VET SYSTEM 30
Veterinary Radiological System
DESCRIPTION Modern and functional table for radiodiagnostic exams, suitable for all main kinds of examinations in veterinary field. The system is completed with a 30 kW H.F. x-ray generator.
Consists of a table with floating examination top, a fixed height column and an integrated monobloc.
TABLE TOP Low absorption laminated table top, floating in the four directions, equipped with a pair of mechanical brakes with pedal release on the structure basis.
Cassette holder with removable tray
COLUMN STAND Compact and easy to install column completed with monobloc covers and integrated LCD touch screen support
OPTIONAL
Removable stationary grid, 103 Lp/inch,R10:1, F=100 cm
“Tie down” clam cheats
Table top with track for clean up during examination
+/- 90° monobloc yoke rotation with handle
1600 mm length table top
VET SYSTEM 30
Veterinary Radiological System
TECHNICAL DATA
X-RAY GENERATOR High Frequency monobloc
Maximum power
Maximum voltage
Maximum mAs
Maximum mA
30 kW (300 mA - 100 kV @ 100 msec)
125 kV
315 mAs
400 mA
Monobloc thermal capacity 600 kJ (800 KHU)
Monobloc maximum thermal dissipation 55 W
Frequency kV ripple
40 kHz
< 2%
Leakage radiation (IEC 601-1-3)
Total filtration
X-RAY TUBE Rotating X-ray tube
< 1 mGy / h
> 2.5 mm Al @ 75 kV
Focal spot size
Maximum voltage
Anode speed
Anode angle
Anode maximum heat dissipation
Maximum continuous anode heat dissipation
0.6 mm - 1.3 mm
130 kV
3000 r.p.m.
15°
80 kJ (107 kHU)
300 W
CONTROL CONSOLE Microprocessor controller console
Display
Type: 5.7” Colour TFT LCD touch screen display
Resolution: 320 × 240 pixels
Alarm
Acoustic and luminous X-ray emission alarm
APR
(Pre set Anatomical Programs) 3 Mode (Dog — Cat — Exotic animals)
Selectable by animal’s thickness or weight
Film screen selection
3 speeds
Languages available
Italian
English
Spanish
French
Portuguese
VET SYSTEM 30
Veterinary Radiological System
SAFETY AND
PROTECTION
Automatic control and protection of the filament current
Over current protection
Overvoltage protection
Xray tube overload protection
Monobloc kHU automatic survey (KHU%)
Errors alarms with language description
COLLIMATOR Manual collimator
Adjustment from 0x0 to 43x43 cm
High intensity lamp (160 lux) for X-ray field simulation and automatic switch off after 30 sec.
Retractable tape
OPERATION MODE 3-point operation (kV – mA – time) kV selection: from 40 to 125 kV [1kV step]
2-point operation (kV – mAs) mA selection: from 25 mA to 400 mA [13 steps]
Time selection: from 0.001 sec to 6,3 sec [35 steps] kVp selection: from 40 kV to 125 kV [1 kV step] mAs selection: from 0.2 mAs to 315 mAs [27 steps]
EXPOSURE CONTROL Two-step pedal
POWER SUPPLY Voltage: 230 Vac Single phase (115Vac Single phase on request)
Frequency: 50/60 Hz
Current consumption: 16 A max.
Total line resistance: 0.40 Ohm max.
Automatic line compensation: ± 10%
MECHANICAL DATA
HORIZONTAL TABLE Table top dimension
Longitudinal travel
Transversal travel
Brakes
Table top– floor distance
Table top-cassette distance
Maximum patient weight
COLUMN STAND Column height
SID
Monobloc rotation on its (X) axis
(optional)
WEIGHT & DIMENSIONS Weight
Dimensions
1500 x 630 mm
720 mm (± 360 mm)
170 mm (± 85 mm)
Mechanical
800 mm
65 mm
100 Kg
1950 mm
1000 mm
± 90°
1600x630 mm (optional)
840 mm (± 420 mm)
170 mm (± 85 mm)
Mechanical
800 mm
65mm
100 Kg
197,5 Kg
150 x 92 x 195 cm
VET SYSTEM 30
Veterinary Radiological System