FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI Fluoroscopy Exam Cheat Sheet SI Units: Absorbed Dose: Gray (Gy )& Rad; 1 Rad equals 100ergs per gram 1 Gray= 1joule/kg –Gy is SI unit 1 Gray= 100 Rads Dose Equivalent: Absorbed Dose X Quality Factor Rem; Sievert (Si) is the SI unit 1 Si= 1 joule/kg 1 Si=100 Rem Electron Volt: Unit of Energy Erg: another unit of energy 10-7 joules Kerma: sum of kinetic energies of all charged items, unit is joule/kg or Gy. Milliampere : mA , measure of X ray tube current, measure of quantity of X ray Eponyms: Stochastic Effect Non Stochastic Effect Photoelectric Effect Compton Effect Law of Bergonie & Tribondeau 1 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI Chapter 1 Input Phosphor traditionally Silver activated Cadmium sulphate 2nd gen CsI Output Phosphor CsI Photocathode is made of antimony and Cesium Electrostatic focusing lens positively charged and invert image Accelerating Anode has a potential of 25kV Electrostatic focusing voltage directly proportional to magnification Exposure directly increases proportionally with magnification to compensate for decreased brightness = normal squared/mag mode squared The mA is automatically increased when unit used in 6 inch magnified mode so patient dose exposure increased (9/6 gen 2.2) Brightness Gain = Intensifer luminance/Patterson B2 luminance Conversion factor= luminance of output phosphor/Input exposure rate Brightness Gain= Flux Gain(inherent) x Minification Gain 1 input phosphor photon= 1 electron ejected= 50 output photons Minification Gain= d1 squared/d0 squared Brightness Gain= Always above 1000-6000 Image quality function of number of absorbed photons Quantum noise/mottle/scintillation- eliminate by higher mA so more X ray quanta, lesser with Zn Cd compared with CsI Resolution 4 1p/mm For input Phosphor; Thicker screen with large particle size= bright but less resolution Thinner screen with small particles= less bright but better resolution Centre of Image brighter, less distortion and better resolution Pincushion Distortion: curved surface of input phosphor/warped image 2 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI Higher kVp increased transmission Brightness = mA X kVp5 Low Contrast images low kV High Contrast images high kV High kVp for reduced exposure and low kVp for best contrast kVp degrades contrast Ouput screen Luminance is measured in mililamberts A brightness gain of 3000 will satisfy light requirements for photopic vision Statistical quality of image can be improved by -increasing X ray beam -increasing Tube Current -increasing quantity of incident Photon -increasing X ray to light conversion efficiency Chapter 2 Objective Lens and Camera Lens Beam Splitter to direct portion of light into cine 70mm or closed circuit television camera F number= relative Aperture/speed= focal length/ diameter Each successive number will require exposure time twice as long Doubling the diameter will quadruple the light gathering ability ; D2 Chapter 3 525 scan line system Vidicon camera surrounded by 2 types of coils: focusing and deflecting coils Target assembly : Glass faceplate/signal plate and target 3 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI Vidicon target: thin film of photoconductive material made from antimony trisulphite suspended as globules in mica matrix Electrons released from electron gun by thermionic emission Signal plate potential 25 volts compared to wire mesh 250 volts Focusing coils and 2 pairs of deflecting coils From vidicon tube to CCU (Camera Control Unit) which synchronizes and Television Monitor Each Television frame is 525 lines Scanned 30 times per sec by electron gun Frame = Field X 2 To avoid flicker effect: interlaced horizontal scanning Bandwidth or Bandpass : frequency range or total number of cycles per second available in the camera; 4.1MhZ Horizontal resolution: Bandwidth or Bandpass Brightness on television screen controlled by control Grid. Frequency per line is a measure of horizontal resolving power of camera pickup tube and camera In the 525 scan line , 4.5MhZ TV system the amount of cycles to image one active line is approx 803 cycles/line Lag or stickness when camera moved Types of Camera Tubes: Vidicon Camera Plumbicon Camera: PbO conductor, less lag and least patient exposure. Image Orthicon Camera: larger, light & temp sensitive, works at lower illumination levels, better resolution, no lag, more expensive, long warm-up time Charged Coupled devices: small, low power, long life and no lag; used in high speed imaging such as cardiac cath. Kell Factor= Vertical Resolution/ scan lines = 0.7 Weakest Link in Image intensified change is TV system 4 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI Most important part of TV monitor is Cathode Ray Tube Chapter 4 Recording head Gap Width 0.001mm Video Disc Recorders used to record single fields, single frames or a short sequence. Low Dose to patient Spot Film Cameras and Cine Cameras record image from output phosphor of image intensifer Video Recordings are obtained from output of television camera tube Underframing: poor optical system, avoid Exact Framing: only 58% of frame used Overframing: Part of Image Lost Total Overframing: increases patient exposure Check Calculation for skin exposure 3.6R/min Cinefluororadiography equipment check once a year Spotfilm Camera : Reduced cost, time and radiation exposure Equipment requirements for fluoroscopy: Fixed Table: Arthrogram, enterclysis, retained stone removal, biopsy and position of tubes Motor Table: ERCP, UGI, Tube CHolangiography, GB spot films, VCUG Using a Spot film Camera film size with greatest dose to patient 105mm Common video disc frame rate 30/sec Cineradiography is influenced by synchronization, frame rate & f # of optical system Chapter 5 5 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI TID (Time Interval Difference Mode) : insensitive to respiratory motion, cardiac activity and motion artifacts K Edge Image intensification: Using Cerium to isolate iodine signal, limited to X ray tube power requirments Subtraction of phase related sharp masks is necessary to minimize the unsubtracted background Logarithmic processing of data prior to subtraction is necessary to ensure that the contrast will produce the same residual signal in the subtracted image Typical X ray exposure for a left heart study is 200R Conventional Temporal resolution Spatial Resolution Contrast +/- + _ + _ +/_ + Cinefluoro Computed Tomography - Chapter 6 Velocity of Light= wavelength X frequency Thermionic Emission Bremsstrahlung radiation 90% Characteristic radiation Increasing voltage , produces X rays of shorter wavelength and high Energy Photoelectric effect: knocks off inner electron Compton Effect: knocks off valence electron and original photon is scattered Brems radiation transforms kinetic energy f the electron into radiant energy Radiation capable of freeing electrons: ionizing radiation Tube Current Permeability of Xray determined by kV 6 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI Chapter 7 Exposure: X ray intensity and quantity proportional to mA or tube current Kilovoltage Peak Kvp determines quality of X ray, more penetrability Collimation decreases exposure & quantum mottle Filtration stops less penetrable rays; at least 2.5mm Al Intensity of X ray Beam at the table top should not exceed 2.2R/min for each mA of operating tube current at 80 kVp With AEC limit of exposure 10R/min Without AEC 5R/min With Boost 5R/min Monitor X ray tube current and potential at least once a week in 9inch water or 7 7/8 lucite phantom 1cm at under table X ray tube and 30cm when measuring overhead X ray tube Target to panel distance should not be less than 18 inches and shall not ebe less than 12 inches For Boost fluoro : special activation at Control panel needed , audible signal, Table top dose limited to 20R/ min unless recording devices used Chapter 8 300 R temporary sterility in females 30R temporary sterility in males Genetically significant dose: future children, X ray exam rate, mean gonadol dose per exam A dose of 50mRem delivered in a single exposure may result in cessation of sperm 7 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI At 25 Rads or Less ordinary laboratory or clinical methods show no indications of biological injury Cancer susceptibility from XRT decreasing order: Female breast, Thyroid, Hemopoetic Tissue, lungs, GI , Bones 50 Rads to fetus involuntary abortion Dose for Cataracts 200-250 Rads The X ray tube housing must be constructed so that the leakage radiation at a distance of 1 m from the target cannot exceed 100mR in 1 hr when the tube is operated at 80kVP Lead Apron/bucky and lead curtain,gloves and thyroid shields 0.25mm Lead Exposure reduction 97% When Fluoro tube working at 100kVP 1ft 500mR/hr, 2 ft 100mR/hr, 3 ft 50mR/hr Mobile Screen desks 1-2mm lead Gonad shield 0.5mm lead 135 degrees safe for operator Pocket Ionization chambers: can be accidently discharged and min range Whole Body Limits for rad equivalent 5Rem, extremities 50 Rem, Lens 15 Rem mA during Spot filming 100-150mA Numbers to Remember - Fluoro xrays tubes operate at 125-150 kVp; 2-5 mA Filtration must be at least 2.5mm Al equivalent Filtration eliminate less penetrating xrays before patients are exposed Intensity of xray beam at tabletop fluoro no greater than 2.2 Rads/min/mA (at 80kVp) Source-to-tabletop = target-to-panel distance shall be >12inch; should be >18 inch Increase image intensifier-to-pt distance increases pt dose Carbon fiber table tops reduce pt dose Primary protective barrier = image intensifier assembly, at least 2mm Pb equiv Bucky slot cover / protective curtain/drapes at least 0.25mm Pb Gonad shield at least 0.5mm Pb 8 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI - - Routine fluoro: max allowable exposure rate = 5 Rad/min With AEC, max exposure rate = 10 Rads/min; w/o AEC = 5 Rads/min When boost is available, max exposure rate = 5 Rads/min High Level (boost) fluoro: (10-50 Rads/min) 2-10 times higher dose than conventional fluoro; max tabletop dose = 20 Rad/min when not using recording devices Cumulative manual-reset time = 5min AEC=ABC: check current and potential qwk using phatom Brightness gain = electronic gain (=flux gain from accelerating anode and output phosphor) x minification gain (input diameter^2 / output screen diameter ^2) Conversion factor is a measure of brightness gain = ratio of intensity of output phosphor to input exposure rate High kVp decreases subject contrast Video Disc Recording (fluoro only on long enough to make 1 frame on TV monitor) reduces pt dose by 95% Spot Films w/ conventional cassettes >100 mA, high resolution, short exposure Photospot camera less quality than conventional, ½ to 1/3 dose of spot films Grids reduce scattered radiation from patient Patient = source of scattered radiation, produced via Compton interaction High kVp, large field size, thick body part increases scattered radiation Magnification use higher voltage from electrostatic lens. 6inch mode mags from 9inch mode & increase pt dose by 2.25 Quantum mottle = quantum noise = scintiliation = more visible in high resolution, high contrast sys Modulation Transfer Function (MTF) expresses resolution (cesium iodide 4 lp/mm) Brightness increases by square of kVp increase Horizontal resolution depends on bandwidth Vertical resolution depends on Kell factor (=vertical resolution / # scan lines [525]) F-number(=focal length/lens diameter); lower f# -> more light, faster lens, less pt radiation exposure Brightness gain and contrast gain tested annually Use lowest mA and highest kVp to minimize dose Monitor TV monitor brightness/contrast daily Wear lead if likely to receive more than 5mRad/hr Overexposure notification: immediate if 25rem total, 75rem eye, 250rem skin; 24hr if 5rem total, 15rem eye, 50rem skin Max occupational exposure = 5 rem/yr (over 18yo; <18yo = 10% of that dose) General public exposure max 0.1 rem/yr = 0.002 rem/hr High radiation area = may receive >0.1 rem/hr at 30cm from source Radiation area = may receive > 0.005 rem/hr at 30cm from source Gray = SI unit of absorbed dose: 1Gy = 100 Rads = 1J/Kg Rad (Radiation Absorbed Dose) = special unit of absorbed dose: 1Rad = 100ergs/gm 9 FLUOROSCOPY EXAM CHEAT SHEET OMAR CHOUDHRI - Rem = specialunite of dose equivalent (Rads x quality factor) : 1rem=0.01 sievert Sievert = SI unit of dose equivalent (Gray x quality factor) Quality factor for xray =1 Photoelectric effect = collision of xray photon with inner orbital electron, knocking electron out of orbit, photon loses all energy Compton scattering = xray photon hits outer orbital electron and making recoil electron and decreasing photon energy Pair production = incident photon is annihilated producing electron/positron pair 10-14 days after onset of menses = improbably for pregnancy Max dose during 9 months of pregnancy = 0.5 rem (+0.05rem if knowledge delayed) Ok if additional dose to embryo no more than 0.05rem during remainder of pregnancy if dose is already 0.5rem at time of declaration of pregnancy Phantom = 9inch water or 7.9 inch lucite 10