BOD Analysis using LCK555 – brief user guide DR2800/ DR5000 and: LZC901 Dilution Water Set, LZC555 Inoculation Material, LCA555 BOD Standard 1. Preparation of Dilution Water Set up the dilution water kit as shown in the diagram. The pump should be placed above the water level in the jar Add 500ml tap water to the jar and aerate for 5 minutes to expel free chlorine Pipette 0.3ml of LCK555 trace element “D” into the jar. Add appropriate inoculation material to the jar. Either: i. ii. 5ml particle-free, municipal wastewater with COD @250mg/l or 20ml particle-free, municipal wastewater with COD @50mg/l Then aerate for 3 days. This water could be kept for up to 8 days Or iii. 0.5ml prepared LZC555 inoculation solution (see box below). Then aerate for 1 hour. This water should be used immediately. Preparation of LZC555 1. Add a level dosing spoon of inoculation material into the reaction tube 2. Pipette 10ml of buffer solution into the tube 3. Shake for 1 minute 4. Leave to stand for 1 hour at room temperature 5. Pipette 0.5ml of the solution into the dilution water jar 2. Sample Preparation Homogenise the sample at 20°C, 700-900rpm for 5 minutes (for larger particulates, homogenise @20,000rpm for 5 seconds) Select the appropriate range for the test, based upon estimated BOD. This will determine volume of sample/ dilution water used in the sample preparation step, and then the dilution factor entered in the evaluation step after 5 days. If it is not possible to estimate the BOD, make an approximation from the COD value: Untreated industrial and municipal wastewater- BOD (mg/l) is @35-65% COD Biologically treated wastewater- BOD (mg/l) is @25% COD Prepare the sample in the dilution tube according to the table. Shake the dilution tube for 1 minute to aerate the sample fully Transfer the appropriate volume into a cuvette and fill to the brim with dilution water eg. For a sample with estimated BOD of 75mg/l – measuring range would be B1 Pipette 1ml of sample and 1 ml of dilution water into the dilution tube. Shake, then pipette 0.5ml into the cuvette and fill with dilution water. Enter dilution factor of 25 in evaluation step Screw on the cuvette cap, ensuring no air bubbles are present At the same time as preparing the samples, fill a cuvette with dilution water only. This will be used as the blank in the evaluation step (1 needed per day’s BOD analysis) Incubate the cuvettes at 20°C in the dark for 5 days using an appropriate incubator (such as LT20) After 5 days, evaluate using DR2800/ DR5000 etc 3. Evaluation using DR2800/ DR5000 Remove the cuvettes from the incubator For each cuvette in turn remove the foil from the dosi-cap, unscrew the cap and mount the funnel Pour in the contents of the dosi-cap (reagent and glass beads) into the cuvette, screw on the cap (again ensuring no air bubbles present) and invert for 3 minutes o nb if liquid meniscus drops, add 2-4 glass beads to make up the volume Wait for a further 3 minutes Insert the blank into the DR2800/ DR5000 Upon obtaining the E1 message, insert the first sample. When prompted, enter the appropriate dilution factor for the sample The instrument will display the result in mg/l Repeat for each sample cuvette 3. Use of LCA555 for AQC (Analytical Quality Control) Pipette 5ml distilled water into one of the LCA555 cuvettes Shake vigorously (upto 3 minutes) to dissolve all of the contents Pipette 2.2ml of this into a clean 50ml volumetric flask Top up with distilled water and shake This is equivalent to a 205mg/l +/-45mg/l BOD standard Using LCK555, this is analysed in measuring range B2 Pipette 0.5ml of this directly into a cuvette and fill with dilution water For evaluation, enter a dilution factor of 50 The result should be between 160- 250mg/l n.b BOD standards can also be prepared using dried glucose/ glutamic acid Note on use of LCK554 - measuring range 0.5- 12mg/l Undiluted – 0.5- 6.0mg/l Diluted (x2) – 1.0- 12.0mg/l Sample preparation: Pour appropriate sample quantity (undiluted – 40ml sample; Diluted – 20ml sample and 20ml chlorine-free drinking water) into a beaker and stir for 5 minutes at 500- 750rpm LCK554 is measured directly - 2 sample cuvettes are required for each sample – cuvette 1 is measured on day 1, cuvette 2 measured on day 5. Label these e.g. A1 and A5 Fill each cuvette to the brim with the prepared sample, and incubate the cuvette 2 sample(s) for 5 days at 20°C (ensure no air bubbles are present) For sample 1 cuvette(s), add the dosi cap contents, invert for 3 minutes and wait 3 minutes Evaluate using spectrophotometer, noting the results for use to calculate the BOD on day 5 On day 5, evaluate the cuvette 2 sample(s) in the same way Calculate BOD using the day 1 and day 5 results: e.g. A1- A5= BOD (mg/l) If a dilution is made, ensure this is included in the calculation Points to Note Peroxides, strong oxidising or reducing agents and high chlorine concentrations can interfere with oxidation, resulting in high or low bias Sample pH should be between 4 and 10 It is important to check the table of interferences in the procedure for ions present in the sample It is recommended to run AQC checks, such as with LCA555, every 10 th sample