Implant Angle Repeatability on Optima MD Robert Rathmell, Jonathan David, and Mark Harris Axcelis Technologies, Inc., 108 Cherry Hill Dr, Beverly, MA 01915 USA Abstract. A very sensitive channeling implant has been used to quantify the repeatability of implant angles for the Optima MD over a period exceeding one month. Sensitivity of sheet resistance to variation in vertical and horizontal beam angles was measured for 500 keV P++ at a dose of 5E13 /cm2 near orientations tilt/twist of 0/0 and 35.26/0 for (100) wafers. Sensitivity to tilt in the vertical plane at both conditions was at least a 10% change in Rs for a 0.5 change in tilt at a tilt of 0.75 away from the channel (i.e. at 0.75/0 and 34.5/0), where the variation in Rs was linear with tilt. Horizontal angle sensitivity resulted in a 7.5% change in Rs for a 0.5 change in angle, but was less sensitive to smaller angle variation, since Rs varied roughly as a second order function near the channel. Sensitivity to tilt is reduced by about a factor of 3 for implants at 0/0 compared to 0.75/0 with a similar response at 35.26/0 compared to 34.5/0. Implants were carried out using wafers from a single boule over a period exceeding one month at the orientation of 0.75/0. Angle control was enabled by in-situ measurement and correction of the beam angles in both the vertical and horizontal planes prior to implant. The repeatability of sheet resistance was used to infer the variation in angle. Attributing all Rs variation to be the result of beam angle variation showed that the standard deviation of beam angles for over 30 wafers is 0.03. Keywords: Implant angle control PACS: 85.40.Ry, 81.05.Ea, 41.85.Qg HARDWARE DESCRIPTION Optima MD angle control hardware was described previously [1 - 3]. A brief summary of the techniques will be given here. The Optima MD is a medium dose, serial implant tool that uses a hybrid scanning system. The beam is scanned electrostatically in the horizontal plane and angles are “corrected” by a curved acceleration gap to form rays that are parallel to the central, unscanned ions. The wafer is scanned mechanically “vertically” in the tilted plane of the implant to pass through the ribbon-like beam to uniformly dose the entire wafer. Beam angles are measured in the horizontal and vertical planes prior to implant and automatically corrected to ensure the implant is performed at the angles specified in the recipe. Horizontal Angle Measurement Horizontal beam angles are measured by a moving profiler behind a mask with 7 slots. The angle corrections ensure that the 7 beamlets spanning the width of the wafer are parallel to each other and that the largest angle satisfies the recipe limit of Maximum Horizontal Beam Angle. Corrections are automatically made by steering the beam prior to the scan plates to correct the average angle or adjusting the parallelizing lens voltage, if necessary, to correct parallelism. . Vertical Angle Measurement The beam is deflected vertically by nominally 15 by an electrostatic energy filter just before entering the end station. Measurement in the vertical plane of the angle of the scanned beam verifies that the actual Vertical Beam Angle, VBA, is known prior to implant. The measurement is made by the VBA monitor, which is a faraday with a thick mask with narrow slots that admits more current when it is aligned with the direction of the ions. The VBA monitor is mounted on the scan arm of the wafer so that it has a fixed offset from the wafer surface as shown in Figure 1. At the start of a process job, the scan arm tilts so that the VBA monitor collects current as it rotates from 5 above to 5 below the nominal angle and then again on the way back to generate two plots of current vs. angle. The two plots are compared to detect noise or beam drop-outs. The flux weighted average angle of the two data sets is the Vertical Beam Angle. The wafer angle is adjusted about an axis parallel to the scanned beam, if necessary. Both measurement schemes are calibrated to crystal planes during machine setup to ensure that a 0 implant is normal to the surface of the wafer. +5 320 -5 Rs, ohms/sq Wafer 300 290 280 30 Tilt Angle, deg FIGURE 2. Average sheet resistance for a series of wafers implanted with 500 keV P++ at a dose of 5E13/cm2 near the 0/0, <100> axis at a range of vertical tilt angles. 360 The primary goal of this experiment was to evaluate the ability of the angle control scheme of Optima MD to accurately and repeatably measure and correct implant angles from batch to batch over a significant period of time. Prior experience has shown that high energy phosphorus implants at a dose of 5E13 atoms/cm2 are very sensitive to channeling effects [3]. Two different crystal orientations were considered for the test. For (100) wafers, an implant at a tilt/twist of 0/0 is aligned with the <100> axis, while an implant at 35.26/0 is aligned with the <112> axial channel. Vertical Angle Sensitivity Implants were made at a range of tilt angles near these two axes to determine the sensitivity. The wafers were measured with a Therma-Wave ThermaProbe and a sheet resistance, Rs, probe. Results for Rs are shown in Figures 2 and 3. Therma Wave results (not shown) showed comparable sensitivity. One can see that the sensitivity to small angle changes is greater at a tilt of about 0.75 away from the channeling axis and has a roughly linear slope. Sensitivity was comparable at both conditions and the 0.75/0 orientation was selected for the repeatability test. Rs, ohms/sq EXPERIMENT Test Conditions 1.5% per 1/4 deg 270 -0.75 -0.50 -0.25 0.00 0.25 0.50 0.75 1.00 VBA FIGURE 1. Scan arm is shown in the position to measure vertical beam angle. The VBA monitor is mounted to the scan arm with a fixed angle to the wafer. 5% per 1/4 deg 310 340 320 6% per 1/4 deg 2% per 1/4 deg 300 280 260 33.50 34.00 34.50 35.00 35.50 36.00 36.50 Tilt Angle, deg FIGURE 3. Average sheet resistance for a series of wafers implanted with 500 keV P++ at a dose of 5E13/cm2 near the 35.26/0, <112> axis at a range of vertical tilt angles. Horizontal Angle Sensitivity In the horizontal plane on the Optima MD, one does not have to implant a series of wafers to produce the “V-Curve” of Rs vs. angle. One can implant a single wafer with the Parallelizing Lens off so that the scanned beam spans the wafer with angles that vary linearly by +/-1.2 degrees at this energy, as shown in Figure 4. The sensitivity of sheet resistance to angle is about the same for the <100> and <112> axes. The Rs increases about 7% at 0.5 from the channel, which, at this angle, is about 70% as sensitive as the vertical axis. Horizontal sensitivity was about the same at a tilt/twist of 0.75/0 (not shown), because the channeling feature is largely planar. Horizontal sensitivity is less for small angle errors near 0. 1.5 1 +/- 0.5 16 12 0.5 0 8 -0.5 4 0 -150 -100 -50 0 50 100 -1 -1.5 150 Angle Repeatability Implants Beam Angle, deg. Change in Rs, % 24 20 Position, mm Rs 0/0 Rs 35.26/0 Angle FIGURE 4. Variation of sheet resistance horizontally across the wafer implanted with 500 keV P+ at a dose of 5E13/cm2 with the P-lens off so that angles vary linearly from one side to the other. Test Angle Correction Accuracy Rs Ohm/Sq 320 +/- 0.55 uncorrected 310 Results of the Angle Repeatability Implants Sheet resistance repeatability of the implants is shown in Figure 6. Dashed lines above and below the data show the equivalent change in Rs corresponding to +/- 0.1 in angle variations. If one assumes that all of the variation is due to angle variation, the results show that the maximum angle variation was less than +/-0.1. It is possible that some of the variation is caused by finite dose repeatability and metrology. 310 Rs, ohms/sq In order to confirm that angle corrections worked, the beam was tuned to cause a significant angle error, and wafers were implanted with and without automatic corrections enabled. Since the beam is deflected in the vertical plane by nominally 15 at the Angular Energy Filter of the Optima MD, the AEF can be adjusted to change the beam angle in the vertical plane. Wafers were implanted with the AEF adjusted to cause the beam to enter the end station at angles approximately +/-0.55 from the nominal 15 bend. In each case, one wafer was implanted with VBA corrections off, and one with corrections enabled. Figure 5 shows that the wafers implanted with corrections enabled fall within <0.1 to the expected Rs value for 0.75/0 orientation. Cassettes of wafers from the same boule used for the implants in Figures 2 to 5 were selected for a repeatability test. Using wafers from a single boule ensured that orientation of the crystal matched the sensitive region near a tilt of 0.75 shown in Figure 2 and minimized changes in the crystal cut error through the test. No attempt was made to compensate for crystal cut error in the selected material, although wafers with a reported crystal cut error of <0.1 were chosen, which is consistent with the curve in Figure 2. Implants were performed daily for over a month using 500 µA of 500 keV P++, at a dose of 5E13 atoms/cm2 and an orientation of 0.75/0 tilt/twist. Recipe limits for horizontal beam angles were set to 0.1 to ensure that horizontal steering or P-lens voltage would be automatically adjusted if any of the seven measured peaks were >0.1. Vertical beam angle auto-correction was enabled so that the wafer angle was adjusted about an axis parallel to the scanned beam to correct for variation in the VBA. 300 305 300 295 290 285 290 280 +/- 0.55 corrected 280 270 -1.00 -0.50 0.00 0.50 1.00 1.50 1 4 7 10 13 16 19 22 25 28 31 34 37 Wafer Rs Ave Average +0.1 deg -0.1 deg Implant angle, deg FIGURE 5. Wafers implanted at a recipe orientation of 0.75/0 with the beam steered by the AEF to cause a vertical angle error of +/- 0.55 with automatic angle corrections on and off are compared with the V-curve. FIGURE 6. Sheet resistance results of daily implants of 500 keV P++ at 5E13 atoms/cm2 at an orientation of 0.75/0 tilt/twist. Upper and lower lines show the change in Rs equivalent to +/- 0.1 degree in tilt. Implant Data Logs for these implants report the measured Horizontal Beam Angle, HBA, and VBA for these implants. Table 1 shows the average value and standard deviation of these angles taken from the IDL’s for all implants. It also shows the average value and standard deviation of the sheet resistance. If the standard deviation of the sheet resistance is all attributed to angle variations, it would convert to a standard deviation of 0.032 for all of the implants. TABLE 1. Data summary for repeatability implants Angle Average Standard standard value deviation deviation based on Rs HBA 0.005° 0.019° NA VBA 0.76° 0.023° NA Rs 295.3 0.74% 0.032° SUMMARY Optima MD has the ability to measure and automatically correct implant angles in the horizontal scan plane and the vertical plane the start of a batch of wafers. A very sensitive channeling implant condition was selected to test repeatability of the implant angle using daily implants for over a month. The orientation of the implant was chosen as 0.75/0 tilt/twist because of the sensitivity (5% change in Rs for 0.25 change in angle) and the near linearity of the response. It was also shown that comparable sensitivity could be obtained by implanting at 34.5/0. The variation in Rs was consistent with total angle variation of less than +/-0.1 with a standard deviation of 0.03 for all implants. ACKNOWLEDGMENTS The authors would like to thank Fernando Silva and Khen Ung for running the implants. REFERENCES 1. K. W. Wenzel, A. M. Ray, B. H. Vanderberg, and R. D. Rathmell, “Optima MD: Mid-Dose, Hybrid-Scan Ion Implanter”, Proc. of 16th Intl. Conf. on Ion Implantation Tech., Marseille, France, AIP Conf. Proc. 866, 2006, pp. 637-640. 2. R. D. Rathmell, B. H. Vanderberg, A. M. Ray, D. E. Kamenitsa, M. Harris, and K. Wu, “Implant Angle Control on the Optima MD”, Proc. of 16th Intl. Conf. on Ion Implantation Tech., Marseille, France, AIP Conf. Proc. 866, 2006, pp. 349-352. 3. R. D. Rathmell, D. E. Kamenitsa, M. L. King, and A. M. Ray, IEEE Proc. of Intl. Conf. on Ion Implantation Tech., Kyoto, Japan, 1998, pp. 392-395