Micropipettes Measuring volumes which are what we scientists call “very, very small” MICROPIPETTE RULES •Don’t set the pipettes past their intended range (high or low), even though you can – set the pipette volume only within the range specified for that micropipette P-20 range is 2 – 20 ul P-100 range is 20 – 100 ul P-200 range is 20 – 200 ul P-1000 range is 200 – 1000 ul •When using a micropipette, first apply a tip. •Check to be sure the tip is on securely. •Only open the tip box when getting a tip – DON’T LEAVE IT OPEN! •NEVER hold pipette so that the tip is higher than the handle or liquid leaks into the barrel (BAD!) •Always hold pipettes vertically when in use. •Feel the stops before you begin taking up or dispensing any liquid. st 1 stop: Exact volume to which the pipette is set. Press to this stop before immersing pipette tip in the desired solution. 2nd stop: Use when dispensing solution. Pushes an extra volume of air to dislodge any solution adhering to the tip. How to pipette •Extract and dispense SLOWLY. •When taking up or dispensing liquid, don’t put the tip any deeper into the liquid than necessary. •When dispensing, touch the tip to the side or bottom of the tube, or if there is liquid already in the tube, touch the tip to the liquid. Mid-range: 2 - 20 µL 10 -100 µL 20 - 200 µL Range is printed on the button or on the side Use medium sized yellow tips We will be using P-20 and P-200 pipettes P-20 has a range of 2 - 20 µL Digits represent Tens (10’s) Ones (1’s) Tenths (1/10’s) 10 -100 µL This pipette is set to its maximum: 100 µL Digits represent: 100’s 10’s 1’s High Range: 200-1000 µL Use large blue tips Digits from top to bottom represent: 1000’s 100’s 10’s The 1’s are indicated by the tiny white tick marks in the dial with the 10’s. Marks are in increments of 2 µL This pipette is set to 1000 µL This pipette is set to 990 µL