Product Specifications: ProTec 1024 Multiwire Gamma Camera Multiwire Detector Physical Characteristics and Imaging Performance: UFOV Active Depth Readout Wires High Speed Delay Line Readout Lead Shielding Pressurization Spatial Resolution Energy Resolution Imaging Energy Range Collimator Sensitivity Spectroscopy Frame Rate Maximum 20% Dead Time loss Maximum Count Rate Frame Processor Data Corrections 10 inch (25cm) diameter circular field 2 inches (5cm) 128 x 128 10 nsec/cm 1.75mm thickness of shielding in head providing <2% leakage up to 80keV Xenon / Methane (90/10) 20-30 psi 2.5mm FWHM Intrinsic At 60keV: 25% FWHM 20keV - 100keV Defined as counts per second detected through collimator per mCi of 178Ta in air – 40 x 103 cps 0 to 100 keV in steps of 1 keV 32 x 32 – 160 frames/ second, 64 x 64 – 40 frames/second >450,000 >1,000,000 cps 68HC11 8 MHz microprocessor Radioactive decay, Field Uniformity - Intrinsically uniform Detector Movement, Demensions and Ambient Operating Requirements: Up/Down Maximum Height (detector facing down) Maximum Height (detector facing up) Maximum Reach Yoke Rotation Head Rotation Length Width Height Weight Power Transformer Temperature Humidity Mobility Motor Driven, 35.4 ipm (90 cm/min) 53.7 inches (136.4 cm) 31 inches (78.7 cm) From side of cart to center of detector, 34 inches (86.4 cm) with manual lock. 360 degrees 360 degrees 30 inches (76.2 cm) 24 inches (61 cm) 36.5 inches (92.7 cm) 525 lbs. 2.2 amp, 115 VAC, 60Hz Built-in isolation transformer 60 – 80 degrees F 20 – 60% relative (noncondensing) Manual transport (8" balloon wheels) Software Highlights: Acquisition *Dynamic Studies *Static Studies *Analog EKG *Analog 0.5 V signals (maximum of 4) *Calibration *Live display in all acquisition modes *Energy Spectra *Multiple image formats – 32 x 32, 64 x 64 Display $*Image File retrieval *Live gray scale and color scale windowing in 1% steps, 256 shades of gray or colors from pointing device. *Multiple color and gray scale selection *Image rotation and flip *Live Cine Display; synchronous side-by-side display with windowing alphanumeric data *Region of Interest; any irregular closed shape *Flexible user save and display of images in Windows® .bmp format Application Software *Patient report generator *Research report generator *Patient demographic data entry *Patient database linked to patient report *Patient database manipulation *Energy spectrum collection *System diagnostics *Archival and retrieval in compressed format Clinical Cardiac Study Software – First Pass, Left and Right Ventricle *Automated with minimal user interventions for Ventricular Analysis *Automated QRS timing point location with user edit *Generates raw data ventricular histogram *Performs temporal smoothing *Corrects for lung background Most functions of the system are controlled by the central notebook computer by Windows™ - based software functions. These include HV level settings, pulse height window, static acquisition of the images and dynamic first-pass acquisition of images. Many variables are monitored from the same computer, including detector pressure, detector temperature and detector HV+ and HV-. Microsoft, MS-DOS, Windows and other names of Microsoft products referenced are trademarks or registered trademarks of Microsoft Corporation. Copyright © 1996 Microsoft Corporation. All Rights Reserved. Microsoft neither endorses nor acknowledges our product or this website.