N191xA Power Meter Range Accuracy Verification test 20dBm (100mW) Page 1 of 7 Table of Contents 1. Overview .............................................................................................................................................................. 3 2. Instrument Accuracy Test......................................................................................................................................... 4 Page 2 of 7 1. Overview This document describes the condition during range accuracy verification using the 11683A Range Calibrator especially at 20dbm (100mW). Page 3 of 7 2. Instrument Accuracy Test Equipment Range Calibrator Power Sensor Cable Agilent 11683A Agilent 11730A Test Setup Figure 1-1 Instrument Accuracy Test Setup Procedure The following procedure should be performed for the Instrument Accuracy Test. 1. Connect the equipment as shown in Figure 1-1. 2. Switch on the power meter. 3. Press Preset, then confirm. 4. Press DISP units W for readings in Watt. 5. Press Disp Resolution to set the resolution to 4. 6. Set the range calibrator as follows: RANGE 3µw Page 4 of 7 POLARITY NORMAL FUNCTION STANDBY LINE ON Note When switching the range calibrator to STANDBY, allow enough time for the range calibrator to settle to its zero value before attempting to zero the power meter. This settling will appear on the power meter display as downward drift. When the drift reaches minimum (typically less than 60 seconds), the range calibrator is settled. 7. On the power meter, press Channel set Measure Averaging to Manual set to 512 and then press enter. 8. On the power meter, press Cal, Zero. Wait approximately 10 seconds for the wait symbol to disappear. Verify that the display reads 0±0.05µW. 9. Set the range calibrator’s FUNCTION switch to CALIBRATE. 10. Set the range calibrator’s RANGE switch to 1mW. 11. Press Cal to calibrate the power meter. 12. Set the range calibrator’s RANGE switch to 100mW. Figure 1-2 Default display when the Range Calibrator is connected to the N1912A Observation: 1. The default display setting for the N1912A at the upper window is Single Numeric. Therefore, you will not be able to see the “over range” message on the screen since there is no scale indication. Page 5 of 7 Figure 1-3 Default display when the Range Calibrator is connected to the N1911A Observation: 1. The default display setting for the N1911A is Single Numeric at the upper window and Analog display at the bottom window. Therefore, you will see an “over range” message at the analog view. 2. The pre-set scale on Analog display is set at min -70dBm (100pW) and max 20dBm (100mW) and the display will show “over range” when tested at 20dBm. This is normal because the reading already exceeds the maximum limit but is still within the range. As long as the measurement is still within the spec limit, “over range” is not an issue and the unit is good. You can change the maximum limit by changing the scale: Press DISP 1 of 2 Anlg Mtr Scaling Enter the desired max limit. Figure 1-4 3. Alternatively, you can either select the Single or Dual numeric display to view the readings without bothering with the max and min scale. Page 6 of 7 Figure 1-5 Dual Numeric Display – without scale Page 7 of 7