Page 1 of 8 Clinical AlarmsPatient Safety Policy MANUAL Care of the Patient SECTION Patient Safety POLICY CODE and NUMBER PS-6 PURPOSE: The purpose of this policy is to appropriately adjust and respond to all clinical alarms to ensure patient safety. POLICY: 1. 2. 3. 4. 5. 6. 7. 8. All clinical alarms will remain active at all times and will be documented in EMR . It is the responsibility of all clinicians in this institution to respond immediately to all clinical alarms. Clinicians will perform corrective action within their scope of practice to resolve the problem. Failure to respond immediately to clinical alarms can be life threatening; therefore failure to comply with this policy may result in termination of employment. Pulse oximetry is a noninvasive method of monitoring the oxygen saturation of arterial blood. This policy is to assure arterial oxygenation for all patients whether breathing spontaneously or on mechanical support. It is indicated for the following clinical situations a)During continuous mechanical ventilation, b)During ventilator or supplemental oxygen weaning or airway decannulation trials, c) During exercise and physical therapy on patients receiving supplemental oxygen, d) During transport of patients on mechanical ventilation or supplemental oxygen, e) During cardiac arrest, f)During bronchoscopy, g)To document the appropriateness of supplemental oxygen, h) and other invasive procedures as directed. This is monitored by Respiratory on a monthly basis. The Medical Gas alarms/bulk oxygen pressure alarm is designed to visually and audibly alert hospital staff to an irregularity in the system. a) The alarm will activate at 40 psig (Low Pressure) or 60 psig (High Pressure). b) This policy is applicable to all maintenance and plant operations personnel who have responsibility in operation and maintenance of the medical gas system. c) This is monitored by Maintenance on a quarterly basis and Respiratory internally on a monthly basis. The Ventilator Support System (invasive and non-invasive) provides respiratory support for a wide range of adult patients for a wide variety of clinical conditions. a) The ventilator’s mixing technique allows to ventilate critically ill patients. b) All ventilator alarms will remain active at all times and will be documented in EMR ventilator assessment c) This is monitored by Respiratory and Maintenance on a monthly basis. The Enteral Pump System is designed to deliver only a liquid enteral feeding product. a) It is a micro-processor controlled pump that provides accurate, controlled enteral feeding. b) The pump used is a volumetric pump that uses a specially designed feeding set, which includes a cassette that delivers measured amounts of enteral products. c) The monitoring by the manufacturer consist of flow calibration and alarms monitored by Nursing. d) This is monitored monthly by Maintenance and Nursing daily. Clinical AlarmsPatient Safety Policy Page 2 of 8 9. The Intravenous Pump System is designed 10. The Telemetry System is intended for use in the ICU type of areas. a) It is easy to operate with analog touch screen, high-resolution display review function such as graphic trends. b) Up to 16 patients can be monitored simultaneously from other areas of the hospital. c) As integrated real-time patient data through bedside monitors of the telemetry receiver. d) This is monitored by Monitor Technicians, ICU Nurses, and Respiratory and is part of EOC Medical Equipment Plan. 11. The Fall Prevention System currently at this facility is the RN+ System. a) It is a wireless system used primarily in hospitals to monitor and detect patient movement and alert nurse to possible unassisted exits. b) The RN+ utilities a bandwidth system. c) Each employee will prioritize and assist with the prevention of patient falls and responds to alarms immediately. 12. The Nurse Call System is designed as an integrating nurse call system with an in-building wireless telephone system that can produce tremendous improvements in staff efficiency and thereby ultimately improve response times to patients. This integrated health care communications products enables caregivers to provide the best patient-focused care possible. a) This system has the ability for a handset delivery communication directly and indirectly with the use of text paging. b) This system also has a management software for tracking and organizing patient and staff activity. PROCEDURE: 1. Pulse oximetry: a. b. c. d. e. f. g. h. Plug pulse oximeter monitors power cables into a red electrical outlet and turn unit on. Follow manufacturer's recommendations for appropriate setup. Select appropriate monitoring site and secure probe on patient. When using continuous pulse oximetry proper alarms setting should be set with HR low at 50 and HR high at 150, Sat O2 low at 85%, unless otherwise noted by physician. Portable or bedside pulse oximetry may be used Pulse oximetry will be monitored Q 12 hours on non-ventilated patients and with Q 2-hour ventilatory checks. Alarm settings and function along with site location will be documented during Q 12-hour ventilator assessment. Results and alarm settings will be charted in EMR. Clinical AlarmsPatient Safety Policy 2. Page 3 of Medical Gas alarms/bulk oxygen pressure alarm The responding technician shall verify that the alarm is a result of the system being switched to the standby sources due to the primary source being exhausted. If the above occurs during others then normal duty hours and the reserve supply is adequate until the following morning, there will be no need to call out a technician to replace the tanks. Upon notification by PBX operator of any of the following alarm conditions a. b. c. d. Oxygen pressure high Oxygen Pressure low Medical air pressure high Medical air pressure low During normal working hours, the responding engineer shall verify the alarm condition, attempt to determine the malfunction and take appropriate action to repair. During other than normal working hours, the responding maintenance person shall verify the alarm condition attempt to determine the cause of the malfunction and notify the appropriate manager or contractor, who shall respond accordingly. If the alarm is determined not to be a system malfunction, but an alarm condition that can wait until the following morning, a contractor does not need to be called out. BULK OXYGEN ALARMS (liquid oxygen) - The bulk oxygen level alarm is designed to visually and audibly alert hospital staff to an irregularity in the system. a. b. c. d. Locate all patients on mechanical ventilation and insure proper function of ventilator. Utilize manual resuscitator to ventilate patient if ventilator is inoperable. Silence the alarm by depressing the alarm silence button on the alarm system control panel. Read the alarm indicator directly below the liquid oxygen level. If the indicator indicates “Emergency Reserve In Use”, notify Plant Engineer, Respiratory Care, Nursing Manager, Administration, and the oxygen supply company IMMEDIATELY. Contact area hospitals for additional cylinders of oxygen if the Emergency Reserve is in use and it is known the supplier cannot refill in adequate time. e. Read pressure and inches of water gauges on bulk oxygen system. e. Check and record readings for the bulk system emergency reservoirs or other back up system. f. Check line pressure downstream of zone valves located throughout the hospital. g. Notify House Supervisor, Respiratory Director Administration, Plant Engineer. Report all current readings. h. Call the oxygen supply company. Inform them of the current alarm status. i. Use cylinder oxygen for resuscitators. j. Assess all other patients receiving oxygen. Provide oxygen via cylinders until problem is corrected. k. Reassure patients. l. Complete Incident Report Form. Note: After the problem has been corrected, place patients back on the mechanical ventilator. Check for proper ventilator and alarm function. Clinical AlarmsPatient Safety Policy 3. Page 4 of Mechanical Ventilators (invasive) Alarm settings will be set as follows: a. b. c. d. e. High-Pressure Alarm: Set 10-20 cm H2O above the highest documented peak inspiratory pressure in the past twelve hours. Low-Pressure Alarm: Set 10-20 cm H2O below the highest documented peak inspiratory pressure in past twelve hours. High-Rate Alarm: Set at 10 breaths per minute above the patient's average stable rate. This high-rate alarm will not be set higher than 40 breaths per minute unless otherwise indicated by the patient's condition. Low tidal volume: 100-200cc below the patient’s ordered Low expired minutes ventilation is to be set 20% below last documented VE If Respiratory Therapist finds it necessary to set alarms outside these parameters, this is documented and the reason for the change each shift. These parameters need to be re-evaluated every shift and readjusted according to patient’s condition as well. BIPAP Vision (non-invasive) Alarm setting will be as follows: 1. 2. 3. 4. 5. High pressure limit (PIP) is to be set 50% above highest documented PIP in past twelve hours not to exceed 10em H2O below. Low-pressure delay is to be set at ten seconds. Apnea – 20 or less seconds delay. Low expiratory minute ventilation is to be set 20% below last documented VE. High respiratory rate is to be set 10-15 above highest RR documented. Listed below are the most frequent alarms of ventilators, probable cause and corrective actions needed to resolve alarms. MANUALLY RESUSCITATE PATIENT UNTIL CAUSE OF ALARM IS RESOLVED. Nursing personnel must notify Respiratory Care personnel of alarm condition. Clinical AlarmsPatient Safety Policy Page 5 of 6. Actions Required for the following: ALARMS Low Pressure High Pressure Limit PROBABLE CAUSE Ventilator Circuit disconnected CORRECTIVE ACTION Begin at patient end of circuit and scan for disconnects. Leak in ventilator circuit Repeat above procedure Cuff Leak Check cuff for proper inflation pressure. Obstruction in patient airway Clear obstruction (suction). Obstruction in ventilator tubing Clear obstruction (drain tubing) Kinked ventilator tubing Straighten ventilator tubing Change in patient status Re-evaluate patient. Re-position artificial airway. Low expired volume minute Artificial airway misplacement Same as low pressure High Expired volume minute Change in patient status Re-evaluate patient Apnea Change in patient status Re-evaluate patient High Rate Change in patient status Re-evaluate patient Air Supply Inadequate compressed air supply to ventilator Replace ventilator compressor. Provide compressed air via cylinders. Notify R.C. Manager. Oxygen Supply Inadequate oxygen supply to ventilator Check oxygen piping pressure and supply. Provide O2 via cylinders. Notify R.C. Manger. Loss of Power Loss of electrical power Insure ventilator connected to emergency electrical outlet. Vent inoperative Microprocessor failure Manual resuscitate patient EST to be done by RT. If necessary replace vent. Same as low pressure Clinical AlarmsPatient Safety Policy 4. Page 6 of Enteral Pump System Intermittent audio alarms are accompanied by messages on the front panel indicating cause of the alarm for all alarm conditions except LOW AUDIO (visual only), which indicates low alarm volume has been selected. Actions Required to be followed: Visual Display Occlusion – flow has stopped Corrective Action clear obstruction, check for kinked tubing, then run Empty – feed or flush bag empty Dose Complete – volume fed is equal to the set dose. Pump stops Set Rate – dial turned to run without entering rate Turn to Run – pump dial in setting other than run or off more than five minutes Door/Cassette – door open, cassette not properly loaded or excess foam or air in cassette Low Battery – pump has approximately 30 minutes of battery operation before complete shutdown Turn dial to hold; refill container; turn to run Turn dial to set dose; press keypad arrows to set dose; to silence alarm, turn to hold, then run. Select flow rate (1 to 300 ml/hr); to slience alarm, turn to hold, then run Turn the dial to hold or run to reset ERR-system malfunction Flush Display Blank with Feed/Flush Set in use – clamp is closed on flush bag 5. Intravenous Pump System Confirm cassette in place and close door; manually reprime to remove foam or air; replace set; if alarm persist, requires service Complete shutdown is preceded by a continuous alarm for one minute; plug power cord into AC outlet, pump will operate normally while battery is charging Must turn pump off to reset, if persist requires service Turn pump off, fill flush bag, open clamp, restart Clinical AlarmsPatient Safety Policy 6. Page 7 of Telemetry System The telemetry system incorporates a priority alarm system with different notifications for different levels of alarm priorities. Levels of alarm priorities ranges from system failure, to life threatening, to cautionary, to treatment needed, and message alarms. Individual parameter alarms consist of ECG, respiration, blood pressure, temperature, and oxygen saturations, among other which currently are not used in this facility. When setting an alarm at the central station, the network function will set the bedside monitor for the same configurations. This applies to ON/OFF/SUSPEND. When the system alarms are on and the alarm factor occurs the message will be displayed and the alarm sound occurs. The measured values for each parameter are displayed on the individual display. The alarm limits are set as follows: Actions Required to be Followed: Parameter Heart rate Respiration Blood Pressure Oxygen Saturation CVP 7. Alarm Limits High limits: 150, Low Limits: 50 High limits: 40, Low limits: 10 Low limits: Systolic <85, Diastolic <100 High limits: Systolic >170, Diastolic >100 Low limit: <85% High limits: 16-20, Low limits: 0-4 Fall Prevention System Patients admitted to the hospital will be assessed by an RN for fall risk factors during admission, daily and as his/her condition warrants. Every attempt will be made to safely support the patient¹s rights for independence and self-determination by creatively activating the least restrictive methods for fall prevention which are effective. Manufacturer recommends that staff conduct system tests upon installation and on periodic basis. The system has three methods of alerting staff battery needs recharging. Red charge battery light on signal unit will illuminate approximately one week prior to dead battery. Signal unit will chirp every ten seconds approximately 24 hours prior to dead battery, and finally console status review light will illuminate approximately one week prior to dead battery. Select a signal unit to be programmed. To test unit prior to placement on a patient, press unit button to determine battery status. Plug sensor into signal unit sensor jack. Both console and signal unit will alarm. Press console silence alarm button. Console will stop alarming, while signal unit will continue to alarm. Place hand flat across sensor and apply pressure. After a short delay, signal unit alarm will stop. Clinical AlarmsPatient Safety Policy Page 8 of 8. Nurse Call System 9. Monitoring of clinical alarm response times is through the Nurse Call System’s Database and reporting System for the following: a) Ventilators b) Pulse Oximetry c) IV Pumps d) RN+ System The data is reviewed monthly and reported quarterly. References: JC-HAS: OTHER: Original Date Review/Revision Date Supersedes all Previous Approved: _____________________ Date