Clinical Laboratory San Frnacisco State University 1600 Holloway Avenue San Francisco, CA CLS Laboratory Procedure #900.099 Section: Chemistry Date of Implementation: April 4, 2011 Revision:_______________________ Glucose Hexokinase Procedure for the quantitative determination of glucose in serum Policy The laboratory’s compliance to CLSI GP2A5 guidelines shall identify work processes in the laboratory’s operational path of workflow and write procedures for pre-analytical, analytical, and post-analytical laboratory activities. Purpose This procedure provides instructions for in vitro testing for the quantitative determination of glucose in human serum and plasma on the Pointe 180 QT analyzer. Responsibility The CLS internship Program Director shall review and approve procedural documents, appoint designees for the implementation and to maintain document control and record. This procedure follows the CLS SOP template # 900.099 to implement SOP and the designated author will perform document revisions. Clinical Significance Glucose is the major carbohydrate present in the peripheral blood. Oxidation of glucose is the major source of cellular energy in the body. Glucose derived from dietary sources is converted to glycogen for storage in the liver or to fatty acids for storage in adipose tissue. The concentration of glucose in blood is controlled within narrow limits by many hormones, the most important of which are produced by the pancreas. The most frequent cause of hyperglycemia is diabetes mellitus resulting from a deficiency in insulin secretion or action. A number of secondary factors also contribute to elevated blood glucose levels. These include pancreatitis, thyroid dysfunction, renal failure and liver disease. Hypoglycemia is less frequently observed. A variety of conditions may cause low blood glucose levels such as insulinoma, hypopituitarism or insulin induced hypoglycemia. Glucose measurement in urine is used as a diabetes screening procedure and to aid in the evaluation of glycosuria, to detect renal tubular defects, and in the management of diabetes mellitus. Glucose measurement in cerebrospinal fluid is used for evaluation of meningitis, neoplastic involvement of meninges and other neurological disorders. CPT code Page 1 of 5 99214 Clinical Laboratory San Frnacisco State University 1600 Holloway Avenue San Francisco, CA Principle CLS Laboratory Procedure #900.099 Section: Chemistry Date of Implementation: April 4, 2011 Revision:_______________________ UV Test Enzymatic reference method with hexokinase1 HK Glucose + ATP----------------------->G6P + ADP G6PDH G6P + NAD------------------------> 6-Phosphogluconate + NADH + H+ Glucose is phosphorylated with adenosine triphosphate (ATP) in the reaction catalyzed by hexokinase (HK). The product, glucose-6phosphate (G6P) is then oxidized with the concomitant reduction of nicotinamide adenine dinucleotide (NAD) to NADH in the reaction catalyzed by glucose-6phosphate-dehygrogenase (G6PDH). The formation of NADH causes an increase in absorbance at 340 nm. The increase is directly proportional to the amount of glucose in the sample. Specimen Requirements For specimen collection and preparation, only use suitable tubes or collection containers. Serum: Use fresh, unhemolyzed serum. Plasma: Unhemolyzed samples from tubes containing oxalate, citrate, EDTA, fluoride or heparin may be used. Serum and plasma must be separated from the red cells promptly to prevent glycolysis. Glucose will decrease approximately 7% per hour when left in contact with red cells.10 The addition of sodium fluoride to the specimen may prevent glycolysis.2 Glucose in serum of plasma is stable for 8 hours at room temperature and 24 hours refrigerated at 2-8°C. Specimen Rejection: Any unlabelled or improperly labeled specimen or sample not collected in accordance with requirements above will be rejected. Page 2 of 5 Clinical Laboratory San Frnacisco State University 1600 Holloway Avenue San Francisco, CA Materials/ Equipment CLS Laboratory Procedure #900.099 Section: Chemistry Date of Implementation: April 4, 2011 Revision:_______________________ Equipment Accurate pipetting devices Spectrophotomet er with ability to read at 340nm Heat block Procedure Calculations Reagents Glucose Hexokinase reagent Supplies Timer Deionized water Test tubes/rack Step Action 1. Reconstitute Glucose Hexokinase reagent with distilled water using the volume stated on the vial label. Swirl gently to dissolve lyophilized material. Do not shake. 2 Pipet 1.0 ml of reagent into the required number of labeled test tubes (eight tubes in total), and incubate for 3-5 minutes. 3 Turn on the spectrophotometer and when asked if you’d like to use a “Stored Curve” select NO. 4 Read a blank tube at 340 nm containing 1.0 ml of Glucose Hexokinase. Remove tube. 5 Add 10 ul each of control and patient to their respective tubes. Mix well and set the timer to incubate in the heat block for three minutes. 6 Read and record the absorbance values of all tubes. 7 If the result exceeds linearity or an “R” flag appears, dilute the flagged sample 1:1 with normal saline, repeat the measurement, and multiply the result by two. 8 To determine results, see “Calculations” section. Standard method to determine concentration of unknown values: Abs. = Absorbance Conc. = Concentration Sample Abs. X Conc. of Standard = Glucose (mg/dL) Standard Abs. Example: If Sample Abs. = 0.150 Standard Conc. = 100 mg/dL Standard Abs. = 0.200 Then: 0.150 X 100 (mg/dL) = 75 mg/dL 0.200 Page 3 of 5 Clinical Laboratory San Frnacisco State University 1600 Holloway Avenue San Francisco, CA Interpretation of Results Clinical Significance Serum Normal Reference intervals Abnormal results Critical statements CLS Laboratory Procedure #900.099 Section: Chemistry Date of Implementation: April 4, 2011 Revision:_______________________ 74-100 mg/dL (adults, fasting) 135-140 mg/dL (nonfasting) Greater than 180-200 mg/dL Less than 60-65 mg/dL < 40 mg/dL or > 400 mg/dL Abnormal results greater than 180-200 mg/dL indicates hyperglycemia. Results less than 60-65 mg/dL is indicative of hypoglycemia. Interferences Serum/plasma Hemolysis: No significant Interference up to a hemoglobin concentration 1000 mg/dL Lipemia: Extremely lipemic samples may give a falsely elevated glucose value for an L index of 1000. Icterus: No significant interference up to an I index of 60 (60 mg/dL) In very rare cases gammopathy, in particular type IgM (Waldenström's macroglobulinemia), may cause unreliable results. Drugs: No interference was found using common drug panels. Expected Values Reported normal range: 65-110 mg/dL It is recommended that each laboratory establish its own reference range. Clinical Applications Urine is not specified as an appropriate sample type per reagent package insert. However, there are many other reagent applications that do accept and give reference ranges for urine. Reporting Analytical test result calculated manually. See “Calculation” section for a sample calculation. Critical Values Serum - less than 40 mg/dL or greater than 400 mg/dL Page 4 of 5 Clinical Laboratory San Frnacisco State University 1600 Holloway Avenue San Francisco, CA CLS Laboratory Procedure #900.099 Section: Chemistry Date of Implementation: April 4, 2011 Revision:_______________________ Protocol to follow if critical value result is obtained: Immediately contact attending physician. See Critical Values Policy. Notes For calibrator use an aqueous glucose standard of 100 mg/dl and quality controls please reference the control procedure. References References and MSDS associated references 1. Tietz NW, ed. Clinical Guide to Laboratory Tests. 3rd ed. Philadelphia: WB Saunders 1995:268-273. 2. Sacks DB. Carbohydrates. In: Tietz NW, ed. Fundamentals of Clinical Chemistry. 4th ed. Philadelphia: WB Saunders 1996:351-374. Author:_Alaina Whitaker, CLS Trainee Date:_April 4, 2011__________________ Approved by:_______________________ Date:______________________________ Page 5 of 5