Measurement and Micropipetting

advertisement
Measurement and Micropipetting
(adapted from Laboratory DNA Science)
Metric Conversions
Become familiar with metric units of measurement and their conversions. We will concentrate on
liquid measurements based on the liter, but the same prefixes (milli- and micro-) also apply to dry
measurements based on the gram. The two most useful measurements in molecular biology are the
milliliter (ml) and the microliter (μl).
1 ml = 0.001 liter
1 μl = 0.000001 liter
1 μl = 0.001 ml
1,000 ml = 1 liter
1,000,000 μl = 1 liter
1,000 μl = 1 ml
Digital Micropipettors
A digital micropipettor is essentially a precision pump fitted with a disposable tip. The volume of
air space in the barrel is adjusted by screwing the plunger farther in and out of the piston, and the
volume displayed on a digital readout. Depressing the plunger displaces the specified volume of air
from the piston; releasing the plunger creates a vacuum, which draws an equal volume of fluid into
the tip. The withdrawn fluid is then expelled by depressing the plunger again.
Take the following precautions when using a micropipettor:
 Never rotate the volume adjustor beyond the upper or lower range of the pipet as stated by the
manufacturer.
 Never use the micropipettor without the tip in place; this could run the piston.
 Never invert or lay the micropipettor down with a filled tip; fluid could run back into the piston.
 Never immerse the barrel of the micropippetor in fluid.
 Never flame the tip of the micropipettor.
 Never reuse tips!
Familiarize yourself with small scale and large scale micropipettors that you will be using in this
laboratory. Locate the digital readouts on your micropipettors. It is located below the plunger. The
digital readout has three number boxes that display the specified volume (read vertically, top
down).
P10 micropipettor
1
0
0
0
1
0
0
5
7
The above digital readouts (left to right) indicate readings of 10.0, 1.0 and 5.7 μl on a 0.1-10 μl
volume micropipettor.
1
P100 micropipettor
1
0
0
0
1
0
0
5
7
The above digital readouts indicate readings of 100, 10 and 57 μl on a 10-100 μl volume
micropipettor.
P1000 micropipettor
1
0
0
0
1
0
0
5
7
The above digital readouts indicate readings of 1000 μl (or 1 mL), 100 and 570 μl on a 100-1000 μl
micropipettor.
Reagents
1 ml of Solution I, colored
1 ml of Solution II, colored
1 ml of Solution III, colored
1 ml of Solution IV, colored
Supplies and Equipment
100-1000 μl micropipettor and tips
10-100 μl micropipettor and tips
0.5-10 μl or 2.0-10μl micropipettor and tips
1.5 ml tubes
microcentrifuge
1.5 tube rack
permanent marker
I. General Use of Micropipettors
1. Rotate the volume adjustor to the desired setting. Note the change in plunger length as the
volume changes. Be sure to properly locate the decimal point when reading the volume
setting.
2. Firmly seat a proper-sized tip on the end of the micropipettor.
3. When withdrawing or expelling fluid, always hold the tube firmly between your thumb and
forefinger. Hold the tube nearly at eye level to observe the change in the fluid level in the
pipet tip. Do not pipet with the tube in the test tube rack. Do not have another person hold
the tube when you are pipetting.
2
4. Each tube must be held in the hand during each manipulation. Open the top of the tube by
flipping the tap with your thumb. During manipulations, grasp the tube body (rather than
the lid), to provide greater control and to avoid contamination of the mouth of the tube.
5. For best control, grasp the micropipettor in your palm and wrap your fingers around the
barrel; work the plunger (piston) with the thumb. Hold the micropipettor almost vertical
when filling it.
6. Most micropipettors have a two-position plunger with friction “stops”. Depressing the fist
stop measures the desired volume. Depressing to the second stop introduces an additional
volume of air to blow out any solution remaining in the tip. Notice these friction stops; they
can be felt with the thumb.
7. To withdraw the sample from a reagent tube:
a. Depress the plunger to the first stop and hold this position. Dip the tip into the
solution to be pipetted, and draw fluid into the tip by gradually releasing the plunger.
Be sure that the tip remains in the solution while you are releasing the plunger.
b. Slide the pipet tip out along the inside wall of the reagent tube to dislodge any excess
droplets adhering to the outside of the tip.
c. Check that there is no air space at the very end of the tip. To avoid future pipetting
errors, learn to recognize the approximate levels to which particular volumes fill the
pipet tip.
d. If you notice air space at the end of the tip or air bubbles within the sample in the tip,
carefully expel the sample back into its supply tube. Coalesce the sample by sharply
tapping the tube on the bench top or pulsing it in a microfuge.
8. To expel the sample into a reaction tube:
a. Touch the tip of the pipet to the inside wall of the reaction tube into which the sample
will be emptied. This creates a capillary effect that helps draw fluid out of the tip.
b. Slowly depress the plunger to the first stop to expel the sample. Depress to the second
stop to blow out the last bit of fluid. Hold the plunger in the depressed position.
c. Manually remove or reject the tip into a beaker kept on the lab bench for this purpose.
The tip is ejected either by depressing the plunger beyond the second stop or by
depressing a separate tip-ejection button.
d. When adding several reagents to the same tube, pipetting the solution in and out of
the pipettor after the addition of the last solution will mix the solution without causing
droplets to for on the inner wall of the tube.
9. To prevent cross-contamination of reagents:
a. Always add appropriate amounts of a single reagent sequentially to all reaction tubes.
If this is the first reagent added to each tube, the same tip can be used for all tubes.
b. Use a fresh tip for each new reagent to be pipetted.
c. When dealing with small amounts, it is important to keep track of your place on the
rack when loading multiple tubes (since visually the volumes in the tubes look the
3
same). Either move the tube to a different row on the rack or angle the lid differently
as an indication that reagent has been added to that tube.
II. Practice with P10 or P20 Micropipettor
1. Use a permanent marker to label two 1.5 mL microfuge tubes A and B.
2. Use the matrix below as a checklist while adding solutions to each microfuge tube.
3. The solutions are merely water (this is for practice only).
Tube
Sol. I
Sol. II
Sol. III
A
2 μl
5 μl
3 μl
B
3 μl
3 μl
4 μl
4. Close the tops. Pool and mix reagents , by the following method:
a. Place in a microfuge and apply a short pulse for several seconds. Make sure that
reaction tubes are placed in a balanced configuration in the microfuge rotor. Spinning
tubes in an unbalanced position will damage the microfuge rotor. (An empty 1.5-ml
tube can be used to balance a sample with a volume of 20 μl or less).
5. A total of 10 μl of reagents was added to each reaction tube. To check the accuracy of your
meaurements, set a pipet to 10μl and very carefully with draw the solution from each tube.
a. Does a small volume of fluid remain in the tube? This indicates an overmeasurement.
b. Is air space left in the end of the tip? This indicates an undermeasurement. (The
actual volume of fluid can be determined simply rotating the volume adjustment to
expel the air and push fluid to the very end of the tip. Then, read the volume directly.)
6. If several measurements were inaccurate, repeat the exercise to obtain nearly perfect results.
III. Practice with a P100 Micropipettor
1. Use a permanent marker to label two 1.5 mL microfuge tubes A and B.
2. Use the matrix below as a checklist while adding solutions to each microfuge tube.
3. The solutions are merely water (this is for practice only).
Tube
Sol. I
Sol. II
Sol. III
A
25 μl
35 μl
30 μl
B
20 μl
40 μl
30 μl
4. A total of 100 ml of reagents was added to each tube. To check the accuracy of your
4
measurements, set the pipettor to 100 μl and very carefully withdraw the solution from each
tube.
IV. Practice with P1000 Micropipettor
1. Use a permanent marker to label two 1.5-ml reaction tubes D and E.
2. Using the table below as a checklist while adding solutions to each reaction tube.
Tube
Solution I
Solution II
Solution III Solution IV
D
100μl
200μl
150μl
550μl
E
150μl
250μl
350μl
250μl
3. A total of 1,000 μl (1ml) of reactants was added to each reaction tube. To check the accuracy
of your measurements, set the micropipettor to 1,000μl and carefully withdraw the solution.
4. If your measurements are inaccurate, repeat the exercise to obtain nearly perfect results.
5
Download