UNITS OF MEASUREMENT 1mm = 1000µm 1µm = 10-3mm (convert mm to µm by multiplying by 1000 = 3 zeros) 1nm = 10-6mm (convert nm to mm by dividing by 1000000 = six zeros) 1000nm = 1µm 0.001µm = 1 nm Protozoa are fairly large single-celled animals. Bacteria are so small, they are measured in µm. Viruses are even smaller, so they are measured in nm. MICROSCOPY: THE INSTRUMENTS VOCABULARY Immersion oil: keeps light from bending and allows lens to be refracted. Resolution: ability of two lenses to distinguish two points. Parfocal: focused in all lenses. Depth of field: how much of the background is in focus at the same time that the foreground is in focus. Refractive Index: a measure of the light-bending ability of a medium Numerical aperture: numerical aperture increases as depth of field decreases. Resolution power: limits the useful magnification of the microscope resolving power. RESOLUTION: the ability of the lenses to distinguish two points. A microscope with a resolving power of 0.4 nm can distinguish between two points greater than or equal to 0.4nm. Shorter wavelengths of light provide greater resolution. When you go to the eye doctor, you look at the chart (Snellen chart) and read it from 20 feet away. If you can read what a normal sighted person can read from 20 feet away, it is called 20/20 vision. If you can’t read it well, your eyesight has less resolution than normal. RESOLVING POWER: the distance between two closely adjacent objects where the objects still appear separate and distinct. The shorter the distance (the smaller the number), the better the resolving power (the sharper the image). To calculate the resolving power (to see how close two objects can be so you can still see them): D = distance (smaller number is better) 0.61 = a constant number, does not change NAobj = numerical aperture of the objective (larger number causes D to decrease, which is better) λ = (lambda): the wavelength (nm) of the light going through the microscope. Convert lambda’s nm to mm by moving the decimal place six places to the left. D = 0.61 λ / NAobj PRISM: a triangular device that breaks up light into its various wavelengths so you can see all the colors of the rainbow (the visible spectrum). The wavelengths of the visible spectrum range from 350-700nm. The visible spectrum of colors starts with violet (350nm), and goes on to indigo, blue, green (550nm), yellow, orange, red (700nm). These colors are picked up with special photoreceptor cells in the human retina called cones, and the information is transmitted to the brain where we perceive the color. Sample Problem When we want to observe the color green (550nm) under an oil-immersion objective lens of a microscope, where the NAobj is 1.25, what is the resolving power (in microns)? Solution * Convert 550 nm to mm by moving the decimal place six places to the left: 550nm = 0.000550 mm D = (0.61)(0.000550mm) / 1.25 D = 0.0002684mm convert this to microns (µm) by multiplying by 1000 D = 0.2684 µm round off to the nearest 1/100 (two decimal places to the right) D = 0.27 µm Sample Problem The NAobj for the high-dry (400x) lens is 0.65 What is the resolving power (D) of this objective when viewing a wavelength of 550nm? Solution D = (0.61)(0.000550) / 0.65 D = 5.1615 mm (Now multiply by 1000 to convert to microns) D = 0.52 µm Therefore, we can see an organism such as E. coli, which is 2µm long and 1µm wide because it is larger than the resolving power. However, we could not see Haemophilus influenza, which is 0.2µm long because it is smaller than our resolving power. Therefore, the resolving power limits the useful magnification of the microscope. Resolution determines the magnification. REFRACTION Refraction is the bending of light caused by the surrounding medium. When you stick a pencil in a glass of water and look at it from the side, the pencil appears to be bent. That is because the water is refracting the light. We need to take that into consideration when viewing things under oil magnification with the microscope. N = Refraction Index of the medium surrounding the lens Air: N= 1 Glass: N = 1.5 Immersion Oil: N = 1.51 (about the same as glass) TYPES OF MICROSCOPES SIMPLE MICROSCOPE: Has only one lens, like an ocular (eyepiece) COMPOUND MICROSCOPE: More than one lens, like an ocular and an objective. An example is the Brightfield microscope. There are two main types of compound microscopes: Light Microscopes and Electron Microscopes. Light Microscopy Light microscopes have an RP of about 0.25 µm, about the size of the very largest viruses and the very smallest cells. They extend the resolving power of the human eye by about 1000x. One major advantage of light microscopes is their low cost and relative ease of use. Another is that it is possible to view living tissue in some microscopes, especially with phase optics. For many applications, light microscopy is the preferred tool, such as in screening tissues for signs of cancer or other pathology. A disadvantage of light microscopes is that they do not have good depth of field; when you look at the foreground, the background is out of focus. Light microscopes can have a variety of configurations. Some popular types are: Dissecting Brightfield Darkfield Phase-contrast Differential Interference contrast Fluorescence DISSECTING MICROSCOPE This type of microscope is for viewing in 3D objects that are very large (visible to the naked eye) but in greater detail than the naked eye can see. It shows a view like a powerful magnifying glass. It typically magnifies 20x to 40x. It requires an external light source (plug in a lamp and face it to the scope). BRIGHTFIELD ILLUMINATION Most student microscopes (including the ones we use in lab) are Brightfield microscopes, requiring that objects to be viewed either have natural contrast or be stained to provide sufficient contrast for viewing. Some stains, called vital stains, can be used on living tissue; most require the sample to be dried and fixed. In general, the Brightfield microscope is not very good for looking at live cells (without stain) because the contrast is not good; cells look white on a white background. You can attempt to look at live cells with this microscope by opening up the iris all the way to let all the light in to improve contrast. This microscope is very good for looking at stained cells. When you apply stain to cells it kills them, but the contrast is improved and you can see them well. This type of light microscope produces the most realistic images of a cell, compared to other light microscopes. DARKFIELD ILLUMINATION This has limited applications, but it is good for looking at live cells because the cells look white against a dark background. You can even see the cilia and flagella. PHASE CONTRAST MICROSCOPY This is the most sensitive type of microscopy. Live cells are visible because contrast is increased. You can even see cilia and flagella. DIFFERENTIAL INTERFERENCE CONTRAST This shows depth in 3D. It has two beams of light and a prism. By moving the prism, you see different colors. This is a good way to view live cells; it has the best resolution of the light microscopes. To get better resolution, you need to use electron microscopy. FLUORESCENCE MICROSCOPY In this technique, ultraviolet light (invisible) is used. Cells are stained with a fluorescent dye call fluorochromes, which energizes electrons to create visible light. You can see the light given off by cells. It is not possible to see live cells with this. It is often used for viewing syphilis and TB. ELECTRON MICROSCOPY Electron microscopes use a beam of electrons rather than light to illuminate an object. Magnification of up to a million-fold is possible. Creating an electron beam requires very high voltages, typically 100,000 volts, and the electron beam works well only in a vacuum, so air must be removed prior to specimen examination. Electrons are not easily scattered by the light atoms of common cellular materials: C, H, O, and N. To improve electron scattering and increase contrast, heavy metals are used, either as coatings over exposed surfaces or as stains applied to tissues. The primary advantage of all electron microscopes is the added resolving power and the extraordinary clarity of the images. Disadvantages include: the cost and complexity of the equipment; the need to isolate them from vibrations (vibrations from a passing truck can be easily visible when viewing an object at 1,000,000x) the inability to study living tissue; the very thin samples required; and the need for heavy metals to obtain good contrast, with possible damage or distortion of the sample. The two major forms of electron microscope, TEM and SEM, are very different. One views surfaces or "outsides," the other views "insides." Together, they reveal an extraordinary world of very small objects. Transmission Electron Microscope (TEM) Transmission electron microscopes (the most common types of electron microscopes) transmit a beam of electrons through the sample. To achieve this, samples must be very small and very thin. Therefore, the specimen is sliced thin using an instrument called an ultramicrotome. Most samples are mounted on tiny circular grids with a very fine mesh screen. The grid is coated with a thin plastic layer, and sample is either adsorbed directly (as in a virus sample) or prepared as an ultra thin section of tissue that has been fixed, stained, dehydrated, and embedded in resin to form a hard block of plastic. This gives a view of sections of the specimen’s cells, revealing the structures of its organelles. This type of microscope uses electromagnetic lenses and creates a flat image, and since there is no light, there is no color. Electrons float better in a vacuum, so it has a vacuum pump. Staining is done with heavy metal salts instead of dyes. Areas with low levels of impregnation of these metal salts produce bright regions on the screen, and heavily impregnated areas produce dark ones (the electrons can't get through). This improves how thing look with this microscope. Advantages of TEM are the ability to view objects at a very high level of resolution, and to see the insides of common biological materials. Disadvantages include the limit of only viewing a small sample at one time, and the need for very thin specimens. If the specimen is thicker than 0.25 µm, too many electrons are adsorbed to obtain a good image. The TEM and the SEM have the highest resolution of all the microscopes. Scanning Electron Microscope (SEM) Scanning electron microscopes work by scanning a beam of electrons across the surface of an object. Each area so illuminated discharges a small shower of electrons, called secondary electrons. These are trapped by a detector that converts electrons into light signals, which in turn are trapped by a photomultiplier and used to drive a cathode-ray tube (similar to a television tube). The resulting image can be viewed and photographed. A unique feature of an SEM is its great depth of focus; objects in both foreground and background appear equally sharp and crisp, giving the best 3D image because it has the highest resolution. By contrast, light microscopes have a very shallow depth of focus, so as one moves to high-power lenses, only a small part of the image is in focus at any one time. Another advantage of SEM is that sample size can be very large. For example, an entire insect or small flower can be viewed at one time. The SEM gives the most realistic view of a specimen. This provides a view of the outside of the cell only; it cannot penetrate to the inside of cells like the TEM. It also needs a vacuum, and cells are never alive. Its resolution is 10x better than a Brightfield microscope. A special type of SEM is called the Scanning Probe Microscope. Scanning Probe Microscope This microscope forms images of surfaces using a physical probe that scans the specimen. An image of the surface is obtained by mechanically moving the probe in a raster scan of the specimen, line by line, and recording the probe-surface interaction as a function of position. A Raster scan, or raster scanning, is the pattern of image detection and reconstruction in television, and is the pattern of image storage and transmission used in most computer image systems. The word raster comes from the Latin word for a rake, as the pattern left by a rake resembles the parallel lines of a scanning raster. In a raster scan, an image is cut up into successive samples called pixels, or picture elements, along scan lines. Each scan line can be transmitted as it is read from the detector, as in television systems, or can be stored as a row of pixel values in an array in a computer system. On a television receiver or computer monitor, the scan line is turned back to a line across an image, in the same order. After each scan line, the position of the scan line is advanced, typically downward across the image in a process known as vertical scanning, and a next scan line is detected, transmitted, stored, retrieved, or displayed. One advantage of scanning probe microscopy is that the resolution of the microscopes is not limited by diffraction, but only by the size of the probe-sample interaction volume. Disadvantages of scanning probe microscopy are that the scanning techniques are generally slower in acquiring images due to the scanning process, and the maximum image size is generally smaller. This type also gives a 3D view of the specimen. COMPARISON OF MICROSCOPES BRIGHTFIELD DARKFIELD PHASECONTRAST DIFFERENTIAL INTERFERENCE CONTRAST FLUORESCENCE TRANSMISSION ELECTRON SCANNING ELECTRON SCANNING PROBE Dark objects are visible against a bright background. Light reflected off the specimen does not enter the objective lens Not for looking at live cells Maximum resolution is 0.2µm and maximum magnification is 2000x Stains are used on specimens Light objects are visible against dark background Used for live cells, cilia, flagella Especially good for spirochetes Uses special condenser with an opaque disc that eliminates all light in the center No staying required Accentuates diffraction of the light that passes through a specimen Good for live cells; good contrast Most sensitive; cilia shows up Not three-dimensional Uses two beams of light Shows three dimensions Has a prism to get different colors Good for live cells (unstained) Best resolution Uses ultraviolet light Stained cells with fluorescent dye; energizes electrons and creates visible light No live cells Quick diagnosis of TB and syphilis Get flat images Have vacuum pumps to allow electrons to float better Stain with heavy metal salts Shows sections of cell, revealing organelles Requires an ultramicrotome Best resolution of all microscopes Surface view only Needs a vacuum No live cells Three-dimensional view Physical probe scans the specimen Raster scan: image is cut up into pixels and transmitted to computer Not limited by diffraction Slower in acquiring images Maximum image size is smaller