CM20 Alignment Procedures Last Revised: 20070508 Gun Alignments Gun Tilt Gun Shift Condenser System Alignments Aperture Centering Aperture Stigmation Sample Alignments Sample Height Sample Orientation Objective Lens Alignments Pivot Points Rotation Centers Objective Aperture Centering Final Alignments Objective Lens Astigmatism Coma Free Alignment Sample Orientation Imaging 1 Gun Alignments It is necessary to align the gun such that the electron beam leaves the gun and travels along the optical axis. This will provide optimum imaging conditions and maximum brightness. Gun Tilt 1. Set the magnification to ~30kX. 2. Condense the beam using the SHIFT and INTENSITY knobs until the beam is approximately half the size of the screen. 3. Press the ALGN button on the right panel and then Gun TILT using the screen buttons 4. Use the MULTIFUNCTION knobs to maximize screen brightness 5. Exit the alignments menu by pressing ALGN Use the exposure time readout or your own eyes to obtain best result Should the beam leave the screen, bring it back to the center using the SHIFT knobs Condense the beam to the smallest possible size Select “Gun Tilt” from the Direct Alignments menu. Use the multifunction knobs to bring the bright spot to the centre of the beam. It may be necessary to move the beam back to centre using the left trackball. Gun Shift 1. Check if the beam for spot size 3 to 9 stays in the same position, if not, perform the alignment 2. Press ALGN and then Gun SHIFT 3. Select spot size 9 and center it using the SHIFT knobs 4. Select spot size 3 and center it using the MULTIFUNCTION knobs 5. Repeat steps 3 and 4 until spot movement is minimized 6. Exit the alignments menu by pressing ALGN 2 Condenser System Alignments The condenser system determines the beam convergence angle of the incident beam and therefore affects the spatial coherency envelope function of the microscope. It also affects the intensity on the screen. Aperture Centering 1. Select the correct C2 aperture – this is typically the second largest 2. Select SPOT SIZE 1or 2 for most imaging applications 3. Select emission 1 or 2 in the PARAMETERS menu 4. Leave this menu by pressing the READY button 5. Set the MAGNIFICATION to ~50kX 6. Center the beam using the SHIFT knobs 7. Center the aperture until the beam can expand and condense concentrically when going through INTENSITY Condenser Stigmation 1. Set the MAGNIFICATION to ~50kX 2. Expand and condense the beam using the INTENSITY knob. Make sure that the beam stays round (not elliptic) and expands concentrically, if not: 3. Press STIG on the right panel and COND using the screen buttons 4. Adjust by using the MULTIFUNCTION knobs 5. Press STIG again to leave the menu 3 Sample Alignments Sample Height 1. Set the MAGNIFICATION to ~100kX 2. Position an area of interest in the center of the screen, an edge is strongly recommended. 3. Expand the beam with the INTENSITY knobs until it covers the screen 4. Set the objective lens current to 9990mA 5. Adjust the sample height using the screw on the goniometer until the image is in focus (minimum contrast). It should be possible to bring the sample in focus +/- 100Å or better If the screw reached the end position before Gaussian focus is obtained, the remove the sample holder, remount the sample upside down and try again. The spherical aberration constant is extremely sensitive to variations in the lens current (focus changes). This should thus be kept reasonably constant. Coarse adjustments in focus should be performed by changing the position of the sample in the microscope with fine adjustments left to adjustments in the lens current. Sample Orientation 1. Set the MAGNIFICATION to ~50kX 2. Position an area of interest in the center of the screen 3. Condense the illumination on that area using the INTENSITY knobs 4. Press D to obtain a diffraction pattern 5. Tilt the sample using the goniometer controls, to a low index zone axis of your choice and orient it such that the intensity of the reciprocal spots is evenly distributed on either side of the transmitted beam. (The low index plane is then normal to the optical axis) The sample usually moves while tilting unless it is rotated in the plane of the sample (which it should be if sample height is at eucentric height, see last alignment). This causes the illuminated area to change and consequently, the diffraction pattern does not correspond to the original selected area. Iterate by moving the sample in real space and tilting in reciprocal space until the area of choice is aligned to the optical axis. 4 Objective Lens Alignments Pivot Points To ensure that shifting the beam does not also tilt the beam, the pivot points must be aligned. 1. Set IMG SHIFT to ZERO with the screen controls 2. Set the MAGNIFICATION to ~50kX 3. Use the FOCUS knob to bring the sample to exact (Gaussian) focus 4. Condense the beam to a minimum and centre it using INTENSITY and SHIFT 5. Press ALGN and Pivot Points X 6. Adjust MULTIFUNCTION knobs until the two spots overlap perfectly 7. Repeat for Pivot Points Y 8. Exit the menu by pressing ALGN Should the beam leave the screen while adjusting the pivot points, use SHIFT to bring the beam back or lower the magnification. Rotation Centers Current and voltage fluctuations should not produce a image shift or defocus the image. The beam should be placed in the axis of rotational symmetry of the objective lens. This is the same as aligning the incident beam to the optical axis. Another benefit from this procedure is that lens aberrations are less distinct near the optical axis. 1. Set the MAGNIFICATION to ~100Kx 2. Position an area of interest in the center of the screen, an edge is strongly recommended. 3. Expand the beam with the INTENSITY knobs until it covers the screen 4. Use the FOCUS knob to bring the sample to exact (Gaussian) focus 5. Check that the objective lens current is still about ~9990mA, otherwise return to “Sample Height” 6. Set FOCUS to step 3 or 4 (this gives the modulation amplitude) 7. Press ALGN and ROTATION CENTER 8. Choose either current or voltage but don’t do both (current is the more suitable choice and voltage the easier…) Adjust MULTIFUNCTION knobs until there is no movement in the image 9. Repeat for successively higher MAGNIFICATION up to ~300kX 10.Leave the menu by pressing ALGN 5 Objective Lens Stigmation Astigmatism causes images to appear diffuse and causes streaks in the defocused image. 1. Set the MAGNIFICATION to ~300kX, if the astigmatism is bad, go to lower magnifications 2. Move to an amorphous edge, preferably one that is thin (glue is not amorphous) 3. Press STIG on the right panel and then OBJ on the screen buttons 4. Use step 3 for coarse adjustments and smaller steps for fine adjustments 5. Bring the image into focus using the FOCUS knob 6. Use the MULTIFUNCTION knobs to make the image sharper, remove streaks or to ensure that the defocus is consistent around edges. 7. Increase the magnification to ~800kX and adjust again 8. Press STIG to leave the menu It is possible to use the CCD for this. Start the program “Digital Micrograph” on one of the standalone computers. REDUCE THE INTENSITY ON THE FLUORESCENT SCREEN UNTIL THE SCREEN IS BLACK. Insert the CCD and press Start View in Digital Micrograph. Increase the intensity SLOWLY until an image is obtained. Use ROI Tools, hold down the ALT key on the keyboard and mark out a square on the image. Then select Process -> Live -> FFT to obtain a Fourier transform. Astigmatism is minimized when the Fourier transform is round. Best focus is when the Fourier transform is largest (and slightly white) 6 Objective Aperture To remove the high spatial frequency oscillations in the contrast transfer function of the objective lens and to reduce the confusion in the image, we introduce an aperture in the back focal plan of the objective lens, which blocks electrons with high spatial frequency 1. 2. 3. 4. Move the sample to an area that is both interesting and thin Bring the sample to exact focus using the FOCUS knob Press D to obtain a diffraction pattern Using the diffraction pattern as a guide for the eye, insert and center the second smallest objective aperture around the transmitted beam 5. Return to imaging mode by pressing D 6. Check with the astigmatism if the objective aperture is centered. If the aperture is off axis the image will again appear astigmatic. Note: Once alignment is completed, dark field imaging may be used by centering the aperture over a non-central diffraction spot. High resolution dark field imaging is obtained by first centering the aperture for bright field imaging then tilting the beam using the DF button and placing a noncentered spot in the center of the aperture. 7 Coma-Free Alignment Coma is an aberration associated with objects near the optical axis. Previous alignments are not sufficient to bring the incident beam close to the optical axis since even beam tilts of 1 mRad can affect the image. This step ensures that the beam is indeed aligned with the optical axis. Deviations from the optimal position are found by inducing slight tilts to either side of the optical axis. A beam tilt causes a slight change in focus of the image and this change should be the same (symmetrical) on either side of the optical axis. 1. 2. 3. 4. 5. Set the magnification to ~500kX Move to an amorphous edge, preferably one that is thin Bring the sample into focus using the FOCUS knob Press ALGN and COMA FREE If there is no image, reduce the magnification and perform the “pivot points” correction again. 6. Press coma x 7. Adjust the MULTIFUNCTION knobs until the two images have the same defocus and appear similar 8. Repeat for coma x 9. Change focus and check the images remain similar 10.Perform the Objective Lens Astigmatism correction again The Fourier transform should be the same for both images once coma is properly corrected. 8 Imaging 1. Move the sample to an area of interest 2. Use the handlebars for coarse adjustments and IMAGE SHIFT for fine adjustments 3. Choose the best defocus for optimal imaging conditions (use “Scherzer” for best point resolution) 4. Check (Objective) Astigmatism 5. Check Coma 6. Check Focus 7. Check the drift in the sample 8. Don’t condense the beam too strongly. A convergent beam gives poor resolution 9. Use a smaller spot size (higher number) – intensity will unfortunately go down 10.Use lower emission (less electrons) 11.Take Images BE CAREFUL WITH THE CCD Good Luck! 9