installation and within every 12 months thereafter; 2. Following any reconfiguration of the system; 3. Following maintenance requiring disassembly or removal of a local component; 4. During maintenance and alignment requiring the presence of a primary Xray beam when a local component has been disassembled or removed; 5. Any time a visual inspection of local components reveals an abnormal condition; and 6. Whenever personnel monitoring devices show a significant increase over Page | 14 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 the previous monitoring period or the readings are approaching the limits specified in 105 CMR 120.211. DMPRS will make surveys according to the following guidelines: 7.0 1. Upon the installation of the X-ray equipment and at least once a year thereafter. 2. Upon any change in the initial arrangement, number, or type of local components in the system. 3. Upon any maintenance requiring the disassembly or removal of a local component in the system. Personnel Monitoring and Control of X-ray Exposure The DMPRS periodically checks and surveys Analytical X-ray devices to ascertain that safety procedures are followed. In addition, personnel who are required to be badged per applicable state regulations (i.e. 105 CMR 120) will be monitored. Dosimeters will be provided to all X-ray personnel working with open beam analytical X-ray equipment to ensure that exposures are maintained ALARA. Employees not working with open beams may request to be monitored on a voluntary basis. Radiation exposure limits are provided in the Massachusetts state regulations 105 CMR 120.211. Any X-ray worker who is pregnant may voluntarily declare her pregnancy and the estimated date of conception in writing to the BU/BMC Division of Medical Physics and Radiation Safety. Thereafter, her occupational exposure shall be limited to 500 millirem for the nine-month gestational period. Declaration of pregnancy forms are available at: http://www.bu.edu/ehs/plans/management-plans/rpo/forms. At the RSO's discretion, an investigation may be performed at any exposure level. 7.1 PROPER USE OF DOSIMETRY DEVICES Page | 15 [BU/ BMC ANALYTICAL X-RAY SAFETY MANUAL] Rev. 3 : 2013 The analytical X-ray equipment users who were assigned dosimeters will wear whole body badges and/or finger rings as assigned by the DMPRS. The monitoring device should be worn such that the body part nearest the primary beam is monitored. Monitoring badges worn on the chest or abdomen may provide an indication as to the amount of stray radiation to the whole body. Operations involving the use of analytical X-ray equipment will be planned so that the exposures are in compliance with the occupational limits shown above. OTHER REQUIREMENTS: 1. Do not take dosimetry badges or rings home with you. 2. Do not leave badges / rings in a high background area. 3. Dosimetry issued by BU/BMC DMPRS can only be used at Boston University and Boston Medical Center. 4. Please contact DMPRS if you lose your badge / ring. 5. Do not wear another person’s badge / ring, or allow someone else to take your dosimetry device. . Page | 16 APPENDIX A DEFINITIONS Page | 17 APPENDIX A DEFINITIONS ANALYTICAL X-RAY EQUIPMENT: Any device, which uses X-rays for the purpose of examining the microstructure of materials. This includes all types of X-ray diffraction and spectrographic equipment. ANALYTICAL X-RAY SYSTEM: A group of components which x rays to determine the elemental composition or to examine the microstructure of materials. ANODE: A positive electrode; in an X-ray tube, it is the target for the accelerated electrons. CATHODE: A negative electrode; it is the filament at which free electrons are produced by thermionic emission. DOSE, ABSORBED (RAD): The amount of energy deposited in medium by a beam of ionizing radiation. The special unit of absorbed dose is the RAD, which is equal to 100 ergs/gm or 0.01 joule/kilogram. DOSE EQUIVALENT (REM): A quantity that expresses the irradiation incurred by exposed persons on a common scale for all radiations. It is defined numerically as the product of the absorbed dose in rads multiplied by the quality factor and the REM. (For radiation protection purposes in this safety program, the dose equivalent in rems may be considered numerically equivalent to the absorbed dose in rads and exposure in roentgens.) EXPOSURE (ROENTGEN): A measure of the ionization produced in air by x or gamma radiation. This special unit of exposure is the roentgen, which is equal to 2.58 E-4 coulomb of charge collected per kilogram of air exposed. FAIL-SAFE DESIGN: One in which all failures of indicator or safety components that can reasonably be anticipated cause the equipment to fail in a mode such that personnel are safe from exposure to radiation. For example: (a) If a light indicating “X RAY ON” fails, the production of X-rays shall be prevented, and (b) if a shutter status indicator fails, the shutter shall close. HALF VALUE LAYER (HVL): The thickness of any material that is required to reduce the intensity of a given beam by one half. INSTALLATION ENCLOSURE: That portion of an X-ray installation, which clearly defines the transition from a non-controlled area to a controlled area, and provides such shielding as may be required to limit the dose rate in non-controlled areas during normal operation. INTERLOCK: A device for precluding access to an area in which radiation is present by automatically reducing the exposure rate upon entry by personnel or parts of their body. Page | 18 kVp (KILOVOLTAGE PEAK): The maximum potential difference applied between the anode and cathode by a pulsating voltage generator. LEAKAGE RADIATION: All radiation coming from within the X-ray tube housing except the primary radiation beam. LOCAL COMPONENT: Part of an analytical X-ray system that includes areas that are struck by x rays such as radiation source housing, port and shutter assemblies, collimator, sample holders, cameras, goniometers, detectors and shielding, but do not include power supplies, transformers, amplifiers, readout devices, and control panels. mAs (MILLIAMPERE SECONDS): A combination unit, which is the product of the tube current (expressed in mA) and the exposure time (expressed in seconds). The total output of an X-ray tube is directly proportional to the mAs (or either of its components). MAXIMUM PERMISSIBLE DOSE EQUIVALENT: The maximum dose equivalent that a person or specified parts thereof shall be allowed to receive in a stated period of time. NORMAL OPERATION: Operation under conditions suitable for collecting data recommended by a manufacturer of the X-ray system. Recommended shielding and barriers shall be in place. OPEN-BEAM CONFIGURATION: An analytical X-ray system in which an individual could accidentally place some part of their body in the primary beam path during normal operation. PERMIT HOLDER (X-RAY): The BU faculty member issued an internal permit for use of analytical X-ray equipment by the Radiation Safety Committee. PRIMARY BEAM: Radiation which passes through an aperture of the source housing by a direct path from either the X-ray tube or a radioactive source located in the radiation source housing, which is either unscattered or undeflected. RADIATION AREA: Any area accessible to personnel in which there exists radiation at such levels that a major portion of the body (whole body, head and trunk, active blood-forming organs, gonads, or eye lenses) could receive in any one hour a dose equivalent in excess of 5 mrem, or in 5 consecutive days a dose equivalent in excess of 100 mrem. SCATTERED RADIATION: Radiation that, during passage through matter, has been deviated in direction. STRAY RADIATION: The sum of leakage and scattered radiation. SYSTEM BARRIER: That portion of an X-ray installation, which clearly defines the transition from a controlled area to a radiation area and provides such shielding as may be required to limit the dose rate in the controlled areas during normal operations. Page | 19 THERMIONIC EMMISSION: The process by which free electrons are produced at the cathode of an X-ray tube when the filament is electrically heated such that the thermal energy imparted to the electrons is sufficient to overcome the forces binding them to the filament. TUBE HOUSING-APPARATUS COMPLEX: Those parts of an analytical X-ray device in which X-rays are produced and utilized. This includes the X-ray tube housing, shutter or port assemblies, collimator, cameras, goniometers, and electronic radiation detectors. This is not intended to include such components as transformers, control panels, or temporary shielding. X-RAY GENERATOR: That portion of an X-ray system, which provides the accelerating voltage and current for the X-ray tube. X-RAY DIFFRACTION EQUIPMENT: An analytical X-ray device in which an X-ray beam (usually monochromatic) is made to strike a specimen, causing a portion of the beam to be diffracted. Measurements of certain parameters of the diffracted beam may be used to provide qualitative and/or quantitative information about the specimen. X-RAY FLUORESCENCE EQUIPMENT: An analytical X-ray device in which a polychromatic X-ray beam is made to strike a specimen, producing X-ray fluorescence, which is characteristic of the specimen. A portion of the fluorescent radiation is directed into an analyzing crystal where it is diffracted. The wavelength of interest may then be monitored by a properly positioned detector to produce qualitative and/or quantitative analysis of the specimen. X-RAY RESTRICTED AREA: A specified area in which exposure of personnel to radiation or radioactive material is controlled and which is under the supervision of a person who has knowledge of the appropriate radiation protection practices, including pertinent regulations. Page | 20 APPENDIX B Boston University/ Boston Medical Center Form XR-1 Application for a Permit to Possess and Use Equipment that Emits Ionizing Radiation Page | 21 Medical Physics and Radiation Safety 72 East Concord St. Boston, Massachusetts 02119-2511 Tel: 617.638.7052 Fax: 617.638.7509 APPLICATION FOR ANALYTICAL X-RAY USE PERMIT (Non-Human Use) New Renewal Amendment Please complete each section of this form and attach additional sheets as necessary. Return the completed application to the Division of Medical Physics and Radiation Safety at the address shown above. The RSO will review your application and notify you of preapproval or if further information is needed. The final approval of your application will be issued by the RadiationGenerating Devices subcommittee. Applications will require a copy of a current: a) CV, b) X-ray related trainings and certificates, c) SOPs, d) Completed user certification, e) X-ray unit inventory, and f) Laboratory layout map SECTION A. PROPOSED PERMIT HOLDER Permit Holder Name: Home Phone: Office Phone: Lab phone: X-ray Supervisor Name: Email: Home Phone: Office Phone: Lab phone: Email: TRAINING AND EXPERIENCE Permit Holder X-ray supervisor Highest Degree Obtained And Specialty Institution Training Where Trained (most recent) Date Hours of Training Please describe your training and experience with x-ray units you are planning on using. PROPOSED X-RAY USER(S) authorized to perform beam alignment (if applicable) Note: List the specific X-ray unit serial number(s) for which the individual has been trained and authorized. User (Last, First) X-Ray System (SNs) Training Date Page | 23 Radiation Protection Office 80 East Concord St. Boston, Massachusetts 02119-2511 Tel: 617.638.7052 Fax: 617.638.7509 SECTION B. PROPOSED UNIT(S) Instructions: All Analytical X-ray units are required to be registered with the Radiation Safety Officer. X-RAY MANUFACTU RER X-RAY MODEL SERIAL NUMBER BU PROPERTY ID EHS REF NUMBER X-RAY TYPE (diffraction, spectrographic, mammography) POWER / ENERGY OUTPUT APPLICATION INSTALLATION DATE OWNERSHIP (BU, other) AGE OF SYSTEM STATUS (Active / Inactive) Page | 24 Radiation Protection Office 80 East Concord St. Boston, Massachusetts 02119-2511 Tel: 617.638.7052 Fax: 617.638.7509 SECTION C. SOP(S) Please attach a copy of proposed SOPs for each unit (form XR-3). SECTION D. X-RAY USERS Please attach an X-ray User Authorization form for each X-ray user (form XR-2). I agree to fully comply with the safety requirements outlined by the Massachusetts Department of Public Health (105 CMR 120.600) and the Boston University/BMC Analytical X-ray Safety Manual. I will operate all analytical X-ray equipment in a safe manner, and I will only operate the equipment for which I have been permitted by the Radiation Safety Committee. Permit Holder Signature: Date: RSO: Date: Page | 25 APPENDIX C Boston University/ Boston Medical Center XR-2 X-ray User Certification Form Analytical X-Ray User Certification Form X-Ray Permit Holder: ___________________________________________________ Lab location and room: _____________________________ Lab telephone number: __________________ Instructions for Completing the X-Ray User Certification Form: 1. Read the responsibilities and certification items listed below. 2. Ensure all conditions of certification have been met. 3. Ensure each person listed provides a signature upon completion of conditions Your responsibilities as an X-Ray Analytical Equipment user are: 1. You must complete an initial Analytical X-Ray Safety training on line via RIMS. 2. You must familiarize yourself with the location and content of your X-Ray Analytical Equipment permit. 3. You must use X-Ray Analytical Equipment at designated locations approved in the permit. 4. Your work must be performed in the manner specified in the SOP (standard operating procedures). There shall be no changes in the approved procedures without the prior approval of the RSO. 5. You are only allowed to use the X-Ray Analytical Equipment system(s) listed on your permit that you have been trained on. 6. You must notify the RSO/CMP of new purchases of any analytical X-Ray systems and register these systems with the RSO/CMP. Routine operation of analytical X-Ray systems may not begin until the RSO has been notified and has conducted a thorough survey and given approval for operation. 7. You must notify the RSO/CMP of any analytical X-Ray system transfer or disposal. 8. You must perform any maintenance or repair to your X-Ray Analytical Equipment system unless proper documentation is presented to the RSO. 9. If you will be aligning an open beam system, you must possess and wear dosimeters provided to you by the DMPRS. 10. You must read, understand, and follow all safety requirements specific to use of your X-Ray systems. 11. You must know how to respond, report, and who to contact in case of an emergency involving an X-Ray system. I certify that: I have read, understood and agree to above stated laser user responsibilities. I have completed the on-line Analytical X-Ray Safety Training. I have been provided training specific to the laboratory and its X-Ray units by the PI. I have read and understood the information specific to laser safety requirements and the laboratory SOP. I will follow all laboratory safety procedures at all times. I will report any accident, potential exposure or safety concerns to my supervisor immediately and the RSO. Signature: __________________________________ Print name: _________________________________ Date: ____________________ BU ID: ___________________ If you need additional information, please contact the Department of Medical Physics and Radiation Safety at 617- 638-7052. Note: Violations of health and safety requirements are considered as serious infractions that may result in the suspension and/or termination of the protocol or an individual’s privileges to work with X-Ray systems. Page | 27 APPENDIX D Boston University/ Boston Medical Center Form XR-3 Template SOP Page | 28 XR-3 Boston University/Boston Medical Center Division of Medical Physics and Radiation Safety SOP for Title Location Contact X-ray Permit Holder: X-ray Supervisor: Office Phone # Emergency Phone # Revision History Revision A Description of Changes Effective Date Author Initial Release: Approvals Name and Title Signature Date Page | 29 A. EQUIPMENT INFORMATION 1. X-ray specifications Type: (diffraction, etc.) Manufacturer: Manufacturer Information: Manuf. Date: Model: Serial #: BU Property #: Status: Installation date: Max. Power: Model: Serial #: Calib. Date: Probes: Model: Serial #: Calib. Date: 2. Survey meter specifications Manufacturer: 3. Dosimetry X-ray users shall wear the whole body dosimetry and / or finger rings as assigned by DMPRS Do not remove badges from the lab or leave badges in a high exposure area. Page | 30 B. GENERAL SAFETY REQUIREMENTS 1. Exposure Risk Do not expose any part of the body to the primary radiation beam. The greatest potential for acute exposure occurs during the manipulation of the sample to be irradiated. Exposure rates of up to 10,000 R/sec can occur at the housing ports of certain X-ray units... Signs of exposure include erythema (reddening) of the skin in less than 0.3 seconds. Serious and permanent injury may occur if exposure lasts more than 0.1 seconds. Scattered radiation can also produce high exposure rates (100s of mr/hr). Immediately report any suspected exposures to the X-ray Permit Holder (xxx-xxxx), the DMPRS (638-7052), and the Research Occupational Health Program (Call the Research Occupational Health Program (ROHP) at 617-414-7647. Boston Medical Center Emergency Department on the Medical Campus 617-414-4075 or go to the nearest hospital emergency room ) 2. Access Controls Describe access controls, including use of interlocks. It is the responsibility of the user to maintain and enforce access control. Access to X-ray lab via card/key entry system. X-ray housing depresses interlocks. Removing screws will defeat interlocks. Do not override interlocks. Interlocks may be defeated for alignment purposes only with permit holder’s approval. 3. Warning Signs & Lights “CAUTION: X-RAY RADIATION. AUTHORIZED PERSONNEL ONLY” signs are posted above the X-ray Unit and on doors leading to any restricted area. “CAUTION X-RAYS—THIS EQUIPMENT PRODUCES X-RAYS WHEN ENERGIZED” signage is posted near any switch that energizes the X-ray tube. “CAUTION: HIGH INTENSITY X-RAY BEAM” signage is affixed to the X-ray source housing. WARNING LIGHTS: Each X-ray unit has a fail-safe warning light to indicate when the X-ray tube is energized. Page | 31 4. Authorized Users: X-ray Users: Authorized users include only those personnel listed in the X-ray Safety Manual / X-ray User Authorization Records. X-ray Users must have: (a) completed the Initial X-ray Safety Training provided by Boston University's Office of Environmental Health and Safety, (b) enrolled in the applicable dosimetry program at BU, and (c) had specific, documented training on this X-ray's SOP. Users are responsible for reviewing and complying with all applicable rules, regulations and University procedures as explained in the BU X-ray Safety Manual and other BU X-ray Safety Program documents. An “X-ray User Authorization Form” for each authorized user shall be documented in the X-ray manual using form XR-2. 5. Alignment procedures: All X-ray users need to keep in mind that the majority of X-ray accidents occur while aligning the X-ray beam. All possible steps will be taken to prevent any such accidents. Alignment procedures are performed in accordance to the manufacturer’s instructions. Notify supervisor prior to commencing with alignment procedures. Notify lab personnel of initiation of alignment procedure. Ensure that there are no unauthorized occupants in the area. Complete “Safe Operating Procedures” as itemized in part C.1. Operator will not leave the area until the X-ray housing is replaced and the interlocks are operational. 6. Actions: Prior to operating the X-ray unit, verify that the following actions have been implemented. Notify the permit holder prior to operating the X-ray equipment if there are any “NO” responses. ACTION YES NO ACTION YES NO Are safety devices operating Are all signs and labels in place as properly? (interlocks, beam noted in B.3? cut-offs) Are the X-ray on/off light(s) Is a survey meter available & functioning? calibrated? Are unused ports on the Is a current written SOP available? radiation source closed? Is the lab secure? Is a written Use log available & in use? Are all necessary interlocks operating properly? Page | 32 C. SAFE X-RAY OPERATING PROCEDURES 1. Prior to operation of the ___________ X-ray System, each of the following safe operating steps shall be implemented. Post section C.1. of the X-ray SOP, “Safe X-ray Operating Procedures” in close proximity to the X-ray Unit. This list of safe operating procedures shall be maintained in clear view of the operator. POWER ON SEQUENCE Record the date and start time on the X-ray Use log, form XSP-3. Perform a radiation survey if required (see instructions on Use log). POWER OFF SEQUENCE (Non-Emergency) Enter stop time and date on the X-ray Use Log, form XSP-3. (Emergency) Hit the emergency power off button for this unit or the entire lab (See floor plan reference # XXXX). 33 Page | 2. Floor plan Attach a diagram of X-ray use area. (A simple block diagram will do. Block diagram should also be posted in a prominent location, either on the lab door or inside the laboratory.) 3. Photograph of X-ray unit: Page | 35 4. Beam Configuration: Yes No N/A Does this device have a configuration such that any portion of a person’s body can enter the primary X-ray path? Is there a functioning safety device that will shut off the beam upon entry to the path? If there is no such safety device, has an exemption been issued by the MDPH? Is there a readily discernable indication of X-ray “on / off”? Is the “on / off” indicator located near each port on the radiation source housing? Is there a shutter “open / closed” status indicator located near each port on the radiation source housing? For open beam devices installed after February 1995, does each port have a shutter that cannot be opened unless a collimator or coupling has been connected to the port? 5. Non-beam hazards, if any: List and describe how they have been addressed (electrical, trip & fall, etc.) Hazard Response Action 6. Hazardous material handling: List hazardous materials involved, such as chemical or biological, including the quantity, handling and labeling requirements. Hazardous Material Quantity Handling and Labeling Requirements Page | 36 7. Hazardous Waste: Identification of hazardous waste generated by the activity (if any) and its disposal. Waste Identification Disposal Requirements 8. Emergency Procedures: Authorized X-ray users will be familiar with the building’s emergency evacuation plan, location of emergency equipment, and emergency procedures for fires or other emergencies. Emergency shut off procedures for lasers consist of shutting off electrical power to the X-ray system. Note locations of main electrical shut off switches to the X-ray. Power off X-ray via: On/Off switch on X-ray control panel Emergency power off (EPO) button on wall. (Reference specific EPO that shuts down this equipment from floor plan.) Follow specific instructions for each piece of equipment. Circuit breaker box Follow BU Safety/ Emergency Instructions on flip chart if there is a need to contact Police, Fire or Ambulance Emergency Response personnel. (BUPD 353-2121) 9. Maintenance: Only OEHS authorized, licensed and certified outside service personnel will maintain the equipment. A current copy of the MDPH registration information for all outside contractors must be on file at OEHS prior to any service. All relief devices, safety interlocks, alarms and other hazard prevention devices will be maintained, calibrated, and tested for functionality on a regular basis in accordance with standard industrial practices and recommendations of the manufacturers. Service Agreement with: Date of last maintenance inspection: Page | 37 APPENDIX E Boston University/ Boston Medical Center XR-4 Analytical X-ray Restricted Laboratory Inspection Checklist XR-4 Boston University/Boston Medical Center Division of Medical Physics and Radiation Safety X-ray Radiation Restricted Laboratory Inspection Checklist Building Address: Permit Holder: Inspected By: Survey Meter Model: Room(s): Serial #: Date: Calib. Date: NOTE: S = Satisfactory; U = Unsatisfactory; NA = Not Applicable. S U NA Item GENERAL X-RAY RADIATION SAFETY 1. Security: Doors are closed & locked when unoccupied; closed when person is physically in that specific room. 2. Personnel respond to requests for bioassays from DMPRS. 3. Personnel are working with X-ray sources in accordance with ALARA. 4. Radiation survey of representative areas result in dose rates < 0.03 mr/hr. 5. X-ray sources stored properly, labeled, locked if necessary. 6. All personnel are registered with DMPRS and training is current. 7. Applicable dosimetry worn when X-ray equipment is in use, and stored on dosimetry rack. RECORDS 8. Radiation surveys are performed as required and records are filed with the X-ray Safety Manual. 9. Physical inventory of equipment is current. 10. Records available for transfer of X-ray equipment to / from other institutions. 11. X-ray Safety Manual has completed User Authorization Form for each user. 12. A current written SOP is posted. 13. Use logs are up to date and posted. 14. A current permit, including an accurate list of Authorized Users, is posted. 15. Service technicians are registered with MDPH and approved by RSO. Comments Page 2 X-ray Radiation Restricted Laboratory Inspection Checklist S U NA Item FACILITIES & EQUIPMENT 16. Egress unobstructed. 17. Appropriate shielding in use. 18. Safety equipment unobstructed. 19. Exits are identifiable. 20. EPO is present and functioning. 21. X-ray generator has protective cabinet that limits leakage radiation as specified. 22. Fire extinguishers are inspected annually & tagged. 23. Survey meters: proper type, adequate number and calibrated annually. POSTINGS, SIGNS, LABELS & WARNINGS 24. X-ray Safety Manual available. 25. “Caution: X-ray Radiation. Authorized Personnel Only.” sign posted on door. 26. “Caution X-rays. This equipment produces X-rays when Energized.” Posted near “ON” switch. 27. MDPH registration is posted and current. 28. “Caution- High intensity X-ray beam” affixed to source housing. 29. Fail safe warning light(s) indicate when the X-ray tube is energized present & functioning. 30. MRCP Form 120.750-1 “Notice to Employees” posted. POSTINGS, SIGNS, LABELS & WARNINGS 36. Emergency contact flip chart posted. 37. Copy of BU X-ray Permit available. 38. Shutter “OPEN / CLOSED” status indicator near each X-ray port. 40. All areas used for X-ray equipment are clearly marked. Comments APPENDIX F Boston University/ Boston Medical Center XR-5 X-ray Equipment Inventory Form XR-5 Boston University/Boston Medical Center Division of Medical Physics and Radiation Safety X-RAY EQUIPMENT INVENTORY Last Update: X-ray Permit Holder: Office Phone: Department: Lab Phone: Home Phone: Alternate Contact: Office Phone: X-RAY MANUFACTURER X-RAY MODEL Bldg. Address: Email: Email: Lab Phone: SERIAL NUMBER EHS REF NUMBER Room #s: Home Phone: BU PROPERTY ID X-RAY TYPE (diffraction, spectrographic, mammography) POWER / ENERGY OUTPUT APPLICATION INSTALLATION DATE AGE OF SYSTEM OWNERSHIP (BU, other) STATUS (Active / Inactive) XR-5 Boston University/Boston Medical Center Division of Medical Physics and Radiation Safety X-RAY EQUIPMENT INVENTORY Page 2 X-RAY MANUFACTURER X-RAY MODEL SERIAL NUMBER EHS REF NUMBER BU PROPERTY ID X-RAY TYPE (diffraction, spectrographic, mammography) POWER / ENERGY OUTPUT APPLICATION INSTALLATION DATE AGE OF SYSTEM OWNERSHIP (BU, other) STATUS (Active / Inactive)