The Effect of Blood Flow Rate on Venous Collapse During clinical processes such as blood donation and blood pressure measurement, the pressure outside of a vessel may become greater than the vessel’s internal pressure, causing the vessel to collapse. In this experiment, the relationship between these pressures and system parameters is analyzed to determine the point at which collapse will occur. To simulate collapse, a collapsible tubing model is created for experimentation, in which the diameter and type of material used as tubing were varied. A dimensionless parameter for these variables is created and verified in relation to Poiseuille’s Law. According to the data and the dimensionless parameter, higher flow rates require a greater pressure difference to cause collapse and therefore the higher the flow rate, the higher the pressure necessary for the tubing to eventually collapse. Betsy Hamme Robert Jenkins Rob Ledger Jess LeMay Group M4 Dept. Of Bioengineering University of Pennsylvania Philadelphia, PA Introduction Flow has collapse, or the cross sectional area may medical become very small. At this point, the lumen Virtually all fluid-carrying is reduced to two narrow channels separated vessels in the body have flexible walls and by a flat region of contact between opposite can collapse when the transmural pressure, walls. This twin-lobed shape is the lowest- (pressure across the vessel wall) ptm, falls energy below a critical value. At large positive flexible values of ptm where the internal pressure is constraints1. numerous applications. in collapsible physiological tubes and configuration wall except for a close uniformly to end greater than the external pressure, collap- For the case of steady-state flow, the sible tubes are distended and stiff and upstream-downstream pressure difference behave as non-collapsible tubes. (p) is related to flow rate and vessel At parameters through Poiseuille’s Law: negative values the tubing may completely 1 Hamme, Jenkins, Ledger, LeMay P 8LQ a 4 fluid, the compliance of the tubing, the flow Equation 1 rate, transmural pressure and internal pressure differences5. As of now, there are where is dynamic viscosity, L is the tubing no simple or conclusive mathematical length, Q is the flow rate, and a is the radius equations modeling pulsatile flow. Reynold’s number is a dimensionless of the tubing. Large values of p tend to variable that is defined as: correspond with large positive values of ptm. Oscillations occur in collapsible Re U l tubes when the external pressure exceeds the Equation 2 internal pressure, which causes the vessel to in which U is the mean fluid velocity, l is a collapse. This difference in pressure occurs length when the flow rate exceeds a “critical” flow diameter), and is the kinematic viscosity rate, at which point p becomes great of the fluid. Collapse can occur in small enough that the internal pressure is at the vessels with low Reynolds number (Re) or minimum needed to balance the given in larger vessels with high Reynolds external pressure and still maintain steady numbers. flow. The tube begins collapse from the accompanied by self-excited, flow-induced downstream end, where pressure it at its oscillations. There are several examples of lowest. Collapse restricts flow, and pressure physiological vessels that are prone to builds up at the upstream end. The pressure vascular collapse. Gravitational decrease in in the tube eventually becomes great enough internal pressure with height causes collapse to overcome the effects of the external in systemic veins above the heart. Systemic pressure and flow is forced through the tube. arteries can become compressed by a With the increase in flow rate, the internal pressure cuff, or within the chest during pressure decreases and the process starts cardiopulmonary resuscitation. The urethra over. and the ureter are both compressible tubes, Higher external pressures require parameter (likely the tube’s In some cases collapse is lower flow rates before oscillations begin, which and likewise higher flow rates for given micturition external pressures cause greater oscillations. respectively. The rate of oscillation increases until the biological occurrences of collapsible tubes, vessel completely collapses. The period of including intramyocardial coronary blood oscillation depends on the viscosity of the vessels during systole, pulmonary blood 2 may be or compressed peristaltic There are during pumping, many other Hamme, Jenkins, Ledger, LeMay vessels in the upper levels of the lung, and the systolic pressure over the diastolic large intrathoracic airways during forced pressure measured in mmHg4. expiration or coughing2. The measure of systolic and diastolic Materials/Apparatus blood pressure relies on the fact that arteries 1. Tygon tubing are collapsible tubes and that Korotkoff 2. Penrose tubing, 16 mm internal diameter sounds are created during oscillation. Study 3. Latex tubing, 2mm internal diameter of these sounds has determined that they are 4. Tank with supports for holding water generated by self-excited tube oscillations 5. Pressure transducer and are specifically due to the increasing 6. Graduated cylinder to collect water steepness of the wave front during the 7. Stop watch propagation of a pressure wave3. When a 8. Flow cutoff valve blood pressure cuff on the arm is highly 9. LabView (flow.vi and pressure.vi) pressurized, the external pressure around the 10. Pressurized Rubbermaid® chamber artery causes the artery to collapse. Nothing 11. Pipettes to act as manometers is heard at this time since there is no flow. 12. Sphygmomanometer The cuff pressure is slowly released, and reaches a specific pressure at which blood Procedure can be forced through the artery during systole. During diastole, the artery is forced The experimental setup is shown in Figure closed again, causing a Korotkoff sound. 1. Water was used as the fluid medium. A The first cuff pressure at which these sounds thin-walled undistended tube was mounted are heard is called the systolic pressure, between which pressurizable chamber (Figure 2). is measured using a rigid couplings We significantly decreased, it becomes low Rubbermaid® container measuring (7.5in x enough that there are no fluctuations in the 4.5in x 2.25in). The inlet end view can be artery wall and blood flow becomes silent seen in Figure 3. and laminar, as it is normally. The lowest identical except that there is no inlet for air. pressure. Blood pressure readings consist of 3 using a constructed sounds are heard is called the diastolic chamber in sphygmomanometer. As the cuff pressure is cuff pressure at which these Korotkoff the brass a The outlet end view is Hamme, Jenkins, Ledger, LeMay Figure 1: Experimental Set-up Figure 2: Pressurizable chamber Figure 3: End view of chamber 4 Hamme, Jenkins, Ledger, LeMay Thus, the chamber could be pressurized. experiment, an initial flow rate was set and The pressure within the chamber provided external external pressure on the tubing (pe) and was atmospheric controlled by an ideally airtight system collapsed and flow ceased; measurements of consisting of a sphygmomanometer pump p and Q were recorded for each new pe. and pressure This was repeated for several initial flow The transducer and Labview rates. The maximum initial flow rate for tubing transducer. connecting to a were used to measure pe. The tubing in the pressure was pressure increased until the from tubing any trial was 40 ml/sec. chamber consisted of both 2mm (internal diameter) latex tubing and 16mm Penrose Results and Discussion tubing with lengths of 14cm and 9cm, respectively. An upstream At low external pressures (.002 - .01 reservoir atm gage), the flow through the latex rubber provided the mean upstream pressure while collapsible tubing was found to obey an upstream valve controlled flow rate (Q) within the tubing. Poiseuille’s Law. Four flow studies fell into Upstream (p1) and this range, with outside gage pressures at downstream (p2) pressures within the tubing .0026, .0045, .0069, and .011 atmospheres. were measured using manometers mounted The internal pressure drop for these flow at the inlet and outlet ends of the chamber. situations varied linearly with the flow rate, Two different experiments were with R-square values ranging from .96 at performed for both types of tubing. In the .011 atm to .998 were plotted against the first, external pressure was held constant at a theoretical Poiseuille flow, given the same pressure and the flow rate was varied. The viscosity, internal change in pressure, p = p1 - p2, was tube at .0069 atm. The pressure/flow relationships length, and tube measured for each flow rate. The flow rates diameter (see Figure 4). were increased until large oscillations of the From this data, it was determined that the tubing occurred and flow became pulsatile. collapsible nature of the tubing does not A new pe was selected and the experiment affect flow until the outside-inside pressure was repeated. The external pressures ranged difference is great enough to cause it to between 262 Pa and 1,114 Pa. In the second 5 Hamme, Jenkins, Ledger, LeMay and 1,113 Pa latex tubing trials. The Volume Flow (m^3/s) 0.00006 critical point was defined as the internal Poiseuille Flow 0.00005 0.00004 264 Pa 0.00003 460 Pa 0.00002 696 Pa 0.00001 1113 Pa pressure difference that corresponded to the first oscillations in flow. Of course, a minor amount of error was induced because the flow itself oscillated with the diameter of the tube, but this was negligible by 0 0 500 1000 1500 2000 design (the critical point was chosen at the 2500 very smallest amount of oscillation). Pressure Drop (Pa) Figure 5 shows the “flutter points” in Fig. 4: Internal pressure/volume flow relationships for low outside pressures Essentially, at these pressure 0.00006 differences (both transmural and internal), 0.00005 Volume Flow (m^3/s) flutter. relationship to the rest of the flow trials. flow functions as it does in rigid tubes. However, as the internal pressure difference increases (i.e. as the flow velocity increases), the transmural pressure difference increases to a critical point. That Poiseuille Flow 0.00004 460 Pa 0.00003 696 Pa 0.00002 0.00001 1113 Pa 0 0 is, as the pressure downstream continues to 500 1000 1500 2000 2500 Pressure Drop (Pa) drop, the pressure across the membrane Fig. 5: Internal Pressure/flow relationship including point of deviation from Poiseuille flow increases (because the external pressure is held constant). At the critical point, the The critical point of oscillation is clearly outside pressure overcomes the elastic force shown as the last point on each of the data of the tube, and the tubing begins to neck. series. The sharp decrease in volume flow Water builds up behind the neck, which re- makes sense, since less fluid is allowed opens the tube. This process causes high- through the tube due to the fluctuations in frequency oscillations of the diameter of the diameter. tube, resulting in the commonly known “bronx-cheer effect.” As the transmural pressure increases, the degree of the effect of the At the critical oscillations also increases (i.e. larger effect pressures, the flow through a collapsible on volume flow). This effect results from tube no longer obeys Poiseuille’s law. This the higher transmural pressure collapsing the phenomenon was observed in the 460, 696, tube to a greater degree. Note also that as 6 Hamme, Jenkins, Ledger, LeMay the transmural pressure increases, the tube relationships were linear, the actual flow collapses at a smaller internal pressure rate was three orders of magnitude lower difference. Collapse occurs because the than the expected flow rate (10-5 vs. 10-2 increased transmural pressure allows the m3/s). The flow rates and internal pressure rubber to collapse at greater internal differences were within the same order of pressures. magnitude as those in the latex trials. That Unfortunately, the latex-rubber is, the radius of the tubing increased by eight tubing could not be collapsed. This failure times, but the flow through the tube did not. was primarily due to the high elastic Because the pressure drop remained the modulus and small diameter of the tubing, in same, and the radius increased, Poiseuille’s conjunction with the limitations of the law predicts a greatly increased flow rate for pressure chamber. The latex tubing required the Penrose trials, ranging from .021 to .274 too large a transmural pressure difference m3/s. for the pressure chamber to handle – while As shown in Figure 6, these predictions were false. 0.00005 Volume Flow (m^3/s) the critical transmural pressure difference would only have been around .5 atm, the pressure upon the lid of the Rubbermaid container was too great to hold such pressures. In an attempt to overcome these 460 Pa 0.00004 0.00003 745 Pa 0.00002 965 Pa 0.00001 0 0 difficulties, a series of experiments was 500 1000 1500 2000 Pressure Drop (Pa) completed using Penrose rubber tubing, experiments Fig. 6: Pressure/flow relationship for Penrose tubing, showing a 10^3 drop in flow rate from Poiseuille previously used on the latex rubber (internal The flow through the Penrose tubing did not pressure vs. volume flow at constant act in accordance with Poiseuille’s law transmural pressure) were carried out using because of errors in the experimental set-up. the Penrose tubing. However, the data for Primarily, the Penrose tubing was wider the Penrose rubber trials did not comply than the upstream and the downstream with same Tygon tubing. In fact, the Penrose tubing transmural pressure differences as with the was around 3 times as wide as the Tygon latex). While the internal pressure vs. flow tubing (Penrose ~ 8mm and Tygon ~ which had a greater diameter and a smaller elastic modulus. Poiseuille’s The Law (at the 7 Hamme, Jenkins, Ledger, LeMay 2.5mm). Furthermore, the fluid flowed through several centimeters of Tygon tubing 0.00005 downstream manometer. Volume Flow (m^3/s) after the upstream manometer and before the This change in radius created most of the errors in predicting flow through the Penrose tube. For example, while the radius of the Penrose was wide enough to allow a 10-2 m3/s 1/3 Highest Vel. 0.00004 0.00003 2/3 Highest Vel. 0.00002 Highest Vel. 0.00001 0 volume flow for a 1000 Pa internal pressure 0 5000 10000 15000 20000 25000 30000 Atmospheric Pressure (Pa) difference, the radius of the Tygon forced the volume flow down to about 10-5 m3/s. Fig. 7: Figure 7 shows the relationship between Furthermore, because the Penrose was so volume flow and atmospheric pressure for Penrose much weaker than the latex tubing, the diameter began to oscillate at very low In Figure 7, the needle valve at 1/3, 2/3, and transmural and internal pressure differences. completely open positions corresponded to In fact, it was nearly impossible to increasing volume flow at near atmospheric distinguish pressure. The flow curves then converge as between the steady and the atmospheric pressure is increased – this oscillating flow in the Penrose tubing trials. trials corresponds to the beginning and increase of conducted on the Penrose tubing involved oscillations of the tube diameter. Finally, opening the needle valve to a given the atmospheric pressure is increased until diameter, and increasing the atmospheric flow is completely stopped. The pressure pressure until the tube completely collapsed. needed to completely collapse the tube Essentially, the pressure behind the tubing increased directly with the size of the valve was kept at a constant value (by keeping the aperture. This result is fairly easy to explain needle valve set and the water level in the –the height of the water in the tank was kept tank constant) and the atmospheric pressure constant, and the size of the valve aperture was increased from zero gage until the flow directly through the tube completely stopped. Three upstream of the tube. An opening of the trials were completed, with the needle valve aperture from 1/3 to 2/3 of it’s maximum at 1/3, 2/3, and completely open. The data is value creates a greater pressure behind the The final experimental presented in Figure 7. 8 corresponded to the pressure Hamme, Jenkins, Ledger, LeMay tube, which in turn increases the necessary behavior of flow at greater transmural transmural pressure for collapse. pressures would be beneficial. For both the It is useful to relate all of the system latex and Penrose, it would be very variables to each other and enable the interesting to investigate the flow behavior system to be a prototype for a biological at atmospheric pressures closer to the point system. two of collapse. One of the major problems with created this study, however, is that once the which together contain all of the necessary diameter begins to oscillate, it is nearly variables. The variables were created using impossible to determine the internal pressure the Buckingham Pi method. difference or flow rate, because pressure and In order dimensionless 1 to parameters L Q 2 do this, were Q 2 Pext D 3 (P) 2 flow rate oscillate as well. For the latex rubber trials, the A relationship between these two parameters experimenters would like to have applied can be derived by plotting one against the pressures great enough to observe complete other. The determined relationship is shown collapse. The total pressures applied were in Figure 8. actually fairly small, but the surface area of the lid of the pressure chamber was too large to remain sealed at higher pressures. With a revised experimental set-up, greater pressures could be applied, and collapse could be observed. For the Penrose tubing trials, the experimenters would like to use tubing with a smaller diameter. If tubes were used that had diameters comparable in size to the Tygon tubing, the deviations from Poisuille Fig. 8: This graph shows the relationship between the two undimensional variables. flow would disappear. Then more thorough, meaningful Further Investigation This experiment studies of pressure/flow relationships could be completed. would benefit greatly from a minor amount of further investigation. Primarily, more study of the 9 Hamme, Jenkins, Ledger, LeMay References 6. Hoffman, JI and Spaan, JA, “Pressureflow Relations in Coronary Circulation.” Physiological Reviews. 70(2): 331-90, 1990 April. 7. Katz, AI, Chen, Y, and Moreno, AH. “Flow Through a Collapsible Tube. Experimental Analysis and Mathematical Model.” Biophysical Journal. 9(10): 1261-79, 1969 Oct. 8. Ku, David, Ziegler, Marvin, and Downing, J.Micah, “One-Dimensional Steady Inviscid Flow through a Stenotic Collapsible Tube.” Journal of Biomechanical Engineering. 112: 444449, November 1990. 9. Jensen, O.E. “Chaotic Oscillations in a Simple Collapsible-Tube Model.” Journal of Biomechanical Engineering. 114: 55-59, February 1992. 1. Bertram, C.D. “The Dynamics of Collapsible Tubes.” Symposia of the Society for Experimental Biology. 49:253-64, 1995. 2. Kamm, R.D. and T.J. Pedley. “Flow in Collapsible Tubes: A Brief Review.” Journal of Biomechanical Engineering. 111: 177-179, 1989 August. 3. Chew, Y.T. and Low, H.T. “Pressure/flow relationships in collapsible tubes: effects of upstream pressure fluxuations.” Medical & Biological Engineering & Computing. 29: 217-221, March 1991. 4. Fox, Stuart Ira. Human Physiology, Fifth Edition. Times Mirror Higher Education Group, Inc., 1996. pp 4123. 5. Bertram, C.D. “The Dynamics of Collapsible Tubes.” Journal for Experimental Biology. P.253-63, July 10, 1995 10