Data Collection Diffraction experiment: X-ray source Synchrotron Lab sources (Sealed tube, rotating anode) Optics Focusing mirrors Monochromators select Kα or Kβ Monochromatic X-ray beam Crystal 30μm 100μm 2D Detector Orthogonal to the rotation axis and typically but not necessary to the incident beam. 2 between diffracted and unaffected beam Standard method of data collection for macromolecules: - Copper rotating anode - rotation method - area detector X-ray sources: Seald tube vacuum A = anode of copper or molybdenum cooled by water (Win – Wout) X = X-rays obtained through Beryllium windows K = Katode emitting a beam of high speed electrons (blue points; energy = 10,000 eV) due to high negative potential with respect to the anode U = 30 to 150kV between K and A– accelerates the electrons Disadvantage: limited intensity because of melting of the anode at higher tension Rotating anode Rotation allows better cooling of the anode and therefore application of higher powers (12-18kV) generating higher intensities. Disadvantage: continuous pumping to maintain the vacuum Synchrotron 85% macromolecular structures rings with UHV (ultra high vacuum) and radius of 10-100m in which electrons circulate at velocities 99,9% of light speed Magnetic fields bend the electron beam (acceleration toward the center of the ring) which then emittes X-rays: wide wavelength region depending on particle energy and field strength 3rd generation elmag. radiation collision of particles Advantages: - - High flux = many photons (macromolecular crystals show weak diffraction due to big dimensions of proteins (fewer in the crystal than of small molecules) and due to small number of electrons in light elements like N, O and C are) High collimated and small beam: instead of (brilliance = ?radiance = Strahlendichte) => small macromolecular crystals Tunablility = ability to choose wavelength (MAD) polarized electromagnetic radiation Time structure (radiation is produced in flashes: allows time resolved experiments i.e. snapshots of reaction intermediates) Devices: - - Bending magnets (guide electrons in their orbit: dependent of their energy and the magnetic field) Wavelength shifter (-> shorter wavelength = higher energy due to increased field strength) Multipole wiggler = a series of wavelength shifter (increasing of intensity due to constructive interference of the radiation of single periods -> fluxtotal = 2N*fluxsingle period with N periods) - Undulator (for collimation on the orbit plane of the electrons, nearly monochromatic radiation = narrow engery bands compared to wiggler due to strong interference because of the moderate magnetic field and poles close together, tunability by adjustment of the pole distance, intensity total= N2*intensitysingle periode) Crystal Mounting: Avoid precession (Kreiselbewegung) upon spindle axis rotation. You have to center the crystal on the goniometer. - - Capillary Mounting: o If crystal decay is no problem o At Room temperature 1. Link a thin-walled capillary tube with a syringe or pipette with the help of a flexible adaptor. 2. Soak up the crystal. 3. Close the end of the capillary needle with resin(Harz)/clay(Lehm). 4. Remove the excess water by a thin glass capillary. 5. Break off the needle and close the second end with resin/clay. 6. Mark the position of the crystal with a felt tip pen. Loop Mountin: o Safe option for synchrotron data collection o Cryocooling: Flash freeze crystal soaked with cryoprotectant Nitrogen steam at 100K and dry air (or nitrogen) at 300K (0°C = 273K) perpendicular to the X-ray beam o Kinds of loops: Litholoops Micromounts o A thin liquid spans the loop whose size approximate the size of the crystal. By this way the crystal is fixed. The surface tension has to be big enough. Tools: - - Manual mount: o Crystalwands (Stäbe) o Cryotons o Reverse pincers (Kneifzange) crystal mounting and transfer: o loops homemade and bought o arcs (Bögen) o tongs o vials and special vials and holders Transfer the crystal to the diffractometer: 1. 2. 3. 4. 5. 6. Precool tongs and plunge (abschrecken) the crystal. Clasp (ergreifen) the mounting pin. Remove the pin holder. Carry it to the diffractometer. Transfer it to the diffractometer. Open the tongs Dewar for storage: - A dewar is a double walled vessel for isolation and vacuum with mirrored glass for storage of very hot or very cold liquid. You fill it with liquid nitrogen. Storage of crystals in vials stored in baskets in a canister. Automated crystal transfer from the dewar on the diffractometer: Reproducible Fast expensive Cryocooling: Resolution and overall data quality can be improved by flash-cooling the crystals to 100K. For this we use liquid nitrogen. We mount the crystal in rayon (Kunstseide) and put it in the nitrogen steam reduced thermal motion reduced conformational disorder enhanced signal-to-noise-ratio higher limiting resolution SUPPRESSION OF RADIATION DAMAGE Ad radiation damage: When X-rays interact with matter free radicals and electrons are formed. This radicals and electrons can attack the protein. They diffuse through the water channels of crystals. This diffusion can be prevented by decreasing the temperature. But it does not inhibit the formation of the radicals and electrons. Cryoprotectants are added before crystallization: - Glycerol Ethylene glycol PEG of low molecular weight Sucrose They prevent ice formation and support vitrification (vitrifies = amorphous ice). increase in viscosity depression of freezing temperature Sudden freezing (flash freezing or shock cooling) also prevents ice formation. Surrounding the cold gas steam by a warm and dry steam of air or nitrogen avoids the accumulation of ice on the crystal and the diffraction instrument. Adjusting the flow speeds of the steams prevents turbulences between them. Further protection against ice formation is enclosing the apparatus in a box. Rotation method: The crystal is fixed on a goniometer for rotation of crystal (Eulerian or Kappa geometry) around an axis orthogonal to the incident beam. 2.0° > Δ > 0.1° Planes of the reciprocal lattice are brought into diffraction conditions. Rotations is repeated for contiguous angles until at least the independent part (not already measured because of symmetry) of the reciprocal lattice is completely scanned. independent A set of reciprocal lattice planes will never meet diffraction conditions. This set is collected as symmetry related. Only if the symmetry axis lies along the or near the rotation axis there will be a blind region. Kappa goniometer can help. Advantage: fast Disadvantage: distorted image of reciprocal lattice planes. Diffraction pattern: lunes To avoid spots superposition the rotation angle has to be small enough so that the lunes are well resolved. Quality Indicators of rotation methode: - - Completeness: indicates the percentage of reflections experimentally determined with respect to the theoretical ones for that resolution (should be as near as possible to 100%). 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑢𝑛𝑖𝑞𝑢𝑒 𝑟𝑒𝑓𝑙𝑒𝑐𝑡𝑖𝑜𝑛𝑠 𝑚𝑒𝑎𝑠𝑢𝑟𝑒𝑑 𝑡𝑜𝑡𝑎𝑙 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑢𝑛𝑖𝑞𝑢𝑒 𝑟𝑒𝑓𝑙𝑒𝑐𝑡𝑖𝑜𝑛𝑠 Check the low resolution completeness. Ideally, every independent (unique) reflection for a chosen resolution should be detected, but in reality there are always a few reflections omitted. <I>/σ<I>: indicates the signal to noise ratio (how strong the diffraction from the crystal is compared to the backround). - - <I> is the mean intensity. σ<I> is the mean standard deviation of the measurements. Hight resolution limit = |I/σ| = for example 2. That means <I> has fallen to twice the mean standard deviation. Redundancy: indicates the number of times that the reflections are measured as such or as symmetry related. The higher redundancy the better the reflection intensity estimation. Rmerge = Σhkl Σi |Ii(hkl) - <I(hkl)>|/Σhkl |<I(hkl)>| Ii(hkl) are measurements of a symmetry related reflection I(hkl). The number of measurement is indicated by i. <I(hkl)> is the mean value of I(hkl). It is ΣIi(hkl)/n R(merge) can be regarded as the mean error of an intensity measurement compared to the mean intensity. Detectors: Measure intensities of diffracted beams (-> amplitude of structure factor) Number of X-photons Energy in a given time Quality Indicators: - - - - Detection Quantum Efficiency: indicates signal to noise ratio (fraction of photons absorbed from the detector and the detector intrinsic noise) = fraction/percentage of the total radiation impinging (auswirkend) on an image receptor that is actually detected by the receptor. the higher the better Dynamic range: The ability to acquire (annehmen) very strong and very weak spots at the same time. the wider the better Detector response: relationship between incoming photons and detected photons linear over the entire dynamic range uniform all over the surface Point spread function. ability to discriminate very close spots Types of detectors: a) Photon counting: - Scintillation counters measurement of individual diffracted beams: one beam at a time - Multi-wire proportional gas chambers = 2D detectors slow fast saturation outdated technology b) Energy in a given time -> 2D detectors: - - Films low sensibility non-linear response narrow dynamic range a lot of work IP = Image plates CCD = Charge-coupled devices Image Plates: 1.) X-ray-photons excite electrons in a thin layer of small crystals of photostimulable phosphor. 2.) Electrons are excited to a conduction band (Leitungsband). only in semiconductor (Halbleiter) and isolators 3.) In the conduction band the electrons are trapped in Br- and F- vacancies into a metastable state. 4.) A laser scans the plate and the visible light stimulates the electrons 5.) The electrons drop back to the valance band and thereby emit photons. 6.) Photons are detected by a photomultiplier. Advantages: - good Detection Quantum Efficiency - high Dynamic Range - good Point Spread Function Disadvantage: Readout time of minutes -> poor duty cycle especially for short exposures Wikipedia: Duty cycle is the proportion of time during which a component, device, or system is operated.[1] Suppose a disk drive operates for 1 second, and is shut off for 99 seconds, then is run for 1 second again, and so on. The drive runs for one out of 100 seconds, or 1/100 of the time, and its duty cycle is therefore 1/100, or 1 percent. Charge coupled Device: 1.) X-rays excite a phosphor layer on a fiberoptic taper. 2.) Phosphor emits visible light. 3.) Light is transferred through the fiberoptic taper to the CCD. 4.) In CCD photons of energy greater than 1.1 eV allows a valance electron of the semiconductor to move into the conduction band. 5.) The free electrons are collected and counted by an electronic system. Advantages: - good Detection Quantum Efficiency - high Dynamic Range - good Point Spread Funktion - fast readout - small, smaller than IP Properties Size Speed of reading Price X-ray-source Exposure time Image Plates Small 10 sec 120,000 Euros Rotating anodes Few to over 10 min Charge coupled devices bigger < few sec. 250,000 – 1,000,000 Euros Synchrotron Few sec