Line Broadening 1) Collect 1st spectrum without compressor running W202 Test Input: 2 Lines OK 1500V, wait 30s, clear Data If required change V to accommodate peaks Set windows on escape peak excluding noise and on 14.4K Lines visible = peaks set Send windows, Δ = sent Exit W302 2 mm/s Monitor V transducer; err ≤ 40 mV PP Count rate ≈800c/mCi = OK Channel 1 Plot, see spectrum, wait 30minutes Save data (in default w302_1.txt), folder should be on desktop Open w302_1.txt, put remarks at bottom, save as yymmdd… 2) Collect data with compressor running + temperature set ambient SAME 3) Run WMOSS to check Line Broadening (Check Reference “A” at Bottom of this text 1st) Click 98 desktop ← Run WMOSS Files Data load Format 4 (Y ASCII) Type in yy*.txt ← + esc 1024 Points Fold data Y Interchange N Previous cal N Remove baseline N Auto fold ← 1 Step 1 (Go below everything to enter remarks) Esc ← Save binary….b Param load Quadr sp ← ←← ← Analysis Menu Edit Param ← …input file …output file Page down, page down, page down Subsite #1 Shift 220 (~mid spect in channels) dEq 200 (splitting) Gamma L 20 Esc Files Param save qp1.pi qp1.po Lower 200 Lower 100 10 Upper 300 Upper 300 50 ← Analysis Menu do Fit Return, utilities, t Add mask <> 1←v <>← 512 ← v <← Esc Analysis Menu do Fit ← R Edit param ← Page down, page down, page down Subsite #1 dEq = 183.641 Gamma L = 27.761 1.68 𝑚𝑚 → 183.6𝑠 = 0.009150 (1.68 known from Fe) →0.009150 ∗ 27.761 = 0.254 𝑚𝑚/𝑠 (=full width @ ½ max) Close WMOSS by Esc, Exit, Y 2 4) Collect 2nd spectrum with compressor running and temperature set at ambient 5) Run WMOSS to check Line Broadening (Check Reference “A” at Bottom of this text 1st) Click 98 desktop ← Run WMOSS ← Files ← Data load ← Format 4 (Y ASCII) ← Type in yy*.txt ← + esc 1024 Points ← Fold data Y Interchange N Previous cal N Remove baseline N Auto fold ← Step 1 (Go below everything to enter remarks) Esc ← Save binary….b ← Param load ←← qp1.pi (from previous file, already in mm) Analysis Menu Fit (no Edit Param because of qp1.pi here above) R = return Edit Param ← Page down, page down, page down Subsite #1 Shift 245.7 dEq 186.5 Gamma L 29.3 → 29.3 ∗ 0.009150 = 0.268𝑚𝑚/𝑠 (0.0091..from previous, Δ should be ≤ 0.1mm/s) Close WMOSS by Esc, Exit, Y 1) Collect 3rd spectrum with compressor running and temperature set at xK …and do line broadening check when ready. 3 A) RUNNING WMOSS IN ‘98 VM Ware Player Open Windows 98 WMOSS Click 98 Desktop + OK Open folder/drive/etc containing spectrum Drop spectrum file on 98 Desktop and put in “Shortcut to Data” Close Data Folder Click 98 desktop Run WMOSS Files Data load … ← ← ← ← 4