3. Common Mode Rejection Ratio: Part I 3.1 Introduction In general, an instrumentation amplifier is required to amplify the difference between two input signals or voltages, V1 and V2 as shown in Fig.1. However, as discussed, in the case of many transducers there is often a potential at both inputs present when the input parameter to be measured is zero. This signal is the same at both inputs and is present regardless of the value of the input parameter. This signal is defined as the Common-Mode input signal, Vic and should make no contribution to the output voltage of the amplifier. On the other hand, when the value of the input parameter is non-zero, the potential at one input terminal will increase, while that at the other input terminal will decrease proportionately, giving a difference in the potentials at the two input terminals. This difference in the two input potentials is defined as the Differential-Input signal, Vid. Fig. 1 below shows the common-mode voltage as applied centrally to both input terminals of the amplifier while the differential voltage is considered split between the two inputs as half the differential potential one each side, V id /2. It is applied as a positive sense signal to the non-inverting input of the amplifier and as a negative sense signal to the inverting input of the amplifier. Vid / 2 + diff amp - ~ ~ Vic ~ -Vid / 2 V2 V1 V O Fig. 1 Differential and Common Mode Inputs to a Differential Amplifier This distribution of common-mode and differential mode signal is illustrated in Fig, 2. The potentials shown represent steady-state or dc levels, but could equally be time-varying or dynamic signals. For example if the signal of interest to be measured were the Electrocardiogram obtained from two electrodes placed on the surface of the body, then this would form the differential input signal. The common-mode signal in this scenario is often composed of mains interference from the electricity supply. This gives rise to an unwanted signal at the input of the amplifier which must be rejected in favour of the wanted differential signal which should be amplified. The ability of the amplifier to discriminate against the common- mode signal and prevent it from making any contribution to the output voltage of the amplifier is termed 1 Common Mode Rejection Ideally, the common-mode input signal should produce no response at the output but in practice it does make some contribution. A figure of merit used to quantify the extent of an amplifier’s ability to reject or supress the common-mode input signal is the Common-Mode-Rejection-Ratio or CMRR of the amplifier. Vid /2 Vid Vid /2 V1 Vic V2 Fig. 2 The Distribution of Differential and Common Mode Input Voltages 3.2 Common-Mode-Rejection-Ratio From the above definitions and from Fig. 1 and Fig. 2 we have: V1 Vic Vid 2 V2 Vic and so that: Vid V1 V2 and Vic Vid 2 V1 V2 2 If the differential amplifier were ideal, it would suppress the common-mode component of the input signal so that the output has no contribution from the common mode input and the output voltage would be simply be given by: Vo Ad (V1 V2 ) AdVid where Ad is the differential gain of the amplifier as given, for example by Eq. 6 of the previous lecture. However, in practice the amplifier does not fully reject the common mode component of the input signal and this consequently makes some contribution to the output. A common-mode gain, Ac can therefore also be specified so that the output voltage of the amplifier is given in practice as: Vo AdVid AcVic Ideally Ac →0 but in practice Ac << Ad but is finite. 2 The measure of the ability of the amplifier to reject the common mode input component, Vic , in favour of the differential component is the Common-ModeRejection Ratio or CMRR of the amplifier. This is defined as: CMRR Ad Ac ; ideally A c 0 and CMRR Then expressing the common-mode gain in terms of the differential gain we have: Ad Ac CMRR In this case: VO A d Vid Ad Vic CMRR The left hand term is the wanted output signal while the right hand term is essentially an error component. The error in the output is then given as the ratio of these components: ε Vic /CMRR 1 Vic x 100% Vid CMRR Vid In order to obtain a desired fractional error ε the required CMRR is: CMRR 1 Vic ε Vid For example, if Vic=1V and Vid =1mV, then for a 1% error we need: CMRR 1 1 x 3 105 100dB 0.01 10 This is substantial but typical of the CMRR required in bio-amplifiers. 3 3.3 Determining Factors of CMRR Consider again the standard 3 op-amp instrumentation amplifier as shown below: + _ R3A R4A A1 V1 VO1 R2A + R1 _ A3 R2B _ VO A2 R4B R3B + VO2 V2 Fig. 3 The standard 3 op-amp Instrumentation amplifier From previous work: R R VO1 1 2 V1 2 V2 R1 R1 and R R VO 2 1 2 V2 2 V1 R1 R1 for the input stage. If V1 Vic Vid 2 V2 Vic and 4 Vid 2 then this gives: R 2 V R V Vic id 2 Vic id VO1 1 R1 2 R1 2 R R V R R V VO1 1 2 Vic 1 2 id 2 Vic 2 id R1 2 R1 R1 2 R1 so that R V VO1 Vic 1 2 2 id R1 2 Similarly: R V VO 2 1 2 Vic id R1 2 VO 2 Vid R2 V ic R 2 1 R2 R 2 Vid R 2 R 2 Vid 1 V 1 V ic 2 R ic R 2 R R 1 1 1 1 R V VO1 Vic 1 2 2 id R1 2 From this it can be seen that the common-mode signal gets amplified by a factor of only unity on both sides of the first stage of this amplifier structure. On the other hand, the differential input component on each side of the amplifier receives a gain of (1+2R2/R1). This improves the overall common mode rejection ratio of the amplifier because it boosts the wanted differential signal compared to the unwanted common-mode signal before being passed to the output stage which performs the differential input-to-single-ended output. In the final stage we have: R VO 4 VO1 VO 2 R3 5 Substituting gives: R 4 R 2 Vid R 2 Vid Vic 1 2 Vic 1 2 VO R1 2 R1 2 R 3 so that finally: R R VO 4 1 2 2 Vid R1 R 3 This shows that in the final stage, under perfect conditions, the commonmode signal is rejected completely by the amplifier, while the differential signal gets a further gain of R4 / R3. This implies that under these ideal conditions this amplifier structure has infinite common-mode rejection ratio. In practice, of course, conditions are not ideal and the CMRR of the amplifier is finite. There are three main factors which act to limit the CMRR attainable: mismatch in the source and input impedances of the amplifier manufacturing tolerances, in the gain-determining resistors finite CMRR of the individual operational amplifiers. 3.4 Source and Input Impedance Mismatch The input stage of any instrumentation amplifier can be modelled as shown in Fig. 4. Each input has an impedance associated with the corresponding source. For, example the impedance of the electrodes used to measure an ECG signal is significant and will also show a variation from one electrode to another, so that there is a mismatch between the two source impedances. The amplifier has an impedance between each input terminal and ground, which is referred to as the common-mode input impedance. This can also be slightly different on each side. Finally, there is a finite impedance between the two input terminals of the amplifier, referred to as the differential input impedance. If a common-mode signal alone is applied to the amplifier inputs, via different source impedances to different input common-mode impedance on each side, then the signal appearing at the two input terminals of the amplifier will be slightly different. This means that the mismatch in the impedances leads to the applied common-mode input signal being effectively converted into a signal having a differential component at the input to the amplifier. This differential component will subsequently be given the full differential gain in passing through the amplifier and will give rise to an error in the overall output signal which is due to the unwanted commonmode input signal which has not been completely rejected. 6 ZCA ZSA V+ + ZSA ZD Amplifier _ V1 ZD V1 ZSB ZCA ZCB ZSB ZCB V2 Fig. 4 V- The Equivalent Circuit of an Instrumentation Amplifier Input The principle of superposition can be used to establish the potential at the inverting and non-inverting inputs of the amplifier. If the input impedance of the amplifier itself is taken as ideal, then no current flows into it but is confined to the network of impedances. Consider the case where the input potential V 1 is applied with V2 = 0V so that the latter input can then be considered grounded. In this case the potential at the non-inverting input, V+ and that at the inverting input due to the input, V1 can be found as: V1 and ZD ZSB // ZCB // ZCA V 1 ZSA ZD ZSB // ZCB // ZCA ZD ZSB // ZCB // ZCA V ZSB // ZCB V1 ZD ZSB // ZCB ZSA ZD ZSB // ZCB // ZCA 1 𝑽 𝟏 Similarly, expressions for the potentials at the inverting and non-inverting inputs due to the input V2 can be obtained by symmetry as: V2 and V2 ZD ZSA // ZCA // ZCB V 2 ZSB ZD ZSA // ZCA // ZCB Z D Z SA // ZCA // ZCB V Z SA // Z CA Z D Z SA // ZCA Z SB Z D Z SA // ZCA // ZCB 2 7 If the common-mode input impedances have a tolerance of ±ΔC but are intended to be purely resistive where possible and the source impedances have a mismatch of ±ΔS then the worst case mismatch arises when: ZSA ZS 1 S ZSB ZS 1 S ZCA R C 1 C ZCB R C 1 C Then with: V1 Vic Vid 2 V2 Vic Vid 2 When these substitutions are made, the potential at the input of the amplifier from the network can be obtained in the form: V V Ac Z ViC AdZVid The CMRR due to input impedance mismatch can then be defined in the usual manner and when the appropriate analysis is carried out it can be shown that: CMRR Z A dZ R CA R CB R C (1 2C ) A c Z R CA ZSB R CB ZSA 2ZS C S With a small value of ΔC this can be expressed in Decibels as: CMRR Z dB 20 log 10 RC 1 20 log 10 ZS 2 C S If all of the other factors influencing the overall CMRR of the amplifier can be taken as ideal then the mismatch in impedance values will determine the CMRR which can be attained by the instrumentation amplifier in rejecting unwanted common mode signals. 8 3.5 Design Example: An ECG recording amplifier is used with high impedance electrodes in a portable application. The smallest component of interest in the ECG signal has an amplitude of 1mV at the point of pick-up. It is desired to maintain a minimum signal-to-interference ratio of 20dB when observing this component. Due to interference in the recording environment mains supply hum at a frequency of 50Hz is present on the patient’s body at an amplitude which can be as high as 2Vrms. If the recording electrodes have an impedance 0.5MΩ ± 20% at 50Hz, determine the value of common-mode input impedance required in the instrumentation amplifier used. Solution: The differential input signal level must be taken as Vid = 1mV peak The common-mode interfering signal is the mains hum at an rms level of 2V The peak value of the common-mode signal is then Vic = 2 x √2 = 2.83V The signal-to-interference ratio required at the output of the amplifier is 20dB which corresponds to an absolute value of 10. This implies that the error in the signal observed should be less than 1 part in 10 or 1/10 = 0.1. Then: CMRR 1 Vic 1 2.83 3 2.83 104 ε Vid 0.1 10 If the CMRR is taken as being limited only by impedance mismatch, then this implies that: CMRRZ RC (1 2C ) 2.83 10 4 2Z S C S The tolerance in the electrode impedance is ±20% so that ΔS = 0.2. If a 5% tolerance in the resistors is assumed then ΔC = 0.05. This gives: RC 4 2 . 83 10 2 0.5 10 6 0.25 So that: RC 0.25 106 2.83 10 4 7 109 9