CAVITY DIAGNOSTIC SYSTEM FOR THE PERFORMANCE TEST OF THE 1.3GHZ SUPERCONDUCTING 9-CELL CAVITY Y. Yamamoto#, H. Hayano, E. Kako, S. Noguchi, M. Sato, T. Shishido, K. Umemori, K. Watanabe, KEK, Tsukuba, 305-0801, Japan, H. Sakai, ISSP, Univ. of Tokyo, Kashiwa, 277-8581, Japan, T.X. Zhao, IHEP, Beijing, 100049, China INTRODUCTION The new test facility for the superconducting cavity (STF) is completed at KEK. It is included the electropolishing (EP) device for the surface treatment of the cavity, the chemical polishing (CP) system around the flanges of the beam pipes, the high pressure rinsing (HPR) with ultra-pure water, the clean-room for the assembly working of the cavity, the pre-tuning system for the frequency tuning of the cavity, and the test facility for the cavity performance test (vertical test). During the vertical test, the temperature mapping (T-mapping) system using the carbon resistor is normally used to identify the heating spot, which is considered as the defect or the contamination. On the other hand, the x-ray monitoring using the PIN diode is effective for the observation of the field emission [1]. New vertical test facility for the superconducting 9-cell cavity (Figure 1) was completed in STF at the beginning of 2008 [2]. One superconducting 9-cell cavity (AES#001) was sent from FNAL for the system check of the facility. This cavity was totally measured seven times in U.S. The achievable accelerating field was 15.2MV/m. The heating was detected at #3 cell by FNAL’s Tmapping system in the vertical test at FNAL. The three suspicious points were found at #3 cell from the result of the observation with Kyoto Camera [3], and one suspicious point was found at #7 cell. As the pilot test of the new vertical test facility, the vertical tests are done totally a few times using this cavity within 2008. introduced to detect the heating during the high power rf measurement. After finding the heating spot, it is also important to inspect the cavity inside using the optical inspection system. The T-mapping system and the optical inspection system are complementary each other and vital to improve the performance of the superconducting cavity, which is used for ILC (International Linear Collier) and ERL (Energy Recovery Linac). NEW VERTICAL TEST FACILITY & NEW T-MAPPING SYSTEM (FISH-BONE) In the new vertical test facility (Figure 2), there are two holes for the two cryostats in the pit, the iron shield for the radiation which is automatically moved, the pumping system composed of 2 rotary pumps and 1 mechanical booster pump with the high exhaust velocity, the clean booth for the assembly of the T-mapping system, the control room and the four cavity stands. The 400W highpower amplifier is used for the RF measurement. Figure 2: The new vertical test facility in STF. Figure 1: The Superconducting 9-cell Cavity. After first CP and HPR at STF, the first vertical test was done on Oct/2008. The new T-mapping system was The new T-mapping and x-ray monitor system is shown in Figure 3. It is called “Fish-Bone” structure. Although the number of the carbon resistors is about 150 now, it will be attained several hundred channels in the future. The resistors will be attached every 30 degrees at each cell and the number per cell is 36. The PIN diodes are attached to the top and bottom flanges for the observation of the field emission, and near the stiffening ring. SUMMARY & FUTURE PLAN Figure 3: The new T-mapping and x-ray monitor system. VERTICAL TEST & RESULT The first vertical test using AES#001 cavity was done in 8-9/Oct./2008. The achievable field was 15.9MV/m and consistent with FNAL’s result. The different points between FNAL and KEK are the coupler type (variable or fixed) and the use of the magnetic shield (with or without). The heating was observed at #3 cell using Fish-Bone. In π, 8π/9, 6π/9 and 5π/9 mode, #3 cell was heating and limited to the achievable field. The heating at the suspicious point was highest and other points were lower than that (Figure 4), when the cavity was quenched. On the other hand, although #6 cell was temporarily heating by the multi-pacting in 7π/9 mode, it eventually disappeared in the end of the measurement [4]. The new vertical test facility was completed in STF and the new T-mapping system was also introduced. The first vertical test in STF was done using it. It was successful to detect the heating at #3 cell, which result is same as that in FNAL. It found that the achievable field was limited by #3 cell in four pass-band modes. On the other hand, it was possible to observe the temporary phenomenon by multipacting. The number of the carbon resistors will be increased more and more in the future, and it will be about 350 by the end of 2008. Similarly, the number of the PIN diodes will be increased. To avoid the use of many cables, it is necessary to introduce the switching circuit and it is under consideration. The T-mapping system and the optical inspection system is complementary each other. Inspecting the cavity before and after each vertical test, and detecting the heating by T-mapping, it will find out whether any suspicious spot is problematic or not. It is crucial for the fabrication of the cavity and helpful for the understanding of the cavity performance [4]. In STF, many cavities will be tested in the future and the systematic understanding for the cavity performance will be advanced. ACKNOWLEDGMENTS The authors wish to express our special thanks to S. Mishra, B. Kephart, M. Champion and C. Ginsburg in FNAL for the rental of AES#001 cavity, H. Padamsee in Cornell, W. Moeller in DESY and Y. Morita in KEK for providing the carbon resistors, T. Okada and M. Iitake for the assembly working of the T-mapping. REFERENCES [1] Y. Yamamoto et al., SRF 2007, Beijing, China, (2007) WEP13. [2] Y. Yamamoto et al., EPAC’08, MOPP044. [3] Y. Iwashita et al., EPAC’08, WEOBM03. [4] https://indico.desy.de/conferenceOtherViews.py? view=standard&confId=946 Figure 4: The heating at #3 cell during the rf measurement. It found that the other cells can achieve around 30MV/m, although #3 cell was limited around 15MV/m, from the result of the measurement using seven pass-band modes. For monitoring the x-ray emission, the x-ray level was below several hundred μSv/h in π mode measurement. From the result of Q-E curves, the trend of the field emission was observed in 8π/9, 7π/9 and 4π/9 mode.