A Simplified Gamma-Ray Self-Attenuation Correction in Bulk Samples A. E. M. Khater,1,2* Y. Y. Ebaid,3 1 2 National Centre for Nuclear Safety and Radiation Control, Atomic Energy Authority, Egypt Physics Department, College of Sciences, King Saud University, Riyadh, Kingdom of Saudi Arabia 3 Physics Department, Faculty of Science, Fayum University, Fayum, Egypt Abstract Gamma-ray spectrometry is a very powerful tool for radioactivity measurements. For accurate gamma-ray spectrometry, certain correction measures should be considered, for instance, uncertainty in the photo-peak efficiency due to the differences between the matrix (density and chemical composition) of the reference and the other bulk samples. Therefore, gamma-ray attenuation correction factors are of major concern for precise gamma-ray spectrometry. Simple practical correction for the photo-peak efficiency, due to discrepancies in both the samples matrices and densities (self-attenuation), is performed in this study. This study suggests a brief measurement of relative photons transmission through both reference and unknown bulk samples where the variations of photon transmissions are linearly correlated to the samples’ densities. Specific correction factors would be produced for each analyzed sample to be considered when their * Corresponding author present address; P.O. Box 2455, 11451 Riyahd, Kingdom of Saudi Arabia. Email: khater_ashraf@yahoo.com (A. Khater); yebaid@yahoo.com (Y. Ebaid) 1 activities are calculated. Practically, the suggested method was verified and succeeded in improving the obtained results. Keywords: Gamma-ray spectrometry; attenuation correction; bulk sample Introduction: Gamma-ray spectrometry using hyper pure germanium (HpGe) detectors has been an essential and principal spectroscopy technique in almost all radioactivity measurements laboratories worldwide. Its major advantages are being non-destructive, multi-elements analysis, simplified regarding sample preparation, i.e. mostly no need for any chemical separation processes, and its applicability for all types of samples, etc. Low energy gamma emitters are used in many scientific studies in a wide range of applications. For example, 210Pb (46.5 keV) is used for determining sedimentation rate in lakes, estuaries and the coastal marine environment during the last century. Also it is used to study the atmospheric fluxes and mixing in the troposphere. Additionally 234 Th (63.3, 92.4 and 92.8 keV) is extensively used as a natural tracer to study biological mixing and particle scavenging processes in recent sediment (San Miguela et al., 2004). However, the precise determination of the activity concentration of each radionuclide requires the determination of full energy efficiency calibration for a given geometry. Therefore, a detection efficiency curve, known as efficiency calibration, over the energy region of interest must be established precisely in advance. The detection efficiency at certain gamma-ray energy and sample geometry is given by; ( E , n) C ( E , n) t. f ( E , n). A 2 Where: C(E,n): net photo-peak count of gamma-ray transition with energy E of radionuclide n, t: counting time, s, f(E,n): branching ratio, number of photon with energy E per hundred disintegration of radionuclide n, A: activity concentration in Bq of radionuclide n. The detection efficiency curve depends not only on a detection system but also on both the sample shape and matrix (Saegusa et al., 2004). Efficiency calibration can be performed theoretically using Monte Carlo computation techniques or semi-empirically using analytical method. Experimentally, the detection efficiency curve can be performed using standard samples that contain a set of radionuclides with known activities and cover the gamma-ray energy range of interest (usually from 35 to 2000 keV). Standard or reference samples should have the closest specifications, regarding geometry and matrices (apparently density and composition), to the analyzed samples (Abbas et al., 2006; Gurriaran etal, 2004; Vargas et al., 2006). The accurate determination of the photopeak efficiency curve for a given sample matrix represent the main challenge in gamma-ray spectrometry (Quindos, 2006). Practically, the samples geometry (shape and sample-detector geometry) can be easily reproduced. However, the major source of possible error in efficiency calibration remains due to the difficulties to reproduce the same matrix and chemical composition of the standard samples as that of the other bulk samples that could have vast verities in both densities and compositions. 3 An important correction applied in gamma-ray spectrometry of bulk samples is the correction for photon attenuation within the source material itself that known as selfattenuation or absorption correction (Boshkova, 2003). For a given geometric setup, the correction factor is expressed as the ratio of efficiency of standard to that of the sample: CS ( E ) ( E , S tan dard ) ( E , sample) In practical applications, standard and sample materials might be entirely different. Environmental samples may greatly differ in their chemical composition, even of the same matrices (such as soil and sediment samples), and their density ranging from extremely low densities up to 2.0 g/cm3. Self-absorption correction can be determined experimentally or using the Monte Carlo computation techniques or using analytical methods. The procedure for self-absorption factor determination includes firstly calculation of CS that is obtained for various densities ρ and photon energies E. Then data collected for each photon energy are fitted to an appropriate function Cs(ρ) or Cs(E, ρ) that are written as Cs(ρ) = a exp(-b ρ) or Cs(E, ρ) = a(E) exp[-b(E) ρ] Where a and b are the adjustable parameters. 4 For low energy gamma-ray (below about 100 keV) these formulae are applicable only for materials of similar composition on account of the relation between the mass attenuation coefficient and the atomic number (Jodlowski, 2006). Gamma-ray photons are known to be attenuated through the material according to the following relation; I I o .e( m . x ) & I I o .e ( m .x. ) Where: Io: the photons, with energy E, intensity without attenuation I: the photons, with energy E, intensity after attenuation μ: the linear attenuation coefficient, cm-1, μm: the mass attenuation coefficient, μm = μ/ ρ, cm-2/g. x: the sample thickness (or effective thickness) Jodlowski (2006) compared the different methods for self-absorption correction in gamma-ray spectrometry of environmental samples and conclude the following (Jodlowski, 2006): - The experimental method is time consuming and inconvenient. It requires that the curves be fitted to a small number of measurement data, which gives rise to a relatively high uncertainty. - Monte Carlo techniques are not so widely adopted in the laboratory conditions, as they require considerable skill and experience in computer simulations. 5 - The exact analytical description of self-absorption is a complex task that is why simplified models are adopted instead. In the first place, it is assumed that the selfabsorption correction is proportional to the term exp (-μ x) or exp (-μm ρ x). Another widely applied analytical formula providing a simplified description of selfabsorption in cylindrical samples is involving the integration of photons of the specified energy coming from subsequent sample layers and reaching the detector. Self-absorption correction, C sa , for the sample with reference to air (matrix μ ≈ 0) is given by the following equation: C sa .x 1 e x Another analytical method was proposed where in order to apply Cs computing techniques, it is required that attenuation coefficients of the standard and sample materials be known before hand. For high energy range (more than 100 or 200 keV) of gamma-ray photons, the mass attenuation coefficient is obtained on the basis of the sample approximate chemical composition, assuming that for a given energy value the value of Cs depends on the sample density exclusively. For low energies (below 100 keV), the mass attenuation coefficient in different materials may vary significantly and the attenuation coefficient has to be determined experimentally. The transmission method developed by Cutshall et al (1983) where a point-like radioactive source is positioned above the sample located on the detector and the number of counts in the full energy peak is measured. The sample self-absorption correction factor Cs with the reference to the standard is given as: 6 Cs ln(I c / I s ) 1 (I c / I s ) Where Ic, Is are the transmission experiment results (detector count rates) for the standard and sample, respectively (Cutshall et al., 1983). The aim of our work is to shed more light on the significant of self-absorption correction for precise gamma-ray spectrometry of bulk samples and a simple practical procedure to correct it. Experimental work Four point gamma-ray emitter sources (241Am, 133 Ba, 137 Cs, and 60 Co) with the following energy transitions (59.5 keV, 81.2 keV, 276 keV, 302.4 keV, 356 keV, 383.9 keV, 661.6 keV, 1173.2 keV and 1332.5 keV) were utilized to perform transmission measurement for standard and other samples. Each sample was placed on the top of the detector, and then the four point sources were put above the sample. Enough distance was left between the detector and the point sources to maintain acceptable dead time. To obtain acceptable low statistical error, counting time was 2000 second - Samples preparation Ten samples of different matrices (water, grass, soil and geological ore), with the same geometry, have been prepared to test the gamma-ray attenuation in relation to samples matrix and density differences. Samples were prepared in identical polypropylene containers and tightly closed. This set of samples cover a wide range of 7 appeared densities of 0.48 – 1.69 g/cm3. Another set of samples was prepared to have the same geometry and matrix, but different appeared densities that ranged from 0.98 to 1.42 g/cm3. Another set of samples was prepared to have the same geometry and appeared density (- 5%), and of different matrices (water, soil and geological ore). - Gamma ray spectrometry: Gamma-ray spectrometer with a GX4019-7500Sl CANBERRA extended range electrode germanium detector with a CANBERRA model 2002CSL preamplifier was used. The hyper pure germanium (HpGe) detector has a relative efficiency of 40% and full width at half maximum (FWHM) of 1.9 keV for 60 Co gamma energy transition at 1332.5 keV. It is connected to 8k computerized multi-channels analyzer where gamma spectrum analysis performed using Genie-2000 gamma-ray spectroscopy software by Canberra. Results and Discussions: Gamma-ray spectrometry based on hyper pure germanium detectors is a very powerful tool that has a very wide range of applications in radiation measurement, generally, and specially in environmental radioactivity measurement in different bulk samples. Many studies have been focused on the accurate detection efficiency calibration using different, experimental, Monte Carlo simulation and analytical, techniques. Other studies have been concerned with background reduction, by passive and/or active shielding, for low level measurement; quality control, optimum sample-detector geometry arrangement, and other aspects. However, the variations of chemical and 8 physical properties of the bulk samples could be the main source of noticeable uncertainty, especially for the samples with high apparent densities and different chemical compositions from that of the standard samples used for efficiency calibration and specially in relatively low gamma-ray transmission energies (below 100-200 keV). In the low gamma-ray energy ranges, correction of self- attenuation (absorption) of gammaray in the bulk samples is essential for accurate results. Different methods have been suggested and applied for self-attenuation corrections. It is using experimental, Monte Carlo simulation and analytical techniques (Sigh et al., 2004; Pilleyre et al., 2006; Sima and Dovlete, 1997; Cincu, 1992; Korun, 2000). All this techniques can overcome the self-attenuation problem in many cases but can not ensure it, especially in low energy range and for some samples of high densities that far from the densities of the calibration sources. In this work we are suggesting a modified practical method for attenuation correction based on the relative gamma-ray transmissions (I/Io) through both standard and the different bulk samples relatively to the air or water samples of the sample geometry. The relative gamma-ray transmissions for a set of bulk samples of different densities (0.48-1.69 g/cm3) and different matrices for different photon energies (59.51332.5 keV) are given in Table 1 and shown in Figure 1. The variation of the relative transmission is very wide especially for low energies (59.5 and 81.2 keV) that could be because of the density and matrix differences. These variations are clearly reduced at the high energies range of gamma rays. Another set of samples, of the same matrix and different densities (0.98-1.42 g/cm3) have been examined. The relative gamma-ray transmission (I/Io) for energy range 59.5-1332.5 keV and for different sample densities are given in Table 2 and shown in 9 Figures 2 and 3. Both figures show the variation of the relative gamma ray transmission as a function of energy and sample density where the variation of the relative transmission is linearly correlated to the sample density and polynomially (exponentially growth) correlated to the photon energy. The last set of samples has the same apparent density and different matrices (i.e. chemical composition). The relative gamma-ray transmission as a function of gamma-ray energy are given in Table 3 and shown in Figure 4. It is obviously clear that the variations in the transmission of gamma-rays of different energies, due to the variation in the chemical composition of the samples, are considerable. This figure show the variation in the gamma-ray self attenuation either due to the density or the chemical composition or both of them. The uncertainty in the gamma-ray spectrometry, due to the samples’ densities or chemical compositions or both, has different patterns. Unless the difference in density and chemical composition is fairly close to that of the reference, the selfattenuation correction of the bulk sample is a necessity to eliminate its effect on the accuracy of the detection efficiency. In the following section of this paper, we shall explain our suggested method for self-attenuation correction without a tedious work of the experimental approach or the need to mathematical skills or the simplified assumption of both Monte Carlo simulation and analytical approaches. In this work, we are suggesting a simple method to overcome the variation in the counting efficiency due to self-attenuation in the bulk samples. This method is based on measuring the relative transmission of gamma-ray of different energies as a ratio of gamma-ray 10 photo-peak count rate pass through the reference standard and other (unknown) samples to that through the air or other samples such as water of the same geometry, (I/I0)std and (I/I0)unk, respectively. To overcome the errors due to sample’s radioactivity contributions to the point source counts, point sources were selected so that they have relatively higher activity than that of the samples themselves. Also, they were counted together for relatively short time. The advantage of this method is that it takes into consideration both the density and matrix effect while correcting for the counts, i.e. detection efficiency. Accordingly, the activity concentration would be corrected using the following formula Aunk CR unk (I ) I 0 std f (I I )unk 0 (4) Where CRunk : The unknown sample’s count rate at the specified photo-peak energy : The photopeak efficiency for certain geometry : The photopeak intensity (I/I0)std : relative transmission of gamma-ray through the standard reference material. (I/I0)unk : relative transmission of gamma-ray through the unknown measured sample. Equation no. 4 could be improved to the following formula; Aunk CRunk f (5) Where is the correction factor calculated by dividing the point source counts passing through the standard reference material by that of unknown sample according to the following equation: 11 (I (I I0 )Std I0 )Unk (6) Application: The IAEA-RGU-1 (uranium ore) reference sample was used to perform the efficiency calibration for the energy transition of 63.3 keV (3.6 %) of the 234 Th. The 234 Th (in equilibrium with its parent 238U) is usually used for the specific activity assessment of 238U. The estimated correction factors were used to calculate the 238 U activities of other ore samples, of the same geometry and of different densities, as well as uranyl nitrate solution with known uranium concentrations. The corrections suggested in this paper were then applied for self-attenuation correction. The results were compared and shown in Table 4. The performed corrections were based on the transmission differences due to the nearest energy transition (59.5 keV) of the 241 Am. Therefore, when the energy transition of interest is not available as point source, it is recommended to produce a fitted correction curve using energies as near as possible to that needed. Using this curve, a correction factor could be easily obtained for most of required energy transitions. It is clearly noticed that a considerable improvements has been encountered in the measuring process, where the percentage of error is dramatically decreased. Conclusions: The suggested procedure introduced in this paper is an innovative, reliable and straightforward method to overcome the errors generally produced due to the difference in samples matrices and densities. It also minimizes the measurement errors. There is no need for tedious experimental work or assumptions for the mathematical approaches. This method could be adopted within the laboratories where they encounter incoming 12 wide varieties of samples for analysis. A simple calculation program could be produced to help the unskilled technicians in the routine measurements of gamma-ray spectrometry for bulk sample. Finally, the applicability of this method is almost unlimited as long as the sample is homogenous. References: Abbas, Mohamoud I., Nafee, Sherif, Selim, Younis S., 2006. Calibration of cylindrical detectors using a simplified theoretical approach. Appl. Radiat. Isot 64, 1057-1064. Aguiar, Julio C., Eduardo Galiano, Jorge Fernandez, 2006. Peak efficiency calibration for attenuation corrected cylindrical sources in gamma ray spectrometry by the use of a point source. Appl. Radiat. Isot 64, 1643-1647. Boshkova, T., 2003. Effective thickness of bulk samples in close measuring gamma-ray spectrometry. Appl. Radiat. and Isot. 59, 1-4. Cincu, Em. , 1992. A practical method for accurate measurement of radionuclide activities in environmental samples. Nucl. Instr. and Meth. A 312, 226230. Cutshall, N., Larsen, , I.L., Olsen, C.R., 1983. Direct analysis of Pb-210 in sediment samples: a self-absorption corrections. Nucl. Instr. and Meth. A 206, 309312. Gurriaran, R., Barker, E., Bouisset, P., Cagnat, X., Ferguson, C., 2004. Calibration of a very large ultra-low background well-type Ge detector for envoironmental sample measurements in an underground laboratory. Nucl. Instr. and Meth. A 524, 264-272. 13 Jodlowski, Pawel, 2006. Slef-absorption correction in gamma-ray spectrometry of environmental samples- an overview of methods and correction values obtained for the selected geometries. Nukleonika 51 (2), S21-S25. Korun, M. , 2000. Calulation of self-attenuation factors in gamma-ray spectrometry for samples of arbitrary shape. Radioanal. Nucl. Chem. 244(3) 685-689. Pilleyre, T., Sanzelle, S., Miallier, D., Fain, J., Courtine, F., 2006. Theoretical and experimental estimation of self-attenuation correction in determination of 210 Pb by gamma-spectrometry with well Ge detector. Radiation Measurements 41, 323-329. Quindós, L. S., Sainz, C., Fuente, I., Nicolás, J., Quindós, L., Arteche,J., 2006. Correction by self-attenuation in gamma-ray spectrometry for environmental samples. Radioanal. Nucl. Chem. 270 (2), 339-343. Saegusa, Jun, Kawasaki, Katsuya, Mihara, Akira, Ito, Mitsuo, Yoshida, Makoto, 2004. Determination of detection efficiency curves of HPGe detectors on radioactivity measurement of volume samples. Appl. Radiat. Isot 61, 13831390. San Miguela, E.G. , Perez-Morenoa, J.P., Bolivara, J.P., Garcia-Tenoriob, R., 2004. A semiempirical approach for determination of low-energy gamma-emmiters in sediment samples with coaxial Ge-detectors. Appl. Radiat. Isot 61, 361–366. Sima, O. , Dovlete, C., 1997. Matrix effects in the activity measurement of environmental samples- implementation of specific corrections in Gammaray spectrometry analysis program. Appl. Radiat. Isot. 48 (1), 59-69. Singh, Charanjeet, Singh, Tejbir, Kumar, Ashok, Mudahar, Gurmel S., 2004. Energy and chemical composition dependence of mass attenuation coefficient of building materials. Annals of Nuclear Energy 31, 1199-1205. 14 Vargas, M. Jurado, Fernandez Timon, A., Cornejo Diaz, N., Perez Sanchez, D., 2002. Monte Carlo simulation of the self-absorption correction for natural samples in gamma-ray spectrometry. Appl. Radiat. Isot 57, 893-898 15 Table 1: Gamma-ray transmissions (I/Io) for a set of samples of different environmental matrices and for different photon energies. Serial Sample types No. Energy, keV Density g.cm -3 59.5 81.2 276 302.4 356 383.9 661.6 1173 1332.5 1 Air 0.00 1.00 1.00 1.00 1.00 1.00 1.00 1.00 1.00 1.00 2 Grass 0.48 0.81 0.86 0.88 0.90 0.89 0.89 0.93 0.92 0.92 3 Soil-1 0.95 0.59 0.75 0.82 0.85 0.85 0.87 0.89 0.90 0.91 4 Ore-1-1 0.97 0.54 0.68 0.83 0.83 0.82 0.86 0.85 0.89 0.88 5 Ore-2-1 0.98 0.65 0.62 0.83 0.84 0.82 0.84 0.87 0.89 0.91 6 Water 1.00 0.68 0.74 0.79 0.83 0.80 0.84 0.86 0.89 0.90 7 Ore-1-2 1.11 0.64 0.68 0.80 0.81 0.78 0.83 0.87 0.90 0.90 8 Ore-2-2 1.25 0.61 0.64 0.80 0.79 0.75 0.82 0.85 0.89 0.88 9 Soil-2 1.37 0.49 0.60 0.77 0.75 0.78 0.81 0.82 0.87 0.87 10 Ore-2-3 1.42 0.57 0.61 0.77 0.77 0.72 0.79 0.83 0.87 0.86 11 Soil-3 1.69 0.43 0.54 0.68 0.72 0.72 0.71 0.78 0.84 0.84 16 Table 2: Gamma-ray transmissions (I/Io) for the same sample matrix (Ore-2) with different densities (g.cm-3) and for different photon energies. Energy, keV Density g/cm3 0.98 1.11 1.25 1.42 59.5 81.2 276 302.4 356 383.9 661.6 1173 1332.5 0.65 0.64 0.61 0.57 0.72 0.68 0.64 0.61 0.83 0.80 0.80 0.77 0.84 0.81 0.79 0.77 0.82 0.78 0.75 0.72 0.84 0.83 0.82 0.79 0.87 0.87 0.85 0.83 0.89 0.90 0.89 0.87 0.91 0.90 0.88 0.86 17 Table 3: Gamma-ray transmissions (I/Io) for a set of different sample matrices with the same densities, ±5% (g.cm-3) and for different photon energies Sample types Ore-1-1 Soil-1 Ore-2-1 Water Energy, keV Density g/cm3 0.97 0.95 0.98 1.00 59.5 0.54 0.59 0.65 0.68 81.2 0.68 0.75 0.62 0.74 276 0.83 0.82 0.83 0.79 302.4 0.83 0.85 0.84 0.83 356 0.82 0.85 0.82 0.80 383.9 0.86 0.87 0.84 0.84 661.6 0.85 0.89 0.87 0.86 1173 0.89 0.90 0.89 0.89 1332.5 0.88 0.91 0.91 0.90 18 Table 4: Specific activity U-238, Bq/kg, calculated for reference samples using the traditional method (without correction) and using present suggested corrections Reference value Calculated Value (g/cm ) (Bq/Sample) (Uncorrected) Ore -2- 1.25 464.7 440.5 -5.2 466.9 0.5 Ore -2 1.42 528.9 486.6 -8.0 545.0 3.0 1.02 437.8 471.2 7.9 442.9 1.1 Sample Code Uranyl Nitrate solution Density 3 Bias % Calculated Value (corrected) Bias % 19 59.5 keV 0.9 81.0 keV 0.9 Y = 0.93 - 0.287 X 0.8 0.9 Y = 0.97 - 0.26 X 0.8 0.8 0.7 0.7 0.7 0.6 0.6 0.6 0.5 0.5 0.5 0.4 0.6 0.8 1.0 1.2 1.4 1.6 0.4 1.8 1.0 1.0 0.9 0.9 0.8 0.8 302.9 keV 0.7 0.6 0.8 1.0 1.2 1.4 1.6 0.6 0.8 1.0 1.2 1.4 1.6 0.9 0.8 1.0 1.2 1.4 1.6 Y = 0.99 - 0.117 X 0.6 0.9 0.9 0.6 0.8 1.0 1.2 1.4 1.6 1.8 0.6 0.8 1.0 1.2 1.4 1.6 1.8 Y = 0.97 - 0.069 X 0.6 0.5 0.4 0.4 0.4 1.8 1332.6 keV 0.7 Y = 0.95 - 0.061 X 0.5 0.4 1.6 0.8 1173.2 keV 0.6 0.5 0.4 1.8 1.0 0.7 661.6 keV 0.7 1.4 0.4 0.6 1.0 0.8 0.8 1.2 0.5 0.4 1.0 1.0 Y = 0.97 - 0.155 x 0.6 0.4 1.8 0.8 356.01 keV 0.7 Y = 0.98 - 0.138 X 0.5 0.4 0.6 0.8 383.9 keV 0.6 0.4 0.4 1.8 0.9 0.7 0.5 Y = 0.97 - 0.149 x 1.0 Y = 0.98 - 0.155 X 0.6 276.4 keV 0.4 0.4 0.4 I/IO 1.0 1.0 1.0 0.4 0.6 0.8 1.0 Denisty, g/ cm 1.2 1.4 1.6 1.8 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 3 Fig 1: Gamma-ray transmissions (I/Io) for a set of different environmental samples of different matrices with different densities (g.m-3)and for different photon energies 20 1.0 Ore-2 samples Energy, keV 59.5 81.0 276.4 302.9 356.01 383.9 661.6 1173.21 1332.6 0.9 I/I0 0.8 0.7 0.6 0.9 1.0 1.1 1.2 1.3 Denisty, g/cm 1.4 1.5 1.6 3 Fig 2: Gamma-ray transmissions (I/Io) for the same sample matrix (Ore-2) with different densities (g.m-3) and for different photon energies 21 0.95 0.90 0.90 2 ± R =0.97 2 R =0.97 0.85 0.85 0.80 0.80 0.75 0.75 0.70 Polynomial regression equation: 2 3 4 Y = A + B1*X + B2*X + B3*X + B4*X A = 0.535 B2= -3.08397E-6 B4= -6.32973E-13 0.65 B1= 0.00169 B3= 2.36547E-9 0.70 Polynomial regression equation: 2 3 Y = A + B X + B X +B3X 1 2 0.65 A = 0.60664 I / I0 B1 = 8.57177E-4 B2 = -8.82923E-7 B3 = 3.06021E-10 0.60 0.60 0 200 400 600 800 1000 1200 1400 0 200 400 600 800 1000 1200 1400 0.90 0.90 0.85 0.85 2 R =0.94 2 0.80 R =0.94 0.80 0.75 0.75 0.70 Polynomial Regression Equation: 0.70 2 3 Y=A+B X+B X +B X 1 2 3 0.65 Polynomial Regression Equation: 0.65 2 3 Y=A+B X+B X +B X 1 2 3 0.60 0 200 400 600 800 0.60 A = 0.56784 B1= 9.60098E-4 B2= -1.01224E-6 B3= 3.54804E-10 1000 1200 0.55 1400 0 200 400 600 800 A = 0.53003 B1 = 0.001 B2 = -1.04263E-6 B3 = 3.61575E-10 1000 1200 1400 Energy, keV Fig 3: Gamma-ray transmissions (I/Io) for the same sample matrix (Ore-2) with different densities (g/cm3) and for different photon energies. 22 1.0 Ore -1 Soil-1 Water Ore-2 Energy, keV 59.5 81.2 276 302.4 356 383.9 661.6 1173.2 1332.5 0.9 I/Io 0.8 0.7 0.6 0.5 0.4 0.94 0.95 0.96 0.97 0.98 0.99 Denisty, g/cm 1.00 1.01 1.02 3 Fig 4: Gamma-ray transmissions (I/Io) for a set of different sample matrices with the same densities, ±5% (g/cm3) and for different photon energies 23