Experiment 4 Calibration of Temperature Sensors I. Objective: The purposes of this experiment are to: (1) acquaint you with common thermometry instruments, (2) impart a fundamental understanding of the physical principles that underpin these devices, and (3) point out the respective advantages and disadvantages that these instruments offer. II. Apparatus 3 ½” Floppy Diskette (Supplied by the student.) Personal Computer Running LabView Keithley 2000 6 ½ Digit Multimeter Equipped with 2000-SCAN MUX Card Julabo F12-MD PID Temperature-Controlled Bath Type-J 0.250” Dia. Sheathed Thermocouple Omega GJ QSS-14G-12 Type-J 0.0625” Dia. Sheathed Thermocouple Omega GJ QSS-116G-12 Type-T 0.250” Dia. Sheathed Thermocouple Omega GT QSS-14G-12 Type-T 0.0625” Dia. Sheathed Thermocouple Omega GT QSS-116G-12 Industrial Thermistor Omega ON-970-44006 (10 k @ 25 C) Industrial Resistance Temperaure Detector (RTD) Rosemount 563792 1, S/N 0693603 Precision Mercury Thermometer (NIST Certified) Omega GT-3554Y (-1-100 C +/- 0.1 C) Standard Alcohol Laboratory Thermometer Fisher 14-997 III. Introduction: Thermometry is an essential element in many industrial processes. For example, when extruding polymers such as polyethylene, the extruder is partitioned into several thermal zones. The temperature of each zone must be maintained within strict limits. If for instance, the temperature in any zone is too great, then the polymer will burn and carbon will be deposited on the interior walls of the extruder. This carbon is nearly impossible to remove and it has an unfortunate tendency to break-off when the extruder is in operation. These chunks of carbon contaminate the end product. The product is subsequently sent to a landfill because it can be neither sold nor recycled. Clearly, control relies upon accurate measurement of the operating temperatures. Thermometers: These devices are the oldest thermometry instruments and their use is well known. Two thermometers are used in this experiment. One should be familiar with the standard alcohol thermometer. The precision thermometer uses mercury and it has been calibrated at NIST. Thermocouples: These devices consist of two dissimilar metals that are joined at the ends. When these junctions are held at different temperatures, a thermoelectric potential difference exists. This is called the Seebeck effect and one may measure the thermoelectric potential ( or electromotive force, emf) to infer temperature junction temperatures. The potential, temperature range, and sensitivities of thermocouples depend upon the combinations of wire used. The Type-T thermocouples employ copper and constantan and the Type-J thermocouples employ iron and constantan. Thermoelectric potentials are typically small (mV). A calibration relation e.g., formula or table, relates temperature to thermoelectric potential. The NIST (formerly NBS) thermocouple tables comprise standard calibrations for the most popular thermocouple types. The accuracy of these tables is approximately +/- 2 degrees C. Please refer to the introductory section of Experiment 3 for more complete information on thermocouples. Resistance Temperature Detectors (RTD): These devices use the property that the electrical resistance of materials varies with temperature. RTDs typically consist of a platinum wire enclosed within a sheath. The Callendar-Van Duesen calibration equation, T (C ) (0.01 T (C ) 1) (0.01 T (C ) , R() R0 () R0 () (1) (0.01 T (C ) 1) (0.01 T (C ) 3 ) relates wire resistance to temperature. The symbols R0 , , , and represent calibration constants. The factory-determined values of these constants are stamped on the metal tag attached to the RTD used in this laboratory. Record these constants and the serial number of the RTD in your lab notebook. Note that the constant is used only when the measurement temperature is below 0 C, i.e., 0 when T 0 C . The manufacturer reports the accuracy of Eqn. (1) to be 0.06 C over the range from 0 to 100 C when their calibration coefficients are used. The calibration constants of the RTD are NIST (National Institute for Standards and Technology) traceable. This implies that the RTD was calibrated against internationally accepted temperature standards. Thermistors: These devices also use the property that the electrical resistance of materials varies with temperature. In contrast to RTDs, thermistors are made of semiconductor materials. Unlike metals, the temperature-resistance dependence of semiconductor materials is strongly non-linear. Consequently among thermometry instruments, thermistors generally have the greatest sensitivity, but their useful range is comparatively small. The calibration equation are typically of the form 1 A Bln R C ln R 3 . T (K ) (2) Julabo F-12MD Constant Temperature Bath: The following diagrams and instructions are exerpted from the Julabo manual. Operating controls and functional elements 9 10 11 12 7 5 6 ! 1a 8 2 3 4 13a 14 16a 15a 19 24a 24b 1b 13b 23 22 25 21 25 16b 15b 17 18 Rear view 1a/1b Mains power switch, illuminated, for circulator / cooling machine 2 Start / stop key 3 Working temperature T1 4 Working temperature T2 5 High temperature warning limit 6 Low temperature warning limit 7 Safety temperature 8 Adjustable excess temperature protection (safety temperature) 9 MULTI-DISPLAY (LED) temperature indication 10 Cursors left/right 11 Edit keys (increase/decrease setting) 12 Enter key (store) Error! Objects cannot be created from editing field codes. Indicator light - Alarm Indicator light - Cooling Error! Objects cannot be created from editing field codes. Indicator light - Heating 13a Connector: Control cable for cooling machine 13b Connector: Cooling machine control 5 14 1 9 6 RS232C interface 15a Mains fuses for circulator 15b Mains fuses for cooling machine 16a Mains power cable for circulator 16b Mains power cable with plug 17 Built-in mains outlet for connection of circulator 18 Selector dial for cooling machine Position "1" for operation with MD circulator. 19 Control cable for cooling machine 21 Removable venting grid 22 Drain tap 23 Drain port 24a Pump connector for feed 24b Pump connector for return Only for water cooled models: 25 Cooling water OUTLET / INLET Filling / draining Filling Take care that no liquid enters the interior of the circulator. Recommended maximum filling level with water as bath liquid: 25 mm below the tank rim Recommended maximum filling level with bath oils: 40 mm below the tank rim ATTENTION: the volume of bath oils will increase due to thermal expansion when the bath temperature rises. Exercise CAUTION when emptying hot bath liquids! Start: Press the start/stop key. - The MULTI-DISPLAY (LED) indicates the actual bath temperature. (example: 21.0 °C) - An illuminated indicator light in the "T1" or "T2" key indicates the activated working temperature. Stop: Press the start/stop key. The MULTI-DISPLAY (LED) indicates the message "OFF". Setting the temperatures Setting the working temperature "T1": Press the setpoint key . The indicator light blinks and the value previously set appears on the MULTI-DISPLAY (LED). Use the cursor keys to move left or right on the display until the numeral you wish to change is blinking. Use the increase/decrease arrows to change the selected numeral (-, 0, 1, 2, 3, ... 9). Press enter to store the selected value (example: -15.0 °C). The working temperature is maintained constant after a short heat-up time (e. g. -15.0 °C). Setting the working temperature "T2": Press the setpoint key Same procedure as with "T1" (example: 25.0 °C). . Selecting the working temperature: Press the setpoint key Press the setpoint key and then enter and then enter . . Technical specifications MD Display resolution °C 0.1 ATC - Absolute Temp. Calibration °C ±3 Heater capacity at 115V at 230 V Pressure pump: pressure/flow rate Watts Watts head max./Lpm Electrical connectors: Alarm output Computer interface 1000 2000 11.5 ft/15 24-0 V DC / max. 25 mA RS232C Mains power connection V/Hz 115/60 V/Hz 208-230/60 All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C Technical changes without prior notification reserved. Working temperature range °C Temperature stability °C ±0.01 Cooling capacity (bath liquid: ethanol) °C Watts450 320 140 30 Refrigerant F34 F12 -34 ... 200 -25 ... 200 ±0.01 +20 0 -20 -30 200 120 25 +20 0 -20 R134a R134a 24x30/15 15x13/15 Bath opening/bath depth: WxD/H Bath volume: from...to liters Dimensions: WxDxH Shipping weigth Mains power connection V/Hz Cm. 14 ... 20 In. Cm. 3 ... 4,5 15x23x25 38x58x61 6x9x10 20x36x55 lbs/kg 106/45 51/23 115/60 115/60 Keithley 2000 DMM: The use of the Keithley 2000 is described in the following section. One should note that the connections and measurement procedures are described to facilitate the uncertainty analysis, but physically, these connections are automatically made by the SCAN-2000 multiplexer card. Voltage Measurements: To measure voltage with the Keithley 2000, high and low leads are connected to the inputs depicted in Figure 1. Sense Input (+) Red Lead Hi (-) Black Lead Lo Figure 1: DC voltage measurement instrument connections. Once the leads have been properly connected, the “DCV” function is selected---one should watch for the DCV LEDS to appear when the Keithley is in scan mode during data acquisition. The Keithley 2000 is a 6 ½ Digit Multimeter. The DC voltage accuracy is listed in Table 1. Table 1. Keithley 2000 DC Voltage Accuracy Range 100.0000 mV 1.000000 V 10.00000 V 100.0000 V 1000.000 V Resolution 0.1 V 1.0 V 10 V 100 V 1 mV Accuracy (ppm RDG + ppm RANGE) 50 + 35 30 + 7 30 + 5 45 + 6 45 + 6 Keithley reports their multimeter accuracies with (ppm of RDG + ppm of RANGE), where 1 ppm=10 -6. For example, suppose 5.00000 V is measured on the 10.00000 V range. Then the accuracy estimate is (30106 5.00000 V+510-610.00000 V)= 200 V. Resistance Measurements: The Keithley 2000 has two modes for measuring resistance. The two-wire mode is illustrated in Figure 2. V1 i V2 Figure 2. Two-wire resistance measurement circuit. One should note that in this mode, the multimeter passes a current through the restive load and it simultaneously measures the voltage across the terminals. This method is employed to measure the thermistor resistance. The two-wire mode is simple, but it has the undesirable consequence that the resistance of the lead wires is included in the measurement. This could introduce substantial error while measuring small resistances. The four-wire mode compensates for lead wire resistance. The circuit is depicted in Figure 3. Input Hi i Hi Sense Lo Input Lo Figure 3. Four-wire resistance measurement circuit. The four-wire technique uses separate wires to apply a current and to measure the voltage drop. Consequently only the voltage drop across the resistive load is measured, and the effects of the lead resistance are eliminated. The fourwire method is used to measure the RTD resistance. The connections are illustrated in Figure 4. Sense Input Red Red Hi White White Lo Figure 4. Four-wire resistance measurement instrument connections. The accuracy of the resistance measurements are estimated from Table 2. Table 2. Keithley 2000 resistance measurement accuracy. Range 100.0000 1.000000 k 10.00000 k 100.0000 k 1.000000 M 10.00000 M 100.0000 M Resolution 100 1 m 10 m 100 m 1 10 100 Accuracy (ppm RDG + ppm RANGE) 100 + 40 100 + 10 100 + 10 100 + 10 100 + 10 400 + 10 1500 + 30 The measurement accurcies listed in Table 2 apply to both two- and four-wire modes. Procedure: 1. Inspect all electrical connections. Verify that the Keithley 2000 will use the rear inputs i.e., the “Inputs” button is depressed. 2. Verify that the liquid level in the circulator bath is within acceptable limits. 3. Prepare the ice bath. Fill the thermos bottle with ice and DI water so that there are no air gaps. Adding water serves two purposes. a) Fills all the air pockets in the ice filled thermos bottle and b) Ice water mixture is at 0 C (which is what we want), whereas just ice could be at a lower temperature than 0 C. 4. Record instrument data e.g., thermometer resolution, RTD calibration constants, and serial numbers. 5. Energize equipment i.e., turn on the F-12 Circulator bath, MD controller, Keithley 2000, computer, and monitor. 6. Activate the Julabo MD controller and adjust the setpoint temperature to 10 C. One should note that it takes approximately 35 minutes for the water bath to cool from room temperature to 10 C. The approximate bath temperature is indicated by the Julabo MD LED display. 7. Start up the PC and the Lab4B data acquisition software should start up on its own. If it doesn’t or if the PC is already booted up, click on the Lab4B icon on the desktop. when prompted choose “Start New Calibration” if you are starting to calibrate for the first time. If the program had crashed midway and you just started it back on, choose “Continue Previous Calibration”. 8. When the bath reaches setpoint temperature, acquire data: a. b. c. d. 9. Select the trial to acquire (1-10). One should note that 10 measurements are suggested. Read the Alcohol and Mercury thermometers and enter the measurements in the appropriate boxes for these instruments. These boxes are located below the Data Table. One can either enter the value in these boxes or set the adjacent slider at the appropriate position. The former is a better option Press the “Acquire” button to obtain the output from the other instruments. Repeat steps a-c for subsequent trials. Write the data to a file: a. b. Press the “Write Data to File” button. Use the pull-down menu to select the directory to which the file will be written. Please choose the “C:\Lab4B_student” directory. This is the default directory that the software prompts you to save in. c. d. e. Name the output file and click “Ok.” One should note that this file name must be unique and that the default extension is “.txt” to ensure that Windows will recognize that it is an ASCII file. Use Windows Explorer to view the data file in notepad. Once you saved the data, you can click on the clear button to clear the data from the screen and proceed to calibrate at other temperatures. 10. Adjust the setpoint temperature and repeat steps 8 and 9 for bath temperatures of approximately 15, 20, 25, 30, and 40 C. 11. Copy the data files to the 3 ½” floppy diskette. 12. Press the “End DAQ” button to stop the VI and close all application windows. 13. Shut down the computer (Using the “Start” menu). 14. Turn off all equipment. 15. See the TA on duty to check-out. 16. Perform data reduction and uncertainty analysis. 17. Write laboratory report. Data Analysis: 1. Consolidate the multiple samples to single measurements and estimate total uncertainties: a. Compute mean measurements of each device at every setpoint temperature. b. Estimate the precision uncertainties of the mean measurements (assume 95% confidence level). c. Estimate the bias uncertainties of the mean measurements. d. Estimate the total uncertainty of each mean measurement. Document the procedure used to complete steps a-d and tabulate the results of each step. After completing step d, tabulate setpoint temperatures, mean meaurements and their appropriate uncertainties. A table for each instrument should be prepared (eight tables). Of note, one must adhere to significant figure reporting conventions to receive credit for this step. Examine the tabulated results of steps a-d. For each instrument, determine whether bias or precision limits the uncertainty of the measurement. Which measurements are the most precise? Which measurements are the most accurate? How could the uncertainties of these measurements be reduced? What factors contribute to the measurement uncertainty? Which can be controlled and which cannot? Could bath temperature fluctuates contribute to uncertainty? Does the data support the assumption that the bath temperature is constant? 2. Use the RTD data to determine the bath temperatures. Please note that the Callendar-Van Duesen equation and the manufacturer’s calibration coefficients are employed. In addition, estimate the uncertainty in the RTD bath temperature measurements. One should note that there are essentially two factors that contribute to the overall uncertainty in this measurement. What are these factors? Which factor dominates? What would one need to do to improve i.e., increase accuracy, of this measurement? 3. One should recall that a calibration standard is by assumption, error-free. In actuality, they have uncertainty and calibrating an instrument is essentially a procedure for substituting the uncertainty of the standard for the uncertainty of the test instrument. Obviously, the uncertainty of standard must be smaller than the test instrument for this to be a worthwhile endeavor. Thus the instrument with the smallest uncertainty is most suitable to be the calibration standard. Determine which instrument yields the smallest temperature uncertainty. Hints: The calibration coefficients of the thermistor are unknown, hence it cannot be a standard. It is better to use a sensor with NIST traceable calibration as the calibration standard against which the other sensors are calibrated. If the thermocouple reference tables and the thermocouple voltages were used to estimate the bath temperature, what would the temperature uncertainties be? Further examination of the tables prepared in step 1d and the results of step 2 should whittle the choices for calibration standard down to two instruments. What are these instruments? Which device is your choice for the calibration standard? 4. Determine calibration equations for each instrument: a. The relationship between the thermistor resistance and temperature is non-linear. The expression R A exp( BT ) , seems to fit the thermistor data well. R is the thermistor resistance and T is the bath temperature (determined by the calibration standard). Determine the coefficients A and B using linear regression. Plot both the thermistor data and the fitted line. Also, plot error bars and estimate the uncertainty in the calibration equation. b. Determine calibration equations for the thermometers---assuming of course, that neither is the calibration standard. Use linear regression analysis and estimate the uncertainty in the calibration equation. Do these instruments have offsets? If so, how do they compare to the instrument resolution? c. Determine calibration equations for the thermocouples. 5. Compute the sensitivity of each instrument. References: The Temperature Handbook Omega Engineering Inc., Stamford, CT 06907-0047 Beckwith, Marangoni, and Lienhard, Mechanical Measurements, Addison-Wesley.