Multiwire Detector Physical Characteristics and Imaging Performance:

Product Specifications: ProTec 1024 Multiwire Gamma Camera
Multiwire Detector Physical Characteristics and Imaging Performance:
UFOV
Active Depth
Readout Wires
High Speed Delay Line Readout
Lead Shielding
Pressurization
Spatial Resolution
Energy Resolution
Imaging Energy Range
Collimator Sensitivity
Spectroscopy
Frame Rate Maximum
20% Dead Time loss
Maximum Count Rate
Frame Processor
Data Corrections
10 inch (25cm) diameter circular field
2 inches (5cm)
128 x 128
10 nsec/cm
1.75mm thickness of shielding in head providing <2%
leakage up to 80keV
Xenon / Methane (90/10) 20-30 psi
2.5mm FWHM Intrinsic
At 60keV: 25% FWHM
20keV - 100keV
Defined as counts per second detected through collimator
per mCi of 178Ta in air – 40 x 103 cps
0 to 100 keV in steps of 1 keV
32 x 32 – 160 frames/ second, 64 x 64 – 40 frames/second
>450,000
>1,000,000 cps
68HC11 8 MHz microprocessor
Radioactive decay, Field Uniformity - Intrinsically
uniform
Detector Movement, Demensions and Ambient Operating Requirements:
Up/Down
Maximum Height
(detector facing down)
Maximum Height
(detector facing up)
Maximum Reach
Yoke Rotation
Head Rotation
Length
Width
Height
Weight
Power
Transformer
Temperature
Humidity
Mobility
Motor Driven, 35.4 ipm (90 cm/min)
53.7 inches (136.4 cm)
31 inches (78.7 cm)
From side of cart to center of detector, 34 inches (86.4 cm)
with manual lock.
360 degrees
360 degrees
30 inches (76.2 cm)
24 inches (61 cm)
36.5 inches (92.7 cm)
525 lbs.
2.2 amp, 115 VAC, 60Hz
Built-in isolation transformer
60 – 80 degrees F
20 – 60% relative (noncondensing)
Manual transport (8" balloon wheels)
Software Highlights:
Acquisition
*Dynamic Studies
*Static Studies
*Analog EKG
*Analog 0.5 V signals (maximum of 4)
*Calibration
*Live display in all acquisition modes
*Energy Spectra
*Multiple image formats – 32 x 32, 64 x 64
Display
$*Image File retrieval
*Live gray scale and color scale windowing in 1% steps,
256 shades of gray
or colors from pointing device.
*Multiple color and gray scale selection
*Image rotation and flip
*Live Cine Display; synchronous side-by-side display with
windowing
alphanumeric data
*Region of Interest; any irregular closed shape
*Flexible user save and display of images in Windows®
.bmp format
Application Software
*Patient report generator
*Research report generator
*Patient demographic data entry
*Patient database linked to patient report
*Patient database manipulation
*Energy spectrum collection
*System diagnostics
*Archival and retrieval in compressed format
Clinical Cardiac Study Software –
First Pass, Left and Right Ventricle
*Automated with minimal user interventions for
Ventricular Analysis
*Automated QRS timing point location with user edit
*Generates raw data ventricular histogram
*Performs temporal smoothing
*Corrects for lung background
Most functions of the system are controlled by the central notebook computer by Windows™ - based
software functions. These include HV level settings, pulse height window, static acquisition of the images
and dynamic first-pass acquisition of images. Many variables are monitored from the same computer,
including detector pressure, detector temperature and detector HV+ and HV-.