(EPA) Room Air Cleaner Energy Star program.

advertisement
Test Method for Rating
Performance of Portable Room Air
Cleaners
BY
Ramona Saar, Director of Standards and Certification
Association of Home Appliance Manufacturers
1111 19th St. NW Suite 402
Washington DC 20036
USA
Phone: 202-872-5955
Fax: 202-872-9354
E-Mail: rsaar@aham.org
Web Site: http://www.AHAM.org
Ramona Saar
AHAM
Page 2
ABSTRACT
Portable electric room air cleaners have become very popular over the last few
decades. In order to measure and compare the performance characteristics of these
appliances, the Association of Home Appliance Manufacturers (AHAM) developed a
rating test method in the late 1980’s. The AHAM test method, prescribed in
ANSI/AHAM Standard AC-1-2002 has been recently revised after careful review by
manufacturers, technical experts and interested stakeholders in the indoor air quality
field.
The AC-1 standard rates the performance of room air cleaners using a metric called
the Clean Air Delivery Rate (CADR), which is a measure of the appliance’s ability to
reduce smoke, dust, and pollen particles, in the 0.10 to 11 micron (μm) size range from
the air. The standard also prescribes a method for sizing room air cleaners to the
appropriate room size, based on the CADR. In addition, it includes a procedure for
measuring energy consumption in accordance with Energy Star. This paper describes
the CADR test method, the revisions recently made to the standard, and the certification
program that AHAM operates to verify the performance claims of room air cleaners.
INTRODUCTION
The main function of a portable household electric room air cleaner is to reduce
indoor airborne particles. There are many types of portable room air cleaners available
in the market today. These products vary in size from small table top units to large
console units and utilize a number of different particle removal technologies, including
mechanical filters, electrostatic plates, and ionization, for example. The majority of
consumers that buy an air cleaner cite the desire to improve indoor air quality as the
reason for purchasing the product.1 By reducing the particles, the potential harmful
effect of those particles is reduced, improving the quality of the indoor air.
How effective are portable household electric room air cleaners at removing
particles? The Association of Home Appliance Manufacturer (AHAM) has developed
an objective and repeatable performance test method for measuring this key
performance characteristic2. The method, ANSI/AHAM AC-1, Method for Measuring the
Performance of Portable Household Electric Room Air Cleaners is designed to evaluate
the performance of the cleaners regardless of the particle removal technology utilized.
ANSI/AHAM AC-1 forms the basis for an industry certification program for room air
cleaners. The resulting performance measurement, called the Clean Air Delivery Rate
(CADR), is placed on product packaging along with a recommended room size. This
information helps consumers make an appropriate purchasing decision based on
performance and needs.
Since its original development in the early 1980s, the ANSI/AHAM AC-1 test method
has increasingly become the credible industry standard for evaluating room air cleaner
particle removal performance. The test method is utilized by the U.S. Environmental
Protection Agency (USEPA) in its Air Cleaner Energy Star Program. Additionally,
Consumers Union3 uses CADR to determine the air cleaner ratings and product
comparison information that are published in Consumer Reports®. Air Quality experts
often refer to CADR in technical papers as well. In a recent research paper on the use
Ramona Saar
AHAM
Page 3
of air cleaners to reduce fine particulate matter during prescribed burns and wildfires in
Colorado, the authors utilized the CADR metric to select appropriate air cleaners for the
study, noting that CADR is a “useful parameter for characterizing effectiveness of an air
cleaner.” 4
HISTORY OF AC-1
Work on a test method for room air cleaners began in 1981. A number of air cleaner
manufacturers came to AHAM requesting the establishment of a program to test and
certify room air cleaners. At that time, there were no test methods for room air cleaners
that approached real-world use conditions. Room air cleaner claims were usually based
on testing conducted in small chambers and then the results extrapolated to indicate
performance claims in larger rooms. In these small chamber tests, particles were often
“fed” into the air cleaner in lieu of allowing the air cleaner to bring the particles into the
system itself. This process would not allow for evaluation to see if the air cleaner’s fanmotor system was powerful enough to bring the contaminants from the room into the
filter media on its own. In addition, the integrity of the air cleaner’s seals and the air flow
pattern set up by the air cleaner were not being evaluated. Such tests measured the
performance of the filtration media but not of the air cleaner system as a whole. Results
of the different test methods were not comparable.
Ramona Saar
AHAM
Page 4
The goal of the AHAM test method development task force was to come up with a
metric that would show the initial performance of the air cleaner on the entire room and
that could be used to compare performance of different units. This metric would
hopefully reduce market confusion over air cleaner performance claims, which was a
concern of the Federal Trade Commission in the early 1980s.
In the fall of 1982, air cleaner manufacturers established a set of testing parameters
that would more closely relate to consumer usage of air cleaners. AHAM sought
technical guidance from the laboratories that had been performing some of the testing
for air cleaner manufacturers and other experts in the field. Technical issues such as
the size of the test chamber, the particulates to be used, the initial starting concentration
of the particles, and how the particles were to be mixed were established through this
initial effort and are consistent in the standard even today.
The initial test procedure was completed through numerous technical meetings and
experimental testing. In September 1984, it became an AHAM standard, AHAM AC-11984. The AHAM certification program was then formulated and it was officially started
in August of 1985.
In March 1986, AHAM presented AHAM AC-1 and the related certification program
information to the FTC for its consideration and comment. The FTC retained a
technically knowledgeable EPA consultant to aid in its review. With additional
comments from the consultant, precision statements were added to improve the
standard. The updated version was granted approval by ANSI as an American National
Standard in December, 1988. In 1989, the FTC confirmed to AHAM that the
ANSI/AHAM AC-1-1988 test method was a reasonable basis for measuring the degrees
of reductions of airborne solid particulate matter from household rooms.5
2004 INDEPENDENT TECHNICAL EXPERT REVIEW
On November 29, 2006, ANSI/AHAM AC-1-2006 was re-approved as an American
National Standard by ANSI6. The changes reflected in the new updated 2006 version
are a result of a unique AHAM-initiated independent technical expert review of the
standard and a subsequent ANSI consensus panel review.
In 2004, in response to comments suggesting that the test method might not be an
appropriate standard for measuring the performance for certain air cleaner
technologies, the AHAM Air Cleaner Council that oversees the AC-1 standard agreed
that AHAM should contact independent experts earlier in the process of the
ANSI/AHAM AC-1-2002 five year revision cycle so that the standard could be reviewed
objectively for its relevancy to today's air cleaner technologies. The independent expert
review was sought to specifically ensure that the method was technically sound and
was not inadvertently and/or unfairly favoring any design or technology. This is
important because AHAM does not promote any specific brand of air cleaner or any
specific technology for air cleaners.
Seven experts from the government, academia and the private sector agreed to
participate in the independent review and submitted comments. The experts included
representatives from Lawrence Berkeley National Laboratories, University of Texas
(Austin), University of Tulsa, Clarkson University, an independent air quality testing
laboratory, and the National Institutes of Standard and Technology (two from NIST).
Ramona Saar
AHAM
Page 5
The experts were asked to review the AHAM standard within the limits of the current
claimed scope (removal of particulate matter) and its application to today’s
technologies.
The most important outcome of the review was that the experts reaffirmed the
technical validity of the test method and its application to all types of room air cleaners.
In addition, the experts agreed with the scope expansion to include plug-ins and wallmounted air cleaner units and also with the inclusion of a new section on calculating
room size. (Previously, room size calculation was a certification program feature, but
was not included in the actual standard.)
A number of comments were also submitted to improve the clarity of the document.
New annexes were added for a derivation of the effective room size calculation and for
example data sheets, a schematic of a test stand for wall-mounted units, a list of the
test sequence steps, and a timeline of the data acquisition.
Following the expert review, AHAM published AHAM AC-1-2005 and submitted the
document for ANSI approval. During the ANSI consensus panel review process,
optional performance tests for air cleaner operating power and standby power were
added and appear in ANSI/AHAM AC-1-2006. The optional power performance tests
are in line with those included in the Environmental Protection Agency (EPA) Room Air
Cleaner Energy Star program.
TEST METHOD - CLEAN AIR DELIVERY RATE (CADR)
Given time, most particles decay out of the air naturally. To be effective, an air
cleaner must be able to remove particles at a rate that is higher than the normal decay
rate of those particles. The faster the air cleaner can remove the particles, the more
effective it is.
The ANSI/AHAM AC-1 test method describes a process for measuring the relative
reduction by the air cleaner of particulate matter suspended in the air in a specified test
chamber. The unique metric utilized to describe the room air cleaner’s particulate
matter removal performance is the Clean Air Delivery Rate (CADR). CADR is defined
as the measure of the delivery of contaminant free air by a portable household electric
room air cleaner, expressed in cubic feet per minute (cfm).
CADR is the rate of
particulate reduction in the test chamber when the air cleaner is turned on, minus the
rate of natural decay when the unit is not running, times the volume of the test chamber
as measured in cubic feet.
The final results are expressed in terms of a CADR rating for each of three
contaminants used (smoke, dust, and pollen). These contaminants are used to
simulate various sizes of particles that would typically exist in a home. The particle size
ranges measured in ANSI/AHAM AC-1 are noted in Table 1.
Table 1
Particulate Matter
Particle Size
Range Measured
Cigarette Smoke
0.10 μm to 1.0 μm
Fine Test Dust
0.5 μm to 3.0 μm
Ramona Saar
AHAM
Page 6
Paper Mulberry Pollen
5 μm to 11 μm
TEST METHOD - THE TESTING CHAMBER & SEQUENCE OF STEPS
The AC-1 standard requires use of 1008 ft 3 (10.5 ft. x 12 ft. x 8 ft.) sealed chamber.
Flooring in the chamber is specified as seamless smooth surface linoleum or vinyl. The
walls must be painted with a washable latex semi-gloss paint. The environmental
control system consists of humidifiers, a HEPA filter, blowers, supply dampers and
return dampers. The chamber also has a ceiling fan and a recirculation fan that are
used for mixing the air. The recirculation fan remains on during the duration of the test.
The particle counters are located in the chamber.
Figure 2 at the end of this document shows a diagram of the test chamber.
TEST METHOD – TEST SEQUENCE
The CADR for each particulate is determined by a separate test sequence in the test
chamber. It is necessary to measure both the natural decay (air cleaner turned off) and
the measured decay (with air cleaner turned on) separately for smoke, dust and pollen.
Prior to the start of the test, the chamber must be thoroughly cleaned. The air cleaner is
then set up in the chamber. Air cleaners under test are normally placed in the center of
the chamber on a table. A separate test stand is provided for mounting plug-in and
wall-mounted air cleaners.
The chamber’s environmental system, the recirculation fan and the ceiling mixing fan
are turned on. The background concentration of particles is monitored, and when an
acceptable low background concentration level is obtained, the chamber’s
environmental control system is turned off. Particles are then injected and monitored to
obtain a natural decay.
Next, a measured decay test is conducted. The particulate matter being tested
(smoke, dust or pollen) is injected into the room to the required concentration level
specified in the standard. The air supply used to inject the particulate matter is then
turned off and the injection valve for the pollutant closed. The pollutant is allowed to mix
for one minute via the recirculation fan and the ceiling mixing fan. The mixing fan is
then turned off and one additional minute is provided for the mixing fan blades to stop.
The particle counter sampling and the data acquisition then begin. There are slightly
different sequences for the smoke, dust and pollen. The data acquisition is done
through a 20 second sampling at one minute intervals. The test for smoke and dust is
20 minutes. For pollen, the test is 10 minutes. The minimum number of acceptable
data points must be obtained (at least 9 for dust and smoke, 5 for pollen). Following
completion of the test sequence, the data analysis is performed.
TEST METHOD - DATA ANALYSIS – DECAY CONSTANT
For each particulate, the decay constant, k, is calculated.
Since the test chamber air exchange is negligible (less than 0.03 air changes per
hour) the air exchange rate is not included in the calculations.
Ramona Saar
AHAM
Page 7
The decay constant, k, for particulate matter is based on the formula:
Cti=Cie-kti
where:
Cti = concentration at time ti (particles/cc)
Ci = concentration at t = 0 minutes
k = decay rate constant (minutes -1)
ti = time (minutes)
The decay constant, k, is obtained using the linear regression on the lnC ti and ti using
the formula:
k=
SXY
SXX
where:
N
Sxy =
Sxx =
N
∑ti lnCti - (1/n) (∑ti)
i =1
i =1
N
N
∑(ti)2 - (1/n)
i =1
N
(∑ln Cti)
i =1
( ∑ti)2
i =1
When the above calculations are used for natural decay measurements, the results
represent the natural decay rate in the room air.
When the above calculations are used for the total particulate matter removal
measurements the results represent the air cleaner particulate matter removal rate,
which includes natural decay of the particles also.
TEST METHOD – CADR CALCULATION
Ramona Saar
AHAM
Page 8
The CADR is then calculated using the following equation:
CADR = V(ke - kn)
where:
CADR = clean air delivery rate (cu. ft/min)
V= volume of test chamber, cu. ft.
ke= total decay rate, min-1
kn = natural decay rate, min
-1
LIMITS OF MEASURABILITY
There are defined limits of measurability for AC-1 based on the statistical accuracy of
the method. Only CADR rates within the limits specified below can be considered
statistically valid:
Dust CADR = 10 cfm to 400 cfm
Smoke CADR = 10 cfm to 450 cfm
Pollen CADR = 25 cfm to 450 cfm
The maximum CADR values stated here were determined through analysis of the
acceptable minimum number of available particles, an average background natural
decay rate (from statistical study), the size of the chamber, and the available minimum
experiment time. CADR values outside the ranges listed will not have the necessary
statistical data required by this method.
PRECISION
The precision of the method, expressed as 2 standard deviation limits (95%) are:
Dust
Smoke
Pollen
CADR = + 10 cfm
CADR = + 10 cfm
CADR = + 25 cfm
DETERMINING ROOM SIZE7
Along with CADR performance ratings for dust, smoke and pollen, ANSI/AHAM AC1-2006 now includes a process for calculating the Suggested Room Size for a specific
Ramona Saar
AHAM
Page 9
room air cleaner. The methodology used for this process was developed in 1992 and
was added to AHAM’s Certification Program following FTC confirmation that the
methodology proposed was scientifically valid8.
The room size is based on several standard construction criteria for rooms and a
history of the natural decay of small particles (cigarette smoke). Cigarette smoke was
selected because smoke particles are the hardest to remove from the air and have the
slowest natural decay rate.
The effective room size is based upon the CADR obtained for cigarette smoke and
the ability of the air cleaner to reduce the concentration of particles in a room at steadystate to a new steady-state concentration 80% less than the original, when the air
cleaner is operating. The 80% reduction criterion was determined by examining
different room sizes within a home or office and observing the significant differentiation
between air cleaners. 80% particle remove was considered realistic for the selection of
an air cleaner. The theoretical assumptions of the room characteristics are based upon
(1) a mixing factor equal to 1.0, (2) an air exchange rate of 1.0 hr -1, (3) a cigarette
smoke particle natural decay equal to the average background natural decay rate, (4) a
ceiling height of 8 ft (2.4 m), and (5) a cigarette smoke particle generation or influx rate
such that a cigarette smoke particle concentration of unity is maintained at the initial
steady state.
A detailed derivation of the room size calculation, as it appears in ANSI/AHAM AC-12006, is included in the following section. Based on the assumptions above, the
equation for the Suggested Room Size is:
Room Size (ft2) = Smoke CADR x 1.55
DERIVATION OF ROOM SIZE CALCULATION
Concentrations of particles in indoor air are dynamic and result from the competition
between various source and removal processes. Steady state can be defined as when
neither the source nor the removal processes are rapidly changing and thus the indoor
concentration is relatively constant. In this situation,
C = source terms
removal terms
(1)
C = indoor sources+outdoor sources
ventilation+air cleaning+deposition
Without air cleaning, particle removal is through ventilation and deposition.
Rearranging these terms and substituting decay parameters for words in equation (1):
Source terms  C  k V  k dep ,

(2)
where kV is removal rate due to ventilation (minutes-1), kdep is the removal rate due to
deposition (minutes-1) and when used in equation (3) below, kAC is the removal rate due
to air cleaning (minutes-1). Steady-state for cleaning is defined by AHAM Air Cleaner
Ramona Saar
AHAM
Page 10
Council as being 20% or less of the initial particle load in a room, or in other words: at
least an 80% continuous removal of smoke particles. This defines a new steady state
particle concentration, now with the air cleaner operating, CAC, which equals 0.2C.
Using the same formula as in equation (2) above,
Source terms  0.2C  k V  k dep  k AC (3)



Since the source terms do not change – only the use of an air cleaner – the two
equations can be combined through the equivalent source terms to give
C  k V  k dep  0.2C  k V  k dep  k AC (4)
Canceling the C’s from both sides, rearranging and solving for the air cleaner
removal rate:
k AC  4k V  k dep .
(5)
Recall that CADR is defined as V  k e  k n , where V is the room volume (ft3) and ke
and kn are the decay rates with and without air cleaner operation, respectively (as
defined in section 8.4).

Effectively then,
air cleaner operating decay = ke =
(kv + kdep + kAC)
(6)
and natural decay = kn = (kv + kdep)
(7)
This difference in decay rates is kAC, thus
CADR  V  k AC  4Vk V  k dep .

The objective is to show what size room results from the minimum air cleaner
performance needed to provide an 80% reduction in steady-state particle
concentrations. For simplicity, it is assumed that the ceiling height is 8 feet; this permits
recasting of the equation in terms of floor area, A, a more commonly known description
of room size than is room volume,
A

(8)
CADR
.
32k V  k dep 
(9)
Ramona Saar
AHAM
Page 11
Commonly accepted values for the two decay parameters are used. Typical air
exchange (ventilation) rates for houses in the U.S. is 1 per hour, or kV = 0.01667 minute1.
For smoke, average deposition rates are 0.0034 minute -1, as provide by AHAM
through actual measurements by a third party independent laboratory. Substituting
these values in equation (9):
A (ft 2) = CADR / [32(0.01667 + 0.0034)] = 1.557 CADR (cfm)
(10)
In a meeting of the AHAM Air Cleaner Council, the value of 1.557 was rounded to
1.55 for simplicity, thereby providing the standard equation (11):
Room Size in Square feet = 1.55 x cfm
Based on this equation, an air cleaner with a CADR rating of 50 cfm can be used to
clean a small room (78 ft2). Likewise a CADR rating of 100 cfm would permit an air
cleaner to be used in a 156 ft2 room (~10 x 15 ft), and so forth. Conversely, this
equation can also be used to determine what the CADR requirement is for a room or
indoor space of a given size. For example, for a room size of 100 ft 2 a CADR rating of
at least 64 cfm is needed.
INDUSTRY CERTIFICATION
Since 1985, AHAM has administered a portable room air cleaner certification
program for industry. Both AHAM and non-AHAM members are eligible to participate in
the program. To reduce market confusion, participants must include all the units they
manufacture in the program. In 2005, there were 24 manufacturers participating in the
program.
Three samples of each air cleaner model entering the program must be submitted for
initial certification testing at the independent AHAM-contracted testing laboratory. Initial
certified CADR results are calculated from the averaged results of the three tests and
the recommended room size is determined form the averaged certified smoke CADR
results.
Air cleaner models included in the program are then independently tested on a
periodic on-going basis to ensure that the units meet the claimed ratings. The models
to be tested are randomly selected and are obtained from either the participant’s
warehouse or the open market. Units that do not successfully pass the on-going
verification process must be re-rated or withdrawn from the market.
AHAM provides all participants with the following CADR ratings seal shown below in
Figure 1.
Ramona Saar
AHAM
Page 12
FIGURE 1. CADR Ratings Seal
Participants are required to include the seal on the packaging of all certified room air
cleaners. The seal provides the consumer with a simple tool to compare the
performance of room air cleaners, making that purchasing decision easier. It includes
the participant’s certified CADR rates for tobacco smoke, dust and pollen, and a
suggested room size that is based on the tobacco smoke CADR results. Using the
ratings seal, consumers can comparison shop, selecting the unit the meets their room
size and performance needs. By knowing the size of their room, consumers can
choose an effective air cleaner.
Certified models are listed in an on-line searchable Directory of Certified Room Air
Cleaners at a dedicated product web site: www.cadr.org.
SUMMARY
ANSI/AHAM AC-1 provides a realistic, effective measurement of the ability of air
cleaners to reduce indoor particulate matter by examining the decay rates of particles in
a sealed chamber when an air cleaner is turned off and when an air cleaner is turned
on.
The test method, developed through AHAM and the manufacturer members,
reviewed in 2004 by an independent panel of technical experts, and re-approved by
ANSI in 2006, has been shown to be a technically sound, reproducible protocol that can
be used for all types of portable household electric room air cleaners. The information
provided through AHAM’s Room Air Cleaner Certification Program ratings seal, and
placed directly on the air cleaner packaging provides the consumer with an objective,
easy way to determine which air cleaner is best for their use.
ABOUT AHAM
AHAM is a full service trade association representing the major appliance, portable
appliance, floor care appliance manufacturers and suppliers to the industry. AHAM
provides its members with service in the following areas: government relations,
business data, communications, technical standards development and product
certification. In the standards development area, AHAM is an accredited standards
development organization (SDO) through ANSI (American National Standards Institute).
Selected AHAM appliance performance standards are processed through the ANSI
consensus review process. This process is designed to ensure that American National
Standards have been developed through a consensus process that included a balance
Ramona Saar
AHAM
Page 13
of interest and lack of dominance by any stakeholder group. The ANSI/AHAM AC-1
standard that is the subject of this paper was re-approved by ANSI at the end of 2005
and is being published as ANSI/AHAM AC-1-2006. In the area of product certification,
AHAM has been operating industry programs since 1967 to independently verify
manufacturer performance claims. The AHAM Room Air Cleaner certification program
that is the subject of this paper has been operating since 1985.
Ramona Saar
AHAM
Page 14
FIGURE 2. AIR CLEANER CHAMBER SCHEMATIC
SCHEMATIC
(O)
(N)
(Q)
(P)
(R)
(S)
(T)
(L)
(M)
(K)
(J)
(F)
(G)
(E)
(H)
(I)
(D)
(C)
(B)
(A)
(A) VOLTAGE REGULATOR
(B) DATA ACQUISITION AND
CONTROL INTERFACE
(C) AIR SUPPLY (FILTER/DRIER)
(D) COMPUTER TERMINAL
(E) CIGARETTE SMOKE POT
(F) POLLEN GENERATOR
(G) DUST AND POLLEN
MONITOR
(H) SMOKE MONITOR
( I ) CIGARETTE SMOKE DILUTER
(J)
(K)
(L)
(M)
(N)
(O)
(P)
(Q)
(R)
(S)
(T)
DUST GENERATOR
TEST UNIT
CEILING MIXING FAN
RETURN AIR DAMPER (2)
RECIRCULATION FAN
HUMIDIFIER
PREFILTER
BLOWER SECTION
HEPA FILTER
ELECTRIC HEATER
SUPPLY AIR DAMPER
Ramona Saar
AHAM
Page 15
REFERENCES
1
2002 consumer research study commissioned by AHAM.
Other room air cleaner performance characteristics such as the ability of the air
cleaner to reduce gases, odors or microbiological components, or the sound and ozone
emissions levels of the product are outside the scope of ANSI/AHAM AC-1. Some of
these additional performance characteristics such as sound and ozone production are
covered under other standards, including ANSI/AHAM AC-2 (Sound), and UL 867
(Section 37 – Ozone).
3 Consumers Union (CU) is a nonprofit organization whose main mission is to ensure
that there is a fair and safe marketplace for consumers. CU tests products and
publishes the product ratings in its magazine.
4 Henderson, Milford and Miller. Prescribed Burns and Wildfires in Colorado: Impacts
of Mitigation Measures on Indoor Air Particulate Matter; Air & Waste Management
Association. 2005, 55, 1516-1526.
5 Sanger, William S, Associate Director for Enforcement, Bureau of Consumer
Protection, FTC, Letter dated June 1989.
6 American National Standards Institute.
7 ANSI/AHAM AC-1-2006
8 Dingfelder, Justin, Assistant Director, Enforcement Division, Bureau of Consumer
Protection, FTC, Letter dated May 21, 1993.
2
Download