XAD-8 & XAD-4 Resin Cleaning Procedure (New Resin) * This procedure is for cleaning new XAD-8 and XAD-4 resin which is often stored in methanol Supplies needed 1 L erlenmeyer flasks ~ 0.1 N NaOH (enough to match the amount of resin to be cleaned) Procedure 1) Rinse resin several times with hot deionized water (~60 oC) to drive off methanol 2) Place resin in a 1 L flask 3) Add 0.1 N NaOH in excess, swirl and let sit for at least 12 hours. 4) Swirl and decant NaOH Allow fines to be decanted since you want to end up with a uniform range of resin size 5) Add fresh NaOH and repeat procedure 3 & 4 for at least 3 days (six decants) 6) Decant NaOH and rinse with MQ water several times 7) Store in MQ water Resin Cleaning Procedure (Used Resin) * This procedure is for cleaning used XAD-8 and XAD-4 resin * If resin is dried out, discard and do not clean Equipment & Supplies needed 2 – 1 L erlenmeyer flasks Methanol (HPLC grade) Acetonitrile (HPLC grade) Cleaning apparatus (Soxhlet extractor) Condenser Clamp 1 Resin cleaning vessel Clamp 2 Resin holding sleeve Cleaning agent holding vessel Heater 02/16/16 1 533563252 Figure 1 Cleaning Procedure 1) Remove any contaminants and resin that is of the other kind (this can also be done after cleaning) 2) If resin has significant sorbed humic substances (brown color), wash with 0.1 N NaOH. 3) Loosen “Clamp 1”, raise the “Condenser”, re-clamp 4) Loosen “Clamp 2” and remove “Resin Cleaning Vessel” 5) Fill “Cleaning Agent Holding Vessel” with methanol to just below the white oval (Figure 2) Fill line Figure 2 6) Clean “Resin Cleaning Vessel” and “Resin Holding Vessel” with methanol squirt bottle. 7) Transfer resin to “Resin Holding Thimble” Check resin level by inserting “Resin Holding Thimble” into the “Resin Cleaning Vessel”. Do not fill the “Resin Holding Thimble” so level is above bubble in glass (Figure 3). 8) Rinse “Resin Holding Thimble” 2-3 times with methanol squirt bottle and reinsert. Liquid fill line Max. resin fill line Resin holding sleeve Overflow/siphon tube Cleaning agent holding vessel 02/16/16 2 533563252 Figure 3 9) Reassemble “Cleaning Agent Holding Vessel”, “Resin Cleaning Vessel” and “Condenser” Make sure the seals are a tight fit. 10) Make sure water is running through the “Condenser”. This should be at a low flow rate to conserve water. The flow may change slightly after a few minutes, so be sure to recheck it. 11) Turn heater to 3.5-4.0 units and wait 10 min. for methanol to be heated 12) Adjust heat so drip rate out of the “Condenser” is ~1 drop/sec. 13) Make sure methanol rises up to the “Liquid fill line”, thus saturating the resin (Figure 3). The liquid level of methanol should be at the top of the overflow/siphon tube. If this does not occur, see if the methanol is condensing in the top corners of the “Resin Cleaning Vessel” and dripping down the larger outside tube. If it is, rotate the setup so the methanol misses the outside tube as it drips down the side. 14) Clean with methanol for at least 48 hours 15) Discard methanol at end of cleaning 16) Repeat steps 2 – 14 with acetonitrile instead of methanol 17) Rinse resin in “resin holding vessel” several times with MQ 18) Transfer resin to an erlenmeyer flask and rinse several times with hot MQ water (~60 oC) 19) Store cleaned resin in MQ water 20) Wash resin cleaning apparatus with methanol to prepare for next run 02/16/16 3 533563252