Instrumentation ( ENT 165 ) Laboratory Module KOLEJ UNIVERSITI KEJURUTERAAN UTARA MALAYSIA INSTRUMENTATION ENT 165 EXPERIMENT # 3 PRESSURE MEASUREMENT UNIT CHARACTERISTICS NAME PROGRAMME signature GROUP DATE MATRIK # 1 Instrumentation ( ENT 165 ) Laboratory Module Pressure Measurement Unit Characteristic 1. OBJECTIVE 1.1 To study the principles of different types pressure measurement unit such as: 2. i. Vacuum Gauge ii. Pressure Gauge iii. Differential Pressure Gauge iv. Inclined Tube Manometer v. Singe Tube Manometer vi. U-Tube Manometer INTRODUCTION Pressure is the normal force exerted by medium, usually a fluid and gas. In engineering, pressure is most often expressed in pascal ( Pa = 1 N/m 2 ) or pounds-force per square inch ( lbf/in2 or psi ). Often pressure is measured by measuring its effect to deflection through use of a pressurized area and either a gravitational or elastic restraining element. A comprehensive classification of basic pressure-measuring methods are difficult to make. However, the measurement unit you will observe in this session lab should suffice for our objectives. Vacuum Gauge Vacuum is pressure measured below atmospheric pressure, referenced to atmospheric pressure. Vacuum is negative gauge pressure. Because gauge pressure is measured relative to local atmospheric pressure, it can be either positive or negative. Perfect vacuum is zero absolute pressure and indicates complete absence of any matter. Figure 3.1(a) shows Vacuum Gauge. Pressure Gauge Pressure gauge is measured relative to local ambient pressure. It’s the difference between the measured pressure and atmospheric pressure, unless the ambient 2 Instrumentation ( ENT 165 ) Laboratory Module pressure surrounding the sensor is different from the atmospheric pressure. Figure 3.1 (b) shows Pressure Gauge. Differential Pressure Gauge Differential pressure is the difference in pressure between two points of measurement. If the pressure at one point in a system is a 150 psig and another point is 75 psig, the differential pressure is 75 psid. The difference is usually the lower pressure. Figure 3.1 (c) shows Differential Pressure Gauge. U-tube Manometer Pressures are applied to both legs of the U, and the manometer fluid is displaced until force equilibrium is attained. Pressures P1a and P2a are transmitted to the manometer legs through some fluid of density ρt, while the manometer fluid has some greater density ρm. In general, a u – tube manometer will have a greater pressure range when a more dense measuring fluid is used and greater sensitivity ( change in height per unit change in pressure ) when a less dense fluid is used. U-Tube Manometer can be seen in Figure 3.1 (d). Single Tube Manometer This is the simple method to measure pressure. Because the fluid density is involved, accurate work will require consideration of temperature variation. Single Tube Manometer can be seen in Figure 3.1 (e). Inclined Tube Manometer An Inclined Tube Manometer is used to measure very small pressures, vacuums or differentials. The slanted scale enables reading 0.01 “wc variations, which is impossible with conventional manometers or gauges. Inclined Tube Manometer can be seen in Figure 3.1 (f). The pressure which measured using above method must be compared to the Pressure Calibrator unit. The error between the gauges reading compare to actual reading ( Pressure Calibrator unit ) need to be calculated. This is done to determine which measurement give the most accurate reading according to the Pressure Calibrator unit. 3 Instrumentation ( ENT 165 ) Laboratory Module The error percentage ; % error = Gauge value – Pressure Calibrator Value x 100 3.1 Pressure Calibrator Value a b c d e f Figure 3.1 : a)Vacuum gauge, b) Pressure Gauge, c) Differential Pressure Gauge, d) U-tube manometer, e) Single tube manometer, f) Inclined tube manometer 4 Instrumentation ( ENT 165 ) 3. Laboratory Module EQUIPMENT Figure 3.2: Assembly Diagram of Pressure Measurement Equipment No. 1. 2. 3. 4. 5. Item Pressure Gauge Vacuum Gauge Differential Pressure Gauge Vacuum Tank Vacuum/Pressure Pump 4.0 PRE-LAB QUESTIONS No. 6. 7. 8. 9. 10. 4.1 What is pressure definition ? Item U-Tube Manometer Single Tube Manometer Inclined Tube Manometer Pressure Tank Pressure Calibrator ( 2 marks) 4.2 Give the brief idea about the fundamental law using in pressure measurement. ( 4 marks) 4.3 What are standard pressure commonly use in industries? State four standards. ( 2 marks ) 4.4 What medium can be used to generate pressure? ( 2 marks) 5 Instrumentation ( ENT 165 ) Laboratory Module 5.0 PROCEDURE No. 1. Start-up procedure Perform a quick inspection to make sure that equipment is in a proper working condition. 2. Be sure that all valves ( V1 to V6 ) are initially closed 3. Ensure that the vacuum tank and pressure tank are connected to the vacuum and pressure port of the pump, respectively (as shown in Figure 3.2). 4. Regulate the vacuum and pressure regulator to minimum. 5. Connect the pump power supply cable. 6. The equipment is now ready to be run. No. Vacuum Gauge Calibration Procedure 1. Perform the general start-up procedures. 2. Connect valve V2 to the vacuum gauge connection using 5x8 mm pneumatic tubing 3. Connect valve V1 to the ‘Lo’ port of pressure calibrator using 4x6mm pneumatic tubing. 4. Open all valve ( V1 to V6) about 10 second to equalize the pressure. Then close back all valves. 5. Open Valve V1 and V2. 6. Ensure that the vacuum is regulated to the minimum, and then switch on the pump. 7. Slowly adjust the vacuum regulator so that the vacuum gauge pointer indicates exactly – 1.0 inHg. 8. Switch on the pressure calibrator and set to inHg unit. 9. Record the gauge and calibrator vacuum readings into the experimental data sheet in Table 3.1. 10. Repeat steps 6 to 8 with decreasing 1.0 inHg up to - 10.0 inHg. 11. Always adjust the pressure and vacuum regulators to the minimum and wait until the pressure equalizes before switch off the pump 6 Instrumentation ( ENT 165 ) No. Laboratory Module Pressure Gauge Calibration Procedure 1. Perform the general start-up procedures. 2. Connect valve V5 to the pressure gauge connection using 5x8mm pneumatic tubing. 3. Connect valve V4 to the ‘Hi’ port of pressure calibrator using 4x6mm pneumatic tubing. 4. Open all valve ( V1 to V6) about 10 second to equalize the pressure. Then close back all valves. 5. Open Valve V4 and V5. 6. Ensure that the pressure is regulated to the minimum, and then switch on the pump. 7. Slowly adjust the pressure regulator so that the pressure gauge pointer indicates exactly 1 psi. 8. Switch on the pressure calibrator and set to psi unit. 9. Record the gauge and calibrator pressure readings into the experimental data sheet in Table 3.2. 10. Repeat steps 6 to 8 with increasing 1 psi up to 10 psi. 11. Always adjust the pressure and vacuum regulators to the minimum and wait until the pressure equalizes before switch off the pump No. Differential Pressure Gauge Calibration Procedure 1. Perform the general start-up procedures. 2. Connect valve V5 to the ‘Hi’ port of the differential pressure gauge using 5x8mm pneumatic tubing. Connect valve V2 to the ‘Lo’ port of the differential pressure gauge using 5x8mm pneumatic tubing. 3. 4. Connect valves V1 and V4 to the ‘Lo’ and ‘Hi’ port of pressure calibrator, respectively using 4x6mm pneumatic tubing. 5. Open all valve ( V1 to V6) about 10 second to equalize the pressure. Then close back all valves. 6. Open Valve V1, V2, V4, and V5. 7. Ensure that the pressure and vacuum is regulated to the minimum, and then switch on the pump. 7 Instrumentation ( ENT 165 ) Laboratory Module 8. Slowly adjust the pressure regulator so that the gauge pointer indicates 1 psi. 9. Switch on the pressure calibrator and set to psi unit. 10. Record the gauge and calibrator pressure readings into the experimental data sheet in Table 3.3. 11. Repeat steps 7 to 9 with increasing 1.0 psi up to 10 psi. 12. Always adjust the pressure and vacuum regulators to the minimum and wait until the pressure equalizes before switch off the pump No. Inclined Tube Manometer Calibration Procedure 1. Perform the general start-up procedures. 2. Connect valve V5 to the ‘Hi’ port of the inclined manometer using 8mm grey pneumatic tubing. Note: Leave the ‘Lo’ port of the inclined tube manometer opened to the atmospheric pressure. 3. Connect valve V4 to the ‘Hi’ port of pressure calibrator using 6mm pneumatic tubing. 4. Open all valve ( V1 to V6) about 10 second to equalize the pressure. Then close back all valves. 5. Open Valve V4, V5 and V6. Note: V6 is opened to reduce pressure due to small range of the inclined tube manometer. 6. Adjust manometer ruler so that the red fluid on the zero scale. 7. Ensure that the pressure is regulated to the minimum, and then switch on the pump. 8. Switch on the pressure calibrator and set to inH2O unit 9. Slowly adjust the pressure regulator and V6 so that the indication liquid shows 0.4 inH2O. 10. Record the manometer and calibrator pressure readings into the experimental data sheet in Table 3.4. 11. Repeat steps 9 to 10 with increasing 0.2 inH2O up to 2.6 inH2O. 11. Always adjust the pressure and vacuum regulators to the minimum and wait until the pressure equalizes before switch off the pump 8 Instrumentation ( ENT 165 ) No. Laboratory Module Single Tube Manometer Calibration Procedure 1. Perform the general start-up procedures. 2. Connect valve V5 to the ‘Hi’ port of the Single Tube Manometer using 8mm grey tubing. 3. Connect valve V4 to the ‘Hi’ port of pressure calibrator using 6mm pneumatic tubing. 4. Open all valve ( V1 to V6) about 10 second to equalize the pressure. Then close back all valves. 5. Open Valve V4, V5 and V6. Note: V6 is opened to reduce pressure due to small range of the Single Tube Manometer. 6. Ensure that the pressure is regulated to the minimum, and then switch on the pump. 7. Slowly adjust the pressure regulator so that the indication liquid shows 0.2 kPa. 8. Switch on the pressure calibrator and set to kPa unit. 9. Record the manometer and calibrator pressure readings into the experimental data sheet in Table 3.5. 10. Repeat steps 6 to 8 with increasing 0.2 kPa up to 2.0 kPa. 11. Always adjust the pressure and vacuum regulators to the minimum and wait until the pressure equalizes before switch off the pump. No. U-Tube Manometer Calibration Procedure 1. Perform the general start-up procedures. 2. Connect valve V5 to the ‘Hi’ port of the U-tube manometer using 6x8mm pneumatic tubing. 3. Connect valve V2 to the ‘Lo’ port of the U-tube manometer using 6x8mm pneumatic tubing. 4. Connect valves V1 and V4 to the ‘Lo’ and ‘Hi’ port of pressure calibrator, respectively using 4x6mm pneumatic tubing. 5. Open all valve ( V1 to V6) about 10 second to equalize the pressure. Then close back all valves. 9 Instrumentation ( ENT 165 ) Laboratory Module 6. Open Valve V1, V2, V4, and V5. 7. Ensure that the pressure and vacuum is regulated to the minimum, and then switch on the pump. 8. Slowly adjust the pressure regulator so that the liquid in Left tube shows 220 mmH 2O height. 9. Switch on the manometer calibrator and set to mH2O unit. 10. Record the gauge (Left and Right tube) and calibrator pressure readings into the experimental data sheet in Table 3.6. 11. Repeat steps 7 to 9 with left tube showing 240, 260, 280……400 mmH2O. 12. Always adjust the pressure and vacuum regulators to the minimum and wait until the pressure equalizes before switch off the pump. No. Shut-down Procedure 1. Disconnect all pneumatic tubings that connect gauges or manometers to tank. 2. Check that the pressure/vacuum regulators are set to minimum. 3. Switch off the pump and the main power. 4. End of experiment 10 Instrumentation ( ENT 165 ) Laboratory Module 6.0 RESULT Observation Table Table 3.1: Vacuum Gauge Gauge reading Actual reading ( inHg) ( inHg) %Error -1.0 -2.0 -3.0 -4.0 -5.0 -6.0 -7.0 -8.0 -9.0 -10.0 * Plot the graph gauge reading against actual reading. Label the graph. Table 3.2 : Pressure Gauge Gauge reading Actual reading ( psi) ( psi) %Error 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 * Plot the graph gauge reading against actual reading. Label the graph. 11 Instrumentation ( ENT 165 ) Laboratory Module Table 3.3 : Differential Pressure Gauge Gauge reading Actual reading ( psi) ( psi) %Error 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 * Plot the graph gauge reading against actual reading. Label the graph. Table 3.4 : InclinedTube Manometer Manometer reading Actual reading ( inH2O) ( inH2O) %Error 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 * Plot the graph gauge reading against actual reading. Label the graph. 12 Instrumentation ( ENT 165 ) Laboratory Module Table 3.5 : Single Tube Manometer Manometer reading Actual reading ( kPa) ( kPa) %Error 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 * Plot the graph gauge reading against actual reading. Label the graph. Table 3.6 : U-Tube Manometer Manometer Actual reading difference reading ( mmH2O) %Error ( mmH2O) 220 240 260 280 300 320 340 360 380 400 * Plot the graph gauge reading against actual reading. Label the graph. 13 Instrumentation ( ENT 165 ) 7.0 Laboratory Module DISCUSSION 7.1 Explain the characteristic of each measurement unit that you had observed in lab session: a) Parts and their function. b) Operation of measurement unit c) Advantages d) Disadvantages 7.2 ( 20 marks ) Discuss the differences between Vacuum gauges, Pressure gauges and Differential Pressure gauges. 7.3 ( 10 marks ) Suggest any transducers you know that suitable to replace the Vacuum gauges, Pressure gauges and Differential Pressure Gauges in this lab session. Explain the transducer’s characteristic. 7.4 ( 15 marks ) Base on the graph that you have plotted, is that every gauge and manometer need to be calibrated again? Why?( Give any references that say so). ( 10 marks ) 7.5 Suggest any method to reduce the error in measuring the pressure in this lab session. 8.0 ( 5 marks ) CONCLUSION ( 5 marks ) (Based on data discussion, make your overall conclusion by referring to experiment objectives) 14