Rev 02-2010 TissueSAFE High Vacuum Biospecimens Transfer System i Operation Manual (MM074) October 2010 TissueSAFE Operation Manual 1/60 Rev 02-2010 Thank you for having selected our system and welcome to the ever growing club of satisfied Milestone users. We invite you to read this operator manual carefully and to keep it in reach for convenient and fast consultation. For any possible clarification or any request for assistance please contact either our Representative in your country : Or: Milestone Srl Via Fatebenefratelli, 1/5 24010 Sorisole (BG) Italy Tel. +39-035-4128264 Fax +39-035-575498 Web site www.milestonemedsrl.com E-mail medical@milestonesrl.com Caution! Please read the user manual before using the instrument! TissueSAFE Operation Manual 2/60 Rev 02-2010 SUMMARY 1. INTRODUCTION .............................................................................................................................. 4 1.1. IMPORTANT NOTES ................................................................................................................ 4 SYMBOLS USED IN THIS MANUAL ................................................................................................ 4 DESIGNATED USES ........................................................................................................................ 4 1.2. SAFETY ..................................................................................................................................... 5 2. SETTING UP THE INSTRUMENT.................................................................................................... 6 2.1. SITE REQUIREMENTS ............................................................................................................. 7 2.2. ELECTRICAL REQUIREMENTS .............................................................................................. 7 2.3. UNIT OVERVIEW ...................................................................................................................... 8 2.4. SWITCH ON AND OFF .............................................................................................................. 9 2.5. FUMES EXTRACTION SYSTEM............................................................................................. 10 3. MILESTONE VACUUM BAGS ........................................................................................................ 11 4. OPERATING TISSUE SAFE........................................................................................................... 13 4.1. CONTROL PANEL .................................................................................................................. 13 4.2. USER LOG IN .......................................................................................................................... 14 4.3. SELECT A PROGRAM ............................................................................................................ 15 4.4. RUN A PROGRAM .................................................................................................................. 15 4.5. MODIFY A PROGRAM ............................................................................................................ 21 4.6. STOP A PROGRAM ................................................................................................................ 23 4.7. SEAL A PLASTIC BAG ........................................................................................................... 23 4.8. SERVICE MENU ...................................................................................................................... 24 4.8.1 Watch ............................................................................................................................... 24 4.8.2 Language .......................................................................................................................... 24 5. ACCESSORIES .............................................................................................................................. 25 5.1. NON MOBILE REFRIGERATOR ............................................................................................ 25 5.2. THERMAL BAG....................................................................................................................... 26 5.3. PORTABLE REFRIGERATOR 26 LITRES ............................................................................. 28 5.4. KIT PORTABLE REFRIGERATOR 65 LITRES ...................................................................... 30 5.5. KIT PRINTER........................................................................................................................... 32 5.6. BAG OPENER ......................................................................................................................... 32 5.7. TEMPERATURE MONITORING SYSTEM .............................................................................. 33 5.8. HOW TO OPERATE THE TEMPERATURE MONITORING SYSTEM ................................... 41 6. CONSUMABLES ............................................................................................................................. 49 6.1. POLYSTYRENE CONTAINERS WITH PERFORATED LIDS ................................................. 50 6.2. POLYSTYRENE CONTAINING TRAYS ................................................................................. 50 7. MAINTENANCE .............................................................................................................................. 51 7.1. CARBON/HEPA FILTER REPLACEMENT............................................................................. 52 8. WASTE DISPOSAL ........................................................................................................................ 53 8.1. WASTE DISPOSAL OF VACUUM BAGS ............................................................................... 53 8.2. WASTE DISPOSAL OF PROTECTION BAGS FOR PDA ...................................................... 53 8.3. WASTE DISPOSAL OF ELECTRICAL AND ELECTRONIC EQUIPMENT ............................ 53 APPENDIX A .......................................................................................................................................... 54 DECONTAMINATION PROCEDURE .................................................................................................... 54 SAFETY PRECAUTIONS ................................................................................................................... 54 DECONTAMINATION PROCEDURE ................................................................................................ 55 APPENDIX B .......................................................................................................................................... 59 APPENDIX C .......................................................................................................................................... 60 TissueSAFE Operation Manual 3/60 Rev 02-2010 1. INTRODUCTION 1.1. IMPORTANT NOTES SYMBOLS USED IN THIS MANUAL An instruction accompanied by this symbol provides a cautionary statement: WARNING. Failure to follow the instruction may endanger the user or cause damage to the instrument. An instruction accompanied by this symbol provides a caution against electric shock. Be sure to follow the instruction to prevent accidents resulting from electric shock. Biohazard Warning, danger of crushing hand CE logo: This instrument complies with European community directives IVD medical device according to 98/79/EC directive Please read carefully this instruction. Please read Service Manual (operation for service only). WEEE European directive symbol (2002/96/EC): Electric/electronic device do not throw out in the environment (3) Figures in brackets refer to item nos. in drawings DESIGNATED USES The Milestone TissueSAFE system is intended for the following laboratory applications: Histological specimen preservation Histological specimen transportation Long term specimen preservation (with an appropriate fixative into the plastic bag) The Milestone TissueSAFE must be operated exclusively as reported in this manual. Any other use of the instrument is considered improper! TissueSAFE Operation Manual 4/60 Rev 02-2010 1.2. SAFETY This paragraph details important safety information. Please read this section carefully: If additional requirements on accident prevention and environmental protection exist in the country of operation, this instruction manual must be supplemented by appropriate instructions to ensure compliance with such requirements. Do not remove any instrument and/or accessory panel or cover. Only authorized and qualified service personnel may repair the instrument and access the instrument’s internal components. It is important that normal standards of safety and good laboratory practices are employed. Always use common sense and the best known practice when operating the instrument. The institution owning the instrument has primary responsibility for accident-free operation together with designated personnel who operate, service or repair it. To avoid damage to the instrument, use only the plastic bags described in this manual, supplied by Milestone s.r.l. Waste disposal of plastic bags Plastic bags supplied by Milestone s.r.l. are preventively irradiated with B rays before the use. After their use, treat empty plastic bags as hazardous. Waste material should be disposed in an approved incinerator in compliance with all federal, provincial and local government regulations. TissueSAFE Operation Manual 5/60 Rev 02-2010 2. SETTING UP THE INSTRUMENT The Milestone TissueSAFE is a precision instrument that must be unpacked and installed with care. Move the TissueSAFE to its final location. Make sure that the floor is level and is made of non-flammable material. The instrument physical dimensions are: W 45cm (18”) x D 60cm (24”) x H 135cm (54”) Extra Space required : Back 15cm (6”), sides 5cm (2”) each TissueSAFE weight is approximately 74Kg (164lb) when empty. The Milestone TissueSAFE system, if dealing with Formalin fixed specimens, must be connected to the fumes extraction system of the Laboratory to guarantee a healthy environment. TissueSAFE Operation Manual 6/60 Rev 02-2010 2.1. SITE REQUIREMENTS Practically vibration-free floor Room temperature constantly between +10°C (50°F) and +40°C (104°F) Relative humidity maximum 80% (non condensing) Avoid vibrations, direct sunlight and heavy variation in temperature Floor must be able to support the weight (74Kg, 164lb) 2.2. ELECTRICAL REQUIREMENTS Check the voltage label on the back of the instrument to ensure that the instrument is set to the correct voltage range. The TissueSAFE requires 115Vac/230Vac 3kWp. The power supply voltage for the instrument is factory-preset and CANNOT be altered by the user. The Milestone TissueSAFE system must be connected to the power supply and grounded to guarantee the optimal performance of all the high voltage components. All Milestone units are assembled to be connected to the ground of the laboratory and operating room electrical system. Please contact our distributor to confirm that the power ground cable is securely grounded. TissueSAFE Operation Manual 7/60 Rev 02-2010 2.3. UNIT OVERVIEW 1 6 7 2 3 8 9 89 8 4 5 Cabinet front 1. Working area cover 2. Control panel 3. Plastic bags drawer 4. Metallic shelf 5. Cover (removable shelves holding vacuum bags inside) Holder PDA Cabinet back for Holder RFID (Temperature reader) 6. HEPA (high efficiency particulate air) filter, that satisfies the EN 1822 normative (filter class H12). 7. Charcoal/Formalin filter. 8. Connector for fume extraction system 9. Pump Oil level indicator Space for the plastic bags TissueSAFE Operation Manual 8/60 Rev 02-2010 2.4. SWITCH ON AND OFF The electrical plug is located on the back of the unit (see red square in the picture below). Connect the cable provided by Milestone S.r.l, as described in the following pictures: It is now possible to switch the unit on by pressing the main switch (A) on top of the control panel (see picture below). The same main switch (A) must be used to switch the unit off. A Make sure that the switch (A) is in OFF position before plugging the power supply cable. TissueSAFE Operation Manual 9/60 Rev 02-2010 2.5. FUMES EXTRACTION SYSTEM The unit can be connected to an external fume extraction system through the built-in connector (see red circle). TissueSAFE Operation Manual 10/60 Rev 02-2010 3. MILESTONE VACUUM BAGS Milestone provides the TissueSAFE system with plastic vacuum bags of four different sizes (not included, to be ordered apart), that are preventively irradiated with B rays. The use of the TissueSAFE system is allowed with Milestone proprietary plastic vacuum bags ONLY. P/N 68047SS P/N 68055SS P/N 68054SS P/N 68046SS Milestone S.r.l. provides these types of Milestone vacuum bags: P/N VACUUM BAGS Double welding for enhanced vacuum seal 68047SS Special vacuum bag 140μ - Size SMALL (Q.ty 200) Dimensions: external 250 x 300 mm – internal: 230 x 125 mm 68055SS Special vacuum bag 140μ -Size MEDIUM (Q.ty 200) Dimensions: external 340 x 385 mm – internal 320 x 150 mm 68054SS Special vacuum bag 140μ -Size LARGE- (Q.ty 200) Dimensions: external 340 x 550 mm – internal 320 x 320 mm 68046SS Special vacuum bag 140μ - Size EXTRA-LARGE (Q.ty 200) Dimensions: external 340 x 550 mm – internal 320 x 500 mm TissueSAFE Operation Manual 11/60 Rev 02-2010 An annotation label is placed on the outside of each bag. Normally such notes are taken by the nurses in the OR while the surgical operation is going on. At the end of the surgical operation, the MD can fill the request form and place it inside the appropriate pocket. Requisition form can be sealed inside the dedicated pocket. In this way requisition form and surgical left over are properly transported with no chance of mistakes. TissueSAFE Operation Manual 12/60 Rev 02-2010 4. OPERATING TISSUE SAFE When the TissueSAFE unit is switched on, the control panel will appear as shown in Fig 1. Fig 1 The operator can access the functionality of the unit through the panel, which controls all the parameters involved in the bio-specimens preservation process. 4.1. CONTROL PANEL The following picture describes the buttons on top of the control panel. Fig 2 1 2 6 3 4 5 Control panel description: 1. Force sealing. 2. Interrupt a vacuum process. 3. Program selection. 4. Minus. 5. Plus. 6. Display. TissueSAFE Operation Manual 13/60 Rev 02-2010 4.2. USER LOG IN When the system is switched on, the operator is logged in as BASIC USER by default. The BASIC USER is allowed to: Run a default program (from P1 to P4). Seal a plastic bag without vacuum, pressing the “” button (1) of fig. 2. Interrupt a vacuum process, pressing the “STOP” button (2) of fig. 2. Only ADMINISTRATOR (or SERVICE) can modify a preset program or create a new program. To log in as Administrator, reboot the unit, keeping pressed the “” button (1) of fig. 2. When the “Welcome” screen disappears, insert the password as explained below: Press: “+”: to increase the digit, “–“: to decrease the digit, “SEL”: to change the digit to insert, Press “SEL” again to confirm. ADMINISTRATOR password issued as a separate document included in this operator manual. TissueSAFE Operation Manual 14/60 Rev 02-2010 4.3. SELECT A PROGRAM TissueSAFE is able to memorize up to 30 different protocols. To move from P01 to another protocol, press the PLUS button (5) to move forward or the MINUS button (4) to go back to previous program (see fig. 2). The right programs selection must be done according to the Vacuum Selection Guide of the Operation Manual (see APPENDIX B). 4.4. RUN A PROGRAM After the user log in, the following screen appears: In the above picture the display is showing P1. It means that the unit is ready to start the vacuum program n.1. After the excision of the organ by the surgeon, it should be placed in the appropriate (by dimension) Milestone plastic bag (fig.3). The operator can use the designed shelf on the right of the vacuum cavity. Fig 3 Fig 4. The plastic bag with the specimen has to be placed with the open side directed toward the operator, exactly as shown in the picture above (Fig 4). TissueSAFE Operation Manual 15/60 Rev 02-2010 Make sure of using plastic bags of the appropriate size (Small/Medium/Large/Extra-large) according to the specimen dimensions. Use of inappropriate plastic bag size (too small) may cause spill out of the specimen. After the organ excision, in case of bleeding organ, remove the excess of blood (or any other organic fluid) with a medical cloth. This to avoid any putrefactive effect caused by the blood/organic fluid stagnation. The plastic bag must be placed with the metallic sealing bar between the two white lines (see picture below). The operator has to manually close the cover (fig. 5). When the cover is closed, the vacuum pump will start creating the vacuum inside the chamber (fig 6). Fig 5 Fig 6 TissueSAFE Operation Manual 16/60 Rev 02-2010 Human Organs : vacuum selection guide P1 P2 P3 P4 Kidney Lung Brain Cyst > 10cm in diameter Breast Placenta Eye Uterus Prostate Cyst <10cm in diameter Liver Gall bladder Muscle Thyroid Thymus Skin Tongue Spleen Colon Bowel Pancreas Bladder Testis This guide is intended for specimens directly packaged with vacuum sealed plastic bags. When sealing very soft specimens (For ex: Ovarian Cyst), we suggest the use of plastic containers P/N 68093-94 with perforated lids (see Par 6 for details). In this case the specimen dimension must not exceed 2/3 of the container inner volume. ONLY perforated lids can be used to seal specimens containers. TissueSAFE Operation Manual 17/60 Rev 02-2010 The vacuum progression inside the chamber is shown on the control panel display. TissueSAFE will reach the pre-set value of vacuum within 20 secs. Fig 7 After the vacuum step, the vacuum bag is sealed. Fig 8 On completion of the sealing procedure, an audible alarm will advise the user and the cover will automatically open (fig. 10). Fig 9 Fig 10 TissueSAFE Operation Manual 18/60 Rev 02-2010 Now the nurse of the Operating Room can annotate the patient data, writing on the label of the plastic bag (fig. 11). Fig 11 Fig 12 At the end of the surgical operation, the MD requesting the histological examination will fill the “Requisition Form”. This document needs to be placed in the external pocket on the right of the label where the nurse reported the first annotation (fig. 12). At this point, the MD can seal the requisition form in order to avoid any accidental separation from the specimen (fig. 13). Fig 13 The plastic bag containing the specimen needs now to be placed into an appropiate refrigerator and transported to the Anatomical Pathology Department under controlled temperature conditions, according to the following scheme. TissueSAFE Operation Manual 19/60 Rev 02-2010 On the base of your Hospital organization, the correct SPECIMENS TRANSPORT PROCEDURE is described in the following scheme: OPERATING ROOM SEALING TRANSPORT CONDITIONS MAX 30MIN HISTOLOGY LAB FRIDGE +4°C GROSSING FRIDGE +4°C GROSSING FRIDGE +4°C GROSSING THERMAL BAG COOL PACKS +4°C SEALING SEALING FRIDGE 2Hr ,+4°C MOBILE REF +4°C MAX 3 HOURS THERMAL BAG COOL PACKS +4°C > 3 HOURS MOBILE REFRIGERATOR +4°C 1. If the time required by the courier for the specimens’ transportation to the Anatomical Pathology Department is less than 30 minutes, Milestone S.rl. recommends putting the specimens in the Thermal bag with the cool packs at +4°C. 2. If the time required by the courier for the specimens transportation to the Anatomical Pathology Department is less than 3 hours, Milestone S.rl. recommends to put the specimens into the refrigerator ), close to the TissueSAFE for at least 2 Hr (pre-cooling procedure) and later to place them in the Thermal bag with the cool packs at +4°C. 3. If the time required for the specimens’ transportation exceeds 3 hours, Milestone S.rl. recommends the use of the Mobile refrigerator connected to an external power supply (12/24Vdc, 115/230Vac). For details on refrigerators models available from Milestone s.r.l., see Chapter 5 TissueSAFE Operation Manual 20/60 Rev 02-2010 Operator must take care that the total transportation time does not exceed 72 hours (including time spent in both O.R:/Histology Lab refrigerators). For a correct preservation of the specimens’ morphology, they have to be placed and maintained at a temperature of +4°C , ±2°C ( 39,2 °F, ±35,6°F). For further details about the complete TissueSAFE system and its accessories, see Paragraph 5 “ACCESSORIES”. 4.5. MODIFY A PROGRAM TissueSAFE can memorize up to 30 protocols. Only ADMINISTRATOR (or SERVICE) is allowed to modify programs as follows: Protocol P01–P03 P04–P30 Administrator User Allowed NOT allowed To modify a program, log in as Administrator (see Cap 4.2 for details). Select the program of interest using the button (4) and (5) shown in fig. 14 and press the “SEL” button (3) to open the program customization screen. Fig 14 1 2 6 3 4 5 TissueSAFE Operation Manual 21/60 Rev 02-2010 The program customization screen will appear as shown in Fig. 15. Move through the parameters using the “SEL” button (3) and set the relative value using the button (4) and (5) shown in fig. 14. Fig 15 Every program is characterized by 5 parameters : 1 VACUUM PERCENTAGE. Insert the correct vacuum value, according to the Organ Selection Guide (see APPENDIX B). 2 SEALING TIME. The plastic bags provided by Milestone s.r.l. are correctly sealed with a sealing time of 2,4 sec. It is strongly recommended NOT to modify this sealing time for use with standard bags provided by Milestone TissueSAFE Operation Manual 22/60 Rev 02-2010 3 PROGRAM ENABLED. If you choose: “Yes” the BASIC USER can use the protocol “No” the BASIC USER cannot use the protocol 4 PROGRAM NAME. It is possible to insert a short name for each program (generally according to the most used organ). You have 16 alphanumeric characters (“0”–“9”, “A”–“B”). Pressing the “SEL” button (3) once more, the system goes back to the main screen and the program is shown with the name you have selected TissueSAFE can now operate the program following the parameter inserted by the Administrator. The instrument is shipped by the manufacturer with 4 factory preset protocols (P1, P2, P3 and P4). From P5 to P30 default parameters are inserted and can be modified. Do not modify any TissueSAFE protocol without Milestone s.r.l. authorization. A protocol with inappropriate parameters may damage tissues. Do not leave the software logged in as Administrator, to prevent any accidental programs modification made by basic users. 4.6. STOP A PROGRAM Press the “STOP” button (2) on the control panel (fig. 14), to stop a program while the vacuum process is already started. The vacuum process will be immediately stopped and the vacuum chamber will be opened. 4.7. SEAL A PLASTIC BAG Press the “” button (1) shown in fig. 14, to seal the plastic bag (and to stop the vacuum cycle) while the process is running. The vacuum cycle will be immediately stopped, the plastic bag will be sealed and the vacuum chamber will be opened. TissueSAFE Operation Manual 23/60 Rev 02-2010 4.8. SERVICE MENU Service menu allows setting “WATCH” and “LANGUAGE” of the unit. Log in as Administrator (see Cap. 4.2 for details) and press the “” button (1) shown in fig. 14, to open the menu. Fig 16 Move through the parameters using the buttons (4) and (5) shown in fig. 14 and press the “SEL” button (3) to open the relative function. 4.8.1 Watch Set the proper value using the buttons (4) and (5) shown in fig. 14 and then press the “SEL” button (3) to confirm the setting and move to the next parameter: Day.....: 12 Month.....:01 Year......:10 Hour......:17 Minutes...:29 The time format is: 2010–01–12, 17.29. 4.8.2 Fig 17 Language Set the proper language using the buttons (4) and (5) shown in fig. 14 and then press the “SEL” button (3) to confirm the setting. Fig 18 On completion of the settings, press the “” button to exit the menu. TissueSAFE Operation Manual 24/60 Rev 02-2010 5. ACCESSORIES The specimens under vacuum must be stored at a temperature of +4°C (39,2°F). Milestone S.r.l. provides a set of optional accessories to maintain temperature under control, according to the Specimen Transport Procedure (see Chap. 4.4) : 5.1. NON MOBILE REFRIGERATOR P/N DESCRIPTION 68064 68090 Refrigerator laboratory type 140 Litres 230Volt complete Refrigerator laboratory type 140 Litres 115Volt complete Milestone suggests positioning the Fig 19 TissueSAFE close to a controlled refrigerator. In Fig 19, the refrigerator is shown on A the left side (A). In this type of refrigerator, the temperature is kept at approx +4°C (39,2°F). The operator can easily place the under vacuum B specimens and the cool packs (B) in the refrigerator. Main Characteristics: ECT-F Control® : with Hi/Low Temp. alarm, door ajar, ON/OFF w/password. Automatic condensate water evaporation. Internal cabinet with shelves/drawers interchangeable. Auto-defrosting type. For further information, see specific instruction manual included in the shipment. TissueSAFE Operation Manual 25/60 Rev 02-2010 5.2. THERMAL BAG P/N DESCRIPTION 68084 Kit Refrigerator box in PS 32 litres complete with N.4 cool packs When the Courier is ready to transport the specimens to the Anatomical Pathology Department, the under vacuum specimens must be placed in the Thermal Bag, appropriately closed and then transported (fig. 19 a/b). Fig 19a Fig 19b The operator must take care of placing at least 1, 2 or 3 Cool Packs at +4°C inside the thermal bag, according to the inner volume of the thermal bag used (fig. 20). Fig 20 TissueSAFE Operation Manual 26/60 Rev 02-2010 For an optimal specimens preservation, Milestone S.r.l. recommend to keep the Cool Packs in the refrigerator close to the TissueSAFE, at a temperature value of +4°C , ±2°C ( 39,2 °F, ±35,6°F). Do not keep and transport any Cool pack inside the Thermal bag at temperature values lower than 2°C (35,6°F). Frozen Cool packs could damage the specimens during their transportation. Cool packs are extremely important to keep the temperature inside the transportation bag. Do not transport any specimens in a Thermal bag without using the cool packs. After transportation, specimens are ready for grossing procedure (fig. 21). Fig 21 TissueSAFE Operation Manual 27/60 Rev 02-2010 5.3. PORTABLE REFRIGERATOR 26 LITRES P/N DESCRIPTION 68083 Kit Portable Refrigerator 26 Litres with trolley with folding handle. Power supply 12-24V & 110-230V. Ideal for long term transportation by car. It is also possible to use the mobile refrigerator (A) as shown in the picture below (Fig. 22). This device is ideally suited for long term transportation by car. See also Table in Appendix C Fig 22 A TissueSAFE Operation Manual 28/60 Rev 02-2010 With this type of portable refrigerator, cool packs are not required because the temperature is monitored by a thermometer. In this case there is an easy-to-use basket that perfectly fits in the refrigerator (fig. 22a,b). Fig 22a Fig 22b The portable refrigerator: P/N 68083: PORTABLE REFRIGERATORS C 26 (capacity: 26 litres). Main Characteristics: Voltage allowed: 12-24V, 110-240V - 50Hz/60Hz. Digital electronic thermostat with audible and visual alarm system (not for power failure). The weight of the C26 mobile refrigerator, when empty, is 15 Kg (33lbs). For further information, see specific instruction manual included in the shipment TissueSAFE Operation Manual 29/60 Rev 02-2010 5.4. KIT PORTABLE REFRIGERATOR 65 LITRES This mobile refrigerator with its mobile cart, as shown in the picture above, is ideally suited for easy transportation inside the Hospital. See also Table in Appendix C NOTE: Model at 230V is supplied also with UPS P/N 68081 68082 DESCRIPTION Kit Portable Refrigerator 230V - 65 litres with trolley and UPS (*) Kit Portable Refrigerator 115V - 65 litres with trolley (NO UPS included) (*) UPS Technical Specifications: Capacity: 1000VA/600W Voltage: 220/230/240VAC – 50/60Hz Recharge Time: For first installation, connect at least for 10 hours. Afterwards, 10 hours to 90% after complete discharge. Autonomy in Battery Mode (UPS): 4 hours with minimum load (tested with a sample of 0.5 litre volume) 1 hour with maximum load (tested with 20 litres organic material at 37°C) Operating Environment: 0-40°c, 0-80% Relative Humidity (non-condensing) Dimensions: (DxWxH) (mm): 405x205x145 Net Weight: 9.6 Kg For further information, see specific instruction manual included in the shipment TissueSAFE Operation Manual 30/60 Rev 02-2010 P/N DESCRIPTION REFRIGERATORS 68084 Kit Refrigerator box in PS 32 litres complete with N.4 cool packs 68083 Kit Portable Refrigerator 26 Litres with trolley with folding handle. Power supply 12-24V & 110-230V. Ideal for long term transportation by car. 68081 68082 Kit Portable Refrigerator 230V - 65 litres with trolley and UPS Kit Portable Refrigerator 115V - 65 litres with trolley (NO UPS included) 68064 68090 Refrigerator laboratory type 140 Litres 230Volt complete Refrigerator laboratory type 140 Litres 115Volt complete TissueSAFE Operation Manual 31/60 Rev 02-2010 5.5. KIT PRINTER The kit allows printing of a label with indication of date and time of packing/sealing of the sample, for enhanced traceability. Label must be positioned on the vacuum bag including specimen P/N DESCRIPTION KIT PRINTER 68120 Kit Printer at 230V for Tissue Safe Bags 68118 Roll of paper for Kit Printer at 230V for Tissue Safe 68119 Spare Blades (Kit 2 pieces) for Kit Printer at 230V for Tissue Safe 5.6. BAG OPENER Suitable to facilitate opening of the vacuum bag for easy insertion of the sample/organ. PN 68112 BagOpener TissueSAFE Operation Manual 32/60 Rev 02-2010 5.7. TEMPERATURE MONITORING SYSTEM P/N DESCRIPTION TEMPERATURE MONITORING KIT 68075 RFID controller plus reader: F Including: A. Palm Pc Touch-screen PDA B. RFID Datalogger card (q.ty 10) C. RFID smart card SDIO reader D. RFID Desktop Reader USB E. Power Supply for PDA F: N. 100 Protection Bags for Palm Pc Touch-screen PDA The software program requires the installation of Microsoft Excel (minimum version 2002 SP3). 68100 RFID Datalogger card 68036 N. 100 Protection Bags for Palm Pc Touch-screen PDA Milestone s.r.l. provides this dedicated system to continuously monitor and document the specimens temperature during their transportation, through a Radio Frequency Identification Device (RFID) smart card (according to the scheme of chapter 4.4). The system ensures a permanent monitoring of temperature-sensitive specimens and enables a fast documentation of process and storage temperatures through the easy handling. It measures temperature data by means of a calibrated sensor and stores them directly on the chip. TissueSAFE Operation Manual 33/60 Rev 02-2010 TEMPERATURE MONITORING KIT INCLUDES : 1. RFID smart card SDIO reader (P/N 68103, fig. 23) and controller (P/N 68101, fig. 24), provided with Milestone user-friendly software interface. Fig 23 Fig 24 Main Characteristics: Connector type: SDIO Operating frequency: 13,56MHz Air Interface protocol: ISO 15693-3 Please make sure that: - Palm PC Touch-Screed PDA (P/N 68101) is operated with specific protection bag. Protection bag is changed after each transportation cycle to avoid contamination For protection bag disposal, see Chapter 8 “Waste Disposal”. 2. RFID smart card data logger (P/N 68100, fig 25). Main Characteristics: Fig 25 Dimensions: 54mm X 86mm. Thickness: 1.80 mm. Construction: multilayer body of different synthetic materials. Back side cover: adhesive backside. Flexible and environmental friendly. Single and multi use. Dynamic bending stress. In accordance to EN ISO/IEC 10536-1 “identification cards – Contactless integrated circuit(s) cards – Close coupled cards” respectively ISO/ IEC JTC/SC17 N 1480. Temperature transponder in ID1-format (credit card size). Storage conditions: at room temperature, higher than 30% relative humidity. TissueSAFE Operation Manual 34/60 Rev 02-2010 Technical data: Data points 700 values (rate automatically adjusted by sw). Operating temperature -20°C up to 50°C Operating frequency 13,56MHz Air Interface protocol ISO 15693-3 Temperature accuracy Typical 0,3 K (-20°C to +30°C) Time accuracy Better than ±3% Memory 8kBit EEPROM Battery MnO2Zn - printed battery (1,5V ... 1,1V) Battery Life Up to 25 recording cycles (4°C), max 1 year TissueSAFE Operation Manual 35/60 Rev 02-2010 3. RFID smart card desktop reader (P/N 68102, fig. 26) and Milestone RFID data recording software. On completion of the recording procedure (at the end of the trip or other monitoring episode), the tag information can be downloaded by the PC of the Histology Lab with RFID desktop shown below (fig 26). Fig 26 Desktop reader main characteristics: Interoperable with ISO 14443 A&B, ISO 15693. USB interface. USB bus powered (no external power source). Supports Windows® 98, Mn, 2000, XP Serial communication driver for easier PC-linked application development. Embedded evaluation kit. Cryptographic security management for picopass® chips. Transparent mode preserves open. Connection with contactless interfaces. Minimum requirements for general use with PC: The PC must have Microsoft Windows XP SP2 (or later)/Vista and Microsoft Excel 2002 SP3 (or later). An USB port and a CD reader are also required. TissueSAFE Operation Manual 36/60 Rev 02-2010 Drivers Installation instructions: To install the Data Recording software please follow the points below: 1) Insert the Installation CD-ROM in the PC. 2) Connect the Card Desktop Reader in one of the USB ports of the PC. 3) Automatically the Windows Wizard opens the following window, select “No, not this time” and then press “Next”. 4) Now select “Installation from a list of specific location” and press “Next”. TissueSAFE Operation Manual 37/60 Rev 02-2010 5) Select “Include this location in the search:” and choose the folder in the CD-ROM named “RFID Desktop Reader Driver”. Click “Next” to continue. 6) Press the icon “Continue anyway” to skip Windows Logo testing. 7) The Wizard has successfully installed the files, press “Finish” to proceed. 8) Repeat again the installation procedure starting from point 3, to install the USB port driver. TissueSAFE Operation Manual 38/60 Rev 02-2010 Data Recording software installation instructions: 1) From the installation CD-ROM, open the folder “PC 1.0.0“. Then select and double click on the icon “Milestone TissueSAFE setup.exe”. 2) The installation Window Wizard will appear. Press “Next” to continue. TissueSAFE Operation Manual 39/60 Rev 02-2010 3) Leave the default installation Folder (C:\Program Files\Milestone TissueSAFE) and press “Next” to continue. 4) Click “Next” again to start the installation procedure. 5) On completion of the procedure, click “Close” to exit from the Wizard. At this point, you will find a new “Milestone TissueSafe” icon on your desktop. By clicking on it the Milestone data recording software will appear as shown below. If you cannot correctly visualize the Main Menu, you need to modify the Macro settings of Microsoft Office. To do this operation, select from the file menu SettingMacroProtection and then set as “Low” the Protection Level. TissueSAFE Operation Manual 40/60 Rev 02-2010 5.8. HOW TO OPERATE THE TEMPERATURE MONITORING SYSTEM A. How to activate the RFID card data logger in the PREOPERATING ROOM: 1. Switch on the PDA, the Tissue SAFE software will automatically start. The main screen will appear, as shown in fig. 27: Fig 27 Please make sure that: - Palm PC Touch-Screed PDA (P/N 68101) is operated with specific protection bag. Protection bag is changed after each transportation cycle to avoid contamination For protection bag disposal, see Chapter 8 “Waste Disposal”. To facilitate explanation of the functionalities, PDA is shown in this chapter without protection bag 2. Click on “Start Cold Chain Log” to activate the smart card (see arrow in fig. 27). 3. Place the smart card data logger close to the smart card reader as shown in the pictures 28a, b. Fig 28a Fig 28b Do not move the card during the reading procedure. TissueSAFE Operation Manual 41/60 Rev 02-2010 4. When the reader has detected the smart card, it activates the temperature/time monitoring with the message shown in Fig. 29. Fig 29 The smart card activation needs to be done upon the packaging of the FIRST specimen received by the Operating Room. If you want to stop the card reading and clear all the data, press on “Clear card” (fig. 27). A message will confirm the operation. In this case the card is still ready for a next data recording procedure. Under “Settings” section, it is possible to set the data recording time of the smart card (fig. 27). Click on “Settings” and insert the password 0039 to edit the data recording time (use the plus/minus buttons to modify the time). Select the record time that you need to monitor and then press OK. TissueSAFE Operation Manual 42/60 Rev 02-2010 A. Indicate the recording time of the card, (1 = one hour). If 1 hour recording time A is selected, temperature is read every 6sec. According to the recording time B C selected, the sampling time is automatically adjusted. D B. Current data and time: press the arrows E F to adjust them. C. If checked before programming the card, the software advises extra information, like completion time of the temperature monitoring. D. If checked, additional information can be added to the graph, like the description of the batch transported. E. If you press Exit the application will be closed. F. Press OK to confirm the modification. This PDA is conceived only for operation with Milestone Software. Do not install other programs on it. It is also possible to check the status of the smart card battery, by pressing “Battery” (fig. 27). TissueSAFE Operation Manual 43/60 Rev 02-2010 When the card has been activated, the operator must place it into the external pocket of the plastic bag of the FIRST specimen that has been sealed under vacuum that day, together with the Requisition Form (fig. 30a,b). Fig 30b Fig 30a Now, the operator can seal the external pocket to prevent any accidental separation of the card from the specimens (fig.31). Fig 31 TissueSAFE Operation Manual 44/60 Rev 02-2010 The plastic bag, containing the smart card, should be now placed into the refrigerator close to the Tissue SAFE or directly into the Thermal Bag (see fig. 32a,b, according to the scheme of the chapter 4.4). Fig 32b Fig 32a For an optimal specimen’s temperature monitoring, Milestone recommends to keep the smart card far away from the cooling elements of the refrigerator. B. SPECIMENS TRASPORTATION: from the surgery suites to the Pathology Lab. When the Courier is ready to transport the specimens to the Anatomical Pathology Department, the smart card together with the under vacuum specimens must be placed in the Thermal Bag, appropriately closed and then transported (fig. 33). Fig 33 TissueSAFE Operation Manual 45/60 Rev 02-2010 The operator must take care of placing at least 1, 2 or 3 Cool Packs at +4°C inside the thermal bag, according to the inner volume of the thermal bag used (fig. 34). Fig 34 For an optimal specimens temperature monitoring, Milestone S.r.l. recommend to keep the smart card among the under vacuum specimens, on the opposite side of the Cool packs. C. How to handle specimens upon their arrival in the HISTOLOGY LAB. Upon its arrival in the Histology Lab, the thermal bag is opened and the RFID card is removed from the plastic bag (fig. 35a,b). Fig 35a Fig 35b TissueSAFE Operation Manual 46/60 Rev 02-2010 Then the card is read on a standard PC through the RFID smart card desktop reader (fig 36 red circle). Fig 36 Please be sure to correctly position the card on the desktop reader as shown in fig. 36. In case of wrong card positioning, the system will not read the RFID card data. Position the RFID on the smart card desktop reader and click on “Read Card” (fig 37, red circle). A single reading takes less than 30 seconds, even for a tag that is completely full of data. Fig 37 On completion of the data reading procedure, the card will be deactivated and the temperature log stopped. TissueSAFE Operation Manual 47/60 Rev 02-2010 The time/temperature curve from the time of the smart card activation until arrival in the Histology Lab is shown in fig. 38. Fig 38 By pressing “Save Report” is possible to save the complete Log (fig. 37, green circle). If required, for documentation purposes, the complete log can be printed. Press the “Print Preview” button to confirm the printing operation (fig. 37, blue circle). Upon arrival, bags are sequentially opened as shown in fig. 39a,b. Once opened, specimens are ready for the grossing procedure. Fig 39a Fig 39b TissueSAFE Operation Manual 48/60 Rev 02-2010 6. CONSUMABLES P/N VACUUM BAGS Double welding for enhanced vacuum seal 68047SS Special vacuum bag 140μ - Size SMALL (Q.ty 200) Dimensions: external 250 x 300 mm – internal: 230 x 125 mm 68055SS Special vacuum bag 140μ -Size MEDIUM (Q.ty 200) Dimensions: external 340 x 385 mm – internal 320 x 150 mm 68054SS Special vacuum bag 140μ -Size LARGE- (Q.ty 200) Dimensions: external 340 x 550 mm – internal 320 x 320 mm 68046SS Special vacuum bag 140μ - Size EXTRA-LARGE (Q.ty 200) Dimensions: external 340 x 550 mm – internal 320 x 500 mm P/N DESCRIPTION VIALS 68058 Kit 200 plastic vials, capacity 1,5 ml with perforated lid Suitable for small needle biopsies. For use only in our Vacuum Bags. POLYSTYRENE CONTAINERS WITH PERFORATED LID 68093 Kit 10 pieces of polystyrene containers, capacity 200ml Suitable for small delicate (soft) specimens. For use only in our Vacuum Bags. 68094 Kit 10 pieces of polystyrene containers, capacity 30ml Suitable for small delicate (soft) specimens. For use only in our Vacuum Bags. 68125 Kit 200 polystyrene containing trays including: - N. 50 of polystyrene trays dim. 235 x 150 x h44 mm - N. 50 of polystyrene trays dim. 255 x 180 x h36 mm - N. 50 of polystyrene trays dim. 290 x 215 x h40 mm - N. 50 of polystyrene trays dim. 330 x 260 x h52 mm For use only in our Vacuum Bags. TissueSAFE Operation Manual 49/60 Rev 02-2010 6.1. POLYSTYRENE CONTAINERS WITH PERFORATED LIDS The use of polystyrene containers and vials with perforated lids is suggested when sealing very soft specimens (For ex: Ovarian Cyst). The container with the sample must be then positioned inside the vacuum bag for sealing. Capacity 200ml Capacity 30ml In this case the specimen dimension must not exceed 2/3 of the container inner volume. 6.2. POLYSTYRENE CONTAINING TRAYS Conceived to facilitate the introduction and the handling of the sample in the vacuum bag, these trays are suggested for sealing of organs like small and large bowel. See details on the organs that can be positioned in these trays in table of Appendix C, column P1 only. NOT suitable for very soft tissues (like brain). Risk of damaging the sample. TissueSAFE Operation Manual 50/60 Rev 02-2010 7. MAINTENANCE Maintenance of the TissueSAFE must be done at two levels. 1. The first is at laboratory level: simple routine cleaning maintenance can prevent the likelihood of more serious problems occurring at a later stage. 2. The second is technically more detailed and is beyond the scope of normal laboratory personnel. Attention: check with local distributor or with Milestone offices directly (customersupport@milestonemedsrl.com) how to perform Preventive Maintenance and regular inspections in order to maintain your Tissue Safe within safe operative standards. OPERATING ROOM PERSONNEL Minimal routine maintenance is required by Operating Room personnel. The following is recommended to ensure optimal performance of the TissueSAFE. A. DAILY: Clean and disinfect the instrument every day, following the procedure described in the APPENDIX A. Ensure that TissueSAFE is generally kept clean after each use, but do not use abrasives or sharp instruments that may damage acid/solvent resistant coating etc. Daily check the covers and the edge surface of the vacuum chamber. Traces of organic residues must be removed to ensure the optimum closing of the cover. The incorrect closing of the cover prevents the possibility of creating the vacuum. Daily sterilization can prevent cross contamination of the specimens. TissueSAFE Operation Manual 51/60 Rev 02-2010 B. MONTHLY: Check the oil level of pump on the back side of the instrument ( see 1 in fig 40). A level under the minimum can cause degradation on performance vacuum level and the damage of the pump. Fig 40 1 When oil became dark muddy and/or emulsify, it loses the lubricant properties and it is required a substitution (service personnel only). 7.1. CARBON/HEPA FILTER REPLACEMENT SERVICE PERSONNEL TissueSAFE is equipped with a Carbon/HEPA filter. This type of filter is used to filter the vapours coming from the biological specimens. The filter should be replaced every 6 months during the Preventive Maintenance action performed by authorized and trained service personnel or when the following warning appears (see picture below) on the Tissue Safe display (in this case, contact your local representative or manufacturer). TissueSAFE Operation Manual 52/60 Rev 02-2010 8. WASTE DISPOSAL 8.1. WASTE DISPOSAL OF VACUUM BAGS The vacuum bags supplied by Milestone s.r.l. are preventively irradiated with β rays before the use. After its use, treat empty vacuum bag as biohazardous. Waste material should be disposed in an approved incinerator in compliance with all federal, provincial and local government regulations. 8.2. WASTE DISPOSAL OF PROTECTION BAGS FOR PDA The protection bags for PDA supplied by Milestone s.r.l. must disposed after each use and must be treated as biohazardous. Waste material should be disposed in an approved incinerator in compliance with all federal, provincial and local government regulations. 8.3. WASTE DISPOSAL OF ELECTRICAL AND ELECTRONIC EQUIPMENT Removal of equipment from use for disposal: 1) Switch OFF the panel PC. Disconnect the main plugs of the panel PC from the main socket outlets 2) Switch off the camera power supply. Disconnect the main plug of the camera power supply from the main socket outlet. 3) Disconnect all accessories (foot pedal, keyboard, bar-code reader, camera, camera power supply). 4) Clean and disinfect all parts of the equipment according to the manufacturer specification. 5) Dispose all electric and electronic parts of the equipment according to the manufacturer: Complying with directive 2002/96/EC of the European Parliament and of the Council of 27 January 2003 on waste electrical and electronic equipment (WEEE), the separate collection environmental managed of equipment is mandatory. It is necessary to return the used equipment to the distributor or to inquire about the presence of a local empowered system for collection and disposal of WEEE. The in-observance of Directive 2002/96/EC or of the local law which acknowledge it can have potential effects on the environment and human health Before the disposal, the unit must be disinfected and oil of the pump must be removed from the unit according to the procedure described in Appendix A. Oil removed must be disposed separately according to your local regulations. This symbol indicates separate collection for electrical and electronic equipment For this and any possible clarification or any request for assistance please contact: customersupport@milestonemedsrl.com TissueSAFE Operation Manual 53/60 Rev 02-2010 APPENDIX A DECONTAMINATION PROCEDURE SAFETY PRECAUTIONS The personnel involved in the TissueSAFE decontamination procedure must wear the appropriate individual protective devices, in particular: Single use, HI-RISK protective gloves (100% hi-risk latex or equivalent ). Individual protection device that meets the standard EN 374 and the Directive 89/686/CEE. Antibacterial mask face for operating room and plastic apron. Individual protection device, meets the standard EN 149 Grade FFP3D and the Directive 89/686/CEE. Protective eyeglasses. Individual protection device, meets the standard EN 166 and the Directive 89/686/CEE. For safety reason, any operation without the individual protective devices must be avoided. TissueSAFE Operation Manual 54/60 Rev 02-2010 DECONTAMINATION PROCEDURE Please follow the instructions below on how to carry out a correct cleaning and decontamination procedure. Do not remove the side panels of the unit during the cleaning operations. The decontamination procedure can be performed only when the unit is cool. Before doing any operation disconnect the power supply. Do not use direct or high-pressure water to clean the instrument. It is recommended the use of enzymatic detergents and polyphenolic based disinfectants or chlorine substitutes to clean and disinfect the instrument. For further information, please contact the manufacturer: medical@milestonesrl.com Germicidal efficacy of polyphenolic based disinfectant agents: Biocide efficacy (in minutes) In alternative, to decontaminate the instrument, the use of a solution of 1:10 Bleach (recommended by the CDC for disinfecting) is also allowed. TissueSAFE Operation Manual 55/60 Rev 02-2010 HOW TO OPERATE: 1. Remove the spacers from the working area and wash them separately with an appropriate detergent. After that, spray the disinfectant solution and leave the solution for at least 5 minutes. Clean the spacers surfaces with a dry cloth or paper. 2. Remove, with an appropriate detergent, any organic residual from the inner surfaces of the chamber and from the external surfaces of the instrument. 3. Remove, with an appropriate detergent, any organic residual from the soldering bar connectors. Remove also polyester residuals from soldering bar (if any). TissueSAFE Operation Manual 56/60 Rev 02-2010 4. Spray the disinfectant solution on the external surfaces and the steel parts of the unit and leave the solution for at least 5 minutes. After that clean the surfaces with a dry cloth or paper. 5. Spray the disinfectant solution on the inner surfaces of chamber and leave the solution for at least 5 minutes. After that, clean the surfaces with a dry cloth or paper. 6. Spray the disinfectant solution on the Plexiglas cover (internal/external surface) and leave the solution for at least 5 minutes. After that, clean the surfaces with a dry cloth or paper. TissueSAFE Operation Manual 57/60 Rev 02-2010 7. Clean also the rubber gasket so that it can correctly adheres to the steel surface of the working area. Maintain also clean the inhaling bushing inside the working area (see red circle). 8. Before removing gloves, wash them with soap, water and disinfectant. 9. Wash your hands. 10. Remove the individual protection devices. 11. Protect the instrument as required by specific protocols for your department and / or service (in containers, medical grade paper, etc.). 12. Make sure the contaminated material is packed and disposed in an approved incinerator in compliance with all federal, provincial and local government regulation. On completion of the cleaning operations, wait at least 5 minutes before using the unit (connecting the power supply cable). TissueSAFE Operation Manual 58/60 Rev 02-2010 APPENDIX B Human Organs : Vacuum Selection Guideline P1 P2 P3 P4 Kidney Lung Brain Cyst > 10cm in diameter Breast Placenta Eye Uterus Prostate Cyst <10cm in diameter Liver Gall bladder Muscle Thyroid Thymus Skin Tongue Spleen Colon Bowel Pancreas Bladder Testis This guide is intended for specimens directly packaged with vacuum sealed plastic bags. When sealing very soft specimens (For ex: Ovarian Cyst), we suggest the use of plastic containers with perforated lids (P/N 68093-94). In this case the specimen dimension must not exceed 2/3 of the container inner volume. ONLY perforated lids can be used to seal specimens’ containers TissueSAFE Operation Manual 59/60 Rev 02-2010 APPENDIX C Specimens Transportation Guidelines OPERATING ROOM SEALING TRANSPORT CONDITIONS MAX 30MIN HISTOLOGY LAB FRIDGE +4°C GROSSING FRIDGE +4°C GROSSING FRIDGE +4°C GROSSING THERMAL BAG COOL PACKS +4°C SEALING SEALING FRIDGE 2Hr ,+4°C MOBILE REF +4°C MAX 3 HOURS THERMAL BAG COOL PACKS +4°C > 3 HOURS MOBILE REFRIGERATOR +4°C 1. If the time required by the courier for the specimens’ transportation to the Anatomical Pathology Department is less than 30 minutes, Milestone S.rl. recommends putting the specimens in the Thermal bag with the cool packs at +4°C. 2. If the time required by the courier for the specimens transportation to the Anatomical Pathology Department is less than 3 hours, Milestone S.rl. recommends to put the specimens into the refrigerator ), close to the TissueSAFE for at least 2 Hr (pre-cooling procedure) and later to place them in the Thermal bag with the cool packs at +4°C. 3. If the time required for the specimens’ transportation exceeds 3 hours, Milestone S.rl. recommends the use of the Mobile refrigerator connected to an external power supply (12/24Vdc, 115/230Vac). For details on refrigerators models available from Milestone s.r.l., see Chapter 5 Milestone s.r.l. TissueSAFE Operation Manual 60/60