Ge 114 - Optical Mineralogy

advertisement
Ge 114 - Optical Mineralogy -Laboratory Exercise 1
INTRODUCTION TO THE POLARIZING MICROSCOPE
Reading: Phillips (1971) Ch. 2. or Nesse (1991) Ch. 2
Phillips (1971) Ch. 3 or Nesse (1991) Ch. 3
Nikon Polarizing Microscope LABOPHOT-POL instructions
Or Leica Microscope instructions
The optical microscope is one of the first tools which is used in the serious
examination of rocks and minerals. These microscopes are designed to study materials
with properties which vary with direction. The features which distinguish them from
"standard" biological microscopes are the use of linearly polarized light which defines
the direction of vibration of the light incident upon the sample, rotating stages which
allow the sample to be freely positioned with respect to the direction of vibrations of the
light, the use of special strain-free objectives which do not distort the polarization
properties of the light, and a variety of special accessories which are used in conjunction
with the polarized light.
The following instructions will guide you through the steps required to
familiarize yourself with the use of the microscope. All questions or procedures in
bold type should be answered and recorded in the lab report.
A) Introduction to the Petrographic Microscope
1. Identify the parts of the microscope and accessory equipment for the
microscope, using p.1-14 in the Nikon manual and Ch. 2 in Nesse. This should include
standard oculars (one with cross hairs), four objectives (mounted on turret nosepiece)
and a combined 1/4  and tint (530nm=1) plate. Check and report any deficiencies or
indication of damage to any of this equipment.
2. Set up the microscope and adjust its position for maximum comfort in
operation (p. 8 Nikon manual). The light sources should already be factory-adjusted for
optimum illumination.
3. Practice focusing on a glass slide. Start with the lowest (5x) objective. Check
the coarse and fine adjustments of the microscope focusing mechanism. Read p.21-22
Nesse concerning the proper technique for focusing with various objectives to avoid
running an objective lens into a thin section! Change objectives by gently turning the
rotating nosepiece. Do NOT grasp the objective lenses- you might misalign and
damage them.
4. If necessary, carefully center all objectives (p. 8 Nikon manual; p. 42 Phillips,
1971). Learn to center the stage if all objectives are misaligned.
February 12, 2016
1
5. Calibrate the ocular micrometer scale: Using a stage micrometer, determine
for each objective the length represented by 1 scale division of a micrometer ocular.
Make a table with the magnification of each objective lens and the corresponding
length per division on the ocular scale. Also determine and include in this table the
length represented by the diameter of the field of view for each ocular. This
information will help later for quick estimates of grain size in mineral powders and in
thin sections.
6. Check the centering of the substage diaphragms (including the light source
diaphragm) of your microscope (p. 9 Nikon manual). For each objective, note what
effect is produced by closing down each diaphragm. Make the tests with condenser
assembly at the top, and then at the bottom of its travel, for the low- and mediumpower objectives. For the high-power objective, carry out these tests both with and
without the movable substage condenser in the optic train. Realize that in general it is
not possible to have all elements perfectly centered under all possible combinations.
7. Verify that the vibration direction of the analyzer (upper polarizer) is set
normal to that of the polarizer (see p.19 Nesse). To do this, insert the upper polarizer.
When the vibration direction of the upper polarizer is perpendicular to that of the lower
polarizer, the field of view is dark: this is called crossed polars. When the upper polar
is parallel to the lower polar, light is transmitted and the field is light. The upper
polarizer may be rotated by loosening a screw on the left side of the polar and rotating
the numbered ring. The default setting is with the “0”’s aligned. You may rotate the
lower polar slightly to maximize extinction in the field of view.
8. Check the (lower) polarizer setting using a slide of biotite. In biotite the
greatest absorption occurs when the trace of the basal cleavage is parallel to the
vibration direction of the polarizer. Why? (see p.44 Phillips, 1971). Record whether
the polarization direction is NS (north/south=up to down) or EW (east/west= left to
right). Compare your result to Fig. 20 on page 13 of the Nikon manual.
B. Relief and Becke Lines
References: Phillips (1971) p.48-50 or Nesse (1991) pp. 25-28
When grains of a mineral are mounted in a medium of a different refractive index, the
apparent contrast between the grain and the mounting medium depends upon the difference
between their indices of refraction. This apparent contrast, commonly called "relief", comes
largely from irregularities in the surface of the mineral grain. When the relief is high, the
mineral grain appears to have a "bumpy" surface. We want to become familiar with the
appearance of minerals of different indices of refraction in a medium of fixed index.
February 12, 2016
2
To familiarize yourself with the Becke line which is commonly used to compare the index of
refraction of adjacent mineral grains, first examine microscope slide Ge114 2-2 (see sketch)
which contains grains of two different minerals, one with an index of refraction lower than
the mounting medium (fluorite) and the other with a higher index (garnet). Make sure you
place this and all other slides on the microsope stage with the cover slip side up. Under
intermediate magnification, examine the grains with the lower index (near the black ink
label). Lower the stage of the microscope to bring the sample out of focus and observe the
Becke line move out of the grain into the mounting medium. Now raise the stage to bring the
grain in focus and observe the Becke line come back into the grain.
Repeat this sequence with the grains of the mineral with the higher index. You should
see the Becke line move into the sample as the sample stage is lowered. Read about Becke
lines in the references above and write a short explanation of this phenomena.
Ge 114 2-2
Becke Line
Mineral grains with
index of refraction
less than the mount
Mineral grains with
index of refraction
greater than the mount
Examine one of the special slides illustrating the relief scale. The slides contain
examples of the range of relief effects observable for mineral grains imbedded in Canada
balsam (n=1.54), from moderately negative (fluorite) to very high positive (garnet). Try to
quantitatively calibrate your sense of mineral refractive index in relation to the relief shown.
Make a table of the behavior of the Becke line and the index of refraction for each
mineral in the slide.
February 12, 2016
3