THE TITLE PAGE Cyromazine GRM029.01A DRAFT COPY Cyromazine (CGA72662) - Analytical Method for the Determination of Residues of Cyromazine and its Metabolite Melamine (C1803) in High water Content Crops. Final Determination by LC-MS/MS Analytical Method AUTHORS: S L Hargreaves EFFECTIVE DATE: DATA REQUIREMENT: EPA OPPTS 860.1340 EC Guidance Document SANCO/3029/99 rev 4 EC Guidance Document SANCO/825/00 rev 7 PERFORMING LABORATORY: Syngenta Ltd. Jealott’s Hill International Research Centre Bracknell, Berkshire, RG42 6EY, UK www.syngenta.com SPONSOR: Syngenta Ltd. Jealott’s Hill International Research Centre Bracknell, Berkshire, RG42 6EY, UK www.syngenta.com Summary of revisions to previous version Version Summary of Revisions GRM029.01A This confirmatory method supersedes method REM174.02 (Reference 1) for analysis of high water content crops. The method includes a cation exchange SPE procedure and final determination is by LC-MS/MS with a confirmatory transition. Authorisation Authorised by : DRAFT GRM029.01A S Hadfield Technical Manager Date Product Metabolism Page 2 of 57 Abbreviations and symbols Abbreviation A Å a.i. amt amu C CAS CFR cm DA[#]A EPA EU FIFRA ft g gal GC GLP GRM ha HPLC i.d. ID in IUPAC kg L lb LC LC-MS/MS LOD LOQ m Definition acre angstrom (10-8 m) active ingredient amount atomic mass unit Celsius or centigrade Chemical Abstract Services Code of Federal Regulations centimetre days after application, [#] = 1, 2, 3 etc., if there are multiple applications Environmental Protection Agency (U.S.) European Union Federal Insecticide, Fungicide and Rodenticide Act (U.S.) foot (feet) gram gallon gas chromatography Good Laboratory Practice Global Residue Method hectare high performance liquid chromatography inside diameter identification inch International Union of Pure and Applied Chemistry kilogram litre pound liquid chromatography liquid chromatography - tandem mass spectrometry/mass spectrometry limit of detection limit of quantification meter/metre DRAFT GRM029.01A Page 3 of 57 Abbreviations and symbols (continued) Abbreviation µg µL µm mbar MDL mg min mL mm mmol mol ms MS MS/MS mV MW m/z na nd ng No. oz PMRA ppb ppm pg psi QAU 2 R (or r2) RSD RT s SD SPE Definition microgram microliter micrometer millibar method detection limit milligram minute millilitre millimetre millimole mole millisecond mass spectrometry tandem mass spectrometry/mass spectrometry millivolt molecular weight mass to charge ratio not applicable not detectable (below limit of detection) nanogram number ounce Pest Management Regulatory Agency (Canada) parts per billion or micrograms per kilogram parts per million or microgram per gram or milligrams per kilogram picogram pounds per square inch quality assurance unit square of correlation coefficient relative standard deviation retention time second standard deviation solid phase extraction DRAFT GRM029.01A Page 4 of 57 Abbreviations and symbols (continued) Abbreviation USDA UV V vol v/v wt w/v Definition United States Department of Agriculture ultraviolet volt volume volume/volume weight weight/volume DRAFT GRM029.01A Page 5 of 57 TABLE OF CONTENTS THE TITLE PAGE 1 TABLE OF CONTENTS 6 1.0 INTRODUCTION 9 1.1 1.2 Scope and chemical structures ................................................................9 Method summary ..................................................................................10 2.0 MATERIALS AND APPARATUS 2.1 2.2 2.3 2.3.1 2.3.2 2.4 Apparatus ..............................................................................................10 Reagents ................................................................................................10 Preparation of analytical standard solutions .........................................10 Fortification solutions ...........................................................................11 Standard solution storage and expiration ..............................................11 Safety precautions and hazards .............................................................11 3.0 ANALYTICAL PROCEDURE 3.1 3.2 3.3 3.4 3.5 3.6 3.7 Modifications and potential problems ...................................................13 Sample preparation................................................................................13 Extraction ..............................................................................................13 Sample Clean-up ...................................................................................14 Time required for analysis ....................................................................15 Method stopping points .........................................................................15 Preparation of calibration standards for LC-MS/MS ............................16 4.0 FINAL DETERMINATION 4.1 4.2 4.3 Instrument description ...........................................................................16 Chromatography conditions ..................................................................17 Mass spectrometer conditions ...............................................................17 5.0 CALCULATION OF RESULTS 5.1 5.2 Single point calibration procedure ........................................................18 Multi point calibration procedure..........................................................19 6.0 CONTROL AND RECOVERY SAMPLES 20 7.0 SPECIFICITY 20 7.1 Matrix interference ................................................................................21 DRAFT GRM029.01A 10 13 16 18 Page 6 of 57 7.2 7.3 Reagent and solvent interference ..........................................................21 Labware interference .............................................................................21 8.0 METHOD VALIDATION 8.1 8.2 8.3 8.4 8.5 Recovery data and repeatability ............................................................21 Limit of quantification (LOQ) ..............................................................21 Limit of detection (LOD) ......................................................................21 Detector linearity ...................................................................................22 Extract stability .....................................................................................22 9.0 LIMITATIONS 22 10.0 CONCLUSIONS 22 11.0 REFERENCES 23 APPENDICES SECTION APPENDIX 1 APPENDIX 2 APPENDIX 3 APPENDIX 4 APPENDIX 5 APPENDIX 6 APPENDIX 7 APPENDIX 8 21 24 APPARATUS .......................................................................................25 REAGENTS ..........................................................................................28 METHOD VALIDATION DATA ........................................................30 REPRESENTATIVE CHROMATOGRAMS ......................................36 DETECTOR LINEARITY GRAPHS ...................................................50 API4000 MS/MS TUNING PROCEDURE..........................................54 MS/MS PRODUCT ION SPECTRA ....................................................55 METHOD FLOWCHART ....................................................................57 DRAFT GRM029.01A Page 7 of 57 1.0 INTRODUCTION 1.1 Scope and chemical structures Analytical method GRM029.01A is suitable for the determination of cyromazine and its metabolite melamine (Figures 1 & 2) in high water content crop matrices. The limit of quantitation (LOQ) of the method has been established at 0.05 mg/kg. This method satisfies EU guidelines SANCO/3029/99 rev. 4, SANCO/825/00 rev. 7 and US EPA guideline OPPTS 850.7100. Figure 1 Common Name : Compound Code Number : CAS Number : IUPAC Name : Molecular Formula : Molecular Mass : Cyromazine CGA72662 66215-27-8 N-cyclopropyl-1,3,5-triazine-2,4,6-triamine C6H10N6 166.19 H N H2N N N N NH2 Figure 2 Common Name Compound Code Number CAS Number IUPAC Name Molecular Formula Molecular Mass : : : : : : Melamine C1803 108-78-1 1,3,5-triazine-2,4,6-triamine C3H6N6 126.12 NH2 H2N N N N NH2 DRAFT GRM029.01A Page 9 of 57 1.2 Method summary High water content crop samples are extracted by homogenisation with 0.025 M KH2PO4: in 0.04 M H3PO4 and by mechanical shaking following the addition of methanol. After centrifugation and filtration, aliquots equivalent to 0.2 g are taken through an Oasis MCX solid phase extraction (SPE) procedure. Final determination is by high performance liquid chromatography with triple quadrupole mass spectrometry detection (LC-MS/MS). The limit of quantification of the method is 0.05 mg/kg. 2.0 MATERIALS AND APPARATUS 2.1 Apparatus The recommended equipment and apparatus are listed in Appendix 1. Equipment with equivalent performance specifications may be substituted. 2.2 Reagents All solvents and other reagents must be of high purity, e.g. glass distilled/HPLC grade solvents and analytical grade reagents. Particular care must be taken to avoid contamination of the reagents used. Reagents of comparable purity may be substituted as long as acceptable performance is demonstrated. A list of reagents used in this method along with details of preparation of solutions is included in Appendix 2. 2.3 Preparation of analytical standard solutions It is recommended that the following precautions should be taken when weighing the analytical materials. 1. 2. 3. 4. Ensure good ventilation. Wear gloves and laboratory coat. Prevent inhalation and contact with mouth. Wash any contaminated area immediately. Prepare individual 200 µg/mL stock solutions for cyromazine and melamine by one of the following methods. Weigh out accurately, using a five figure balance, sufficient cyromazine and melamine analytical standard into separate amber “Class A” volumetric flasks (50 mL). Dilute to the mark with acetonitrile for cyromazine and melamine to give separate 200 µg/mL stock solutions. Alternatively, the appropriate volume of solvent to add to a known amount of standard material may be determined using the equation below. The standard concentration is corrected for its chemical purity. DRAFT GRM029.01A Page 10 of 57 W P 1000 C P = Standard purity in decimal form (P(%)/100) V = Volume of acetonitrile required W = Weight, in mg, of the solid analytical standard C = Desired concentration of the final solution, (g/mL) 1000 = Unit conversion factor V In this case, the standard material is weighed directly into an appropriate storage vessel. 2.3.1 Fortification solutions The stock solutions should then be diluted by serial dilution to 0.01 g/mL in acetonitrile and used for fortification and for preparation of calibration standards for LC-MS/MS analysis. Mixed cyromazine and melamine standards may be prepared similarly. It is recommended that the following solutions are prepared as required: 1.0 µg/mL, 0.1 µg/mL and 0.01 µg/mL. The preparation of LC-MS/MS calibration standards is discussed in Section 3.7. 2.3.2 Standard solution storage and expiration All standard solutions should be stored in a refrigerator or freezer when not in use to prevent decomposition and/or concentration of the standard. Standard solutions should be allowed to equilibrate to room temperature before use. An expiry date of six months is recommended unless additional data are generated to support a longer expiry date. 2.4 Safety precautions and hazards The following information is included as an indication to the analyst of the nature and hazards of the reagents used in this procedure. If in any doubt, consult the appropriate MSDS or a monograph such as ‘Hazards in the Chemical Laboratory’, edited by S G Luxon, The Chemical Society, London (Reference 2). DRAFT GRM029.01A Page 11 of 57 Solvent and Reagent hazards Methanol Acetonitrile KH2PO4 Harmful Vapour Highly Flammable Harmful by Skin Absorption Irritant to respiratory system and eyes Causes severe burns Syngenta Hazard Category SHC-C, S SHC-C, S N/A* OES Short Term (mg/m3) 310 105 N/A 260 70 N/A 3 OES Long Term (mg/m ) o-phosphoric acid Ammonium hydroxide (30% w/v) Acetic acid Harmful Vapour Highly Flammable Harmful by Skin Absorption Irritant to respiratory system and eyes Causes severe burns Syngenta Hazard Category SHC-C, S SHC-C, S SHC-C, S OES Short Term (mg/m3) N/A 24 37 N/A 17 25 3 OES Long Term (mg/m ) N/A - Not known. N/A* - Not known. Assume Syngenta Hazard Category SHC-C In all cases avoid breathing vapour. Avoid contact with eyes and skin. Cyromazine has been designated as Syngenta Hazard Category SHC-A and melamine as SHC-C. The hazard category scale rates highly toxic chemicals as SHC-E and non-toxic chemicals as SHC-A. An additional hazard category of S indicates the compound is a severe skin and eye irritant. DRAFT GRM029.01A Page 12 of 57 3.0 ANALYTICAL PROCEDURE 3.1 Modifications and potential problems a) Melamine is widely used in a number of applications and is found as a contaminant in wastewater (Reference 3). Melamine is a ubiquitous compound and may be present at significant concentrations in soil even where cyromazine has not been applied. It is recommended that prior to sample preparation, soil is removed as far as possible from crops which have had contact with soil e.g. potato crops. b) Spot contamination may also be an issue and it is recommended that equipment such as the macerator heads especially, vacuum manifold inlets and needles and sample concentrator needles are taken apart and cleaned thoroughly with 0.1 M HCl, ultra pure water and methanol before use. Disposable labware should be used where possible. c) Bottled HPLC grade ultra pure water is used to prepare the LC mobile phase, which produces a lower background noise in the MS/MS chromatograms than water taken from a laboratory water purification system. d) The condition of the LC-MS/MS instrument is also important to ensure optimum chromatography and sensitivity. Where sensitivity is poor, it is recommended that the mass spectrometer front plate and ionisation source be cleaned. 3.2 Sample preparation Samples should be prepared using an approved method of sample preparation for residue analysis. It is recommended that prior to sample preparation, soil is removed as far as possible from crops which have had contact with soil e.g. potato crops. 3.3 Extraction In order to verify method performance and allow recovery corrections to be made (if appropriate), fortified control samples should be included with each sample set. To each pre-weighed control crop sample, add the appropriate amount of standard solution containing cyromazine and melamine in acetonitrile. Let each sample stand for at least five minutes after fortification to allow the spiking solution to soak into the matrix before proceeding with the extraction procedure. At least one untreated control and two fortified control samples should be analysed with each sample set. The method is summarized in flow chart form in Appendix 8. The following extraction procedure is based on that given in REM179.02 (Reference 1). a) Accurately weigh a representative amount of crop (10 g) into separate plastic centrifuge bottles (250 mL size) and fortify recovery samples as required. DRAFT GRM029.01A Page 13 of 57 b) Add 0.025 M KH2PO4 in 0.04 M H3PO4 (50 mL) (50 mL minus the water content of the samples). Homogenise at high speed for 1 minute. Note: Estimate the percentage water content in each matrix type and hence the total volume of water in the 10 g sub-sample. E.g. for a 10 g sub-sample with 90% natural water content add 50 mL – (10 x 90/100) mL = 41 mL extraction solution. It is sufficient to round the natural water content to the nearest ten percent value. Any volume contraction due to mixing organic solvents with water and evaporation loss during extraction is considered to be negligible. The water content of matrices can be obtained from published sources. The relevant information can be obtained from the following USDA web site: http://www.nal.usda.gov/fnic/cgi-bin/nut_search.pl Alternatively, where information is not available from such sources, it may be necessary to determine the moisture content experimentally, following a suitable moisture content determination procedure. c) Rinse the macerator head with methanol (53 mL) to remove as much residual sample as possible into the sample container. Note: 53 mL methanol is the volume of methanol required to prepare a 100 mL mixture of 50:50 v/v methanol water, due to the contraction of solvents on mixing. d) Secure the lids and shake the samples on a flat bed mechanical shaker at a speed which visibly agitates the contents for 30 mins at room temperature. e) Centrifuge samples at a speed which separates the soil from the supernatant (e.g.3500 rpm for 5 minutes). The sample concentration is now 0.1 g/mL and 2 mL of extract is equivalent to 0.2 g crop. f) Using a disposable plastic syringe (10 mL size) carefully draw approximately 5-10 mL of the extract from Section 3.3 (e) above into the syringe barrel. Attach a disposable 25 mm 4.5 m PTFE syringe filter for aqueous samples to the tip of the syringe. Carefully push the extract through the filter, collecting the filtrate into a clean disposable glass test tube. 3.4 Sample Clean-up a) Take one Oasis MCX SPE cartridge (30 mg size, with 3 mL reservoir) for each sample to be analysed and place on a suitable vacuum manifold (e.g. IST Vacmaster). Add methanol (2 mL) and allow to percolate through under gravity or draw through under vacuum to the level of the top frit at a rate of approximately 1 mL/min, discarding the column eluate. Do not allow the cartridges to become dry. Add 0.1 M hydrochloric acid (2 mL) to the top of each cartridge and allow to percolate through under gravity or draw through DRAFT GRM029.01A Page 14 of 57 under vacuum to the level of the top frit at the same rate, again discarding the column eluates. Do not allow the cartridges to become dry. b) Using e.g. a Gilson p5000 pipetteman , accurately transfer a 2 mL aliquot of the sample from Section 3.3 (f) onto the cartridge and allow to percolate through under gravity or draw through under vacuum to the level of the top frit at a rate of approximately 1 mL/min. Discard the column eluate. c) Sequentially add 0.1 M HCl (aq) (2 mL), acetonitrile (2 mL) and methanol (2 mL) to the top of each cartridge and allow to percolate through under gravity or draw through under vacuum to the level of the top frit at a rate of approximately 1 mL/min. Discard the column eluates. d) Place suitable disposable glass test tubes (e.g. 15 mL size) under each port, as required, in the manifold rack. Add 95:5 methanol:30% ammonium hydroxide solution (2 mL) to the top of each cartridge and allow to percolate through under gravity or draw through under vacuum to the level of the top frit at a rate of approximately 1 mL/min. Collect the column eluate. e) Close the manifold taps. Add a further portion of 95:5 methanol:30% ammonia solution (2 mL) to the top of each cartridge and leave to stand for at least 5.0 minutes to allow equilibration of the slow ionic interactions to take place. Open the manifold tap and collect the column eluate. f) Evaporate the samples to dryness under a stream of air, in a sample concentrator at 35 oC. g) Add ultra pure water (5 mL) and ultrasonicate samples briefly to mix thoroughly. The final sample concentration is now 0.04 g/mL. h) Transfer aliquots into suitable autosampler vials for analysis by LC-MS/MS. Note: The above SPE procedures have been developed using columns from the stated manufacturer; however, it is possible to carry out the procedure using similar columns from other manufacturers. In all cases it is strongly recommended that the elution profile is checked prior to commencing analysis. This will rule out any variation between manufacturers’ products and between batches. Each batch of cartridges should be checked for interference prior to use. 3.5 Time required for analysis The methodology is normally performed with a batch of up to 15 samples. One person can complete the analysis of up to 15 samples in 1 day (8 working hour period). 3.6 Method stopping points The analytical procedure can be stopped at various points for overnight and weekend breaks unless otherwise specified in the analytical procedure. Acceptable method DRAFT GRM029.01A Page 15 of 57 recoveries will validate any work flow interruptions. Samples should be stored refrigerated in sealed containers where the analysis cannot be completed in a single day. 3.7 Preparation of calibration standards for LC-MS/MS Some matrix suppression of the instrument response for cyromazine and melamine has been observed in the crops tested using the above procedure in this laboratory (Table 5, Appendix 3). Matrix matched standards may be used to compensate for these effects, at the discretion of the study director. It is recommended however, that samples are quantified using reagent matrix matched calibration standards. Calibration standards for LC-MS/MS analysis are prepared from the fortification standards described in Section 2.3.1. To prepare e.g. a 0.002 g/mL mixed cyromazine and melamine reagent-matched calibration standard, a method blank (2 mL of 50:50 v/v methanol:0.025 M KH2PO4 in 0.4 M H3PO4) is taken through the SPE procedure as described in Section 3.4 to point 3.4 (f). Add 100 L of 0.1 g/mL mixed cyromazine and melamine in acetonitrile followed by 4.9 mL ultra-pure water. Mix thoroughly by ultra-sonicating the sample briefly and transfer to a suitable autosampler vial for final determination by LC-MS/MS. A calibration curve may also be generated to quantify cyromazine and melamine residues. Standards over an appropriate concentration range should be prepared as described above, using the requisite volumes of mixed cyromazine and melamine standard in acetonitrile. 4.0 FINAL DETERMINATION The following instrumentation and conditions have been found to be suitable for this analysis. Other instrumentation can also be used, though optimisation may be required to achieve the desired separation and sensitivity. The operating manuals for the instruments should always be consulted to ensure safe and optimum use. The method has been developed for use on the Applied Biosystems API 4000 LC-MS/MS. Final determination by LC-MS/MS with 2 transitions is considered to be highly specific and no further confirmatory conditions are included. 4.1 Instrument description Pump : Agilent 1100 series quaternary pump model number G1311A Degasser : Agilent 1100 series model number G1322A Column Oven : Agilent 1100 series model number G1316A fitted with column switching valve Detector : Applied Biosystems API 4000 triple quadrupole mass spectrometer with Analyst™ DRAFT GRM029.01A Page 16 of 57 software version 1.4.1 Autosampler : Agilent 1100 series model number G1313A Gas Supply : Peak Scientific NM20ZA gas station 4.2 Chromatography conditions Column Column Oven Temperature Injection volume Stop Time Injection protocol Mobile phase : : : : : Phenomenex Luna CN 5 µm 150 x 2.0 mm i.d. 40C 10 L 5.5 minutes Analyse calibration standard after 3 to 4 sample injections : Solvent 1 = Acetonitrile Solvent 2 = Acetic acid (0.2%, v/v) in ultra pure water Isocratic Mobile Phase Time (min) 0.0 5.5 % Solvent 1 30 30 % Solvent 2 70 70 Flow (mL/min) 0.5 0.5 Under these conditions the retention times of melamine and cyromazine are approximately 2.0 and 2.3 minutes respectively. The retention time of both cyromazine and melamine can vary over a period of time, depending on the condition of the HPLC column. 4.3 Mass spectrometer conditions Interface : TurboIonSpray Polarity : Positive Curtain gas (CUR) : Nitrogen set at 17 (arbitrary units) Temperature (TEM) : 500C Ionspray voltage : 4500V Collision gas setting (CAD) : Nitrogen set at 4 (arbitrary units) Gas 1 (GS1) : Air set at 60 (arbitrary units) Gas 2 (GS2) : Air set at 60 (arbitrary units) Interface heater (ihe) : On Scan type : Multiple reaction monitoring (MRM) DRAFT GRM029.01A Page 17 of 57 Cyromazine (primary) MRM Conditions Cyromazine Melamine (confirmatory) (primary) Melamine (confirmatory) Q1 m/z : 167 167 127 127 Q3 m/z : 85 68 85 68 Dwell Time : 150 ms 150 ms 150 ms 150 ms Resolution Q1 : Unit Unit Unit Unit Resolution Q3 : Unit Unit Unit Unit Declustering Potential (DP) : 66 V 66 V 66 V 66 V Entrance Potential (EP) : 10 V 10 V 10 V 10 V Collision Energy (CE) : 29 V 49 V 27 V 43 V Collision Cell Exit Potential : (CXP) 4V 2V 4V 4V Note: Either transition for cyromazine and melamine may be used for quantification. Typical chromatograms are shown in Appendix 4. 5.0 CALCULATION OF RESULTS 5.1 Single point calibration procedure Residues may be calculated in mg/kg for each sample using a mean standard response from each of the injections bracketing the sample as follows. a) Make repeated injections of a mixed standard containing cyromazine and melamine at an appropriate concentration into the LC-MS/MS operated under conditions as described in Section 4. When a consistent response is obtained, measure the peak areas obtained for the analytes. b) Make an injection of each sample solution and measure the areas of the peaks corresponding to the analytes. c) Re-inject the standard solution after a maximum of four injections of sample solutions. d) Calculate the residues in the sample, expressed as mg/kg, using a mean standard response from each of the injections bracketing the sample as follows. DRAFT GRM029.01A Page 18 of 57 Residue (mg/kg ) PK area (SA) PK area (STD) Standard Conc. Sample Conc. = = = = PK area (SA) Standard Conc. PK area (STD) Sample Conc. Peak response for sample Average peak response for bracketing standards Concentration of standard (g/mL) Sample concentration (g/mL) If residues need to be corrected for average percentage recovery e.g. for storage stability studies, then the equation below should be used. Corrected Residue = Residue 100 (mg/kg) Average percentage Recovery Although single point calibration may be used to quantify residues it is recommended that a calibration curve is generated with each analytical run to demonstrate the linearity of instrument response (Reference 4). 5.2 Multi point calibration procedure Residues may be calculated in mg/kg for each sample as follows. a) Prepare mixed standard solutions of cyromazine and melamine over a concentration range appropriate to the expected residues in the samples (for example, 50% LOQ to 10 x LOQ). An appropriate number of different concentrations within this range should be prepared (at least four). b) Make an injection of each sample solution and measure the areas of the peaks corresponding to the two analytes. Calibration standard solutions should be interspersed throughout the analysis, after a maximum of four injections of sample solutions. c) Generate calibration curve parameters using an appropriate regression package. d) The following equation can be rearranged and used to calculate residues as follows: y mx c Where y is the instrument response value, x is the standard concentration, m is the gradient of the line of best fit (“X-variable 1” in MS Excel) and c is the intercept value. An example of this equation generated using the experimental values of m and c should be included in the raw data, as should the “R-Squared” value for the regression. Re-arrangement for x gives DRAFT GRM029.01A Page 19 of 57 yc m Alternatively (depending on the regression analysis software available) a quadratic equation may be used to fit the data. In this case the following general equation should be re-arranged and used to calculate residues: x e) y a bx cx 2 Where y is the instrument response value, x is the standard concentration and a, b, c are constants. f) Calculate the residues of cyromazine and melamine in the sample, expressed as mg kg-1, as follows Analyte found ( g/mL ) Residue (mg/kg ) Sample conc. (g/mL ) Where analyte found (g/mL) is calculated from the standard calibration curve and sample conc. is the final sample concentration in g/mL. If residues need to be corrected for average percentage recovery e.g. for storage stability studies, then the equation below should be used. Corrected Residue = 6.0 Residue 100 (mg/kg) Average percentage Recovery CONTROL AND RECOVERY SAMPLES Control samples should be analysed with each set of samples to verify that the sample used to prepare recovery samples is free from contamination. A minimum of one control should be analysed with each batch of samples. At least two recovery samples (control samples accurately fortified with known amounts of cyromazine and melamine in acetonitrile) should also be analysed alongside each set of samples. Provided the recovery values are acceptable they may be used to correct any residues found. The fortification levels should be appropriate to the residue levels expected. Recovery efficiency is generally considered acceptable when the mean values are between 70% and 110% and with a relative standard deviation of <20%. 7.0 SPECIFICITY It is recommended that two method blanks should be analysed with each batch of samples and that prior to sample preparation, soil is removed as far as possible from crops which have had contact with soil e.g. potato crops. Melamine is a ubiquitous compound DRAFT GRM029.01A Page 20 of 57 and may be present at significant concentrations in soil even where cyromazine has not been applied. Additionally, untreated control samples should be screened for melamine residues prior to use in any recovery experiments. 7.1 Matrix interference LC-MS/MS is a highly specific detection technique. Interference arising from the matrices tested has not been observed. 7.2 Reagent and solvent interference Using high purity solvents and reagents no interference has been found. 7.3 Labware interference This method uses disposable labware. All reusable glassware should be detergent washed and then rinsed with HPLC-grade methanol, acetone or acetonitrile prior to use. 8.0 METHOD VALIDATION 8.1 Recovery data and repeatability Method validation has been carried out on the procedures described in Section 3. The method validation data are reported in ??? (Reference 5), and a summary is included in Appendix 3. 8.2 Limit of quantification (LOQ) The limit of quantification of the method is defined as the lowest analyte concentration in a sample at which the methodology has been validated and a mean recovery of 70-110% with a relative standard deviation of 20% has been obtained. Generally, for accurate quantification, the response for an analyte peak should be no lower than four times the mean amplitude of the background noise in an untreated sample at the corresponding retention time. The limit of quantification has been set at 0.05 mg/kg for cyromazine and melamine. 8.3 Limit of detection (LOD) The limit of detection of the method is defined as the lowest analyte concentration detectable above the mean amplitude of the background noise in an untreated sample at the corresponding retention time. An estimate of the LOD can be taken as three times background noise. Note that the LOD may vary between runs and from instrument to instrument. DRAFT GRM029.01A Page 21 of 57 8.4 Detector linearity For accurate quantification of residue concentrations, analyses should be carried out within the linear range of the detector. Detector linearity graphs are given in Appendix 5. The linearity of the LC-MS/MS detector response for cyromazine and melamine was tested in the range from 0.001 g/mL to 0.05 g/mL (equivalent to 10 pg to 500 pg injected on column when using a 10 L injection volume) and was found to be linear for both transitions in cyromazine and melamine. If a residue beyond the tested concentration range is expected, dilute the sample appropriately to bring it within the tested linear range prior to quantitation. Standards at ? different concentration levels (n = ?) were injected and the response plotted against amount injected on column, using Microsoft Excel 2003. Detector linearity graphs are given in Appendix 5. 8.5 Extract stability Cyromazine and melamine were proven to be stable in the final extracts for a period of up to ? days when stored at a temperature of 7°C. Storage stability data are presented in Appendix 3. 9.0 LIMITATIONS The method has been tested on representative high water content crop matrices. It can reasonably be assumed that the method can be applied to other crop matrices not tested in this study, provided successful recovery tests at the relevant levels validate the suitability of the method. 10.0 CONCLUSIONS Method GRM029.01A has been demonstrated to be a reliable and accurate procedure for the determination of cyromazine and melamine in high water content crop matrices, using commercially available laboratory equipment and reagents. The limit of quantification of the method is 0.05 mg/kg for cyromazine and melamine. This method complies with EU guidelines SANCO/3029/99 rev. 4, SANCO/825/00 rev. 7 and US EPA guideline OPPTS 860.1340. DRAFT GRM029.01A Page 22 of 57 11.0 REFERENCES 1. Cyromazine (CGA72662): REM174.02 Residue Analytical Method for the Determination of Cyromazine and its Metabolite Melamine (C1803) in Crops by High Performance Liquid Chromatography. 2. Luxon S G (1992): Hazards in the Chemical Laboratory 5th Edition. The Royal Society of Chemistry. Thomas Graham House, The Science Park, Cambridge, CB4 4WF, UK. ISBN 0-85186-229-2. 3. Yokley, R et al : Analytical Method for the Determination of Cyromazine and Melamine Residues in Soil Using LC-UV and GC-MSD. Journal of Agricultural and Food Chemistry (2000), 48(8), 3352-3358 4. Cardone M J, Palermo P J and Sybrand L B: Potential error in single point ratio calculations based on linear calibration curves with a significant intercept. Anal Chem., 52 pp 1187-1191, 1980 5. ? (2007): Cyromazine (CGA72662): Validation of a Residue Method for the Determination of Cyromazine (CGA72662) and Melamine in High Water Content Crop. Report No. xxxxx. DRAFT GRM029.01A Page 23 of 57 APPENDICES SECTION DRAFT GRM029.01A Page 24 of 57 APPENDIX 1 APPARATUS UK suppliers Equipment for the initial preparation of samples e.g. Robot Coupe R20 bowl chopper available from Lockhart Catering Equipment, Lockhart House, Brunel Road, Theale, Reading, Berkshire, RG7 4XE. General glassware, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire LE11 5RG Plastic centrifuge bottles, 250 mL size, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire LE11 5RG. High speed homogeniser for extraction of samples e.g. Janke and Kunkel Ultra Turrax T25, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire LE11 5RG. Mechanical shaker, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire LE11 5RG. Laboratory centrifuge e.g. MSE Mistral 1000 series, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire LE11 5RG, UK. Disposable borosilicate glass test tubes, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire, LE11 5RG. Plastic disposable pipettes, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire, LE11 5RG. Isolute Vacmaster-20™ sample processing station, available from Biotage, Tir-y-Berth Industrial Estate, New Road, Hengoed, Mid Glamorgan, CF8 8AU. Oasis MCX solid phase extraction columns, 3 mL 60 mg size, available from Waters Ltd., 730-740 Centennial Court, Centennial Park, Elstree, Hertfordshire, WD6 3SZ. Techne Dri-block 3D sample concentrator, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire, LE11 5RG. Ultrasonic bath e.g. Ultrawave U300/D, available from Fisher Scientific UK, Bishop Meadow Road, Loughborough, Leicestershire, LE11 5RG Crimp cap autosampler vials and caps, available from Agilent Technologies UK Limited, Chemical Analysis Group, Lakeside Heath, Cheadle Royal Business Park, Stockport, Cheshire, SK8 3GR. API 4000 LC-MS/MS system equipped with a TurboIonSpray source, available from Applied Biosystems, 120 Birchwood Boulevard, Warrington, Cheshire, WA3 7PB. Agilent 1100 HPLC system equipped with quaternary pump, vacuum degasser and column compartment with column switching valve, available from Agilent Technologies DRAFT GRM029.01A Page 25 of 57 UK Limited, Chemical Analysis Group, Lakeside Heath, Cheadle Royal Business Park, Stockport, Cheshire, SK8 3GR. CTC HTS PAL autosampler, available from Presearch Ltd, System House, 59-61 Knowlpiece, Hitchin, Herts, SG4 0TY. HPLC column, Phenomenex Luna CN 5 m 150 mm 2.0 mm i.d., available from Phenomenex, Queens Avenue, Hurdsfield Ind. Est., Macclesfield, Cheshire, SK10 2BN. Peak Scientific NM20ZA gas station, available from Peak Scientific Instruments Ltd., Fountain Crescent, Inchinnan Business Park, Inchinnan, Renfrew, PA9 4RE US suppliers Equipment for the initial preparation of samples e.g. Tecator homogeniser available from Perstorp Analytical inc., 12101 Tech Road, Silver Spring, Maryland 20904. General glassware, available from Fisher Scientific UK, Liberty Lane, Hampton, NH 03842. Plastic centrifuge bottles, 250 mL size, available from Fisher Scientific UK, Liberty Lane, Hampton, NH 03842. High speed homogeniser for extraction of samples e.g. Janke and Kunkel Ultra Turrax T25, available from Fisher Scientific, Liberty Lane, Hampton, NH 03842. Mechanical shaker, available from Fisher Scientific, Liberty Lane, Hampton NH 03842 Laboratory centrifuge e.g. Heraeus Instruments model 17RS, available from Heraeus Instruments, 111-A Corporate Blvd, South Plainfield, NJ 07080. Disposable borosilicate glass test tubes, available from Fisher Scientific, Liberty Lane, Hampton, NH 03842. Plastic disposable pipettes, available from Fisher Scientific, Liberty Lane, Hampton, NH 03842. Isolute Vacmaster-20™ sample processing station, available from Biotage Ltd., PO Box 280 329, Lakewood, Colorado, 8022-0329. Oasis™ MCX solid phase extraction columns, 3 mL 60 mg size, available from Waters Corporation, 34 Maple Street, Milford, Massachusetts, 01757-3696 Techne Dri-block 3D sample concentrator, available from Fisher Scientific, Liberty Lane, Hampton, NH 03842. Ultrasonic bath available from Fisher Scientific, Liberty Lane, Hampton, NH 03842. Crimp cap auto sampler vials and caps, available from Agilent Technologies, 395 Page Mill Road, Palo Alto, CA 94304. DRAFT GRM029.01A Page 26 of 57 API 4000 LC-MS/MS system equipped with a TurboIonSpray source, available from Applied Biosystems, 850 Lincoln Center, Foster City, CA 94404-1128. Agilent 1100 HPLC system equipped with quaternary pump, vacuum degasser and column compartment with column switching valve, available from Agilent Technologies, 395 Page Mill Road, Palo Alto, CA 94304. CTC HTS PAL autosampler, available from LEAP Technologies Inc., P.O. Box 969, Carrboro, NC 27510. HPLC column, Phenomenex Luna CN 5 m 150 mm 2.0 mm i.d., available from Phenomenex, 411 Madrid Avenue, Torrance, CA 90501-1430. Peak Scientific NM20ZA gas station, available from Peak Scientific Instruments, 1300 West Belmont Ave., Chicago, IL 60657. DRAFT GRM029.01A Page 27 of 57 APPENDIX 2 REAGENTS UK suppliers Solvents: Ultra pure water (HPLC grade), methanol and acetonitrile available from Rathburn Chemicals Ltd., Walkerburn, EH43 6AU. Analytical grade potassium dihydrogen phosphate (KH2PO4), o-phosphoric acid, hydrochloric acid, ammonium hydroxide solution (30% w/v) and acetic acid available from Sigma-Aldrich, The Old Brickyard, New Road, Gillingham, Dorset, SP8 4XT or www.sigmaaldrich.com Cyromazine and melamine analytical standards, available from Syngenta Crop Protection, GLP Testing Facility WMU, CH-4333, Munchwilen, Switzerland. US suppliers Solvents: Analytical grade methanol, acetonitrile and acetone available from B & J Brand Solvents, from Scientific Products Division of Baxter Healthcare Corporation. Ultra pure HPLC grade water from e.g. Fluka via Sigma-Aldrich www.sigmaaldrich.com Analytical grade Analytical grade potassium dihydrogen phosphate KH2PO4, ophosphoric acid, hydrochloric acid, ammonium hydroxide solution (30% w/v) and acetic acid available from www.sigmaaldrich.com Cyromazine and melamine analytical standards, available from Syngenta Crop Protection Inc., P.O. Box 18300, Greensboro, NC 27419-8300. Preparation of Reagents a) 1 M phosphoric acid: Carefully add 11.5 g H3PO4 (85% w/w) to ultra pure water (~ 50 mL) in a 100 mL volumetric flask. Adjust to the mark with ultra pure water. Stopper flask securely and mix thoroughly by shaking. b) 0.025 M potassium dihydrogen phosphate in 0.04M phosphoric acid: Dissolve 3.41 g KH2PO4 in ultra pure water (~200 mL) in a 1 L volumetric flask. Add 1 M H3PO4 (40 mL). Adjust to the mark with ultra pure water. Stopper flask securely and mix thoroughly by shaking. c) 0.1 M Hydrochloric acid Carefully add 8.4 g concentrated HCl to ultra pure water (~ 500 mL) in a 1 L volumetric flask. Adjust to the mark with ultra pure water. Stopper flask securely and mix thoroughly by shaking. d) 95:5 v/v methanol:30% ammonium hydroxide Carefully add 5 mL ammonium hydroxide solution (30% w/v) to methanol (~50 mL) DRAFT GRM029.01A Page 28 of 57 in a 100 mL volumetric flask. Adjust to the mark with methanol. Stopper flask securely and mix thoroughly by shaking. e) 0.2% v/v glacial acetic acid in ultra pure water: Carefully add 2 mL acetic acid to ultra pure water (~500 mL) in a 1 L volumetric flask. Adjust to the mark with ultra pure water. Stopper flask securely and mix thoroughly by shaking. DRAFT GRM029.01A Page 29 of 57 APPENDIX 3 Table 1. : Matrix * ** METHOD VALIDATION DATA Cyromazine recovery data obtained during method validation. Primary transition m/z 167 68 Fortification Level (mg/kg) Recovery (%)** n 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 Mean (%) RSD (%) Range (%) Limit of quantitation, defined by the lowest validated fortification level. Two control samples and a reagent blank were analysed with each analytical batch. No residues were measured at or above 30% of the LOQ in any of the samples. All recovery data were generated using reagent matched standards. DRAFT GRM029.01A Page 30 of 57 Table 2. : Matrix * ** Cyromazine recovery data obtained during method validation. Confirmatory transition m/z 167 85 Fortification Level (mg/kg) Recovery (%)** n 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 Mean (%) RSD (%) Range (%) Limit of quantitation, defined by the lowest validated fortification level. Two control samples and a reagent blank were analysed with each analytical batch. No residues were measured at or above 30% of the LOQ in any of the samples. DRAFT GRM029.01A Page 31 of 57 Table 3. : Matrix * ** Melamine recovery data obtained during method validation Primary transition m/z 127 68 Fortification Level (mg/kg) Recovery (%)** n 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 Mean (%) RSD (%) Range (%) Limit of quantitation, defined by the lowest validated fortification level. Two control samples and a reagent blank were analysed with each analytical batch. No residues were measured at or above 30% of the LOQ in any of the samples. DRAFT GRM029.01A Page 32 of 57 Table 4. : Matrix * ** Melamine recovery data obtained during method validation Primary transition m/z 127.0 85.0 Fortification Level (mg/kg) Recovery (%)** n 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 0.05* 5 0.5 5 Overall 10 Mean (%) RSD (%) Range (%) Limit of quantitation, defined by the lowest validated fortification level. Two control samples and a reagent blank were analysed with each analytical batch. No residues were measured at or above 30% of the LOQ in any of the samples. . DRAFT GRM029.01A Page 33 of 57 Table 5. : Determination of LC-MS/MS Matrix Effects The effect of crop matrices on the LC-MS/MS response was assessed by preparing standards in the presence of matrix and comparing the peak areas of cyromazine and melamine against reagent-matrix standards at an equivalent concentration. Matrix and reagent matched standards were prepared as described in Section 3.7. Matrix effects (enhancement or suppression) on the instrument response were considered not to be significant and reagent-matched calibration standards should be used. Matrix matched standards may be used for calibration however, to compensate for these effects at the discretion of the study director. Matrix Table 6. : Matrix Matrix Effect Cyromazine Transition m/z 167 68 Matrix Effect Cyromazine Transition m/z 167 85 Matrix Effect Melamine Transition m/z 127 68 Matrix Effect Melamine Transition m/z 127 85 Cyromazine recovery data obtained after ? days storage Calculated using the primary transition m/z 167 68 Fortification Level (mg/kg) Recovery (%)** Mean (%) RSD (%) Range (%) 0.05* 0.5 0.05* 0.5 0.05* 0.5 0.05 0.5 *Limit of quantification, defined by the lowest validated fortification level **Residues in control samples and reagent blanks were less than 30% of the LOQ. DRAFT GRM029.01A Page 34 of 57 Table 7. : Matrix Melamine recovery data obtained after ? days storage Calculated using the primary transition m/z 127 68 Fortification Level (mg/kg) Recovery (%)** Mean (%) RSD (%) Range (%) 0.05* 0.5 0.05* 0.5 0.05* 0.5 0.05 0.5 *Limit of quantification, defined by the lowest validated fortification level **Residues in control samples and reagent blanks were less than 30% of the LOQ. DRAFT GRM029.01A Page 35 of 57 APPENDIX 4 REPRESENTATIVE CHROMATOGRAMS Figure 3: 0.002 g/mL Cyromazine Standard m/z 167 68 Figure 4: 0.002 g/mL Cyromazine Standard m/z 167 85 DRAFT GRM029.01A Page 36 of 57 Figure 5: 0.002 g/mL Melamine Standard m/z 127 68 Figure 6: 0.002 g/mL Melamine Standard m/z 127 85 DRAFT GRM029.01A Page 37 of 57 Figure 7: Control Crop 1 Sample Concentration 0.04 g/mL, Cyromazine Residue <LOQ m/z 167 68 Figure 8: Control Crop 1 Sample Concentration 0.04 g/mL, Cyromazine Residue <LOQ m/z 167 85 . DRAFT GRM029.01A Page 38 of 57 Figure 9: Control Crop 1 Sample Concentration 0.04 g/mL, Melamine Residue <LOQ m/z 127 68 Figure 10: Control Crop 1 Sample Concentration 0.04 g/mL, Melamine Residue <LOQ m/z 127 85 DRAFT GRM029.01A Page 39 of 57 Figure 11: Control Crop 1 Fortified with 0.05 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 68 Figure 12: Control Crop 1 Fortified with 0.05 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 85 DRAFT GRM029.01A Page 40 of 57 Figure 13: Control Crop 1 Fortified with 0.05 mg/kg Melamine Sample Concentration 0.04 g/mL, Recovery = ? m/z 127 68 Figure 14: Control Crop 1 Fortified with 0.05 mg/kg Melamine Sample Concentration 0.04 g mL-1, Recovery = ? m/z 127 85 DRAFT GRM029.01A Page 41 of 57 Figure 15: Control Crop 1 Fortified with 0.5 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 68 Figure 16: Control Crop 1 Fortified with 0.5 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 85 DRAFT GRM029.01A Page 42 of 57 Figure 17: Control Crop 1 Fortified with 0.5 mg/kg Melamine Sample Concentration 0.04 g/mL, Recovery = ? m/z 127 68 Figure 18: Control Crop 1 Fortified with 0.5 mg/kg Melamine Sample Concentration 0.04 g/mL, Recovery = ? m/z 127 85 DRAFT GRM029.01A Page 43 of 57 Figure 19: Control Crop 2 Sample Concentration 0.04 g/mL, Cyromazine Residue <LOQ m/z 167 68 Figure 20: Control Crop 2 Sample Concentration 0.04 g/mL Cyromazine Residue <LOQ m/z 167 85 DRAFT GRM029.01A Page 44 of 57 Figure 21: Control Crop 1 Sample Concentration 0.04 g/mL, Melamine Residue <LOQ m/z 127 68 Figure 22: Control Crop 1 Sample Concentration 0.04 g/mL, Melamine Residue <LOQ m/z 127 85 DRAFT GRM029.01A Page 45 of 57 Figure 23: Control Crop 2 Fortified with 0.05 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 68 Figure 24: Control Crop 2 Fortified with 0.05 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 85 DRAFT GRM029.01A Page 46 of 57 Figure 25: Control Crop 2 Fortified with 0.05 mg/kg Melamine Sample Concentration 0.04 g/mL, Recovery = ? m/z 127 68 Figure 26: Control Crop 2 Fortified with 0.05 mg/kg Melamine Sample Concentration 0.04 g/mL, Recovery = ? m/z 127 85 DRAFT GRM029.01A Page 47 of 57 Figure 27: Control Crop 2 Fortified with 0.5 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 68 Figure 28: Control Crop 2 Fortified with 0.5 mg/kg Cyromazine Sample Concentration 0.04 g/mL, Recovery = ? m/z 167 85 DRAFT GRM029.01A Page 48 of 57 Figure 29: Control Crop 2 Fortified with 0.5 mg/kg Melamine Sample Concentration 0.04 g/mL, Recovery = ? m/z 127 68 Figure 30: Control Crop 2 Fortified with 0.5 mg/kg Melamine Sample Concentration 0.04 g/mL, Recovery = ? m/z 127 85 DRAFT GRM029.01A Page 49 of 57 APPENDIX 5 Figure 31: DETECTOR LINEARITY GRAPHS LC-MS/MS Calibration Graph for Cyromazine (No Intercept Set) m/z 167 68 DRAFT GRM029.01A Page 50 of 57 Figure 32: LC-MS/MS Calibration Graph for Cyromazine (No Intercept Set) m/z 167 85 DRAFT GRM029.01A Page 51 of 57 Figure 33: LC-MS/MS Calibration Graph for Melamine (No Intercept Set) m/z 127 68 DRAFT GRM029.01A Page 52 of 57 Figure 34: LC-MS/MS Calibration Graph for Melamine (No Intercept Set) m/z 127 85 DRAFT GRM029.01A Page 53 of 57 APPENDIX 6 API4000 MS/MS TUNING PROCEDURE Calibration of instrument The instrument must be mass-calibrated on a regular basis using polypropylene glycol (PPG) solutions according to the manufacturer’s instructions. Calibrate both massresolving quadrupoles (Q1 and Q3). The instrument must be mass calibrated on a regular basis using polypropylene glycol (PPG) solutions according to the manufacturer’s instructions. Calibrate both mass resolving quadrupoles (Q1 and Q3). Tuning instrument for Cyromazine and Melamine Infuse separate standard solutions of cyromazine and melamine (0.001 to 0.1 µg/mL) in mobile phase (see section 4.2) directly into the mass spectrometer interface at a rate at of approximately 10-20 µL/min. Roughly adjust interface parameters (sprayer position, spray, heater/auxiliary gas flows, as well as voltages of spray, orifice and focusing ring) for a sufficiently high parent ion signal at m/z = 167 for cyromazine and m/z = 127 for melamine Using the Analyst software quantitative optimisation routine, tune the instrument for cyromazine and melamine, ensuring that the correct ions are selected. If desired, manual tuning of the ion optics and collision energy can be carried out to ensure maximum sensitivity. Finally, connect the LC-pump via the autosampler directly to the MS/MS instrument. Perform repetitive flow injection of a mixed cyromazine and melamine standard using mobile phase at the flow rate to be used. Tune the interface parameters (sprayer position, spray and heater gas flows, spray, orifice and focusing ring voltages) and the collision gas flow for maximum sensitivity. In positive ionisation mode, the protonated molecular ions generated in the ion source (m/z = 167 for cyromazine and m/z = 127 for melamine) are selected and subjected to further fragmentation by collisional activation. The two most sensitive daughter ions (m/z = 85 and m/z = 68 for cyromazine and melamine are then selected and used for quantitative analysis. The most sensitive daughter ion (m/z = 85) corresponds to the loss of a cyanamine group from the 1, 3, 5-trazine-2, 4, 6-triamine molecular ion. A second transition (m/z = 68) corresponding protonated imidazole may be used to confirm any cyromazine and melamine residues detected. Final determination by LC-MS/MS with two transitions is considered to be highly specific; hence no further confirmatory conditions are included. DRAFT GRM029.01A Page 54 of 57 APPENDIX 7 MS/MS PRODUCT ION SPECTRA Figure 35: Initial product ion Scan for Cyromazine (positive ionisation mode). File:Cyromazine_20070208081733 This scan was not produced as part of the validation study DRAFT GRM029.01A Page 55 of 57 Figure 36: Initial product ion Scan for melamine (positive ionisation mode). File:melamine_20070208082010 This scan was not produced as part of the validation study DRAFT GRM029.01A Page 56 of 57 APPENDIX 8 METHOD FLOWCHART Extract crop sample (10 g) with by maceration with 0.025 M KH2PO4 in 0.025 M H3PO4 Add methanol to sample and shake for 30 mins using a mechanical shaker Take aliquot (0.2 g) through cation exchange SPE clean-up procedure Collect and evaporate column eluate to dryness Redissolve in ultra pure water Final determination by LC-MS/MS DRAFT GRM029.01A Page 57 of 57