This is new section 6, undulator performance. Redo of old section 5. 6.1 OK 6.2.1 line 13 ‘K=0.17’ Line 14 ‘0.34 photons/beam electron.’ Omit 3’rd line from end. Replace with ‘Due to losses from collimator and undulator misalignment E166 measures only a lower limit of 0.2 photons/electron for the undulator photon flux. 6.2.2 Photons from the undulator are measured by Silicon Tungsten detectors S1 and PRT whose absolute calibration is determined by GEANT simulation as described in section 3.4.2. The four PRT detectors are located immediately upstream at the face of the collimator C2 with edges tangent to the 3 mm diameter collimator aperture as shown in Figure X.(use photograph from old version) These detectors sample that portion of the flux that does not enter the collimator aperture. They also provide a measurement of the centering of the undulator beam. Photons exiting the 15 cm long collimator strike the conversion target producing positrons and are then measured by detector S1 prior to incidence on the Fe analyzing magnet where the photon transmission asymmetry is determined. With suitable correction the sum of these fluxes yields the number of photons per beam electron produced by the undulator. The angular and energy distribution of the undulator photons is described in Section 2. Combining this distribution with the measured vertical and horizontal emittance of the electron beam it is possible to simulate the photon distribution at the collimator face. Figure XX shows the simulated acceptance as a function of aperture diameter assuming that the photon beam is centered on the collimator so that 93% of the photons should enter the collimator. 5000 4000 3500 3000 2500 2000 1500 1000 500 0 0.97685 0.96042 0.94766 0.92975 0.89863 0.85147 0.77116 0.62685 0.36335 0 Collimator Acceptance Fraction of Photons into C2 Aperture 1.2 1 0.8 0.6 Series1 0.4 0.2 0 0 1000 2000 3000 4000 5000 6000 Collimator Aperture (microns) The conversion target removes an additional 7.3% of photons before they are counted by S1. The total detected photon flux is therefore given by sum of PRT and corrected S1 signal. However, throughout the data taking period it was necessary to steer the electron beam so that it passed cleanly through the undulator to minimize background noise in the CsI detector caused by beam tails striking the undulator protection collimator C1. The steering resulted in a small displacement of the photon beam from the aperture center resulting in reduced transmission through C2. Displacement is manifested by inequality in the PRT counter signals. As an example Figure YY shows the left right and up down variation for 420 runs at spectrometer setting 100 amperes Series 1 up/down Series 2 left/right PRT Assymetry 4 3.5 3 2.5 Series1 Series2 2 1.5 1 0.5 0 0 50 100 150 200 250 Run Sequence Number 300 350 400 450 The beam displacement is primarily in the x direction and from simulations corresponds to displacements of from 500 to 1100 microns at the collimator. The average y displacement is less than 100 microns. The correspondence with S1 and total PRT signal is shown in Figure ZZ . Series 1 Collimator Aperture Series 2 Stopped by Collimator Series 3 Total Undulator Photons per Beam Electron for 420 Runs at Spectrometer Setting 100 Amperes 0.25 0.2 0.15 Series1 Series2 Series3 0.1 0.05 0 0 50 100 150 200 250 300 350 400 450 Run Sequence Number As expected, increasing left right ratio leads to reduction in S1 signal and a corresponding increase in PRT signal with, however, the sum remaining fairly constant at about 0.2 photons per electron about 60% of the expected undulator performance. Averaged over the entire experiment we find the undulator flux to be 0.199 ± 0.008 photons per beam electron. We attribute this discrepancy to imperfect collimator alignment. While the collimator aperture diameter was chosen sufficiently large to accommodate collimator tilt for a centered photon beam, displacement of the beam toward the aperture edge causes serious loss of transmission due to tilt. Simulation shows that a misalignment of 3.3 milliradians results in a 3% loss if the beam is centered but is 47% for beam displaced by 1000 microns. Therefore, we know that the E166 undulator production is greater than 0.2 photons per beam electron from the E166 undulator but we are unable to directly confirm the design value of 0.34. Last paragraph line 3 ‘of the photon flux reaching the conversion target by nearly 30% from 0.061 to 0.079 photons/electron. (see figure nn) Photons/beam electron S1 photon flux.Spectrometer setting 180 amperes. Ferrofluid introduced at run 132 0.1 0.09 0.08 0.07 0.06 0.05 0.04 0.03 0.02 Series1 0.01 0 0 50 100 150 200 250 300 350 Sequence number Photns/beam Electron S1,PRT, Total 6.2.3 Intensity ……. Use as is. Add at the end. ‘The K factor is indeed approximately linear but the absolute magnitude of K reflects the fact that we measure only a lower limit for the undulator flux.’ 0.25 0.2 0.15 Series1 Series2 Series3 0.1 0.05 0 0 500 1000 1500 -0.05 Undulator Current 2000 2500 K value. Upper curve total flux, Lower curve photons through collimator Dependence of Undulator K value on Current 0.14 0.12 0.1 0.08 0.06 0.04 0.02 0 0 500 1000 1500 2000 2500 Undulator Current 6.2.4 Undulator Steering Wording OK. Omit last line Beam deflection X,Y BPM 6130 DEFLECTION (MICRONS) 5 0 -5 0 500 1000 1500 2000 2500 -10 Series1 Series2 -15 -20 -25 -30 -35 Undulator current