CHARACTERISATION OF AN X-RAY IMAGING SETUP FOR DIGITAL MAMMOGRAPHY BASED ON A SILICON MICROSTRIP DETECTOR Tadej Mali1, Vladimir Cindro1, 2, Marko Mikuž1, 2, Dejan Križaj3, Danilo Vrtačnik3, Urban Zdešar4 1 Jožef Stefan Institute, Ljubljana, Slovenia 2 Faculty of Physics and Mathematics, University of Ljubljana, Slovenia 3 Faculty of Electrical Engineering, University of Ljubljana, Slovenia 4 Institute of Occupational Safety, Ljubljana, Slovenia Table of contents Introduction Materials and Methods Setup Modulation transfer function (MTF) measurement Noise power spectrum Detective quantum efficiency Results Summary Abstract: A silicon microstrip detector was designed and fabricated. "Edge-on" geometry was used to assemble it into a slit-scanning X-ray imaging system, resulting in a linear pixel detector with a pixel size of 0.100×0.220 mm2. Single photon counting was realized using the CASTOR 1.0 ASIC. The line response function in the direction of scanning was measured and the corresponding modulation transfer function (MTF) calculated. Noise power spectrum was obtained and the detective quantum efficiency (DQE) evaluated. Introduction Several approaches to digital image acquisition are under consideration. The first approach utilises integration mode detectors, e. g. photostimulable phosphorus, amorphous silicon and scintillator-coupled CCD (overview given in Säbel and Aichinger 1996, Yaffe 1998). There the charge generated by impinging radiation is integrated over exposure time. Another approach is to use detectors operating in single photon counting mode, with individual photons detected and counted. There are several groups implementing this technique either using silicon microstrip detectors (Arfelli et al. 1998, Beuville et al. 1998, Mali et al. 1999a, 1999b) or gallium arsenide pixel detectors (Amendolia et al. 1997). The advantage of single photon counting systems compared to integrating ones is that using appropriate readout electronics, the acquisition can be practically free of detector and electronic noise. The only source of noise in the image are fluctuations in the number of impinging photons with a Poissonian distribution. The response of silicon detectors is linear over an effectively infinite dynamic range (Turchetta et al. 1998). Together with high efficiency of semiconductor detectors, this contributes to good sensitivity of such detectors as well as to a reduction of dose. Materials and methods Absorption length of 20 keV photons in silicon is approx. 1 mm. Therefore 3 mm of silicon are sufficient to absorb 95 % of radiation. Since silicon microstrip detectors are produced on thin (several hundred microns) wafers, they must be used in an "edge-on" geometry (Arfelli et al. 1998, Beuville et al. 1998, Mali et al. 1999a, 1999b), where photons hit the detector from the side and the photons are absorbed along the whole length of the strip (figure 1). In this way only the dead volume between the edge and the active volume limits the quantum efficiency (QE). The detector acts as a linear pixel detector with pixel dimensions given by the strip pitch and the detector thickness. Images of large objects are obtained by scanning in the direction perpendicular to the detector plane. This approach is called scanning slit radiography or slot scanned radiography and has the additional benefit of efficiently suppressing scattered radiation detection (Yaffe 1993, Barnes et al. 1993, Stres et al. 2000). Figure 1: Silicon microstrip detector used in the ‘edge-on’ geometry. X-ray photons emerge from the source on the right hand side of the figure. Setup The detector with a strip pitch of 0.100 mm was fabricated on 0.220 mm thick, high resistivity silicon, resulting in a pixel size of 0.100×0.220 mm2. Details of design and fabrication are described elsewhere (Vrtačnik et al. 1999). A total of 32 strips were wire bonded to the CASTOR 1.0 readout integrated circuit, covering a total area of 3.2×0.22 mm2. The readout circuit CASTOR 1.0 is a multichannel, mixed analog-digital, low-noise integrated circuit. It enables simultaneous photon detection and counting without false photon hits for photon energies above 12 keV (Comes et al. 1996). Phantom and test objects were positioned on a table moved by a stepping motor, while the position of the detector and source were kept fixed. The X-ray source was an X-ray tube with a tungsten anode operated at single phase 32 kV peak voltage, the tube current was 5 mA. The first half value of aluminum was measured to be 0.8 mm. The source to detector distance was 60 cm. Modulation transfer function (MTF) measurement Modulation transfer function characterizes the spatial resolution of the system. For an ideal pixel detector with dimensions x y one expects: sin xu sin yv MTF u, v xu yv where u and v are spatial frequencies along x and y-axis. Since MTF falls off inversely proportional to pixel dimensions, the spatial resolution increases as the pixel size is reduced. In the case of silicon microstrip detectors, the pixel size can be reduced by smaller strip pitch (denoted y), while the pixel size in the direction of scanning (denoted x) is limited by wafer thickness to approx. 0.200 mm. Therefore our research is focused into possible improvements of the spatial resolution in the direction of scanning (Mali et al. 1999b). Since only the MTF in the direction of scanning is investigated, it can be obtained from the line spread function (LSF), which was obtained as the derivative of measured edge response function (ERF). A thin (0.05 mm) lead foil, embedded into plastic was used to obtain the ERF. Noise power spectrum (NPS) The noise power spectrum can be calculated analytically. Suppose that on average N photons are detected in a pixel at intervals of x (scanning step x needs not be equal to pixel width x ). The noise in the image is a random variable x , describing fluctuations in the number of detected photons N / x at each scan position x. It has zero mean and standard deviation equal to the square root of the number of detected photons per scan: N / x . The space function of measured quantum fluctuations can be represented as a piecewise constant function over an interval of x . The values in neigbouring intervals are totally uncorrelated, therefore the autocorrelation function (ACF) of the noise equals: N x , if x x 1 X /2 2 1 ACF x lim x x x dx x x X X X / 2 0, elsewhere To obtain the NPS, the Fourier transform of the ACF must be calculated. It should be stressed that in digital systems the image is sampled at discrete points separated by x . According to the sampling theorem, the noise components with frequency above 1 / 2x will be mapped into lower frequencies interval. The discrete Fourier transform of the above ACF is: NPS (u ) ACR(nx) exp inxux n The result is NPS u N / x , since only ACF(0) is different from 0. This result implies that white noise is expected in single photon counting systems. Detective quantum efficiency (DQE) DQE is defined as: QE 2 MTF 2 N in / x DQE NPS where QE equals quantum efficiency of the detector with the number of incident photons N in connected to the number of detected photons via N QE N in . If the image noise follows the predicted constant power spectrum, DQE differs from MTF 2 only by the factor of QE. Note that in digital systems DQE vanishes at frequencies larger than Nyquist frequency, because the high frequency signal components are aliased to into the low frequency interval. Results To accurately measure the MTF, it is necessary to reduce the sampling step to such extent, that the MTF is negligible outside the Nyquist interval. In our study the scanning step was set to 0.025 mm, resulting in Nyquist frequency of 20 cycles/mm. The measured LSF of the system has a full width at half maximum of 0.227 mm, which is close to the pixel dimensions in the direction of scanning (0.220 mm). The corresponding MTF is shown in figure 2, together with a MTF of an ideal pixel detector with a pixel size of 0.220 mm. It can be seen, that the system MTF is lower than the ideal one. Effectively the measured MTF can be described as a product of the ideal MTF and a Gaussian curve with 4.6 cycles/mm . This corresponds to a LSF of an ideal pixel detector blurred with a Gausian with 0.03 mm . It has been shown (Mali et al. 1999b) that X-ray test patterns are stil visible at frequencies, where the MTF drops to 0.01. In addition, using appropriate image processing, the distortions due to LSF could also be restored up to frequencies where MTF=0.01 (Mali et al. 1999b). Therefore our conclusion is, that the current system is capable of showing details with spatial frequencies above 10 cycles/mm, with some additional room for improvement. Figure 2: The modulation transfer function of the system. In the contrast to MTF measurements, in the case of NPS the scanning step can not be reduced to such extent as to prevent aliasing, since the ACF depends on the sampling rate x and its Fourier transform is always non-negligible outside the Nyquist interval. To evaluate the NPS a 1024 steps long flood image was taken. It was corrected for gain and tube current variations. After these corrections the mean and standard deviation of pixel values were estimated. The Poissonian relation between mean and standard deviation was confirmed. Then the NPS was measured in the following way. The 1024 samplings of each of the 30 pixels were Fourier transformed (using the fast Fourier transform algorithm) and the transforms were averaged over the pixels. In the resulting spectrum 8 neighbouring bins were averaged. The resulting error in NPS estimate is then 6.5%. Within this error the NPS is indeed constant over the whole frequency interval from 0 to Nyquist frequency (figure 3). Figure 3. Noise power spectrum of the system. Average number of detected photons per pixel is approx. 4100. Due to a frequency independent NPS, the DQE of the system is simply OE MTF 2 as explained previously. Assuming a 0.200 mm thick dead region, the resulting QE at 20 keV amounts to 0.80. Summary A silicon microstrip detector was designed, fabricated and assembled into a X-ray imaging system. The system was characterized by means of MTF, NPS and DQE. Comparison to previous results (Mali et al. 1999b) shows, that the system is capable of resolving details with spatial frequency above 10 cycles/mm, when an appropriate scanning step is chosen and image processing applied. References Amendolia, S. R., E. Bertolucci, U. Bottigli, M. A. Ciocci, A. Cola, M. Conti, P. Delogu, M. E. Fantacci, G. Magistrati, E. Pernigotti, N. Romeo, P. Russo, A. Stefanini and S. Stumbo. 1997. A Project for Digital Mammography based on a GaAs Pixel Detector and on a SelfTriggering Single Photon Counting Acquisition System. Physica Medica. 13:157-165. Arfelli, F., V. Bonvicini, A. Bravin, G. Cantatore, E. Castelli, L. Dalla Palma, M. Di Michiel, R. Longo, A. Olivo, S. Pani, D. Pontoni, P. Poropat, M. Prest, A. Rashevsky, G. Tromba and A. Vacchi. 1998. Mammography of a phantom and breast tissue with synchrotron radiation and a linear-array detector. Radiology 208:709-715. Barnes, G. T., X. Wu and A. J. Wagner. 1993. Scanning slit mammography. Med. Prog. Technol. 19:7-12 Beuville, E., B. Cederström, M. Danielsson, L. Luo, D. Nygren, E. Oltman and J. Vestlund. 1998. High resolution X-ray imaging using a silicon strip detector. IEEE Trans. Nucl. Sci. 46:3059-3063. Comes, G., F. Loddo, Y. Hu, J. Kaplon, F. Ly, R. Turchetta. 1996. CASTOR a VLSI CMOS Mixed Analog-Digital Circuit for Low Noise Multichannel Counting Applications, Nucl. Instr. Meth. A. 377:440-445. Mali, T., V. Cindro, M. Mikuž M, U. Zdešar and B. Jančar. 1999. Evaluation of silicon microstrip detectors as X-ray sensors in digital mammography. Radiol. Oncol. 33:237-244. Mali, T., V. Cindro and M. Mikuž. 1999. Silicon microstrip detectors for digital mammography - evaluation and spatial resolution study. Presented at 1st Int. Workshop on Radiation Imaging Detectors, Sundsvall, Sweeden; to appear in Nucl. Instr. Meth. A. Turchetta, R., Y. Hu, Y. Zinzius, C. Colledani and A. Loge. 1998. A Low-power low-noise mixed mode VLSI ASIC for infinite dynamic range imaging applications. Proc. of the SPIE's Int. Symp. on Optical Science, Engineering and Instrumentation. (San Diego) 3345:318-328 Säbel, M. and H. Aichinger. 1996. Recent developments in breast imaging. Phys. Med. Biol. 41:315-368. Stres, Š., T. Mali, M. Mikuž and V. Cindro. 2000. Design optimization of a breast imaging system based on silicon microstrip detectors. Accepted for publication in Phys. Med. Biol. Vrtačnik, D., D. Križaj, T. Mali, D. Resnik, U. Aljančič and S. Amon. 1999. Side-illuminated 100µm pitch X-ray detector for digital radiology. Accepted for publication in Sensors and Actuators A Yaffe M. J. 1993. Direct digital mammography using a scanned-slot CCD imaging system. Med. Prog. Technol. 19:13-21. Yaffe M. J. 1998. Development of full field digital mammography. Digital Mammography, Proc. 4th Int. Workshop on Digital Mammography (Nijmegen) ed. N. Karssemeijer, M. Thijssen et al. (Dordrecht: Kluwer Academic Publishers) 3-10.