ANALYZE software manual During practical classes, a custom-made software (ANALYZE) will be used to measure and visualize different physiological variables as well as to analyze effects induced by experimental manipulations. Physiological variables are transformed into electric signals by appropriate sensors (strain gauge, blood pressure probe, etc.) connected to the amplifier. Amplified (analog) signals are digitalized by an analog-to-digital converter card (A/D converter) installed into the computer. Digitalization means that the converter measures the input voltage at preset intervals (sampling time or sampling rate) and passes the value in numeric form to the computer. The A/D converter we use, generates 12-bit binary numbers between -2048 and 2047 (-212 – 212-1). ANALYZE program consists of a measuring and an analysis module. To switch from one module to the other, the “Analyze” or “Measure” menu points should be selected, depending on which module is active. The software is able to measure and analyze three signals in parallel. While many different physiological parameters can be monitored by the program, their measurement (and analysis) fits into one of two basic types. In continuous mode, input signals are digitalized at the preset frequency and the resulting numeric values are stored and displayed on the screen in parallel with measurement, in real-time. This mode emulates an ink-writer, that records signals on a paper strip. Maximal sampling rate depends on the number of measured signals and on computer performance. Considerably higher sampling rate can be achieved in block mode. Measurement of input signals is started by user command. The time-consuming display of the acquired data is postponed until the preset number of data points is taken. Thus, measurement is not continuous, it occurs in separate blocks. During analysis, data acquired in blocks can be treated as continuous and vice versa. In this way, special features provided by the two types of analysis can be used independently of the method by which the data was collected. All necessary information for using the program is provided on the pages below describing user screens and menu commands for continuous and block modes during acquisition and analysis. Before hints and special information to complete the first experiment, “Calibration” are given, a short overview of the connections between the amplifier and computer follows. Input “BNC” sockets (1-8) of the A/D converter are found on the top of a small black box, which is connected to the converter installed in the computer through a gray flat cable. These input sockets should be connected to the outputs of the amplifier. The output (“jack”) sockets are located on the rear panel of the amplifier (SG – strain gauge, etc.). For the sake of simplicity, if one signal is measured it should be connected to the first socket. In the case of two input signals, the first two sockets should be used, and so on. Experiment 1. Calibration The goal of this experiment to get familiar with the components of the data acquisition system: the displacement sensor (strain gauge), the amplifier and the software. a/ In continuous measurement mode click on the downward arrow of the first (white) channel to activate the drop-down list and select the first input socket (1.ch) to which the strain gauge (SG) output of the amplifier is connected. Gain should be left at the default value (1x). Three signals can be measured simultaneously by the program, but there are only one set of buttons to control the position and scaling of the three curves (white, yellow, blue) on the display. On the right side of the display, there are three small, color-coded buttons to assign the controls to one of the three signals. The white one should be selected. Assignment of the buttons to another signal is indicated by a change in the color of controls and y-axis. b/ Set scaling on y-axis to the maximal useful range (-2000 – +2000). Sampling rate should be 50 Hz, screen length 30 s. Start sampling and displaying of the signal by pressing “Monitor” button. c/ Move the curve to the bottom of the display by turning the offset dial on the amplifier. Make sure that the curve can still move a little bit downward if you carefully touch the lever of the displacement sensor with your finger. If amplification or offset is too large then the curve is locked to the upper or lower limit and is insensitive to the displacement of the sensor. If this phenomenon is caused by inappropriate scaling of the y-axis, then data stored into the disk file is still usable. However, if the amplified signal at the input of the A/D converter exceeds ±10 V, which is the maximal value that the converter can pass to the computer as ±2048, the measurement should be repeated. d/ Before the actual experiment, determine which is the largest weight that does not cause locking of the curve to the upper limit (2048) of the display. Please note that the curve on the display is displaced upward when the weight is pulling the lever of the sensor downward. During experiments in the following weeks, always test the relationship between the displacement of the lever and the curve on the display by carefully moving the lever of the strain gauge. This information is crucial for the evaluation of your observations. Finding out if the frog’s heart stopped in contracted or relaxed state, has a large impact on interpretation. e/ After the largest weight to be used has been determined, arrange the weight in ascending order and open a data file by pressing the “Record” button. It is advisable to give the files unique, meaningful names. You do not have to specify the extension of the file (the part after the “.” character), the program will set it automatically to “.bin”. File length should be the default value of 180 s. Let’s recording and displaying baseline data for a few seconds (10-20), then put on the first weight carefully. Wait until the transients are over, record the displacement caused by the weight for a few seconds, then remove the weight. After a short baseline recording, put on the next weight and repeat the procedure until the largest weight has been measured. Close the data file by pressing the “Stop” button. f./ Select “Analyze” from the menu and open the data file (“File/Open”). If you press the left mouse button over the blue display area within the range of the curve, a cursor line appears. Position the cursor over the first baseline period. Press “Add” button. You are prompted for a label by the program. The label will be transferred together with the mean and standard deviation values to MS Word, where the report can be written. Give the value of the first weight as label (e.g. “5 g”). Move the cursor to the period where the effect of the first weight is seen and press “Add” again. Go back to the baseline and repeat at least three times the measurements. Note that the “Add” button is always selected (a dotted frame within the button marks the selection), thus pressing the “Enter” on the keyboard has the same effect as clicking on the button with the mouse. Therefore, it is easier and faster to use the mouse only to position the cursor, while activating the “Add” button from the keyboard. In this way, there is no need to repeatedly move the mouse pointer between the curve and the “Add” button. When three data pairs have been measured and entered into the table at the bottom of the screen, press the “Newblk” button to terminate analysis of the first weight’s effect. Mean and standard deviation is automatically calculated from the data. g./ Repeat these measurements for all weights. When ready, put labels on the curve above each displacement, indicating the values of the weights. This can be done by activating the “Text/Insert” menu point. A new textbox will appear in the top left corner. After typing the text, the box can be dragged to the appropriate position with the mouse. Labels can be reselected with the mouse and edited if necessary. Empty textboxes are automatically deleted. The “Text/Delete” menu point removes the selected label. When the figure is ready, select the “Copy/Picture” menu point to copy it to the clipboard. Locate and launch MS Word. Paste the figure into a new document, and switch back to Analyze. Make sure that when you switch between the running applications, you do not restart a new instance of programs. Use the “Alt-Tab” key sequence, or another method. Launching several copies of the same program, especially Analyze can crash the system resulting in possible loss of data. Select the “Copy/Block means” menu point in Analyze to copy mean and standard deviation values to the clipboard. Paste these values into the same document in Word and then everything is ready for the preparation of the report. h./ Using weight values and mean displacement data, a calibration curve can be constructed, preferably in MS Excel and inserted into the report. Make sure that you select “Scatter” diagram in Excel and not “Line”, otherwise weight values will be ignored and displacement values will be evenly distributes along the x-axis. Menu points and settings in continuous measurement: File Exit Display Block mode Analyze terminate program no menu point can be selected change to block mode, specifying block length switch to analysis module increase amplitude move curve upward select between channels display mode (read only) move curve upward decrease amplitude monitor w/o storing measure and store marker on/off stop measurement give stimulus select channel to measure sampling frequency gain of the A/D converter displayed epoch maximal stored length Menu points and settings in block measurement: File Display Exit terminate program Single/Average display single curve with white/yellow/cyan color and average with dark gray in the background Average/Single display averaged curve with white/yellow/cyan color and single with dark gray in the background Single display only single curves with white/yellow/cyan color Average display only averaged curves with white/yellow/cyan color Block length change block length Continuous change to continuous mode Analyze switch to analysis module add single curve to average clear average save last single curve into the previously opened data file open data file stop acquisition, close data file number of averaged curves delay to first stimulus interval between stimuli number of stimuli to deliver Menu points of analysis in continuous mode: File Open Exit open data file terminate program Copy Picture Block means Whole table (Excel) Whole table (Word) Curve points copy figure to clipboard copy statistical data (mean, standard deviation, n) to clipboard copy individual cells of the table to the clipboard in Excel compatible form copy individual cells of the table to the clipboard in Word compatible form copy curve points to the clipboard Zoom Expand Compress Zoom selection displayed curves are expanded (x-axis range is decreased to half) displayed curves are compressed (x-axis range is doubled) epoch selected with the right mouse button is expanded to fill the display Display no menu point can be selected Invert invert displayed curve (multiply by –1) Marks off clears cursors positioned with the right mouse button Text Insert Delete insert label to the figure delete label from the figure Auto peaks automatic location of peaks in the epoch selected with the right mouse button taking into account lower and upper limits set with the mouse pointer Block mode change to block mode specifying block length Measure switch to continuous measurement mode User screen of analysis in continuous mode: step backward half screen step forward half screen put selected data point into the table clear table close analysis epoch delete last line in table x1 y1 x2 y2 ampl. tone period time of the first extreme value (peak or trough) amplitude of the peak or trough time of the next, opposite extreme value (trough or peak) amplitude of the trough or peak absolute difference of the two amplitudes (peak-to-peak amplitude) average of the two amplitudes, i.e. middle point of the wave time from the previous x1, i.e. distance between the two waves Menu points of analysis in block mode: File Open Exit open data file terminate program Copy Picture Whole table (Excel) Whole table (Word) Curve points copy figure to clipboard copy individual cells of the table to the clipboard in Excel compatible form copy individual cells of the table to the clipboard in Word compatible form copy curve points to the clipboard Zoom Expand Compress Zoom selection displayed curves are expanded (x-axis range is decreased to half) displayed curves are compressed (x-axis range is doubled) epoch selected with the right mouse button is expanded to fill the display Display Single/Average Average/Single Single Average Overplot Raw curves PSD display single curve with white/yellow/cyan color and average with dark gray in the background display averaged curve with white/yellow/cyan color and single with dark gray in the background display only single curves with white/yellow/cyan color display only averaged curves with white/yellow/cyan color display blocks under each other (waterfall) curves are displayed as measured, in the time frame Power Spectrum Density) histograms of the curves are displayed Invert invert displayed curve (multiply by –1) Marks off clears cursors positioned with the right mouse button Text Insert Delete insert label to the figure delete label from the figure Block length change block length Continuous change to continuous mode Measure switch to continuous measurement mode User screen of analysis in block mode: step back one block step forward one block t A t-dly p-to-p dist time of the extreme value (peak or trough) from the beginning of the curve amplitude of the peak or trough calculated from the baseline (prestimulus period) time of the peak or trough from the stimulus (delay) signed amplitude difference of the peak or trough from the previous extreme value (A2-A1) time from the previous extreme value (t2-t1)