STEAM-Burst: a single-shot, multi-slice imaging sequence without rapid gradient switching Y. Crémillieux1, C. Wheeler-Kingshott2, A. Briguet1 and S.J. Doran2 1Laboratoire de RMN, CNRS UPRESA 5012, UCB Lyon 1CPE, 69622 Villeurbanne Cedex, France 2Department of Physics, University of Surrey, Guildford, Surrey, GU2 5XH, England Running Head: Single-shot, multi-slice imaging using STEAM-Burst Proofs to: Dr. S.J. Doran Department of Physics, University of Surrey, Guildford, GU2 5XH, England Tel. : Fax: e-mail: +44 1483 259413 +44 1483 259501 S.Doran@surrey.ac.uk 1 Abstract The STEAM-Burst sequence is a single-shot, multi-slice imaging technique which does not involve rapid gradient switching. A Burst excitation pulse train is followed by a 90 hard pulse and, after a mixing time, by a 90 slice-selective pulse. A read gradient refocuses a set of stimulated echoes, which can be phase-encoded to form an image. By repeating the selective pulse N times, each time with the carrier frequency offset differently, it is possible to sample N slices in a single-shot. A comparison is made of the sequence with other 3-D single-shot methods. Experiments implementing the technique on a 3 T whole-body imaging system and a 2 T, 31 cm bore animal imager are described. Both phantom and brain images are presented. The principal advantages of the new sequence are its speed, the absence of rapid gradient switching and corresponding freedom from artifacts, its insensitivity to static magnetic field inhomogeneities and its low acoustic noise. The main disadvantages are the low signal-to-noise ratio of the images produced and the concomitant limitation in resolution. Key words: 3D imaging, ultra-fast imaging, Burst, STEAM. 2 Introduction As MRI tackles ever more challenging applications, the role of ultra-rapid imaging sequences is becoming more prominent. The undoubted success of fast single-slice techniques has turned attention towards the development of new 3-D and multi-slice sequences, in order to meet the demands of imaging extended 3-D regions of the body in times as short as a few hundred milliseconds. Ultra-rapid multi-slice imaging will be a key development in the study of the thorax and abdomen, where cardiac and respiratory motion cause artifacts in the images acquired with conventional sequences. Both the diagnosis of heart disease and the study of the gastro-intestinal system will benefit. Another area of research demanding fast multi-slice sequences is the study of the perfusion of pathological tissues (for example, in the brain). To perform dynamic-contrast bolus-tracking experiments, it is necessary to match one’s imaging speed to the transit times, typically a few seconds, which characterise the perfusion of contrast agent into the tissues. Finally, functional MRI is an immense domain of investigation. The activated areas of the brain do not necessarily lie in the 2-D planes which are examined by today’s ultra-rapid sequences and a 3-D, single-shot, multi-slice technique would prove invaluable. Burst (1) is the generic name for a class of pulse sequences involving the application of multiple r.f. pulses under a constant gradient and the subsequent refocusing of a set of echoes. The possibility of using Burst to obtain ultra-fast images without the need for rapidly switched gradients is a very exciting one. Whilst Burst sequences are well known to produce images with poor signal-to-noise, they have a number of advantages over the standard fast imaging methods such as EPI 3 (2), Turbo-FLASH (3), RARE (4) and GRASE (5): they are free from artifacts due to inhomogeneity of the main magnetic field B0; they do not involve a high r.f. power deposition; they can be made acoustically very quiet. Previous ultra-fast 3-D imaging sequences We define as “single-shot” techniques only those which are able to acquire the whole set of data “in one repetition time”, without the need to wait for magnetisation recovery by T1 relaxation between the sampling of successive lines or planes of kspace. For 3-D imaging, only three pulse sequences satisfying this criterion have been found in the literature: (i) Echo-Volumar Imaging (EVI) (6) is an EPI-based experiment, which has recently been applied to the study of gastric filling and emptying (7). Single-shot abdominal images have been obtained (64328 voxels in 102 ms, with a resolution per plane of 6410 mm3). However, the large number of gradient switches involved in this technique make it suitable only for specially designed systems with resonant or ultrashort ramp-time gradients. (ii) URGEVI is a recently introduced hybrid of the URGE (Ultra Rapid Gradient Echo) variant of Burst (8) with switched, EVI-like gradients. Heid describes, in abstract form (9), the acquisition of 16 partial Fourier images of 5664 in 132 ms, with a resolution of 5.75 mm2. As in EVI, the images produced are T2* weighted and the method is suitable for bolus-tracking, abdominal scanning and kinematic joint movement. 4 (iii) Frequency-shifted Burst (10) combines a gradient-refocused Burst experiment with a standard 3-D phase-encoding. Whilst the acquisition speed is an order of magnitude slower than the above methods (484848 voxels in 3.1 s, with a resolution of 4.54.54.7 mm3) and whilst the sequence is repeated for each phaseencode step in the third dimension, the frequency shifting principle means that no T1 relaxation delay is required between repetitions. This technique, too, has been used successfully for bolus-tracking experiments (11). Against this background, the STEAM-Burst (12) imaging sequence is a promising, alternative technique, giving single-shot, multi-slice images. It is the combination of the STimulated Echo Acquisition Mode (13) and the Burst excitation pulse train. Here, we present phantom and human brain images to demonstrate that this new sequence is well suited for ultra-fast multi-slice acquisition. Results are shown from both a highfield whole-body scanner and a conventional small-bore animal imaging system. 5 Theory The sequence on which STEAM-Burst is based is the DUFIS (DANTE Ultra Fast Imaging Sequence) variant of Burst (14). A DANTE (15) excitation pulse train of n low flip angle -pulses is applied in the presence of a constant gradient in the read direction. We modify DUFIS as shown in Fig. 1 by replacing the slice-selective 180 refocusing pulse with a non-selective 90. N slice-selective 90 pulses are then used to read out N separate echo trains, with each train of n stimulated echoes corresponding to one image. We have also changed the location of the phaseencoding gradient; by making it coincide with the excitation instead of with the repeated acquisition, we reduce the number of gradient switches. In addition, kspace is now sampled parallel to kread and not at an angle, as in DUFIS. In Fig. 1 the time between the last pulse and the centre of the 90° hard pulse is indicated by T0. The magnetisation stays along the longitudinal axis for the mixing time, TMj, until a 90° soft pulse selects a slice and tips its magnetisation back into the transverse plane. The length of TMj is different for each slice, j. During this delay, a gradient along the slice direction dephases any residual transverse magnetisation. The 90° soft pulse is applied together with a slice-selective gradient followed immediately by a slice-refocusing lobe. (This change from the original STEAM-Burst proposal (12), in which the slice-refocussing lobe was placed after the excitation, reduces the signal attenuation due to diffusion, but it increses the number of gradient switches.) The readout gradient refocuses one stimulated echo from each excitation pulse. Sequence timings are arranged such that the time between the (n-i+1)th pulse and the centre of the 90° hard pulse is equal to the time from the centre of the selective 90° pulse to the i th echo. Each echo corresponds to a different line of the kspace. The dotted box of Fig. 1 represents the multi-slice loop which is repeated N 6 times to sample N images. Each time the program goes through this loop, the transmit frequency is shifted according to the position of the slice to be acquired. In order to avoid disruptive interference of higher order echoes with the primary echoes, must be small. This leads to a low S/N ratio, which is particularly significant when acquiring stimulated echoes which give only half the signal of the corresponding DUFIS spin echoes. This problem can be reduced by phase-modulating the Burst pulses, as suggested by previous authors (16, 17) and, here, we have used Zha and Lowe’s “two-phase OUFIS” modulation scheme. The signal obtained is affected by decay processes. The major problem arises from the diffusion of spins in the presence of the strong excitation and readout gradients (18). In addition, T1 and T2 relaxation affect the signal decay with consequent influence on the weighting of each image. The amplitude Aij of the i th echo in the j th slice is reduced by these decays as described in the following form: 2T 0 i TM j 2 Aij exp D G read i2 ij i exp exp , 3 T2 T1 [1] where the D is the self-diffusion coefficient, Gread the amplitude of the excitation and readout gradients, ij and ij the Stejskal-Tanner (19) parameters as applied to the STEAM-Burst experiment (see Fig. 1) and the distance between the centres of two successive excitation pulses or echoes. T0 is determined by the echo time for the first echo, chosen by the user. i depends on the echo number, i, and TMj depends on the slice number, j, while ij depends both on the echo position and on the slice number as follows: 7 TM j TM1 ( j 1)Ts i 1 i 2 2 t ramp [2] ij 2T0 TM j i 1 . TM0 is the mixing chosen by the user for the first slice, tramp is the gradient ramp time and Ts is the slice repetition time, indicated in Fig. 1, which is the interval between any two corresponding points of two successive slice acquisitions. Equation [1] represents the application of Tanner’s standard formula (20) to the case of Burst. Equation 1 is stated for a uniform sample, for which D, T1 and T2 do not vary with position. However, for a more complex sample, we may simply regard Aij as being the contribution to the echo amplitude from an isochromat at position (x,y,z). The resultant amplitude is found by integrating Eq. [1] over the slice considered. All the images have the same T2 weighting, but a different diffusion and T1 weighting . Moreover, each of the echoes forming an image has a different diffusion and T2 attenuation, but the same T1 attenuation. This gives rise to an image contrast and blurring which is a complicated function of T1, T2 and D, particularly where these parameters are spatially dependent. The acquisition bandwidth, BW, has a role in determining the decay behavior of the signal, since it determines the values of Gread, i and ij and it is closely related to the final imaging speed. Figure 2 plots the decay of the signal amplitude, Aij, described by Eq. [1], as a function of the echo number; the joined lines correspond to the theoretical curves of slices 1, 3 and 5, determined using an acquisition bandwidth of 200 kHz (sample period = 5 s per complex point), while the dotted curves are the 8 corresponding decays for a bandwidth of 100 kHz (sample period = 10 s per complex point). Note the gain in using the higher BW, though against this must be set the fact that S/N 1/ BW. 9 Method and Results The STEAM-Burst imaging sequence has been implemented on two different MRI scanners: a whole-body 3 T system (Surrey Medical Imaging Systems (SMIS), Guildford, England) and a 2 T, small-animal imager (Oxford Magnet, SMIS console). Hardware and experimental details have been summarised in Table 1. The predictions of Eq. [1] were tested by acquiring STEAM-Burst data without phase-encoding. In Fig. 3, we have plotted the normalised values of the signal amplitude from slices 1 () and 5 () out of a set of 6, together with the corresponding decays (solid lines), as predicted from Eq. [1]. The data were acquired on a water phantom, using the 3 T system with the phase encode gradient turned off. The 64 echoes were Fourier transformed in the read direction and, then, for a given point on the resulting profiles, Eq. [1] was fitted to the decay data using the LevenbergMarquardt least-squares algorithm. For the curves shown, we obtained values of D = 1.99×10-5 cm2 s-1 () and D = 2.16×10-5 cm2 s-1 (). It is important to note that the head gradient set used for these measurements gives rise to a field gradient whose magnitude varies by up to 5% over the imaging volume. Calculation of D using a nominal value of G can lead to errors. In addition, the equipment does not currently include facilities for monitoring or stabilising the phantom temperature. For these reasons we did not attempt a rigorous comparison with literature values of diffusion coefficient. We are, as yet, unclear why the data for the first slice () appear noisier then for the subsequent ones () and why the earlier echoes appear noisier than subsequent ones. In all the imaging studies presented below, image acquisition was truly single-shot; no averaging was performed. Figure 4 shows a set of six full k-space 64 64 STEAM- 10 Burst images of the water phantom used to test the sequence on the 3 T system, whilst Fig. 5 shows the results of using the same sequence on a human head. In both cases, all six images were acquired in a total time 180 ms (i.e., 20 ms per image, with the remaining time being taken by the various slice-selective pulses, gradient ramps, etc.). The image resolution is 4 4 mm2 for the phantom images and 5 5 mm2 for the head images. Figure 6 shows six half k-space images of a lemon, on the 2 T instrument; the nominal resolution of these images is 1.25 1.6 mm2. All the other imaging parameters are given in Table 1. Discussion We have described the implementation of the STEAM-Burst sequence on two different MR scanners, demonstrating its performance as a single-shot multi-slice technique. Good, distortion-free and artefact-free images are obtained both on a water phantom, where diffusion is heavily present, and on the human brain, where diffusion is reduced, but T2 decay is increased. The images on the 2 T animal system demonstrate that the STEAM-Burst sequence can be used for higher resolution studies, but they highlight a number of problems, which will be discussed further below. All the images presented here show a good definition of the samples consistent with the nominal pixel resolutions given in Table 1. White and gray matter are distinguishable in the human brain images, where it is possible to recognise internal structures. The only major image artifact is a vertical line, in the head data, arising from local radio frequency interference, which will, in future, be screened out. 11 The STEAM-Burst sequence has a number of attractive features: (i) Speed — We acquire multi-slice data very rapidly (66464 full k-space head images in an acquisition period of 180 ms — 210 ms including excitation period). (ii) Gradient Performance — As a general rule, the higher the data sampling rate, the stronger will be the gradient needed. This is true for most fast imaging sequences and Burst is no exception. Here, γGread = BW / FOVread and so for our ultra-fast experiments ( 20 ms / image) we do need strong gradients (14.6 mT m-1 for our head images). However, the fact that the gradient does not need to be switched rapidly means that Burst can be implemented on systems without specialised “EPI” gradient hardware and that the images do not suffer from artifacts due to eddy currents. Notice in Table 1 that the head images were acquired with a gradient ramp time of 1 ms — i.e. only 16 T m-1 s-1. (iii) Inhomogeneities in B0 — Since each echo is independently r.f. refocussed (i.e., a true stimulated echo), images do not exhibit distortion or signal “drop-out” in regions of inhomogeneous B0. STEAM-Burst is particularly robust in regions of poor shimming and/or rapid changes in susceptibility across the sample. In Fig. 5, we obtain good images from low down in the brain, where EPI images would show considerable signal loss around the air spaces in the sinuses. Notice that this advantage would render the implementation of STEAM-Burst described here unsuitable for functional imaging, which uses T2* contrast. However, by adding a negative gradient lobe in the read-direction immediately after the acquisition, one can obtain a T2*-weighted image in addition to the standard T2-weighted one, albeit with a further reduced S/N. 12 (iv) Acoustic Noise — Preliminary data indicate that Burst can be made very quiet. In certain frequency bands, we have observed Burst to be 15 dB quieter than an equivalent EPI sequence. However, two major problems exist with Burst sequences: (i) Low Signal-to-Noise — To overcome the low S/N typical of all Burst experiments, we find that a phase-modulation scheme for the -pulses is essential. With the implementation of Zha and Lowe’s optimised two phase OUFIS scheme (17), our 3T head images have a S/N which varies between 15:1 to 20:1, whilst the lemon images at 2 T have S/N of approximately 6:1. (ii) Signal Decay — Equation [1] shows that D, T1 and T2 are all potentially important factors governing the viability of the technique. Due to the acquisition of stimulated echoes, T2* does not contribute to the observed signal decay. Table 2 shows the signal attenuation caused by each of these three factors, calculated with Eq. [1], on the middle echo of successive slices for brain tissue and water respectively, assuming an ideal experiment. In the brain, T2 is the most significant factor and limits the maximum length of the echo train to approximately 80 ms; by contrast, T1 does not have a major effect over the images, due to its long value (more than 1.5 s at 3 T) compared with the total acquisition time. 13 Limits to resolution in STEAM-Burst Ultimately, the most severe limit on resolution is low signal-to-noise. Without taking any other complicating factors into account, simply reducing the voxel size from 5510 mm3 as in the current head images to 2.52.55 mm3, would entail a reduction in S/N by a factor of 8 (i.e., images with S/N 2). In addition, we must consider the signal decay and the need to add more r.f. pulses to the Burst train. The resolution of Burst images is limited both along the read and the phaseencode direction. To achieve a better resolution along kread, a higher number of sample points per echo should be acquired, while to improve the resolution along kphase it would be necessary to increase the number of excitation pulses. In both the cases, according to Eq. [1], this would mean longer acquisition times, with a more evident signal decay within each slice and, also, a quicker degradation of the signal as a function of the slice number. (Increasing the number of pulses also has the undesirable consequence that one must reduce the elementary flip angle , which will further reduce the signal-to-noise of the image.) The signal decay due to diffusion is particularly significant for small-bore systems (e.g., in animal imaging), because of the larger gradient values used, and may well prove to be a fundamental limit to the resolution attainable using the technique. The 2 T lemon images were acquired using 25 r.f. pulses to shorten the sequence and so lessen the signal decay, increasing S/N by using a larger flip angle . To overcome the resulting problem of a poor resolution along the phase encode direction it is possible to use asymmetric k-space scanning strategies. The lemon images demonstrate this; a matrix of 64 50 has been reconstructed from the 25 echoes of a half k-space acquisition. 14 The number of pulses for the 3T experiments (Fig. 4 and 5) has been set at 64 and k-space has been sampled symmetrically throughout, so the images have a matrix size of 6464. An asymmetric sampling could increase the size to 64128. At 3 T it should be possible to extend the number of sample points per echo to 96, without paying too much in S/N and signal decay, so that our best matrix size would be about 96 128. Experiments to achieve this are still in progress. The signal decay process limits the number of slices which can be acquired. At 3 T it is possible to scan 8–10 slices before late echoes start to be unacceptably attenuated in the final slice. Unfortunately, there is little that can be done to overcome this problem, apart from trying to optimise all the timings to make the sequence as short as possible. In practice, the large slice width and the finite active volume of the head coil can also limit the number of slices. Conclusions In this study, we propose a novel method combining stimulated echo acquisition and BURST excitation. This new sequence has been applied to ultra-fast multi-slice acquisition and has been demonstrated on phantom as well as in vivo experiments. Multi-slice STEAM-Burst represents one approach to optimise volume acquisition in a single-shot with BURST-type sequences. It has proved to be suitable for ultra-fast imaging on small bore as well as on whole-body systems. The technique has two main limitations: low signal-to-noise and signal decay due to T2 and diffusion. At present it seems that 8–10 is the maximum number of slices which may be acquired at 3T whilst maintaining adequate S/N in the later images. Image weighting can be introduced with preparation pulses, prior to the STEAMBurst experiment. The diffusion decay, which is a problem for this method, can be 15 exploited in experiments to measure self-diffusion coefficients, as previous work has demonstrated (21). The STEAM-Burst sequence offers the possibility for a wide range of experiments which are still under investigation. Preliminary results (not shown) indicate that by using an acquisition bandwidth of 400 kHz, we can reduce the time for 6 images of 6464 pixels to 118 ms. References 1. J. Hennig, M. Hodapp, Burst imaging. MAGMA 1, 39-48 (1993). 2. P. Mansfield, Multi-planar image formation using NMR spin echoes. J. Phys. C: Solid State 10, L55-L58 (1977). 3. A. Haase, J. Frahm, D. Matthaei, W. Hänicke, K.-D. Merboldt, FLASH imaging . Rapid NMR imaging using low flip-angle pulses. J. Magn. Reson. 67, 258-266 (1986). 4. J. Hennig, A. Nauerth, H. Freidburg, RARE imaging: a fast imaging method for clinical MR. Magn. Reson. Med. 3, 823-833 (1986). 5. K. Oshio, D. A. Feinberg, GRASE (gradient- and spin-echo) imaging: a novel fast MRI technique. Magn. Reson. Med. 20, 344-349 (1991). 6. P. Mansfield, P. R. Harvey, M. K. Stehling, Echo Volumar Imaging. MAGMA 2, 291-294 (1994). 7. P.R. Harvey, P. Mansfield, Echo-Volumar Imaging (EVI) at 0.5 T: First wholebody volunteer studies. Magn. Reson. Med. 35, 80-88 (1996). 8. O. Heid, M. Deimling, W. J. Huk, Ultra-Rapid Gradient Echo Imaging. Magn. Reson. Med. 33, 143-149 (1995). 16 9. O. Heid, Ultra Rapid Gradient Echo Planar/Volumar Imaging, in “Proc., SMRM, 3rd Annual Meeting, Nice, 1995,” p. 98. 10. J.H. Duyn, P. van Gelderen, G. Liu, C.T.W. Moonen, Fast volume scanning with frequency-shifted Burst MRI. Magn. Reson. Med. 32, 429-432 (1994). 11. J.H. Duyn, P. van Gelderen, P. Barker, J.A. Frank, V.S. Mattay, C.T.W. Moonen, 3D bolus tracking with frequency-shifted BURST MRI. J. Comput. Asst. Tomog. 18 (5), 680 (1994). 12. Y. Crémillieux, B. Sola, O. Beuf, A. Briguet, STEAM-Burst: Application to sub-second multi-slice acquisition, in “Proc., SMRM, 4th Annual Meeting, New York, 1996,” p. 1500. 13. J. Frahm, K. Merboldt, W. Hänicke, A. Haase, Stimulated echo imaging, J. Magn. Reson., 64, 81–93 (1985) 14. I. J. Lowe, R. E. Wysong, DANTE Ultrafast Imaging Sequence (DUFIS). J. Magn. Reson. Ser. B 101, 106-109 (1993). 15. G.A. Morris, R. Freeman, Selective excitation in Fourier transform nuclear magnetic resonance. J. Magn. Reson. 29, 433-462 (1978). 16. P. van Gelderen, J.H. Duyn, C.T.W. Moonen, Analytical solution for phase modulation in Burst imaging with optimum sensitivity. J. Magn. Reson. Ser. B 107, 78-82 (1995 ). 17. L. Zha, I.J. Lowe, Optimized Ultra-Fast Imaging Sequence (OUFIS). Magn. Reson. Med. 33, 377-395 (1995). 18. S.J. Doran, M. Décorps, A robust, single-shot method for measuring diffusion coefficients using the “Burst” sequence. J. Magn. Reson. Ser. A 117, 311-316 (1995). 17 19. E.O. Stejskal, J.E. Tanner, Spin diffusion measurements: Spin echoes in the presence of a time-dependent field gradient. J. Chem. Phys. 42 (1), 288-292 (1965). 20. J. E. Tanner, Use of the stimulated echo in NMR diffusion studies. J. Chem. Phys. 52 (5), 2523-2526 (1970) 21. K.-D. Merboldt, W. Hänicke, J. Frahm, Diffusion imaging using stimulated echoes. Magn. Reson. Med. 19, 233-239 (1991). 18 T0 T0 TMj 90° 90° RF and Signal n -pulses n STE Gslice i ij Gread Gphase Ts N slices Fig. 1. The STEAM-Burst sequence diagram. Multi-slice acquisition is based on the repetition of the right part of the sequence (rectangular box). The timing parameters are explained further in the text with the corresponding values summarised in Table 1. 19 Fig. 2. Theoretical signal decay in a single-shot multi-slice STEAM-Burst experiment. The calculated attenuation of the signal takes into account diffusion, T1 and T2 decays as predicted by Eq. [1], assuming the parameters of the experiment at 3T. Dotted line 100 kHz acquisition bandwidth; solid line 200 kHz acquisition bandwidth. The curves are numbered according to the slice they refer to. All curves are normalised to the amplitude of echo 1 of slice 1 with a 200 kHz bandwidth. 20 Fig. 3. Plot showing the signal decay versus echo number in a STEAM-Burst experiment. Data were acquired without phase encoding and each echo was Fourier transformed separately. The “amplitude” on the y-axis is the magnitude of a given point on the resulting profile. Similar results would be obtained if the magnitude of the corresponding echo were used. = data from the slice 1, = data from the slice 5. The solid curves represent the fit of Eq. [1] to the data. Both curves are normalised to the theoretical (fit) value at echo 1 for slice 1. The reduced initial value for echo 1 of slice 5 is approximately that predicted by Eq. [1]. An exact comparison is not possible because of the (slight) spatial inhomogeneity of the phantom, which meant that slice 5 contained a different amount of material from slice 1. 21 Fig. 4. Single-shot, multi-slice STEAM-Burst images of a water phantom at 3T. The sample diameter is 200 mm and the field of view 250 mm. 6 slices of 5 mm slice thickness have been acquired in a total acquisition time of 180 ms. Other parameters are as in Table 1. The original 64 64 images have been interpolated to 256 256. 22 Fig. 5. Single-shot, multi-slice STEAM-Burst images of a human brain at 3T. 6 slices of 10 mm slice thickness have been acquired in a total acquisition time of 180 ms. Other parameters are as in Table 1. The original 64 64 images have been interpolated to 256 256. 23 Fig. 6. Single-shot, multi-slice STEAM-Burst images of a lemon at 2 T. 6 slices of 5 mm slice thickness have been acquired in a total acquisition time of 267 ms. Other parameters are as in Table 1. The original 64 50 images have been interpolated to 128 128. 24 Table1. Hardware details and experimental parameters System 1 System 2ystem 2 Hardware 80-cm, whole-body, 3T (127 MHz) SMIS imager 30 cm horizontal bore, 2 T (85.1 MHz) magnet, SMIS console Gradient strength (Gread) 14.6 mT m-1 (human brain) 12 mT m-1 (phantom images) 15 mT m-1 Ramp time (tramp) 1 ms 1 ms No of Burst pulses (= number of echoes) 64 25 (half k-space) Burst pulse length 10 s (phantom images) 17 s (human brain) 27 s Phase scheme two-phase OUFIS two-phase OUFIS Flip Angle () 8 15 Inter-pulse spacing ( 320 s 1280 s Total excitation time = acq. time per image 20.5 ms 32 ms T0, TM1, Ts 7.7 ms, 3 ms, 32 ms 7.6 ms, 4.1 ms, 44 ms Acquisition bandwidth 200 kHz 50 kHz Sample points / echo 64 64 Processed image matrix Size 64646 64506 Nominal Image Resolution 44 mm2 (phantom images) 55 mm2 (head images) 1.251.6 mm2 Slice thickness 5 mm (phantom images) 10 mm (head images) 5 mm Number of slices 6 6 25 Total acquisition time 180 ms (210 ms including excitation period) 267 ms (311 including excitation period) Number of averages 1 1 Signal to noise ratio 19:1 to 24:1 (phantom images) 6:1 15:1 to 20:1 (head images) 26 Table 2. Attenuation of the Burst signal caused by T1, T2 and diffusion. Results are presented in terms of the amplitude of the ith echo compared with the amplitude of echo 1 (i.e., the echo attenuation). The intra-slice attenuation is given for echo 32 during acquisition of slice 1 to assess the signal at the point corresponding to the “zero phase-encode step” and for echo 64 of slice 1 to show the amplitude at the end of the echo train. The inter-slice attenuation is given between echoes 1 of slice 1 and slice 5 to illustrate the reduction in S/N of successive images. The separate evaluation of the contributions of T1, T2 and diffusion decay is based on Eq. [1], using the data values D = 2.2 10-5 cm2 s-1, T1 = 1.5 s, T2 = 1.3 s for water and D = 0.85 10-5 cm2 s-1, T1 = 1.35 s, T2 = 0.1 s for brain tissue. Water Phantom Decay process Brain Tissue Attenuation Echo 32 Attenuation Echo 64 Inter-slice attenuation Attenuation Attenuation Echo 32 Echo 64 Inter-slice attenuation T1 1 1 0.91 1 1 0.90 T2 0.98 0.97 1 0.81 0.66 1 D 0.94 0.72 0.73 0.97 0.88 0.88 Overall 0.92 0.70 0.66 0.79 0.58 0.79 27 Figure and Table Legends Fig. 1. The STEAM-Burst sequence diagram. Multi-slice acquisition is based on the repetition of the right part of the sequence (rectangular box). The timing parameters are explained further in the text with the corresponding values summarised in Table 1. Fig. 2. Theoretical signal decay in a single-shot multi-slice STEAM-Burst experiment. The calculated attenuation of the signal takes into account diffusion, T1 and T2 decays as predicted by Eq. [1], assuming the parameters of the experiment at 3T. Dotted line 100 kHz acquisition bandwidth; solid line 200 kHz acquisition bandwidth. The curves are numbered according to the slice they refer to. All curves are normalised to the amplitude of echo 1 of slice 1 with a 200 kHz bandwidth. Fig. 3. Plot showing the signal decay versus echo number in a STEAM-Burst experiment. Data were acquired without phase encoding and each echo was Fourier transformed separately. The “amplitude” on the y-axis is the magnitude of a given point on the resulting profile. Similar results would be obtained if the magnitude of the corresponding echo were used. = data from the slice 1, = data from the slice 5. The solid curves represent the fit of Eq. [1] to the data. Both curves are normalised to the theoretical (fit) value at echo 1 for slice 1. The reduced initial value for echo 1 of slice 5 is approximately that predicted by Eq. [1]. An exact comparison is not possible because of the (slight) spatial inhomogeneity of the phantom, which meant that slice 5 contained a different amount of material from slice 1. Fig. 4. Single-shot, multi-slice STEAM-Burst images of a water phantom at 3T. The sample diameter is 200 mm and the field of view 250 mm. 6 slices of 5 mm slice thickness have been acquired in a total acquisition time of 180 ms. Other parameters are as in Table 1. The original 64 64 images have been interpolated to 256 256. 28 Fig. 5. Single-shot, multi-slice STEAM-Burst images of a human brain at 3T. 6 slices of 10 mm slice thickness have been acquired in a total acquisition time of 180 ms. Other parameters are as in Table 1. The original 64 64 images have been interpolated to 256 256. Fig. 6. Single-shot, multi-slice STEAM-Burst images of a lemon at 2 T. 6 slices of 5 mm slice thickness have been acquired in a total acquisition time of 267 ms. Other parameters are as in Table 1. The original 64 50 images have been interpolated to 128 128. Table1. Hardware details and experimental parameters Table 2. Attenuation of the Burst signal caused by T1, T2 and diffusion. Results are presented in terms of the amplitude of the ith echo compared with the amplitude of echo 1 (i.e., the echo attenuation). The intra-slice attenuation is given for echo 32 during acquisition of slice 1 to assess the signal at the point corresponding to the “zero phase-encode step” and for echo 64 of slice 1 to show the amplitude at the end of the echo train. The inter-slice attenuation is given between echoes 1 of slice 1 and slice 5 to illustrate the reduction in S/N of successive images. The separate evaluation of the contributions of T1, T2 and diffusion decay is based on Eq. [1], using the data values D = 2.2 10-5 cm2 s-1, T1 = 1.5 s, T2 = 1.3 s for water and D = 0.85 10-5 cm2 s-1, T1 = 1.35 s, T2 = 0.1 s for brain tissue. 29 List of symbols TM0 italic , zero subscript Tmj italic , j subscript N italic , caps T1 italic , 1 subscript T2 italic , 2 subscript T2* italic , 2 subscript, * superscript n italic , small caps greek alpha kread italic kphase italic T0 italic , zero subscript Ts italic , s subscript Aij italic , ij subscript ith italic , th superscript jth italic , th superscript D italic, caps Gread italic , read subscript i italic , i subscript, greek delta, small caps ij italic , ij subscript, greek delta, caps italic , greek tau BW italic , greek gamma italic , caps 30