107 Wolfe Run Road, Burgw, North Carolina 28425 - Phone:(910)259-7366 - email: sbruce@sdproficiency.com Final Test Report (participant's Co. name) (address) (address) (Test Date) (Report Date) Start Date of Test Cycle: End Date of Test Cycle Proficiency Test: Flow of Gas 50sccm to 50slpm Participant ID: (Participant contact name) Participants's Cert #: Test Artifact # NC12012015 (cal date: 09/10/2015) Final Report # (xxxxxx) Test Coordinator: S. Bruce Benson The following tables and graphs represent all participants test results. Participants are referenced only by their participant number assigned at the beginning of the test cycle. Reference values and uncertainties are traceable to the SI. They are reported by a laboratory that is accredited by A2LA or one of A2LA's mutual recognition partners in a certificate bearing the logo of their accrediting body. Test result numbers including the calculated En values are calculated using SD Proficiency Testing's procedure QF005.2 "Participant En Calculations. En numbers outside +/- 1.0 (indicated in red) indicate a difference between your measured value and the reference value larger than predicted by your uncertainty. Please review these results carefully as they may require corrective action on your part. This report shall not be reproduced, except in full without written approval of SD Proficiency Testing, LLC. We always treat participant results as confidential, and welcome your feedback and suggestions for improvement of this test and our operation in general. Page 1 of 4 Date Prepared 12/2/2015 Service and calibration of reference standards used in this testing scheme are performed by a supplier of critical services, who is technically proficient and fully accredited to ISO/IEC-17025:2005 and ANSI/NCSL Z540-1-1994. SD Proficiency Testing maintains full oversight and responsibility of this work 107 Wolfe Run Road, Burgw, North Carolina 28425 - Phone:(910)259-7366 - email: sbruce@sdproficiency.com 50 slpm Participant # Ref N Ref N Ref N Ref N Ref N 53301 En # 42608 En # 31921 En # 21506 En # 10655 En # 1 53484 0.382 42813 0.429 32198 0.58 21619 0.235 11017 0.759 2 53400 0.42 42866 0.58 32180 0.69 21620 0.45 11010 0.8 3 53400 0.42 42866 0.58 32180 0.69 21620 0.45 11010 0.8 4 53400 0.42 42866 0.58 32180 0.69 21620 0.45 11010 0.8 5 53400 0.42 42866 0.58 32180 0.69 21620 0.45 11010 0.8 6 53400 0.42 42866 0.58 32180 0.69 21620 0.45 11010 0.8 7 53411 0.68 42688 0.45 32810 1.25 21650 0.8 11020 0.7 8 53411 0.68 42688 0.45 32810 0.68 21650 0.68 11020 0.68 9 53411 0.68 42688 0.45 32810 0.68 21650 0.68 11020 0.7 10 53411 0.68 42688 0.45 32810 0.68 21650 1.1 11020 0.78 11 53411 0.68 42688 1.25 32810 0.98 21650 0.8 11020 0.68 12 53411 1.25 42688 1.25 32810 1.3 21650 0.68 11020 0.68 53420 1.4 53415 1.2 53410 1 0.8 53405 Participant Reference 0.6 53400 En # 0.4 53395 0.2 53390 0 1 2 3 4 5 6 7 8 9 10 11 12 100% data point Page 2 of 4 Date Prepared 12/2/2015 Service and calibration of reference standards used in this testing scheme are performed by a supplier of critical services, who is technically proficient and fully accredited to ISO/IEC-17025:2005 and ANSI/NCSL Z540-1-1994. SD Proficiency Testing maintains full oversight and responsibility of this work 107 Wolfe Run Road, Burgw, North Carolina 28425 - Phone:(910)259-7366 - email: sbruce@sdproficiency.com 5 slpm Page 3 of 4 Date Prepared 12/2/2015 Service and calibration of reference standards used in this testing scheme are performed by a supplier of critical services, who is technically proficient and fully accredited to ISO/IEC-17025:2005 and ANSI/NCSL Z540-1-1994. SD Proficiency Testing maintains full oversight and responsibility of this work 107 Wolfe Run Road, Burgw, North Carolina 28425 - Phone:(910)259-7366 - email: sbruce@sdproficiency.com 500 sccm Page 4 of 4 Date Prepared 12/2/2015 Service and calibration of reference standards used in this testing scheme are performed by a supplier of critical services, who is technically proficient and fully accredited to ISO/IEC-17025:2005 and ANSI/NCSL Z540-1-1994. SD Proficiency Testing maintains full oversight and responsibility of this work