Questions of Suppliers concerning procurement part No. 4: System of flue gas analyzers for laboratory application. Question 1: Is it necessary to supply SO2 analyser with analysis method UV fluorescence or would it be possible to offer analyser with NDIR analysis method (the same like for CO-CO2 module)? Analysers with UV fluorescence method are used in ambient air quality measurements! UV fluorescence analysis method is not compliant with any standard methods for Emission measurement => it is much more efficient to add SO2 module into the NDIR analyser with CO + CO2 measurement. Answer: We would like to mention that the main purpose of procuring the system of flue gas analysers is to purchase an equipment intended for investigation of combustion processes of different kind of biofuel and solid recovered fuel. Therefore the range of emissions and its’ concentrations could be very variable. To be sure that we will be able to detect the emissions in quite wide concentrations range we worked out technical specification of equipment that could operate in wide range of detection, would be sensible and would apply detection method, which has the lowest interference from other possible emissions. In this case the UV fluorescence method for SO 2 analyses in combustion products meets our basic requirements for the equipment. According to the literature of measuring technique, our own experience and experience of other similar laboratories, UV fluorescence and NDIR methods are most suitable for SO2 analyses. However UV fluorescence method has advantages when low values of SO 2 are measured and has wider measuring ranges. We should particularly mention that all analyzers (except CO and CO 2) should operate individually (separately) and not be part of each other, in order to have independent operation in case if one or another analyser has encountered technical problems, or is under technical supervision or repair. Question 2: Would it be possible to offer heated line with 4/6 mm tube instead of 6/8 mm? Solution with 4/6 mm is better because of shorter response time. 6/8 mm heated line doesn’t provide any advantages. Answer: It is possible to offer either 4/6 mm or 6/8 mm heated line, however this choice should not interfere in reliable and appropriate function of the whole offered system. Question 3: Our solution doesn’t need any additional pump except if distance is longer than 50 m. Could you specify the length of heated line? Is it possible to omit additional pump (item 3) case if the length of heated line will be shorter than 50 m? Answer: We have doubt that it will be possible to control and supply the flue gas for five analysers with one pump due to different inlet conditions of every analyser. The relevant answer is presented below in an answer to question 5. Question 4: Whether data acquisition system and software for data collection and data acquisition shall be delivered together with the equipment? Answer: The data acquisition system and software for data collection is not required in the tender technical specification. Question 5: Quantity of sample probes, sample lines and analyser protection cabinets: In our opinion it is not recommended to integrate all requested analysers into one common analyser protection cabinet. We would like to suggest to apply 2 cabinets, the first one for integration of the "hot" analysers (FID and Chemiluminescence) and the second one for the cold path of analysers (NDIR, UV and paramagnetic). Reasons: The hot 19"-analysers cannot be mounted one onto the other compactly because of the need of heat dissipation from hot surfaces. In summary we would have to implement 3 housings into the analyser protection cabinet, each of them consisting hot parts: - Housing of the FID analyser (177 mm high, hot surface); - Chemiluminescence analyser (133 mm high, hot surface of the converter); - 19"-Module of pumps and ozone scrubber (approx. 222 mm high, hot surface of the ozone scrubber). It is recommended to install the 19"-Module of pumps and ozone scrubber at the bottom of the cabinet and the analysers at a comfortable height but with a suitable space between both analyser housings for safe heat dissipation. If the cold analytical path consisting of 3 analysers, a sample gas cooler, a sample pump, filters and valves, would be integrated also into this protection cabinet then it would be very tight and overloaded. We fear that the heat dissipation would not be possible. This could be a problem. Please, advise the quantity of cabinets. pages 1 of2 It will be difficult to use one single sample probe and one sample line to feed all analysers from one source because of 4 different sample pumps (1 pump of the FID, 2 pumps of the Chemiluminescence analyser and a global sample gas pump for the cold analysers) will have an influence one to the others. This could also be a problem. Please, advise, if one probe is mandatory or if 2 probes are requested. Answer: We have not specified integration schemes of analysers into common system due to possible variety of analysers and this task is left for own decision of supplier or producer of the system. Technical specification of the system contains only basic requirements in accordance to measurements, which should be performed by the system. In this particular case we require that analysers integrated in such system would operate properly and reliably as long as possible without any failure caused by incorrect design of the system. Your point of view on system design by separating it in to two separate cabinets with two separate sample lines and probes is reasonable and acceptable. However, other suppliers may suggest other acceptable and reliable solutions for integration of analysers. Question 6: Protection class of the analyser mounting cabinet: Please advice, which kind of analyser mounting cabinet is mandatory for all bidders. This item is not described in the specification of the tender. In principle, there are some different solutions which can be realized: - 19"-rack(s) without side walls, without backside wall, without front door, without swing frame; - 19"-rack(s) with side walls, but without backside wall, without front door, without swing frame; - 19"-cabinet(s) with side walls, with backside wall, with swing frame, without front door; - 19"-cabinet(s) with side walls, with backside wall, with swing frame, with front door made of steel (IP 54); - 19"-cabinet(s) with side walls, with backside wall, with swing frame, with front door made of glass (IP54). Please, note that a cabinet with IP54 must be cooled by an electric air conditioning unit. Answer: Basic requirements for analysers mounting cabinet or cabinets are: - XX"-rack(s) with side walls, but without backside wall, without front door, without swing frame (IP20) Question 7: Analysis method for SO2 analysis: The specified UV fluorescence method is a typical emission monitoring method, mainly applied in the USA. Only one analyser is able to measure in the higher ppm range (Teledyne Model 6400EH). All other known analysers are designed for ppb ranges and small ppm ranges only and cannot be applied for this tender. Common methods of SO2 analysis used in Europe are NDIR and NDUV. These methods are used in thousands of emission monitoring systems in European countries. A lot of European manufacturers are producing NDIR analysers. NDIR analysers are robust, very stable over a long time and need only low maintenance efforts. Disadvantages of NDIR and NDUV are the limited possibility to measure low values of SO2 and the limitations concerning the measuring ranges (depends on the internal optical path length in the sample cell). Answer: This question is already analyzed in answer to Question 1. pages 2 of2