Supplementary Material Membrane-based Continuous Remover of Trifluoroacetic Acid in Mobile Phase for LC-ESI-MS Analysis of Small Molecules and Proteins Zhigui Zhou, Jialing Zhang, Jiawei Xing, Yu Bai, Yiping Liao, and Huwei Liu* Beijing National Laboratory for Molecular Sciences, the Key Laboratory of Bioorganic Chemistry and Molecular Engineering of Ministry of Education, Institute of Analytical Chemistry, College of Chemistry and Molecular Engineering, Peking University, Beijing 100871, China Correspondence to: Huwei Liu, Institute of Analytical Chemistry, College of Chemistry and Molecular Engineering, Peking University, Beijing 100871, China, fax: +86-10-62751708, e-mail: hwliu@pku.edu.cn. Experimental Chemicals and Reagents Ribonuclease A (RNase A), cytochrome c (CYC, equine heart), and lysozyme (LYS) were obtained from Merck (NJ, U.S.A.), and myoglobin (MYG, equine) was from Sigma-Aldrich (MO, U.S.A.). N--benzoyl-L-arginine ethyl ester (99%, BAEE) and N--benzoyl-L-arginine (99%, BA) were purchased from Alfa Aesar (MA, U.S.A.). Trifluoroacetic acid (TFA) was from Acros Organics (NJ, U.S.A.). HPLC grade acetonitrile (ACN) and formic acid (FA) were purchased from Dikma Technologies Inc. (CA, U.S.A.). Purified water was provided by Wahaha Group Co., Ltd. (Hangzhou, China). All the reagents were used without further purification. TFA Remover fabrication As shown in Figure S1, the TFA remover was constructed with two electrodes, two polycarbornate (PC) slices with a channel (40 mm long, 1 mm wide, 1 mm deep) in the middle to deliver running buffer, a bipolar membrane, a anion exchange membrane, and a poly(tetrafluoroethylene) (PTFE) film with a micro-channel (40 mm long, 1 mm wide, 50 μm deep) in the middle to deliver LC mobile phase. The bipolar membrane with sulfonic and quaternary ammonium functional groups as ion exchangers was purchased from Beijing Tingrun Membrane Technology Development Co., Ltd. (Beijing, China), and the anion exchange membrane was from Unisplendour Co., Ltd. (Beijing, China). The channels in the PC slices and PTFE film were mechanically fabricated through the slices and film. The PTFE film was sandwiched between the bipolar membrane, anion exchange membrane, and PC slices as shown in Figure S1. Two additional channels with the same size (400 μm wide, 400 μm deep, 12 mm long) were mechanically fabricated on both ends of the PC slice at the side of the anion exchange membrane. Two corresponding channels on the anion exchange membrane were made by mildly pressing the membrane into the channels with a capillary to leave two prints. Two PEEK capillaries (360 μm o.d, 100 μm i.d.) were embedded and fixed into the two channels by epoxy resin glue to deliver mobile phase through the micro-channel. Two electrodes were placed at the outer sides of the PC slices, and connected with the power supply. All these units were clamped between two aluminum plates and fixed by four screws. Holes were drilled in the electrodes and aluminum plates to deliver running buffer (20 mM NH4HCO3) through the channels in the PC slices for TFA removal. Instrument Setup of LC-TFA remover-MS System All LC experiments were performed on an Agilent 1100 series HPLC system (Agilent Technologies, CA, U.S.A.). As shown in Figure 1(a), the inlet of the TFA remover was coupled with LC column via a commercial T-connection for mobile phase splitting if necessary. The splitting ratio was controlled by the length and inner diameter of the tubing that connected the T-connection and the waste bottle. The outlet tubing of the remover was directly connected to the ESI source of MS. Sample Preparation The stock solution of BAEE and BA mixture was prepared in 50 mM NH4HCO3 at the concentration of 2 mM each. Working solutions (0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, and 200 μM) were prepared by diluting the stock solution with 50 mM NH4HCO3 to the required concentration. Proteins (RNase A, CYC, LYS, and MYG) were respectively dissolved in 50 mM NH4HCO3 at the concentration of 1 mg/mL and stored at -20 °C. The stock solution of protein mixture (0.5 mg/mL each) was prepared by mixing the individual protein solutions in 50 mM NH4HCO3. The stock solution was diluted with 50 mM NH4HCO3 to obtain the working solutions (0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, and 500 μg/mL). All stock solutions were stored under refrigerated conditions (4°C), and all solutions were filtered through 0.22 μm cellulose membrane before injection. LC and MS conditions An Agilent HC-C8 column (150 mm × 2.1 mm i.d., 5 μm particle size, Agilent Technologies, CA, U.S.A.) was utilized for the separation of BAEE and BA. The mobile phase was water and ACN with TFA as modifier. UV at 254 nm was used for the detection. Proteins were analyzed on a Zorbax 300SB-C18 column (150 mm × 2.1 mm i.d., 5 μm particle size, Agilent Technologies, CA, U.S.A.). The mobile phase consisted of water (containing 0.1% FA or TFA, solvent A) and ACN (containing 0.1% FA or TFA, solvent B), and the flow rate was 0.2 mL/min. The optimized gradient was: 0-25 min 30%-65% B, 25-35 min 65% B, 35-50 min 30% B. For UV detection, the wavelength was set to 210 nm. All MS experiments were performed on an Agilent XCT ion trap mass spectrometer (Agilent Technologies, CA, U.S.A.). Positive mode was used for all experiments, and the parameters were set as follows: nebulizer 35 psig, dry gas temperature 325 °C, and dry gas flow 8 L/min. The MS data were acquired at the rate of 1.03 spectra/s in the mass range of m/z 100–600 for BAEE and BA and m/z 100–2200 for proteins. MS data was processed using DataAnalysis 3.3 (Bruker Daltonik GmbH) for Agilent ion trap MS. The extracted ion chromatograms (EICs, ±0.5 m/z expansion) of BAEE (m/z 307.4), BA (m/z 279.3), RNase A (m/z 1053.5), CYC (m/z 816.3), LYS (m/z 1101.3), and MYG (m/z 848.6) were extracted and integrated, respectively. Figure S1. Expanded view of the designed TFA Remover. Figure S2. Total conductivities and pH values of the tested mobile phases flowing out of the remover under different flow rates and voltages: (a) and (b) water (containing 0.5% TFA); (c) and (d) ACN/water (50/50, v/v, containing 0.5% TFA); (e) and (f) water (containing 0.1% TFA); (g) and (h) ACN/water (50/50, v/v, containing 0.1% TFA) Figure S3. Total conductivities of water (a) and ACN/water (50/50, v/v, b) flowing out of the remover under different flow rates and voltages Figure S4. (a) (b) Typical chromatograms of BAEE and BA without (a) and with (b) TFA as modifier in the mobile phase; (c) typical chromatogram of BAEE and BA using LC-TFA remover-UV system; (d) combined EICs of BAEE and BA obtained by LC-TFA remover-MS system Figure S5. Typical chromatograms of four proteins using FA (a) and TFA (b) as modifier in the mobile phase Figure S6. Combined EICs of four proteins using LC-TFA Remover-MS system under different remover voltages: (a) 5 V; (b) 8 V; (c) 11 V Table S1. Peak Areas of BAEE and BA Using UV Detector With and Without TFA Removal (n = 3) Without TFA Removal With TFA Removal BA 2561 3870 BAEE 2524 2139 Table S2. Quantitative Analysis Results of BAEE and BA Using LC-TFA remover–MS System (n=3)* BA Concentration BAEE μM Average area RSD% Average area RSD% 0.2 91850736 21.9 96676626 9.8 0.5 103516183 2.7 236306144 1.5 1 153746374 1.7 336916671 1.1 2 300060670 1.0 457646998 0.2 5 669831969 0.1 1128752728 1.4 10 1284170156 2.9 2048397682 0.8 20 2296288389 1.2 3452519634 1.4 50 5159487972 0.5 8639170653 0.3 100 8434985861 0.4 15457365482 0.9 200 13258694913 1.5 26353756576 1.4 *The injection volume was 2 μL. The voltage of TFA remover was 10.5 V. The extracted ion chromatograms (EICs, ±0.5 m/z expansion) of BAEE (m/z 307.4) and BA (m/z 279.3) were extracted and integrated. Table S3. Quantitative Analysis Results of Four Proteins Using LC-TFA remover–MS System (n=3) * Concentration RNase A CYC μg/mL Average area RSD% Average area RSD% 500 133470469 1.6 5346055936 1.5 200 58162085 6.3 2390222152 2.3 100 27153302 2.0 1085214485 1.2 50 13582525 7.9 554323892 0.8 10 4047666 6.3 75835380 0.8 5 2739096 17.2 33554511 7.1 (Continued) LYS Concentration MYG μg/mL Average area RSD% Average area RSD% 500 45288935 1.0 7209702274 0.4 200 16100213 1.0 2971252906 0.8 100 8694860 3.5 1531549823 1.0 50 5383699 3.4 760094700 0.7 10 1993456 1.3 153376556 2.1 5 781522 (LOD) - 82652611 4.0 1 Not detected - 17665235 4.0 0.5 Not detected - 6581537 1.5 *The injection volume was 5 μL. The voltage of TFA remover was 6 V. The extracted ion chromatograms (EICs, ±0.5 m/z expansion) of RNase A (m/z 1053.5), CYC (m/z 816.3), LYS (m/z 1101.3), and MYG (m/z 848.6) were extracted and integrated.