Hellenic Accreditation System Annex F2/7 to the Certificate No. 559-2 SCOPE of ACCREDITATION of the Calibration Laboratory of the DEKA S.A. Measurand / Calibration item Range of measurement Calibration & Measurement Capability (k=2)* Remarks Temperature measurements Temperature/ Analog and digital direct reading thermometers Temperature/ Analog and digital direct reading thermometers Temperature/ Liquid-in-glass thermometers of total and partial immersion Temperature/ Analog, digital direct reading thermometers for measurements of air temperature, temperature data loggers Page 1 of 3 - 40 oC ... 250 oC 0,047 oC 250 oC ... 420 oC 0,17 oC 420 oC ... 600 oC 1,3 oC 600 oC ... 1000 oC 1,8 oC - 30 oC ... 150 oC 0,10 oC 150 oC ... 420 oC 0,17 oC - 40 oC... 150 oC 0,047 oC 150 oC ... 250 oC 0,047 oC - 40 oC ... 70 oC 0,36 oC Annex F2/7 to the Certificate ESYD No. 559-2 Calibration method based on COFRAC LAB GTA 08-2005 Calibration method based on COFRAC LAB GTA 08-2005 Calibration is performed on site. Calibration method based on NIST : “Liquid in – Glass Thermometer Calibration Service” 091988 Comparison method 25.06.2015 Measurand / Calibration item Temperature / Temperature transmitter with analog signal output mA or V connected with temperature sensor Temperature / Temperature transmitter with analog signal output mA or V connected with temperature sensor for measurements of air temperature Temperature/ Climatic and isothermal chambers, ovens and incubators, refrigerators, freezers, with internal volume up to 2000 l, with or without air circulation Range of measurement - 40 oC ... 250 oC - 40 oC ... 70 oC - 40 oC ... 250 oC Calibration & Measurement Capability (k=2)* 0,050 oC 0,40 oC 0,20 oC Remarks Calibration method based on DKD-R 5-1 2003 Calibration method based on DKD-R 5-1 2003 According to technical guide DKD-R 5-7 07/2004, using Pt 100 platinum resistance thermometer. Calibration is carried out in the useful volume or in selected locations in the chamber. Calibration can also be performed on site. Relative humidity measurements Relative humidity / Analog and digital hygrometers / Humidity transmitter with analog signal output mA or V connected with humidity sensor for measurements of relative humidity Page 2 of 3 15 % RH … 25 % RH 1,8 % RH 25 % RH … 40 % RH 2,1 % RH 40 % RH … 60 % RH 2,5 % RH 60 % RH … 75 % RH 3,3 % RH 75 % RH … 95 % RH 3,9 % RH Annex F2/7 to the Certificate ESYD No. 559-2 Calibration is performed at 25 oC. Calibration method based on NPL , “A guide to the measurement of humidity”, 1996 25.06.2015 Measurand / Calibration item Range of measurement Calibration & Measurement Capability (k=2)* Remarks Pressure measurements Pressure/ Analog and digital gauge pressure measuring instruments of direct reading (gas medium) Pressure/ Analog and digital gauge pressure measuring instruments of direct reading (deionised water medium) Pressure/ Analog and digital gauge pressure measuring instruments of direct reading (oil medium) 5 kPa ... 40 kPa 20 Pa 40 kPa ... 160 kPa 80 Pa 160 kPa … 400 kPa 0,20 kPa 400 kPa … 1000 kPa 0,50 kPa 1 MPa … 2,5 MPa 1,3 kPa 2,5 MPa … 6 MPa 3,0 kPa 6 MPa … 16 MPa 8,0 kPa 16 MPa … 40 MPa 20 kPa 40 MPa … 100 MPa 50 kPa 1 MPa … 10 MPa 1,563·10-4 · p – 2,20·10-5 in MPa 10 MPa … 100 MPa 1,621·10-4 · p – 2,63 ·10-4 in MPa Calibration according to DKD-R 6-1 /2003 Using pressure balance Calibration according to DKD-R 6-1 /2003 p in MPa Pressure/ Analog and digital negative gauge pressure measuring instruments of direct reading (gas medium) Pressure/ Analog and digital absolute pressure measuring instruments, including barometers, of direct reading (gas medium) Pressure/ Analog and digital absolute pressure measuring instruments of direct reading (gas medium) * -100 kPa ... -12,5 kPa 50 Pa 5 kPa abs ... 160 kPa abs 80 Pa 0,16 MPa abs … 1,6 MPa abs 0,80 kPa Calibration according to DKD-R 6-1 /2003 Where uncertainty is accompanied by the corresponding unit, it is absolute, while where it is not accompanied by a unit, it is relative. Site of assessment: Permanent laboratory premises, 21, Aristidou str., Piraeus, Attiki, Greece. Approved Signatories: S. Kritikos, C. Athanasopoulos. This Scope of Accreditation replaces the previous one dated 23.07.2013. The Accreditation Certificate No. 559-2, to ELOT EN ISO/IEC 17025:2005, is valid until 31.05.2017. Athens, 25.06.2015 Ioannis Sitaras Director of the Laboratories Accreditation Division Page 3 of 3 Annex F2/7 to the Certificate ESYD No. 559-2 25.06.2015