GMSU_AdminManual_000..

advertisement

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 1 of 48

Gubbs Mass Spec Utilities 3.x.x

GMSU 3 and earlier

Supported Platforms

Sciex® Analyst™ 1.3.1 - 1.4.2

Thermo® XCalibur™ 2.0.5 and earlier

Administration and Installation

Manual

Gubbs, Inc.

265 Blue Spruce Circle

Alpharetta, GA 30005

P: 770-573-0169

F: 508-453-1338 gubbs@gubbsinc.com

www.gubbsinc.com

Version 2.0 (June 4, 2007)

D:\726884342.doc

2

3

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

Table of Contents

Section

1

Page 2 of 48

Page Number

INTRODUCTION ........................................................................................................................................................ 5

4

MANUAL HISTORY ................................................................................................................................................... 5

HIGH THROUGHPUT (HT) VS LOW THROUGHPUT (LT) ACQUISITION ................................................... 6

DEFINITIONS .............................................................................................................................................................. 6

5 INSTALLATION OVERVIEW .................................................................................................................................. 7

5.1

I NSPECT SYSTEM REQUIREMENTS AND PREREQUISITES ............................................................................................... 7

5.2

E XECUTE S ETUP .

EXE ................................................................................................................................................... 7

5.3

D ETERMINE THE PHYSICAL LOCATION OF THE GMSU DATABASE G UBBS _01.

MDB ( IF APPLICABLE ) .......................... 7

5.4

P REPARE W INDOWS

®

DIRECTORY SHARE AND DIRECTORY PERMISSIONS ( IF APPLICABLE ) ........................................ 7

5.5

C ONFIGURE DATABASE AND DSN ( IF NEEDED ) ........................................................................................................... 7

5.6

C ONFIGURE GMSU_ XX .

INI ......................................................................................................................................... 7

5.7

E XECUTE U PDATE GMSUD ATABASE .

XLS ( IF APPLICABLE ) ......................................................................................... 7

5.8

C ONFIGURE GMSU C ONFIGURATION D EFAULT P ARAMETERS ................................................................................... 7

6 SYSTEM REQUIREMENTS ...................................................................................................................................... 8

7 PRE-REQUISITES ...................................................................................................................................................... 8

7.1

S CIEX ENVIRONMENT .................................................................................................................................................. 8

7.2

T HERMO F INNIGAN ENVIRONMENT ............................................................................................................................. 9

8

9

SETUP EXECUTION .................................................................................................................................................. 9

8.1

D IRECTORY CREATION ................................................................................................................................................ 9

8.2

S YSTEM DSN ODBC FILE CREATION .......................................................................................................................... 9

8.3

GMSU_ XX .

INI FILE CREATION .................................................................................................................................. 10

8.4

G UBBS _01.

MDB FILE CREATION ................................................................................................................................ 10

PHYSICAL LOCATION OF GMSU DATABASE GUBBS_01.MDB .................................................................. 10

10 WINDOWS SHARE PERMISSIONS AND DIRECTORY PERMISSIONS ....................................................... 11

10.1

L OCAL D ATA D IRECTORY ......................................................................................................................................... 11

10.2

S HARED N ETWORK D ATA D IRECTORY ( IF APPLICABLE ) ........................................................................................... 12

10.3

S HARED L OCAL D ATA D IRECTORY ( IF APPLICABLE ) ................................................................................................. 13

11 DATABASE CONFIGURATION – MICROSOFT® ACCESS ............................................................................. 14

12 DSN CONFIGURATION – MICROSOFT® ACCESS .......................................................................................... 15

12.1

ODBC D ATA S OURCE A DMINISTRATOR ................................................................................................................... 15

12.2

C REATE N EW D ATASOURCE ..................................................................................................................................... 15

12.3

ODBC S ETUP ............................................................................................................................................................ 16

12.4

S ELECT DATABASE .................................................................................................................................................... 16

13 DATABASE CONFIGURATION – MICROSOFT® SQL SERVER ................................................................... 18

13.1

C REATE A SQL S ERVER DATABASE USING THE MS A CCESS U PSIZING W IZARD ....................................................... 18

13.2

C REATE A GMSU DATABASE ACCOUNT .................................................................................................................... 22

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 3 of 48

Table of Contents

Section Page Number

13.3

R ECORD THE USER ACCOUNT AND PASSWORD INFORMATION ................................................................................... 22

13.4

L OCAL D ATA S OURCE N AME (DSN) G UBBS _100 ..................................................................................................... 22

14 DSN CONFIGURATION – MICROSOFT® SQL SERVER ................................................................................. 23

14.1

ODBC D ATA S OURCE A DMINISTRATOR ................................................................................................................... 23

14.2

C REATE NEW DATA SOURCE ...................................................................................................................................... 24

14.3

C REATE A NEW DATA SOURCE ................................................................................................................................... 25

14.4

E NTER CONNECTION PARAMETERS ............................................................................................................................ 26

14.5

A DDITIONAL PARAMETERS ........................................................................................................................................ 26

14.6

T EST CONNECTION .................................................................................................................................................... 27

15 GMSU_XX.INI ........................................................................................................................................................... 28

15.1

MS A CCESS GMSU_ XX .

INI CONFIGURATION ........................................................................................................... 28

15.2

SQL S ERVER GMSU_ XX .

INI CONFIGURATION .......................................................................................................... 29

16 GUBBS INC PASSWORD ENCRYPTION TOOL ................................................................................................. 29

17 GMSU DATABASE UPDATE .................................................................................................................................. 29

18 GMSU CONFIGURATION DEFAULT PARAMETERS ...................................................................................... 31

19 CONSOLE ................................................................................................................................................................... 31

20 CONFIGURATION UTILITY .................................................................................................................................. 32

20.1

C ACO G LOBAL ........................................................................................................................................................ 32

20.2

C ACO

– A NALYSIS T IME P OINTS ............................................................................................................................... 34

20.3

C ACO

– C HROMATOGRAPHY P EAK S ETS ................................................................................................................... 35

20.4

C ACO

– C ONSTANTS ................................................................................................................................................. 36

20.5

F LAGS ....................................................................................................................................................................... 37

20.6

G ENERIC V IEWER ...................................................................................................................................................... 38

20.7

G LOBAL .................................................................................................................................................................... 40

20.8

H EPATIC C LEARANCE C ALCULATOR ......................................................................................................................... 42

20.9

H EPATIC C LEARANCE C ALC .

E XPT .

C ONSTANTS

– E XISTING E XPERIMENTS T AB ..................................................... 43

20.10

H EPATIC C LEARANCE C ALC .

E XPT .

C ONSTANTS

– A DD S PECIES /T YPE T AB ............................................................. 44

20.11

H EPATIC C LEARANCE C ALC .

E XPT .

C ONSTANTS

– E XPERIMENT T IME P OINTS T AB ................................................. 45

20.12

M ODIFY M ETHOD ..................................................................................................................................................... 47

21 REFERENCES ........................................................................................................................................................... 48

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 4 of 48

Table of Figures

Figure Page Number

F IGURE 1 N ETWORK BROWSE

F IGURE 2 MS A CCESS DNS D ATA S OURCE A DMINISTRATOR

F IGURE 3 MS A CCESS DNS C REATE A N EW D ATA S OURCE W IZARD

F IGURE 4 MS A CCESS DNS ODBC M ICROSOFT A CCESS S ETUP

F IGURE 5 MS A CCESS DNS S ELECT D ATABASE

F IGURE 6 MS A CCESS U PSIZING W IZARD

F IGURE 7 MS A CCESS U PSIZING W IZARD C HOOSE C REATE NEW DATABASE

F IGURE 8 MS A CCESS U PSIZING W IZARD C ONNECTION SETTINGS

F IGURE 9 MS A CCESS U PSIZING W IZARD E XPORT ALL TABLES

F IGURE 10 MS A CCESS U PSIZING W IZARD

– E XPORT ATTRIBUTES

F IGURE 11 MS A CCESS U PSIZING W IZARD

– N O APPLICATION CHANGES

F IGURE 12 MS A CCESS U PSIZING W IZARD F INISH

F IGURE 13 MS SQL S ERVER GMSU ACCOUNT ROLES

F IGURE 14 SQL S ERVER DNS D ATA S OURCE A DMINISTRATOR

F IGURE 15 SQL S ERVER DNS C REATE A N EW D ATA S OURCE W IZARD

F IGURE 16 SQL S ERVER DNS C REATE A NEW DATA SOURCE

F IGURE 17 SQL S ERVER DNS E NTER CONNECTION PARAMETERS

F IGURE 18 SQL S ERVER DNS E NTER ADDITIONAL PARAMETERS

F IGURE 19 SQL S ERVER DNS E NTER ADDITIONAL PARAMETERS

F IGURE 20 SQL S ERVER DNS T EST CONNECTION

F IGURE 21 SQL S ERVER DNS T EST CONNECTION

F IGURE 22 GMSU D ATABASE U PDATE U TILITY F IRST TIME USE SCREEN

F IGURE 23 GMSU D ATABASE U PDATE U TILITY

– A FTER E XECUTE

F IGURE 24 GMSU C ONSOLE

F IGURE 25 C ONFIGURATION U TILITY

– C ACO (P ERMEABILITY C ALCULATOR ) – G LOBAL

F IGURE 26 C ONFIGURATION U TILITY

– C ACO (P ERMEABILITY C ALCULATOR ) – A NALYSIS T IME P OINTS

F IGURE 27 C ONFIGURATION U TILITY

– C ACO (P ERMEABILITY C ALCULATOR ) – C HROMATOGRAPHY P EAK S ETS

F IGURE 28 C ONFIGURATION U TILITY

– C ACO (P ERMEABILITY C ALCULATOR ) – C ONSTANTS

F IGURE 29 C ONFIGURATION U TILITY F LAGS

F IGURE 30 C ONFIGURATION U TILITY

– G ENERIC V IEWER

F IGURE 31 C ONFIGURATION U TILITY G LOBAL

F IGURE 32 C ONFIGURATION U TILITY

– H EPATIC C LEARANCE C ALCULATOR

F IGURE 33 C ONFIGURATION U TILITY

– H EPATIC C LEARANCE C ALC .

E XPT .

C ONSTANTS

– E XISTING E XPERIMENTS

34

35

36

37

30

30

31

32

26

27

27

28

23

24

25

26

38

40

42

19

19

20

20

21

21

22

14

15

16

16

17

18

43

44

T AB

F IGURE 34 C ONFIGURATION U TILITY

– H EPATIC C LEARANCE C ALC .

E XPT .

C ONSTANTS

– A DD S PECIES /T YPE T AB

F IGURE 35 C ONFIGURATION U TILITY

– H EPATIC C LEARANCE C ALC .

E XPT .

C ONSTANTS

– E XPERIMENT T IME P OINTS

T AB I NITIAL

F IGURE 36 C ONFIGURATION U TILITY

– H EPATIC C LEARANCE C ALC .

E XPT .

C ONSTANTS

– E XPERIMENT T IME P OINTS

T AB E DIT

F IGURE 37 C ONFIGURATION U TILITY

– M ODIFY M ETHOD

45

46

47

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

1

Page 5 of 48

Introduction

Gubbs Mass Spec Utilities (GMSU) is a suite of utilities that enhance the ability of scientists to perform high volume actions in mass spectrometer data acquisition systems. This is useful in laboratories engaged in high-throughput studies.

GMSU uses either Microsoft® (MS) Access or MS SQL Server as a datastore. MS Access may be used to quickly setup a GMSU instance for user testing or production use. MS SQL Server may be used if a more secure and possibly more efficient environment is desired. The MS Access database can be directly upgraded to SQL Server without loss of data.

GMSU may be used in a multi-user environment. There are several configuration possibilities for multi-

user environment which are described in more detail in Section 0. Briefly, the GMSU database and

acquired mass spectrometer raw data files may be stored in a common location accessible to others (either a shared network directory or a shared local directory). Multiple installations of GMSU may be configured to use the database and access the raw data files located in the common location. This allows users in the analytical department to use GMSU to generate and optimize chromatographic integration acquired data. Once the chromatographic integration data is approved, users in the pharmacokinetics department may be notified and can immediately generate pharmacokinetic data.

GMSU data are stored in a database. This data may be accessed or generated from multiple installations of GMSU.

2 Manual History

To view the changes between this version and the previous version of this manual, please do the following:

Open the MS Word version of the manual o The most recent versions of manuals are available at www.gubbsinc.com

Show the Reviewing toolbar in the Word document

In the Show button of the Reviewing toolbar, select Insertions and Deletions

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

3 High Throughput (HT) vs Low Throughput (LT) Acquisition

Page 6 of 48

One of the salient features of GMSU is that it can process HT acquired data as well as LT acquired data.

Typically, mass spectrometry data acquisition systems require that compound(s) of interest occur once in a chromatographic analysis. Therefore, in order to analyze a 12-point assay, scientists had to acquire twelve separate ‘samples’ within a data file. This is referred to as LT acquisition in the GMSU system.

In HT acquisition, users open the data acquisition window for as long as needed and inject and acquire all assay time points in a single chromatographic run. The benefits of HT acquisition 1 - 4 include:

The ability to view all chromatographic peaks simultaneously to check for possible instrumental or sample handling problems and/or trends

When combined with multiple-column switching techniques, increases throughput by decreasing the time between sample injections and minimizes communication problems that can occur between instrument and data acquisition system.

NOTE: Currently HT acquisition data processing is supported on the Sciex platform in the Hepatic

Clearance and Permeability modules.

4 Definitions

Term

CDS

DSN

GMSU

HT

LT

Chromatography Data System

Data Source Name

Gubbs Mass Spec Utilities

High Throughput

Low Throughput

Definition

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

5 Installation Overview

Page 7 of 48

GMSU Setup includes the following steps. Please refer to each referenced section for a detailed discussion of each step

Step Section

5.1

Inspect system requirements and prerequisites

Sections 6 - 7

Section 8

5.2

Execute Setup.exe

5.3

Determine the physical location of the GMSU database

Gubbs_01.mdb (if applicable)

5.4

Prepare Windows® directory share and directory permissions (if applicable)

Note: Does not need to be performed if upgrading an existing

GMSU installation

Thermo users please pay special attention to Section 10.3

5.5

Configure database and DSN (if needed)

Note: Does not need to be performed if upgrading an existing

GMSU installation

Microsoft Access

Microsoft SQL Sever

5.6

Configure GMSU_xx.ini

Notes: Does not need to be performed if upgrading an existing

GMSU installation

5.7

Execute UpdateGMSUDatabase.xls (if applicable)

To be performed with GMSU uprades

5.8

Configure GMSU Configuration Default Parameters

Admins must configure several default values for proper GMSU usage.

Notes: Does not need to be performed if upgrading an existing

GMSU installation

Section 9

Sections 10

Sections 11 - 12

Sections 13 - 14

Section 15

Section 17

Section 18

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

6 System Requirements

Page 8 of 48

The following describes the environments on which GMSU was tested. The user is free to install GMSU on workstations of differing configuration with the understanding that GMSU should be tested before implementing.

OS

CPU (GHz)

RAM (MHz)

Sciex Analyst 1,4

Thermo Excalibur 2,4

Microsoft® Office 3

UpdateGMSUDatabase.xls

Client Workstations

MS Windows XP SP2

1.8

512

1.4.1, 1.4.2

2.0.5

Office 2003 SP2

MS Excel 2003 SP2

GubbsIncPswd_01.xls

1

MS Excel 2003 SP2

GMSU supports Sciex Analyst 1.3.x and Analyst 1.4.2

2 GMSU supports XCalibur 2.0.5 and previous versions

3 At a minimum, Word and Excel must be installed. GMSU is compatible with Office XP or later.

4

Not required for data reporting (see Section 7)

SQL Server

OS

CPU (GHz)

RAM (MHz)

1 If SQL Server is to be used

MS SQL Server Server 1

SQL Server 2000 SP4, SQL Server 2005

MS Windows Server 2003 SP1

1.3

1250

GMSU uses Microsoft Office to generate reports in Excel and Word.

7 Pre-Requisites

7.1

Sciex environment

7.1.1

Data Processing

In order to view chromatography and process PK data, Sciex Analyst must be installed on the workstation on which the GMSU client is being installed. GMSU uses several Sciex

Analyst Active X objects and .dll’s to ensure that information provided by GMSU is identical to information if it were to be generated on Sciex Analyst.

If Sciex Analyst is not installed on the workstation, GMSU will not allow the user to open data processing modules.

7.1.2

Data Reporting

Sciex Analyst is not required to be installed to report data. All data reporting modules are available.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

7.2

Thermo Finnigan environment

Page 9 of 48

7.2.1

Data Processing

The Thermo software development kit called ‘SDK Plus’ must be installed on the workstation on which the GMSU client is being installed. GMSU uses several XDK Plus

Active X objects and .dll’s.

If XDK Plus is not installed on the workstation, GMSU will not allow the user to open data processing modules.

7.2.2

Data Reporting

XDK Plus is not required to be installed to report data. All data reporting modules are available.

8 Setup Execution

Run the GMSU Setup.exe executable.

In addition to installing the GMSU software, Setup.exe will perform the following tasks:

8.1

Directory creation

Setup.exe will create a C:\GubbsInc\GMSU\ directory and several subdirectories related to

GMSU operation (see Section 10.1). If the directories already exist, then they will not be re-

created and existing files within these directories will not be deleted or modified.

8.2

System DSN ODBC file creation

Setup.exe will create a System DSN ODBC file used by GMSU to communicate with the GMSU database.

8.2.1

Original Installation

If the GMSU installation is an original installation on a workstation, Setup.exe will prompt the user to choose MS Access or MS SQL Server as the GMSU database.

If MS Access is chosen, no further DSN configuration is needed.

If MS SQL Server is chosen, then further database and DSN configuration is needed

as described in Section 13 and 14

NOTE: If for some reason Setup.exe cannot create the System DSN, Setup.exe will inform the user that DSN creation was unsuccessful. If this should happen,

then the DSN file will need to be created manually according Section 12 for

MS Access or Section 14 for MS SQL Server.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

8.2.2

Upgrade Installation

Page 10 of 48

If the GMSU installation is an upgrade, then the System DNS will not be created unless it does not exist for reason.

8.3

GMSU_xx.ini file creation

8.3.1

Original Installation

Setup.exe will create a GMSU_xx.ini file in C:\GubbsInc\GMSU\.

8.3.2

Upgrade Installation

If GMSU_xx.ini already exists in C:\GubbsInc\GMSU\. Setup.exe will not overwrite the file.

8.4

Gubbs_01.mdb file creation

8.4.1

Original Installation

Setup.exe will create a Gubbs_01.mdb file in C:\GubbsInc\GMSU\Database\.

8.4.2

Upgrade Installation

If Gubbs_01.mdb already exists in C:\GubbsInc\GMSU\Database\, Setup.exe will not overwrite the file.

9 Physical Location of GMSU database Gubbs_01.mdb

If MS SQL Server is to be used as the GMSU database, then this section may be ignored.

In order for GMSU to be used in a multi-user environment, the Gubbs_01.mdb file must be located in a shared directory to which all GMSU users have access. The preferable location for Gubbs_01.mdb is on a network directory, but it can also reside on a local directory that has been shared out to all GMSU users.

If a network directory will be used, it is at this point that the network directory should be defined for the

next installation step Section 10.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

10 Windows Share Permissions and Directory Permissions

Page 11 of 48

The person installing GMSU must be logged onto the workstation with an account that has administrative privileges to the local workstation.

10.1

Local Data Directory

The GMSU setup.exe will create the following GMSU directories.

C:\GubbsInc\GMSU\

C:\GubbsInc\GMSU\Database\ 1,2

C:\GubbsInc\GMSU\Export\ 3

C:\GubbsInc\GMSU\CompoundLists\ 3

C:\GubbsInc\GMSU\ChromatographyMethods\ 4

C:\GubbsInc\GMSU\Manuls\

The folder

The folder

…\Database\ is the location in which the GMSU database is originally stored 1,2 .

GMSU setup.exe places the GMSU database ‘Gubbs_01.mdb’ here.

…\Export\ is the location in which text file reports will be stored (for Sciex HT

The folder

Acquisition Tools).

…\CompoundLists\ is the location in which Automaton™ compound lists are

The folder stored (for Sciex HT Acquisition Tools).

…\ChromatographyMethods\ is the location in which Permeability Calculator

The folder chromatographic methods are stored (if chromatographic methods are generated by the user in an HT environment).

…\Manuals\ is the location in which GMSU Manuals are stored. GMSU setup.exe places current manuals in this directory.

1 If Microsoft Access is to be used as the backend database

2

If MS Access is to be used in a multi-user environment, see Section 10.2

3 Used in a Sciex environment only

4 Used only by the Permeability Calculator module

Notes:

If the GMSU database is to be shared by multiple instances of GMSU, it is HIGHLY

RECOMMENDED that all local installation GMSU directory structures are not modified once created by GMSU setup.exe. The database stores some local directory structure variables and assumes that all local installation directory structures are identical.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

10.2

Shared Network Data Directory (if applicable)

Page 12 of 48

If GMSU is to be configured such that multiple instances of GMSU will share the same database and access the same acquired data, a shared directory must be configured according to the following guidelines. This directory can be a directory on a network file server, or the appropriate local directories of a single data acquisition workstation can be shared. The following instructions assume that a network file server directory will be configured (Thermo users should see

Section 10.3).

Identify a location to which GMSU files will be stored. At this location, generate the following directories:

NOTE: The easiest way to perform this process is to simply copy a local \GubbsInc\... directory set to the network.

\\[network]\GubbsInc\GMSU\

\\[network]\GMSU\Database\ 1

\\[network]\GMSU\Export\ 2,3

\\[network]\GMSU\CompoundLists\ 2,3

\\[network]\GMSU\Data\ 2

\\[network]\GMSU\ChromatographyMethods\ 3,4

The folder

The folder

The folder

The folder

\\…\Database\ is the location in which the GMSU database is stored

\\…\Export\ is the location in which text file reports are stored.

1 .

\\…\CompoundLists\ is the location in which compound lists are stored.

\\…\Data\ is the location in which acquired .wiff files are copied. Note that if a network repository of acquired data already exists, that is acceptable. The acquired data directory location is configured as a variable within the GMSU

Configuration Module (for Sciex environment only).

The folder \\…\ChromatographyMethods\ is the location in which Permeability Calculator chromatographic methods is stored (if a chromatographic method is generated by the user).

1 If Microsoft Access is to be used as the backend database

2 Sciex environment only

3 Optional. If this information will be specific to an instrument, then local directories may be configured in GMSU Configuration (assumes local data directories are configured as

described in Section 10.1).

4 Used only by the Permeability Calculator module

10.2.1

Copy files

At a minimum, copy GMSU_01.mdb to \\[network]\GubbsInc\GMSU\Database\.

10.2.2

Share settings

Share the …\GMSU\ directory and configure GMSU users to have a minimum of

CHANGE share permissions.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

10.2.3

Directory permission settings

Page 13 of 48

Configure directory permissions at the …\GMSU\ and configure the subdirectories to inherit. At a minimum, users must be able to read and write to these directories

(MODIFY permissions).

Note: If the directory to which acquired data is stored is something other than a GMSU directory, users must be able to read and write to this directory (Psuedo HT acquisition chromatogram text files are written to this directory).

10.2.4

GubbsSciex_100 System DSN ODBC file

Remember that the network Gubbs_01.mdb will be used when configuring the MS

Access GubbsSciex_100 System DNS ODBC file described in Section 12.

10.3

Shared Local Data Directory (if applicable)

GMSU is especially useful when it is configured in a manner that all users have access to the raw chromatographic data (.wiff, .raw, .sld). This type of access is most easily attained when the raw data are stored on a centralized network server. However, it is recognized that the Thermo chromatographic raw data file structure mostly is not amenable to manual movement of data to a network directory. In addition, some Sciex administrators may not wish to include a manual raw data copy step in their data reporting process.

Therefore, if it is desired to use GMSU in a manner that all users have access to local chromatographic raw data, administrators must perform the following steps:

10.3.1

Share the raw data directory (e.g. C:\XCalibur\ or D:\Analyst Data\Projects\)

The raw data directory must be shared to GMSU users.

Give the share an intuitive name, such as ‘XCalibur’ or ‘Analyst’

Users must be granted a minimum of CHANGE share permissions.

10.3.2

Configure raw data directory security permissions

The directory security permissions of the raw data directory must be configured such that

GMSU users have a minimum of MODIFY permissions.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

10.3.3

GMSU Configuration Module – Generic Viewer

Page 14 of 48

Configure the appropriate Generic Viewer settings according to the table below.

NOTE: When configuring Default Data Directory, browse to the MICROSOFT

WINDOWS NETWORK to configure the share (see Figure 1).

Setting Value

Default Data Directory With the Browse button, browse to the NETWORK

share of the raw data directory (see Figure 1)

The result would be a UNC path configured as

Default Data Directory, e.g.:

\\gubbs03\XCalibur\ or \\gubbs03\Analyst\

Select Default CDS Environment Thermo(R) ‘for Thermo users’

Sciex(R) ‘for Analyst users’

Use UNC Path Yes

Figure 1 Network browse

11 Database Configuration – Microsoft® Access

If MS Access is to be used as the backend database, follow the instructions in this section.

Setup will place a copy of the Gubbs_01.mdb in the C:\GubbsInc\GMSU\Database\ directory.

Configure an appropriate DSN File according to Section 12.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

12 DSN Configuration – Microsoft® Access

Page 15 of 48

If Setup.exe could not create the GubbsSciex_100 System DSN ODBC file, then the administrator must manually create the DSN file according to this section.

GMSU communicates with the MS Access database via a system DSN named GubbsSciex_100. Create the DSN according to the following procedure:

12.1

ODBC Data Source Administrator

Open the Data Source Administrator and select the System DSN tab

Windows XP: via Control Panel – Administrative Tools – Data Sources

Figure 2 MS Access DNS - Data Source Administrator

12.2

Create New Datasource

Click on the Add… button to start the Create a New Data Source Wizard, select “Driver do

Microsoft Access (*.mdb), and click on Finish

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 16 of 48

Figure 3 MS Access DNS - Create a New Data Source Wizard

12.3

ODBC Setup

At the ODBC Microsoft Access Setup window, enter the Data Source Name and Description as

shown in Figure 4. IMPORTANT: The Data Source Name must be entered exactly as shown.

Figure 4 MS Access DNS - ODBC Microsoft Access Setup

12.4

Select database

Click on the Select… button, navigate to the Gubbs_01.mdb, and click on OK

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 17 of 48

Figure 5 MS Access DNS - Select Database

Click on the OK button to complete the DSN configuration process.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

13 Database Configuration – Microsoft® SQL Server

Page 18 of 48

If Microsoft SQL Server is to be used as the backend database, follow the instructions in this section.

13.1

Create a SQL Server database using the MS Access Upsizing Wizard

Setup will place a copy of the Gubbs_01.mdb in the [drive]:\Program Files\Gubbs Mass Spec

Utilities\ directory (or installation location configured by the user).

Open this file in MS Access and run the Upsizing Wizard

Figure 6 MS Access Upsizing Wizard

The following figures show the parameters to choose when performing the Upsize action

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 19 of 48

Figure 7 MS Access Upsizing Wizard - Choose Create new database

Figure 8 MS Access Upsizing Wizard - Connection settings

Here, enter a login ID and password of an account with sufficient permissions to create a database

(e.g. sa). A GMSU user account will be generated at a later step. The database name can be of your choosing.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 20 of 48

Figure 9 MS Access Upsizing Wizard - Export all tables

Figure 10 MS Access Upsizing Wizard – Export attributes

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 21 of 48

Figure 11 MS Access Upsizing Wizard – No application changes

Figure 12 MS Access Upsizing Wizard - Finish

At this point the Upsizing Wizard will generate a report which may be discarded.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

13.2

Create a GMSU database account

Page 22 of 48

In SQL Server, create a GMSU database account. The account name and password can be of your choosing.

Add the account to the GMSU database and assign the account the following roles:

Figure 13 MS SQL Server GMSU account roles

13.3

Record the user account and password information

Record the user account and password information on a sheet of paper. The user account and

password information is needed to update the GMSU_xx.ini file described in Section 15.

13.4

Local Data Source Name (DSN) Gubbs_100

GMSU communicates with the SQL Server database via a system DSN. Create the DSN

according to the procedure described in Section 14.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

14 DSN Configuration – Microsoft® SQL Server

Page 23 of 48

If MS SQL Server is to be used as the backend database, follow the instructions in this section.

If Setup.exe could not create the GubbsSciex_100 System DSN ODBC file, then the administrator must manually create the DSN file according to this entire section.

If Setup.exe was successful in creating the GubbsSciex_100 System DNS ODBC file, then the

administrator must complete the configuration starting with Step 14.3 of this section.

GMSU communicates with the SQL Server database via a system DSN named GubbsSciex_100. Create the DSN according to the following procedure:

14.1

ODBC Data Source Administrator

Open the Data Source Administrator and select the System DSN tab

Windows XP: via Control Panel – Administrative Tools – Data Sources

Figure 14 SQL Server DNS - Data Source Administrator

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

14.2

Create new data source

Page 24 of 48

Click on the Add… button to start the Create a New Data Source Wizard, select “SQL Server”, and click on Finish

Figure 15 SQL Server DNS - Create a New Data Source Wizard

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

14.3

Create a new data source

Page 25 of 48

At the Create a new data source window, enter ‘GubbsSciex_100’ as Name and Description as shown. Choose the server on which the GMSU SQL Server database was generated, then click

Next

IMPORTANT: The Data Source Name must be entered exactly as shown.

Figure 16 SQL Server DNS - Create a new data source

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

14.4

Enter connection parameters

Page 26 of 48

Configure the next window as shown. Enter the userid and password combination created in

Section 13.2. Then click Next.

Figure 17 SQL Server DNS - Enter connection parameters

14.5

Additional parameters

Enter additional parameters as shown in the next two figures, then click Next and Finish:

NOTE: Ensure that the default database is the GMSU database.

Figure 18 SQL Server DNS - Enter additional parameters

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 27 of 48

Figure 19 SQL Server DNS - Enter additional parameters

14.6

Test connection

Test the connection by clicking on the ‘Test Data Source’ button

Figure 20 SQL Server DNS - Test connection

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 28 of 48

Figure 21 SQL Server DNS - Test connection

If the last line does note read “TESTS COMPLETED SUCCESSFULLY”, then contact your

GMSU administrator before proceeding.

Click on the OK button to complete the DSN configuration process.

15 GMSU_xx.ini

Setup will create an .ini file (named GMSU_xx where ‘xx’ is a number) located in the directory

C:\GubbsInc\GMSU\.

15.1

MS Access GMSU_xx.ini configuration

The default .ini content is for a MS Access installation. Please note that the UserID and Pswd entries of the .ini file may be ignored. Ensure that the .ini file has the following content: boolAccess boolSQLServer

True

False

UserID

Pswd

[USERID]

[PASSWORD]

Note: UserID is ignored

Pswd is ignored

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

15.2

SQL Server GMSU_xx.ini configuration

Page 29 of 48

If SQL Server is to be used as database, ensure Ensure that the .ini file has the following content: boolAccess boolSQLServer

False

True

UserID

Pswd

[USERID]

[PASSWORD]

Note:

[USERID] is the SQL Server account generated in Section 13.2.

[PASSWORD] is the password of the SQL Server account generated in Section 13.2

IMPORTANT: The Gubbs Inc Password Encryption Tool described in Section 16 MUST be used

to encrypt the password. Copy the encrypted password form the Password Encryption Tool into the Pswd portion of the .ini file.

16 Gubbs Inc Password Encryption Tool

Setup will place a file named GubbsIncPswd_01.xls in the installation directory. Open this file to display the Gubbs Inc Password Encryption Tool. Copy the encrypted password to the .ini file if necessary.

17 GMSU Database Update

This utility should be run when GMSU is upgraded from a previous version. If GMSU database is a shared database or MS SQL Server database, the GMSU Database Update utility needs to executed only once. The utility does not need to be run when the system is initially installed and configured.

NOTE: Ensure that Excel Macro Security (Tools – Macro – Security) is set to Medium. If Macro Security is set to High, then the UpdateGMSUDatabase utility will not run.

Setup will create a file called UpdateGMSUDatabase.xls located in the installation directory. Open the file to display the GMSU Database Update Utility.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 30 of 48

Figure 22 GMSU Database Update Utility - First time use screen

At the displayed window, click on Execute Update.

Figure 23 GMSU Database Update Utility – After Execute

If the Notes report that the GMSU technical representative should be contacted, please contact Gubbs Inc.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

18 GMSU Configuration Default Parameters

Page 31 of 48

Please note that this section may be ignored if the GMSU installation is an upgrade to an existing GMSU installation.

Several GMSU configuration parameters must be configured in order that GMSU functions properly upon initial use. These parameters are defined in the following sections of the Administration Manual.

Note: If the GMSU Module will not be used, then the configuration section may be skipped.

GMSU Module Configuration Section

Caco (for Permeability Calculator Module)

Administration Manual

Reference

Section 20.1

Generic Viewer (for Generic Chromatographic Viewer Module)

Global

Hepatic Clearance

Section 20.6

Section 20.7

Section 20.8

19 Console

The GMSU Console is the module from which users navigate to other modules. The Console contents may differ slightly from what is shown.

Figure 24 GMSU Console

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 32 of 48

Users click on the appropriate buttons either to open a utility or to navigate to an additional hierarchical level.

20 Configuration Utility

Administrators use the Configuration Utility to configure global variables and contents of module dropdown lists. Users choose the appropriate module to configure by choosing an item in the ‘Choose a

Category’ list.

The initial view of the Configuration Utility is a non-editable mode. In order to edit a value, the user must select the value, then click the Edit button. The user will be prompted to enter a new value, after which the user will click the Save button.

20.1

Caco - Global

The Caco – Global window contains default paths and values specific to the Permeability

Calculator. The parameters must be configured in order for proper GMSU operation follow the figure

Figure 25 Configuration Utility – Caco (Permeability Calculator) – Global

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.1.1

Initial use parameter configuration

Page 33 of 48

The following are parameters that must be configured upon initial GMSU installation.

Note: If this module will not be used, then this section may be ignored.

20.1.1.1

Default Method Chromatography Path

This directory is where Permeability Chromatographic Method text files

(for HT Acquisition) are stored.

This directory is created by GMSU Setup.exe (see Section 10.1):

C:\GubbsInc\GMSU\ChromatographyMethods\

20.1.1.2

Default Data Directory

This directory is the root directory of where raw data is stored. This directory is shown initially when the Browse… button is clicked in the

Permeability Calculator module.

20.1.1.3

Show Low Throughput/High Throughput Option Window

Yes/No to view the LT/HT option window in the Permeability Calculator module.

If both LT and HT acquisition data (see Section 3 for a description of LT

and HT acquisition) will be viewed by users, then set this value to Yes.

If not, set this value to No

20.1.1.4

LT or HT Data Acquisition

Set value to LT if only LT acquisition data will be viewed.

Set value to HT if only HT acquisition data will be viewed.

 This value is ignored if ‘Show Low Throughput/High Throughput

Option Window’ is set to Yes

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.2

Caco – Analysis Time Points

Page 34 of 48

The Analysis Time Points window is used to create Time Point Sets that will be assigned to analysis data files. ‘Time Points’ are defined as the assay time points in minutes.

Users click on the Edit button to configure a new Time Point Set. Users click on the Save button to actually create the Time Point Set. Once a Time Point Set has been created, it cannot be modified.

Figure 26 Configuration Utility – Caco (Permeability Calculator) – Analysis Time Points

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.3

Caco – Chromatography Peak Sets

Page 35 of 48

The Chromatography Peak Sets window is used to create Chromatography Peak Sets that will be assigned to analysis data files. These sets provide identities of the chromatographic peaks of the acquired data.

Users click on the Edit button to edit an existing Chromatography Peak Set or configure a new

Chromatography Peak Set.

Peak identities are assigned by selecting the appropriate Peak Name in the Peak Name list, then double-clicking the appropriate row in the

Users click on the Create New Set button to configure a new Chromatography Peak Set. Users click on the Execute button to actually create the Chromatography Peak Set, then click the Save button to save the set.

Figure 27 Configuration Utility – Caco (Permeability Calculator) – Chromatography Peak Sets

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.4

Caco – Constants

Page 36 of 48

The Constants window is used to create Constants Sets that will be assigned to analysis data files.

These sets include the various constants used in the calculation of Permeability Experiments.

Users click on the Edit button then Create New Set button to configure a new Constants Set.

Users click on the Save button to actually create the Constants Set. Once a Constants Set has been created, it cannot be modified.

Figure 28 Configuration Utility – Caco (Permeability Calculator) – Constants

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.5

Flags

Page 37 of 48

The Flags window configures the contents of the ‘Chromatogr. Review’ dropdown box of the

Hepatic Clearance Calculator. ‘Results Flags’ are not configurable.

Figure 29 Configuration Utility - Flags

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.6

Generic Viewer

Page 38 of 48

Figure 30 Configuration Utility – Generic Viewer

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 39 of 48

The Generic Viewer window contains global parameters concerning default directory paths of system components.

20.6.1

Initial use parameter configuration

The following are parameters that must be configured upon initial GMSU installation.

Note: If this module will not be used, then this section may be ignored.

20.6.1.1

Default Data Directory

This directory is the root directory of where raw data is stored. This directory is shown initially when the Browse… button is clicked in the

Generic Chromatographic Viewer module.

20.6.1.2

Select Default CDS Environment

Used only if users will be viewing both Thermo and Sciex data. This setting governs the order in which ‘File Type Filter’ are displayed when the

Browse… button is clicked.

 If the setting is ‘Thermo(R)’, then the default choice is ‘.sld’.

 If the setting is ‘ABI(R)’, then the default choice is ‘.wiff’.

20.6.1.3

Use UNC Path

If it is desired that raw data be viewed directly from the data acquisition workstation, then this value must be set to Yes.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.7

Global

Page 40 of 48

The Global window contains global variables concerning default directory paths of system components.

Figure 31 Configuration Utility - Global

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 41 of 48

The Global window contains global parameters concerning default directory paths of system components

20.7.1

Initial use parameter configuration

The following are parameters that must be configured upon initial GMSU installation.

20.7.1.1

Analyst Acquisition Data Path

Note: This parameter is related to Sciex HT Acquisition Tools. If these tools will not be used, then this section may be ignored.

This directory is the root directory of where Analyst raw data is stored.

20.7.1.2

Assay Request File Path

Note: This parameter is related to Sciex HT Acquisition Tools. If these tools will not be used, then this section may be ignored.

This directory is the root directory of where Assay Request Files (or may be referred to as Compound Lists) are stored. The Compound List is essentially the text file used by Automaton™.

This directory is generated by the GMSU Setup.exe (see Section 10.1):

C:\GubbsInc\GMSU\CompoundLists\

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.8

Hepatic Clearance Calculator

Page 42 of 48

The Hepatic Clearance Calculator window contains default paths and values specific to the

Hepatic Clearance Calculator Utility. Such values include labels for the Species and Results sections.

Figure 32 Configuration Utility – Hepatic Clearance Calculator

20.8.1

Initial use parameter configuration

The following are parameters that must be configured upon initial GMSU installation.

Note: If this module will not be used, then this section may be ignored.

20.8.1.1

Default Data Directory

This directory is the root directory of where raw data is stored. This directory is shown initially when the Browse… button is clicked in the

Permeability Calculator module.

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.9

Hepatic Clearance Calc. Expt. Constants – Existing Experiments Tab

Page 43 of 48

Administrators use this window to configure species- and experiment-specific constants that are used during Hepatic Clearance value calculations. The contents of the Species dropdown boxes in

the Hepatic Clearance Calculator are obtained from the Experiment list shown in Figure 33.

As an aid to administrators when configuring new Experiments, users may copy data from one experiment to another.

To add additional experiments, use the Add Species/Type tab (see Section 20.10).

Figure 33 Configuration Utility – Hepatic Clearance Calc. Expt. Constants – Existing Experiments

Tab

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.10

Hepatic Clearance Calc. Expt. Constants – Add Species/Type Tab

Page 44 of 48

Here users add species and/or types. Once added, new experiments will be generated automatically.

Figure 34 Configuration Utility – Hepatic Clearance Calc. Expt. Constants – Add Species/Type Tab

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.11

Hepatic Clearance Calc. Expt. Constants – Experiment Time Points Tab

Page 45 of 48

The Analysis Time Points window is used to create Time Point Sets that will be assigned to analysis data files. ‘Time Points’ are defined as each chromatographic peak (for HT acquisition) or each chromatographic injection (for LT acquisition).

Users click on the Edit button to configure a new Time Point Set. Users click on the Execute button to actually create the Time Point Set and the Save button to save the set. Once a Time

Point Set has been created, it cannot be modified.

In order to aid internal information technology (IT) department efforts to generate internal reports or transfer data to internal laboratory information management systems (LIMS), users may label peaks not used in actual regression analysis that are needed for additional internal process control information.

Figure 35 Configuration Utility – Hepatic Clearance Calc. Expt. Constants – Experiment Time Points

Tab - Initial

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 46 of 48

Figure 36 Configuration Utility – Hepatic Clearance Calc. Expt. Constants – Experiment Time Points

Tab - Edit

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual

20.12

Modify Method

Page 47 of 48

The Modify Method window contains the contents of the LC Sync dropdown box of the Modify

Methods Utility. These contents are specific to Sciex instrumentation at this point in time.

Figure 37 Configuration Utility – Modify Method

Version 2.0 (June 4, 2007)

D:\726884342.doc

®

Gubbs Mass Spec Utilities 3.x.x Administration and Installation Manual Page 48 of 48

21 References

1 R. O. Cole. Quantitation of Biofluids with High Speed HPLC/MS/MS: Early Discovery Applications.

Invited Presentation, at “Early ADME and Toxicology in Drug Discovery: Techniques for accelerating and Optimizing Drug Candidate Selection”, Berkley, CA, October 1998.

2 Adam H. Brockman, Donna L. Hiller, and Roderic O. Cole. High Speed HPLC/MS/MS Analysis of

Biological Fluids in Support of ADME Screens: A Practical Review. Current Opinion in Drug Discovery,

4(3), (2000), 432.

3 Janiszewski JS, Rogers KJ, Whalen KM, Cole MJ, Liston TE, Duchoslav E, et al: A high capacity

LC/MS system for the bioanalysis of samples generated from plate-based metabolic screening. Anal

Chem 2001; 73:1495-1501.

4 Z. Yan, J. Wu, L.E. Elvebak, A. Brockman: Validation of a Totally Commercially Available High

Throughput -ADME System and Results for 60 Literature Compounds. Rapid Communications in Mass

Spectrometry, Rapid Commun. Mass Spectrom. 2005; 19: 1191–1199.

Version 2.0 (June 4, 2007)

D:\726884342.doc

Download