Human IgG-Fc ELISA Quantitation Kit Manual Catalog number: 40-288-20083F Kit Contents: (Enough for ten 96-well plates) Coating Antibody (Catalog # 15-288-20083F) Affinity-purified Chicken anti-Human IgG-Fc Quantity: 200 pg Calibrator (Catalog # 10-288-20083F) Pure Human IgG-Fc Antigen Quantity: 3 pg HRP Detection Antibody (Catalog # 27-288-20083F) Affinity-purified Chicken anti-Human IgG-Fc – HRP Conjugate Quantity: 5 pg Buffers, Substrate and Plates not included. Notes: Range of Detection: 300 – 0.412 ng/ml Shelf life: One year from date of receipt. Storage: -20°C. Assay Condition: The kit performance has been optimized for the stated protocol using the materials listed and standard dilutions from 300 – 0.412 ng/ml of human IgG-Fc. For alternative assay conditions, the operator must determine appropriate dilutions of reagents. ELISA assay reactivity is sensitive to any variation in operator, pipetting and washing techniques, incubation time or temperature, composition or reagents, and kit age. Adjustments may be required to position the standard curve and/or samples in the desired detection range. Specificity: By immunoelectrophoresis and ELISA the antibodies in this kit react specifically with human IgG-Fc, not with other human serum proteins. The following factors prepared in the detection range of this kit, 300 – 0.412 ng/ml were assayed and exhibited no cross-reactivity or interference. Human Serum Albumin Human Transferrin Human IgM Human IgA Country of 0rigin: United States of America Assay Use: For in vitro laboratory use only. Not for any clinical, therapeutic, or diagnostic use in humans or animals. Not for animal or human consumption. For research use only; not for diagnostic or therapeutic use. Page 1 Human IgG-Fc Quantitative ELISA Protocol Standard Curve Buffer Preparation 3 2.5 1. Prepare the A. o a t i n g 2 1.5 1 B u f f e r , 0.5 0.1 1 10 100 1000 Concentration y = ( (A - D)/(1 + (x/C)^B ) ) + D: Std (Standards: Concentration vs MeanValue) A 0.417 B 1.002 C 10.594 D 2.998 R^2 1 0 .05 M Carbonate-Bicarbonate, pH 9.6 B. Wash Solution, 0.05% Tween 20 in PBS, pH 7.4 C. Blocking Solution, 50 mM Tris, 0.14 M NaCl, 1% BSA, pH 8.0 D. Dilution buffer, 50 mM Tris, 0.14 M NaCl, 1% BSA, 0.05% Tween 20, pH 8.0 E. Enzyme Substrate, TMB (KPL, Cat # 50-76-00) F. Stopping Solution, 2 M H2SO4 or other appropriate solution Step-by-Step Method (Perform all steps at room temperature, 25 o C 1. Coat with Capture Antibody A. B. C. D. E. . Blocking (Postcoat) A. B. C. 3. Determine the number of single wells needed. Standards, samples, blanks and/or controls should be analyzed in duplicate. Insert the required number of microtiter well strips into a holder. Dilute the capture antibody with dilution buffer to a concentration of 2 Hg/ml. Coat each well with 100 Hl of capture antibody dilution. Incubate coated plate for 60 minutes. After incubation, aspirate the Capture Antibody solution from each well. Wash each well with W ash Solution as follows: 1. Fill each well with Wash Solution 2. Remove Wash Solution by aspiration 3. Repeat for a total of 3 washes. Add 200 Hl of Blocking Solution to each well. Inc u ba te 30 m in ut es . After incubation, remove the Blocking Solution and wash each well three times as in Step 1.E. Standards and Samples For research use only; not for diagnostic or therapeutic use. Page 2 C A. Dilute the calibrator to a concentration of 300 ng/ml. Do 1:3 serial dilutions (1 volume sample into 2 volumes diluent) down to a concentration of 0.412 ng/ml. Dilute the samples, based on the expected concentration of the analyte, to fall within the concentration range of the standards. Transfer 100 Hl of standard and sample to assigned wells. Inc u ba te pl a te 6 0 m in ut es . After incubation, remove samples and standards and wash each well 5 times as in Step 1.E. B. C. D. E. 4. HRP Detection Antibody A. Dilute the HRP conjugate in dilution buffer to a concentration of 50 ng/ml. (Note: Adjustments in dilution may be needed depending on substrate used, incubation time, and age of kit). B. Tr ansfer 100 Hl to each well. C. Inc u ba te 60 m in ut es . D. 5. Enzyme A. B. C. D. 6. After incubation, remove H RP Conjugate and wash each well 5 times as in Step 1.E. Substrate Reaction Prepare the substrate solution according to the manufacturer’s recommendation. Transfer 100 Hl of substrate solution to each well. Incubate plate so that the high calibrator is at an OD450 of 2-3. (Approximately 3-10 minutes depending on age of kit.) To stop the TMB reaction, apply 100 Hl of 2 M H2SO4to each well. If using another substrate, use the stop solution recommended by manufacturer. Plate Reading Using a microtiter plate reader, read the plate at the wavelength that is appropriate for the substrate used (450 nm for TMB). Calculation of Results 1. 2. 3. 4. Average the duplicate readings from each standard, control, and sample. Subtract the zero reading from each averaged value above. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. Other curve fits may also be used. A standard curve should be generated for each set of samples. See example below: Standards (ng/ml) Sample St01 Concentration 300 St02 100 St03 33.333 St04 11.111 St05 3.704 St06 1.235 BackCalcConc 200.562 267.358 485.07 105.378 90.365 122.977 31.805 32.608 35.212 10.555 11.334 11.143 3.597 3.836 3.928 1.112 Wells A1 A2 A3 B1 B2 B3 C1 C2 C3 D1 D2 D3 E1 E2 E3 F1 Values 2.869 2.9 2.943 2.763 2.728 2.794 2.354 2.366 2.402 1.705 1.751 1.74 1.07 1.102 1.114 0.661 For research use only; not for diagnostic or therapeutic use. MeanValue Std.Dev. 2.904 0.037 CV% 1.3 2.762 0.033 1.2 2.374 0.025 1.1 1.732 0.024 1.4 1.095 0.023 2.1 0.676 0.013 1.9 Page 3 St07 0.412 1.223 1.223 0.346 0.502 0.434 F2 F3 G1 G2 G3 0.683 0.683 0.498 0.533 0.518 0.516 0.018 3.4 Smallest standard value: 0.516 Largest standard value: 2.904 Technical Hints 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. The Capture antibody should be diluted with coating buffer immediately prior to its addition to the wells. Coated plates are stable overnight at 4ºC when covered. Change pipette tips between each addition of standard, sample and reagents to avoid crosscontamination. Standards and samples should be pipetted to the bottom of the wells and all other reagents should be added to the side of the wells to avoid contamination. Ensure that all buffers are not contaminated or expired. When troubleshooting ELISA results, it is recommended to prepare all new buffers in new vessels. Do not add Sodium Azide to any of the buffers. Sample and Conjugate dilutions should be made shortly before use. Wash buffer should be aspirated from wells, as pouring wash buffer from wells may cause crosscontamination. When preparing dilutions, wipe excess antibody/analyte from pipette tips to ensure accurate dilutions. Incubation time of the Enzyme Substrate will depend on the substrate used and the intensity of the color change. The high standard should have an O.D. reading of between 2 and 3. The low standard should have an O.D. reading above background. The Stopping solution should be added to the wells in the same order as the Enzyme Substrate. For research use only; not for diagnostic or therapeutic use. Page 4 Troubleshooting The following are some common problems encountered with the use of ELISA kits, and some of the causes of these problems. 1. Problem: Low absorbance Incorrect dilutions or pipetting errors. Improper incubation times Improper mixing of the TMB substrate. Each component is mixed in equal parts. Wrong filter on microtiter reader. Wavelength should be 450 nm for TMB, 490 nm for OPD, or 405 nm for ABTS. Kit materials or reagents are contaminated or expired. Incorrect reagents used. . Problem: High Absorbance Cross contamination from other samples or positive control. Incorrect dilutions or pipetting errors. Improper washing. Wrong filter on microtiter reader. Contaminated buffers or enzyme substrate. Improper incubation times. Kit materials or reagents are contaminated or expired. 3. Problem: Poor Duplicates Poor mixing of specimens. Incorrect dilutions or pipetting errors. Technical error. Inconsistency in following ELISA protocol. Inefficient washing. 4. Problem: All wells are positive Contaminated buffers or enzyme substrate. Incorrect dilutions or pipetting errors. Kit materials or reagents are contaminated or expired. Inefficient washing. 5. Problem: All wells are negative Procedure not followed correctly. Contaminated buffers or enzyme substrate. Contaminated conjugate. Kit materials or reagents are contaminated or expired. For research use only; not for diagnostic or therapeutic use. Page 5