How to Use a Pipette Objective: This document will guide you through proper use of automatic and glass pipette use. These techniques will maximize your reproducibility, precision, and accuracy by minimizing avoidable errors. It is strongly encouraged that you inform yourself further by reading on-line literature on general pipette use. (See end of document for suggested further reading.) Relevant sections from Exploring Chemical Analysis. Harris, D.C. 4th ed. Ch. 2-6 pp. 50-51. Automatic Pipettes The following diagrams are from the Fisher Finnpipette® Digital user manual. Basic Features and Principles. The automatic pipettes we will be using operate on the principle of air displacement through an internal adjustable plunger. This type of pipette is intended for use with polypropylene tips. The sample is loaded into the disposable polymer tip rather into the pipette shaft. Never allow solutions to enter the pipette body as they will damage the piston seals and render the device inoperable until serviced. Figure 1. Image of an automatic pipette used in class. Features of note are: a volume display, a plunger that also serves to adjust the volume of the pipetter as well as load and dispense the sample, Be sure to familiarize yourself with these features. and a thumb operated tip ejector. Basic Operation. The accuracy and precision of a pipette does not depend solely on factory specifications and the volume displayed (Image 1). It depends to a considerable degree on the pipette position, tip conditioning, plunger release, and dispensing technique. The first three of these factors will be discussed below. Position: The pipette should always be held in a vertical position when loading the tip with sample. o When aspirating a sample be sure that the tip is always 2-5 mm below the surface of the solution. Tip conditioning: Tips should generally be pre-wetted or rinsed with the sample to be delivered. This is especially critical when delivering multiple aliquots of the same volume or volumes below 10 L. o If a new volume is dialed in on a pipette, a new tip should be used and conditioned. o Be sure the tip is well sealed to the pipette by inspecting the visible “o-ring” that forms where the tip contacts the pipette. Plunger operation: The plunger should always be depressed or released in a smooth manner. Abrupt movements should be avoided. Figure 2. Schematic of a negative displacement pipette o Inconsistencies in plunger depression or release with proper alignment relative to speed can affect reproducibility when performing sample. repetitive pipetting of the same volume. (http://www.ebbep.org/docs/basics/r o Never move the plunger at a rate faster than the flow aininpipettetechniques.ppt) of the sample into or out of the tip. This rate will vary depending on the viscosity of the sample. o Abrupt plunger release can cause liquid to splash up into the pipette, contaminating the pipette and future samples. o Abrupt plunger release can also cause some of the sample to adhere to the tip above the meniscus affecting the accuracy in delivered volume. There are two techniques for accurately and precisely delivering a given volume as well as a proper method for repetitive pipetting. The choice of which technique to employ depends on the properties and volume of the solution being dispensed. Some general considerations: Never dial the pipette beyond the operating range. Change the volume setting smoothly and gently. Violent or rapid rotation of the plunger dial can damage the micrometer and adversely affect pipette performance. Always dial down to the final volume. After loading the tip with sample, remove any liquid droplets adhered to the outside of the tip by lightly dragging the tip along the side of the container. There are three plunger positions described in all techniques: (A) fully released, (B) first stop – be aware that this stop is a soft stop and can be overshot if too much pressure is applied, and (C) second stop – a hard stop. Technique I: Forward Technique. Used with samples having physical properties - density, viscosity, and vapor pressure - similar to water. This is not the ideal method for aqueous surfactant and protein solutions (refer to Technique II). 1) With the pipette tip out of the solution, press the plunger to position B. 2) Immerse the tip in the solution and slowly release the plunger to position A. 3) Expel the desired volume of sample by immersing the tip into the receiving solution (if performing a dilution) or touching the tip to the side of the receiving container and a. pressing the plunger to position B and waiting a second before b. further pressing the plunger to position C. c. If the tip was immersed in a solution remove the tip along the side of the container. 4) AFTER removing the tip from the sample, return the plunger to position A 5) Eject the tip into an appropriate waste container. Technique II: Reverse Technique. Used with viscous, foamy, and volatile samples. This is also the preferred technique when dispensing volumes less than 10 L. 1) With the pipette tip out of the solution, press the plunger to position C. 2) Immerse the tip in the solution and slowly release the plunger to position A. 3) Expel the desired volume of sample by pressing the plunger to position B and waiting a second. a. DO NOT go beyond plunger position B. b. If the tip was immersed in a solution remove the tip along the side of the container. 4) Clear the tip into a waste container. 5) Return the plunger to position A and eject the tip into an appropriate waste container. Repetitive pipetting. This is the preferred technique when delivering multiple equal aliquots of a sample. 1) With the pipette tip out of the solution, press the plunger to position C. 2) Immerse the tip in the solution and slowly release the plunger to position A. 3) Expel the desired volume of sample by pressing the plunger to position B and waiting a second. a. Be sure that the tip is in contact with the wall of the receiving container above the solution. b. DO NOT go beyond plunger position B. c. DO NOT immerse the tip in a solution other than the stock solution as this will cross contaminate your stock. 4) Repeat steps 2 and 3 as necessary. a. Using this method you will always have some sample left in the tip between each round. Do not worry, this volume should never be included in the delivery volume. b. Be sure to keep the plunger depressed at position B until the tip is returned to the stock solution for the next round. Technical Data. Manual Pipettes Pipettes (volumetric and serological) can be calibrated “to contain” (TC) or “to deliver” (TD) a given volume of water at a certain temperature (typically 20°C). o These pipettes operate with an acceptable accuracy ± 5°C from the calibration temperature. Figure 3. Information inscribed on a pipette. Like any other piece of volumetric glassware the volume is read from the bottom of the meniscus. o Be sure to be looking perpendicular and level with the meniscus or you will incorrectly assess its position. o Precision and accuracy You can deliver a given volume by two methods depending on the pipette you are using (TD or TC and volumetric or serological). o Differential pipetting applies only to serological pipettes. In this method you deliver the desired amount as the difference between two volume graduations on the pipette barrel. o If you are delivering the entire pipette volume (volumetric and serological pipettes) be sure to follow the guidelines for a TD or TC pipette regarding blowing out the last bit of solution in the pipette. NOTE: Regardless of whether you use a pump or a bulb always check the seal between the pipette and bulb/pump. o This can be done easily by loading some solution into the pipette and observing if any leaks out. Roller piston pump use 1. Securely attach your pipette to the pump. a. Be sure there is an even seal between the pipette and the pump by gently rotating the pipette a couple of turns. b. The pump piston should be all the way down. 2. Insert the pipette approximately 1 cm below the surface of the solution and maintain the pipette perpendicular to the ground. 3. Load the pipette with solution by slowly rotating the knurled knob in a downward direction. a. Be sure NOT to overfill the pipette and allow Figure 4. Various sizes of pipette pumps liquid to enter the pump. IF this happens you for different size pipettes. will have assume the system is contaminated and rinse the pump. These pumps are not fun to clean as they are difficult to dry once rinsed. b. Fill the pipette slightly above the desired volume. 4. Slowly rotate the knurled knob in an upward direction to set the meniscus to the desired mark. 5. Remove the pipette from the solution and briefly drag it along the side of the container to remove any liquid adhering to the outside of the pipette. a. If necessary, you may use a Kim-Wipe to clean the outside of the pipette (not a sterile technique!). Be sure not to touch the open end of the tip. 6. Transfer the pipette to the receiving container and touch the tip to the side just above the solution (if performing a dilution). 7. Slowly rotate the knurled knob in an upward direction and deliver the desired volume. Bulb use 1. Securely attach your pipette to the bulb. a. Be sure there is an even seal between the pipette and the bulb by gently rotating the pipette a couple of turns. 2. Press A button and squeeze bulb. a. Once bulb is fully decompressed, release A button. b. This creates a negative pressure differential between inside the bulb and ambient pressure thus generating your liquid aspiration capability. 3. Insert the pipette approximately 1 cm below the surface of the solution maintain the pipette perpendicular to the ground. 4. Load the pipette with solution by gently pressing Figure 5. Valved pipette bulb. the S button. a. Be sure NOT to overfill the pipette and allow liquid to enter the bulb. IF this happens you will have to assume the system is contaminated and rinse the bulb. These bulbs are not fun to clean as they are difficult to dry once rinsed. b. Fill the pipette slightly above the desired volume. c. Release the S button to stop sample aspiration. 5. You can slowly release the solution from the pipette by gently pressing the E button. a. Keep a careful eye on the meniscus and release the E button once it has reached the desired volume. 6. Remove the pipette from the solution and briefly drag it along the side of the container to remove any liquid adhering to the outside of the pipette. a. If necessary, you may use a Kim-Wipe to clean the outside of the pipette (not a sterile technique!). Be sure not to touch the open end of the tip. 7. Transfer the pipette to the receiving container and touch the tip to the side just above the solution (if performing a dilution). 8. Press the E button and deliver the desired volume.