TEP 5.4.48 -01 Qualitative X-ray fluorescence analysis of solutions Related topics Bremsstrahlung, characteristic X-radiation, energy levels, fluorescent yield, solubility, solubility product, semiconductor energy detectors and multi-channel analysers. Principle Various saturated solutions are subjected to polychromatic X-rays. The energy of the resulting fluorescence radiation is analysed with the aid of a semiconductor detector and a multi-channel analyser. The energy of the corresponding characteristic X-ray fluorescence lines is determined. The elements of the samples are identified by comparing the line energies with the corresponding table values. Equipment 1 1 1 1 1 1 1 1 1 1 1 XR 4.0 expert unit 35kV XR 4.0 Goniometer for X-ray unit, 35 kV XR 4.0 Plug-in module with W X-ray tube Diaphragm tube d = 1 mm Diaphragm tube d = 2 mm Multi-channel analyser X-ray energy detector XR 4.0 XRED cable 50 cm Screened cable, BNC, l = 750 mm Universal crystal holder for the X-ray unit Balance, OHAUS TA 501, 500 g / 0.1 g 09057-99 09057-10 09057-80 09057-01 09057-02 13727-99 09058-30 09058-32 07542-11 09058-02 49243-93 2 Beaker, polypropylene, 100 ml 1 Spoon and spatula, steel, l = 120 mm Glass rod, BORO 3.3, l = 200 mm, d = 3 2 mm 1 Macro-cuvettes, 4 ml, PS, 100 pieces 1 Lead(II) chloride, 250 g 1 Potassium bromide, 50 g 1 Software for the multi-channel analyser 36011-01 46949-00 40485-01 35663-10 31117-25 30258-05 14452-61 PC, Windows® XP or higher This experiment is included in the upgrade package: XRM 4.0 X-ray material analysis. Fig. 1: P2544801 www.phywe.com P2544801 PHYWE Systeme GmbH & Co. KG © All rights reserved 1 TEP 5.4.48 -01 Qualitative X-ray fluorescence analysis of solutions Safety instructions When handling chemicals, you should wear suitable protective gloves, safety goggles, and suitable clothing. Please refer to the appendix for detailed safety instructions. Tasks 1. Calibrate the semiconductor energy detector with the aid of the characteristic radiation of the tungsten X-ray tube. 2. Record the fluorescence spectra of saturated potassium bromide and lead chloride solutions. 3. Determine the energy values of the correspond- Fig. 2: Connections in the experimentation area ing fluorescence lines and comparison with the corresponding table values. Note Please comply with the relevant safety instructions when handling and disposing of chemicals. Fig. 3: Connection at the external panel of the XR 4.0 XSet-up ray expert unit to the MCA - Screw the adapter ring onto the inlet tube of the energy detector. X-ray energy - Connect the signal and supply cables to the cordetector responding ports of the detector with the aid of the right-angle plugs. - Connect the signal and supply cables from the MCA to the appropriate connections in the experiment chamber of the X-ray unit (signal cable: red, supply cable: green (see Fig. 2)). Left position of the - Connect the external ports for the X RED of the goniometer x-ray unit (signal cable red, supply cable green, see Fig. 3) to the multi-channel analyse (MCA). Universal crystal Connect the signal cable via a screened BNCholder with filled cucable to the “Input” port of the MCA and the vette supply cable to the “X-Ray Energy Det.” port of the MCA. - Secure the energy detector in the holder of the swivel arm of the goniometer. Lay the two cables with sufficient length so that the goniometer can be swivelled freely over the entire swivelling range. - Connect the multi-channel analyser and computer with the aid of the USB cable. Fig. 4: Set-up at the goniometer 2 PHYWE Systeme GmbH & Co. KG © All rights reserved P2544801 Qualitative X-ray fluorescence analysis of solutions - TEP 5.4.48 -01 Insert the tube with the 2-mm-aperture. Bring the goniometer block and the detector to their respective end positions on the left. Bring the detector to the 90° position in the 2:1 coupling mode (Fig. 4). Calibration of the multi-channel analyser (if there is no other already existing calibration that can be used) - Bring the goniometer block and the detector to their respective end positions on the right. Fig. 5: calibration of the multi-channel analyser - Insert the tube with the 1 mm-aperture into the exit tube of the X-ray tube. - With the X-ray unit switched on and the door locked, bring the detector to the 0° position. Then, shift the detector by some tenths degree out of the zero position in order to reduce the total rate. - Operating data of the tungsten X-ray tube: Select an anode voltage UA = 25 kV and an anode current IA = 0.02 mA and confirm these values by pressing the “Enter” button. - Switch on the X-radiation - In the MEASURE program, select “Multi channel analyser” under “Gauge”. Then, select “Settings and calibration”. After the “Calibrate” button has been clicked, a spectrum can be measured. The counting rate should be < 300 c/s. Energy calibration settings: - 2-point calibration, - Unit = keV, Gain = 2 – Set the offset so that low-energy noise signals will be suppressed (usually a few per cent are sufficient), See Fig 5. - Measuring time: 5 minutes. Use the timer of the X-ray unit. - Make the two coloured calibration lines congruent with the line centres of the two characteristic X-ray lines. The corresponding energy values (see e.g. P2544701) E(L3M5/L3M4) = 8,41keV and E(L2N4) = 9,69 keV are entered into the corresponding fields, depending on the colour. (Note: Since a separation of the lines L3M5 and L3M4 Lines is not possible, the mean value of both lines is entered as the energy of the line). - Name and save the calibration. Procedure Sample preparation In order to produce high-intensity fluorescence lines, saturated solutions must be prepared. To do so, add approximately 0.5 g of PbCl2 into 50 ml of water, and approximately 32 g of KBr into another 50 ml of water. Then, fill the plastic cuvettes to ¾ with the saturated solutions. Spectrum recording - Insert the tube with the 2-mm-aperture. - Bring the goniometer block and the detector to their respective end positions on the left. - Bring the detector to the 90° position in the 2:1 coupling mode (the sample holder must then be aligned to 45°) - Fasten a filled plastic cuvette to the universal holder with some double-sided adhesive tape so that a flat side of the cuvette is hit by the primary x-ray beam (see Fig 4). - Operating data of the tungsten X-ray tube: Adjust an anode voltage Ua = 35 kV and an anode current so that the counting rate is ≤ 200 c/s. - Measuring time: 10 minutes (use the timer of the X-ray unit). www.phywe.com P2544801 PHYWE Systeme GmbH & Co. KG © All rights reserved 3 TEP 5.4.48 -01 Qualitative X-ray fluorescence analysis of solutions Evaluation of the measurement curves - In order to determine the line energy, switch from the bar display to the curve display. To do so, click “Display options” and then “Interpolation and straight lines”. - Extend the relevant line section with the zoom function . Open the window “Function fitting - Then, select the curve section with normal distribution”. - Find the line centroid with the function “Survey” ”. Then, select “Scaled . Theory and Evaluation When X-rays interact with matter, they lose energy. In the range of energy that is available during this experiment, the photoelectric effect plays the most important role. This means that in the atom on one of the lower shells an electron is ejected due to the absorbed photon energy. The now free space is taken by an electron from one of the higher shells. The energy that is produced during this process can be used for the ejection of another electron from one of the higher shells (Auger effect) or for the generation of a photon (fluorescence radiation). Since the energy of the energy levels that are involved in this process is atom-specific, the type of the emitting atom can be determined based on the energy of the fluorescence radiation. In order to determine the type of atom, the experimental energy values are then compared to the corresponding table values (e.g. “Handbook of Chemistry and Physics”-CRC-Press, Inc. USA). During the assignment of the fluorescence lines, it must be taken into consideration that the relaxations that follow the primary ionisation process can only take place if they fulfil the quantum-mechanical selection rules Δj = 0, ± 1 and Δl = ± 1 (j = total angular momentum, l = orbital angular momentum). Task 2 and 3: Record the fluorescence spectra of saturated potassium bromide and lead chloride solutions and determine the energy values of the corresponding fluorescence lines and comparison with the corresponding table values. Figure 6 shows the fluorescence spectrum of potas1 2 3 4 5 sium bromide. Only the high-intensity characteristic ↓↓ ↓ ↓ ↓ Kα- and Kβ-radiation of bromide (line 3 and 4) can be clearly identified. The energy of the characteristic fluorescence radiation of potassium is close to the sensitivity limit of the energy detectors, which is why it cannot be identified in this experiment. The lines 1 (E = 7.41 keV), 2 (E = 8.04 keV) and 5 (E = 16.35 Fig. 6: Kα- und Kβ-fluorescence lines of bromine keV), which are just visible, are caused by nickel and copper. The scattered primary radiation generates fluorescence radiation on the material components nickel and copper of the detector housing. This fluorescence radiation is also detected by the detector. Figure 7 shows the fluorescence spectrum of the lead chloride solution. The lines 1, 3, and 7 can be reassigned to the elements nickel, copper, and molybdenum. The other lines are part of the characteristic L-radiation of lead. The energy of the fluorescence radiation of chlorine is below the detection sensitivity of the detector. Table 1 shows the evaluation of the spectra of Figure 2 and Figure 7. The Lγ1-line (line 6) clearly becomes asymmetric towards the high-energy side. This is due to the lowintensity Lγ2,3-lines of lead with a slightly higher energy, which cannot be separated clearly in this experiment. 4 PHYWE Systeme GmbH & Co. KG © All rights reserved P2544801 TEP 5.4.48 -01 Qualitative X-ray fluorescence analysis of solutions Table 1: Assignment of the characteristic fluorescence lines of potassium bromide and lead chloride Atomic number Z 35 Element Line Eexp./ keV Elit. / keV Transition Br 82 Pb 3 4 3 4 5 6 11,83 13,20 9,18 10,52 12,58 14,84 11,92 13,29 9,18 10,45 / 10,55 12,61 14,76 KL2,3 - Kα KM3 - Kβ L3M1 - Ll L3M4,5 - Lα1,2 L2M4 - Lβ1 L2N4 - Lγ1 Fig. 7: Zoomed representation with a fitted normal distribution of the Kα- and Kβ-line of bromide. 1 2 3 4 5 6 7 ↓ ↓ ↓ ↓ ↓ ↓ ↓ Fig. 8: L-fluorescence lines of lead www.phywe.com P2544801 PHYWE Systeme GmbH & Co. KG © All rights reserved 5 TEP 5.4.48 -01 Qualitative X-ray fluorescence analysis of solutions Disposal Do not dispose of heavy-metal-containing waste via household waste. Appendix Hazard symbol, signal word Hazard statements Precautionary statements H315: Causes skin irritation H319: Causes serious eye irritation H335: May cause respiratory irritation P261: Avoid breathing dust/fume/gas/mist/vapo urs/spray. P305 + P351 + P338: IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present and easy to do. Continue rinsing. Potassium bromide (KBr) Lead(IV) oxide (PbO2) Danger 6 H302: Harmful if swallowed H332: Harmful if inhaled H360: May damage fertility or the unborn child H373: Causes damage to organs through prolonged or repeated exposure H410: Very toxic to aquatic life with long lasting effects PHYWE Systeme GmbH & Co. KG © All rights reserved P201: Obtain special instructions before use. P273: Avoid release to the environment. P308 + P313: IF exposed or concerned: Get medical advice/attention. P2544801