TEP 5.4.50 -01 Quantitative X-ray fluorescence analysis of alloyed materials Related topics Bremsstrahlung, characteristic X-radiation, energy levels, fluorescent yield, Auger effect, coherent and incoherent photon scattering, absorption of X-rays, edge absorption, matrix effects, semiconductor energy detectors and multi-channel analysers. Principle Various alloyed materials are subjected to polychromatic X-rays. The energy of the resulting fluorescence radiation is analysed with the aid of a semiconductor detector and a multichannel analyser. The energy of the corresponding characteristic X-ray fluorescence lines is determined. In order to determine the concentration of the alloy constituents, the intensity of their respective fluorescence signals is compared to that of the pure elements. Equipment 1 1 1 1 1 1 1 1 XR 4.0 expert unit 35kV XR 4.0 Goniometer for X-ray unit, 35 kV XR 4.0 Plug-in module with W X-ray tube Diaphragm tube d = 1 mm Diaphragm tube d = 2 mm XR 4.0 XRED cable 50 cm Screened cable, BNC, l = 750 mm Multi-channel analyser 09057-99 09057-10 09057-80 09057-01 09057-02 09058-32 07542-11 13727-99 1 X-ray energy detector 1 Universal crystal holder for the X-ray unit Set of samples for the quantitative 1 X-ray fluorescence analysis, set of 4 Set of metal samples for the X-ray 1 fluorescence analysis, set of 7 1 Software for the multi-channel analyser PC, Windows® XP or higher 09058-30 09058-02 09058-34 09058-31 14452-61 This experiment is included in the upgrade package: XRM 4.0 X-ray material analysis. Fig. 1: P2545001 www.phywe.com P2545001 PHYWE Systeme GmbH & Co. KG © All rights reserved 1 TEP 5.4.50 -01 Quantitative X-ray fluorescence analysis of alloyed materials Tasks 1. Calibrate the semiconductor energy detector with the aid of the characteristic radiation of the tungsten X-ray tube 2. Recording of the fluorescence spectra that are produced by the alloyed samples. 3. Recording of the fluorescence spectra that are produced by the pure metals. 4. Determination of the energy values of the corresponding fluorescence lines. 5. Calculation of the concentration levels of the alloy constituents. Set-up and procedure Set-up (Fig. 1) Screw the adapter ring onto the inlet tube of the Fig. 2: Connections in the experimentation area energy detector. Connect the signal and supply cables to the corresponding ports of the detector with the aid of the right-angle plugs. Connect the signal and supply cables from the MCA to the appropriate connections in the experiment chamber of the X-ray unit (signal cable: red, supply cable: green (see Fig. 2)). Connect the external ports for the X RED of the Fig. 3: Connection at the external panel of the XR 4.0 Xx-ray unit (signal cable red, supply cable green, ray expert unit to the MCA see Fig. 3) to the multi-channel analyse (MCA). Connect the signal cable via a screened BNCcable to the “Input” port of the MCA and the X-ray energy supply cable to the “X-Ray Energy Det.” port of detector the MCA. Secure the energy detector in the holder of the swivel arm of the goniometer (Fig 4). Lay the two cables with sufficient length so that the goniometer can be swivelled freely over the entire swivelling range. Connect the multi-channel analyser and computer with the aid of the USB cable. Insert the tube with the 2-mm-aperture. Left position of the Bring the goniometer block and the detector to goniometer Universal crystal their respective end positions on the left. Bring holder with metal the detector to the 90° position in the 2:1 cousample pling mode (Fig. 4). Calibration of the multi-channel analyser (if there is no other already existing calibration that can be used) - Bring the goniometer block and the detector Fig. 4: Set-up at the goniometer 2 PHYWE Systeme GmbH & Co. KG © All rights reserved P2545001 Quantitative X-ray fluorescence analysis of alloyed materials - - - - - TEP 5.4.50 -01 to their respective end positions on the right. Insert the tube with the 1-mm-aperture into the exit tube of the X-ray tube. With the X-ray unit switched on and the door locked, bring the detector to the 0° position. Then, shift the detector by some tenths degree out of the zero position in order to reduce the total rate. Operating data of the tungsten X-ray tube: Select an anode voltage UA = 25 kV and an anode current IA = 0.02 mA and confirm Fig. 5: calibration of the multi-channel analyser these values by pressing the “Enter” button. Switch on the X-radiation In the MEASURE program, select “Multi channel analyser” under “Gauge”. Then, select “Settings and calibration”. After the “Calibrate” button has been clicked, a spectrum can be measured. The counting rate should be < 300 c/s. Energy calibration settings: - 2-point calibration, - Unit = keV, Gain = 2 – Set the offset so that low-energy noise signals will be suppressed (usually a few per cent are sufficient), See Fig 5. Measuring time: 5 minutes. Use the timer of the X-ray unit. Make the two coloured calibration lines congruent with the line centres of the two characteristic X-ray lines. The corresponding energy values (see e.g. P2544701) E(L3M5/L3M4) = 8,41 keV and E(L2N4) = 9,69 keV are entered into the corresponding fields, depending on the colour. (Note: Since a separation of the lines L3M5 and L3M4 Lines is not possible, the mean value of both lines is entered as the energy of the line). Name and save the calibration. Spectrum recording - Insert the tube with the 2mm-aperture. - Bring the goniometer block and the detector to their respective end positions on the left. Bring the detector to the 90° position in the 2:1 coupling mode. - Insert the sample with the universal crystal holder (sample at 45°). - Operating data of the molybdenum X-ray tube: Adjust an anode voltage Ua = 35 kV and an anode current so that the counting rate is ≤ 200 c/s. - Measuring time: 10 minutes (use the timer of the X-ray unit). Evaluation of the measurement curves - In order to determine the line energy, switch from the bar display to the curve display. To do so, click “Display options” and then “Interpolation and straight lines”. - Extend the relevant line section with the aid of the zoom function . - Then, select the curve section with . Open the window “Function fitting” “Scaled normal distribution” and confirm. - Find the line centroid of the normal distribution with “Peak analysis” function “Survey” (see Fig. 7). . Then, select or determine it with the www.phywe.com P2545001 PHYWE Systeme GmbH & Co. KG © All rights reserved 3 TEP 5.4.50 -01 Quantitative X-ray fluorescence analysis of alloyed materials Theory In order to determine the concentration of the elements in a sample with the aid of X-ray fluorescence analysis, a qualitative analysis must first be performed. During the assignment of the fluorescence lines, it must be taken into consideration that the relaxations that follow the primary ionisation process can only take place if they fulfil the quantum-mechanical selection rules Δj = 0, ±1 and Δl = ±1 (j = total angular momentum, l = orbital angular momentum). In addition, it should be noted that every element has groups of X-ray lines that have a certain intensity relation. If, for example, one considers a specific line as the Kα-line of an element, it should be possible to detect the corresponding Kβ-line in the correct intensity relation, provided that it is not overlaid by a line of another element. When the lines have been assigned to the elements, the line intensity allows for conclusions to be drawn about the concentration of the elements. In general, matrix effects (see the appendix) make it difficult to determine the concentration directly. This is why, in practical applications, quantitative analysis is performed via a comparison with the calibration functions that are stored in the computer and that have been created with certified reference samples. If there are no matrix effects, the concentration ca of an element a of a sample can be determined based on the relation between the intensity Ia of a line of the element and the intensity Ie of the same line of the pure element, i.e.: ca Ia Ie (1) As a first approximation, the intensity of a fluorescence line can be stated based on its peak value. Evaluation Evaluation of the constantan sample Figure 6 shows the fluorescence spectrum of the constantan sample and the corresponding spectra of the pure elements. Based on the Kα-lines, Figure 7 shows the method for evaluating these curves. The corresponding values are shown in Table 1. The pulse numbers that are stated in the table have been cleared up in terms of the background intensity (approx. 5 pulses). Ni Cu Kα Ni KαCu Kα Kα Kβ Kβ Fig. 6: Fluorescence lines of constantan, copper, and nickel 4 PHYWE Systeme GmbH & Co. KG © All rights reserved P2545001 TEP 5.4.50 -01 Quantitative X-ray fluorescence analysis of alloyed materials Kα Ni KαCu Kα Ni KαCu Fig. 7: Spectrum evaluation method: Kα-lines of constantan, copper, and nickel with a scaled normal distribution Table 1: Constantan sample Sample Pure element A B Linie Cu Ni Kα Kα C E /keV D Ie / Imp. E Ia / Imp. F ca / % 8,0 183 518 57,8 7,4 172 378 42,2 Column F in the table shown above includes the experimental concentration values of the constantan sample. According to these values, the sample consists of 57.8% (≈ 55%) of copper and 42.2 (≈ 45%) of nickel. (Data provided by the manufacturer of the resistance alloy constantan: Cu55Ni45) Evaluation of a brass sample Figure 8 shows the fluorescence spectra of a brass sample (CuZn39Pb3) and those of the corresponding pure elements. Again, only the α-lines are used for determining the concentration levels. The corresponding evaluation, with the background intensity being taken into consideration, can be found in Table 2. Table 2: Brass sample (CuZn39Pb3) Pure Element A Sample B Linie Cu Zn Pb Kα Kα Lα C E /keV D Ie / Imp. E Ia / Imp. F ca / % 8,0 974 546 56,0 8,6 1037 420 40,5 10,5 235 6 2,6 www.phywe.com P2545001 PHYWE Systeme GmbH & Co. KG © All rights reserved 5 TEP 5.4.50 -01 Quantitative X-ray fluorescence analysis of alloyed materials Cu Zn Zn/Pb Pb Pb Lα Zn Cu Kα Kβ Lβ Pb Kα Kβ Fig. 8: Fluorescence lines of brass (CuZn39Pb3), copper, zinc, and lead Appendix Normally, the fluorescence intensity of an element A, with the same concentration in alloys with various alloy constituents, is not identical. This shows that the fluorescence intensity of an element not only depends on the concentration, but also on the combination of elements that form the so-called matrix of the element that is to be analysed. If the energy of the fluorescence radiation of an element A is high enough to stimulate the fluorescence radiation in element B, the radiation coming from B not only depends on the primary intensity, but also on the concentration of element A. Vice versa, element A can absorb the radiation of element B. These additional effects are known as matrix effects. 6 PHYWE Systeme GmbH & Co. KG © All rights reserved P2545001