Supplemental Information Intercomparisons of mobility size spectrometers and condensation particle counters in the frame of the Spanish atmospheric observational aerosol network Francisco J. Gómez-Moreno1,*, Elisabeth Alonso1, Begoña Artíñano1, Vanesa JuncalBello2, Silvia Iglesias-Samitier2, María Piñeiro Iglesias2, Purificación López Mahía2, Noemí Pérez3, Jorge Pey3#, Anna Ripoll3, Andrés Alastuey3, Benito A. De La Morena4, M. Isabel García5,6, Sergio Rodríguez5, Mar Sorribas7,8, Gloria Titos7,8, Hassan Lyamani7,8, Lucas Alados-Arboledas7,8, Enrique Latorre9, Torsten Tritscher10, Oliver F. Bischof10 1 Department of Environment, CIEMAT, Madrid, Spain Grupo Química Analítica Aplicada, Instituto Universitario de Medio Ambiente (IUMA), Departamento de Química Analítica, Facultade de Ciencias, Universidade da Coruña, A Coruña, Spain 3 Institute of Environmental Assessment and Water Research (IDAEA-CSIC), Barcelona, Spain 4 Atmospheric Sounding Station 'El Arenosillo', INTA, Mazagón-Huelva, Spain 5 Izaña Atmospheric Research Centre, (IARC/CIAI), AEMet, Santa Cruz de Tenerife, Spain 6 Department of Analytical Chemistry, Nutrition and Food Science, University of La Laguna, Spain 7 Andalusian Institute for Earth System Research, IISTA-CEAMA, University of Granada, Granada, Spain 8 Applied Physics Department, University of Granada, Granada, Spain 9 Álava Ingenieros, Madrid, Spain 10 TSI GmbH, Aachen, Germany 2 -----------------------------------------------------------------*Corresponding author: CIEMAT, Av. Complutense 40, E-28040 Madrid, Spain. E-mail: fj.gomez@ciemat.es #Current address: Laboratory of Environmental Chemistry LCE-IRA, Aix Marseille University, Marseille, France January 2015 S1. CPC intercomparisons During the 2010 intercomparison campaign, five different models of particle counters were compared. Ambient air was sampled during 14 hours and the results are shown in Figure S1, presenting the total particle concentration measured as a function of time and the deviation of each counter from the mean value. Particle concentration was low until midnight when there was a change in wind direction, increasing the particle levels in the measurement site from 2000 to 14000 cm-3. During both periods, the TSI-CPC-2B (3025A), TSI-CPC-2A (3775) and TSI-CPC-5 (3776) measured similar concentrations, the TSI-CPC-4 (3010) the lowest and the TSI-CPC-3 (3785) the highest. For the TSI-CPC-4 (3010), the difference compared with the TSI-CPC-2A (3775) was -4% during the first period and -11% during the second period. One main difference between this CPC and the other CPC models that also use butanol as their working fluid is the lower detection limit (Hermann et al., 2007). The other three CPCs have a particle size detection limit of around 3 nm, while the model CPC 3010 can only detect particles 10 nm and larger. If the share of nucleation mode particles below 10 nm increases, this CPC model will report a lower number concentration. It should also be pointed out that the CPC 3010 is suitable for monitoring concentrations up to 10,000 particles/cm3. Although it can measure above this value, the coincidence error begins to get very important. When there was a high concentration in the nucleation mode, bigger differences between measurements with different CPC models were observed. In the first period, there was a typical nocturnal distribution at the site, with very few particles below 10 nm and relatively low concentration, so differences were small. When the wind direction changed and there was particle transport from a nearby industrial area, higher concentrations for particles in the nucleation mode were reached. The corresponding concentrations were above the coincidence limit for the CPC 3010. In summary, both the less sensitive lower size limit and the upper concentration limit were the likely reasons for the larger differences during this last period. The TSI-CPC-3 (3785), which has a 5 nm detection limit, measured between 10-12% more than the TSI-CPC-2A (3775). The higher concentration measured only during the second period with air transported from the industrial area, could be explained by the lower accuracy of the photometric mode calibration for this water-based CPC. This has since been addressed in the replacement model WCPC (TSI, model 3787) by the manufacturer, which no longer requires a photometric mode but is still able to detect concentrations up to 250,000 cm-3. CPC-2A (3775) CPC-2B (3025) CPC-3 (3785) CPC-4 (3010) CPC-5 (3776) -3 Particle number concentration (cm ) 16000 14000 12000 10000 8000 6000 4000 2000 0 16:00 04:00 23/04/2010 22:00 22/04/2010 Figure S1a. Difference with the mean value (%) 20 15 CPC-2A (3775) CPC-2B (3025) CPC-3 (3785) CPC-4 (3010) CPC-5 (3776) 10 5 0 -5 -10 -15 -20 16:20 20:30 22/04/2010 0:40 04:50 23/04/2010 Figure S1b Figure S1. CPC intercomparison during the 2010 campaign (a) particle concentration measurements, (b) difference between each counter and the mean value during the measurements During the 2012 intercomparison these procedures were repeated with the counters available (TSI-CPC-1, TSI-CPC-2A, TSI-CPC-2B, TSI-CPC-3, TSI-CPC-5A and TSICPC-5B, corresponding to 3772, 3775, 3025A, 3785, and two 3776 models). All the instruments were connected to the external sampling line using tubes with identical length. Atmospheric aerosol was sampled for 18 hours and the results are shown in Figure S2. There were two clear groups of measurements, above and below the average value. The first group consisted of those CPC models that make use of a photometric mode (CPC models 3775 and 3785). The second group corresponds to those counters with lower particle size detection limits down to 3 nm. As these CPCs are able to measure a wider range of sizes, they should measure the same or higher particle concentration, which was not the case. It is not clear what causes the differences. Some problems in the photometric mode or in the internal flow issues (especially for the CPC 3776) could contribute with a slight drift. No problems were observed with the zero level which could explain these differences. The TSI-UFPM-3031 was also compared with a TSI-CPC-5A (3776) during this campaign and the main results obtained can be found in Figure S2c. Three important factors must be taken into account, the fact that the UFPM is not actually a single particle count instrument, the different lower size limits and the different kernel matrixes used in the UFPM. The measurement time had three different periods where two different kernel matrices were applied. The environmental matrix was used during periods 1 and 3 and the ammonium sulfate matrix during period 2. During periods 1 and 3 with the environmental matrix applied, the UFPM frequently measured higher concentrations than the CPC. During period 2 with the ammonium sulfate matrix used, the UFPM measured concentrations below the TSI-CPC-5A (3776). This seems to indicate that this last matrix might be better suited under the conditions present during our measurements. Due to the limited number of measurements further research is needed to better estimate the influence of the kernel matrix on the particle concentrations measured. 20000 CPC-1 (3772) CPC-2A (3775) CPC-2B (3025) CPC-3 (3785) CPC-5A (3776) CPC-5B (3776) -3 Particle concentration (cm ) 18000 16000 14000 12000 10000 8000 6000 4000 2000 0 13:15 19:05 12/11/2012 Figure S2a 0:55 06:45 13/11/2012 CPC-1 (3772) CPC-2A (3775) CPC-2B (3025) CPC-3 (3785) CPC-5A (3776) CPC-5B (3776) Difference with the mean value (%) 60 50 40 30 20 10 0 -10 -20 -30 -40 13:15 19:05 12/11/2012 06:45 13/11/2012 0:55 Environmental Matrix 15000 (NH4)2SO4 Matrix 20000 CPC-5A (3776) UFP Environmental Matrix -3 Particle number concentration (cm ) Figure S2b 10000 Period 1 5000 08:00 09:00 Period 2 10:00 11:00 Period 3 12:00 Time Figure S2c Figure S2. CPC and UFPM intercomparison during the 2012 campaign (a) CPC particle concentration measurements, (b) difference between each counter and the mean value during the measurements (c) CPC and UFPM concentration measurements S2. SMPS intercomparison During the 2010 campaign, the complete SMPS systems were installed after checking of the CPC behavior. The initial idea was to perform the size distribution measurements using the same flow rates that every instrument uses at its regular location. Nevertheless, the use of different flow rates results in different losses for the smallest particles, as diffusion losses are a function of the particles’ residence time inside the instrument and in the sampling lines. For this reason the diffusion correction factor in the software (TSI, model AIM 3936) was selected, as well as the multiple charge correction (Reineking and Porstendörfer, 1986; Wiedensohler, 1988). No correction was applied for possible differences in particle losses in the sampling lines. For accumulation mode particles, the losses must be very similar for all the flow rates used during the comparison. There were several groups of flow rates during the campaign: (a) three instruments used 0.3 lpm for the aerosol flow, which correspond to the low flow rate of CPC models 3775 and 3776; (b) an instrument used 0.6 lpm, which corresponds to the inlet flow of the CPC 3010; (c) and two instruments used 1 lpm, which corresponds to the inlet flow of CPC models 3772 and 3785. In all cases, the ratio between the aerosol and the sheath flows in the DMAs was 1:10. The measurements obtained with three SMPS systems using CPC models 3775 (TSISMPS-2), 3010 (TSI-SMPS-4) and 3776 (TSI-SMPS-5) as their detectors are shown in Figure S3. All of them used a Nafion dryer for the sample flow and a Nafion or a diffusion dryer in the sheath flow. The measurements were carried out for almost two days (41 hours) and during this time air masses of different origins, and hence different types of particles (Sorribas et al., 2011), reached the sampling point, as was also seen in the meteorological data measured in situ. In this figure an example of two different particle size distributions measured is shown. In Figure S3a, only particles in the accumulation mode were observed, while in Figure S3b a change in the air mass had taken place and a nucleation mode also appeared. These different size distributions are very interesting for this intercomparison because particles belonging to each mode have different properties. In both graphs the particle size corresponding to the modes were very similar in the three instruments. In the case of the accumulation mode, TSI-SMPS-2 and TSI-SMPS-4 measured almost indistinguishable particle number concentrations while TSI-SMPS-5 reported slightly lower values, probably due to losses in the total system. In the case of the nucleation mode (Figure S3b), the situation is different. TSI-SMPS-2 and TSI-SMPS-5 measured the same particle concentrations while the TSI-SMPS-4 measured higher values for the nucleation mode. A possible explanation is that TSI-SMPS-2 and TSI-SMPS-5 worked with aerosol to sheath flow ratios of 0.3 to 3, while the TSI-SMPS-4 worked with a ratio of 0.6 to 6 by means of a flow equalizer. In this latter case the higher aerosol flow produces lower losses in the entire sampling line, so that the measured particle number concentration is also higher for the smaller sizes that are most affected by diffusion losses (Karlsson and Martinsson, 2003). At any rate, the concentrations measured were within the +/- 10% tolerance band and only for the smallest particles they were slightly above this tolerance for TSI-SMPS-4. TSI-SMPS-3 was later compared with TSI-SMPS-2 and very good results were obtained, measuring the same diameters and very similar concentrations that were distinctly within the +/- 10% tolerance. 12000 SMPS-2 SMPS-4 SMPS-5 Average Average +/-10% 21/04/2010 7:00h -3 dN/dLog Dp (cm ) 10000 8000 6000 4000 2000 0 10 100 1000 Dp (nm) Figure S3a 21/04/2010 11:00h SMPS-2 SMPS-4 SMPS-5 Average Average+/-10% 4000 -3 dN/dLog Dp (cm ) 5000 3000 2000 1000 0 10 100 1000 Dp (nm) Figure S3b Figure S3. Particle size distributions measured by the SMPSs during the 2010 campaign During the 2011 intercomparison, the SMPS systems available were the instruments designated as 1, 2, 3 and 5. It was observed that the TSI-SMPS-1 measured in a similar way as the other instruments for the smaller particle sizes, but significantly different above 7080 nm, showing lower values of concentration. The reason for this deviation could be the high-voltage source (Wiedensohler et al., 2012) as the sheath flow was kept constant over time. This case reinforces the necessity of checking the high voltage during the routine maintenance. An example of the results obtained with the other three instruments can be found in Figure S4, where average measurements from 22:00 to 24:00 on October 6, 2011 are shown. The concentrations measured were similar except for the mode of the distribution, where differences exceeded 10%. Such deviation was larger than the differences previously observed among the CPCs. These maximum occurred around 100 nm, so diffusion losses in the sampling lines caused by different aerosol flow rates were discarded as a possible reason. The TSI-UFPM was also compared with the SMPS systems. The comparison with TSISMPS-5 for four selected times is shown in Figure S4b. The SMPS data were average over the size bins used in the UFPM (20-30, 30-50, 50-70, 70-100 and 200-450 nm) for comparison. The kernel matrix used during the measurement was the one for ammonium sulfate. During most of the time, the shapes of the size distribution were similar in both instruments, showing the maximum at the same diameter, although the concentrations measured by the TSI-UFPM were higher. Only for the biggest diameters the concentrations in the SMPS were above those measured by the UFPM. This seems contradictory to the comparison with the CPCs when the UFPM measured lower concentrations, but the CPCs could reach smaller sizes than the UFPM. This is likely due to the fact that the UFPM uses a unipolar corona charger rather than a bipolar aerosol neutralizer. The diffusion charger used in the UFPM delivers a mean charge level per particle that is proportional to its diameter for the size range from 10 to 100 nm (Kaufman et al., 2002), but slightly flattens out above 100 nm which leads to a higher uncertainty in the reported concentrations of an electrometer detector for larger particles. 3500 SMPS-2 SMPS-3 SMPS-5 Average Average +/-10% 2500 -3 dN/dLog Dp (cm ) 3000 2000 1500 1000 500 0 10 100 Dp (nm) Figure S4a 1000 5/10/2012 13:00GMT SMPS-5 UFP 8000 5/10/2012 21:00GMT SMPS-5 UFP 6000 -3 dN/dLog Dp (cm ) -3 dN/dLog Dp (cm ) 4000 4000 2000 3000 2000 1000 0 0 100 100 Dp (nm) Dp (nm) Figure S4b1 7000 6/10/2012 5:00GMT Figure S4b2 SMPS-5 UFP 25000 SMPS-5 UFP 6/10/2012 11:00GMT 20000 -3 dN/dLog Dp (cm ) -3 dN/dLog Dp (cm ) 6000 5000 4000 3000 15000 10000 2000 5000 1000 0 0 100 100 Dp (nm) Dp (nm) Figure S4b3 Figure S4b4 Figure S4. Particle size distributions measured by the SMPSs and UFPM during the 2011 campaign S3. Particle deposition in the dryers SMPS-B Flow splitter Air Common dryer Test dryer SMPS-A Figure S5. Set-up used during the particle deposition study on dryers REFERENCES Hermann, M., Wehner, B., Bischof, O.F., Han, H.-S., Krinke, T.J., Liu, W., Zerrath, A.F. and A. Wiedensohler (2007). Particle counting efficiencies of new TSI condensation particle counters. J. Aerosol Sci., 38:674-682. Kaufman S.L., Medved, A., Pöcher, A., Hill, N., Caldow, R. and Quant, F.R. (2002). An electrical aerosol detector based on the corona-jet charger, Poster PI2-07, Abstracts of 2002 AAAR, Charlotte, NC, p. 223. Karlsson, M.N.A., Martinsson, B. G. (2003). Methods to measure and predict the transfer function size dependence of individual DMAs. J. Aerosol Sci., 34:603-625. Reineking, A. and Porstendörfer, J. (1986). Measurements of Particle Loss Functions in a Differential Mobility Analyzer (TSI, Model 3071) for Different Flow Rates. Aerosol Sci. Technol., 5:483-486 Sorribas, M., de la Morena, B. A., Wehner, B., López, J. F., Prats, N., Mogo, S., Wiedensohler, A.and Cachorro, V. E. (2011). On the sub-micron aerosol size distribution in a coastal-rural site at El Arenosillo Station (SW-Spain). Atmos. Chem. Phys., 11:11185– 11206. Wiedensohler, A. (1988) Technical Note: An Approximation of the Bipolar Charge Distribution for Particles in the Submicron Range. J. Aerosol Sci., 19:387-389. Wiedensohler, A., Birmili, W., Nowak, A., Sonntag, A., Weinhold, K., Merkel, M., Wehner, B., Tuch, T., Pfeifer, S., Fiebig, M., Fjäraa, A. M., Asmi, E., Sellegri, K., Depuy, R., Venzac, H., Villani, P., Laj, P., Aalto, P., Ogren, J. A., Swietlicki, E., Williams, P., Roldin, P., Quincey, P., Hüglin, C., Fierz-Schmidhauser, R., Gysel, M., Weingartner, E., Riccobono, F., Santos, S., Grüning, C., Faloon, K., Beddows, D., Harrison, R., Monahan, C., Jennings, S. G., O'Dowd, C. D., Marinoni, A., Horn, H.G., Keck, L., Jiang, J., Scheckman, J., McMurry, P. H., Deng, Z., Zhao, C. S., Moerman, M., Henzing, B., de Leeuw, G., Löschau, G., and Bastian, S. (2012). Mobility particle size spectrometers: harmonization of technical standards and data structure to facilitate high quality long-term observations of atmospheric particle number size distributions. Atmos. Meas. Tech., 5:657-685.