Predicting dose uncertainty in moving organs at risk Joep Stroom*, Joos Lebesque*, Marcel van Herk*, and Ben Heijmen Radiotherapy Division, The Netherlands Cancer Institute - Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands. Department of Radiation Oncology, Erasmus MC – Daniel den Hoed Cancer Center, Rotterdam, The Netherlands. * Abstract Since the use of safety margins for geometrical uncertainties appears inappropriate for organs at risk (OARs), two methods are proposed to quickly determine the effect of systematic errors on the DVHs of OARs. Special DVHs are constructed in two different ways, which indicate the confidence intervals (CIs) on the volumes for each dose bin in the DVHs. The CIs determined by the new approach were validated with Monte Carlo simulations for the rectum in 40 prostate plans. Keywords Organs At Risk (OAR), geometrical uncertainties, confidence interval, systematic errors, dose volume histogram (DVH) Introduction The geometrical uncertainties in the position of the clinical target volume (CTV) during radiotherapy are normally taken into account by adding a margin to this volume during planning, hence yielding the planning target volume (PTV) [1]. The margins should be such that the planned dose in the PTV is representative for the dose in the “moving” CTV. PTV-margin rules have been developed, which guarantee a low probability of CTV underdosage [2,3]. For organs at risk (OARs), a similar approach has been proposed [4,5]. However, since dose evaluation for OARs depends normally on the percentage of volume that receives a specific dose, margins appear less feasible; margin addition normally increases the volume of the OAR significantly and the dose in the expanded volume hence cannot be made representative for the moving OAR. Therefore, other ways of evaluating (during planning) the effect of geometrical uncertainties on the dose in OARs must be procured. The most straightforward way is to sample translations and rotations from their respective error distributions, moving the OAR accordingly, calculate the dose, and display the distribution of possible results [6,7]. In Fig. 1 a volume(%) 100 0 CTV rectum dose (%) 100 Figure 1: Example of distribution of possible DVHs due to systematic errors for a rectum and CTV in a three-field prostate plan. Thin black lines are original DVHs, grey lines are 100 sampled curves, and black dotted curves the mean. The PTV margin ensures adequate CTV coverage for alle curves, wheras the spread in rectum DVHs is obvious. possible distribution of DVHs is shown as a result of geometrical errors. However, many samples are required to obtain a reliable result, which still requires too much time to be used in practice, especially if this must be incorporated in inverse planning. Therefore, in this paper two methods to calculate the spread of possible results due to geometrical errors will be proposed and compared. At the moment, only systematic errors are considered, which are of 3 to 4 times greater importance than the random errors [2,3]. The systematic errors normally consist of organ motion and set-up error and are in this abstract represented by a normal distribution with standard deviation ; non-isotropic error distributions, deformations, and rotations are ignored in this abstract. Material and methods The coverage probability (CP) method Previously, a method to quickly evaluate the dose in moving CTVs was proposed [2]. It appeared that convolution of the CTV with the distribution of systematic errors and taking the DVH of this convolved, smeared-out, CTV, yielded the mean DVH (mDVH, i.e. the mean volume in each dose bin) for the possible systematic errors (Fig. 2a). The convolved volume was called the coverage probability (CP) volume because the voxel values indicate the probability that the particular voxel is covered by the moving CTV. The mDVH was obtained by weighing each CTV voxel by its CP-value when filling the different dose bins. Since the high dose region is normally planned around the CTV, a CTV motion can only decrease the CTV dose and the mDVH will give an indication of the average underdosage. An OAR is normally not (completely) in the high dose region. Motion of an OAR can therefore increase as well as decrease the OAR dose. Consequently, the mDVH can be almost equal to the original static DVH because individual DVHs cancel out; no information about the spread of DVHs is obtained (see Fig. 1). Therefore, the CP-method is extended. Instead of using the whole error distribution in the convolution, parts of the distribution in different directions can be used. The CTV/OAR -distribution CP-matrix (m)DVHs a) CTV OAR The limited sampling (LS) method b) ventral dorsal c) Figure 2: demonstration of the CP-method for the prostate case from Fig. 1. In a) the volume is convolved with the distribution of systematic errors yielding mDVHs (dotted) for CTV and rectum. Original DVHs are solid. In b) only the ventral cone of the distribution is used yielding a spread in the rectum curves (the dorsal mDVH curve is also indicated). In c) only the outer 10% rim of this cone is used, yielding more extreme mDVHs. DVHs subsequently acquired are the mDVHs for that subgroup of possible systematic motions (see Fig. 2b,c). To get an indication of the spread for all dose levels, subgroup mDVHs in different directions can be combined into two encompassing DVHs (eDVHs, i.e. the minimum and maximum volume for each dose bin in the multiple mDVHs). For instance, the ventral and dorsal mDVHs in Fig. 3 only reflect the DVH spread for the higher dose levels. Variation at the lower levels is due to cranio-caudal translations. The eDVHs in this abstract were constructed from the mDVHs of the cones in the 6 orthogonal directions and of the 6 cones rotated 45 degrees around the lateral axis. The eDVHs can be used to get an indication of a confidence interval (CI) on volumes for each dose bin in the DVHs. Crudely, a mDVH representing a subgroup of motions has about half its curves above and half its curves below the mean. If the different subgroups used to compose the eDVH each 100 cranial caudal volume(%) represented X % of the total distribution, the CI represented by the eDVHs should then be about 100-X %. Taking a smaller part of the cones varies the size of the expected CI (Fig. 2b,c). Instead of the CP-method, an alternative way to determine eDVHs that can represent CIs is to sample translations at specific points in the distribution, move the OAR, and calculate the DVHs. For instance, a translation T of an OAR equal to 1 implies that about 16% of possible systematic translations are larger than T (i.e. conform a 1D normal distribution). Assuming that the dose from a point in the OAR to the high dose region increases continuously, a larger translation in that direction will cause a higher dose in the OAR. So, if there is one dominant direction towards a specific high-dose isosurface D, the volume with at least dose D (V D) corresponding to T=1 then should have about 16% of possible DVHs above it. The DVH curve corresponding to T=-1 (i.e. in the opposite direction) will consequently have about 16% of possible curves below it. Now, to cover the 3D translation space effectively, translations in a limited number of directions will yield multiple DVHs, which can again be used to construct the eDVHs. The CI associated with these eDVHs depends on the size of the selected translations T. For T=1, CI=68%, for T=2, CI=95%, etc. The eDVHs used in this study were constructed from 14 isotropically distributed translations; the 6 orthogonals plus the diagonals in all the octants. For example, the eDVHs corresponding to the 68% CI were constructed from 14 DVHs with translations ,0,0; 0,,0; 0,0,; 1/3,1/31/3; and all possible combinations with –. Verification of the methods with Monte Carlo As a first OAR, the rectum in 3-field prostate plans was considered for 40 patients. The systematic errors were assumed to be 3.5 mm. The various CIs as predicted by the two methods were compared with Monte Carlo (MC) simulations of possible translations, each consisting of 200 samples. To verify the predicted CIs, the number of sampled curves above and below the eDVHs as determined by the two methods was counted for dose levels varying from about 30 to 80 Gy ( Dmax). Subsequently, the average (over the dose) difference between the predicted and simulated CIs was calculated. ventral Results The first results for the LS and CP-method are summarized in dorsal 0 dose (%) 100 Figure 3: indication that the CP-method using subgroups can indicate DVH spread due to systematic errors. Original static DVH in black, 100 MC simulated curves in grey, mDVHs of ventral and dorsal cones dashed, and mDVHs of cranial and caudal cones dotted. Each cone represents about 17% of the possible systematic errors. Hence, the encompassing curves contain about 83% of curves for each dose level. T () 2.5 2 1.5 1 0.5 CI (%) 98.8 95.4 86.6 68.3 38.3 (%) -0.3 -0.7 -0.6 -0.3 -0.5 SD (%) 0.8 1.4 2.4 3.5 3.5 Table 1: Summary of the results for rectum motion of 40 prostate patients with the LS method. T is the translation used to determine the confidence interval CI, the mean (of 40 cases) difference between simulated and expected CI, and SD the standard deviation of the differences. F (%) 10 25 50 75 100 (%) 0.2 0.4 -0.1 0.2 0.4 CI (%) 99.1 97.1 93.0 88.5 82.0 SD (%) 0.7 1.0 1.7 2.3 2.5 Table 2: Summary of the results for rectum motion of 40 prostate patients with the CP method. F is the (outer) fraction of the cones that were taken, for the rest see Table 1. Tables 1 and 2, respectively. Various eDVHs have been calculated (representing different CIs) by varying the translation T or outer fraction of the cones F, respectively. For both methods, the average difference (over 40 cases) between predicted and simulated CIs are smaller than 1%. The results for the LS-methods are graphically shown in Fig. 4. The deviations that are seen are partly due to the limited number of simulations performed and to the low doses used in the averaging; contrary to the lower doses, the higher doses are normally located at one spot (around the CTV) so that there is indeed one dominant direction that determines the spread. 100 be adjusted based on the patient geometry, i.e., the position and shape of the CTV (or high dose region) with respect to the OAR. However, initial experience for the spinal cord in lung and head and neck plans also appears to yield satisfying results with the current settings. The eDVHs indicate the CI for each dose bin and can therefore be used to evaluate the plan. On the other hand, perhaps not so much the CI for a certain DVH is clinically important, but the probability that the possible DVHs lay above a critical DVH point. Only the outer eDVH is required for this. To use these new methods however, new, probability-based criteria must be developed. Conclusion A method to quickly calculate the effect of systematic errors on the DVH for a mobile organ at risk has been suggested. The spread of possible outcomes can be predicted by construction of special DVHs, which indicate the confidence intervals for each dose bin in the DVHs. First results for the rectum are encouraging and other cases will be studied in the future. T = 2.5 T = 2 References T = 1.5 [1] International Commission on Radiation Units and Measurements. 1993 ICRU report 50: Prescribing, recording, and reporting photon beam therapy, ICRU Publications, Bethesda MD. 90 CI (%) 80 70 T = 1 60 50 40 T = 0.5 30 1 4 7 10 13 16 19 22 25 patient nr 28 31 34 37 40 Figure 4: the predicted (straight lines) and simulated CIs for moving rectums = 3.5 mm) of 40 prostate patients. Different CIs have been calculated by varying the size of the translations used to determine the eDVHs. Discussion The differences between the new approach and the simulations will decrease by increasing the number MC samples and the number of mDVHs to construct the eDVHs. On the other hand, the optimal directions to predict the spread of the DVH curves will depend on the dose distribution and on the position and shape of the OAR. This direction might be automatically obtained from the treatment geometry, which will reduce the number of (m)DVHs required to construct a reliable eDVH. For the case studied so far, both methods are able to predict the spread of possible treatment outcomes quite accurately. So both might be used as an alternative to MC-simulations, which yields significant time benefit. The difference between the two methods is that the CP method averages multiple DVHs so that the most extreme translations are included. The LS method consists of single DVHs and assumes continuous dose increment or decrement in one direction. It remains therefore to be seen how both methods will perform for other dose distributions (e.g. IMRT) and OARs (e.g. lung). Possibly, the expected CI associated with values for T and F might have to [2] Stroom JC, de Boer HC, Huizenga H, Visser AG. 1999 Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys. 43 905-19. [3] van Herk M, Remeijer P, Rasch C, Lebesque JV. 2000 The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 47 1121-35. [4] International Commission on Radiation Units and Measurements. 1999 ICRU Report 62: Prescribing, recording, and reporting photon beam therapy (Supplement to ICRU Report 50), ICRU Publications, Bethesda, MD. [5] McKenzie A, van Herk M, Mijnheer B. 2002 Margins for geometric uncertainty around organs at risk in radiotherapy. Radiother Oncol. 62 299-307. [6] Killoran JH, Kooy HM, Gladstone DJ, Welte FJ, Beard CJ. 1997 A numerical simulation of organ motion and daily setup uncertainties: implications for radiation therapy. Int J Radiat Oncol Biol Phys. 37 213-21. [7] Mageras GS, Kutcher GJ, Leibel SA, Zelefsky MJ, Melian E, Mohan R, Fuks Z. 1996 A method of incorporating organ motion uncertainties into three-dimensional conformal treatment plans. Int J Radiat Oncol Biol Phys. 35 333-42.