Activity 1.1.3 – Instrumentation Calibration Introduction This activity is a study in accuracy and precision. Both of these concepts are important tools used in biotechnology engineering. It is very important to know which is more appropriate for the task at hand. In terms of arrows shot toward a bull’s eye of a target, it is possible to see a clear example of accuracy compared to precision. The three arrows with solid lines represent: precision The three arrows with broken lines represent: accuracy When using instruments in the lab, it is critical that they provide reliable results (precise) and are representative (accurate) of what they are supposed to measure. Lab instruments must be calibrated frequently to ensure accuracy. Alternatively, certain lab techniques (e.g., pipetting with serological pipettes) have inherent variability that must be taken into account. Equipment • • • • • • • • Computer Excel Electronic top loading microbalance Electronic Mettler microbalance Micropipettes (p20, p100, p1000) Pipetteman Serological pipettes (5 and 10 ml) Large and small plastic weigh boats Project Lead The Way, Inc. Copyright 2010 PLTW TM - BE – Unit 1 – Lesson 1.1 – Activity 1.1.3 – Instrumentation Calibration – Page 1 Procedure Student teams have been contacted by a local biotech company that needs its pipettes calibrated. You and your team will run the pipettes through a series of tests to determine if they are calibrated. The local company also requests that your teams compare the accuracy and precision of micropipettes compared to serological pipettes. Step 1: Create an Excel spreadsheet that looks identical to the one below. Step 2: Using the table below, transfer the amount of water shown using the correct pipette. Place the water onto a weigh boat, tare the microbalance, and record the weight. Remember that 1 ml of water should weigh 1 mg. Repeat the process 10 times for each volume of water. Actual Measured Amounts (g) Pipette **Micropipette P1000** Serological pipette 5 Serological pipette 5 Serological pipette 10 Volume 40 (μl) 1 (ml) 4 (ml) 4 (ml) Trial 1 .040 Trial 2 .040 Trial 3 .040 Trial 4 .040 Trial 5 .040 .9 .8 .9 .8 .9 3.7 3.8 3.7 3.8 3.8 3.6 3.8 3.9 3.8 3.9 Table 1 * NOTE: The Mettler microbalance must be used. Project Lead The Way, Inc. Copyright 2010 PLTW TM - BE – Unit 1 – Lesson 1.1 – Activity 1.1.3 – Instrumentation Calibration – Page 2 Step 3: Using an Excel spreadsheet, find the following for each data set (YOU WILL NEED TO CREATE A NEW TABLE FOR THESE CALCULATIONS): • Average • Standard deviation • Percentage error (difference from expected) Project Lead The Way, Inc. Copyright 2010 PLTW TM - BE – Unit 1 – Lesson 1.1 – Activity 1.1.3 – Instrumentation Calibration – Page 3 Actual Measured Amounts (g) Pipette Volume Trial 1 Trial 2 Trial 3 Trial 4 Trial 5 Average % Error STDEV 0.04 0.04 0.04 0.04 0.04 0.04 0% **Micro pipette P1000** 40 (μl) 0 Serologi cal pipette 5 1 (ml) 0.9 Serologi cal pipette 5 4 (ml) 3.7 Serologi cal pipette 10 4 (ml) 3.6 0.8 0.9 0.8 0.9 0.86 14% 0.054772 2557505 166 3.8 3.7 3.8 3.8 3.76 6% 0.054772 2557505 164 3.8 3.9 3.8 3.9 3.8 5% 0.122474 4871391 59 Project Lead The Way, Inc. Copyright 2010 PLTW TM - BE – Unit 1 – Lesson 1.1 – Activity 1.1.3 – Instrumentation Calibration – Page 4 Step 4: Graph the standard deviation for each data set and the percentage error. Step 5: Using the graphs you created, answer the following questions: 1. Explain the best way to pipette 1 ml? The best way to pipette 1 mL is using the serological pipette 5 because it can hold at least 1 mL, but it cannot measure a lot more than 1 mL. It also has a relatively low percent error, which means that the serological pipette 5 is an accurate way to measure 1 mL of water (the experimental measurements are close to the goal measurement. It also has a low standard deviation, which means it is a precise way to measure 1 mL of water (the measurements are consistent). 2. Explain how the results match or do not match what you expected. The results are close to the expected results. It was expected that the Micropipette would be the most accurate and precise because it is mechanized and it measures a very small amount of water, meaning that it has less room for error. It is very accurate (it had a percent error of 0% meaning that there was no measurable error in the measurements) and it was very precise (it had a standard deviation of 0, meaning that the results were consistent). It was surprising that the serological pipette 10 was more accurate than the serological pipette 5 in measuring 4 mL of water. The serological pipette 10 had a percent error for measuring 4 mL of water of 5%, while the serological pipette 5 had a percent error of 6%. This means that the serological pipette 10 was more accurate (the results were closer to the goal results). Typically, the instrument with the capacity closest to (but not under). Possible explanations for this include-- different people were measuring the serological pipette 10 and the serological pipette 5 for the 4 mL trials, and instruments that must be read manually (the serological pipettes) can be difficult to read. This results in human error. Predictions were correct that the micropipette was the most accurate and precise. It had a percent error of 0% and a standard deviation of 0. This can be explained because it is mechanized, meaning it measures water mechanically and decreases the opportunity for human error. It was accurate because the percent error was 0% and it was also precise because the results were consistent (there was a standard deviation of 0). 3. How can you determine if all of the micropipettes are calibrated? The micropipettes are calibrated when the results are consistent. Consistency in results is also known as precision. The results must have a very low standard Project Lead The Way, Inc. Copyright 2010 PLTW TM - BE – Unit 1 – Lesson 1.1 – Activity 1.1.3 – Instrumentation Calibration – Page 5 deviation- which measures how much the results vary from each other. The results must also be accurate- the measured values must be close to the actual goal values. Accuracy can be measured by the precent error. 4. What is an acceptable error for pipetting? An acceptable percent error for pipetting is about 10%. Conclusion 5.Using this activity as a guide, describe in your own words the difference between precision and accuracy. Accuracy is how close the results are to the goal measurement. Accurate results are close to the goal. Precision, however, is how close the results are to each other, or the consistency. If results are far away for the goal value, but they are incorrect by the same amount each time- they are precise, but not accurate. 6. Which is more important for an industry: instrumentation, precision, or accuracy? Why? Instrumentation is the most important because it allows the user to know the measurements of accuracy and precision and to “pick the right tool for the job.” Project Lead The Way, Inc. Copyright 2010 PLTW TM - BE – Unit 1 – Lesson 1.1 – Activity 1.1.3 – Instrumentation Calibration – Page 6